ID |
Date |
Author |
Type |
Category |
Subject |
1192
|
Thu Dec 18 12:52:00 2008 |
Alberto | Configuration | SUS | Mode Cleaner Cavity Alignment |
This morning I found the MC locked to the 10 mode. When I locked it on the 00 mode, it was unstable and eventually it always got locked to the wrong mode.
I looked at the Drift Mon MEDM screen, which shows a reference record for position, pitch and yaw of each mirror, and I found that the MC optics were in a different status. Moving the sliders of the mirrors' actuators, I brought them back to the reference position. Then the lock got engaged and it was stable, although the MC reflection from the photodiode, with the wave front sensors (WFS) off, was about 2V. That's higher than the 0.5V the it could get when we aligned the cavity and the input periscope last time.
With the WFS on, the reflection dropped to 0.3V and, so far, the the cavity has been stably locked. |
1193
|
Thu Dec 18 19:15:54 2008 |
Alberto, Yoichi | Configuration | SUS | Mode Cleaner Cavity Alignment |
Quote: | This morning I found the MC locked to the 10 mode. When I locked it on the 00 mode, it was unstable and eventually it always got locked to the wrong mode.
I looked at the Drift Mon MEDM screen, which shows a reference record for position, pitch and yaw of each mirror, and I found that the MC optics were in a different status. Moving the sliders of the mirrors' actuators, I brought them back to the reference position. Then the lock got engaged and it was stable, although the MC reflection from the photodiode, with the wave front sensors (WFS) off, was about 2V. That's higher than the 0.5V the it could get when we aligned the cavity and the input periscope last time.
With the WFS on, the reflection dropped to 0.3V and, so far, the the cavity has been stably locked. |
This evening the mode cleaner was again locking on a higher mode so we tweaked the mirrors' actuators by their sliders on the MEDM screen until we improved the reflection to 0.3V.
Then we went inside and, on the AS table, we centered the beam on the wave front sensors.
Now the mode cleaner is locked, the reflection is less than 0.3V and the transmission about 3V, tha is it is in ideal conditions. We'll see if it holds. |
1234
|
Fri Jan 16 18:29:08 2009 |
Yoichi | Update | SUS | Oplevs QPDs centered |
Kakeru centered ITMX and BS optical levers with the help of Jenne on the walkie-talkie. |
1314
|
Mon Feb 16 22:58:51 2009 |
rana, yoichi | Configuration | SUS | Hysteresis in SUS from Misalignments |
WE wondered if there was some hysteresis in the SUS alignments. When we leave the optics misaligned for a
long time it seems to take awhile for the optic to settle down. Possibly, this is the slow deformation of
the wires or the clamps.
The attached PNG shows the plot of the bias sliders for a few days. You can see that we misalign some of the
optics much more than the others. This must be stopped.
Kakeru is going to use his nearly complete optical lever calibrations to quatify this by stepping the optics
around and measuring the effect in the optical lever. Of course, the misalignment steps will be too large to
catch on the OL, but he can calibrate the align-sliders into radians to handle this. |
Attachment 1: a.png
|
|
1420
|
Tue Mar 24 09:04:02 2009 |
steve | Update | SUS | 4.8 mag earthquake |
SRM, ITMX, ETMX, ITMY and ETMY lost damping at 4:55am this morning from 4.8 magnitude earthquake.
Their damping were restored.
C1:SUS-ITMX_URSEN_OUTPUT swich was found in off position. It was turned on.
MZehnder and MC were locked.
The WFS qpd spot needs recentering |
1422
|
Tue Mar 24 13:54:49 2009 |
Jenne | Update | SUS | Op Levs Centered |
ITMX, ITMY, BS, SRM, PRM op levs were all recentered. ETM's looked okay enough to leave as-is. |
1439
|
Sun Mar 29 13:44:27 2009 |
steve | Update | SUS | ETMY sus damping restored |
ETMY sus damping was found to be tripped.
It was retored.
All fluorecent light were turned off. Please try to conserve some energy. |
1440
|
Sun Mar 29 17:54:41 2009 |
Yoichi | Update | SUS | MC1 drift investigation continued |
The attached plots show the trend of the MC OSEM signals along with the voltages across the output resistors of the bias current buffers.
The channel assignments are:
MC_TMP1 = LL coil
MC_DRUM1 = UL coil
OSA_APTEMP = UR coil
OSA_SPTEMP = LR coil
Although the amplitude of the drift of MC1 is much larger than that of MC2 and MC3, the shape of the drift looks like a daily cycle (temperature ?).
This time, I reduced the MC1 bias currents to avoid saturation of the ADCs for the channels measuring the voltages across the output resistors.
This may be the reason the MC1 has been non-glitchy for the last day.
OSA_APTEMP (UR Coil) shows a step function like behavior, although it did not show up in the OSEM signals.
This, of course, should not happen.
Today, I went to the MC1 satellite box and found that the 64-pin IDE like connector was broken.
The connector is supposed to sandwich the ribbon cable, but the top piece was loose.
The connector is on the cable connecting the satellite box and the SUS rack.
I replaced the broken connector with a new one. I also swapped the MC1 and MC3 satellite boxes to see if the glitches show up in the MC3.
I restored the bias currents of the MC1 to the original values.
The probes to monitor the voltages across the output resistors are still there. For OSA_SPTEMP, which was saturating the ADC, I put a voltage divider before the ADC. Other channels were very close to saturation but still within the ADC range.
Please leave the MC unlocked at least until the Monday morning.
Also please do not touch the Pomona box hanging in front of the IOO rack. It is the voltage divider. The case is connected to the coil side of the output resistor. If you touch it, the MC1 bias current will change.
|
Attachment 1: Drift1.pdf
|
|
1441
|
Mon Mar 30 09:07:22 2009 |
rana | Update | SUS | MC1 drift investigation continued |
Maybe we can temporarily just disconnect the bias and just use the SUS sliders for bias if there's enough range? |
1444
|
Mon Mar 30 13:29:40 2009 |
Yoichi | Update | SUS | MC1 drift investigation continued |
Quote: | Maybe we can temporarily just disconnect the bias and just use the SUS sliders for bias if there's enough range? |
We could do this, but I'm suspicious of the cables between the coil driver and the coils (including the satellite box). In this case, disabling the bias won't help.
Since the MC1 has been quiet recently, I will just lock the MC and resume the locking. |
1583
|
Wed May 13 21:15:04 2009 |
rana | Summary | SUS | Channel Hopping: That ancient enemy (MC problems) |
The MC side problem could also be the side tramp unit problem. Set the tramp to 0 and see if that helps. |
1584
|
Thu May 14 00:15:39 2009 |
rob | Summary | SUS | Channel Hopping: That ancient enemy (MC problems) |
Quote: | The MC side problem could also be the side tramp unit problem. Set the tramp to 0 and see if that helps. |
This started around April 23, around the time that TP1 failed and we switched to the cryopump, and also when there was a mag 4 earthquake in LA. My money's on the EQ. But I don't know how. |
Attachment 1: sidemon.png
|
|
1586
|
Thu May 14 15:28:28 2009 |
steve | Summary | SUS | April 24 earthquake effect on MC2 |
Quote: |
Quote: | The MC side problem could also be the side tramp unit problem. Set the tramp to 0 and see if that helps. |
This started around April 23, around the time that TP1 failed and we switched to the cryopump, and also when there was a mag 4 earthquake in LA. My money's on the EQ. But I don't know how. |
Only MC2 moved in this earth quake. Was the MC alignment touched up since than?
Have you guys swapped satellite amp of MC3 yet? |
Attachment 1: eq042409.jpg
|
|
1587
|
Thu May 14 16:07:20 2009 |
pete | Summary | SUS | Channel Hopping: That ancient enemy (MC problems) |
Quote: |
Quote: | The MC side problem could also be the side tramp unit problem. Set the tramp to 0 and see if that helps. |
This started around April 23, around the time that TP1 failed and we switched to the cryopump, and also when there was a mag 4 earthquake in LA. My money's on the EQ. But I don't know how. |
I wonder if this is still a problem. It has been quiet for a day now. I've attached a day-long trend. Let's see what happens. |
Attachment 1: mc3_5days.jpg
|
|
1588
|
Fri May 15 00:02:34 2009 |
pete | Update | SUS | ETMX coils look OK |
I checked the four rear coils on ETMX by exciting XXCOIL_EXC channel in DTT with amplitude 1000@ 500 Hz and observing the oplev PERROR and YERROR channels. Each coil showed a clear signal in PERROR, about 2e-6 cts. Anyway, the coils passed this test.
|
1603
|
Mon May 18 21:34:18 2009 |
rana | Configuration | SUS | ETMY f2pRatio script run |
Now that the ETMY optical lever is not so bad, I ran the f2pRatio script and it seems to have worked.
I cleaned up the script a little also. Updated in the SVN.
ETMY's A2L scripts have to be run to reduce the A2L noise once the arm is locked again. Might also need
to set the OL UGF too. |
1605
|
Tue May 19 12:30:41 2009 |
rob | Configuration | SUS | ETMY f2pRatio script run |
Quote: | Now that the ETMY optical lever is not so bad, I ran the f2pRatio script and it seems to have worked.
I cleaned up the script a little also. Updated in the SVN.
ETMY's A2L scripts have to be run to reduce the A2L noise once the arm is locked again. Might also need
to set the OL UGF too. |
Just to show, in part, what the script does.
The F2A filters are turned on at 12:21, and the oplev no longer responds to large LSC drives in ETMY. |
Attachment 1: f2ademo.png
|
|
1616
|
Thu May 21 18:05:03 2009 |
pete | Update | SUS | ETMX coils look OK |
Quote: |
I checked the four rear coils on ETMX by exciting XXCOIL_EXC channel in DTT with amplitude 1000@ 500 Hz and observing the oplev PERROR and YERROR channels. Each coil showed a clear signal in PERROR, about 2e-6 cts. Anyway, the coils passed this test.
|
I also made xfer fctns of the 4 piston coils on ETMY and ETMX with OL_PIT. (I looked at all 4 even though the attached plot only shows three.) So it looks ike the coils are OK. |
Attachment 1: etmx_etmy_coils.pdf
|
|
1620
|
Fri May 22 01:27:14 2009 |
pete | Update | SUS | 200 days of MC3 side |
Looks like something went nuts in late April. We have yet to try a hard reboot. |
Attachment 1: mc3_side_200days.png
|
|
1796
|
Mon Jul 27 14:12:14 2009 |
rana | Summary | SUS | TM Coil Noise Spectra |
Rob noticed that the ITMY DAC channels were saturating occassionally. Looking at the spectrum we can see why.
With an RMS of 10000 cts, the peak excursions sometimes cause saturations.
There's a lot of mechanical noise which is showing up on the ITM oplevs and then going to the mirror via
the oplev servo. We need to reduce the mechanical noise and/or modify the filters to compensate. The ITM
COIL_OUT RMS needs to be less than ~3000. |
Attachment 1: Coils.pdf
|
|
2024
|
Tue Sep 29 23:43:46 2009 |
rob | Update | SUS | ITMY UL OSEM |
We had a redo of elog entry 975 tonight. The noisy OSEM was fixed by jiggling the rack end of the long cable. Don't know exactly where--I also poked around the OSEM PD interface board.
In the attached PDF the reference trace is the noisy one. |
Attachment 1: ITMYosemBAD.pdf
|
|
2031
|
Thu Oct 1 08:37:43 2009 |
steve | Update | SUS | suspention damping restored and MZ HV stuck |
Earthquake of magnitude 5.0 shakes ETMY loose.
MC2 lost it's damping later. |
Attachment 1: eq5oct1.jpg
|
|
2034
|
Thu Oct 1 11:39:47 2009 |
Jenne | Update | SUS | MC2 damping restored again |
Quote: |
The EQ did not change the input beam pointing. All back to normal, except MC2 wachdogs tripped again.
|
Round 3 for the day of MC2 watchdogs tripping. |
2036
|
Thu Oct 1 14:22:28 2009 |
rob | Update | SUS | all suspensions undamped |
Quote: |
Quote: |
The EQ did not change the input beam pointing. All back to normal, except MC2 wachdogs tripped again.
|
Round 3 for the day of MC2 watchdogs tripping.
|
I've watchdogged all the suspensions while I mess around with computers. If no one else is using the IFO, we can leave them undamped for a couple of hours to check the resonant frequencies, as long as I don't interrupt data streams with my computer hatcheting. |
2039
|
Thu Oct 1 19:18:24 2009 |
Koji | Update | SUS | all suspensions undamped |
Ops. I restored the damping of the suspensions at around 16:30.
Quote: |
Quote: |
Quote: |
The EQ did not change the input beam pointing. All back to normal, except MC2 wachdogs tripped again.
|
Round 3 for the day of MC2 watchdogs tripping.
|
I've watchdogged all the suspensions while I mess around with computers. If no one else is using the IFO, we can leave them undamped for a couple of hours to check the resonant frequencies, as long as I don't interrupt data streams with my computer hatcheting.
|
|
2193
|
Fri Nov 6 12:56:30 2009 |
Haixing | Update | SUS | Magnet has been levitated |
In this experiment, we used a feedback control to create a stable trap for a NdFeB permanent magnet. The block diagram is the following:

The displacement of the magnet is sensed by the Hall-effect sensor, whose output voltage is proportional to the magnetic flux produced
by the permanent magnet. It has a flat response within the frequencies we are interested in. It is driven by a 5 V power supplier and its
output has a DC voltagle of 2.5 V. We subtracted the DC voltage and used the resulting signal as the error signal. This was
simply achieved by using two channels "A" and "B". The output is "A-B" with a gain equal to one. We then put the error
signal into another SR560 as a low-pass filter with a gain of 100 above 30 Hz. We used the "DC" coupling modes in both
preamplifers. The output is then used to drive a coil. The coil has a dimension of 1.5 inch in diameter and 2 inch in length.
The inductance of the coil is around 0.5 H and the resistance is 4.7 Om. Therefore, it has a corner frequency aournd 10/2pi Hz.
The coil has a iron core inside to enhance the DC force to the permanent magnet. The low-frequency 1/f response of the magnet is produced
by the eddy current damping of the aluminum plane that is below the magnet. This 1/f response is essential for a stable configuration. In the
next stage, we will remove the aluminum plane, and instead we will use a filter to create similar transfer function. At high-frequencies, it behaves as
a free-mass and has a 1/f^2 response. Finally, the magnet is stably levitated.
|
Attachment 1: DSC_0964.JPG
|
|
2204
|
Sun Nov 8 14:18:25 2009 |
Alberto | Update | SUS | ETMY Watchdogs tripped |
This afternoon I re-enabled the ETMY coils after I found that the watchdogs for the mirror had tripped last night at 2:06am. |
2207
|
Mon Nov 9 08:43:16 2009 |
steve | Update | SUS | MC2 damping restored |
MC2 damping restored, MZ locked and the arms are flashing now. |
2246
|
Thu Nov 12 01:18:34 2009 |
haixing | Update | SUS | open-loop transfer function of mag levi system (comparison between simulink and measurement) |
I built a Simulink model of the magnetic levitation system and try to explain the dip in the open-loop transfer function that was observed.
One can download the model in the svn. The corresponding block diagram is shown by the figure below.

Here "Magnet" is equal to inverse of the magnet mass. Integrator "1/s" gives the velocity of the magnet. A further integrator gives the displacement of the magnet.
Different from the free-mass response, the response of the magnet is modified due to the existence of the Eddy-current damping and negative spring in the vertical
direction, as indicated by the feedback loops after two integrals respectively. The motion of the magnet will change the magnetic field strength which in turn will pick
up by the Hall-effect sensor. Unlike the usual case, here the Hall sensor also picks up the magnetic field created by the coil as indicated by the loop below the mechanical
part. This is actually the origin of the dip in the open-loop transfer function. In the figure below, we show the open-loop transfer function and its phase contributed by both
the mechanical motion of the magnet and the Hall sensor with the black curve "Total". The contribution from the mechanical motion alone is shown by the magenta curve
"Mech" which is obtained by disconnecting the Hall sensor loop (I rescale the total gain to fit the measurement data due to uncertainties in those gains indicated in the figure).
The contribution from the Hall sensor alone is indicated by the blue curve "Hall" which is obtained by disconnecting the mechanical motion loop. Those two contributions
have the different sign as shown by the phase plot, and they destructively interfere with each other and create the dip in the open-loop transfer function.

In the following figure, we show the close-loop response function of the mechanical motion of the magnet.

As we can see, even though the entire close loop of the circuit is stable, the mechanical motion is unstable around 10 Hz. This simply comes from the fact that
around this frequency, the Hall sensor almost has no response to the mechanical motion due to destructive interference as mentioned.
In the future, we will replace the Hall sensor with an optical one to get rid of this undesired destructive interference.
|
2274
|
Mon Nov 16 15:18:10 2009 |
haixing | Update | SUS | Stable magnetic levitation without eddy-current damping |
By including a differentiator from 10 Hz to 50 Hz, we increase the phase margin and the resulting
magnetic levitation system is stable even without the help of eddy-current damping.
The new block diagram for the system is the following:

Here the eddy-current damping component is removed and we add an additional differential
circuit with an operational amplifier OP27G.
In addition, we place the Hall sensor below the magnet to minimize the coupling between
the coil and the Hall sensor.
The resulting levitation system is shown by the figure below:

|
2280
|
Tue Nov 17 11:09:43 2009 |
Koji | Configuration | SUS | ETMY suspension conencted to megatron ADC/DAC |
I have connected ETMY sus electronics to megatron ADC/DAC.
We continue this state until 15:00 of today. (Restored 13:00) |
2281
|
Tue Nov 17 13:39:37 2009 |
Koji | Configuration | SUS | ETMY suspension conencted to megatron ADC/DAC |
0) Now the connection for the ETMY suspension was restored in a usual state. It damps well.
1) I thought it would be nice to have dataviewer and DTT working.
So far, I could not figure out how to run daqd and tpman .
- I tried to configure
/cvs/cds/caltech/target/fb/daqdrc
/cvs/cds/caltech/target/fb/master
/cvs/cds/caltech/chans/daq/C1TST.ini (via daqconfig )
- I also looked at
/cvs/cds/caltech/targetgds/param/tpchn_C1.par
but I don't understand how it works. The entries have dcuids of 13 and 14 although C1TST has dcuid of 10.
The file is unmodified.
I will try it later when I got a help of the experts.
2) Anyway, I went ahead. I tried to excite suspension by putting some offset.
It seems to have no DAC output. I checked the timing signal. It seems that looks wrong clock.
I looked at DAC output by putting 5000,10000,15000,20000,25000cnt to UL/UR/LR/LL/SD coils.
I could not find any voltage out of the DAC in any channels.
Then, I checked the timing signal. This clock seems to have wrong frequency.
What we are using now is a clock with +/-4V@4MHz. (Differential)
Maybe 4194304Hz (=2^22Hz)?
I went to 1Y3 and checked the timing signal for 16K. This was +/-4V@16kHz. (Diffrential)
The possible solution would be
- bring a function generator at the end and try to input a single end 4V clock.
- stretch a cable from 1Y3 to 1Y9. (2pin lemo)
Quote: |
I have connected ETMY sus electronics to megatron ADC/DAC.
We continue this state until 15:00 of today.
|
|
2282
|
Tue Nov 17 15:23:06 2009 |
Koji | Configuration | SUS | ETMY suspension conencted to megatron ADC/DAC |
OK. Now, Timing/ADC/DAC are working. It's almost there.
1) As a temporaly clock, I put a function generator at the back side of the ETMY.
Set it to the rectangular +/-4V@16384Hz. Connect it to D060064 PCIX Timing Interface Board in the IO Chasis.
That is a line receiver to feed the TTL signal into ADCs/DACs.
I confirmed the actual sampling clock is supplied to the ADC/DAC boards by looking at the SMB output of the D060064.
2) Restarted the realtime code.
3) I looked at DAC output by putting 5000,10000,15000,20000,25000cnt to UL/UR/LR/LL/SD coils again.
Yes! I could see the DAC channels are putting DC voltages.
4) Then I connected DAC CH0 to ADC CH0 using SCSI breaking up boards.
Yes! I could see the coil output switching change the ADC counts!
Now, we are ready to see the suspension damped. Check it out.
|
2285
|
Tue Nov 17 21:10:30 2009 |
Koji | Configuration | SUS | ETMY suspension conencted to megatron ADC/DAC |
Koji, Rana
The megatron DAC was temporaly connected to the suspension electronics for the DAC test. We went down to ETMY as we could not excite the mirror.
The DAC is putting correct voltages to the channels. However, the anti imaging filter test output does not show any signal.
This means something wrong is there in the DAC I/F box or the cables to the AI circuit. We will check those things tomorrow.
The ETMY was restored to the usual configuration. |
2287
|
Tue Nov 17 21:21:30 2009 |
rob | Update | SUS | ETMY UL OSEM |
Had been disconnected for about two weeks. I found a partially seated 4-pin LEMO cable coming from the OSEM PD interface board. |
2290
|
Wed Nov 18 11:27:33 2009 |
Koji, josephb | Configuration | SUS | ETMY suspension conencted to megatron ADC/DAC |
Quote: |
Koji, Rana
The megatron DAC was temporaly connected to the suspension electronics for the DAC test. We went down to ETMY as we could not excite the mirror.
The DAC is putting correct voltages to the channels. However, the anti imaging filter test output does not show any signal.
This means something wrong is there in the DAC I/F box or the cables to the AI circuit. We will check those things tomorrow.
The ETMY was restored to the usual configuration.
|
It appears the front panel for the DAC board is mis-labeled. Channels 1-8 are in fact 9-16, and 9-16 are the ones labeled 1-8. We have put on new labels to reduce confusion in the future. |
2291
|
Wed Nov 18 12:33:30 2009 |
Koji, josephb | Configuration | SUS | ETMY suspension conencted to megatron ADC/DAC |
Hurraaaah!
We've got the damping of the suspension.
The Oplev loops has also worked!
The DAC channnel swapping was the last key!
DataViewer snapshot to show the damping against an artificial excitation was attached
Quote:
|
Quote: |
Koji, Rana
The megatron DAC was temporaly connected to the suspension electronics for the DAC test. We went down to ETMY as we could not excite the mirror.
The DAC is putting correct voltages to the channels. However, the anti imaging filter test output does not show any signal.
This means something wrong is there in the DAC I/F box or the cables to the AI circuit. We will check those things tomorrow.
The ETMY was restored to the usual configuration.
|
It appears the front panel for the DAC board is mis-labeled. Channels 1-8 are in fact 9-16, and 9-16 are the ones labeled 1-8. We have put on new labels to reduce confusion in the future.
|
|
Attachment 1: Untitled.png
|
|
2293
|
Wed Nov 18 16:24:25 2009 |
pete | Configuration | SUS | ETMY suspension conencted to megatron ADC/DAC |
/cvs/cds/caltech/target/fb/daqd -c daqdrc
This starts the FB.
Now the dataviewer and DTT work!
Quote: |
0) Now the connection for the ETMY suspension was restored in a usual state. It damps well.
1) I thought it would be nice to have dataviewer and DTT working.
So far, I could not figure out how to run daqd and tpman .
- I tried to configure
/cvs/cds/caltech/target/fb/daqdrc
/cvs/cds/caltech/target/fb/master
/cvs/cds/caltech/chans/daq/C1TST.ini (via daqconfig )
- I also looked at
/cvs/cds/caltech/targetgds/param/tpchn_C1.par
but I don't understand how it works. The entries have dcuids of 13 and 14 although C1TST has dcuid of 10.
The file is unmodified.
I will try it later when I got a help of the experts.
2) Anyway, I went ahead. I tried to excite suspension by putting some offset.
It seems to have no DAC output. I checked the timing signal. It seems that looks wrong clock.
I looked at DAC output by putting 5000,10000,15000,20000,25000cnt to UL/UR/LR/LL/SD coils.
I could not find any voltage out of the DAC in any channels.
Then, I checked the timing signal. This clock seems to have wrong frequency.
What we are using now is a clock with +/-4V@4MHz. (Differential)
Maybe 4194304Hz (=2^22Hz)?
I went to 1Y3 and checked the timing signal for 16K. This was +/-4V@16kHz. (Diffrential)
The possible solution would be
- bring a function generator at the end and try to input a single end 4V clock.
- stretch a cable from 1Y3 to 1Y9. (2pin lemo)
Quote: |
I have connected ETMY sus electronics to megatron ADC/DAC.
We continue this state until 15:00 of today.
|
|
|
2307
|
Fri Nov 20 11:32:48 2009 |
Haixing | Update | SUS | Magnetic levitation |
I added an integrator to increase the gain at low frequencies (below 5 Hz). In addition, I increased
the band of the differentiator. The schematics for both integrator and differentiator are the following:

The magnetic is stably levitated.

I turned off the light to get rid of 60Hz noise on the photodiode. I tried to measured the
open-loop transfer function of this setup, but somehow the SR560 is always saturate
when I injected the signal from SR785, which produces some weird results at
low-frequencies.
In addition, I found out that when the light is turned on, the levitation
can be stable even when I inverted the sign of the control loop. The control signal
on the osciloscope is the following:

This oscillator is around 120 Hz, which should be the harmonics of 60 Hz from light pollution.
I am not sure exactly why it is stable when the control-loop sign is flipped. This could
be similar to the Pauli trap in the iron trap, because the coil not only provides a force
but also provides the rigidity. The sign of such rigidity depends on the sign of the control
current. If such oscillating rigidity changes at a frequency much higher than the response
frequency of the magnet, it will stablize the system simply by significantly increasing
the inertial of the magnet.More investigations are essential to completely understand it.
For information about Pauli trap, one can look at the wikipedia:
http://en.wikipedia.org/wiki/Quadrupole_ion_trap
|
2309
|
Fri Nov 20 16:18:56 2009 |
rob | Configuration | SUS | watchdog rampdown |
I've changed the watchdog rampdown script so it brings the SUS watchdogs to 220, instead of the 150 it previously targeted. This is to make tripping less likely with the jackhammering going on next door. I've also turned off all the oplev damping. |
2313
|
Mon Nov 23 11:03:00 2009 |
steve | Update | SUS | jackhammer special well under control |
Quote: |
I've changed the watchdog rampdown script so it brings the SUS watchdogs to 220, instead of the 150 it previously targeted. This is to make tripping less likely with the jackhammering going on next door. I've also turned off all the oplev damping.
|
Saturday's special event of braking up the large concrete pieces in CES bay was un event full. |
Attachment 1: dig10d.png
|
|
Attachment 2: P1050740.JPG
|
|
2331
|
Wed Nov 25 12:28:22 2009 |
Jenne | Update | SUS | MC2 tripped |
Just felt a big "kerplunk" type ground-shaking, presumably from all the antics next door. MC2's watchdog tripped as a result. The watchdog has been reenabled. |
2363
|
Tue Dec 8 03:53:49 2009 |
kiwamu | Update | SUS | Free swinging spectra of ETMX |
In this night, I checked the free swinging spectra of ETMX to make sure nothing wrong with ETMX by the wiping.
Compared with the past (Aug.6 2008), the spectra of ETMX doesn't show significant change.
Successfully the wiping activity didn't change its configuration so much and didn't bring bad situations.
(bad situation means for example, the suspended components hit some others).
The spectra of ETMX by DTT are attached. Also you can see the past spectra in Yoichi's entry.
Yoichi's data was taken during the air-pressure condition, so it's good for comparing.
Actually I compared those data by my eyes, because I could not get the past raw data somehow.
The resonant frequencies and their typical height changed a little bit, but I think those are not significant.
NOTE: In the figure, pitch and yaw modes (~0.57Hz and ~0.58Hz) look like having a smaller Q-factor than the past.
|
Attachment 1: ETMX_air.png
|
|
2368
|
Tue Dec 8 23:13:32 2009 |
kiwamu | Update | SUS | free swinging spectra of ETMY and ITMX |
The free swinging spectra of ETMY and ITMX were taken after today's wiping, in order to check the test masses.
These data were taken under the atmospheric pressure condition, as well as the spectra of ETMX taken yesterday.
Compared with the past (see Yoichi's good summary in Aug.7 2008), there are no significant difference.
There are nothing wrong with the ETMY and ITMX successfully.
--
By the way I found a trend, which can be seen in all of the data taken today and yesterday.
The resonances of pitch and yaw around 0.5Hz look like being damped, because their height from the floor become lower than the past.
I don't know what goes on, but it is interesting because you can see the trend in all of the data.
|
Attachment 1: SUS-ETMY.png
|
|
Attachment 2: SUS-ITMX.png
|
|
2369
|
Wed Dec 9 00:23:28 2009 |
Koji | Update | SUS | free swinging spectra of ETMY and ITMX |
Where is the plot for the trend?
It can be either something very important or just a daydream of you.
We can't say anything before we see the data.
We like to see it if you think this is interesting.
... Just a naive guess: Is it just because the seismic level got quiet in the night?
P.S.
You looks consistently confused some words like damping, Q, and peak height.
- Q is defined by the transfer function of the system (= pendulum).
- Damping (either active or passive) makes the Q lower.
- The peak height of the resonance in the spectrum dy is determined by the disturbance dx and the transfer function H (=y/x).
dy = H dx
As the damping makes the Q lower, the peak height also gets lowered by the damping.
But if the disturbance gets smaller, the peak height can become small even without any change of the damping and the Q.
Quote: |
By the way I found a trend, which can be seen in all of the data taken today and yesterday.
The resonances of pitch and yaw around 0.5Hz look like being damped, because their height from the floor become lower than the past.
I don't know what goes on, but it is interesting because you can see the trend in all of the data.
|
|
2372
|
Wed Dec 9 17:51:03 2009 |
kiwamu | Update | SUS | watchdogs |
Please do not touch the watchdogs for all SUSs except for MCs,
because I am going to measure the free swinging spectra for ITMs, ETMs, BS, PRM, SRM tonight.
Today, it is good chance to summarize those data under atmospheric pressure.
thank you.
|
2374
|
Wed Dec 9 21:09:28 2009 |
kiwamu | Update | SUS | Re: free swinging spectra of ETMY and ITMX |
Okay, now the data are attached. At that time I just wanted to say like the follower.
- - -
In the free-swinging spectra around ~0.5Hz, you can see the two resonances, which come from pitch and yaw mode of the pendulum.
Note that, the vertical and the horizontal axis are adjusted to be the same for the two plots in the figure .
And I found that
* the floor levels are almost the same (the factor of about 1.5 or something like that) compared to the past.
* however the peak heights for two resonances are several 10 times smaller than the past.
* this tendency are shown in all of the data (ITMX, ETMX, ETMY).
If such variation of the peak heights is cased by the seismic activity, it means the seismic level change by several 10 times. It sounds large to me.
Quote: |
Where is the plot for the trend?
It can be either something very important or just a daydream of you.
We can't say anything before we see the data.
Quote: |
By the way I found a trend, which can be seen in all of the data taken today and yesterday.
The resonances of pitch and yaw around 0.5Hz look like being damped, because their height from the floor become lower than the past.
I don't know what goes on, but it is interesting because you can see the trend in all of the data.
|
|
|
Attachment 1: Pitch-Yaw_modes.png
|
|
2375
|
Thu Dec 10 00:46:15 2009 |
Koji | Update | SUS | Re: free swinging spectra of ETMY and ITMX |
Well, I get the point now. It could be either seismic or change in the suspension Q.
The pendulum memorizes its own state for a period of ~ Q T_pend. (T_pend is the period of the pendulum)
If the pendulum Q is very high (>104), once the pendulum is excited, the effect of the excitation can last many hours.
On the other hand, in our current case, we turned on the damping once, and then turned off the damping.
Again it takes ~Q T_pend to be excited.
In those cases, the peak height is not yet before in equilibrium, and can be higher or lower than expected.
So, my suggestion is:
Track the peak height along the long time scale (~10hrs) and compare between the previous one and the current one.
This may indicate whether it is equilibrium or not, and where the equilibrium is.
Quote: |
If such variation of the peak heights is cased by the seismic activity, it means the seismic level change by several 10 times. It sounds large to me.
|
|
2391
|
Thu Dec 10 17:13:36 2009 |
kiwamu | Update | SUS | Free swiging spectra under the atmospheric pressure |
The free swinging spectra of ITMs, ETMs, BS, PRM and SRM were measured last night in order to make sure that nothing wrong have happened by the wiping.
I think there are nothing wrong with ITMs, ETMs, BS, PRM and SRM successfully.
For the comparison, Yoichi's figure in his elog entry of Aug.7 2008 is good, but in his figure somehow PRM spectrum doesn't look correct.
Anyway, compared with his past data, there are no significant changes in the spectra. For PRM which has no counterpart to compare with, its shape of spectra looks similar to any other spectra. So I think PRM is also OK. The measured spectra are attached below.
Indeed the current data are still showing smaller peak height for their pitch and yaw modes (roughly factor of 10 ).
|
Attachment 1: summary_freeswing.pdf
|
|
2405
|
Sun Dec 13 17:43:10 2009 |
kiwamu | Update | SUS | free swinging spectra (vacuum) |
The free swinging spectra of ITMs, ETMs, BS, PRM and SRM were measured under the vacuum-condition. The attachment are measured spectra.
It looks there are nothing wrong because no significant difference appear from the past data and the current data (under atmosperic pressure).
So everything is going well. |
Attachment 1: summary_FreeSwinging_vacuum.pdf
|
|