ID Date Author Typeup Category Subject Text Attachments
  13830   Thu May 10 11:38:19 2018 gautamUpdateGeneralITMY ULLooking at Steve's plot, I was reminded
of the ITMY UL OSEM issue. The numbers don't
make sense to me though - 300um of DC shift
 ITMY_UL.pdf 
  13831   Thu May 10 14:13:22 2018 gautamUpdateGeneralMore refinement of DARM control signal projectionSummary:


It
seems that after a x10 increase in the coil
 darmProj.pdfdarmOLTF.pdfDWcomparison.pdf 
  13833   Fri May 11 13:58:42 2018 ranaUpdateGeneralMore refinement of DARM control signal projectionI think "OLG" trace is not labeled
right; it would be good to see the actual
OLG in addition to whatever that trace actually
  
  13834   Fri May 11 18:17:07 2018 gautamUpdateDetCharAUX laser PLL setup[koji,
gautam]

As discussed at the meeting earlier
  
  13835   Fri May 11 19:02:52 2018 gautamUpdateGeneralMore refinement of DARM control signal projectionI was a bit hasty in posting the earlier
plots. In the earlier plot, the "OLG"
trace was OLG * anti dewhitening as Rana
 darmProj.pdfdarmOLTF.pdfDWcomparison.pdfDARMsuppression.pdfControlLoop.pdf 
  13836   Sat May 12 10:02:03 2018 ranaUpdateGeneralMore refinement of DARM control signal projectionGood question! I've never calculated
what the resonance frequency would be if
had an inductive resistor with our cable
  
  13837   Sun May 13 15:15:18 2018 gautamUpdateGeneralCDS crashI found the c1lsc
machine to be completely unresponsive today.
 vertexFEs_crashed.png 
  13838   Sun May 13 17:31:51 2018 gautamUpdateGeneralCDS crashAs suspected, this was indeed a path problem.
Johannes will elog about it later, but in
short, it is related to some path variables
 CDS_overview_20180513.pngAS_1210293643.jpeg 
  13839   Sun May 13 20:48:38 2018 johannesUpdateGeneralCDS crashI think the root of the problem is that
the /opt/rtapps/
  
  13841   Mon May 14 18:58:32 2018 KevinUpdatePonderSqueezeSqueezing with no SRM


Quote:


 homodyne_heatmap.pdfdisplacement_noise.pdfnoise_budget.pdf 
  13842   Tue May 15 10:42:14 2018 KojiUpdatePSLModulation depth measurement for the 3IFO aLIGO EOM and aftermathThe marconi RF output was turned on and
thus the RF generator condition was restored
to the nominal state on Friday 11th.
  
  13843   Tue May 15 13:34:30 2018 SteveUpdatesafetysurf safety trainingPooja and Keirthana received 40m specific
basic safety training.
  
  13846   Tue May 15 21:56:57 2018 gautamUpdateGeneralStack measurement setup decommissioned[steve,koji,gautam]

Since we think we already know
  
  13847   Tue May 15 22:11:38 2018 gautamUpdateGeneralIFO maintenanceSince there have been various software/hardware
activity going on (stack weighing, AUX laser
PLL, computing timing errors etc etc), I
 OL_RIN_2018_05_15.pdfOLsums.png 
  13849   Wed May 16 21:02:22 2018 KevinUpdatePEMADC common mode rejection with new seismometer connectionsAs described in this
elog, the ADC for the seismometers now
has the signals wired directly to the ADC
 ADC_noise.pdf 
  13850   Wed May 16 21:47:17 2018 KevinUpdatePEMSeismometer Noise SpectraEarlier today Kira and I reconnected the
EX seismometer. I just took some spectra
of all three seismometers, shown in the attachments,
 BS_vel.pdfEX_vel.pdfEY_vel.pdf 
  13851   Thu May 17 09:14:38 2018 SteveUpdateGeneralStack measurement setup decommissionedThe final set-up of stack measurment with
3 load
cells and 4 leveling
 Load_Cell_Measurement_Set_Up.jpgETMX_stack_up_down.png 
  13852   Thu May 17 11:56:37 2018 gautamUpdateGeneralEPICS process died on c1iooThe EPICS process on the c1ioo front
end had died mysteriously. As a result, MC
  
  13859   Thu May 17 15:38:19 2018 KiraUpdatePEMtest setup with seismometerI've moved my setup to the actual seismometer.
I attached the temperature sensor to the
seismometer (attachment 1) with duct tape,
 IMG_20180517_144420.jpgIMG_20180517_145754.jpgIMG_20180517_151956.jpgIMG_20180517_145121.jpg 
  13860   Thu May 17 18:05:01 2018 gautamUpdateSUSSR785 near 1X5I'm working near 1X5 and there is an
SR785 adjacent to the electronics rack with
some cabling running along the floor. I plan
 B55CE985-B703-4282-B716-3144957C7372.jpeg 
  13861   Fri May 18 07:41:01 2018 steveUpdateSUSclipping ITMX oplevThe ITMX oplev
still clipping

  
  13862   Fri May 18 09:13:41 2018 PoojaUpdateSUSColored GigE imageTo obtain a colored version with good contrast
of the grayscale image of scattering of light
by dust particles on the surface of test
 Image__2017-11-14__08-25-13_100k100g1V_colored.pngImage__2017-11-14__08-25-13_100k100g1V.tiff 
  13864   Fri May 18 14:33:34 2018 KiraUpdatePEMtest setup with seismometerHere is the result of my test. I think
I'll leave the can on over the weekend
because there's a long period of time
 seis_temp_can_on.png 
  13866   Fri May 18 19:10:48 2018 keerthanaUpdateGeneralCode for adjusting the oscillator frequency remotlyTarget: Phase locking
can be acheived by giving a scan to the oscilator
frequency. This frequency is now controlled
 frequency_set.jpg 
  13868   Fri May 18 20:03:14 2018 PoojaUpdateCamerasTelescopic lens solution for GigEAim: To find telescopic lens solution to
image test mass onto the sensor of GigE camera.

I wrote a python code to find an
 8x 
  13869   Sun May 20 17:43:01 2018 ranaUpdateElectronicsHow to choose resistorsArticle
from EE Times, describing why metal foil
(NOT metal film) resistors are really better
  
  13870   Sun May 20 23:43:50 2018 gautamUpdateIOOCoil driver noise re-measurementSummary:

In the IMC actuation chain, it
looks like the MC1/MC3
 MC1_coilDriver.pdfMC2_coilDriver.pdfMC3_coilDriver.pdfMC_coilDriverNoises.tgzActuationChainNoiseMeas.pdf 
  13871   Mon May 21 10:15:35 2018 gautamUpdatePEMtest setup with seismometerI guess it's fine for now while we
are still finalizing the setup at EX, but
we should eventually line up the seismometer
  
  13872   Mon May 21 14:17:28 2018 KiraUpdatePEMtest setup with seismometerI have attached the graph for the seismometer
temperature fluctuations over 3 days. As
we can see, there is a noticeable fluctuation
 seis-temp.png 
  13874   Mon May 21 17:36:00 2018 poojaUpdateCamerasGigE camera image of ETMXToday Steve and I tried to to capture the
image of scattering of light by dust particles
on the surface of ETMX using GigE camera.
 Image__2018-05-21__17-34-15_125k100g.tiff 
  13875   Mon May 21 18:02:55 2018 keerthanaUpdateGeneralTesting of the new mini-circuits frequency counterToday, I tested the new mini-circuit frequency
counter by connecting it with the beat signal
output. The frequency counter works fine.
 frequency_counter.jpg 
  13877   Tue May 22 14:49:03 2018 KiraUpdatePEMtest setup with seismometerIt appears that one of the wires was disconnected
overnight or this morning so I wasn't
able to gather data over a full 24 hour period.
 seis-temp-2.png 
  13878   Tue May 22 17:26:25 2018 gautamUpdateIOOMC1 Coil Driver pulled outI have pulled out MC1
coil driver board from its Eurocrate, so
  
  13879   Tue May 22 17:29:27 2018 keerthanaUpdateelogMEDM Diagram for the auxilary laser system control and display.(keerthana, gautam, jon)

In the morning, Jon gave me an
overview of the Auxiliary laser system which
 MEDM_aux.png 
  13880   Tue May 22 23:28:01 2018 gautamUpdateIOOMC1 Coil Driver pulled outThis work is now complete.
MC1 coil driver board has been reinstalled,
  
  13882   Wed May 23 14:50:33 2018 KiraUpdatePEMtest setup with seismometerThis time the test went without issue.
The first attachment is the data for the
past 24 hours and the second attachment is
 seis-temp-3.pngseis-temp-full-2.png 
  13883   Wed May 23 17:58:48 2018 gautamUpdateIOOMC1 Coil Driver pulled out
Marked up schematic + photo post
changes uploaded to DCC
page.
There was a capacitor
 MC1_monitorTF.pngMC1_ULnoise.pdf 
  13885   Thu May 24 10:16:29 2018 gautamUpdateGeneralAll models on c1lsc frontend crashedAll models on the c1lsc
front end were dead. Looking at slow trend
 c1lsc_crashed.png 
  13887   Thu May 24 15:18:12 2018 KiraUpdatePEMPID loop restartedRana said that it wasn't necessary
to gather more data on the temperature fluctuations
so I have reconnected the heater circuit
  
  13888   Thu May 24 16:08:12 2018 KiraUpdatePEMparts listWe will need to order a few things for
our final setup.


1U box to place the heater
 front_panels.zip 
  13893   Fri May 25 14:55:33 2018 Jon RichardsonUpdateCamerasStatus of GigE Camera Software FixesThere is an effort to switch to an all-digital system
for the GigE camera feeds similar to the
one running at LLO, which uses Joe Betzwieser's
  
  13894   Tue May 29 16:22:43 2018 KiraUpdatePEMparts listI've updated the parts list to be an
excel document and included every single
part we will need. This is ony a first draft
 parts_list_-_Sheet1.pdf1U-panel-2.fpdEX-can-panel.fpd 
  13895   Tue May 29 16:33:04 2018 SteveUpdatePEMair cond filters replacedChris replaced some air condition filters
and ordered some replacement filter today.

  
  13896   Wed May 30 10:17:46 2018 gautamUpdateIOOMC1 Coil Driver pulled out[rana,gautam]

Summary:

Last night, Rana fact-checked my
 EM_pickup.pdfcoilDriverNoiseComparison.pdf 
  13897   Wed May 30 12:13:13 2018 ranaUpdateComputersNODUS: rsyncd + framesTo get our rsync back to LDAS back up,
I followed instructions from Dan Kozak:


  
  13899   Wed May 30 23:57:08 2018 gautamUpdatePEMBurning smell in office area / control room[koji, gautam]

We noticed quite a strong burning
smell in the office area and control room
  
  13900   Thu May 31 02:04:55 2018 johannesUpdatePSLAUX laser state of mindThe AUX laser is down to 5.4 mW output
power 
 IMG_20180530_220058190.jpg 
  13901   Thu May 31 10:19:42 2018 gautamUpdateSUSMC3 glitchySeems like as a result of my recent poking
around at 1X6, MC3 is more glitchy than usual
(I've noticed that the IMC lock duty
 MC3_glitchy.png 
  13902   Thu May 31 15:36:59 2018 gautamUpdateGeneralNew camera channelsJon informed me that there are some EPICS
channels that JoeB's camera server code
looks for that don't exist. I thought
  
  13903   Thu May 31 15:48:16 2018 KiraUpdatePEMrunning PID script with seismometerI have attached the result of running the
PID script on the seismometer with the can
on. The daily fluctuations are no more than
 can-on-pid.png 