  519   Wed Jun 4 16:57:12 2008 josephbConfigurationCamerasDark images from cameras (electronics noise measurement)The attached pdfs are 1 second and 1 millisecond
long integrations from the GC650 and GC750
cameras with a cap in place - i.e. no light.
 GC650_1sec_dark.pdfGC650_1msec_dark.pdfGC750_1sec_dark.pdfGC750_1msec_dark.pdfGC650_1sec_dark_zoom.pdf 
  520   Thu Jun 5 10:46:26 2008 josephbConfigurationCamerasApproximately uniform reflected white lightIn an attempt to investigate the structures
seen in previous images for the GC750, I
aimed it at a relatively clean section of
 GC650_white_light.pdfGC750_white_light.pdf 
  521   Thu Jun 5 13:35:23 2008 josephbConfigurationCamerasGC750 looking at 1064nm scattered lightI've taken 200 images of the GC750 (CMOS)
camera while holding it by hand up to a beam
card (also held by hand) in the path of ~5mW
 GC750_HeNe_scatter_avg.pdfGC750_HeNe_scatter_avg2.pdfGC750_HeNe_scatter_avg3.pdfGC750_scatter_avg.pdfGC750_nitrogen_white.pdf 
  525   Fri Jun 6 16:47:04 2008 josephbConfigurationCameras GC650 scatter images of 1064nm lightTook images similar to the scattered light
images from earlier, except with the CCD
GC650 camera.  The first three attached plots
 GC650_scatter_200.pdfGC650_scatter_100a.pdfGC650_scatter_100b.pdf 
  530   Wed Jun 11 15:30:55 2008 josephbConfigurationCamerasGC1280The trial use GC1280 has arrived.  This is
a higher resolution CMOS camera (similar
to the GC750). Other than higher resolution,
 GC1280_60000E_scatter_2d.pdfGC1280_60000E_scatter_3d.pdf 
  558   Tue Jun 24 17:12:10 2008 josephb, EricConfigurationCamerasGC750 setup, 1X4 Hub connected, ETMX imagesThe GC750 camera has been setup to look at
ETMX.  In addition, the new 1X4 rack mounted
switch (131.215.113.200) has been connected
 GC750_ETMX_E30000_single.pdfGC750_ETMX_E30000_avg.pdfGC750_ETMX_E30000_std.pdf 
  566   Wed Jun 25 12:25:28 2008 EricSummaryCameras2D Gaussian Fitting CodeI initially wrote a script in MATLAB that
takes pictures of the laser beam's profile
and fits them to a two dimensional gaussian
  
  622   Wed Jul 2 10:35:02 2008 EricSummaryCamerasGeneral SummaryI finished up the 2D Gaussian fitting code,
and, along with Joe, integrated into the
Snap software so that it automatically does
 attemptedSinglePeakFitWithoutOffset.tiffattemptedSinglePeakFitWithoutOffsetZoomed.tiffattemptedPeakSubtraction.tiffOMCSweepSinglePeak.tiff 
  657   Thu Jul 10 23:27:57 2008 JohnMetaphysicsCamerasSecret handshakesRob and I have joined the ranks of the illuminati
and exercised our power.
  
  660   Fri Jul 11 20:16:01 2008 EricDAQCamerasTaking data from the GC 750 CameraMafalda has been set up with a background
process to constantly take data from the
GC 750 camera (at the end of the x-arm) for
  
  671   Tue Jul 15 10:09:42 2008 EricDAQCamerasDid anyone kill the picture taking process on Mafalda?Did anyone kill the process on Mafalda that
was taking pictures of the end mirror of
the x-arm last Friday?  I need to know whether
  
  678   Wed Jul 16 10:50:55 2008 EricSummaryCamerasWeekly SummaryFinished unwrapping, cleaning, baking, wrapping,
wrapping again, packing, and shipping the
baffles.  
 AttemptedPeakFit3.tiff 
  681   Wed Jul 16 15:59:04 2008 josephb, EricConfigurationCamerasPMC trans camera pathIn order to reduce saturation, we placed a
Y1 plate (spare from the SP table) in transmission
just before the GC650 camera looking at the
  
  693   Fri Jul 18 12:24:15 2008 josephb, EricConfigurationCamerasChanged Lenses on GC750 at ETMXWe removed the giant TV zoom lens and replaced
it with a much smaller fixed zoom lens. 
Currently it views the entire optic.  We
  
  722   Wed Jul 23 12:42:23 2008 EricSummaryCamerasWeekly SummaryI finally got the ezcaPut command working.
 The camera code can now talk directly to
the EPICS channels.  However, after repeated
 PMCScanSinglePeak.tifPMCScanWide.tifPMCSlowSinglePeak.tif 
  769   Wed Jul 30 13:52:41 2008 EricSummaryCamerasWeekly SummaryI tracked the tendency for ezcaPut to fail
and sometimes seg-fault in the camera code
to a conflict between the camera API and
  
  809   Thu Aug 7 11:54:26 2008 josephbConfigurationCamerasNew code + gstreamer allows for easy saving and compression of imagesModified the CamSnap code to output the image
data stream to standard out.  This can then
be piped into a gstreamer plugin and then
 testVideo.ogm 
  812   Fri Aug 8 09:54:10 2008 ranaUpdateCamerasNew code + gstreamer allows for easy saving and compression of images[quote]Modified the CamSnap code to output
the image data stream to standard out.  This
can then be piped into a gstreamer plugin
  
  813   Fri Aug 8 10:58:05 2008 josephbConfigurationCamerasCameras and gstreamerIn regards to camera failure:
1) I forgot to reconnect that particular
camera to the network (my fault) so thats
  
  828   Tue Aug 12 12:21:13 2008 josephbConfigurationCamerasVariation in fit over 140 images for GC650 and GC750Used matlab to calculate Gaussian fits on
145 GC650 images and 142 GC750 images.  These
were individual images (no averaging) looking
  
  835   Thu Aug 14 15:51:35 2008 josephbSummaryCamerasFOUND! The Missing Standoff!We used a zoom lens on the GC750 to take this
picture of the standoff while inside a plastic
rubber-glove bag.  The standoff with bag
 standoff.png 
  839   Fri Aug 15 11:52:32 2008 josephbConfigurationCamerasMulti-computer display and recording of digital camera outputThrough the magic of gstreamer, I've been
able to live play on one machine, compress
the image, send it to another machine via
  
  847   Mon Aug 18 15:32:18 2008 josephbConfigurationCamerasHow to multicast with gstreamer and Gige CamerasIn order to get multicasting to work, one
simply needs to understand the address scheme.
  
  861   Wed Aug 20 12:39:11 2008 EricSummaryCamerasWeekly SummaryI attempted to model the noise produced by
the mirror defects in the ETMX images, in
order to better assure that the fit to the
 OriginalSignImage.pngModifiedSignImage.pngOriginalTVImage.pngModifiedTVImage.png 
  880   Mon Aug 25 14:42:09 2008 EricConfigurationCamerasETMX Digital CameraI changed the lens on the camera looking at
the ETMX to a 16mm, 1:1.4 zoom lens.  This
is in preparation to measure a couple parameters
  
  891   Wed Aug 27 12:09:10 2008 EricSummaryCamerasWeekly SummaryI added a configuration file parser to the
Snap code.  This allows all command line
parameters (like exposure time, etc.) to
 originalETMX.pngrotatedETMX.png 
  914   Wed Sep 3 12:26:49 2008 EricSummaryCamerasWeekly SummaryFinished up simulating the end mirror error
in order to test the whether the fitting
code still provides reasonable answers despite
 SimulatedErrorComparison.png 
  1321   Wed Feb 18 21:03:22 2009 ranaUpdateCamerasETMY Camera work not elogged!The control room video is showing us a
false ETMY image. Who worked on the ETMY
camera or video today??!!
  
  1333   Mon Feb 23 16:42:08 2009 josephbConfigurationCamerasCamera Beta TestingI've setup the GC650 camera (ID 32223)
to look at the mode cleaner transmission. 
I've also added an alias to the camera server
  
  1355   Wed Mar 4 17:20:04 2009 josephbUpdateCamerasCamera code upgradesI've updated the digital camera python
code as well as changed the network topology.
At the moment, both cameras are connected
  
  1385   Wed Mar 11 11:30:15 2009 josephbConfigurationCameras I modified the Video.db file used by c1aux
located in  /cvs/cds/caltech/target/c1aux.
I added the following channels to
  
  1496   Sun Apr 19 11:34:33 2009 josephbHowToCamerasUSB Frame Grabber - How toTo use the Sensoray 2250 USB frame grabber:
Ensure you have the following packages
installed: build-essential, libusb-dev
  
  1497   Sun Apr 19 11:51:05 2009 josephbUpdateCamerasMafalda may need an updateI tried installing libusb-dev on mafalda
in order to try getting the usb frame grabber
to work on it, but could not as it could
  
  1499   Mon Apr 20 11:57:27 2009 robUpdateCamerasMafalda may need an update


   
        
            
  
  1501   Mon Apr 20 18:36:37 2009 ranaUpdateCamerasMafalda may need an updateSadly, the sensoray crap doesn't seem to build
on CentOS. I too would prefer a homogenous
solution, 
  
  1581   Wed May 13 12:41:14 2009 josephbUpdateCamerasTiming and stability tests of GigE Camera codeAt the request of people down at LLO I've
been trying to work on the reliability and
speed of the GigE camera code.  In my
 newCodeFPS.pngnewCodeFPS_hist.png 
  1590   Fri May 15 16:47:44 2009 josephbUpdateCamerasImproved camera codeAt Rob's request I've added the following
features to the camera code.
The camera server, which can be started
  
  1697   Wed Jun 24 12:04:22 2009 ZachUpdateCamerasSURF entryThis week, I've been reading some literature
concerning PLL and familiarizing myself with
LINUX, MATLAB, and high-pass filter circuits. 
  
  1712   Wed Jul 1 11:04:27 2009 ZachUpdateCamerasGigE Phase CameraThis past week, I have building a sine
wave rectifier and trying to write a simple
program that displays a ccd image to familiarize
 trigger.jpgfig1b.pdf 
  1721   Wed Jul 8 11:08:43 2009 ZachUpdateCamerasGigE Phase CameraThe plan for the optical setup has been
corrected after it was realized that it would
be impossible to isolate a 29.501 MHz frequency
 fig1.pdffig2.pdf 
  1751   Wed Jul 15 14:42:31 2009 ZachUpdateCamerasGigE Phase CameraLately, I have been able to externally
trigger the camera using a signal generator
passing through the op-amp circuit that I
 fig1koji.pdf 
  1753   Wed Jul 15 18:22:15 2009 KojiUpdateCamerasRe: GigE Phase Camera


   
        
            
  
  1778   Wed Jul 22 14:44:57 2009 ZachUpdateCamerasGigE Phase CameraThis past week, I have mostly been debugging
my software.  I have tried to use the
fluorescent lights to test the camera, but
  
  1807   Wed Jul 29 14:22:33 2009 ZachUpdateCamerasGigE Phase CameraThis week, Joe and I have been setting
up the laser and optics.  The mephisto
laser is emitting a very ugly beam that we
  
  1822   Mon Aug 3 18:56:59 2009 ZachUpdateCamerasGigE Phase CameraWhile aligning the optics, we tried to
start up the CCD.  Although nothing
should have changed since the last time I
  
  1824   Tue Aug 4 11:45:29 2009 ZachUpdateCamerasGigE Phase CameraThe camera wasn't working because the router
has no built-in dhcp server.  We had
to manually start the server after rebooting
  
  1861   Fri Aug 7 17:46:21 2009 ZachUpdateCamerasThe phase camera is sort of workingShown below are the plots of the amplitude
and phase of the Mephisto laser light modulated
with a chopper as a square wave at about
 ampAv.pngphaseAv.png 
  1862   Fri Aug 7 17:51:50 2009 ZachUpdateCamerasCMOS vs. CCDThe images that I just posted were taken
with the CMOS camera.  We switched from
the CCD to the CMOS because the CCD was exhibiting
  
  2120   Mon Oct 19 18:14:28 2009 robUpdateCamerasvideo switch brokenThe Chameleon HB (by Knox) video switch
that we use for routing video signals into
the control room monitors is broken. 
  
  2304   Fri Nov 20 00:18:45 2009 ranaSummaryCamerasVideo MUX Selection Wiki pageSteve is summarizing the Video Matrix
choices into this Wiki page:
http://lhocds.ligo-wa.caltech.edu:8000/40m/Electronics/VideoMUX
  
ELOG V3.1.3-