ID |
Date |
Author |
Type |
Category |
Subject |
17449
|
Sun Feb 5 10:25:49 2023 |
Anchal | Summary | PowerShutdown | Main laser tripping |
Our main laser has tripped suspiciously twice since the power shutdown. The last time it happened on Thursday Feb 2 night (the day of power outage happened). Next morning Paco turned the laser back on, I'm not sure if he did anything else other than turning the diode current driver ON. Paco, please add anything else you did.
Chris reported that the laser tripped again last night on Feb 4th around 6 pm. I came today to see the same situation, laser diode turned OFF. After a discussion with a friend in the weekend, it turned out that sometimes when brief power outages happen, the TEC circuit for mainitaing laser tremperature turns OFF while the current driver keeps running. I'm not sure if that is the case for us but that can cause such tripping due to over heating. So today instead of simply turning on the current driver again, I power cycled the laser controller. Laser is back on and mode cleaner is locked with fewer counts though since PMC transmission dropped. I don't have time to realign PMC today and I think it might be the case that the transmission would increase once laser has reached a steady state. On Monday, we need to consult with people with more experience and understand why our laser is tripping. I hope it is not sick. |
17451
|
Mon Feb 6 07:58:00 2023 |
JC | Summary | PowerShutdown | Main laser tripping |
I came in today and it seems like the main laser tripped again yesterday around 3:00 pm.
There was a series of earthquakes in Turkey today, but all our suspension seem to be okay. |
17452
|
Mon Feb 6 11:50:44 2023 |
Anchal | Summary | PowerShutdown | Main laser tripping |
[Anchal, JC]
During shimmer test yesterday, the man laser tripped again. This morning, JC and I went to inspect the situation closer. We figured that if we can take a look inside the controller, we can get the replacement fan part number. Attached are some photos of inside. To open the controller, all you need to do is take out the two standoffs that are at the edges in the back side, then the top or botttom cover can slide out. Inside, all heatsinks of heat generating ICs are clamped to a larged metal heat sink which is covered on all side but one at the rear end. Through two holes in the top of this cavity, two fan push air through the heat sink to the back. This prompted us to understand that if the externam cooling fan direction is wrong, it would be pshing in air rather than pulling it out. So we decided to try the configuration where the fan is pulling out air. We think the direction of the fan was wrong till now which might be causing the laser controller to shut down. Now we need to wait and watch. For now, the laser is up and running.
Bt the way, we could not get any part number from the fans inside. We are still looking around in the lab if we have the replacement fans in hand as steve said they procured some.
Quote: |
I came in today and it seems like the main laser tripped again yesterday around 3:00 pm.
|
|
Attachment 1: IMG_6489.JPG
|
|
Attachment 2: IMG_6483.JPG
|
|
4503
|
Fri Apr 8 01:05:45 2011 |
Suresh | Update | RF System | RF Source Harmonics |
The measured power levels of the RF source harmonics are given below:


We are considering inclusion of bandpass filters centered on 11 and 55 MHz to suppress the harmonics and meet the requirements specified in Alberto's thesis (page 88).
|
Attachment 1: RF_Source_Harmonics_Sheet1.pdf
|
|
4511
|
Mon Apr 11 19:09:59 2011 |
Suresh | Update | RF System | Installed low pass filters in the demod boards |
As part of the RF system upgrade some of the demod boards in the lab were modfied. The filter U5 (see the circuit schematic) was replaced. These changes are tabulated below.
Filters installed in the demod boards
Serial number |
Old name of the card |
New name of the card |
Filter installed |
Remarks |
107 |
POY33 |
REFL33 |
SCLF-33+ |
R14=50Ohm |
118 |
AP133, ASDD133 |
REFL55 |
SCLF-65 |
|
114 |
PO199 |
REFL165 |
SCLF-190 |
R14=50Ohm |
120 |
PO133 |
POP110 |
SCLF-135 |
|
123 |
SP133 |
POP55 |
SCLF-65+ |
AT1 removed, R14=50Ohm |
122 |
SP199, REFLDD199 |
AS165 |
SCLF-190 |
|
121 |
SP166, REFL16 |
POP11 |
SCLF-10.7 |
|
116 |
AP199 199 MHz |
POP165 |
SCLF-190 |
|
126 |
AS166 33.3 MHz |
POX11 |
SCLF-10.7 |
|
119 |
POX 33.3 MHz |
POY11 |
SCLF-10.7 |
|
021 |
24.5 MHz (LLO) |
REFL11 |
SCLF-10.7 |
|
020 |
24.5 MHz SCLF-45 |
POP22 |
SCLF-21.4 |
|
022 |
24.5 MHz SCLF-45 |
AS11 with amp |
SCLF-10.7 |
|
029 |
24.5 SCLF-f5 |
AS55 with amp |
SCLF-65 |
|
Next, I and Q phase has to be checked for orthogonality. And noise levels of the cards have to measured.
|
4514
|
Mon Apr 11 23:35:02 2011 |
rana | Update | RF System | Installed low pass filters in the demod boards |
I am a little concerned about using these low pass filters so close to the band edge. Recall that there is no on-board preamp for the RF input to the mixer.
So, if the input impedance of the filters is not 50 Ohms, we will get some unwanted reflections and sensitivity to cable length.
I think its worth while to check the impedance or S-parameters of these things with the LO activated to find out if we need to remove them or not. |
4547
|
Wed Apr 20 21:53:01 2011 |
Suresh | Configuration | RF System | RF system: Stray heliax cable |
We found a stray unused heliax cable running from the LSC rack 1Y2 to a point between the cabinets 1X3 and 1X4. This cable will need to be redirected to the AS table in the new scheme. It is labled C1LSC-PD5 The current situation has been updated as seen in the layout below

|
Attachment 1: rogue_cable_1.png
|
|
4548
|
Wed Apr 20 22:29:07 2011 |
suresh | Update | RF System | Plan for LSC rack |
The suggested layout of the 1Y2 Rack is shown below.
To simplify the wiring, I have largely kept demod boards with the same same LO frequency close to each other.
The Heliax cables land on the top and bottom of the of subracks. These are currently flexible plastic sheets. Steve has agreed to replace them with something more rigid. It would be good to have eight N-type connectors on the top and eight at the bottom. As demod boards occur in sets of eight per subrack. So it would be convenient if the 11 and 55 Mhz Heliax cables land on the top and the rest at the bottom. In the layout I have shown the current situation.
The LO signals to the boards come from the RF Distribution box and this is kept in the middle so that cables to both the subracks can be kept short.
The outputs of the AA filter boards from both subracks have to be connected to the SCSI Interface board with a twisted pair ribbon cable.

|
4551
|
Thu Apr 21 14:39:43 2011 |
steve | Update | RF System | new strain relieved N connectors at AP |
New right angle PVC, 2 x 2 x 1/4" installed at the AP table to strain relief the 1/4" spiral corrugated RF coaxes. |
Attachment 1: P1070562.JPG
|
|
Attachment 2: P1070564.JPG
|
|
4557
|
Fri Apr 22 09:05:53 2011 |
Suresh | Update | RF System | RF Source Harmonics |
As seen in the previous measurement the first harmonic of both the 11 MHz and 55 MHz outputs are about 30dB
higher than desired. In an attempt to attenuate these and higher harmonics I introduced SBP-10.7 filters into
the 11MHz outputs and SLP-50 filters into the 55 MHz outputs.
Then I measured the height of the harmonics again and found that they were suppressed as expected. Now harmonic
at 22 MHz is 58dB lower than the 11 MHz fundamental. And the 110 MHz is lower by 55 dB compared to the 55 MHz
fundamental. None of the higher harmonics are seen => they are below 70dB
SLP-50 has an insertion loss(IL) of 4.65 dB and Return Loss(RL) of 3dB. It would be better to use SBP-60
(IL=1.4 dB and RL=23dB)
The filter on the 11 MHz lines is okay. The SBP-10.7 has IL=0.6 dB and RL=23 dB. |
4558
|
Fri Apr 22 09:25:43 2011 |
Suresh | Update | RF System | RF Source: Temperature sensor relocated |
RF Amp operating temperature
Earlier measurement reported by Alberto in LIGO-T10004-61-v1 based on the LM34 temperature sensor were lower than that shown by placing a calibrated thermocouple sensor directly on the heat sink by about 5deg C. The difference probably arose because the LM34 was located on a separate free-hanging copper sheet attached to the RF Amp by a single screw, resulting in a gradient across the copper strip. I tried to move the LM34 which was glued down, but broke the leads in the process. I then replaced it with another one mounted much closer to the heat sink and held it down with a copper-strip clamp. There is no glue involved and there is heatsink compound between the flat surface of the LM34 and the heatsink. Picture attached.
The picture also shows the new filters which have been put in place to reduce the harmonics. Note that the SBP-10.7 which was to go on the 11 MHz Demod output is located much farther upsteam due to space constraints.

|
4559
|
Fri Apr 22 10:28:22 2011 |
rana | Update | RF System | RF Source Harmonics |
You should be able to resolve the other harmonics by decreasing the IF BW or RBW on the analyzer. Even though
they're OK, its useful to have the final measurement of all of them in some kinds of physical units (like dBm, but
not dBm/Hz or dB or dBcubits). |
4578
|
Thu Apr 28 06:46:30 2011 |
Suresh | Update | RF System | RF Source installed |
RF Source box has been mounted in the 1X2 rack.

Heliax cables have been directly attached to the box and anchored on the side of the 1X2 rack. Here is a list of Helix cables which have been connected so far.
Cables |
old name |
New name |
From -> To |
1 |
133 MHz |
11 Mhz Demod |
1X2 to 1Y2 rack |
2 |
199 MHz |
55 MHz Demod |
1X2 to 1Y2 rack |
3 |
166 EOM |
11 MHz EOM |
1X2 to PSL table |
4 |
33 EOM |
55 MHz EOM |
1X2 to PSL table |
5 |
REFL 33 |
AS11 |
AS table to 1Y2 |
|
4579
|
Thu Apr 28 07:14:34 2011 |
Suresh | Update | RF System | RF Distribution box installed |
RF Distribution box has been mounted in the 1Y2 rack and is ready for use.

The box receives 11 and 55 MHz Demod Signals from the RF source located in the 1X2 rack. |
4591
|
Fri Apr 29 18:24:05 2011 |
Suresh | Update | RF System | RF system: 1X2 Rack cabling |
[Joe, Jamie, Suresh]
We have installed the IDE to SCSI adaptor module into the 1X2 rack and have connected the AA filter outputs to it.

We have removed the following cables running between the 1X2 and 1X3 racks.
The long twisted pair ribbon cable which previously carried the ADC signals.
1X2-ASC 6, 1X2-ASC 47, 1X2-ASC 9, 1X2-ASC 8, 1X2-ASC 10, 1X2-ASC 7,
CAB-1X2-LSC 42, CAB 1X2-LSC 56, CAB 1X2-LSC 41, CAB 1X2-LSC 43
1X3-2 ASC 47
We have also removed the following by mistake. We will put them back them on Monday
1X2-LSC 21, 1X2-LSC-20.
We have also removed the ASC QPD cables and moved the QPD cards which were present in the middle Eurocate (#2) to the unused Eurocrate at the bottom position (#3).
The binary input cables at the back of the cards require to be supported so that their weight does not pull them out of the sockets at the back of the crates.
Some of the slots where we plan to plug in Demod boards (the 165 MHz boards) are not currently connected to any binary output on the C1:LSC computer. We need these binary controls for the fitlter modules on the cards.
When we eventually begin to use the 15PDs as planned, then we will occupy 30 ADC channels (I & Q outputs). Currently we have just one ADC card installed on the C1:LSC providing 32 ADC channels. Joe found another 16bit 32 channel ADC card in his stash but we need to get a timing+adaptor board for it. In general we are going to need the third Eurocrate.
A platform for the power supply of the RF Distribution box needs to be built and the power supply needs to be moved into the 1X2 rack rather than sit on top of 1X2 rack.
|
4622
|
Wed May 4 12:07:48 2011 |
Suresh | Update | RF System | REFL55 installed on the AP table |
REFL55 has been installed on the AP table. REFL11 has been moved to make space for a 50% beam splitter. The reflected beam from this splitter is about 30% of the transmitted beam power. The reflected beam goes to REFL11 in the current configuration. The DC levels are 1.2V on REFL 11 and 3.5V on the REFL55.
I redid some of the cabling on the table because the we need to choose the heliax cables such that they end up close to the demod board location. As per the 1Y2 (LSC) rack layout given here, some of the PD signals have to arrive at the top and others at the bottom of the LSC rack.
Currently the PDs are connected as follows:
REFL11 PD --> Heliax (ASDD133) (arriving at the top of LSC rack) --> REFL11 Demod Board
REFL55 PD --> Heliax (REFL166) (arriving at the top of LSC rack) --> AS55 Demod Board
AS55 PD --> Heliax (AS166) (arriving at the top of the LSC rack) --> not connected.
We are waiting for the Minicircuits parts to modify the rest of the demod boards.
The heliax cables arriving at the LSC rack are not yet fixed properly. I hope to get this done with Steve's help today.
|
4628
|
Wed May 4 15:39:32 2011 |
Suresh | Update | RF System | RF Source Harmonics |
I have measured the RF source harmonics in dBm using the HP 8591E spectrum analyser. There is a small discrepancy (< 1 dBm) in the value of RF power shown by the power meter and the HP8591E. This is probably due to the loss of calibration over time.
Initial problem I faced was that when we try to measure the weak harmonics, many below -50dBm we have to choose a small band as advised by Rana. However due to the large amplitide of the fundamental typically around 15dBm or so, the preamp on the spectrum analyser becomes saturated and nonlinear. This gives rise spurious harmonics not present in the source but are rather an aritifact of measurement. The power in harmonics to avoid this I used filters to selectively attenuate the fundamental component (11 or 55 MHz) and then measure the weak harmonics.
However the filters proved difficult to use, because over their stop-band they do not have an input impedance of 50 Ohm. As a result they produce unreliable power measurements for those frequency components which are within the stop band.
To get around this problem I used a suitable attenuator so that even when the internal attenuation is decreased the preamp does not saturate
All the measurements are recorded in the attached document. Pages 4 and 5 give the reliable measurements with the attenuator.
Notes:
1) At times we can see the 29.5 MHz component reflected back from the triple resonant EOM driver.
2) In the 29.5 MHz source output there is a forest of peaks around 100 MHz, which disappear after passing through the AM stabiliser. This suggests that they are associated with AM modulation and have been removed by the stabilizer. But I did not check this further.
Quote: | You should be able tosd resolve the other harmonics by decreasing the IF BW or RBW on the analyzer. Even though
they're OK, its useful to have the final measurement of all of them in some kinds of physical units (like dBm, but
not dBm/Hz or dB or dBcubits). |
|
Attachment 1: RF_Harmonics_Sheet1.pdf
|
|
4644
|
Thu May 5 15:33:37 2011 |
steve | Configuration | RF System | LSC rack |
New right angle PVC front panel with SMA bulkhead connectors are in place. The connections are still lose. It is ready for Suresh to finalise his vision on it. |
Attachment 1: P1070641.JPG
|
|
Attachment 2: P1070639.JPG
|
|
4650
|
Fri May 6 06:36:18 2011 |
Suresh | Update | RF System | PD DC signals at each port connected |
We now have the DC signal from three PDs available in the ADC channels 14,15 and 16. The signals are from REFL55, AS55 and POY photodiodes respectively. As the DC signals on all the other PDs of the same port (REFL, AS and PO) have the same information we do not need to monitor more than one DC PD at each port.
The LSC PD Interface Card, D990543 - Rev B, can take 4 PDs and provides the DC signals of the PDs on the connector P2 (the lower of the two) on the back plane of the chassis. An adaptor card, D010005-00, plugs into the back plane from the rear of the Eurorack and provides the four DC signals on two-pin lemo sockets.
I have connected the three DC signals from the relevant RF PDs (above) to a DC whitening filter, D990694-B-1 which is associated with the channels 9 to 16 of the ADC card.
The cables are in a bit of a mess right now as some of the PD power supply lines are too short to reach up the the Interface card in the top Eurocart. Steve and I plan to redo some of the cabling later today
|
4657
|
Sat May 7 10:59:11 2011 |
Suresh | Update | RF System | RF Source filters changed |
The SLP-50 filters which were on the 55 MHz lines have been replaced with the SBP-60. Their respective characteristics are given below:
at 55MHz |
Insertion loss (dB) |
Return Loss (dB) |
SLP-50 |
4.65 |
1.5 |
SBP-60 |
1.36 |
23 |
SBP-60 has lower insertion loss and higher return loss.
This may however change the phase of I and Q in the demod boards and they will therefore need to be readjusted. Currently the output power level of 55 MHz demod is at 2dBm, whereas it ought to be at 6dBm. I have not yet corrected that. Once that is completed Kiwamu will adjust the phases.
I shifted the temperature sensor to a new location. See the photograph below. I noticed that the higher temperature is reached on the side where there are two RF Amps. So it would be better to check the temperature of that area and make sure that it remains well below 65 deg. The operating maxium is 65deg C
Here is a picture of the new RF source layout.

And here is a photograph of it

|
4670
|
Mon May 9 17:23:25 2011 |
Suresh | Update | RF System | RF Cables near LSC Rack |
[Steve, Suresh]
We started to clean up the RF cables (heliax and PD interface cables) at the LSC rack.
We have pulled out all the RF cables from the small hole on the side-board close to floor. Passing the cables through this hole makes some of the cables much too short for good strain relief. So we removed the side panel on the vacuum tube side and are going to pass the cables into the rack from there at about waist height. We now have plenty of cable lengths to tie them off to the rack at several points.
We have traced all the available Heliax cables and have attached blank tags to them. We have allocated some cables to REFL11, REFL55 and AS55. These are therefore back in working order. We have also taken stock of the available PD interface cables. They do not have consistent names on both ends of the cable and we will identify and label the ends tomorrow.
MC is locked. The auto-locker works fine.
Handing over the system for night time interferometer work now. Will continue with the cabling tomorrow.
|
4675
|
Tue May 10 01:39:41 2011 |
Suresh | Update | RF System | RF source troubles |
Today after Steve and I finished the RF cabling work for the day, Kiwamu noticed that there were no RF signals to be seen. The problem was traced to disconnected 11 and 55 MHz Demod lines from the RF source. But reconnecting them did not restore the signals. It turned out that one of the Heliax cables had a loose N-type connector at its end and it finally came off while we were tightening it into place.
We replaced the damaged heliax with another (we have two spare running from 1X2 IOO rack to the 1Y2 LSC rack. The new cable is used to be the LO 33. It seems to have a 1.5dB loss. Have to check this again tomorrow.
In the mean time I noticed that the power output of the 55MHz Demod port of the source was less than about -12dB. So I opened the source to take a look and found that all the voltage stabilisers were supplying 15V. Even those which were supposed to be supplying 24V. This was traced to a mistake in wiring the external power supply. The wires had been labeled wrongly and as a result the 18V input line was connected to 28V source and vice versa.
After fixing this problem I reassembled the source checked the power output on all the ports and found everything was functioning as expected. However after installation once again the unit failed. The blue light on the power supply was not lighting up when switched on. Suspecting a power supply problem I opened the unit again and found that a weak solder joint on one of the RF amplifiers had come loose and had overloaded one of the 24V stabilisers. We, found a spare and replaced it. The unit has been reassembled and is functioning fine. The output power levels are
11MHz Demod -- 6dBm
55MHz Demod -- 5.5 dBm
11MHz EOM -- 24dBm
55MHz EOM -- 28dBm
The Marconi is serving as the 11MHz source. The Wenzel 11MHz source is giving 13.3 dBm and is okay. But it needs to be checked for its performance as it may have been exposed to higher than rated power supply levels.
The 29.5MHz source is giving 7dBm. It is supposed to be giving 13dBm.
The Laboratory DC power supplies currently used for both the RF source and Distribution boxes need to be replaced with rack mounted Sorensen power supplies available in the lab.
|
4698
|
Thu May 12 02:31:07 2011 |
Suresh | Update | RF System | installation of RF splitters in Demod boards |
[Jamie, Koji, Suresh]
We replaced splitters in several demod boards as per the table given below:
Demod boards in which splitter has been replaced on 11May2011
Demod Board |
S. No. |
Power Splitter |
Frequency range of splitter (MHz) |
Phase unbalance from datasheet (deg) |
Amplitude Unbalance from datasheet (dB) |
REFL55 |
118 |
PQW-2-90 |
30 to 90 |
90.14 |
0.92 |
AS11 |
021 |
PSCQ-2-51W |
5 to 50 |
87.49 |
0.1 |
POY11 |
119 |
PSCQ-2-32 |
3.2 to 32 |
87.58 |
0.05 |
POY22 |
020 |
PSCQ-2-32 |
3.2 to 32 |
90.26 |
0.02 |
POY110 |
120 |
PSCQ--120 |
80 to 120 |
90.88 |
0.58 |
While doing a rough check of the boards I noticed that the REFL11 demod board had no signal on the Q output.
Rana also advised that we must use the boards which have the piggy-back amplifiers on those signals which are most useful. We referred to Alberto's thesis and chose POY55 (MICH and SRCL), REFL11(PRCL) and AS55 (DARM) as the most useful signals. We currently have these amps on AS11, REFL11 and AS55. We need to convert either AS11 or REFL11 into a POY55. Since we need to troubleshoot REFL11, I thought we might as well modify that and in the process also fix its Q output. So I renamed AS11 as REFL11 and will convert the old REFL11 into POY55 tomorrow.
Power splitter of different types have different pin-outs. The way we mount a splitter depends on which type we are using. I will detail the mounting scheme in a separate elog tomorrow.
|
4702
|
Thu May 12 10:27:02 2011 |
rana | Update | RF System | installation of RF splitters in Demod boards |
Quote: |
Rana also advised that we must use the boards which have the piggy-back amplifiers on those signals which are most useful. We referred to Alberto's thesis and chose POY55 (MICH and SRCL), REFL11(PRCL) and AS55 (DARM) as the most useful signals. We currently have these amps on AS11, REFL11 and AS55. We need to convert either AS11 or REFL11 into a POY55. Since we need to troubleshoot REFL11, I thought we might as well modify that and in the process also fix its Q output. So I renamed AS11 as REFL11 and will convert the old REFL11 into POY55 tomorrow.
|
I think we should leave them as is; the AS11 was made by taking into account the SB levels at the AS port and should not become REFL11. We should instead convert one of the old 25 or 33 MHz diodes into a POY55. |
4703
|
Thu May 12 16:50:22 2011 |
Suresh | Update | RF System | installation of RF splitters in Demod boards |
We have no plan to change the AS11 PD. I was referring to the AS11 Demod board which currently has the "Demodulator Preamp" circuit installed as a piggy back. In future I will append "_Demod" when I am referring to a demod board, to avoid confusion.
Quote: |
Quote: |
Rana also advised that we must use the boards which have the piggy-back amplifiers on those signals which are most useful. We referred to Alberto's thesis and chose POY55 (MICH and SRCL), REFL11(PRCL) and AS55 (DARM) as the most useful signals. We currently have these amps on AS11, REFL11 and AS55. We need to convert either AS11 or REFL11 into a POY55. Since we need to troubleshoot REFL11, I thought we might as well modify that and in the process also fix its Q output. So I renamed AS11 as REFL11 and will convert the old REFL11 into POY55 tomorrow.
|
I think we should leave them as is; the AS11 was made by taking into account the SB levels at the AS port and should not become REFL11. We should instead convert one of the old 25 or 33 MHz diodes into a POY55.
|
|
4708
|
Thu May 12 23:50:10 2011 |
Suresh | Update | RF System | POY55_Demod board Hardware change completed |
The Demod board with S. No. 022 (being used earlier as REFL11) has been modified. It now has SCLF-65 as its input LP filter on the PD input line and a PQW-2-90 power splitter. The unit functioning okay (I and Q signals are 90 deg apart.
The loss of Q output was traced to a possible loose solder joint and we now have both the I and Q signals after resoldering all components in the vicinity of U7 (Ref Schematic of D990511)
There is a strong oscillation around 350Hz present on I and Q signals of both REFL55_Demod and POY55_Demod. Don't know the source.
We have run out of power splitters to continue with the Demod board modification. We do not currently have an AS11_Demod board. All the others are in place and ready for the I<->Q phase angle measurement.
In summary we now have the following Demod boards in place:
[ REFL11, POY11, REFL55, AS55, POY55, POY22, POY110]_Demod
|
4711
|
Fri May 13 01:51:56 2011 |
Suresh | Update | RF System | RF Status update |
I have posted the attached RF status update and 1Y2 rack layout to the svn. |
Attachment 1: 1Y2_Rack_Layout.pdf
|
|
Attachment 2: RF_Work_Status.pdf
|
|
4714
|
Fri May 13 22:45:37 2011 |
Suresh | Update | RF System | The full set of 8 Demod boards is ready for testing |
We have Completed the hardware changes to the full set of 8 demod boards. The last one completed today is AS11. I have collected the info on all the demod boards available so far in the table below. As we measure the actual phase and amplitude unbalance we will expand this table to include new info.
The set of 8 demod boards
Demod Board |
S. No. |
Power Splitter |
Frequency range of splitter (MHz) |
Phase unbalance from datasheet (deg) |
Amplitude Unbalance from datasheet (dB) |
AS11 |
121 |
PSCQ-2-51W |
5 to 50 |
87.49 |
0.1 |
REFL11 |
021 |
PSCQ-2-51W |
5 to 50 |
87.49 |
0.1 |
POY11 |
119 |
PSCQ-2-32 |
3.2 to 32 |
87.58 |
0.05 |
AS55 |
029 |
PSCQ-2-51W |
5 to 50 |
no info |
no info |
REFL55 |
118 |
PQW-2-90 |
30 to 90 |
90.14 |
0.92 |
POY11 |
119 |
PSCQ-2-32 |
3.2 to 32 |
87.58 |
0.05 |
POY22 |
020 |
PSCQ-2-32 |
3.2 to 32 |
90.26 |
0.02 |
POY110 |
120 |
PSCQ--120 |
80 to 120 |
90.88 |
0.58 |
|
4715
|
Fri May 13 23:04:58 2011 |
Suresh | Update | RF System | DC power supply on RF distribution box has been replaced. |
[Steve, Koji, Suresh]
We shifted two Sorensen power supplies from the Auxiliary rack next to 1X2 to 1Y2. And have installed them there (pic below). The local ground reference was picked up from the racks ground reference. A shielded cable with two twisted pairs was used to make a new power cable for the RF rack. Since we are using three of the four conductors (+18,+28 and ground), one of them is not connected to anything. This situation can be improved in a future iteration when, for example, we might wish to relocate the Sorensens to a different rack.
We are still working on changing the power supply to the RF source. Will complete this early next week

|
4716
|
Sat May 14 14:12:16 2011 |
Koji | Update | RF System | DC power supply on RF distribution box has been replaced. |
Key points of the power supply installation
- We followed the grounding configuration for KEPCO except for the signal ground connection
- AC power supply has been obtained from the local power strip. This also provides chassis earthing (for safety)
- The chassis is connected to the shieldin of the DC supply cable. The other end should be isolated.
- The low voltage side of Sorensen's DC outputs are connected in order to share the same reference level.
- The ground level is provided from the cross connect. The cable is connected between the cross connect ground to the sorencen.
Unlike the KEPCO case, this cable does not have the current return, but just is to define the voltage level of those Sorensens.
- New AC&DC cables have been nicely strain-relieved.
Quote: |
[Steve, Koji, Suresh]
We shifted two Sorensen power supplies from the Auxiliary rack next to 1X2 to 1Y2. And have installed them there (pic below). The local ground reference was picked up from the racks ground reference. A shielded cable with two twisted pairs was used to make a new power cable for the RF rack. Since we are using three of the four conductors (+18,+28 and ground), one of them is not connected to anything. This situation can be improved in a future iteration when, for example, we might wish to relocate the Sorensens to a different rack.
We are still working on changing the power supply to the RF source. Will complete this early next week
|
|
Attachment 1: sorensen.png
|
|
4736
|
Wed May 18 07:13:00 2011 |
Suresh | Update | RF System | Demod board measurements |
I measured the amplitude and phase imbalances of the demod boards which have been modified. This is just a basic health check. We hope to use the script that Kiwamu is developing for a more accurate test. The script can also use these measurements as a sanity check. POP110 requires some further attention.

The RF distribution box outputs corresponding to the demod board (eg. AS55_LO --> AS55_demod) were used as LO sources. The RF signal was generated with a Marconi and held a kHz away from the LO frequency. The amplitude and phase unbalance were measured with SR785. The RF Power meter was used to check the LO power in each case.
|
Attachment 1: Demod_Board_measurements.png
|
|
4737
|
Wed May 18 07:18:15 2011 |
Suresh | Update | RF System | POY55_Demod board Hardware change completed |
The ~350 Hz noted in the elog below was traced to an RF modulation of the 11 MHz sideband. This modulation was set up in the Marconi which is currently supplying the 11 MHz local oscillator signal to the RF source. lt was used during the MC length study completed last week by Valera and Ryan. The frequency measured was 322 Hz.
As we do not require this any longer, I have switched off this modulation.
Quote: |
The Demod board with S. No. 022 (being used earlier as REFL11) has been modified. It now has SCLF-65 as its input LP filter on the PD input line and a PQW-2-90 power splitter. The unit functioning okay (I and Q signals are 90 deg apart.
The loss of Q output was traced to a possible loose solder joint and we now have both the I and Q signals after resoldering all components in the vicinity of U7 (Ref Schematic of D990511)
There is a strong oscillation around 350Hz present on I and Q signals of both REFL55_Demod and POY55_Demod. Don't know the source.
We have run out of power splitters to continue with the Demod board modification. We do not currently have an AS11_Demod board. All the others are in place and ready for the I<->Q phase angle measurement.
In summary we now have the following Demod boards in place:
[ REFL11, POY11, REFL55, AS55, POY55, POY22, POY110]_Demod
|
|
4738
|
Wed May 18 15:54:50 2011 |
Koji | Update | RF System | DC power supplies for the RF generation box in place |
[Koji, Steve]
DC power supplies for the RF generation box are now in place. They are the top two of the 6 Sorensens in the OMC short rack next to 1X2.
We made the connections as we did for the RF distribution box, the power supplies labele, and the cables strain-relieved.
The power supply is not yet connected to the actual RF generation box. This should be done by Suresh or someone with the supervision of him.
Note:
We have two +18V supply on the short OMC rack, in total. One is for the RF source, the other is for the OMC PZTs, whitening, etc.
This is to avoid unnecessary ground loop although the grounding situation of the OMC side is not known to me. |
4739
|
Wed May 18 16:52:23 2011 |
Suresh | Update | RF System | Cables for AS11 PD are in place |
[Larisa, Suresh]
All the cables needed for the AS11 PD are in place... the heliax cable runs from the AS table to the PSL rack. The LO and RF cables to demod board as well as the I and Q cables into the LSC Whitening board are connected.
The cables get rather densely packed when the LSC Whitening filter sits between the PD Interface Board and the LSC AA filter board. This makes it difficult to access the SMA connectors on the LSC whitening filter. So we shifted the LSC Whitening and AA Filter boards one slot to the right. The LSC rack looks like this just now. We have also shifted the binary cables at the back of the Eurocart by one slot so the same cables are associated with the cards.

|
4740
|
Wed May 18 17:06:39 2011 |
Suresh | Update | RF System | DC power supplies for the RF generation box in place |
I have checked the voltages on the connector. They are okay and I have plugged in the Sorensen power into the RF Source. The ground reference for the Sorensens comes from the 1X2 Rack ground reference lines on the south side of the rack.
I looked for the OMC ground reference. Could not find one on either of the the OMC half racks.
Quote: |
[Koji, Steve]
DC power supplies for the RF generation box are now in place. They are the top two of the 6 Sorensens in the OMC short rack next to 1X2.
We made the connections as we did for the RF distribution box, the power supplies labele, and the cables strain-relieved.
The power supply is not yet connected to the actual RF generation box. This should be done by Suresh or someone with the supervision of him.
Note:
We have two +18V supply on the short OMC rack, in total. One is for the RF source, the other is for the OMC PZTs, whitening, etc.
This is to avoid unnecessary ground loop although the grounding situation of the OMC side is not known to me.
|
|
4751
|
Thu May 19 19:41:17 2011 |
Suresh | Update | RF System | DAQ channel Assignments for RF PDs |
[Kiwamu, Suresh]
In trying to keep the wiring of the LSC rack as neat as possible, we came up with the following channel assignments of the RF PD signals.
PDI = PD Interface, The PD Interface D-type connectors (#1 to #12) are numbered: Top -> Bottom and Left -> Right in ascending order.
The Analog channels on the LSC Whitening Filter boards are numbered similarly, 1 to 32 in four sets.

|
4755
|
Fri May 20 05:41:22 2011 |
Suresh | Update | RF System | Demod board measurements |
The POP110 board which had the large Amplitude and Phase unbalances was examined today. It turned out that there was some stray solder which had connected the Sum port of the PSCQ-2-120 splitter to its body (ground). After I removed that the amplitude unbalance was 0.3dB however the phase was 105deg. The phase reduced to 90 deg only if the power on the splitter is around 19 dBm. So removed the AT1 (10dB attenuator) and the phase unbalance dropped to 91 deg. However this is not a sustainable solution as the ERA-5 max output is about 19.5 dBm.
As this is a side band power monitor (and not a length sensing RFPD), we can make do with a poorer phase. I will therefore replace the AT1 and adjust the residual phase with cable delay lines.
Quote: |
I measured the amplitude and phase imbalances of the demod boards which have been modified. This is just a basic health check. We hope to use the script that Kiwamu is developing for a more accurate test. The script can also use these measurements as a sanity check. POP110 requires some further attention.

The RF distribution box outputs corresponding to the demod board (eg. AS55_LO --> AS55_demod) were used as LO sources. The RF signal was generated with a Marconi and held a kHz away from the LO frequency. The amplitude and phase unbalance were measured with SR785. The RF Power meter was used to check the LO power in each case.
|
|
4804
|
Fri Jun 10 12:04:57 2011 |
Jenne | Update | RF System | Bad RF connections!! |
I am in the process of calibrating AS55's shot noise, and I noticed that the AS55 PD input to the demod board was only finger-tight. I then checked all of the other SMA connections in the set of RF PD demod boards, and found several more that were loose, including all of the REFL55 connections. This is no good!!!! RF connections need to be tightened! I went through and tightened all of the offending connections with my personal Snap-on SMA wrench. |
4957
|
Fri Jul 8 19:50:19 2011 |
Suresh | Update | RF System | LSC rack channel assignment |
[Jamie, Suresh]
We looked at the ADC channel assignments in the LSC model and wanted to make sure that the LSC rack wiring and the LSC model are in agreement with each other. So the plan is to wire the rack as shown below. I will also post this file on svn so that we can keep it updated in case there are changes.

|
5249
|
Tue Aug 16 16:59:20 2011 |
Anamaria | Update | RF System | AM in the PM |
Kiwamu, Keiko, Anamaria
Looking at the I and Q signals coming from REFL11 and REFL55 we saw large offsets, which would mean we have amplitude modulation, especially at 11MHz. We checked the PD themselves with RF spectrum analyzer, and at their frequencies we see stationary peaks (even if we look only at direct reflection from PRM). We changed the attenuation of the PSL EOM, and saw the peak go down. So first check is beam out of PSL EOM, to make sure the input beam is aligned to the crystal axis and is not giving AM modulation in adition to PM. |
5251
|
Wed Aug 17 02:48:56 2011 |
kiwamu | Update | RF System | Re: AM in the PM |
[Keiko / Suresh / Anamaria / Kiwamu]
The AM components do exist also on the beam after the EOM.
The peaks were found at 11, 29 and 55 MHz, where the PM are supposed to be imposed.
Suresh and Keiko minimized them by rotating the HWP, which is in front of the EOM.
Also Anamaria and I tried minimizing them by adjusting the EOM crystal alignment.
However everytime after we minimized the AM peaks, they grew back in a time scale of ~ 1 min.
Potentially it could be a problem of the HWP and/or EOM alignment.
Since we wanted to proceed the in-vac work anyways, we stopped investigating it and decided to postpone it for tomorrow.
We again adjusted the incident power to 20 mW.
-- P.S.
The incident power going to MC went down to 7 mW for some reasons. This was found after ~ 6 hours from our works on the PSL table.
We haven't touched anything on the PSL table since the daytime work.
Possibly the angle of the HWP is drifting (why?) and changed the amount of the P-polarizing beam power.
Suresh locked the angles of two HWPs, which are the one just after the EOM and the one after the attenuation PBS.
Quote from #5249 |
So first check is beam out of PSL EOM, to make sure the input beam is aligned to the crystal axis and is not giving AM modulation in adition to PM.
|
|
5278
|
Mon Aug 22 20:37:43 2011 |
Anamaria | Configuration | RF System | Plan for install of 3f PDs |
I made a quick sketch of how to include two more RF PDs on the REFL beam, given the space we have on the table. We want to install REFL33 and REFL165, 3f signals for the the two modulation frequencies we are using. The point is to make the distance from first beam splitter the same to all PDs so that we can use only one lens before this BS to make the beam the right size. Currently there are 2 PDs on the refl beam, REFL11 and REFL55, predictably. So the drawing shows 4 PDs. Drawing is to scale but is a bit coarse. Hopefully we'll take pictures once we're done.
Reference from current BS splitting beam to the existing PDs. |
Attachment 1: 40m3fReflPdLayout.pdf
|
|
5921
|
Thu Nov 17 11:04:02 2011 |
Jenne | Update | RF System | Stochmon? |
Is there an update on Stochmon? Are the signals acquired somewhere already? What's the current deal-io? The new EOM mount should be here later today, and I'm jazzed to start checking how my EOM box helps (hopefully) the amount of RFAM we see.
I'll start making the adapter plate while I wait... |
5923
|
Thu Nov 17 11:35:27 2011 |
Koji | Update | RF System | Stochmon? |
The Stochmon channels for 11&55MHz have been reasonably working since last night.
The output is not yet calibrated as the RF power detector has a strange scaling.
I am analyzing the calibration data. |
5926
|
Thu Nov 17 14:38:16 2011 |
Zach | Update | RF System | Stochmon? |
It turns out that we don't have all the parts I would need to do a full prototype of the precision temperature controller. I am guessing that we won't want to sit around and wait for the parts given the upcoming TAC meeting, so I'll do the next best thing:
- Standard DC temperature readout using an AD590.
- More-or-less complete heater driver
Does anyone have a suggestion for how this thing will be packaged? I.e., should it be in a box or should it be mounted in a rack, etc. In the end, a real board will be printed and stuffed, so this need not be a really professional job in the short term.
Quote: |
Is there an update on Stochmon? Are the signals acquired somewhere already? What's the current deal-io? The new EOM mount should be here later today, and I'm jazzed to start checking how my EOM box helps (hopefully) the amount of RFAM we see.
I'll start making the adapter plate while I wait...
|
|
5951
|
Fri Nov 18 19:07:07 2011 |
Jenne | Update | RF System | Foam house on EOM |
[Jenne, Zach, Frank]
Frank helped Zach and I cable up at PT-100 RTD, and make sure it worked with the Newport Model 6000 Laser Diode Controller. We're using this rather than the Newport 3040 Temperature Controller because Dmass says the output of that isn't working. So we're using just the temp control part of the Laser Diode controller.
The back of the controller has a 15-pin D-sub, with the following useful connections. All others were left Not Connected.
1 & 2 (same) - Pin 2 is one side of TEC output (we have it connected to one side of a resistive heater)
3 & 4 (same) - Pin 4 is the other side of the TEC output (connected to the other side of the resistive heater)
7 - connected to one side of PT-100 temp sensor
8 - connected to other side of PT-100 temp sensor
I used aluminum tape to attach the sensor and heater to the 40m's EOM, and we plugged in the controller. It seems to be kind of working. Zach figured out the GPIB output stuff, so we can talk to it remotely.
|
5954
|
Sat Nov 19 00:09:02 2011 |
Zach | Update | RF System | Foam house on EOM |
Quote: |
I used aluminum tape to attach the sensor and heater to the 40m's EOM, and we plugged in the controller. It seems to be kind of working. Zach figured out the GPIB output stuff, so we can talk to it remotely.
|
I stole the Prologix wireless GPIB interface from the SR785 that's down the Y-Arm temporarily. The address is 192.168.113.108. (Incidentally, I think some network settings have been changed since the GPIB stuff was initially configured. All the Prologix boxes have 131.215.X.X written on them, while they are only accessible via the 192.168.X.X addresses. Also, the 40MARS wireless router is only accessible from Martian computers at 192.168.113.226---not 131.215.113.226).
In any case, the Newport 6000 is controllable via telnet. I went through the remote RTD calibration process in the manual, by measuring the exact RTD resistance with an ohmmeter and entering it in. Despite this, when the TEC output is turned on, the heating way overshoots the entered set temperature. This is probably because the controller parameters (gain, etc.) are not set right. We have left it off for the moment.
Here are a couple command examples:
1. Turning on the TEC output
nodus:~>telnet 192.168.113.108 1234
Trying 192.168.113.108...
Connected to 192.168.113.108.
Escape character is '^]'.
TEC:OUT on
TEC:OUT
TEC:OUT?
++read eoi
1
2. Measuring the current temperature
TEC:T?
++read eoi
32.9837
3. Reading and then changing the set temperature
TEC:SET:T?
++read eoi
34.0000
TEC:T 35.0
TEC:SET:T?
++read eoi
35.0000
4. Figuring out that the temperature is unstable and then turning off the TEC (this is important)
TEC:T?
++read eoi
36.2501
TEC:OUT off
TEC:OUT?
++read eoi
0
(The "++read eoi" lines are the commands you give the Prologix to read the controlled device output.)
As I understand, Frank has some code that will pull data in realtime and put it into EPICS. This would be nice. |
5955
|
Sat Nov 19 00:34:36 2011 |
Zach | Update | RF System | why the Newport 6000 isn't working |
I just figured out why the Newport 6000 isn't stabilizing the temperature. It is designed to drive a TEC, so that when the temperature is too high, it just applies a negative current. Of course, this doesn't work with a resistive heater; it just keeps heating it up more.
I'm not sure if Frank has actually used this with a restive heater before, but it doesn't appear that you can limit the low-current level or add an offset. |
5976
|
Tue Nov 22 06:12:43 2011 |
Zach | Update | RF System | Prototype temperature controller |
Tonight I built a simpler version of what will be the new general-purpose precision temperature controller. This one is built on a breadboard and will be used for RFAM testing at the 40m until a better version is made. Some differences between this version and the final one:
- In the interest of time, this controller senses temperature using a DC wheatstone bridge, instead of the audio-frequency bridge of the final controller.
- I eschewed the more complicated transistor current source in favor of a simple current buffer. In effect, using a constant-current source is not absolutely necessary, since we are not interested in constant current but rather a constant system temperature. In this sense, it doesn't matter if we have a transistor current source or a transistor voltage source or a current-buffered op amp voltage source; the loop will simply drive the heater with the proper current to keep the error signal nulled.
So, how it works:
- The DC bridge drive voltage is supplied by a voltage-divided and buffered AD587 (low-noise 10-V reference).
- The reference resistors are just 1% metal film leaded resistors, but I have put some effort into making them quiet:
- Each resistor's body is wrapped in Al tape, and then all the resistors are taped together using Al tape, as well. This is to strongly couple them to each other thermally.
- All the reference resistors are embedded in some foam I found in the Bridge sub-B hallway. It's nothing fancy, but it keeps large advection currents from causing thermal drifts.
- The sensing element is a PT100 100-ohm RTD. Tempco is ~0.0037 1/K
- The bridge differential voltage is read out by an AD620 instrumentation amplifier with G = 100
- The AD620 output is fed directly to an OP27 with G = 0-20
- This is fed to an LF356 (FET-input op amp, to reduce the effects of bias current when the integrator is on) with a single pole at 0.1 Hz, switchable via jumper to DC for true integration
- This is summed with an offset via an OP27 summer (the offset determines the heater current with no signal---half the maximum current of ~120 mA is optimal)
- The summer output is buffered with a BUF634, which can provide well over the maximum current we can push through our heater, and the BUF634 directly drives the heater
- Between the BUF634 and the heater is a back-biased diode to ground. This is to prevent the current from going negative when the error signal is well below zero.
I have tested the circuit using a spare resistive heater and a potentiometer to simulate the RTD. First I tested the sensing and drive circuits separately, then I connected the sensor output to the drive input and modulated the potentiometer resistance while monitoring the current. The circuit behaved as expected.
When I got to the 40m, it struck me that the resistance I had chosen (115 ohms) corresponded to 40 C, which I realized might be above what we could reach with the current we can provide. I used the Newport 6000 via telnet to drive the heater at several current values and see what the resistance became. I found that with I = Imax/2 ~ 0.6, the resistance was around 113 ohms (it was ~111 at room temp). So, I switched the reference resistor in the leg above the PT100 from 115 -> 113.
I then plugged everything in while monitoring the heater current and AD620 output (error signal), and it seemed not to do anything. I was tired so I figured I'd leave it for tomorrow.
Here is a sketch of the schematic, as well as some pictures:
  

|
5984
|
Wed Nov 23 00:30:14 2011 |
Zach | Update | RF System | EOM temperature controller trials |
[Jenne, Zach]
We did some testing of the prototype temperature controller. When I left it late last night, it was not working in conjunction with the real heater and PT100 mounted to the EOM, but had been tested using simulated loads (a spare heater and a potentiometer for the RTD).
We measured each of the reference resistors carefully, as I should have done in the first place since they are only 1% tolerance (I am using 100-ohm ones in series with ~15-ohm ones, so they have a variation of +/- an ohm or so, which is consequential). We calculated the estimated zero-signal resistance of the RTD, then used a trimpot to verify that the AD620 output behaved as expected. We realized that I didn't tie the 620's reference to ground, so the output was floating around by a lot. Once we did that, the readout was still not working properly, but eventually magic happened and we got an appropriate signal. I did find that there was a discrepancy between the estimated zero-signal resistance and that measured across the trimpot with the readout nulled---this may be caused by a small offset in the 620, but is not important so long as the output still scales properly.
Before trying it out again on the real McCoy, I tested the whole, closed-loop circuit with the spare EOM on Jenne's desk. The temperature oscillated at first, but a reduction of gain at the input stage of the driver allowed it to stabilize. The temperature of the EOM (sitting on the electronics bench) was kept constant with a control current that varied from ~40 - 70 mA, depending on how many people were around it, etc. This is pretty much perfect for the quiescent level, but that means that we might have to increase the baseline operating resistance of the PT100 (by changing the reference resistors) once it is sitting in a hot foam box. Otherwise, we will have no gain on the cooling side. I tested the circuit response by cupping my hands over the EOM to increase the temperature and ensuring that the current dropped so as to null the error signal. It worked pretty well, with a thermally-limited bandwidth of I would estimate around 0.1 Hz.
I went to try it out on the PSL table, but again it didn't work. It turned out that this time I had broken one of the soldered connections from the broken-out D-sub cable to the (tiny) wires going to the PT100, so there was no temperature signal. I resoldered it, but I forgot that there is a thin insulating layer on the wire, so no connection was made. Frank tutored Jenne on how to properly strip these wires without damaging the core, but alas I didn't pay attention.
The RTD/heater/D-sub package lies in wait on Jenne's desk, where I have left an apologetic note. Once it is fixed, we should be able to finally hook it up for realz. |