  12901   Thu Mar 23 01:44:53 2017 gautamUpdateIMCWFS sensing matrix measurementsThanks to Koji's nice MATLAB script
using DttData functions, I was able to quickly
analyze the TF data. Essentially, this measurement
 IMC_WFS_segment_TF.pdfIMC_WFS_channels_TF.pdfTFsummary.pdfIMC_WFS_170322.xlsx.zip 
  12902   Thu Mar 23 08:43:11 2017 ranaUpdateIMCWFS sensing matrix measurementsFor sensing matrix, better to use single
frequency sine response. We don't want
to measure around the bounce or above the
  
  12903   Thu Mar 23 23:38:58 2017 gautamUpdateIMCMC SUS damping gains stepped downI've reduced the gains of the damping
on all 3 MC SUS by a factor of 3 for overnight
observation as part of the ongoing feedforward
  
  12904   Fri Mar 24 11:26:57 2017 gautamUpdateIMCMC SUS damping gains restoredI've restored the damping loop gains
to their nominal values. Analysis of the
coherence between MCL and seismometer channels
  
  12906   Fri Mar 24 19:04:18 2017 gautamUpdateIMCSeismic feedforward and WFS[valera, gautam]

On Wednesday at the meeting, we
were discussing why we aren't able to
 MCL_comparison.pdfseis_comparison.pdf 
  12908   Mon Mar 27 15:36:27 2017 SteveUpdatePEMparticle counter inside of PSL enclousure 




 HEPA_flow_effects.png 
  12909   Mon Mar 27 16:01:55 2017 SteveUpdatePEMX arm AC set to 68F 

The X arm air conditioner was not
regulating properly. The arm temp was warmer
  
  12911   Mon Mar 27 20:41:21 2017 rana, gautamUpdatePSLPMC DAQ assay for feed-forward integrationWe are thinking to use the PMC signals
to help us in figuring out the feedback /
feedforward stuff and making it better.
 TEK00000.PNG20170327_194931.jpg20170327_204554.jpg 
  12913   Tue Mar 28 16:47:40 2017 SteveUpdatesafetyProjector bulb is out againThree replacement
bulbs ordered

Rana can discribe
  
  12915   Wed Mar 29 09:24:28 2017 SteveUpdateloresummery pagesThe summery pages are working at a slow
motion speed. It's response time 12 minutes.
  
  12916   Wed Mar 29 11:41:19 2017 gautamUpdatePSLPMC DAQ assay for feed-forward integrationThe C1IOO frontend machine that resides
in 1X1/1X2 has 2 ADCs, ADC0 and ADC1. The
latter has 28 out of 32 channels unused at
 C1ALS_PSLnamespace.pngPMCcheckout.pdfD980352-A-40m_151119.pdf 
  12918   Thu Mar 30 00:16:09 2017 gautamUpdatePSLPSL NPRO PZT calibrationAs part of the ongoing effort to try and
calibrate the PMC DAQ channels into physical
units, I tried to get a calibration for the
  
  12920   Thu Mar 30 18:11:01 2017 ranaUpdatePSLPMC DAQ assay for feed-forward integrationWhat you have drawn looks good to me: the
cut should be between TP3 and pin3 of the
AD620. This should maintain the DC coupled
  
  12921   Fri Mar 31 10:16:07 2017 SteveUpdatesafetylaser safety glasses annual inspectionLaser safety glasses cleaned in "
Dawn Ultra " mild soap - water solution
and measured for 1064 nm transmission at
 march2017.jpg2017march.jpg 
  12922   Fri Mar 31 16:10:33 2017 SteveUpdateTreasureLes Guthman Les Guthman interviews  gradstudent
Graig.

Main laser emergency shut off was
  
  12923   Sun Apr 2 23:14:30 2017 ranaUpdateComputer Scripts / Programsnodus update/upgrade/rebootI just did remote apt-get update, apt-get
upgrade, and then reboot on nodus. ELOG started
up by itself.
  
  12924   Mon Apr 3 17:09:47 2017 gautamUpdateCDSC1PSL burt-restoredWhen I came in this morning, Steve had
re-locked the PMC and IMC - but I could see
a ~1Hz intensity fluctuation on the PMC REFL video
  
  12925   Mon Apr 3 17:25:13 2017 gautamUpdatePSLPSL NPRO PZT calibrationSummary:

By sweeping the laser frequency
and looking at the PMC PDH error signal,
 PDHerr.pdfNPROcalib.pdf 
  12926   Mon Apr 3 23:07:09 2017 gautamUpdatePSLPMC DAQ assay for feed-forward integrationI made some changes to the DAQ path on
the PMC servo board, as per the plan posted
earlier in this thread. Summary of changes:
  
  12927   Tue Apr 4 11:24:21 2017 SteveUpdatesafetyProjector bulb is out againShipped out for repair.




  
  12928   Tue Apr 4 17:27:58 2017 ranaUpdatePSLPSL NPRO PZT calibrationgood cal. I wonder if this data also gives
us a good measurement of the cavity pole
or if the photo-thermal self-locking effect
  
  12929   Wed Apr 5 16:05:47 2017 gautamUpdateGeneralNB code checkout[evan, gautam]

We spent some time trying to get
the noise-budgeting code running today. I
  
  12930   Thu Apr 6 15:35:44 2017 SteveUpdatePEM envioirmental noiseBuilding:        
Central Engineering Services

     
  
  12931   Fri Apr 7 13:46:23 2017 SteveUpdateSUSETMX enclosure feedthoughETMX enclosure feedtrouh cabeling corrected.  bad.jpggood.jpg 
  12932   Mon Apr 10 09:49:32 2017 SteveUpdateOptical Leversoplev laser RIN test planningWe are planning to test 3 identical 1103Ps
RIN with  continous temp monitoring
and control later.
  
  12934   Mon Apr 10 14:21:57 2017 ranaUpdateOptical Leversoplev laser RIN test planningI'm suspicious of this temperature
sensor comparison. Usually, what they mean
by accuracy is not the same as what we mean.
  
  12936   Mon Apr 10 15:37:11 2017 gautamUpdateCOCRC folding mirrors - v3 of specs uploaded Koji and I have been going over these calculations
again before we send a list of revised requirements
to Ramin. I've uploaded v3 of the specs
 8x 
  12937   Mon Apr 10 16:00:26 2017 rebeccaUpdateCamerasPylon installation warningWhen trying to install an older version
of Pylon packaged for Debian that Joe B.
had sent, it gave the warning that the package
  
  12938   Mon Apr 10 18:42:09 2017 johannesUpdateCamerasPylon installation warningIt looks like we may not need to use this
old Pylon version after all. Gautam and I
looked into installing SnapPy with the makefile
  
  12939   Tue Apr 11 00:38:37 2017 gautamUpdatePSLPMC demod moved off servo boardAs discussed at the Wednesday meeting last
week, I tried moving the demodulation of
the PMC error signal off the PMC servo board,
 PMC_Ctrl_spec.pdf 
  12940   Wed Apr 12 00:36:53 2017 gautamUpdatePSLPMC demod moved off servo boardHere is a more detailed comparison of the
spectra of the signals at the front panel
DAQ LEMO output, measured with the Agilent
 PMC_digitalSpec.pdfPMC_DAQ_spectra.pdf 
  12941   Thu Apr 13 09:48:37 2017 SteveUpdateSUSITMY-UL and ETMX sensorsWhy ITMY UL can not see this earth quake?
SRM and PRM are misaligned. ETMX is still
not well.
 ITMY.pngITMY-UL.pngETMX?.png 
  12942   Thu Apr 13 19:54:07 2017 ranaUpdateDAQcheckup on minute trendsOur minute trends are still not available
through NDS2 from the outside world due to
the bad config of the DAQ, but I can confirm
 BRLMS-trend.png 
  12944   Tue Apr 18 01:01:03 2017 gautamUpdatePSLPMC OLTF measured, DAQ channels calibratedQuick entry, details to follow in the AM
tomorrow.


I calibrated the PMC DAQ
 PMCspectra_calibrated.pdfPMC_MEDMs.png 
  12945   Tue Apr 18 16:10:00 2017 gautamUpdatePSLPMC OLTF measured, DAQ channels calibratedHere are the details:


PMC OLTF:


the
procedure used was identical to what Koji
 PMC_OLTF_170418.pdf 
  12946   Tue Apr 18 23:37:15 2017 ranaUpdatePSLPMC OLTF measured, DAQ channels calibratedWhat's
the reasoning behind setting the the gain
  
  12947   Wed Apr 19 15:13:30 2017 gautamUpdatePSLPMC/MCL multicoherenceI used a one hour stretch of data from
last night to look at coherence between the
PMC control signal and MCL, to see if the
 PMC_MCL_multicoherence.pdf 
  12948   Wed Apr 19 15:46:24 2017 gautamUpdateGeneral1611/1811 inventory checkI looked through the lab area to do a fast
photodiode inventory check, as we may need
to buy some for the higher order mode spectroscopy
  
  12950   Tue Apr 25 19:35:41 2017 gautamUpdateGeneralIPCS -qDataviewer wouldn't launch on pianosa
- it seemed to work fine on Donatella though.
Rana suggested using the ipcs -q command.
  
  12951   Wed Apr 26 01:00:23 2017 gautamUpdateGeneralDRMI locking Since we'd like to get back to DRSE
locking, I tried locking the DRMI tonight.
I did the following:
 DRMI_1f_26April2017.pngIMG_7373.JPGIMG_7372.JPG 
  12952   Thu Apr 27 16:41:13 2017 Eric GustafsonUpdateLSC Status of the 40 m PD Frequency Response Fiber SystemThere two reports in the DCC describing
the state of the system as of October 2014
including: (1)  Alex Cole’s “T1300618
  
  12953   Thu Apr 27 17:55:33 2017 SteveUpdateCameraswhich camera to use for IR scatterring picturesYesterday I failed to take good pictures
of ETMY  resonant arm of 1064 nm with
Cannon Rebel T3i in RAW 22-27Mp & JPG
 P4270081RAWolym.jpgImg0344.jpg 
  12954   Fri Apr 28 02:04:36 2017 gautamUpdateGeneralDRMI locking I got a couple of ~30min long DRMI lock
stretches today. The settings I used are
essentially the same as what I had back in
 20170428_DRMI.pdf 
  12955   Fri Apr 28 13:56:26 2017 ranaUpdateGeneralDRMI locking one of these signals does not look like
the others: explanation?
  
  12956   Fri Apr 28 18:01:56 2017 rebeccaUpdateCamerasAttempting to Load Camera ClientUsing /ligo/apps/linux-x86-64/camera/bin/camera_client.py
-c  /opt/rtcds/caltech/c1/scripts/GigE/SnapPy/L1-CAM-MC1.ini,
the Python script was able to run without
  
  12957   Fri Apr 28 19:32:06 2017 gautamUpdateGeneralDRMI locking - PRCL angular FFI took a closer look at the POP QPD/ PRC
angular feedforward situation yesterday.
I thought it would be useful to have a 
 PRC_angularFF.pdfPRC_TFs.pdf 
  12958   Fri Apr 28 22:50:35 2017 johannesUpdateCamerasAttempting to Load Camera ClientYou'll likely have to run camera_server.py
using the same ini file first before you
can use the client. Since the pylon installation
  
  12959   Sun Apr 30 13:24:00 2017 ranaUpdateCamerasAttempting to Load Camera ClientWe ought to put the camera software on
the shared disk; I don't think there's
any speed reasons that it needs to be local.
  
  12960   Mon May 1 16:29:51 2017 gautamUpdateGeneralDRMI locking For the traces I posted, I had not turned
on the whitening for the SRCL sensing PD
(REFL55). However, I took a spectrum on a
 LSC_sensingNoise.pdfIMG_7378.JPG 
  12961   Mon May 1 17:14:58 2017 SteveUpdateCamerasETMY & MC2 ccd cameras removedMC2 ccd camera is replaced by Olympus 570
zoom temporarly.

So as the ETMY ccd camera is replaced
  