  40m Log, Page 274 of 355  Not logged in ELOG logo
ID Date Author Type Category Subjectdown Text Attachments
  9163   Thu Sep 26 01:49:28 2013 MasayukiSummaryGreen LockingFPMI noise caused by ARM locking 
 Measurement
 MICH_OLTF.pdfBS-RS55Q.pngETMX-RS55.pngETMY-RS55.pngplot.pdf 
  9167   Thu Sep 26 23:02:40 2013 ranaUpdateLSCFPMI noise caused by ARM lockingHidden in Nakano-kun's previous entries
was that the phase margin of the X-Arm was
only 9 degrees!! This extremely close to
 Xarm_sweep_130926.pdflsc.pdferr.png 
  9168   Fri Sep 27 00:48:53 2013 MasayukiUpdateLSCFPMI noise caused by ARM locking

    

       
            
 phase_badget.pngFPMI_noises.png 
  17069   Tue Aug 9 19:54:31 2022 yutaSummaryLSCFPMI locking tonight[Tega, Anchal, Yuta]

We resored FPMI locking settings.
  
  17012   Mon Jul 18 16:39:07 2022 PacoSummaryLSCFPMI locking procedure using REFL55 and AS55[Yuta, Paco]

In summary, we locked FPMI using
REFL55_I, REFL55_Q, and AS55_Q. The key to
 DARM_07_18_2022_FMPI.pdfCARM_07_18_2022_FPMI.pdfMICH_07_18_2022_FPMI.pdffpmi_darm_nb_2022_07.pdf 
  17016   Mon Jul 18 21:41:42 2022 AnchalSummaryLSCFPMI locking procedure using REFL55 and AS55Now that you have found a working configuration,
I suggest we update CARM and DARM filter
banks so that they are used in locking those
  
  14926   Wed Oct 2 23:15:02 2019 gautamUpdateLSCFPMI lockingSummary:

I was able to lock the FPMI. The
lock was quite stable. However,
 FPMIlocked.png 
  17369   Wed Dec 21 12:18:44 2022 PacoSummaryLSCFPMI locked, but FPMI BHD not locked[Yuta, Paco]

We tried restoring the FPMI BHD
readout from yesterday but 
 FPMI_calibrated_noise_20221221.pdf 
  17367   Tue Dec 20 18:27:14 2022 AnchalSummaryLSCFPMI locked, DARM calibration data taken, FPMI BHD locked![Anchal, Paco, Yuta]

FPMI
 PXL_20221221_013126861.MP.jpg 
  17366   Tue Dec 20 09:20:18 2022 PacoSummaryLSCFPMI locked and cal [Anchal, Yuta, Paco]

Late elog (from yesterday) -- We
locked FPMI, YAUX, and turned on calibration
  
  7240   Tue Aug 21 01:54:09 2012 JenneUpdateLockingFPMI locked - arms locked with IRI (for the first time personally) locked
the FPMI.  I have data for the POX11I,
POY11I, AS55Q error signals for each arm
  
  8170   Tue Feb 26 11:55:11 2013 ManasaUpdateLockingFPMI locked [Yuta, Manasa]

FPMI locked.
  
  8171   Tue Feb 26 15:32:47 2013 JenneUpdateLockingFPMI locked

    

       
            
  
  9157   Tue Sep 24 22:19:57 2013 ManasaUpdateGeneralFPMI locked [Masayuki, Manasa]
We locked FPMI and measured the FPMI
noise (power spectrum of error signal - MICH_IN1)
  
  6914   Wed Jul 4 21:11:53 2012 yutaUpdateLockingFPMI in vacuum is backI aligned FPMI and greens. There's no recognizable
difference between before and after the vent.
 ALL_1025495266.pngMEDMscreenshotswithCOW_20120704.png 
  17359   Wed Dec 14 17:01:39 2022 PacoSummaryLSCFPMI in the post-cds upgrade era[Paco, Anchal, Koji]

A possible long-hidden
bug of MC2 dewhitening switching was found
  
  17414   Tue Jan 24 11:21:59 2023 yutaSummaryBHDFPMI displacement noise with 60 Hz harmonics side lobesJust to show how bad 60 Hz noise is, I
compared FPMI displacement noise with pre-BH44
era (measured on Jan 13, 40m/17400).
 FPMI_calibrated_noise_20230124_60Hz.pdfScreenshot_2023-01-24_11-27-22_FPMIndscope.png 
  17415   Tue Jan 24 16:12:07 2023 ranaSummaryBHDFPMI displacement noise with 60 Hz harmonics side lobes1) do a comparison with the whtiening before
the ADC on/off. This will tell us if it is
pickup before the whtiening filter or not.
  
  17593   Tue May 16 08:42:14 2023 PacoUpdateASCFPMI differential ASC - IIWe
aligned the AS WFS and measured some angular
motion from ETMX
 PXL_20230515_173719320~2.jpgASWFS_test0_Screenshot_2023-05-16_18-38-39.png 
  17589   Wed May 10 17:11:34 2023 PacoUpdateASCFPMI differential ASC - II
began commissioning the AS WFS for (PR)FPMI
configurations
 PXL_20230510_233850237~2.jpgPXL_20230511_000946357~2.jpgPXL_20230511_000941172~2.jpg 
  17089   Thu Aug 18 14:49:35 2022 YehonathanSummaryLSCFPMI Sensitivity{Yuta, Yehonathan}

We wrote a notebook found on Git/40m/measurements/LSC/FPMI/NoiseBudget/FPMISensitivity.ipynb
for calculating the MICH, DARM (currently
 Sensitivity_Plot_1344133503.pdf 
  17091   Thu Aug 18 18:10:49 2022 KojiSummaryLSCFPMI SensitivityThe overlapping plot of the calibrated
error and control signals gives you an approximately
good estimation of the freerun fluctuation,
  
  4180   Thu Jan 20 22:17:12 2011 ranaSummaryLSCFPMI Displacement NoiseI found this old plot in an old elog entry
of Osamu's (original
link).
 osamu-1140657006.pdf 
  17499   Wed Mar 8 18:32:22 2023 AnchalConfigurationCalibrationFPMI DARM calibration run set to happen at 1 amOn rossa in tmux session name FPMI_DARM_Cal,
a script is running to take FPMI DARM calibration
data at 1:00 am on March 9th. Please do not
  
  17502   Thu Mar 9 19:20:44 2023 AnchalConfigurationCalibrationFPMI DARM calibration run set to happen at 1 amRunning this test again tonight. Will probably
run it every night now.

 
  
  17492   Sat Mar 4 18:57:18 2023 PacoConfigurationCalibrationFPMI DARM calibration run Locked FPMI, measured DARM and CARM OLTFs,
locked YAUX and measured the analog loop
TF at the cal line frequencies. Turned the
  
  17468   Thu Feb 16 14:44:06 2023 yutaUpdateBHDFPMI BHD with BH55 recoveredFPMI BHD with LO phase locked using BH55
is recovered after 60 Hz frequency noise
saga.
 FPMI_calibrated_noise_20230216.pdfFPMI_calibrated_noise_20230216_60Hz.pdf 
  17395   Thu Jan 12 12:00:09 2023 PacoSummaryLSCFPMI BHD sensitivity curve with higher resolution upto 6.9 kHzHere's the same sensitivity plot from
yesterday (40m/17392),
but with higher frequency resolution, upto
 FPMI_calibrated_noise_20230112.pdfFPMI_calibrated_noise_20230112_HF.pdf 
  17506   Mon Mar 13 19:53:36 2023 yutaUpdateBHDFPMI BHD sensing matrix measurement with individual linesFPMI BHD sensing matrix was measured by
an updated method with updated RF demodulation
phases for REFL55 and AS55.
  
  17392   Wed Jan 11 16:56:57 2023 yutaSummaryLSCFPMI BHD recovered, LO phase noise not limiting the sensitivity[Paco, Yuta]

We recovered FPMI BHD, and sensitivity
 6x 
  835   Thu Aug 14 15:51:35 2008 josephbSummaryCamerasFOUND! The Missing Standoff!We used a zoom lens on the GC750 to take this
picture of the standoff while inside a plastic
rubber-glove bag.  The standoff with bag
 standoff.png 
  11110   Fri Mar 6 14:29:59 2015 manasaUpdateGeneralFOL troubleshooting[EricQ, Manasa]

Domenica:
Since Domenica
  
  11123   Mon Mar 9 14:43:31 2015 manasaUpdateGeneralFOL troubleshootingFiguring out the problem with
frequency counter readouts:

 FOL_trouble.pdfIMAG0341.jpgIMAG0342.jpg 
  11128   Tue Mar 10 17:48:07 2015 manasaUpdateGeneralFOL troubleshooting[Koji, Manasa]

This problem was solved by changing
the Frequency Counter range settings.
  
  11135   Wed Mar 11 19:48:25 2015 KojiUpdateGeneralFOL troubleshootingThere is a frequency counter code written
by the summer student.
The code needed some cleaning up.
  
  11137   Thu Mar 12 11:57:38 2015 KojiUpdateGeneralFOL troubleshootingBTW, during this trouble shoot, we looked
at the IR beatnote spectrum between the Xend
and the PSL.
  
  11650   Tue Sep 29 19:38:09 2015 gautamUpdateGeneralFOL fiber box revampThe new 2x2 fiber couplers arrived today
so Eric gave me an overview on the changes
to be made to the existing configuration
 IMG_0017.JPGFOL_schematic.pdfIMG_0018.JPGIMG_0016.JPG 
  11652   Wed Sep 30 13:07:13 2015 gautamUpdateGeneralFOL fiber box revampEric pointed out that the 1x2 couplers
that were used in the previous arrangement
and which I recycled, were in fact NOT
 FOL_schematic.pdf 
  11670   Tue Oct 6 16:56:40 2015 gautamUpdateGeneralFOL fiber box revamp[gautam, ericq]

We had a look at the IR beat (PSL+Xarm)
today using the new FOL fiber box, and compared
 20151006_Xbeat_psd.pdf20151006_Xbeat_coherence.pdf 
  11102   Thu Mar 5 12:27:27 2015 manasaUpdateGeneralFOL fiber box on the PSL tableWorking
around the PSL table

I have put the FOL
  
  10352   Fri Aug 8 14:27:18 2014 AkhilUpdateComputer Scripts / ProgramsFOL Scripts The scripts written for interfacing
the FC with R Pi, building EPICS database,
piping data into EPICS channels,PID loop
  
  10376   Wed Aug 13 16:12:55 2014 HarryUpdateGeneralFOL Layout DiagramPer Q's request, I've made up a diagram
of the complete FOL layout for general reference.
 FOLLayout2.png 
  10273   Fri Jul 25 17:28:31 2014 HarryUpdateGeneralFOL Box and PER Update Purpose
We're putting together a box to go
into the 1X2 rack, to facilitate the frequency
 FOLBox.png 
  1498   Mon Apr 20 05:18:42 2009 YoichiConfigurationLockingFM6 and FM10 of LSC-MC were restoredDuring tonight's locking work, I realized
that FM6 and FM10 (both resonant gains around
20Hz) were actually activated by cm_step.
  
  10226   Thu Jul 17 02:57:32 2014 AndresUpdate40m Xend Table upgradeFInish Calculation on Current X-arm mode MatchingData and Calculation for the
Xarm Current Mode Matching
 RawDataForTheModeGreenScan.pngResultForModeMatching.pngDataAndCalculationOfModeMismatch.zip 
  10237   Fri Jul 18 16:52:56 2014 AndresUpdate40m Xend Table upgradeFInish Calculation on Current X-arm mode Matching

    

       
            
 OldAndNewSetupPlotsOfDisplacementAndAngleAtTheETM.pngOldSetUpDisplacementAndNewSetup.m.zip 
  17711   Mon Jul 24 23:01:10 2023 KojiSummaryCDSFIXED: rossa can't bootCompared the network settings between some
of the machines.

It seemed that we can write network
  
  17738   Mon Jul 31 22:25:07 2023 KojiUpdatePEMFIXED: Strange behavior of ITMX and ITMY probably due to DAC issue[Hiroki, Yehonathan, Yuta, Koji]

Summary


The
 PXL_20230731_233855623.MP.jpgPXL_20230731_235323818.jpgPXL_20230801_033401354.MP.jpgPXL_20230801_033332906.MP.jpg 
  17746   Wed Aug 2 14:50:22 2023 KojiUpdatePSLFIXED: Remote switching of PSL shutter is not working[JC Koji]
c1vac was hard rebooted. This fixed
the PSL shutter issue.
  
  17786   Tue Aug 15 17:11:17 2023 KojiUpdateIOOFIXED: PMC issueSummary

The problem in the PMC
Frequency Reference Card (35.5MHz) was fixed.
 7x 