Sun Jun 26 18:35:16 2011 |
rana | HowTo | SUS | free swing all optics |
I used scripts/SUS/freeswing-all.csh to give the optics a kick and then turn off their watchdogs and collect the free swinging data. Final script end time = 993173551. Start taking data ~ 993173751
I had to fix up the script a little: it had amateur stuff in there, such as undefined variables.
It still doesn't work that well. On the new Ubuntu workstations, pianosa, it fails by just not setting some of the EPICS variables using the EZCA stuff.
On Allegra, it failed on ~1 out of 10 commands by returning "epicsThreadOnce0sd epicsMutexLock failed" ???
On Pianosa, it sometimes says, instead, "epicsThreadOnceOsd: pthread_mutex_lock returned Invalid argument. ". Ah...now I understand?
So finally, I had to run the script on op340m to get it to actually run all of its commands. That's right; I used a 15 year old Solaris 9 Blade 150 because none of our fancy new Linux machines could do the job reliably.
Fixing our EZCA situation is a pretty high priority; if the locking scripts fail to run ~1 command every hour its going to completely derail the lock acquisition attempts.
If you want to use the IFO tonight, just run the script again on op340m again when you're done. |
Attachment 1: ringdown.png
|
|
4892
|
Tue Jun 28 01:18:53 2011 |
rana | HowTo | SUS | free swing all optics |
Chris Wipf tells me that the EPICS Mutex Jumbo Mumbo can be overcome by upgrading our EPICS. We should get one of Jamie's assistants to get this going on one of the Ubuntu workstations. |
5204
|
Fri Aug 12 04:11:13 2011 |
kiwamu | Update | SUS | free swinging |
Excited all optics - -
Fri Aug 12 03:34:12 PDT 2011
997180467
|
5220
|
Sat Aug 13 02:11:33 2011 |
kiwamu | Update | SUS | free swinging again |
I am leaving all of the suspensions free swinging. They will automatically recover after 5 hours from now.
--
Excited all optics
Sat Aug 13 02:08:07 PDT 2011
997261703
--
FYI : I ran a combination of two scripts: ./freeswing && ./opticshutdown |
5234
|
Sun Aug 14 22:48:37 2011 |
kiwamu | Update | SUS | free swinging again |
Excited all optics
Sun Aug 14 20:22:33 PDT 2011
997413768
|
5246
|
Tue Aug 16 04:50:17 2011 |
kiwamu | Update | SUS | free swinging again |
Since Suresh and I changed the DC biases on most of the suspension, the free swingning spectra will be different from the past.
- -
EXcited ETMX ETMY ITMX ITMY PRM SRM BS
Tue Aug 16 04:48:02 PDT 2011
997530498
|
2405
|
Sun Dec 13 17:43:10 2009 |
kiwamu | Update | SUS | free swinging spectra (vacuum) |
The free swinging spectra of ITMs, ETMs, BS, PRM and SRM were measured under the vacuum-condition. The attachment are measured spectra.
It looks there are nothing wrong because no significant difference appear from the past data and the current data (under atmosperic pressure).
So everything is going well. |
Attachment 1: summary_FreeSwinging_vacuum.pdf
|
|
2368
|
Tue Dec 8 23:13:32 2009 |
kiwamu | Update | SUS | free swinging spectra of ETMY and ITMX |
The free swinging spectra of ETMY and ITMX were taken after today's wiping, in order to check the test masses.
These data were taken under the atmospheric pressure condition, as well as the spectra of ETMX taken yesterday.
Compared with the past (see Yoichi's good summary in Aug.7 2008), there are no significant difference.
There are nothing wrong with the ETMY and ITMX successfully.
--
By the way I found a trend, which can be seen in all of the data taken today and yesterday.
The resonances of pitch and yaw around 0.5Hz look like being damped, because their height from the floor become lower than the past.
I don't know what goes on, but it is interesting because you can see the trend in all of the data.
|
Attachment 1: SUS-ETMY.png
|
|
Attachment 2: SUS-ITMX.png
|
|
2369
|
Wed Dec 9 00:23:28 2009 |
Koji | Update | SUS | free swinging spectra of ETMY and ITMX |
Where is the plot for the trend?
It can be either something very important or just a daydream of you.
We can't say anything before we see the data.
We like to see it if you think this is interesting.
... Just a naive guess: Is it just because the seismic level got quiet in the night?
P.S.
You looks consistently confused some words like damping, Q, and peak height.
- Q is defined by the transfer function of the system (= pendulum).
- Damping (either active or passive) makes the Q lower.
- The peak height of the resonance in the spectrum dy is determined by the disturbance dx and the transfer function H (=y/x).
dy = H dx
As the damping makes the Q lower, the peak height also gets lowered by the damping.
But if the disturbance gets smaller, the peak height can become small even without any change of the damping and the Q.
Quote: |
By the way I found a trend, which can be seen in all of the data taken today and yesterday.
The resonances of pitch and yaw around 0.5Hz look like being damped, because their height from the floor become lower than the past.
I don't know what goes on, but it is interesting because you can see the trend in all of the data.
|
|
5282
|
Tue Aug 23 01:09:44 2011 |
kiwamu | Update | SUS | free swinging test |
excited all the optics ---
Tue Aug 23 01:08:00 PDT 2011
998122096 |
5287
|
Tue Aug 23 11:57:22 2011 |
kiwamu | Update | SUS | free swinging test during lunch time |
excited all the optics. (with ITMY WTF OFF)
Tue Aug 23 11:52:52 PDT 2011
998160788 |
5290
|
Tue Aug 23 17:21:45 2011 |
kiwamu | Update | SUS | free swinging test for ETMY |
Excited ETMY
Tue Aug 23 17:20:45 PDT 2011
998180460
|
5375
|
Sat Sep 10 02:28:45 2011 |
kiwamu | Update | SUS | free swinging test in vacuum condition |
All the optcs were excited
Sat Sep 10 02:14:11 PDT 2011
999681266
|
5421
|
Thu Sep 15 18:12:21 2011 |
Jenne | Update | SUS | free swinging test in vacuum condition |
Quote: |
All the optcs were excited
Sat Sep 10 02:14:11 PDT 2011
999681266
|
Optic |
The Plot |
Input Matrix |
BADness |
ITMX |
 |
pit yaw pos side butt
UL 0.601 0.680 1.260 -1.009 0.223
UR 0.769 -1.254 -0.175 -0.179 0.581
LR -1.231 0.065 0.566 -0.480 0.252
LL -1.399 2.000 2.000 -1.310 -2.944
SD -0.580 0.868 2.451 1.000 -1.597
|
7.95029 |
ITMY |
 |
pit yaw pos side butt
UL 1.067 0.485 1.145 -0.195 0.929
UR 0.548 -1.515 0.949 -0.142 -1.059
LR -1.452 -0.478 0.855 -0.101 1.051
LL -0.933 1.522 1.051 -0.153 -0.962
SD -0.530 0.903 2.115 1.000 0.142 |
3.93939 |
ETMX |
 |
pit yaw pos side butt
UL 0.842 1.547 1.588 -0.018 1.026
UR 0.126 -0.453 1.843 0.499 -1.173
LR -1.874 -0.428 0.412 0.511 0.934
LL -1.158 1.572 0.157 -0.006 -0.867
SD 1.834 3.513 -0.763 1.000 -0.133 |
5.39825 |
ETMY |
 |
pit yaw pos side butt
UL -0.344 1.280 1.425 -0.024 0.903
UR 1.038 -0.720 1.484 -0.056 -1.161
LR -0.618 -1.445 0.575 -0.040 0.753
LL -2.000 0.555 0.516 -0.007 -1.184
SD -0.047 -0.038 0.986 1.000 0.083 |
4.15747 |
BS |
 |
pit yaw pos side butt
UL 1.549 0.655 0.393 0.263 0.997
UR 0.192 -1.345 1.701 -0.063 -0.949
LR -1.808 -0.206 1.607 -0.085 0.952
LL -0.451 1.794 0.299 0.241 -1.101
SD 0.724 0.293 -3.454 1.000 0.037 |
5.66432 |
PRM |
 |
pit yaw pos side butt
UL 0.697 1.427 1.782 -0.337 0.934
UR 1.294 -0.573 0.660 -0.068 -0.943
LR -0.706 -1.027 0.218 0.016 0.867
LL -1.303 0.973 1.340 -0.254 -1.257
SD 0.369 -0.448 -0.496 1.000 0.456 |
5.1026 |
SRM |
|
Can't invert....need to fix the peak-finding. |
|
MC1 |
 |
pit yaw pos side butt
UL 0.872 0.986 0.160 0.054 0.000
UR 0.176 -0.752 0.917 0.018 0.000
LR -1.824 -2.000 1.840 0.002 3.999
LL -1.128 -0.262 1.083 0.038 -0.000
SD 0.041 0.036 -0.193 1.000 -0.001
|
5.31462 |
MC2 |
 |
pit yaw pos side butt
UL 1.042 0.767 0.980 0.131 0.928
UR 0.577 -1.233 1.076 -0.134 -0.905
LR -1.423 -0.640 1.020 -0.146 1.050
LL -0.958 1.360 0.924 0.120 -1.117
SD -0.073 -0.164 -0.702 1.000 -0.056 |
4.07827 |
MC3 |
 |
pit yaw pos side butt
UL 1.595 0.363 1.152 0.166 1.107
UR 0.025 -1.629 1.135 0.197 -0.994
LR -1.975 0.008 0.848 0.105 0.904
LL -0.405 2.000 0.865 0.074 -0.995
SD -0.433 0.400 -1.624 1.000 0.022 |
3.64881 |
|
5480
|
Tue Sep 20 15:23:16 2011 |
Jenne | Update | SUS | free swinging test in vacuum condition |
This is using data for the SRM from: 20 Sept 2011 03:20:00 PDT = 1000549215
You can see that there are still some funny peaks between Pit and Yaw, but I finnessed the peak-finding, and I was able to fit all of the correct peaks, and invert the matrix:
SRM now has its new matrix, and is damping happily.
Optic |
The Plot |
Matrix |
Badness |
SRM |
 |
pit yaw pos side butt
UL 0.877 0.983 1.105 -0.288 1.092
UR 1.010 -1.017 1.123 -0.145 -1.055
LR -0.990 -1.002 0.895 -0.091 0.848
LL -1.123 0.998 0.877 -0.234 -1.006
SD 0.089 0.064 3.752 1.000 -0.009 |
4.4076 |
|
5481
|
Tue Sep 20 15:39:57 2011 |
Koji | Update | SUS | free swinging test in vacuum condition |
Can't we use Yuta's auto-Q adjust script?
http://nodus.ligo.caltech.edu:8080/40m/3723
Edit by KI :
Of course we can use it but first we have to fix some pynds sentences since his script was written for the OLD pynds. |
5346
|
Tue Sep 6 17:56:12 2011 |
Jenne | Update | SUS | free swinging test on ITMX |
Quote: |
Tue Sep 6 17:48:02 PDT 2011
999391697
|
Kiwamu excited ITMY (which Suresh had already started). I just kicked ITMX:
Tue Sep 6 17:55:21 PDT 2011
999392136 |
5345
|
Tue Sep 6 17:48:57 2011 |
kiwamu | Update | SUS | free swinging test on ITMY |
Tue Sep 6 17:48:02 PDT 2011
999391697
|
5594
|
Sun Oct 2 02:21:27 2011 |
kiwamu | Update | SUS | free swinging test once more |
The following optics were kicked:
MC1 MC2 MC3 ETMX ETMY ITMX ITMY PRM SRM BS
Sun Oct 2 02:13:40 PDT 2011
1001582036
They will automatically get back after 5 hours. |
5240
|
Mon Aug 15 17:23:55 2011 |
jamie | Update | SUS | freeswing script updated |
I have updated the freeswing scripts, combining all of them into a single script that takes arguments to specify the optic to kick:
pianosa:SUS 0> ./freeswing
usage: freeswing SET
usage: freeswing OPTIC [OPTIC ...]
Kick and free-swing suspended optics.
Specify optics (i.e. 'MC1', 'ITMY') or a set:
'all' = (MC1 MC2 MC3 ETMX ETMY ITMX ITMY PRM SRM BS)
'ifo' = (ETMX ETMY ITMX ITMY PRM SRM BS)
'mc' = (MC1 MC2 MC3)
pianosa:SUS 0>
I have removed all of the old scripts, and committed the new one to the SVN. |
6311
|
Fri Feb 24 04:12:44 2012 |
kiwamu | Update | SUS | freeswing test |
The following optics were kicked:
MC1 MC2 MC3 ETMX ETMY ITMX ITMY PRM SRM BS
Fri Feb 24 04:11:15 PST 2012
1014120690
Steve (or anyone), can you restore the watchdogs when you come to the lab in the morning ?
|
Kiwamu (or whoever is here last tonight): please run the free-swing/kick script (/opt/rtcds/caltech/c1/scripts/SUS/freeswing) before you leave, and I'll check the matrices and update the suspensions tomorrow morning.
|
|
2737
|
Wed Mar 31 02:57:48 2010 |
kiwamu | Update | Green Locking | frequency counter for green PLL |
Rana found that we had a frequency counter SR620 which might be helpful for lock acquisition of the green phase lock.
It has a response of 100MHz/V up to 350MHz which is wide range and good for our purpose. And it has a noise level of 200Hz/rtHz @ 10Hz which is 1000 times worse than that Matt made (see the entry).
The attached figure is the noise curve measured while I injected a signal of several 100kHz. In fact I made sure that the noise level doesn't depends on the frequency of an input signal.
The black curve represents the noise of the circuit Matt has made, the red curve represents that of SR620. |
Attachment 1: FCnoise.png
|
|
2741
|
Wed Mar 31 12:30:31 2010 |
rana | Update | Green Locking | frequency counter for green PLL |
Its a good measurement - you should adjust the input range of the 620 using the front panel 'scale' buttons to see how the noise compares to Matt's circuit when the range is reduced to 1 MHz. In any case, we would use it in the 350 MHz range mode. What about the noise of the frequency discriminator from MITEQ? |
2751
|
Thu Apr 1 15:21:12 2010 |
rana | Update | Green Locking | frequency counter for green PLL |
|
2718
|
Sun Mar 28 17:28:26 2010 |
matt, kiwamu | Update | Green Locking | frequency discriminator for green PLL |
Last Friday, Matt made a frequency discriminator circuit on a bread board in order to test the idea and study the noise level. I think it will work for phase lock acquisition of Green locking.
As a result a response of 100kHz/V and a noise level of 2uV/rtHz @ 10Hz are yielded. This corresponds to 0.2Hz/rtHz @ 10Hz.
The motivation of using frequency discriminators is that it makes a frequency range wider and easier for lock acquisition of PLLs in green locking experiment.
For the other possibility to help phase lock acquisition, Rana suggested to use a commercial discriminator from Miteq.
(principle idea)
The diagram below shows a schematic of the circuit which Matt has built.

Basically an input signal is split into two signals right after the input, then one signal goes through directly to a NAND comparator.
On the other hand another split signal goes through a delay line which composed by some RC filters, then arrive at the NAND comparator with a certain amount of delay.
After going through the NAND comparator, the signal looks like a periodic pulses (see below).
If we put a signal of higher frequency we get more number of pulses after passing through the NAND.

Finally the pulse-signal will be integrated at the low pass filter and converted to a DC signal.
Thus the amplitude of DC signal depends on the number of the pulses per unit time, so that the output DC signal is proportional to the frequency of an input signal.
(result)
By putting a TTL high-low signal, an output of the circuit shows 100kHz/V linear response.
It means we can get DC voltage of 1 V if a signal of 100kHz is injected into the input.
And the noise measurement has been done while injecting a input signal. The noise level of 0.2Hz/rtHz @ 10 Hz was yielded.
Therefore we can lock the green PLL by using an ordinary VCO loop after we roughly guide a beat note by using this kind of discriminator.
 |
Attachment 1: DSC_1407.JPG
|
|
Attachment 2: FD.png
|
|
Attachment 3: FDnoise.png
|
|
2728
|
Mon Mar 29 15:19:33 2010 |
mevans | Update | Green Locking | frequency discriminator for green PLL |
Thanks for the great entry!
In order to make this work for higher frequencies, I would add Hartmut's suggestion of a frequency dividing input stage. If we divide the input down by 100, the overall range will be about 200MHz, and the noise will be about 20Hz/rtHz. That might be good enough... but we can hope that the commercial device is lower noise!
Quote: |
Last Friday, Matt made a frequency discriminator circuit on a bread board in order to test the idea and study the noise level. I think it will work for phase lock acquisition of Green locking.
As a result a response of 100kHz/V and a noise level of 2uV/rtHz @ 10Hz are yielded. This corresponds to 0.2Hz/rtHz @ 10Hz.
The motivation of using frequency discriminators is that it makes a frequency range wider and easier for lock acquisition of PLLs in green locking experiment.

|
|
6553
|
Fri Apr 20 23:02:25 2012 |
Den | Update | Adaptive Filtering | frequency domain filter |
DFT-LMS is a frequency domain adaptive filter that demonstrates faster convergence compared to the time-domain LMS filter. I've tested Discrete Fourier Transform (DFT-LMS) filter. It converts witness signal to the frequency domain using DFT and corrects the eigenvalues of the covariance matrix to make them as equal to each other as possible (does pre-whitenning of the witness signal).
Left plot compares learning curves for time domain LMS and DFT-LMS algorithms on the simulated data from seismometers and mcl (number of averages = 30) Right plot shows the evolution of the filter coefficients norm (Euclidean norms of the coefficient vector). Though LMS algorithm works in the time domain and DFT-LMS in the frequency domain, the coefficient vectors must have the same length, because we Fourier Transform is achieved by applying a unitary operator => vector norm must not change.

Plots show that both algorithms converge to the same coefficients vector norm, but DFT-LMS does it much faster then LMS.
Online realization:
Good news: algorithm complexity is linear in filter length. Though the algorithm does Fourier transform, its complexity is still O(M), M - number of coefficients. Simulations show that DFT-LMS is ~8-9 times slower then LMS. This is not so bad, may be we can do even slightly better.
Bad news: downsample process is not simple. Due to Fourier transform, the filter needs the whole witness signal vector before calculating the output. This is sad and in contrast with LMS algorithm where we could start to calculate the new output immediately after computing the previous output. We either need to calculate the whole output immediately or introduce delay in the output or approximate Fourier transform with some previous witness signal values.
Realization in the kernel: I asked Alex about complex numbers, exponents, sin and cos functions in the kernel c and he answers that we do not have complex numbers, about exp, cos, sin he is not sure. But for DFT-LMS algorithm we are able to get round of these difficulties. Complex numbers will be presented as 2 real numbers. Then exp (a) = cos(a) + i*sin(a). All what we need for DFT-LMS are sin(2 * pi * k / M) and cos(2 * pi * k / M), k=0,1,2,...,M-1. Fortunately, M - (filter length) is big enough, typical value pi/M ~ 0.001 and we can calculate sin(2*pi/M) and cos(2*pi/M) using Taylor series. As the argument is small, 5-6 terms will be enough to get precision ~1e-20. Then we build the whole table of cos and sin according to induction cos(2*pi/M*k) = cos(2*pi/M*(k-1))cos(2*pi/M) - sin(2*pi/M*(k-1))sin(2*pi/M), sin(2*pi/M*k) = cos(2*pi/M*(k-1))sin(2*pi/M) + sin(2*pi/M*(k-1))cos(2*pi/M). We should do it only once, so the algorithm will build these values in the beginning during first several iterations, then will use them.
The main problem is downsampling. I need to think more about it. |
7000
|
Sat Jul 21 18:04:02 2012 |
Den | Update | Adaptive Filtering | frequency domain filter |
I've implemented online frequency domain filter and applied it to MC_F.
+
Magnitude of the filter output at 1 Hz is the same as MC_F. This means that it is not hard for FIR to match the resonance. The problem is with the phase. We can not match the resonance exactly. If the resonance is at f0 and we match at f0 +/- df then in the frequency range (f0, f0 +/- df) the phase is not matched for 180. I guess the filter does not diverge because df is small but also the filter can not account for this huge phase lag. We need to slightly change the simulated actuator TF and see how the filter will react.
|
2076
|
Fri Oct 9 16:36:13 2009 |
rob | Update | IOO | frequency noise problem |
I used the XARM as a reference to measure the frequency noise after the MC. It's huge around 4kHz--hundreds of times larger than the frequency noise the MC servo is actually squashing. This presents a real problem for our noise performance.
An elog search reveals that this noise has been present (although not calibrated till now) for years. We're not sure what's causing it, but suspicion falls on the piezojena input PZTs.
I didn't bother too much about it before because we previously had enough common mode servo oomph to squash it below other DARM noises, and I didn't worry too much about stuff at 4kHz.. Now that we have a weaker FSS and thus much weaker CM servo, we can't squash it, and the most interesting feature of our IFO is at 4kHz.
I'll measure the actual voltage noise going to the PZTs. I remember doing this before and concluding it was ok, but can't find an elog entry. So this time maybe I'll do it right. |
Attachment 1: freqnoiseaftermc.png
|
|
2172
|
Tue Nov 3 03:45:04 2009 |
rob | Update | IOO | frequency noise problem |
Quote: |
I used the XARM as a reference to measure the frequency noise after the MC. It's huge around 4kHz--hundreds of times larger than the frequency noise the MC servo is actually squashing. This presents a real problem for our noise performance.
An elog search reveals that this noise has been present (although not calibrated till now) for years. We're not sure what's causing it, but suspicion falls on the piezojena input PZTs.
I didn't bother too much about it before because we previously had enough common mode servo oomph to squash it below other DARM noises, and I didn't worry too much about stuff at 4kHz.. Now that we have a weaker FSS and thus much weaker CM servo, we can't squash it, and the most interesting feature of our IFO is at 4kHz.
I'll measure the actual voltage noise going to the PZTs. I remember doing this before and concluding it was ok, but can't find an elog entry. So this time maybe I'll do it right.
|
This level of frequency noise has not changed, but we now have increased common mode servo gain and so it's not as huge of a deal, although we should still probably do something about it.
Attached is a plot of the piezojena noise measurement, estimated into Hz, along with another measurement of frequency noise as described above.
To get the piezojena voltage noise into Hz, I estimated the PZTs within have a flat 2 micron/V response (based on a rough knowledge of their geometry and assuming a 10 milliradian / 150V steering range). This is the voltage noise with the PZTs operating in closed loop mode, which is how we normally run them. This plot also ignores the transfer function of the Pomona box, as we are mainly looking at noise in the kHz band. I think this plot shows that these PZTs are a good candidate for creating this frequency noise, especially near their mechanical resonances (the manual says the unloaded resonances are in the 3-4kHz range).
I've been operating one DOF of the piezojenas in open loop mode for a couple of weeks now, and the feared drift has not been a problem at all. If we plan to keep using these after the upgrade, we should definitely put some big resistors in series at the outputs and operate them in open loop mode.
Also attached is a plot of RF DARM noise, with a frequency noise spectrum. That spectrum is a REFL 2I spectrum put into DARM units using a measured TF (driving MC_L and measuring REFL 2I and DARM_ERR), and then put into meters using the same DARM calibration as used for the DARM curve. |
Attachment 1: noise.png
|
|
Attachment 2: spectra.pdf
|
|
2357
|
Sat Dec 5 17:34:30 2009 |
rob | Update | IOO | frequency noise problem |
There's a large broadband increase in the MC_F spectrum. I'm not totally sure it's real--it could be some weird bit-swapping thing. I've tried soft reboots of c1susvme2 and c1iovme, which haven't helped. In any case, it seems like this is preventing any locking success today. Last night it was fine. |
Attachment 1: mcf.png
|
|
2359
|
Sat Dec 5 22:31:52 2009 |
rob | Update | IOO | frequency noise problem |
Quote: |
There's a large broadband increase in the MC_F spectrum. I'm not totally sure it's real--it could be some weird bit-swapping thing. I've tried soft reboots of c1susvme2 and c1iovme, which haven't helped. In any case, it seems like this is preventing any locking success today. Last night it was fine.
|
Rebooting c1iovme (by keying off the crate, waiting 30 seconds, and then keying it back on and restarting) has resolved this. The frequency noise is back to the 'usual' trace. |
14086
|
Thu Jul 19 04:44:09 2018 |
Annalisa, Terra | Summary | Thermal Compensation | frequency shift observed with heating! |
Annalisa, Gautum, Koji, Terra
Summary: with the reflector setup, we measured a frequency shift of the first and second order modes! First looks of shifts show 1st HOM shift ~-10 kHz, 2nd HOM shift ~-18 kHz (carrier ~4 kHz). We saw no shift with the cylinder/lenses set up.
- - - - -
Tonight we modified the cavity scan setup: the LO is provided by the Marconi which, at the same time, is also used to scan the AUX laser frequency instead of the Agilent. In order to get rid of the free running noise between Marconi and Agilent, the Marconi frequency was scanned and, point by point, the Agilent center frequency was changed accordingly. In order to speed up the process, the whole procedure was automated. The script is called AGfast.py and can be found in /users/annalisa/postVent.
One thing that helped in improving the data quality of the phase information was to set the Agilent IF bandwidth @1kHz. Not yet clear why, but it was better than having a lower bandwidth. To be further investigated.
With this setup, we made some coarse scan of the full FSR and then we "zoomed" around the main peaks in order to increase the resolution and get a more precise information about the peak frequency.
Here are the frequency ranges that we scanned:
- carrier - central frequency: 31.73MHz; range: [31.68MHz - 31.78MHz]
- HOM1 - central frequency: 32.88MHz; range: [32.84MHz - 32.93MHz]
- HOM2 - central frequency: 34.03MHz; range: [33.95MHz - 34.06MHz]
- HOM3 - central frequency: 35.18MHz; range: [35.09MHz - 35.25MHz]
We powered the heater of the lenses setup @4:55 am at 14.4V and 0.9A. Then we slightly increased the power @5:05am and the final "hot state" configuration is with heater powered at 16V and 0.9A.
With this setup we couldn't see any frequency shift
Then, at around 6:30 am we turned on the reflector setup and we measured a frequency shift of the first and second order modes. First scans show 1st HOM shift ~10 kHz, 2nd HOM shift ~18 kHz. First attachment shows carrier hot/cold, second attachment shows HOM2 hot/cold. We started to get plauged by high seismic noise. Heaters turned off at 7:45 am. Lots of scans and actual analysis to go.
gautam: about the questionable plotting -
- 10 faint (alpha~0.3) lines are individual measurements with the reflector doing its heating. (AG4395A, 0 span, single frequency measurements plotted together).
- charcoal line, labelled mean, is the mean of the 10 above lines.
- bright green ("Reference") is the mean of a coarse scan (cold ETM) overlaid for comparison.
- "cold" - self explanatory.
My personal favourite plot is Attachment #3, which is a 5 MHz scan (cold) to identify positions of the various peaks. The power of including phase information in the analysis is clear. The second FSR on the right edge of the plot is not as prominent as the first is because the arm transmission was degrading throughout the measurement. For future measurements, we should consider locking the IMC length to the arm cavity - this would eliminate such alignment drifts, and maybe also make the PLL control signal RMS smaller. |
Attachment 1: scanning_fine_2018-07-19-07-32-08_parsed.pdf
|
|
Attachment 2: scanning_fine_2018-07-19-06-57-47_parsed.pdf
|
|
Attachment 3: Yscan_scanning_parsed_2am.txt.pdf
|
|
9224
|
Wed Oct 9 15:00:48 2013 |
Steve | Update | PEM | freshmen tour |
[Alan, Koji, Manasa]
Record number of 50 freshmen were given tour of the 40m this afternoon. |
Attachment 1: freshmen2013.jpg
|
|
7493
|
Fri Oct 5 16:21:48 2012 |
Steve | Update | General | freshmen visiters |
40 plus freshmen visited the 40m today |
Attachment 1: IMG_1693.JPG
|
|
Attachment 2: IMG_1694.JPG
|
|
6203
|
Tue Jan 17 02:27:49 2012 |
kiwamu | Update | LSC | fringe tests : all the suspensions are innocent |
I did a quick test to check a hypothesis that PRM is interfering with some other optics in the single bounce configuration.
I shook all the suspensions (except the MC mirrors) at 3 Hz in POS, PIT and YAW with an amplitude of 1000 counts.
No effects were found in the REFL demod signals.
So it is NOT a fringe effect caused by the other suspended mirrors.
Quote from #6202 |
The REFL11 and REFL55 demod signals show high frequency noise depending on how big signals go to the POS actuator of PRM.
Is PRM making some fringes with some other optics ??
|
|
6088
|
Thu Dec 8 11:59:53 2011 |
Zach | Update | RF System | fringing indeed |
Here is a trend of 11 & 55 I&Q, along with the EOM temperature and PSL RMTEMP signals. You can see that there is definitely some fringe-like behavior for monotonic changes in temperature. This is consistent with what I have seen on the gyro table in the past.
Some other notes:
- The EOM temperature (or at least the sensor temperature) seems to track RMTEMP almost exactly when there is no foam box on the EOM. I have verified that the max-min swing here is the same for both signals (~0.77 K).
- Something crazy appears to happen at ~10:15, and all the RAM signals get much noisier. Does anyone know what happened at this time (2:15am local)?
We ought to get to the bottom of the fringing. The CTE of LiNbO3 is ~2 ppm/K, so given that the wavelength is on the order of 0.5 K, this is probably not caused by the etalon effect (2ppm/K * 0.5K * ~1cm << 1064nm).

|
7197
|
Wed Aug 15 17:23:22 2012 |
jamie | Update | CDS | front end IOP models changed to reflect actual physical hardware |
As Rolf pointed out when he was here yesterday, all of our IOPs are filled with parts for ADCs and DACs that don't actually exist in the system. This was causing needless module error messages and IOP GDS screens that were full of red indicators. All the IOP models were identically stuffed with 9 ADC parts, 8 DAC parts, and 4 BO parts, even though none of the actual front end IO chassis had physical configurations even remotely like that. This was probably not causing any particular malfunctions, but it's not right nonetheless.
I went through each IOP, c1x0{1-5}, and changed them to reflect the actual physical hardware in those systems. I have committed these changes to the svn, but I haven't rebuilt the models yet. I'll need to be able to restart all models to test the changes, so I'm going to wait until we have a quiet time, probably next week. |
9074
|
Tue Aug 27 19:34:36 2013 |
Jamie | Configuration | CDS | front end IPC configuration |
So the IPC situation on the front end network is not so great right now. For various no-longer-valid reasons, c1lsc had no RFM card, all the IPC connections were routed through the c1rfm model on c1sus, and routed to c1lsc via dolphin PCIe as needed. As things grew, c1rfm became overloaded. Koji tried to fix the situation by breaking things out of c1rfm to make direct connections where we could. This cleared up c1rfm a bit, but not c1mcs is overloading.
Reminder: PCIe (dolphin) is faster and higher bandwidth than RFM. The more things we can put on PCIe the better.
Attached is a graph of my rough accounting of the intended direct IPC connections between the front ends. By "intended direct" I mean what should be direct connections if we had all the appropriate hardware. Right now the actual connection graph is more convoluted than this since things are passing through c1rfm. I note this graph was NOT particularly easy to make, which is very unfortunate. I had to manually look through every model and determine the ultimate source of every incoming IPC. Kind of a pain in the butt. It would be nice if there was a simple way to represent this.
Here are some various solutions to the problem as I see it:
a) put c1lsc on the RFM network
This would allow c1lsc to talk to c1ioo, c1iscex, and c1iscey without having to go through c1sus, thereby eliminating c1rfm altogether. I'm not sure why we didn't just do this originally.
Requires:
b) put c1ioo on the PCIe network (and move c1sus's RFM card to c1lsc)
This is probably the most robust solution.
b1) There are roughly 8 IPCs going from c1ioo to c1sus, and 4 going the other way, and 3 IPCs from c1ioo to c1lsc. If we put c1ioo on PCIe all of these now RFM connections would become direct PCIe connections, which would be a big win.
At this point only the end station front ends would be on RFM, and most of the connections to those come from c1lsc, so it would make sense to give c1lsc the RFM card, thereby eliminating a lot of stuff from c1rfm.
Requires:
- dolphin card for c1ioo (do the old sun machines support these? if they don't we could swap the old sun machine with a new spare aLIGO-approved supermicro machines, which we have spares of)
- dolphin fibre to go to dolphin switch in 1X3 rack
b2) OR, we could move c1ioo to 1X4 with c1lsc and c1sus, and get a OneStop fibre cable to connect to its IO chassis. We would still need a dolphin card, but we could use coper instead of fibre. This is my preferred solution, since it moves c1ioo out of 1X1, where it's really in the way and making a lot of noise. It would also be easier to manage all the machines if they're together in one rack.
Requires:
- dolphin card for c1ioo
- dolphin coper cable for c1ioo
- OneStop fibre for c1ioo
c) put another cpu in c1sus
c1sus is (I believe) able to support another 6-core cpu. If we added more cores to c1sus, we could break up c1rfm into c1rfm0, c1rfm1, etc. This is a less elegant solution imho, but it would probably do the job.
Requires:
|
Attachment 1: hosts.png
|
|
9076
|
Tue Aug 27 20:43:34 2013 |
Koji | Configuration | CDS | front end IPC configuration |
The reason we had the PCIe/RFM system was to test this mixed configuration in prior to the actual implementation at the sites.
Has this configuration been intesively tested at the site with practical configuration?
Quote: |
Attached is a graph of my rough accounting of the intended direct IPC connections between the front ends.
|
It's hard to believe that c1lsc -> c1sus only has 4 channels. We actuate ITMX/Y/BS/PRM/SRM for the length control.
In addition to these, we control the angles of ITMX/Y/BS/PRM (and SRM in future) via c1ass model on c1lsc.
So there should be at least 12 connections (and more as I ignored MCL).
I personally prefers to give the PCIe card to c1ioo and move the RFM card to c1lsc.
But in either cases, we want to quantitatively compare what the current configuration is (not omitting the bridging by c1rfm),
and what the future configuration will be including the addtional channels we want add in close future,
because RFM connections are really costly and moving the RFM card to c1lsc may newly cause the timeout of c1lsc
just instead of c1sus. |
9086
|
Wed Aug 28 19:47:28 2013 |
jamie | Configuration | CDS | front end IPC configuration |
Quote: |
It's hard to believe that c1lsc -> c1sus only has 4 channels. We actuate ITMX/Y/BS/PRM/SRM for the length control.
In addition to these, we control the angles of ITMX/Y/BS/PRM (and SRM in future) via c1ass model on c1lsc.
So there should be at least 12 connections (and more as I ignored MCL).
|
Koji was correct that I missed some connections from c1lsc to c1sus. I corrected the graph in the original post.
Also, I should have noted, that that graph doesn't actually include everything that we now have. I left out all the simplant stuff, which adds extra connections between c1lsc and c1sus, mostly because the sus simplant is being run on c1lsc only because there was no space on c1sus. That should be corrected, either by moving c1rfm to c1lsc, or by adding a new core to c1sus.
I also spoke to Rolf today and about the possibility of getting a OneStop fiber and dolphin card for c1ioo. The dolphin card and cable we should be able to order no problem. As for the OneStop, we might have to borrow a new fiber-supporting card from India, then send our current card to OneStop for fiber-supporting modifications. It sounds kind of tricky. I'll post more as I figure things out.
Rolf also said that in newer versions of the RCG, the RFM direct memory access (DMA) has improved in performance considerably, which reduces considerably the model run-time delay involved in using the RFM. In other words, the long awaited RCG upgrade might alleviate some of our IPC woes.
We need to upgrade the RCG to the latest release (2.7) |
13138
|
Mon Jul 24 19:28:55 2017 |
Jamie | Update | CDS | front end MX stream network working, glitches in c1ioo fixed |
MX/OpenMX network running
Today I got the mx/open-mx networking working for the front ends. This required some tweaking to the network interface configuration for the diskless front ends, and recompiling mx and open-mx for the newer kernel. Again, this will all be documented.
controls@fb1:~ 0$ /opt/mx/bin/mx_info
MX Version: 1.2.16
MX Build: root@fb1:/opt/src/mx-1.2.16 Mon Jul 24 11:33:57 PDT 2017
1 Myrinet board installed.
The MX driver is configured to support a maximum of:
8 endpoints per NIC, 1024 NICs on the network, 32 NICs per host
===================================================================
Instance #0: 364.4 MHz LANai, PCI-E x8, 2 MB SRAM, on NUMA node 0
Status: Running, P0: Link Up
Network: Ethernet 10G
MAC Address: 00:60:dd:43:74:62
Product code: 10G-PCIE-8B-S
Part number: 09-04228
Serial number: 485052
Mapper: 00:60:dd:43:74:62, version = 0x00000000, configured
Mapped hosts: 6
ROUTE COUNT
INDEX MAC ADDRESS HOST NAME P0
----- ----------- --------- ---
0) 00:60:dd:43:74:62 fb1:0 1,0
1) 00:30:48:be:11:5d c1iscex:0 1,0
2) 00:30:48:bf:69:4f c1lsc:0 1,0
3) 00:25:90:0d:75:bb c1sus:0 1,0
4) 00:30:48:d6:11:17 c1iscey:0 1,0
5) 00:14:4f:40:64:25 c1ioo:0 1,0
controls@fb1:~ 0$
c1ioo timing glitches fixed
I also checked the BIOS on c1ioo and found that the serial port was enabled, which is known to cause timing glitches. I turned off the serial port (and some power management stuff), and rebooted, and all the c1ioo timing glitches seem to have gone away.
It's unclear why this is a problem that's just showing up now. Serial ports have always been a problem, so it seems unlikely this is just a problem with the newer kernel. Could the BIOS have somehow been reset during the power glitch?
In any event, all the front ends are now booting cleanly, with all dolphin and mx networking coming up automatically, and all models running stably:

Now for daqd... |
3292
|
Mon Jul 26 12:31:36 2010 |
kiwamu | Update | CDS | front end machine for the X end |
A brief report about the new front end machine "C1ISCEX" installed on the X end (old Y end).
Still the DAC is not working.
- Timing card
It's working correctly.
The 1PPS clock signal is supplied by the vertex clock distributer via a 40m long fiber.
- ADC
All 16 channels are working well.
We can see the signals in the medm screen while injecting some signals to the ADC by using a function generator.
-DAC
All 16 channels do NOT work.
We can not see any signals at the DAC SCSI cable while digitally injecting a signal on the medm screen. |
16302
|
Thu Aug 26 10:30:14 2021 |
Jamie | Configuration | CDS | front end time synchronization fixed? |
I've been looking at why the front end NTP time synchronization did not seem to be working. I think it might not have been working because the NTP server the front ends were point to, fb1, was not actually responding to synchronization requests.
I cleaned up some things on fb1 and the front ends, which I think unstuck things.
On fb1:
- stopped/disabled the default client (systemd-timesyncd), and properly installed the full NTP server (ntp)
- the ntp server package for debian jessie is old-style sysVinit, not systemd. In order to make it more integrated I copied the auto-generated service file to /etc/systemd/system/ntp.service, and added and "[install]" section that specifies that it should be available during the default "multi-user.target".
- "enabled" the new service to auto-start at boot ("sudo systemctl enable ntp.service")
- made sure ntp was configured to serve the front end network ('broadcast 192.168.123.255') and then restarted the server ("sudo systemctl restart ntp.service")
For the front ends:
- on fb1 I chroot'd into the front-end diskless root (/diskless/root) and manually specifed that systemd-timesyncd should start on boot by creating a symlink to the timesyncd service in the multi-user.target directory:
$ sudo chroot /diskless/root
$ cd /etc/systemd/system/multi-user.target.wants
$ ln -s /lib/systemd/system/systemd-timesyncd.service
- on the front end itself (c1iscex as a test) I did a "systemctl daemon-reload" to force it to reload the systemd config, and then restarted the client ("systemctl restart systemd-timesyncd")
- checked the NTP synchronization with timedatectl:
controls@c1iscex:~ 0$ timedatectl
Local time: Thu 2021-08-26 11:35:10 PDT
Universal time: Thu 2021-08-26 18:35:10 UTC
RTC time: Thu 2021-08-26 18:35:10
Time zone: America/Los_Angeles (PDT, -0700)
NTP enabled: yes
NTP synchronized: yes
RTC in local TZ: no
DST active: yes
Last DST change: DST began at
Sun 2021-03-14 01:59:59 PST
Sun 2021-03-14 03:00:00 PDT
Next DST change: DST ends (the clock jumps one hour backwards) at
Sun 2021-11-07 01:59:59 PDT
Sun 2021-11-07 01:00:00 PST
controls@c1iscex:~ 0$
Note that it is now reporting "NTP enabled: yes" (the service is enabled to start at boot) and "NTP synchronized: yes" (synchronization is happening), neither of which it was reporting previously. I also note that the systemd-timesyncd client service is now loaded and enabled, is no longer reporting that it is in an "Idle" state and is in fact reporting that it synchronized to the proper server, and it is logging updates:
controls@c1iscex:~ 0$ sudo systemctl status systemd-timesyncd
â— systemd-timesyncd.service - Network Time Synchronization
Loaded: loaded (/lib/systemd/system/systemd-timesyncd.service; enabled)
Active: active (running) since Thu 2021-08-26 10:20:11 PDT; 1h 22min ago
Docs: man:systemd-timesyncd.service(8)
Main PID: 2918 (systemd-timesyn)
Status: "Using Time Server 192.168.113.201:123 (ntpserver)."
CGroup: /system.slice/systemd-timesyncd.service
└─2918 /lib/systemd/systemd-timesyncd
Aug 26 10:20:11 c1iscex systemd[1]: Started Network Time Synchronization.
Aug 26 10:20:11 c1iscex systemd-timesyncd[2918]: Using NTP server 192.168.113.201:123 (ntpserver).
Aug 26 10:20:11 c1iscex systemd-timesyncd[2918]: interval/delta/delay/jitter/drift 64s/+0.000s/0.000s/0.000s/+26ppm
Aug 26 10:21:15 c1iscex systemd-timesyncd[2918]: interval/delta/delay/jitter/drift 128s/-0.000s/0.000s/0.000s/+25ppm
Aug 26 10:23:23 c1iscex systemd-timesyncd[2918]: interval/delta/delay/jitter/drift 256s/+0.001s/0.000s/0.000s/+26ppm
Aug 26 10:27:40 c1iscex systemd-timesyncd[2918]: interval/delta/delay/jitter/drift 512s/+0.003s/0.000s/0.001s/+29ppm
Aug 26 10:36:12 c1iscex systemd-timesyncd[2918]: interval/delta/delay/jitter/drift 1024s/+0.008s/0.000s/0.003s/+33ppm
Aug 26 10:53:16 c1iscex systemd-timesyncd[2918]: interval/delta/delay/jitter/drift 2048s/-0.026s/0.000s/0.010s/+27ppm
Aug 26 11:27:24 c1iscex systemd-timesyncd[2918]: interval/delta/delay/jitter/drift 2048s/+0.009s/0.000s/0.011s/+29ppm
controls@c1iscex:~ 0$
So I think this means everything is working.
I then went ahead and reloaded and restarted the timesyncd services on the rest of the front ends.
We still need to confirm that everything comes up properly the next time we have an opportunity to reboot fb1 and the front ends (or the opportunity is forced upon us).
There was speculation that the NTP clients on the front ends (systemd-timesyncd) would not work on a read-only filesystem, but this doesn't seem to be true. You can't trust everything you read on the internet. |
2982
|
Tue May 25 16:32:26 2010 |
kiwamu | HowTo | Electronics | front ends are back |
[Alex, Joe, Kiwamu]
Eventually all the front end computers came back !! 
There were two problems.
(1): C0DCU1 didn't want to come back to the network. After we did several things it turned the ADC board for C0DCU1 didn't work correctly.
(2): C1PEM1 and C0DAQAWG were cross-talking via the back panel of the crate.
(what we did)
* installed a VME crate with single back panel to 1Y6 and mounted C1PEM1 and C0DAQAWG on it. However it turned out this configuration was bad because the two CPUs could cross-talk via the back panel.
* removed the VME crate and then installed another VME crate which has two back panels so that we can electrically separate C1PEM1 and C0DAQAWG. After this work, C0DAQAWG started working successfully.
* rebooted all the front ends, fb40m and c1dcuepics.
* reset the RFM bypath. But these things didn't bring C0DCU1 back.
* telnet to C0DCU1 and ran "./startup.cmd" manually. In fact "./startup.cmd" should automatically be called when it boots.
* saw the error messages from "./startup.cmd" and found it failed when initialization of the ADC board. It saids "Init Failure !! could not find ICS"
* went to 1Y7 rack and checked the ADC. We found C0DCU1 had two ADC boards, one of two was not in used.
* disconnected all two ADCs and put back one which had not been in used. At the same time we changed the switching address of this ADC to have the same address as the other ADC.
* powered off/on 1Y7 rack. Finally C0DCU1 got back.
* burtrestored the epics to the last Friday, May 21st 6:07am |
7477
|
Thu Oct 4 14:04:21 2012 |
jamie | Update | CDS | front ends back up |
All the front end machines are back up after the outage. It looks like none of the front end machines came back up once power was restored, and they all needed to be powered manually. One of the things I want to do in the next CDS upgrade is put all the front end computers in one rack, so we can control their power remotely.
c1sus was the only one that had a little trouble. It's timing was for some reason not syncing with the frame builder. Unclear why, but after restarting the models a couple of times things came back.
There's still a little red, but it mostly has to do with the fact that c1oaf is busted and not running (it actually crashes the machine when I tried to start it, so this needs to be fixed!). |
6171
|
Wed Jan 4 16:40:52 2012 |
Jamie | Update | Computers | front-end fb communication restored |
Communication between the front end models and the framebuilder has been restored. I'm not sure exactly what the issue was, but rebuilding the framebuilder daqd executable and restarting seems to have fixed the issue.
I suspect that the problem might have had to do with how I left things after the last attempt to upgrade to RCG 2.4. Maybe the daqd that was running was linked against some library that I accidentally moved after starting the daqd process. It would have kept running fine as was, but if the process died and was attempted to be started again, it's broken linking might have kept it from running correctly. I don't have any other explanation.
It turns out this was not (best I can tell) related to the new year time sync issues that wer seen at the sites. |
13133
|
Sun Jul 23 22:16:55 2017 |
Jamie, gautam | Update | CDS | front-end now running with new OS, RCG |
All front ends and model are (mostly) running now

All suspensions are damped:

It should be possible at this point to do more recovery, like locking the MC.
Some details on the restore process:
- all models were recompiled with the new RCG version 3.0.3
- the new RCG does stricter simulink drawing checks, and was complaining about unterminated outputs in some of the SUS models. Terminated all outputs it was concerned about and saved.
- RCG 3.0 requires a new directory for doing better filter module diagnostics: /opt/rtcds/caltech/c1/chans/tmp
- had to reset the slow machines c1susaux, c1auxex, c1auxey
The daqd is not yet running. This is the next task.
I have been taking copious notes and will fully document the restore process once complete.
c1ioo issues
c1ioo has been giving us a little bit of trouble. The c1ioo model kept crashing and taking down the whole c1ioo host. We found a red light on one of the ADCs (ADC1). We pulled the card and replaced it with a spare from the CDS cabinet. That seemed to fix the problem and c1ioo became more stable.
We've still been seeing a lot of glitching in c1ioo, though, with CPU cycle times frequently (every couple of seconds) running above threshold for all models, up to 200 us. I tried unloading every kernel module I could and shutting down every non-critical process, but nothing seemed to help.
We eventually tried stopping the c1ioo model altogether and that seemed to help quite a bit, dropping the long cycle rate down to something like one every 30 seconds or so. Not sure what that means. We should look into the BIOS again, to see if there could be something interacting with the newer kernel.
So currently the c1ioo model is not running (which is why it's all white in the CDS overview snapshot above). The fact that c1ioo is not running and the remaining models are still occaissionly glitching is also causing various IPC errors on auxilliary models (see c1mcs, c1rfm, c1ass, c1asx).
RCG compile warnings
the new RCG tries to do more checks on custom c code, but it seems to be having trouble finding our custom "ccodeio.h" files that live with the c definitions in USERAPPS/*/common/src/. Unclear why yet. This is causing the RCG to spit out warnings like the following:
Cannot verify the number of ins/outs for C function BLRMS.
File is /opt/rtcds/userapps/release/cds/c1/src/BLRMSFILTER.c
Please add file and function to CDS_SRC or CDS_IFO_SRC ccodeio.h file.
This are just warnings and will not prevent the model form compiling or warning. We'll figure out what the problem is to make these go away, but they can be ignored for the time being.
model unload instability
Probably the worst problem we're facing right now is an instability that will occaissionally, but not always, cause the entire front end host to freeze up upon unloading an RTS kernel module. This is a known issue with the newer linux kernels (we're using kernel version 3.2.35), and is being looked into.
This is particularly annoying with the machines on the dolphin network, since if one of the dolphin hosts goes down it manages to crash all the models reading from the dolphin network. Since half the time they can't be cleanly restarted, this tends to cause a boot fest with c1sus, c1lsc, and c1ioo. If this happens, just restart those machines, wait till they've all fully booted, then restart all the models on all hosts with "rtcds start all" . |
13205
|
Mon Aug 14 19:41:46 2017 |
Jamie | Update | CDS | front-end/DAQ network down for kernel upgrade, and timing errors |
I'm upgrading the linux kernel for all the front ends to one that is supposedly more stable and won't freeze when we unload RTS models (linux-image-3.2.88-csp). Since it's a different kernel version it requires rebuilds of all kernel-related support stuff (mbuf, symmetricom, mx, open-mx, dolphin) and all the front end models. All the support stuff has been upgraded, but we're now waiting on the front end rebuilds, which takes a while.
Initial testing indicates that the kernel is more stable; we're mostly able to unload/reload RTS modules without the kernel freezing. However, the c1iscey host seems to be oddly problematic and has frozen twice so far on module unloads. None of the other hosts have frozen on unload (yet), though, so still not clear.
We're now seeing some timing errors between the front ends and daqd, resulting in a "0x4000" status message in the 'C1:DAQ-DC0_*_STATUS' channels. Part of the problem was an issue with the IRIG-B/GPS receiver timing unit, which I'll log in a separate post. Another part of the problem was a bug in the symmetricom driver, which has been resolved. That wasn't the whole problem, though, since we're still seeing timing errors. Working with Jonathan to resolve. |
13215
|
Wed Aug 16 17:05:53 2017 |
Jamie | Update | CDS | front-end/DAQ network down for kernel upgrade, and timing errors |
The CDS system has now been up moved to a supposedly more stable real-time-patched linux kernel (3.2.88-csp) and RCG r4447 (roughly the head of trunk, intended to be release 3.4). With one major and one minor exception, everything seems to be working:

The remaining issues are:
- RFM network down. The IOP models on all hosts on the RFM network are not detecting their RFM cards. Keith Thorne thinks that this is because of changes in trunk to support the new long-range PCIe that will be used at the sites, and that we just need to add a new parameter to the cdsParameters block in models that use RFM. Him and Rolf are looking into for us.
- The 3.2.88-csp kernel is still not totally stable. On most hosts (c1sus, c1ioo, c1iscex) it seems totally fine and we're able to load/unload models without issue. c1iscey is definitely problematic, frequently freezing on module unload. There must be a hardware/bios issue involved here. c1lsc has also shown some problems. A better kernel is supposedly in the works.
- NDS clients other than DTT are still unable to raise test points. This appears to be an issue with the daqd_rcv component (i.e. NDS server) not properly resolving the front ends in the GDS network. Still looking into this with Keith, Rolf, and Jonathan.
Issues that have been fixed:
- "EDCU" channels, i.e. non-front-end EPICS channels, are now being acquired properly by the DAQ. The front-ends now send all slow channels to the daq over the MX network stream. This means that front end channels should no longer be specified in the EDCU ini file. There were a couple in there that I removed, and that seemed to fix that issue.
- Data should now be recorded in all formats: full frames, as well as second, minute, and raw_minute trends
- All FE and DAQD diagnostics are green (other than the ones indicating the problems with the RFM network). This was fixed by getting the front ends models, mx_stream processes, and daqd processes all compiled against the same version of the advLigoRTS, and adding the appropriate command line parameters to the mx_stream processes.
|