ID Date Author Type Categoryup Subject Text Attachments
  967   Thu Sep 18 23:31:26 2008 ranaUpdatePSLISS: Saturating too often at nominal gainThe ISS has been saturating whenever the MC
relocks and puts the gain up to +8dB. I have
lowered the gain to +1 dB for now to stop
 FIRE_BLOWER.jpg 
  971   Fri Sep 19 08:09:55 2008 steveUpdatePSLpsl HEPAs turned onI have just turned on the PSL HEPA filters
at 60% operational speed.
  
  978   Mon Sep 22 18:54:54 2008 JenneUpdatePSLPMC transfer functions with various brick-on-top configurationsAttached below is a graphical summary of different
things that I have tried putting on the PMC
to reduce the noise in the loop.  The motivation
 PMC_OLG_100Hz_to_100kHz.pngPMC_OLG_5kHz_to_30kHz.png 
  980   Mon Sep 22 21:30:06 2008 ranaConfigurationPSLbad FSSThe MC refl power was going up and the FSS
PC drive was so large that I had to turn
up the FSS
 DAQ.png 
  982   Mon Sep 22 22:24:19 2008 ranaConfigurationPSLbad FSS[quote]The MC refl power was going up and
the FSS PC drive was so large that I had
to turn up the FSS...[/quote]
  
  984   Tue Sep 23 11:17:59 2008 steveUpdatePSLPMC scattering spotThe PMC output side has a new madly scattering
spot at chamfer 2 o'clock position
 rainbow.pngpmcclip.png 
  986   Tue Sep 23 15:28:06 2008 peteConfigurationPSLnew 21.5 MHz FSS reference installedThe new 21.5 MHz FSS reference is now installed
in the rack with the 7 Sorensen PS.  Both
outputs give 18.7 dBm.  The MC seems happy.
 fss_ref_001.jpgfss_ref_013.jpg 
  989   Thu Sep 25 02:35:21 2008 ranaSummaryPSLFAST is moving alotIt looks like the FAST signal has started
moving a lot - this is partly what inspired
us to tune the SLOW loop.
 Untitled.pngUntitled.png 
  991   Thu Sep 25 10:48:29 2008 YoichiUpdatePSLFSS calibration againI did a calibration of the FSS error signal
again with a different method.
This time, I swept the laser frequency with
 PDTrans.pngPDHsignal.png 
  995   Fri Sep 26 00:19:54 2008 JenneUpdatePSLFilter-action with the PMCWritten, but not posted on 24Sept2008:

 
  
  1000   Fri Sep 26 18:35:17 2008 JenneUpdatePSLPMC filter is out for tuningThe PMC's new Pomona Box filter is out for
tuning.  I'd like to get the notch right
on the 18.3kHz, rather than being off in
  
  1001   Fri Sep 26 19:08:43 2008 ranaConfigurationPSLRefcav Trans: PD + Camera + DumpsI went out to improve the Refcav trans path.

I removed all ND filters to get rid of the
  
  1003   Mon Sep 29 01:19:40 2008 ranaSummaryPSLLaser chiller running a little hotI looked at it some last night and my suspicion
was the ISS. Whenever the ISS switch came
on the FAST got a kick.
  
  1005   Mon Sep 29 13:23:40 2008 robSummaryPSLLaser chiller running a little hot[quote]I looked at it some last night and
my suspicion was the ISS. Whenever the ISS
switch came on the FAST got a kick.
  
  1007   Mon Sep 29 15:09:36 2008 steveUpdatePSLalmost 4 yrs plot of power & tempsThe water chiller is normally running 1.5
C warmer than the laser head temp.
When control room temp is stable and PEM-count_temp
 3.5y.jpg 
  1008   Mon Sep 29 17:53:33 2008 YoichiUpdatePSLISS updateISS has been saturating easily.
Today I opened the PSL enclosure to inspect
the ISS box. Then I found that the sensor
 IMG_1671.JPG 
  1010   Tue Sep 30 19:50:27 2008 JenneUpdatePSLQuicky Summary - more details laterQuicky summary for now, more details later
tonight / tomorrow morning:
 mcf.png 
  1013   Wed Oct 1 02:47:53 2008 ranaUpdatePSLPSL ERR & LODET: Too much offsetLooks like there is an anomolous mixer offset
correlated with the increase in the LO level.
This may be leading to crazy offset locking
 Untitled.png 
  1017   Wed Oct 1 23:05:14 2008 YoichiUpdatePSLISS RIN spectraStefan, Yoichi

We took relative intensity noise (RIN) spectra
 RIN.pngOPLTF.png 
  1018   Wed Oct 1 23:21:03 2008 YoichiConfigurationPSLReference cavity reflection cameraI re-centered the reference cavity reflection
camera, which has been mis-aligned for a
while.
  
  1023   Fri Oct 3 15:09:58 2008 robUpdatePSLFAST/SLOW
Last night during locking, for no apparent
reason (no common mode), the PSL FAST/SLOW
 FASTSLOW.png 
  1034   Wed Oct 8 19:17:55 2008 YoichiConfigurationPSLLaser power is slowly recoveringThis afternoon we (rich, steve, yoichi) shutdown
the laser for the DC-DC converter installation.
  
  1035   Wed Oct 8 21:26:20 2008 YoichiUpdatePSLAttempt to replace the DC-DC converter (aborted)Rich, Steve, Yoichi

We opened the MOPA box and inspected our
 DCDC.JPGNPRO.JPG 
  1037   Wed Oct 8 23:18:23 2008 YoichiUpdatePSLCorrelation between the Sorensen switching noise and the FSS error signalI took some spectra and coherence function
of the FSS error signal and the +24V Sorensen
power line.
 PowerLineSpe.pngCoherence.png 
  1042   Mon Oct 13 11:32:50 2008 YoixhiUpdatePSLMOPA is in trouble nowSteve, Alberto, Yoichi

A quick update.
 powerdrop.png 
  1043   Mon Oct 13 13:51:49 2008 peteConfigurationPSLattempt to measure FSS ref phaseOn Friday I began a measurement of the FSS
reference phase.  The setup involves the
following:
  
  1044   Mon Oct 13 13:56:03 2008 YoichiUpdatePSLMOPA is not that much in trouble nowThe problem was found to be all to do with
the shutter.
The shutter started to work again, after
  
  1045   Mon Oct 13 18:59:39 2008 YoichiUpdatePSLNPRO EMI and FSS error signal correlationI made a simple loop antenna to measure the
electro-magnetic inteference (EMI) around
the master oscillator NPRO.
 IMG_1692.JPGSpectrum.pngSpectrumFar.pngCoherence.png 
  1046   Tue Oct 14 14:19:36 2008 peteConfigurationPSLFSS ref phaseToday I made several measurements which should
yield the optimized phase for the FSS 21.5
MHz reference.  I made two sets of measurements,
  
  1048   Tue Oct 14 19:24:34 2008 YoichiConfigurationPSLFSS light power reducedRana, Yoichi

To change the gain distribution in the FSS,
  
  1049   Wed Oct 15 17:40:50 2008 ranaUpdatePSLPMC Offset adjustedI set the PMC servo input offset: closed the
MOPA shutter, zeroed the mixer output with
the offset slider, 
  
  1050   Wed Oct 15 22:07:52 2008 peteConfigurationPSLFSS ref phase measurementsOptimizing the FSS LO/RF phase at 500 kHz,
above the servo band, proved to be noisy
and those measurements were useless.  Today
 phasedelay.png 
  1051   Thu Oct 16 09:44:49 2008 YoichiUpdatePSLBad cable for FSSYesterday arount 1:30PM, we lost the LO signal
for the FSS.
I found it was caused by a bad cable connecting
  
  1052   Thu Oct 16 09:47:49 2008 YoichiConfigurationPSLFSS ref phase measurements[quote] I fit the bottom (quadratic) portion
of this curve, and found an optimum delay
of 25.8 ns, which can be implemented as 25.81
  
  1053   Thu Oct 16 13:12:58 2008 peteConfigurationPSLphase between FSS reference outputsI verified the phase between the FSS reference
outputs (used for LO and RF) using matched
BNC cables.  I measured 0.95 degree (average
  
  1054   Thu Oct 16 16:26:26 2008 peteConfigurationPSLFSS phase matching cable installedRG 405 cable has a solid teflon dielectric,
and a velocity factor of 0.69.  To get the
8.2 degrees of additional phase on the LO
  
  1061   Mon Oct 20 20:50:09 2008 YoichiConfigurationPSLFSS board chip replacementA quick update.

I changed two AD797s on the FSS board to
  
  1063   Tue Oct 21 16:17:45 2008 YoichiUpdatePSLAD797 Oscillation in the FSS boardI checked each op-amp's output in the FSS
board to see if any indication of slew-rate
saturation can be found.
  
  1064   Tue Oct 21 17:52:30 2008 ranaSummaryPSLFSS Photo: early OctoberThis is a photo of the FSS board before Yoichi
did his surgery - it was taken with the D40
in macro mode, sitting on the big Gorilla
 fss.jpg 
  1078   Thu Oct 23 20:47:28 2008 peteConfigurationPSLFSS LO calibration for MEDMToday I took a quick series of measurements
to calibrate the FSS LO power measurement
in the MEDM.  This was done by using the
 fss_lo_calibration.png 
  1079   Thu Oct 23 21:52:27 2008 YoichiUpdatePSLFSS UGF now 450kHzI measured the open loop transfer function
of the FSS, for the first time after I mitigated
the oscillation.
 FSS_OPLTF.png 
  1080   Thu Oct 23 23:09:18 2008 YoichiUpdatePSLMC UGF is now 75kHzI measured three loop transfer functions of
the MC servo.
The blue curve in the first attachment is
 MC_Loop.pngOldAndNew.png 
  1083   Fri Oct 24 11:21:26 2008 peteConfigurationPSLFSS LO input calibrated in dBmBased on the measurements described in my
previous elog, I created a new calc record
in the file /cvs/cds/caltech/target/c1psl/psl.db
  
  1095   Mon Oct 27 14:48:27 2008 YoichiConfigurationPSLEO shutter installed to the reference cavityI'm now preparing for cavity ring down measurements
of the reference cavity.
An EOM for polarization rotation is installed
  
  1099   Wed Oct 29 12:23:04 2008 YoichiConfigurationPSLMZ alignment touched and the alarm level changedSince the MZ reflection is alarming all the
time, I tried to improve the MZ alignment
by touching the folding mirror.
  
  1107   Mon Nov 3 09:59:47 2008 steveUpdatePSLPSL HEPAs turned onThe psl [COLOR=chocolate]enclosure[/COLOR]
HEPAs were tuned on.
  
  1114   Tue Nov 4 17:58:42 2008 AlbertoDAQPSLMC temperature sensorI added a channel for the temperature sensor
on the MC1/MC3 chamber: C1:PSL-MC_TEMP_SEN.
To do that I had to reboot the frame builder.
  
  1124   Fri Nov 7 18:38:19 2008 AlbertoDAQPSLMC temperature sensor hooked upAlberto, Rana,
we found that the computer handling the signals
from ICS-110B was C1IOVME so we restarted
  
  1130   Wed Nov 12 11:14:59 2008 CarynDAQPSLMC temp sensor hooked up incorrectlyMC Temperature sensor was not hooked up correctly.
It turns out that for the 4 pin LEMO connections
on the DAQ like J13, J14, etc. the channels
  
  1136   Fri Nov 14 19:20:42 2008 YoichiUpdatePSLReference cavity ring downThanks to Bob making the high-voltage BNC
cables for the HV pulse generator, I was
able to operate the EOM in front of
 RingDown.png 