ID Date Author Type Categoryup Subject Text Attachments
  1023   Fri Oct 3 15:09:58 2008 robUpdatePSLFAST/SLOW
Last night during locking, for no apparent
reason (no common mode), the PSL FAST/SLOW
 FASTSLOW.png 
  1034   Wed Oct 8 19:17:55 2008 YoichiConfigurationPSLLaser power is slowly recoveringThis afternoon we (rich, steve, yoichi) shutdown
the laser for the DC-DC converter installation.
  
  1035   Wed Oct 8 21:26:20 2008 YoichiUpdatePSLAttempt to replace the DC-DC converter (aborted)Rich, Steve, Yoichi

We opened the MOPA box and inspected our
 DCDC.JPGNPRO.JPG 
  1037   Wed Oct 8 23:18:23 2008 YoichiUpdatePSLCorrelation between the Sorensen switching noise and the FSS error signalI took some spectra and coherence function
of the FSS error signal and the +24V Sorensen
power line.
 PowerLineSpe.pngCoherence.png 
  1042   Mon Oct 13 11:32:50 2008 YoixhiUpdatePSLMOPA is in trouble nowSteve, Alberto, Yoichi

A quick update.
 powerdrop.png 
  1043   Mon Oct 13 13:51:49 2008 peteConfigurationPSLattempt to measure FSS ref phaseOn Friday I began a measurement of the FSS
reference phase.  The setup involves the
following:
  
  1044   Mon Oct 13 13:56:03 2008 YoichiUpdatePSLMOPA is not that much in trouble nowThe problem was found to be all to do with
the shutter.
The shutter started to work again, after
  
  1045   Mon Oct 13 18:59:39 2008 YoichiUpdatePSLNPRO EMI and FSS error signal correlationI made a simple loop antenna to measure the
electro-magnetic inteference (EMI) around
the master oscillator NPRO.
 IMG_1692.JPGSpectrum.pngSpectrumFar.pngCoherence.png 
  1046   Tue Oct 14 14:19:36 2008 peteConfigurationPSLFSS ref phaseToday I made several measurements which should
yield the optimized phase for the FSS 21.5
MHz reference.  I made two sets of measurements,
  
  1048   Tue Oct 14 19:24:34 2008 YoichiConfigurationPSLFSS light power reducedRana, Yoichi

To change the gain distribution in the FSS,
  
  1049   Wed Oct 15 17:40:50 2008 ranaUpdatePSLPMC Offset adjustedI set the PMC servo input offset: closed the
MOPA shutter, zeroed the mixer output with
the offset slider, 
  
  1050   Wed Oct 15 22:07:52 2008 peteConfigurationPSLFSS ref phase measurementsOptimizing the FSS LO/RF phase at 500 kHz,
above the servo band, proved to be noisy
and those measurements were useless.  Today
 phasedelay.png 
  1051   Thu Oct 16 09:44:49 2008 YoichiUpdatePSLBad cable for FSSYesterday arount 1:30PM, we lost the LO signal
for the FSS.
I found it was caused by a bad cable connecting
  
  1052   Thu Oct 16 09:47:49 2008 YoichiConfigurationPSLFSS ref phase measurements[quote] I fit the bottom (quadratic) portion
of this curve, and found an optimum delay
of 25.8 ns, which can be implemented as 25.81
  
  1053   Thu Oct 16 13:12:58 2008 peteConfigurationPSLphase between FSS reference outputsI verified the phase between the FSS reference
outputs (used for LO and RF) using matched
BNC cables.  I measured 0.95 degree (average
  
  1054   Thu Oct 16 16:26:26 2008 peteConfigurationPSLFSS phase matching cable installedRG 405 cable has a solid teflon dielectric,
and a velocity factor of 0.69.  To get the
8.2 degrees of additional phase on the LO
  
  1061   Mon Oct 20 20:50:09 2008 YoichiConfigurationPSLFSS board chip replacementA quick update.

I changed two AD797s on the FSS board to
  
  1063   Tue Oct 21 16:17:45 2008 YoichiUpdatePSLAD797 Oscillation in the FSS boardI checked each op-amp's output in the FSS
board to see if any indication of slew-rate
saturation can be found.
  
  1064   Tue Oct 21 17:52:30 2008 ranaSummaryPSLFSS Photo: early OctoberThis is a photo of the FSS board before Yoichi
did his surgery - it was taken with the D40
in macro mode, sitting on the big Gorilla
 fss.jpg 
  1078   Thu Oct 23 20:47:28 2008 peteConfigurationPSLFSS LO calibration for MEDMToday I took a quick series of measurements
to calibrate the FSS LO power measurement
in the MEDM.  This was done by using the
 fss_lo_calibration.png 
  1079   Thu Oct 23 21:52:27 2008 YoichiUpdatePSLFSS UGF now 450kHzI measured the open loop transfer function
of the FSS, for the first time after I mitigated
the oscillation.
 FSS_OPLTF.png 
  1080   Thu Oct 23 23:09:18 2008 YoichiUpdatePSLMC UGF is now 75kHzI measured three loop transfer functions of
the MC servo.
The blue curve in the first attachment is
 MC_Loop.pngOldAndNew.png 
  1083   Fri Oct 24 11:21:26 2008 peteConfigurationPSLFSS LO input calibrated in dBmBased on the measurements described in my
previous elog, I created a new calc record
in the file /cvs/cds/caltech/target/c1psl/psl.db
  
  1095   Mon Oct 27 14:48:27 2008 YoichiConfigurationPSLEO shutter installed to the reference cavityI'm now preparing for cavity ring down measurements
of the reference cavity.
An EOM for polarization rotation is installed
  
  1099   Wed Oct 29 12:23:04 2008 YoichiConfigurationPSLMZ alignment touched and the alarm level changedSince the MZ reflection is alarming all the
time, I tried to improve the MZ alignment
by touching the folding mirror.
  
  1107   Mon Nov 3 09:59:47 2008 steveUpdatePSLPSL HEPAs turned onThe psl [COLOR=chocolate]enclosure[/COLOR]
HEPAs were tuned on.
  
  1114   Tue Nov 4 17:58:42 2008 AlbertoDAQPSLMC temperature sensorI added a channel for the temperature sensor
on the MC1/MC3 chamber: C1:PSL-MC_TEMP_SEN.
To do that I had to reboot the frame builder.
  
  1124   Fri Nov 7 18:38:19 2008 AlbertoDAQPSLMC temperature sensor hooked upAlberto, Rana,
we found that the computer handling the signals
from ICS-110B was C1IOVME so we restarted
  
  1130   Wed Nov 12 11:14:59 2008 CarynDAQPSLMC temp sensor hooked up incorrectlyMC Temperature sensor was not hooked up correctly.
It turns out that for the 4 pin LEMO connections
on the DAQ like J13, J14, etc. the channels
  
  1136   Fri Nov 14 19:20:42 2008 YoichiUpdatePSLReference cavity ring downThanks to Bob making the high-voltage BNC
cables for the HV pulse generator, I was
able to operate the EOM in front of
 RingDown.png 
  1137   Fri Nov 14 20:35:47 2008 ranaUpdatePSLReference cavity ring downTo make the DEI pulser make a fast pulse on
the EO shutter EOMs, we had to make sure:
  
  1138   Fri Nov 14 22:40:51 2008 YoichiUpdatePSLReference cavity ring down[quote]
To make the DEI pulser make a fast pulse
on the EO shutter EOMs, we had to make sure:
  
  1140   Mon Nov 17 15:07:06 2008 YoichiUpdatePSLReference cavity ring downI used MATLAB's system identification tool
box to estimate the response of the reference
cavity, i.e. cavity pole.
 Fit.png 
  1149   Thu Nov 20 09:37:39 2008 steveUpdatePSLMZ vs tempThis 12 days plot shows that it can hold lock
if the daily temp variation is not more than
3/4 of a degree C
 mztemp.jpg 
  1151   Fri Nov 21 16:11:26 2008 rana, alberto, robUpdatePSLMach Zender alignmentThe Mach Zender's dark port DC voltage had
gone up too high (~0.5 V)and was turning
yellow
 PB210051-1.JPG 
  1155   Fri Nov 21 20:29:43 2008 rana, alberto, robUpdatePSLMach Zender alignment[quote] So now the MZ alignment seems good;
Alberto is on the MC periscope alignment
like a cheap suit.[/quote]
  
  1160   Mon Nov 24 17:14:44 2008 ranaUpdatePSLMach Zender trendsIt looks like the MZ has gotten less drifty
after the alignment on Friday.
 Untitled.png 
  1162   Tue Nov 25 18:38:03 2008 Alberto, RobUpdatePSLMC Periscope AlignmentThis morning when I came in I found the MC
cleaner unlocked and the autolocker script
could not lock it. The reflected beam was
  
  1166   Tue Dec 2 17:56:56 2008 Alberto, RanaConfigurationPSLMC AlignmentIn the attempt to maximize the Mode Cleaner
transmission and minimize the reflection
from the steering mirrors of the MC periscope,
  
  1169   Wed Dec 3 11:58:10 2008 AlbertoUpdatePSLMC AlignmentRana, Alberto,

more details on the MC alignment we did yesterday.
  
  1190   Fri Dec 12 22:51:23 2008 YoichiUpdatePSLReference cavity ring down measurement againBob made new HV-cables with HV compatible
coaxes. The coax cable is rated for 2kV,
which was as high as Bob
 RC-Ringdown.png 
  1191   Tue Dec 16 19:06:01 2008 YoichiUpdatePSLReference cavity ring down repeated many timesToday, I repeated the reference cavity ring
down measurement many times to see how much
the results vary.
 RC_Ringdown_Estimates.pngCavity_Pole_Histogram.png 
  1228   Wed Jan 14 15:53:32 2009 steveDAQPSLMC temperature sensor[quote]I added a channel for the temperature
sensor on the MC1/MC3 chamber: C1:PSL-MC_TEMP_SEN.
To do that I had to reboot the frame builder.
  
  1244   Thu Jan 22 11:54:09 2009 AlbertoConfigurationPSLMach Zehnder Output Beam QPDI rotated by 180 degrees the 10% beam splitter
that it is used to fold the beam coming from
the Mach Zehnder (directed to the MC) on
  
  1246   Thu Jan 22 14:38:41 2009 carynDAQPSLMC temperature sensor[quote][quote]I added a channel for the temperature
sensor on the MC1/MC3 chamber: C1:PSL-MC_TEMP_SEN.
To do that I had to reboot the frame builder.
  
  1248   Fri Jan 23 10:00:21 2009 steveUpdatePSLPMC transmission is downThe PMC transmission is going down.
I have not relocked the PMC yet.
 pmc4d.jpg 
  1250   Fri Jan 23 14:00:02 2009 YoichiUpdatePSLPMC transmission is down[quote]The PMC transmission is going down.
I have not relocked the PMC yet.[/quote]
 LaserPower.png 
  1254   Wed Jan 28 12:42:51 2009 YoichiUpdatePSLMOPA dyingYoichi, Jenne, Peter

As most of you know, the MOPA output power
 Recent10Days.png126_MOPA.png 
  1256   Wed Jan 28 19:08:50 2009 YoichiUpdatePSLLaser is back (sort of)Yoichi, Peter, Jenne

Summary:
 Improving.png 
  1257   Thu Jan 29 13:52:34 2009 YoichiUpdatePSLLaser is back (sort of)Here is what I think has happened to the laser.

After the chiller line to the NPRO base clogged,
 SlowDC.pdf 