40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 277 of 344  Not logged in ELOG logo
ID Date Author Type Categoryup Subject Text Attachments
  15179   Thu Jan 30 17:41:10 2020 gautamUpdatePSLErrant FSS_INOFFSET changeYou can trend the data for the past few
hours and see what the appropriate value.
I think
  
  15184   Mon Feb 3 15:22:39 2020 JonUpdatePSLc1psl progress/Acromag ADC groundingI tested the c1psl AO channels on the electronics
bench on Friday. While I found all the wiring
to be correct, some of the channels exhibited
  
  15186   Tue Feb 4 18:13:01 2020 YehonathanUpdatePSLBench testing of PSL ai channels{Yehonathan, Jon, Jordan}

I tested the ai channels of the
new PSL Acromag by looping an already-tested
 c1psl_wire_testing_-_By_Connector.pdf 
  15187   Wed Feb 5 08:57:11 2020 YehonathanUpdatePSLBench testing of PSL ai channelsI checked the failed channels against
the EPICS database definitions and the yaml
file inputted to IFOTest. The channels where
  
  15189   Wed Feb 5 21:04:10 2020 YehonathanUpdatePSLBench testing of PSL ai channels{Yehonathan, Jon}

We retested the failed ai channels.
Most of them got fixed by applying the inverse
  
  15194   Thu Feb 6 21:54:13 2020 JonUpdatePSLc1psl bench testing completeToday I engineered the last piece
of the new c1psl system: the multi-bit
binary output (mbbo) channels that control
  
  15202   Mon Feb 10 10:07:20 2020 gautamUpdatePSLPMC re-lockedI found the PMC unlocked this morning.
It was re-locked using the usual procedure.
I feel like this has been happening more
 PMCunlock.png 
  15204   Mon Feb 10 15:54:47 2020 JordanUpdatePSLCompleted Acromag WiringAll spare channels on the PSL acromag chassis
are connected with ~12in of spare wiring
for future use.
  
  15205   Mon Feb 10 15:55:46 2020 JordanUpdatePSLPMCTRANSPD[Gautam, Jordan]

Gautam showed me how the PMCTRANSPD
signal was reading zero, and he suspected
 ADC1.jpg 
  15231   Thu Feb 27 17:50:36 2020 gautamUpdatePSLc1psl setup setup[many people]

in prep for the install tomorrow,
we did the following:
 20200227_173535.jpg20200227_173454_HDR.jpg20200227_172659.jpg 
  15234   Fri Feb 28 08:05:22 2020 gautamUpdatePSLc1psl setup setupAnd so it begins.

  
  15235   Fri Feb 28 10:04:41 2020 gautamUpdatePSLc1psl setup setupSummary:

There are several problems evident
already.
  
  15236   Fri Feb 28 19:37:18 2020 gautamUpdatePSLNew c1psl installed
The new c1psl Acromag crate is
now interfaced to the Eurocrate electronics
in 1X1 (formerly VME c1psl) and 1X2 (formerly
  
  15247   Wed Mar 4 11:16:37 2020 gautamUpdatePSLPMC realignmentI realigned the input pointing into the
PMC this morning. Usually, the way I do this
is to minimize any discernible mode structure
 PMCrealignment.png 
  15253   Wed Mar 4 22:38:31 2020 JonUpdatePSLc1psl communications problem resolvedI investigated the problem
reported earlier today with the BIO1 channels.
By logging the systemd messages generated
  
  15256   Thu Mar 5 19:45:23 2020 JonSummaryPSLC1PSL in-situ test resultsWe've completed
almost all of the in-situ testing of the
c1psl channels. During this
 c1psl_feedthrough_wiring_-_By_Connector_(3).pdf 
  15266   Wed Mar 11 18:12:53 2020 gautamSummaryPSLWFS Demod board modifications[koji, gautam]

 D980233-B_Mar2020Mods.pdf 
  15272   Thu Mar 12 16:13:16 2020 gautamUpdatePSLTemperature sensors connected to Acromag[jordan,
gautam]
 P_20200317_153736_vHDR_On_2.jpg 
  15275   Fri Mar 13 14:28:39 2020 gautamUpdatePSLAdded tees to PMC Trans and REFLI want to monitor the PMC TRANS and
REFL levels on the PSL table - previously
there were some cables going to the oscilloscope
  
  15310   Wed Apr 22 17:29:14 2020 gautamUpdatePSLFSS debugging attemptsSummary:

On Monday, I hooked up an AG4395
to the PMC error point (using the active
 FSSspec.pdf 
  15312   Thu Apr 23 10:42:02 2020 ranaUpdatePSLFSS debugging attemptsI had set up the 4395 to do this automatically
a few years ago, but it looked
at the FSS/IMC instead. When the PCDRIVE
  
  15570   Tue Sep 15 10:57:30 2020 gautamUpdatePSLPMC re-locked, RGA re-enabledThe PMC has been unlocked since September
11 sometime (summary pages are flaky again).
I re-locked it just now. I didn't mess
 RGAscan.pdf 
  15592   Mon Sep 21 16:40:52 2020 gautamUpdatePSLHEPAs are no longer runningThe HEPA filters on the PSL enclosure are
no longer running. I tried cycling the power
switch on the NW corner of the enclosure,
  
  15662   Fri Nov 6 14:08:44 2020 gautamUpdatePSLPMC re-lockedThe PMC servo railed and so I re-locked
it at ~half range. I've been noticing
that the diurnal drift of the PZT control
  
  16016   Mon Apr 12 08:32:54 2021 Anchal, PacoSummaryPSLPMC unlocked at 2pm on Sunday; ~ RestoredPMC lost lock between 21:00 and 22:00 UTC
on April 11th as seen in the summary pages:

https://nodus.ligo.caltech.edu:30889/detcharsummary/day/20210411/psl/#gallery-4
  
  16019   Mon Apr 12 18:34:26 2021 YehonathanSummaryPSLPMC unlocked at 2pm on Sunday; ~ RestoredPMC lost lock again at around 16:00 April
12. I was able to lock it again but the transmission
is only 0.6 now and REFL is 0.14.
  
  16023   Tue Apr 13 19:24:45 2021 gautamUpdatePSLHigh power operationsWe (rana, yehonathan and i) briefly talked
about having high power going into the IFO.
I worked on some calcs a couple of years
 intensityDist.pdfclippingLoss.pdf 
  16024   Tue Apr 13 20:45:16 2021 YehonathanUpdatePSLLaser amplifier{Yehonathan, Rana}

We unpacked the content of the
amplifier crate in front of the water fountain
 20210413_204721.jpg20210413_203300.jpg20210413_204940.jpg20210413_205549.jpg 
  16032   Wed Apr 14 19:48:18 2021 gautamUpdatePSLLaser amplifierA couple of years ago, I got some info
about the amplifier setup at the sites from
Terra - sharing here in case there is some
  
  16034   Thu Apr 15 09:46:24 2021 YehonathanUpdatePSLLaser amplifierSome more relevant documents provided by
Matt:

Phase
  
  16044   Fri Apr 16 18:21:36 2021 YehonathanUpdatePSLLaser amplifierI surveyed a bit the 1X1/2 area to plan
for the installation of the laser amplifier.

There is a vacancy at the bottom
 20210416_143642.jpg20210416_145408.jpg20210416_145448.jpg20210416_181324.jpg 
  16046   Sun Apr 18 21:29:55 2021 ranaUpdatePSLLaser amplifier
Ideally, we put the chiller outside
of the interferometer area. The PSL chiller
used to be in the control room near the door
  
  16056   Wed Apr 21 00:08:15 2021 KojiUpdatePSLPSL Table (sort of) covered / HEPA "chimney"Shutdown Procedure:
PSL Shutter closed / MC Autolocker
disabled / PSL mechanical shutter closed
 20210420235324_IMG_0560.jpeg20210420235304_IMG_0559.jpeg20210420235243_IMG_0558.jpeg20210420235344_IMG_0562.jpeg 
  16057   Wed Apr 21 01:14:03 2021 KojiUpdatePSLPSL Table (sort of) covered / HEPA "chimney"I also located the (possible) HEPA filters
in the lab. (Attachments 1~3)

Oh! This is NO-NO! We can't
 P_20210421_005056.jpgP_20210421_005114.jpgP_20210421_005302.jpg 
  16062   Wed Apr 21 11:09:57 2021 yehonathanUpdatePSLLaser amplifierI went to the TCS lab to take a look at
the chillers lying around. I spotted two
chillers:
 20210420_171606.jpg20210420_171621.jpg20210420_171611.jpg20210420_171629.jpg20210420_171702.jpg 
  16068   Wed Apr 21 19:28:03 2021 AnchalUpdatePSLPSL/IFO recovery[Anchal, Koji]

Removed the top sheet


Opened
first from the door side so that any dust
  
  16069   Wed Apr 21 19:43:20 2021 KojiUpdatePSLNew HEPA speed controlThe new HEPA speed controllers are attached
at the middle of the HEPA unit (not at the
edge of the unit)... (Attachment
 P_20210421_193637.jpgP_20210421_193627.jpg 
  16074   Thu Apr 22 14:41:55 2021 ChubUpdatePSLNew HEPA speed controlWhen adjusting the blower speed, give the
blower at least 30 seconds to speed up or
slow down to the set speed.  The flywheel
 40M_PSL_HEPA.jpg 
  16076   Thu Apr 22 15:15:26 2021 gautamUpdatePSLPMC transmissionI was a bit surprised by these numbers
suggesting the PMC transmission is only 50-60%.
I went to the table today and confirmed that
  
  16077   Thu Apr 22 15:34:54 2021 AnchalUpdatePSLPMC transmissionKoji mentioned that the mode of the laser
is different for lower diode currents. So
that might be the reason why we got less
  
  16080   Thu Apr 22 17:28:34 2021 YehonathanUpdatePSLLaser amplifierAccording to the aLIGO
70W amplifier interlock concept the flow
rate of the chiller should be between 5 and
 20210422_124940.jpgDrillRoomSchematics.pdf20210422_125240_1.png 
  16082   Fri Apr 23 18:00:02 2021 gautamUpdatePSLHEPA speed loweredI will upload some plots later - but in
summary, I set the HEPA speed to ~40%. I
used (i)IMC transmission RIN, (ii) Arm cavity
 HEPAdiag.pdf 
  16083   Fri Apr 23 19:26:58 2021 KojiUpdatePSLHEPA speed loweredI believe that there is an internal setting
for the minimum flow, so the flow is not
linear ("0%" is not zero), but
  
  16141   Fri May 14 17:45:05 2021 ranaUpdatePSLHEPA speed raisedThe PSL was too hot, so I turned on the
south HEPA on the PSL. The north one was
on and the south one was off (or so slow
  
  16142   Sat May 15 12:39:54 2021 gautamUpdatePSLNPRO tripped/switched offThe NPRO has been off since ~1AM this morning
it looks like. 
 NPRO.png 
  16144   Tue May 18 00:52:38 2021 ranaUpdatePSLHEPA speed raisedLooks like the fan lowered the temperature
as expected. Need to get a few more days
of data to see if its stabilized, or if that's
 Untitled.png 
  16145   Tue May 18 20:26:11 2021 ranaUpdatePSLHEPA speed raisedFluke. Temp fluctuations
are as usual, but the overall temperature
is still lower. We ought to put some temperature
  
  16400   Thu Oct 14 09:28:46 2021 YehonathanUpdatePSLPMC unlockedPMC has been unlocked since ~ 2:30 AM.
Seems like the PZT got saturated. I moved
the DC output adjuster and the PMC locked
  
  16401   Thu Oct 14 11:25:49 2021 YehonathanUpdatePSLPMC unlocked{Yehonathan, Anchal}

I went to get a sandwich around
10:20 AM and when I came back BS was moving
 BS_OSEM_Sensor_PSD.pdfBS_OSEM_Sensor_PSD_AfterReconnectingCables.pdf 
  16886   Thu Jun 2 20:05:37 2022 yutaConfigurationPSLIMC input power recovered to 1W, some alignment works[Paco, Yuta]

We have
 Before.JPGAfter.JPG 
ELOG V3.1.3-