2186
|
Thu Nov 5 23:09:34 2009 |
Alberto | Update | LSC | Y Arm Cavity Transfer Function |
As for the X Arm, this the transfer function I measured for the Y arm cavity.

This time I'm using a different photodiode than the PDA255 on the Y end table.
The diode I'm using is the PDA520 from where TRY comes from.
I'm going to repeat the measurement with PDA255. |
2185
|
Thu Nov 5 22:30:09 2009 |
Alberto | Update | LSC | X Arm Cavity Transfer Function |
It seems that just repeating the measurement was enough to get a good transfer function of the x arm cavity. Here's what I got.

I'm going to fit the data on matlab, but at first sight, the pole seems to be at about 1.7KHz (that is where the phase is 45deg): as expected.
Probably it was useful to realign the beam on the Transmission PD. (btw, I'm using the PDA255 that was still on the X end table since the AbsL experiemtn that measured the arm length) |
2184
|
Thu Nov 5 19:25:11 2009 |
Alberto | Update | LSC | X Arm Cavity transfer Function |
Quote: |
I would have guessed that you have to calibrate the detectors relative to each other before trying this. Its also going to be tricky if you use 2 different kinds of ADC for this (c.f. today's delay discussion in the group meeting).
I think Osamu used to look at fast transmission signals by making sure the PD at the end had a 50 Ohm output impedance and just drive the 40m long cable and terminate the receiving end with 50 Ohms. Then both PDs go into the SR785.
|
On Rana's suggestion I measured the trasfer function between the two photodiodes PDA255 that I'm using.
I took the one that I had on the end table and put it on the PSL table. I split the MC transmitted beam with a 50% beam splitter and sent the beams on the two diodes. (Rana's idea of picking off the beam and interposing the PDs before the ISS PDs was not doable: ISS PDs would be too close and there would be no room to install the PDA255 before them). See picture with the final setup.

The transfer function also includes the 40m long cable that I was using for the Arm Cavity measurement.
Here's what I got. It looks rather flat. Yesterday the calibration was probably not the problem in that measurement.

I'm now going to install the PD back on the end table and measure the TFs between the excitation and several points of the loop.
(Trivia. At first, the PDs were saturating so Koji attached attenuation filters on to them. Suddenly the measurement got much nicer) |
2183
|
Thu Nov 5 16:41:14 2009 |
josephb | Configuration | Computers | Megatron's personal network |
In investigating why megatron wouldn't talk to the network, I re-discovered the fact that it had been placed on its own private network to avoid conflicts with the 40m's test point manager. So I moved the linksys router (model WRT310N V2) down to 1Y9, plugged megatron into a normal network port, and connected its internet port to the rest of the gigabit network.
Unfortunately, megatron still didn't see the rest of the network, and vice-versa. I brought out my laptop and started looking at the settings. It had been configured with the DMZ zone on for 192.168.1.2, which was Megatron's IP, so communications should flow through the router. Turns out it needs the dhcp server on the gateway router (131.215.113.2) to be on for everyone to talk to each other. However, this may not be the best practice. It'd probably be better to set the router IP to be fixed, and turn off the dhcp server on the gateway. I'll look into doing this tomorrow.
Also during this I found the DNS server running on linux1 had its IP to name and name to IP files in disagreement on what the IP of megatron should be. The IP to name claimed 131.215.113.95 while the name to IP claimed 131.215.113.178. I set it so both said 131.215.113.178. (These are in /var/named/chroot/var/ directory on linux1, the files are 113.215.131.in-addr.arpa.zone and martian.zone - I modified the 113.215.131.in-addr.arpa.zone file). This is the dhcp served IP address from the gateway, and in principle could change or be given to another machine while the dhcp server is on. |
2182
|
Thu Nov 5 16:30:56 2009 |
pete | Update | Computers | moving megatron |
Joe and I moved megatron and its associated IO chassis from 1Y3 to 1Y9, in preparations for RCG tests at ETMY. |
2181
|
Thu Nov 5 16:24:59 2009 |
Koji | Update | CDS | ETMY CDS test stuff |
Joe, Peter, Jay, Koji, Rana
We put the new CDS stuff at Y end 1Y9 rack.
Items
- megatron
- wireless router
- IO chasis (black)
- Extention cable (between megatron & IO chasis)
- 1 ADC card
- 1 DAC card
- 1 BIO card
- The adapter box for ADC
- The adapter box for DAC
- The adapter box for BIO
- 2x IDC-DB37 cable for the ADC box - AA chasis
- 1x IDC cable for the DAC box - Pentek
- 1x DB cable for the BIO box
- 1x +/-15V cable for the BIO box
|
2180
|
Thu Nov 5 16:24:40 2009 |
Jenne | Update | General | Drill Press is down for the count |
The on/off switch for the drill press is broken. Replacement parts should be here tomorrow. |
2179
|
Thu Nov 5 12:34:26 2009 |
kiwamu | Update | Computers | elog rebooted |
I found elog got crashed. I rebooted the elog daemon just 10minutes before. |
2178
|
Thu Nov 5 05:07:22 2009 |
rana | Update | LSC | X Arm Cavity transfer Function |
I would have guessed that you have to calibrate the detectors relative to each other before trying this. Its also going to be tricky if you use 2 different kinds of ADC for this (c.f. today's delay discussion in the group meeting).
I think Osamu used to look at fast transmission signals by making sure the PD at the end had a 50 Ohm output impedance and just drive the 40m long cable and terminate the receiving end with 50 Ohms. Then both PDs go into the SR785.
|
2177
|
Wed Nov 4 23:17:51 2009 |
Alberto | Update | LSC | X Arm Cavity transfer Function |
I measured the transfer function between MC_TRANS and TRX and I'm attaching the result.

That looks quite strange. Something's wrong. I'll repeat it tomorrow.
for the night I'm putting everything back. I'm also reconnecting the OMC_ISS_EXC and opening again the test switch on the ISS screen.
The RFAM monitor remains disable |
2176
|
Wed Nov 4 21:46:18 2009 |
Alberto | Update | LSC | Arm Cavity Finesse Measurement |
Quote: |
Quote: |
I'm going to work on the X arm to measure the arm cavity finesse.
The idea is to measure the cavity transfer function to estimate the frequency of its cavity pole. That should be a more accurate measurement than that based on the cavity decay time.
I'm starting now and I'm going to inject a swept sine excitation on the OMC_ISS_EXC input cable laying on the floor nearby the AP table (see pic).

In orderf to do that I disconnected the cable from the OMC breakout box laying on the floor. I'm going to plug the cable back in as soon as I'm done.
|
Since I need to measure the transfer function between TRX and MC_TRANS_DC I picked off the beam going to RFAM PD to send it to a PDA255 photodiode (cannibalized from the AbsL's PLL) which I installed on the PSL table.
I centerd the beam on the PD and I was able to see the injected signal.
I think I'm ready to measure the transfer function.
Except for the RFAM PD everything is as before.
I'm gonna go grab dinner and I should be back to keep working on that in about one hour.
|
Back from dinner. Taking measurements. |
2175
|
Wed Nov 4 18:35:19 2009 |
Alberto | Update | LSC | Arm Cavity Finesse Measurement |
Quote: |
I'm going to work on the X arm to measure the arm cavity finesse.
The idea is to measure the cavity transfer function to estimate the frequency of its cavity pole. That should be a more accurate measurement than that based on the cavity decay time.
I'm starting now and I'm going to inject a swept sine excitation on the OMC_ISS_EXC input cable laying on the floor nearby the AP table (see pic).

In orderf to do that I disconnected the cable from the OMC breakout box laying on the floor. I'm going to plug the cable back in as soon as I'm done.
|
Since I need to measure the transfer function between TRX and MC_TRANS_DC I picked off the beam going to RFAM PD to send it to a PDA255 photodiode (cannibalized from the AbsL's PLL) which I installed on the PSL table.
I centerd the beam on the PD and I was able to see the injected signal.
I think I'm ready to measure the transfer function.
Except for the RFAM PD everything is as before.
I'm gonna go grab dinner and I should be back to keep working on that in about one hour. |
2174
|
Wed Nov 4 16:49:32 2009 |
Alberto | Update | LSC | Arm Cavity Finesse Measurement |
I'm going to work on the X arm to measure the arm cavity finesse.
The idea is to measure the cavity transfer function to estimate the frequency of its cavity pole. That should be a more accurate measurement than that based on the cavity decay time.
I'm starting now and I'm going to inject a swept sine excitation on the OMC_ISS_EXC input cable laying on the floor nearby the AP table (see pic).

In orderf to do that I disconnected the cable from the OMC breakout box laying on the floor. I'm going to plug the cable back in as soon as I'm done. |
2173
|
Tue Nov 3 12:47:01 2009 |
Koji | Configuration | CDS | 1Y9 Rack configuration update |
For the CDS upgrade preparation I put and moved those stuff at the rack 1Y9:
Placed 1Y9-12 ADC to DB44/37 Adapter LIGO D080397
Placed 1Y9-14 DAC to IDC Adapter LIGO D080303
Moved the ethernet switch from 1Y9-16 to 1Y9-24
Wiki has also been updated. |
2172
|
Tue Nov 3 03:45:04 2009 |
rob | Update | IOO | frequency noise problem |
Quote: |
I used the XARM as a reference to measure the frequency noise after the MC. It's huge around 4kHz--hundreds of times larger than the frequency noise the MC servo is actually squashing. This presents a real problem for our noise performance.
An elog search reveals that this noise has been present (although not calibrated till now) for years. We're not sure what's causing it, but suspicion falls on the piezojena input PZTs.
I didn't bother too much about it before because we previously had enough common mode servo oomph to squash it below other DARM noises, and I didn't worry too much about stuff at 4kHz.. Now that we have a weaker FSS and thus much weaker CM servo, we can't squash it, and the most interesting feature of our IFO is at 4kHz.
I'll measure the actual voltage noise going to the PZTs. I remember doing this before and concluding it was ok, but can't find an elog entry. So this time maybe I'll do it right.
|
This level of frequency noise has not changed, but we now have increased common mode servo gain and so it's not as huge of a deal, although we should still probably do something about it.
Attached is a plot of the piezojena noise measurement, estimated into Hz, along with another measurement of frequency noise as described above.
To get the piezojena voltage noise into Hz, I estimated the PZTs within have a flat 2 micron/V response (based on a rough knowledge of their geometry and assuming a 10 milliradian / 150V steering range). This is the voltage noise with the PZTs operating in closed loop mode, which is how we normally run them. This plot also ignores the transfer function of the Pomona box, as we are mainly looking at noise in the kHz band. I think this plot shows that these PZTs are a good candidate for creating this frequency noise, especially near their mechanical resonances (the manual says the unloaded resonances are in the 3-4kHz range).
I've been operating one DOF of the piezojenas in open loop mode for a couple of weeks now, and the feared drift has not been a problem at all. If we plan to keep using these after the upgrade, we should definitely put some big resistors in series at the outputs and operate them in open loop mode.
Also attached is a plot of RF DARM noise, with a frequency noise spectrum. That spectrum is a REFL 2I spectrum put into DARM units using a measured TF (driving MC_L and measuring REFL 2I and DARM_ERR), and then put into meters using the same DARM calibration as used for the DARM curve. |
Attachment 1: noise.png
|
|
Attachment 2: spectra.pdf
|
|
2171
|
Mon Nov 2 21:09:15 2009 |
Sanjit | Update | Adaptive Filtering | More work on saving coeffs on the OAF screen |
I made some changes in the code (all commented in the installed and SVN version) to print the filter coefficients. I got crazy output. Sometimes memory bugs lead to such crazy behavior. So far I could not find any bugs, but will have to spend more time on it.
|
2170
|
Mon Nov 2 15:27:08 2009 |
Jenne | Update | PEM | Gur2 cables have been moved |
The cables labeled "Gur2" which were connected to channels 2,3,4 of the PEM-ADCU have been moved to the PCIX ADC which is connected directly to the ASS machine. This means that until I (a) put the cables back or (b) figure out how to route channels from the ASS ADC to the RFM, we won't be able to use these channels for environmental monitoring, nor will they be saved.
The Gur2_X, Gur2_Y, Gur2_Z channels are now connected to the 2nd, 3rd and 4th ADC channels respectively, on the ASS ADC (the first channel / TP1 is ADC0_0, which is the 1pps signal.). The sketchy thing about the setup is the connection between the cables and the new ADC board. The PCIX card is connected to the ASS via a white ribbon cable, and the board is just sitting on top of the computer box. The 1pps (which has been hooked up for a long time) goes into the board via clip-doodles. The regular channels have a SCSI cable connector, so I used a SCSI cable to connect up the ADC tester board, and connected my seismometer inputs to this tester board via more clip doodles. Clearly this is a sketchy solution, and not okay for more than a day or so. But we'll see how it goes.
I'm going to, on the SimuLink Diagram, change the input source of these channels, from the RFMN to the ADC. Then we'll see if that fixes our timing problem, and magically makes everything relating to the OAF work, and subtract huge amounts of noise. |
2169
|
Mon Nov 2 13:34:36 2009 |
kiwamu | Configuration | PSL | removed multiply resonant EOM |
I removed the multiply resonant EOM that has been set by a SURF student from PSL table.
I will use it for checking the resonant circuit. |
2168
|
Mon Nov 2 13:00:55 2009 |
Koji | Update | IOO | PMC aligned, MC WFS aligned |
The beam to PMC aligned. The beam to MC WFS cameras aligned.
PMC Trans increased from 2.73 to 2.75 (+1%).
MC Trans increased from 7.80 to 7.87 (+1%). |
2167
|
Mon Nov 2 10:56:09 2009 |
kiwamu | Update | LSC | cron job works succesfully & no timing jitter |
As I wrote on Oct.27th, the cron job works every Sunday.
I found it worked well on the last Sunday (Nov.1st).
And I can not find any timing jitter in the data, its delay still stay 3*Ts. |
2166
|
Sun Nov 1 17:58:44 2009 |
Jenne | Update | General | Update on Video Switch |
The current update on the Chameleon video switch is: no progress.
I connected the old laptop that Rob/Steve acquired via RS-232 serial to the back of the video switch. I'm using P2, the same serial port that the C1AUX computer was connected to just in case there's something good about P2 vs. P1.
I used HyperTerminal to (try to) talk to the switch. Settings were: COM1, bits per second = 9600, data bits = 8, parity = none, stop bits = 1, flow control = none. I can successfully send/get back responses to the basic commands, I (inquiry as to the type of equipment), and H (help - spits out the list of acceptable commands). But when I try to do an actual command to do some video switching, everything hangs. The front panel's rolling display (which just echos the company name) stops, then starts up again after ~20sec. The hyperterminal display doesn't change. I get neither the "DONE" answerback, which would indicate that the command executed successfully, nor do I get the "ERROR" answerback, which would indicate that something is wrong. It just hangs. If I disconnect, and restart the connection, and instead of trying a real command, but instead just send 'blahblahblah', then it will answerback 'ERROR' the first time, and then if I try to send another garbage message, everything hangs again. So, I can sort of talk to the video switch, but I can't make it do anything yet.
I'm leaving the laptop connected instead of C1AUX, since the video EPICS screen doesn't work anyway for now. If you want to start up the connection, either input the settings quoted above, or open "40m Video", which should have these connection settings saved in HyperTerminal. |
2165
|
Fri Oct 30 10:52:56 2009 |
Jenne | Update | PSL | HEPAs |
Zach found the HEPA switch on the PSL table OFF. He turned them on. |
2164
|
Fri Oct 30 09:24:45 2009 |
steve | HowTo | MOPA | how to squeeze more out of little |
Quote: |
Here is the plots for the powers. MC TRANS is still rising.
What I noticed was that C1:PSL-FSS_PCDRIVE nolonger hit the yellow alert.
The mean reduced from 0.4 to 0.3. This is good, at least for now.
|
Koji did a nice job increasing light power with some joggling. |
Attachment 1: 44to34.jpg
|
|
2163
|
Fri Oct 30 04:41:37 2009 |
rob | Update | Locking | working again |
I never actually figured out exactly what was wrong in entry 2162, but I managed to circumvent by changing the time sequence of events in the up script, moving the big gain increases in the common mode servo to the end of the script. So the IFO can be locked again. |
2162
|
Thu Oct 29 21:51:07 2009 |
rob | Update | Locking | bad |
Quote: |
Quote: |
Lock acquisition has gone bad tonight.
The initial stage works fine, up through handing off control of CARM to MCL. However, when increasing the AO path (analog gain), there are large DC shifts in the C1:IOO-MC_F signal. Eventually this causes the pockels cell in the FSS loop to saturate, and lock is lost.
|
This problem has disappeared. I don't know what it was.
The first plot shows one of the symptoms. The second plot is a similar section taken from a more normal acquisition sequence the night before.
All is not perfect, however, as now the handoff to RF CARM is not working.
|
The problem has returned. I still don't know what it is, but it's making me angry. |
Attachment 1: itsback.png
|
|
2161
|
Thu Oct 29 20:21:14 2009 |
Koji | Update | PSL | NPRO LTMP lowered 9.5deg |
Here is the plots for the powers. MC TRANS is still rising.
What I noticed was that C1:PSL-FSS_PCDRIVE nolonger hit the yellow alert.
The mean reduced from 0.4 to 0.3. This is good, at least for now. |
Attachment 1: PSL_MC.png
|
|
2160
|
Thu Oct 29 18:25:33 2009 |
Sanjit | Update | Adaptive Filtering | More work on saving coeffs on the OAF screen |
Quote: |
[Sanjit,Jenne]
Sanjit has been working today on trying to get the OAF coefficients to save properly. Alex got us most of the way, but right now it's looking like the filter that is being saved is totally constant (all the values are the same). We're poking around trying to figure out why this is.
Also, we're starting again (as we should have been for the last week or so since Alex came in to help us) to check in the TOP_XFCODE whenever we make changes to it, and when we recompile the front end code.
|
We are manually restarting assepics, but the terminal logs us out after sometime and ass may crash. I set autologout=0 in the terminal for the time being. Once the testing process is over, assepics will start automatically when the computer is turned on, so we wont have to worry about this.
(if ass crashes tonight, it is not unexpected!)
|
2159
|
Thu Oct 29 18:04:02 2009 |
Jenne | Update | Adaptive Filtering | More work on saving coeffs on the OAF screen |
[Sanjit,Jenne]
Sanjit has been working today on trying to get the OAF coefficients to save properly. Alex got us most of the way, but right now it's looking like the filter that is being saved is totally constant (all the values are the same). We're poking around trying to figure out why this is.
Also, we're starting again (as we should have been for the last week or so since Alex came in to help us) to check in the TOP_XFCODE whenever we make changes to it, and when we recompile the front end code. |
2158
|
Thu Oct 29 13:48:32 2009 |
Koji | Update | PSL | NPRO LTMP lowered 9.5deg |
13:00 Found MC TRANS less than 7.
13:50 Go into the PSL table.
14:20 Work done. Now I am running SLOWscan script.
15:10 SLOWscan finished. It was not satisfactory. I go into the table again.
15:15 Running SLOWscan again.
16:00 SLOWscan done. Lock PMC. Adjust NPRO current so as to maximize PMC TRANS.
16:10 Lock RC, PMC, MZ, MC. Align PMC / MZ on the table. Align MC WFS beams on the QPDs.
16:30 Work done.
New FSS-SLOWDC nominal is -4.0
Now MC TRANS is 7.9. This is +12% increase. ENJOY!
HEPA is on at 90%. Light is off.
---------
NPRO TEMP trimmer adjustment
o PSL NPRO TEMP trimmer at the back of the laser head was turned 6.5 times in CW.
o It reduced NPRO crystal temp by 9.5deg. (43.5deg -> 34.0deg for FSS_SLOWDC -5.5)
To revert the previous setting, refer to the former measurement
c.f. http://nodus.ligo.caltech.edu:8080/40m/2008
NPRO Thermal scan
o 2 scans are performed.
o I selected the colder side of the second scan. i.e. SLOWDC=-4.0
NPRO Current adjustment
o Tweaked C1:PSL-126MOPA_126CURADJ while looking at PMC TRANS.
o CURADJ was changed from -2.25 to -1.9. This corresponds to change of C1:PSL-126MOPA_CURMON from 2.503A to 2.547A. |
Attachment 1: 091028_PSL.png
|
|
2157
|
Wed Oct 28 17:20:21 2009 |
rana | Summary | COC | ETM HR reflectivity plot |
This is a plot of the R and T of the existing ETM's HR coating. I have only used 1/4 wave layers (in addition to the standard 1/2 wave SiO2 cap on the top) to get the required T.
The spec is a T = 15 ppm +/- 5 ppm. The calculation gives 8 ppm which is close enough. The calculated reflectivity for 532 nm is 3%. If the ITM reflectivity is similar, the signal for the 532 nm locking of the arm would look like a Michelson using the existing optics.

|
2156
|
Wed Oct 28 14:39:10 2009 |
steve | Update | PEM | after the tour of the 40m |
Illuminators and PSL lights turned off.
HEPA filter speed increased from 20 to 100% |
2155
|
Wed Oct 28 09:12:18 2009 |
steve | Update | PSL | PMC power on the rise? |
The PMC power is seems to be on the rise, ( MOPA_AMPMON is dropping ?) but I do not think it is real. We have Santa Anna wind condition, when the relative humidity drops and ......
There is an other funky think. The room temp became rock solid. The PSL HEPAs running at 20% and IFO-room ACs are also in normal operational mode. |
Attachment 1: pmcprising.jpg
|
|
2154
|
Wed Oct 28 05:02:28 2009 |
rob | Update | Locking | back |
LockAcq is back on track, with the full script working well. Measurements in progress. |
2153
|
Tue Oct 27 19:37:03 2009 |
kiwamu | Update | LSC | cron job to diagnose LSC-timing |
I set a cron job on allegra.martian to run the diagnostic script every weekend.
I think this routine can be helpful to know how the trend of timing-shift goes
The cron runs the script on every Sunday 5:01AM and diagnostics will take about 5 min.
! Important:
During the running of the script, OMC and DARM can not be locked.
If you want to lock OMC and DARM in the early morning of weekend, just log in allegra and then comment out the command by using 'crontab -e'
|
2152
|
Tue Oct 27 18:19:14 2009 |
rob | Update | Locking | bad |
Quote: |
Lock acquisition has gone bad tonight.
The initial stage works fine, up through handing off control of CARM to MCL. However, when increasing the AO path (analog gain), there are large DC shifts in the C1:IOO-MC_F signal. Eventually this causes the pockels cell in the FSS loop to saturate, and lock is lost.
|
This problem has disappeared. I don't know what it was.
The first plot shows one of the symptoms. The second plot is a similar section taken from a more normal acquisition sequence the night before.
All is not perfect, however, as now the handoff to RF CARM is not working. |
Attachment 1: MCF_issue.png
|
|
Attachment 2: no_MCF_issue.png
|
|
2151
|
Tue Oct 27 18:01:49 2009 |
rob | Update | PSL | hmmm |
A 30-day trend of the PCDRIVE from the FSS. |
Attachment 1: pcdrive_trend.png
|
|
2150
|
Tue Oct 27 17:58:25 2009 |
Jenne | Configuration | PEM | Unknown PEM channels in the PEM-ADCU? |
Does anyone know what the channels plugged in to the PEM ADCU, channels 5,6,7,8 are? They aren't listed in the C1ADCU_PEM.ini file which tells the channel list/dataviewer/everything about all the rest of the signals which are plugged into that ADCU, so I'm not sure if they are used at all, or if they're holdovers from some previous time. The cables are not labeled in a way that makes clear what they are. Thanks! |
2149
|
Tue Oct 27 15:55:04 2009 |
Koji | Update | General | ISS injection work / HEPA is on |
I was working on the ISS excitation to take TFs.
I used ISS IL excitation, stealing from a small box on the floor for the OMC.
All the configuration was restored except that the HEPA is on. |
2148
|
Tue Oct 27 01:45:02 2009 |
rob | Update | Locking | MZ |
Quote: | Tonight we also encountered a large peak in the frequency noise around 485 Hz. Changing the MZ lock point (the spot in the PZT range) solved this. |
This again tonight.
It hindered the initial acquisition, and made the DD signal handoff fail repeatedly. |
2147
|
Mon Oct 26 23:14:08 2009 |
Koji | Update | PSL | laser power is down |
I adjusted the steerings to the PMC and gained 7%. Now the MC_TRANS 7.0 has been recovered.
Actually I need another 7% to get MC_TRANS 7.5.
But I couldn't find how I can recover 126MOPA-AMPMON to 2.8ish.
Quote: |
The laser power is down 5-6%
|
|
Attachment 1: PSL091026.png
|
|
2146
|
Mon Oct 26 19:12:50 2009 |
kiwamu | Update | LSC | diagnostic script for LSC timing |
The diagnostic script I've written is available in the 'caltech/users/kiwamu/work/20091026_OMC-LSC-diag/src'.
To run the script, you can just execute 'run_OMC_LSC.sh' or just call the m-file ' OMC_LSC_timinig.m' from matlab.
NOTES:
The script destructs the lock of DARM and OMC, be careful if you are working with IFO. |
2145
|
Mon Oct 26 18:49:18 2009 |
kiwamu | Update | LSC | OMC-LSC timing issue |
According to my measurements I conclude that LSC-signal is retarded from OMC-signal with the constant retarded time of 92usec.
It means there are no timing jitter between them. Only a constant time-delay exists.
(Timing jitter)
Let's begin with basics.
If you measure the same signal at OMC-side and LSC-side, they can have some time delay between them. It can be described as followers.

where tau_0 is the time delay. If the tau_0 is not constant, it causes a noise of the timing jitter.
(method)
I have injected the sine-wave with 200.03Hz into the OMC-LSC_DRIVE_EXC. Then by using get_data, I measured the signal at 'OMC-LSC_DRIVE_OUT' and 'LSC-DARM_ERR' where the exciting signal comes out.
In the ideal case the two signals are completely identical.
In order to find the delay, I calculated the difference in these signals based on the method described by Waldman. The method uses the following expression.

Here the tau_s is the artificial delay, which can be adjusted in the off line data.
By shifting tau_s we can easily find the minimal point of the RMS, and at this point we can get tau_0=tau_s.
This is the principle of the method to measure the delay. In my measurement I put T=1sec. and make the calculation every 1sec. in 1 min.
(results)
Attachment is the obtained results. The above shows the minimum RMS sampled every 1sec. and the below shows the delay in terms of number of shifts.
1 shift corresponds to Ts (=1/32kHz). All of the data are matched with 3 times shift, and all of the minimum RMS are completely zero.
Therefore I can conclude that LSC-signal is retarded from OMC-signal with constant retarded times of 3*Ts exactly, and no timing jitter has been found.
|
Attachment 3: OMC_LSC60sec.png
|
|
2144
|
Mon Oct 26 18:15:57 2009 |
rob | Update | IOO | MC OLG |
I measured the mode cleaner open loop gain. It's around 60kHz with 29 degs of phase margin. |
2143
|
Mon Oct 26 17:45:34 2009 |
Jenne | Update | Adaptive Filtering | New changes to the OAF fe code |
[Alex, Jenne, Sanjit]
Alex came to the 40m today, and did several awesome things in OAF-land.
We discovered that there is, in fact, an ADC board connected to the ASS machine. The tricky bit is that it only has a ribbon cable connector, so before we can use this ADC, we need to figure out how to make a breakout board/cable/something to connect the seismometer/accelerometer/microphone BNCs to this little board. This is the same little board that connects the timing slave to the ASS machine. For good or for ill, the timing slave is connected to this board via clip-doodles. Potentially we can connect an ADC tester board to this board, and go from seismometer BNCs to clipdoodles to the tester board, but I'm not in love with the idea of utilizing clipdoodles as a semi-permanent solution until the upgrade. I emailed Ben to see if he has a better idea, or (better yet) some spare hardware now that's the same as we'll use after the upgrade. If we can use this ADC, it may solve our timing problem which is caused by the 110B ADC used by the PEM computer. Alex showed Sanjit and I how to connect the ASS's ADC card to the simulink diagram, when we're ready for that.
We also poked around in the code, and it seems that we can now save and restore OAF coefficients at will. I added buttons to the OAF (ASS) screen, and Alex made it so the OAF coefficients are saved in RFM shared memory whenever you click the "save coeffs" button, and are restored when you click the "restore coeffs" button. The buttons are the same as the 'Reset' button which has been there for a long time, so they seem to maybe have a similar problem in that you have to hold the button for a while in order for the code to realize that the button has been depressed. We couldn't fix this easily, because it looks like our SimuLink cds stuff is a little out of date. Some day (before/when Joe and Peter make new screens for the new 40m), we need to update these things. Alex was concerned that it might take a while to do this, if the update broke some of the blocks that we're currently using. Also, Sanjit and I now need to check that the coefficient-saving is going as planned. When I have DTT open, and the OAF running, I see a certain shape to the signal which is sent to MC1 to correct for the seismic motion. This shape includes at least several peaks at resonant frequencies that exist in our stacks/suspensions. I can then save the coefficients, reset the active filter, and then restore the coefficients. When I do this while watching DTT, it seems as though the general shape of the filter is restored, but none of the detailed features are. The reason for this is still under investigation.
The code-modifications involved a few iterations of 'remaking the ass'. |
2142
|
Mon Oct 26 15:40:01 2009 |
steve | Update | PSL | laser power is down |
The laser power is down 5-6% |
Attachment 1: laserpowerdown.jpg
|
|
2141
|
Mon Oct 26 03:57:06 2009 |
rob | Update | Locking | bad |
Lock acquisition has gone bad tonight.
The initial stage works fine, up through handing off control of CARM to MCL. However, when increasing the AO path (analog gain), there are large DC shifts in the C1:IOO-MC_F signal. Eventually this causes the pockels cell in the FSS loop to saturate, and lock is lost. |
2140
|
Sun Oct 25 14:29:45 2009 |
haixing, kiwamu | Configuration | General | SR785 spectrum analyzer |
In this morning, we have disconnected SR785 which was in front of 1X2 rack, to measure a Hall sensor noise.
After a while, we put back SR785 and re-connected as it has been.
But the display setup might have been changed a little bit.
|
2139
|
Sat Oct 24 04:57:33 2009 |
rana | Update | lore | marconi phase |
Quote: |
So, it appears that one doesn't even have to change the Marconi set frequency to alter the phase of the output signal. It appears that other front panel actions (turning external modulations on/off, changing the modulation type) can do it as well. At least that's what I conclude from earlier this morning, when after setting up the f2 Marconi (166MHz) for external AM, the double-demod handoff in the DRMI no longer worked. Luckily this isn't a real problem now that we have the setDDphases and senseDRM scripts.
|
The real problem is that we are using frequency synthesizers to make the beat signals (133 and 199) instead of mixers. Luckily, the future 40m will not use beat signals (?) or synthesizers. |
2138
|
Fri Oct 23 15:02:00 2009 |
rob | Update | lore | marconi phase |
So, it appears that one doesn't even have to change the Marconi set frequency to alter the phase of the output signal. It appears that other front panel actions (turning external modulations on/off, changing the modulation type) can do it as well. At least that's what I conclude from earlier this morning, when after setting up the f2 Marconi (166MHz) for external AM, the double-demod handoff in the DRMI no longer worked. Luckily this isn't a real problem now that we have the setDDphases and senseDRM scripts. |
2137
|
Fri Oct 23 09:13:45 2009 |
steve | Summary | VAC | RGA scan |
Pump down #66 is 435 days old. RGA scan is normal. New maglev is fine. New UPS is in place but not hooked up to communicate.
V1 has bare minimum interlock. Pirani vacuum gauges PTP1 and PRP do not communicate with readout system.
There is no emergency dial out in case of vacuum loss. Our existing vacuum dedicated desk top computer is dead.
New cold cathodes, Pirani gauges and gauge controller should be added.
In general: vacuum system needs an upgrade !
|
Attachment 1: pd66md435.jpg
|
|
Attachment 2: pd66d435ptt.jpg
|
|