ID |
Date |
Author |
Type |
Category |
Subject |
2733
|
Tue Mar 30 06:37:32 2010 |
rana | Configuration | PSL | Reference Cavity PD Noise Spectrum |
Some more words about the RFAM: I noticed that there was an excess RFAM by unlocking the RC and just looking at the RF out with the 50 Ohm input of the scope. It was ~100 mVp-p! In the end our method to minimize the AM was not so sensible - we aligned the waveplate before the EOM so as to minimize the p-pol light transmitted by the PBS cube just ahead of the AOM. At first, this did not minimize the RFAM. But after I got angry at the bad plastic mounting of the EOM and re-aligned it, the AM seemed to be small with the polarization aligned to the cube. It was too small to measure on the scope and on the spectrum analyzer, the peak was hopping around by ~10-20 dB on a few second timescale. Further reduction would require some kind of active temperature stabilization of the EOM housing (maybe a good SURF project!).
For the EOM mount we (meaning Steve) should replace the lame 2-post system that's in there with one of the mounts of the type that is used in the Mach-Zucker EOMs. I think we have spare in the cabinet next to one of the arms. 
After the RFAM monkeying, I aligned the beam to the RC using the standard, 2-mirror, beam-walking approach. You can see from the attached plot that the transmission went up by ~20% ! And the reflection went down by ~30%. I doubt that I have developed any new alignment technique beyond what Yoichi and I already did last time. Most likely there was some beam shape corruption in the EOM, or the RFAM was causing us to lock far off the fringe. Now the reflected beam from the reference cavity is a nice donut shape and we could even make it better by doing some mode matching! This finally solves the eternal mystery of the bad REFL beam (or at least sweeps it under the rug).
At the end, I also fixed the alignment of the RFPD. It should be set so the incident angle of the beam is ~20-40 deg, but it was instead set to be near normal incidence ?! Its also on flimsy plastic legs. Steve, can you please replace this with the new brass ones? |
Attachment 1: rc.png
|
|
2742
|
Wed Mar 31 15:31:53 2010 |
steve | Update | PSL | Reference Cavity RF PD base upgraded |
Quote: |
Some more words about the RFAM: I noticed that there was an excess RFAM by unlocking the RC and just looking at the RF out with the 50 Ohm input of the scope. It was ~100 mVp-p! In the end our method to minimize the AM was not so sensible - we aligned the waveplate before the EOM so as to minimize the p-pol light transmitted by the PBS cube just ahead of the AOM. At first, this did not minimize the RFAM. But after I got angry at the bad plastic mounting of the EOM and re-aligned it, the AM seemed to be small with the polarization aligned to the cube. It was too small to measure on the scope and on the spectrum analyzer, the peak was hopping around by ~10-20 dB on a few second timescale. Further reduction would require some kind of active temperature stabilization of the EOM housing (maybe a good SURF project!).
For the EOM mount we (meaning Steve) should replace the lame 2-post system that's in there with one of the mounts of the type that is used in the Mach-Zucker EOMs. I think we have spare in the cabinet next to one of the arms. 
After the RFAM monkeying, I aligned the beam to the RC using the standard, 2-mirror, beam-walking approach. You can see from the attached plot that the transmission went up by ~20% ! And the reflection went down by ~30%. I doubt that I have developed any new alignment technique beyond what Yoichi and I already did last time. Most likely there was some beam shape corruption in the EOM, or the RFAM was causing us to lock far off the fringe. Now the reflected beam from the reference cavity is a nice donut shape and we could even make it better by doing some mode matching! This finally solves the eternal mystery of the bad REFL beam (or at least sweeps it under the rug).
At the end, I also fixed the alignment of the RFPD. It should be set so the incident angle of the beam is ~20-40 deg, but it was instead set to be near normal incidence ?! Its also on flimsy plastic legs. Steve, can you please replace this with the new brass ones?
|
Teflon feet removed and heavy brass-delrin pd base installed. Ref-cavity reflected light remains to be beautiful doughnut shape on camera. |
Attachment 1: brspdbs.JPG
|
|
2759
|
Sat Apr 3 11:35:47 2010 |
rana | Configuration | PSL | Reference Cavity PD Noise Spectrum |
The units on this plot are completely bogus - we know that the thermal noise from the resonant part of the circuit is just V = sqrt(4*k*T*Z) ~ 3nV/rHz. Then the gain of the MAX4107 stage is 10. The output resistor is 50 Ohms, which forms a divide by 2 with the input impedance of the spectrum analyzer and so the bump in the dark noise should only be 15 nV/rHz and not microVolts.
Quote: |
[Rana, Alberto]
This evening we measured the noise spectrum of the reference cavity PD used in the FSS loop. From that we estimated the transimpedance and found that the PD is shot-noise limited. We also found a big AM oscillation in correspondence of the FSS modulation sideband which we later attenuated at least in part.
This plot shows the spectrum noise from the RF output of the photodetector.
|
|
2760
|
Sat Apr 3 16:07:40 2010 |
Alberto | Configuration | PSL | Reference Cavity PD Noise Spectrum |
I was aware of a problem on those units since I acquired the data. Then it wasn't totally clear to me which were the units of the data as downloaded from the Agilent 4395A, and, in part, still isn't.
It's clear that the data was in units of spectrum, an not spectral density: in between the two there is a division by the bandwidth (100KHz, in this case). Correcting for that, one gets the following plot for the FSS PD:

Although the reason why I was hesitating to elog this other plot is that it looks like there's still a discrepancy of about 0.5dBm between what one reads on the display of the spectrum analyzer and the data values downloaded from it.
However I well know that, I should have just posted it, including my reserves about that possible offset (as I'm doing now).
Quote:
|
The units on this plot are completely bogus - we know that the thermal noise from the resonant part of the circuit is just V = sqrt(4*k*T*Z) ~ 3nV/rHz. Then the gain of the MAX4107 stage is 10. The output resistor is 50 Ohms, which forms a divide by 2 with the input impedance of the spectrum analyzer and so the bump in the dark noise should only be 15 nV/rHz and not microVolts.
Quote: |
[Rana, Alberto]
This evening we measured the noise spectrum of the reference cavity PD used in the FSS loop. From that we estimated the transimpedance and found that the PD is shot-noise limited. We also found a big AM oscillation in correspondence of the FSS modulation sideband which we later attenuated at least in part.
This plot shows the spectrum noise from the RF output of the photodetector.
|
|
|
2805
|
Mon Apr 19 05:54:50 2010 |
rana | Configuration | PSL | RC Temperature Servo Turned OFF temporarily |
In order to measure the transfer function of the RC cavity's foam, I've turned off the servo so that the room temperature noise can excite it.
The attached plot shows a step response test from 2 weeks ago. Servo is nominally still working fine. |
Attachment 1: Untitled.png
|
|
2810
|
Mon Apr 19 16:31:42 2010 |
Kevin | Update | PSL | Innolight 2W Laser |
Koji and Kevin
We unpacked the Innolight 2W laser, took an inventory, and scanned the operations manual.
[Edit by KA]
The scanned PDFs are placed on the following wiki page
http://lhocds.ligo-wa.caltech.edu:8000/40m/Upgrade_09/PSL
We will measure the P-I curve, the mode profile, frequency actuator responses, and so on. |
2812
|
Tue Apr 20 07:48:42 2010 |
steve | Update | PSL | ion pump HV turned on |
We found ref-cavity HV was off yesterday afternoon. It was turned back on. |
2822
|
Tue Apr 20 20:15:37 2010 |
Kevin | Update | PSL | Innolight 2W Output Power vs Injection Current |
Koji and Kevin measured the output power vs injection current for the Innolight 2W laser.
The threshold current is 0.75 A.
The following data was taken with the laser crystal temperature at 25.04ºC (dial setting: 0.12).
Injection Current (A) |
Dial Setting |
Output Power (mW) |
0.000 |
0.0 |
1.2 |
0.744 |
3.66 |
1.1 |
0.753 |
3.72 |
4.6 |
0.851 |
4.22 |
102 |
0.954 |
4.74 |
219 |
1.051 |
5.22 |
355 |
1.151 |
5.71 |
512 |
1.249 |
6.18 |
692 |
1.350 |
6.64 |
901 |
1.451 |
7.08 |
1118 |
1.556 |
7.52 |
1352 |
1.654 |
7.92 |
1546 |
1.761 |
8.32 |
1720 |
1.853 |
8.67 |
1855 |
1.959 |
9.05 |
1989 |
2.098 |
9.50 |
2146 |
|
Attachment 1: PvsI_2W.jpg
|
|
2828
|
Wed Apr 21 21:56:27 2010 |
Kevin | Update | PSL | Innolight 2W Vertical Beam Profile |
Koji and Kevin measured the vertical beam profile of the Innolight 2W laser at one point.
This data was taken with the laser crystal temperature at 25.04°C and the injection current at 2.092A.
The distance from the razor blade to the flat black face on the front of the laser was 13.2cm.
The data was fit to the function y(x)=a*erf(sqrt(x)*(x-x0)/w)+b with the following results.
Reduced chi squared = 14.07
x0 = (1.964 +- 0.002) mm
w = (0.216 +- 0.004) mm
a = (3.39 +- 0.03) V
b = (3.46 +- 0.03) V |
Attachment 1: bp2.jpg
|
|
Attachment 2: bp2.dat
|
razor height (mm) Voltage (V)
2.75 6.89
2.50 6.90
2.30 6.89
2.25 6.89
2.20 6.75
2.15 6.47
2.13 6.20
2.10 6.05
2.07 5.88
... 17 more lines ...
|
2829
|
Wed Apr 21 22:11:48 2010 |
rana | Update | PSL | Innolight 2W Vertical Beam Profile |
Back in Gainesville in 1997, I learned how to do this using the chopper wheel. We had to make the assumption that the wheel's blade was moving horizontally during the time of the chop.
One advantage is that the repetitive slices reduces the random errors by a lot - you can trigger the scope and average. Another advantage is that you can download the average scope trace using USB, floppy, or ethernet instead of pencil and paper.
But, I never analyzed it in enough detail to see if there was some kind of nasty systematic error. |
2830
|
Wed Apr 21 23:35:37 2010 |
Koji | Update | PSL | Innolight 2W Vertical Beam Profile |
Good fit. I assumed sqrt(x) is a typo of sqrt(2).
Quote: |
Koji and Kevin measured the vertical beam profile of the Innolight 2W laser at one point.
This data was taken with the laser crystal temperature at 25.04°C and the injection current at 2.092A.
The distance from the razor blade to the flat black face on the front of the laser was 13.2cm.
The data was fit to the function y(x)=a*erf(sqrt(x)*(x-x0)/w)+b with the following results.
Reduced chi squared = 14.07
x0 = (1.964 +- 0.002) mm
w = (0.216 +- 0.004) mm
a = (3.39 +- 0.03) V
b = (3.46 +- 0.03) V
|
|
2834
|
Thu Apr 22 21:42:24 2010 |
Alberto | Update | PSL | Innolight 2W Vertical Beam Profile |
What kind of fit did you use? How are the uncertainties in the parameters obtained? |
2837
|
Sat Apr 24 15:05:41 2010 |
Kevin | Update | PSL | 2W Vertical Beam Profile |
The vertical beam profile of the Innolight 2W laser was measured at eight points along the axis of the laser.
These measurements were made with the laser crystal temperature at 25.04°C and the injection current at 2.091A. z is the distance from the razor blade to the flat black face of the front of the laser.
The voltage from a photodiode was measured for the razor at a number of heights. Except for the first two points, one scan was made with the razor moving down and a second scan was made with the razor moving up. This data was fit to
y = a*erf(sqrt(2)*(x-x0)/w) + b with the following results:
z(cm) (±0.1cm) |
w(mm) |
chi^2/ndf |
3.9 |
0.085 ± 0.006 |
77.09 |
6.4 |
0.130 ± 0.004 |
12.93 |
8.8 down |
0.145 ± 0.008 |
66.57 |
8.8 up |
0.147 ± 0.008 |
18.47 |
11.6 down |
0.194 ± 0.010 |
64.16 |
11,6 up |
0.214 ± 0.009 |
27.23 |
14.2 down |
0.177 ± 0.008 |
49.95 |
14.2 up |
0.183 ± 0.007 |
29.85 |
16.6 down |
0.205 ± 0.006 |
18.35 |
16.2 up |
0.203 ± 0.007 |
17.16 |
19.2 down |
0.225 ± 0.007 |
18.92 |
19.2 up |
0.238 ± 0.011 |
25.56 |
21.7 down |
0.292 ± 0.006 |
11.30 |
21.7 up |
0.307 ± 0.008 |
11.85 |
The values for w and its uncertainty were estimated with a weighted average between the two scans for the last six points and all eight points were fit to
w = w0*sqrt(1+(z-z0)2/zR2) with the following results:
chi^2/ndf = 17.88
w0 = (0.07 ± 0.13) mm
z0 = (-27 ± 121) mm
zR = (65 ± 93) mm
It looks like all of the data points were made in the linear region so it is hard to estimate these parameters with reasonable uncertainty. |
Attachment 1: vbp.jpg
|
|
2838
|
Sat Apr 24 15:50:47 2010 |
Koji | Update | PSL | re: 2W Vertical Beam Profile |
1. The vertical axis should start from zero. The horizontal axis should be extended so that it includes the waist. See Zach's plot http://nodus.ligo.caltech.edu:8080/40m/2818
2. Even if you are measuring only the linear region, you can guess w0 and z0, in principle. w0 is determined by the divergence angle (pi w0/lambda) and z0 is determined by the linear profile and w0. Indeed your data have some fluctuation from the linear line. That could cause the fitting prescision to be worse.
3. Probably the biggest reason of the bad fitting would be that you are fitting with three parameters (w0, z0, zR) instead of two (w0, z0). Use the relation ship zR= pi w0^2/lambda. |
2846
|
Mon Apr 26 16:51:37 2010 |
Kevin | Update | PSL | re: 2W Vertical Beam Profile |
I tried Koji's suggestions for improving the fit to the vertical beam profile; however, I could not improve the uncertainties in the fit parameters.
I started retaking the data today with the same laser settings used last time and noticed that the photodiode was saturating. We were using an ND 4.0 neutral density filter on the photodiode. Koji and I noticed that the coating on the filter was reduced in the center and added an additional ND 0.6 filter to the photodiode. This seemed to fix the photodiode saturation.
I think that the photodiode was also saturating to a lesser extent when I took the last set of data. I will take another vertical beam profile tomorrow.
[Edit by KA: Metallic coating started being evaporated and the ND filters reduced their attenuation. We decided to use absorptive one as the first incident filter, and put a thinner one behind. This looked fine.] |
2847
|
Mon Apr 26 17:34:31 2010 |
Koji | Update | PSL | re: 2W Vertical Beam Profile |
Give me the plot of the fit, otherwise I am not convinced.
Quote: |
I tried Koji's suggestions for improving the fit to the vertical beam profile; however, I could not improve the uncertainties in the fit parameters.
|
|
2851
|
Tue Apr 27 15:29:16 2010 |
Kevin | Update | PSL | re: 2W Vertical Beam Profile |
I thought that the micrometer I was using to move the razor through the laser beam was metric; however, it is actually english.
After discovering this mistake, I converted my previous measurements to centimeters and fit the data to
w = sqrt(w0^2+lambda^2*(z-z0)^2/(pi*w0)^2) with the following results:
reduced chi squared = 14.94
z0 = (-4.2 ± 1.9) cm
w0 = (0.013 ± 0.001) cm |
Attachment 1: vbp.jpg
|
|
Attachment 2: vbp_residuals.jpg
|
|
2857
|
Wed Apr 28 14:22:36 2010 |
Kevin | Update | PSL | re: 2W Vertical Beam Profile |
I used the Mathematica CurveFit package that we use in Ph6/7 to make the fits for the beam profile data. I wrote two functions that use CurveFit shown in the attachment to make the fits to the error function and square root. |
Attachment 1: BeamFit.nb.tar
|
2859
|
Wed Apr 28 16:15:02 2010 |
Kevin | Update | PSL | Accelerometer Calibration |
Koji, Steve, and Kevin looked into calibrating the Wilcoxon accelerometers. Once calibrated, the accelerometers will be used to monitor the motion of the PSL table.
We want to use the shaker to shake each accelerometer and monitor the motion with an OSEM. We will make a plate to attach an accelerometer to the shaker. A flag will also be mounted on this plate.The OSEM will be mounted on the table next to the shaker and positioned so that the flag can block the LED light as the plate moves up and down. We will then measure the motion of the accelerometer as it is shaken from the OSEM signal. The OSEM signal will be calibrated by keeping the plate and the flag still and moving the OSEM down along the flag a known distance with a micrometer. |
2873
|
Mon May 3 17:49:41 2010 |
rana | Configuration | PSL | RC Temperature Servo Turned OFF temporarily |
Quote: |
In order to measure the transfer function of the RC cavity's foam, I've turned off the servo so that the room temperature noise can excite it.
The attached plot shows a step response test from 2 weeks ago. Servo is nominally still working fine.
|
I've just now re-enabled the temperature control of the reference cavity can. Trend of the last 8 days is attached. |
Attachment 1: rct.png
|
|
2875
|
Tue May 4 02:28:38 2010 |
rana | Configuration | PSL | RC Temperature Servo Turned OFF temporarily |
My attempt to passively measure the transfer function of the foam failed fantastically.
As it turns out, the room temperature fluctuations inside the PSL box reach the 1 mK/rHz noise floor of the AD590 (or maybe the ADC) at ~1-2 mHz. Everything at higher frequencies is noise.
So to see what the foam is doing we will have to do something smarter - we need a volunteer to disable the RC temperature servo from the EPICS screen and then cycle the PSL table lights every hour in the morning.
We'll then use our knowledge of the Laplace transform to get the TF from the step responses. |
2876
|
Tue May 4 06:32:58 2010 |
alberto | Configuration | PSL | RC Temperature Servo Turned OFF temporarily |
Quote: |
My attempt to passively measure the transfer function of the foam failed fantastically.
As it turns out, the room temperature fluctuations inside the PSL box reach the 1 mK/rHz noise floor of the AD590 (or maybe the ADC) at ~1-2 mHz. Everything at higher frequencies is noise.
So to see what the foam is doing we will have to do something smarter - we need a volunteer to disable the RC temperature servo from the EPICS screen and then cycle the PSL table lights every hour in the morning.
We'll then use our knowledge of the Laplace transform to get the TF from the step responses.
|
more detailed instructions needed.... |
2881
|
Wed May 5 02:37:55 2010 |
rana | Configuration | PSL | RC Temperature Servo Turned OFF temporarily |
Quote: |
more detailed instructions needed....
|
I showed Kiwamu and Alberto how to turn the lights on and off in the PSL. This is why Caltech is such a fine institution: most schools would have TAs delivering this kind of optics instruction.
We've turned off the RC temperature stabilization and the lights will supply the quasi-random heat input to the table and the cavity. Alberto and Kiwamu will be turning the lights on and off at random times.
The attached plot is the spectrum of temperature fluctuations of the room and the vacuum can with no stabilization from this weekend. I think the rolloff above 10 mHz is kind of fake - I had the .SMOO parameter set to 0.99 for both of these channels. I've just now set the .SMOO to 0 for both channels, so we should now see the true ADC or sensor noise level. It should be ~1 mK/rHz. |
Attachment 1: Picture_7.png
|
|
2883
|
Wed May 5 16:58:21 2010 |
Koji | Update | PSL | 2W hooked up to the interlock service |
Ben, Steve, and Koji
Ben came to the 40m and hooked up a cable to the main interlock service.
We have tested the interlock and confirmed it's working.
[Now the laser is approved to be used by persons who signed in the SOP.]
The RC, PMC, and MZ were unlocked during the interlock maneuver.
Now they are relocked. |
2898
|
Fri May 7 21:55:59 2010 |
kiwamu | Update | PSL | remove Mach-Zehnder |
[Koji, Kiwamu]
The Mach-Zehnder on the PSL table was removed.
A path for 166 MHz modulation in the Mach-Zehnder (MZ) was completely removed, the setup for another path remains the same as before.
Also the photo detector and the CCD for the PMC transmittion were moved to behind the PZT mirror of PMC.
Before removing them, we put an aperture in front of the PD for MC REFL so that we can recover the alignment toward MC by using the aperture.
After the removal we tried to re-align the EOM which imposes the sideband of 29MHz for MC.
We eventually got good alignment of 97% transmissivity at the EOM ( the power of the incident beam is 1.193W and trans was 1.160W )
And then we aligned the beam going to MC by guiding the reflected beam to the aperture we put. This was done by using the steering mirrors on the periscope on the corner of the PSL table.
Now MC got locked and is successfully resonating with TEM00.
|
Attachment 1: NO_MachZehnder_s.jpg
|
|
2976
|
Mon May 24 16:34:22 2010 |
Kevin | Update | PSL | ND Filters for 2W Beam Profile |
I tried to measure the beam profile of the 2W laser today but ran into problems with the ND filters. With the measurements I made a few weeks ago, I used a reflective ND 4.0 filter on the PD. The PD started to saturate and Koji and I noticed that a lot of the metallic coating on the filter had been burnt off. Koji told me to use an absorptive ND 4.0 filter in front of a reflective ND 0.6 filter. I tried this today but noticed that a few holes were being burned into the absorptive filter and that the coating on the reflective filter behind it was also being burned off in spots. I don't think we wanted to use a polarizing beam splitter to reduce the power before the PD but I didn't want to ruin any more filters. |
3014
|
Sun May 30 13:26:07 2010 |
rana, kiwamu | Update | PSL | new HIGH-LOW value for PMC_TRANS |
We changed the HIGH/LOW values of the PMC_TRANS.
The edited file was updated on the svn.
Since the PMC_TRANSPD was replaced behind the pzt mirror (see the entry), its nominal value were reduced to something like ~1V from the previous value of ~2V.
In the medm screen C1PSL_PMC.adl the PMC_TRAN always indicated red because the value were low compared with the previous one.
We went to /cvs/cds/caltech/target/c1psl, then edited psl.db
- Here are the new parameters we set up in the file.
grecord(ai,"C1:PSL-PMC_PMCTRANSPD") {
field(LOW,"0.98")
field(LOLO,"0.93")
field(HIGH,"1.15")
field(HIHI,"1.3")
}
- - - -
These values are based on ~4days trend of the PMC_TRAN.
Then we manually updated those numbers by using ezcawrite in order not to reboot C1PSL.
So now it nicely indicates green in the medm screen. |
3016
|
Sun May 30 15:36:22 2010 |
Alberto | Configuration | PSL | IMC periscope shutter |
Two days ago I opened the PSL shutter by switching the switch on the shutter driver. That caused the shutter's switch on the medm screen to work in reversed mode: open meant closed and closed meant open.
I fixed that. Now the medm screen switch state is correct. |
3028
|
Tue Jun 1 20:40:03 2010 |
Koji | Update | PSL | new HIGH-LOW value for PMC_TRANS |
The alarm had kept crying. I reduced the LOW to be 0.90 and the LOLO to be 0.85 both in psl.db and with ezcawrite .
Quote: |
We changed the HIGH/LOW values of the PMC_TRANS.
The edited file was updated on the svn.
Since the PMC_TRANSPD was replaced behind the pzt mirror (see the entry), its nominal value were reduced to something like ~1V from the previous value of ~2V.
In the medm screen C1PSL_PMC.adl the PMC_TRAN always indicated red because the value were low compared with the previous one.
We went to /cvs/cds/caltech/target/c1psl, then edited psl.db
- Here are the new parameters we set up in the file.
grecord(ai,"C1:PSL-PMC_PMCTRANSPD") {
field(LOW,"0.98")
field(LOLO,"0.93")
field(HIGH,"1.15")
field(HIHI,"1.3")
}
- - - -
These values are based on ~4days trend of the PMC_TRAN.
Then we manually updated those numbers by using ezcawrite in order not to reboot C1PSL.
So now it nicely indicates green in the medm screen.
|
|
3030
|
Wed Jun 2 03:24:22 2010 |
Kevin | Update | PSL | 2W Beam Profile |
[Rana, Kiwamu, Kevin]
The Innolight 2W beam profile was measured with the beam scan. A W2-IF-1025-C-1064-45P window was used to reflect a small amount of the main beam. A 5101 VIS mirror was used to direct just the beam reflected from the front surface of the W2 down the table (the beam reflected from the back surface of the W2 hit the optic mount for the mirror). A razor blade beam dump was used to stop the main transmitted beam from the W2. The distance from the laser was measured from the front black face of the laser to the front face of the beam scan (this distance is not the beam path length but was the easiest and most accurate distance to measure). The vertical and horizontal beam widths were measured at 13.5% of the maximum intensity (each measurement was averaged over 100 samples). These widths were divided by 2 to get the vertical and horizontal radii.
The mirror was tilted so that the beam was close to parallel to the table. (The center of the beam fell by approximately 2.1 mm over the 474 mm that the measurement was made in).
The measurement was taken with an injection current of 2.004 A and a laser crystal temperature of 25.04°C.
This data was fit to w = sqrt(w0^2+lambda^2*(x-x0)^2/(pi*w0)^2) with lambda = 1064nm with the following results
For the horizontal beam profile:
reduced chi^2 = 4.0
x0 = (-138 ± 3) mm
w0 = (113.0 ± 0.7) µm
For the vertical beam profile:
reduced chi^2 = 14.9
x0 = (-125 ± 4) mm
w0 = (124.0 ± 1.0) µm
In the following plots, the blue curve is the fit to the vertical beam radius, the purple curve is the fit to the horizontal beam radius, * denotes a data point from the vertical data, and + denotes a data point from the horizontal data. |
Attachment 1: profile.png
|
|
Attachment 2: errors.png
|
|
Attachment 3: Layout.jpg
|
|
3032
|
Wed Jun 2 04:27:02 2010 |
Koji | Update | PSL | 2W Beam Profile |
This is what I already told to Kevin and Rana:
A direct output beam is one of the most difficult measurements for the mode profiling.
I worried about the thermal lensing.
Since most of the laser power goes through the substrate (BK7) of the W2 window, it may induce thermal deformation on the mirror surface.
An UV fused silica window may save the effect as the thermal expansion coefficient is 0.55e-6/K while BK7 has 7.5e-6.
In addition to the thermal deformation issue, the pick-off setup disables us to measure the beam widths near the laser aperture.
I rather prefer to persist on the razor blade then use the pick off between the blade and the PD.
I also confess that the description above came only from my knowledge, and not from any scientific confirmation including any calculation.
If we can confirm the evidence (or no evidence) of the lensing, it is a great addition to my experience.
Quote: |
[Rana, Kiwamu, Kevin]
The Innolight 2W beam profile was measured with the beam scan. A W2-IF-1025-C-1064-45P window was used to reflect a small amount of the main beam. A 5101 VIS mirror was used to direct just the beam reflected from the front surface of the W2 down the table (the beam reflected from the back surface of the W2 hit the optic mount for the mirror). A razor blade beam dump was used to stop the main transmitted beam from the W2. The distance from the laser was measured from the front black face of the laser to the front face of the beam scan (this distance is not the beam path length but was the easiest and most accurate distance to measure). The vertical and horizontal beam widths were measured at 13.5% of the maximum intensity (each measurement was averaged over 100 samples). These widths were divided by 2 to get the vertical and horizontal radii.
The mirror was tilted so that the beam was close to parallel to the table. (The center of the beam fell by approximately 2.1 mm over the 474 mm that the measurement was made in).
The measurement was taken with an injection current of 2.004 A and a laser crystal temperature of 25.04°C.
This data was fit to w = sqrt(w0^2+lambda^2*(x-x0)^2/(pi*w0)^2) with lambda = 1064nm with the following results
For the horizontal beam profile:
reduced chi^2 = 4.0
x0 = (-138 ± 3) mm
w0 = (113.0 ± 0.7) µm
For the vertical beam profile:
reduced chi^2 = 14.9
x0 = (-125 ± 4) mm
w0 = (124.0 ± 1.0) µm
In the following plots, the blue curve is the fit to the vertical beam radius, the purple curve is the fit to the horizontal beam radius, * denotes a data point from the vertical data, and + denotes a data point from the horizontal data.
|
|
3040
|
Wed Jun 2 22:25:39 2010 |
Kevin | Update | PSL | Low Power 2W Beam Profile |
Koji is worried about thermal lensing introducing errors to the measurement of the 2W beam profile so I measured the profile at a lower power.
I used the same setup and methods used to measure the profile at 2W (see entry). This measurement was taken with an injection current of 1.202 A and a laser crystal temperature of 25.05° C. This corresponds to approximately 600 mW (see power measurement).
The data was fit to w = sqrt(w0^2+lambda^2*(x-x0)^2/(pi*w0)^2) with the following results
For the horizontal beam profile:
reduced chi^2 = 2.7
x0 = (-203 ± 3) mm
w0 = (151.3 ± 1.0) µm
For the vertical beam profile:
reduced chi^2 = 6.8
x0 = (-223 ± 6) mm
w0 = (167.5 ± 2.2) µm
In the following plots, the blue curve is the fit to the vertical beam radius, the purple curve is the fit to the horizontal beam radius, * denotes a data point from the vertical data, and + denotes a data point from the horizontal data.
The differences between the beam radii for the low power and high power measurements are
Δw0_horizontal = (38.3 ± 1.2) µm
Δw0_vertical = (43.5 ± 2.4) µm
Thus, the two measurements are not consistent. To determine if the thermal lensing is in the laser itself or due to reflection from the W2 and mirror, we should measure the beam profile again at 2W with a razor blade just before the W2 and a photodiode to measure the intensity of the reflection off of the front surface. If this measurement is consistent with the measurement made with the beam scan, this would suggest that the thermal lensing is in the laser itself and that there are no effects due to reflection from the W2 and mirror. If the measurement is not consistent, we should do the same measurement at low power to compare with the measurement described in this entry.
|
Attachment 1: profile_low.png
|
|
3041
|
Wed Jun 2 22:58:04 2010 |
Kevin | Update | PSL | 2W Second Reflected Beam Profile |
[Koji, Kevin]
The profile of the Innolight 2W was previously measured by measuring the reflected beam from the front surface of a W2 window (see entry). To investigate thermal effects, Rana suggested also measuring the profile of the beam reflected from the back surface of the W2.
I used the same setup and methods as were used in the first measurement. The mirror was moved so that only the beam reflected from the back surface of the W2 was reflected from the mirror. This beam was reflected from both the front of the mirror and the back of the mirror. An extra beam dump was positioned to block the reflection from the back of the mirror.
This measurement was made with 2.004 A injection current and 25.04°C laser crystal temperature.
The data was fit to w = sqrt(w0^2+lambda^2*(x-x0)^2/(pi*w0)^2) with the following results
For the horizontal beam profile:
reduced chi^2 = 5.1
x0 = (-186 ± 6) mm
w0 = (125.8 ± 1.4) µm
For the vertical beam profile:
reduced chi^2 = 14.4
x0 = (-202 ± 11) mm
w0 = (132.5 ± 2.7) µm
In the following plots, the blue curve is the fit to the vertical beam radius, the purple curve is the fit to the horizontal beam radius, * denotes a data point from the vertical data, and + denotes a data point from the horizontal data.
The differences between the beam radii for the beam reflected from the front surface and the beam reflected from the back surface are
Δw0_horizontal = (12.8 ± 1.6) µm
Δw0_vertical = (8.5 ± 2.9) µm
So the two measurements are not consistent. This suggests that the passage through the W2 altered the profile of the beam. |
Attachment 1: profile_2nd.png
|
|
3042
|
Thu Jun 3 00:47:17 2010 |
Kevin | Update | PSL | 2W Beam Profile of Second Reflected Beam |
[Koji, Kevin]
The profile of the Innolight 2W was previously measured by measuring the reflected beam from the front surface of a W2 window (see entry). To investigate thermal effects, Rana suggested also measuring the profile of the beam reflected from the back surface of the W2.
I used the same setup and methods as were used in the first measurement. The mirror was moved so that only the beam reflected from the back surface of the W2 was reflected from the mirror. This beam was reflected from both the front of the mirror and the back of the mirror. An extra beam dump was positioned to block the reflection from the back of the mirror.
This measurement was made with 2.004 A injection current and 25.04°C laser crystal temperature.
The data was fit to w = sqrt(w0^2+lambda^2*(x-x0)^2/(pi*w0)^2) with the following results
For the horizontal beam profile:
reduced chi^2 = 5.1
x0 = (-186 ± 6) mm
w0 = (125.8 ± 1.4) µm
For the vertical beam profile:
reduced chi^2 = 14.4
x0 = (-202 ± 11) mm
w0 = (132.5 ± 2.7) µm
In the following plots, the blue curve is the fit to the vertical beam radius, the purple curve is the fit to the horizontal beam radius, * denotes a data point from the vertical data, and + denotes a data point from the horizontal data. |
Attachment 1: profile_2nd.png
|
|
3109
|
Wed Jun 23 18:05:00 2010 |
Koji | Configuration | PSL | FSS SLOWDC should be ~-4.0 |
FSS SLOWDC slider is at around 0.
Please someone relock this at ~-4.0 to exploit some last juice of the fruit.
See this entry for the details of the operating point.
|
Attachment 1: C1PSL_FSS.png
|
|
3110
|
Wed Jun 23 23:08:30 2010 |
rana | Configuration | PSL | FSS SLOWDC should be ~-4.0 |
|
3163
|
Wed Jul 7 00:15:29 2010 |
tara,Rana | Summary | PSL | power spectral density from RefCav transmitted beam |
I measured the RC transmitted light signals here at the 40m. I made all connections through the PSL patch panel.
Other than two steering mirrors in front of the periscope, and the steering mirror for the RFPD which were used to steer
the beam into the cavity and the RFPD respectively, no optics are adjusted.
We re-aligned the beam into the cavity (the DC level increased from 2 V to 3.83V) (Fig2) (We could not recover the power back to what it was 90 days ago)
and the reflected beam to the center of the RFPD.
I measured the spectral density of the signal of the transmitted beam behind RefCav in both time and frequency domain.
This will be compared with the result from PSL lab later, so I can see how stable the signal should be.
I did not convert Vrms/rtHz to Hz/rtHz because I only look at the relative intensity of the transmitted beam which will be compared to the setup at PSL lab.
We care about this power fluctuation because we plan to measure
photo refractive noise on the cavity's mirros
(this is the noise caused by dn/dT in the coatings and the substrate,
the absorption from fluctuating power on the coating/mirror changes
the temperature which eventually changes the effective length of the cavity as seen by the laser.)
The plan is to modulate the power of the beam going into the cavity,
the absorption from ac part will induce frequency noise which we want to see.
Since the transmitted power of the cavity is proportional to the power inside the cavity.
Fluctuations from other factors, for example, gain setting, will limit our measurement.
That's why we are concerned about the stability of the transmitted beam and made this measurement.
|
Attachment 1: RIN_rftrans.png
|
|
Attachment 2: tara.png
|
|
3164
|
Wed Jul 7 10:42:29 2010 |
Koji | Summary | PSL | power spectral density from RefCav transmitted beam |
How do you calibrate this to Hz/rtHz?
Quote: |
I measured the RC transmitted light signals here at the 40m. I made all connections through the PSL patch panel. No optics/PD were touched.
I measured the spectral density of the signal of the transmitted beam behind RefCav in both time and frequency domain.
This will be compared with the result from PSL lab later, so I can see how stable the signal should be.
We re-aligned the beam into the cavity (the DC level increased from 2 V to 3.83V)
and the reflected beam to the center of the RFPD.
|
|
3189
|
Fri Jul 9 20:16:19 2010 |
rana | Summary | PSL | Things I did to the PSL today: Refcav, PMC, cameras, etc. |
I re-aligned the beam into the PMC. I got basically no improvement. So I instead changed the .LOW setting so that PMCTRANS would no longer go yellow and make the donkey sound.
I did the same for the MOPA's AMPMON because its decayed state is now nominal.
Steve and I removed the thermal insulation from around the reference cavity vacuum chamber. It wasn't really any good anyways.
Here are the denuded photos:
Steve and I are now planning to replace the foam with some good foam, but before that we will wrap the RC chamber with copper sheets like you would wrap a filet mignon with applewood bacon.
This should reduce the thermal gradients across the can. We will then mount the sensors directly to the copper sheet using thermal epoxy. We will also use copper to cover most of this hugely
oversized window flange - we only need a ~1" hole to get the 0.3 mm beam out of there.
My hope is that all of this will improve the temperature stability of this cavity. Right now the daily frequency fluctuations of the NPRO (locked to the RC) are ~100 MHz. This implies
that the cavity dT = (100 MHz) / (299792458 / 1064e-9) / (5e-7) = 1 deg. That's sad....
I also changed the RC_REFL cam to manual gain from AGC. I cranked it to max gain so that we can see the REFL image better. |
3190
|
Sun Jul 11 20:11:48 2010 |
rana | Summary | PSL | RC trend after the insulation removal |

|
3196
|
Mon Jul 12 14:22:36 2010 |
Jenne | Summary | PSL | Things I did to the PSL today: Refcav, PMC, cameras, etc. |
Quote: |
I re-aligned the beam into the PMC. I got basically no improvement. So I instead changed the .LOW setting so that PMCTRANS would no longer go yellow and make the donkey sound.
I did the same for the MOPA's AMPMON because its decayed state is now nominal.
|
[Jenne, Chip]
The alarm was still going, because the LOLO setting was higher than the LOW, which is a little bit silly. So we changed the .LOLO setting to 0.80 (the LOW was set to 0.82)
We also changed psl.db to reflect these values, so that they'll be in there the next time c1psl gets rebooted. |
3210
|
Tue Jul 13 21:04:49 2010 |
tara,rana | Summary | PSL | Transfer function of FSS servo |
I measured FSS's open loop transfer function.
For FSS servo schematic, see D040105-B.
4395A's source out is connected to Test point 2 on the patch panel.
Test Point 2 is enabled by FSS medm screen.
"A" channel is connected to In1, on the patch panel.
"R" channel is connected to In2, on the patch panel.
the plot shows signal from A/R.
Note that the magnitude has not been corrected for the impedance match yet.
So the real UGF will be different from the plot.
-------------------------
4395A setup
-------------------------
network analyzer mode
frequency span 1k - 10MHz
Intermediate frequency bandwidth 100Hz
Attenuator: 0 for both channels
Source out power: -30 dBm
sweep log frequency
------------------------------
medm screen setup
-----------------------------
TP2: enabled
Common gain -4.8 dB
Fast Gain 16 dB |
Attachment 1: TF_FSS_ser.png
|
|
3232
|
Thu Jul 15 19:27:04 2010 |
rana | Summary | PSL | RC trend after the insulation removal |

As you can see, there was not much (if any) worsening of the laser frequency fluctuation from removing the RefCav insulation. The plots below are zooomed in:
 
I have used the same peak-to-peak scale so that its easy to compare the fluctuations before (LEFT) and after (RIGHT).
As you can clearly see, the laser frequency moves just as much now (the SLOW_DC) as it did before when it had the insulation. Only now the apparent (i.e. fake) RC temperature fluctuations are much larger. So this sensor is fairly useless as configured. |
3240
|
Fri Jul 16 20:25:52 2010 |
Megan | Update | PSL | Reference Cavity Insulation |
Rana and I
1) took the temperature sensors off the reference cavity;
2) wrapped copper foil around the cavity (during which I learned it is REALLY easy to cut hands with the foil);
3) wrapped electrical tape around the power terminals of the temperature sensors (color-coded, too! Red for the out of loop sensor, Blue for the first one, Brown for the second, Gray for the third, and Violet for the fourth. Yes, we went with an alphabetical coding system, excluding the out of loop sensor);
4) re-wrapped the thermal blanket heater;
5) covered the ends of the cavities with copper, ensuring that the beam can enter and exit;
6) took pretty pictures for your enjoyment!
We will see if this helps the temperature stabilization of the reference cavity.

The end of the reference cavity, with a lovely square around the beam.

The entire, well-wrapped reference cavity! |
3241
|
Fri Jul 16 23:53:27 2010 |
Rana | Update | PSL | Reference Cavity Insulation |
From the trend, it seems that the Reference Cavity's temperature servo is working fine with the new copper foil. I was unable to find the insulating foam anywhere, but that's OK. We'll just get Frank to make us a new insulation with his special yellow stuff.
The copper foil that Steve got is just the right thickness for making it easy to form around the vacuum can, but we just have to have the patience to wrap ~5-10 more layers on there. We also have to get a new heater jacket; this one barely fits around the outside of the copper wrap. The one we have now seems to have a good heating wire pattern, but I don't know where we can buy these.
I also turned the HEPA's Variac back down to the nominal value of 20. Please remember to turn it back up to 100 before working on the PSL. |
3260
|
Wed Jul 21 15:43:38 2010 |
Megan | Summary | PSL | Copper Layer Thickness on the Reference Cavity |
Using the equation for thermal resistance
Rthermal = L/(k*A)
where k is the thermal conductivity of a material, L is the length, and A is the surface area through which the heat passes, I could find the thermal resistance of the copper and stainless steel on the reference cavity. To reduce temperature gradients across the vacuum chamber, the thermal resistance of the copper must be the same or less than that of the stainless steel. Since the copper is directly on top of the stainless steel, the length and width will be the same for both, just the thickness will be different (for ease of calculation, I assumed flat, rectangular strips of the metal). Assuming we wish to have a thermal resistance of the copper n times less than that of the stainless steel, we have
RCu = RSS/n
or
L/(kCu*w*tCu) = L/(kSS*w*tSS*n)
so that
tCu/tSS = n*kSS/kCu
We know that kSS = 401 W/m*K and KCu = 16 W/m*K, so
tCu/tSS = 0.0399*n
By using the drawings for the short reference cavity vacuum chamber (the only one I could find drawings for online) I found a thickness of the walls of 0.12 in or 0.3048 cm. So for the same thermal resistance in both metals, the copper must be 0.0122 cm thick and for a thermal resistance 10 times less, it must be 0.122 cm thick. So we will have to keep wrapping the copper on the vacuum chamber! |
3268
|
Thu Jul 22 14:07:20 2010 |
kiwamu | Update | PSL | PSL front end machine |
It looks like something wrong happened around the PSL front end. One of the PSL channel, C1:PSL-PMC_LOCALC, got crazy.
We found it by the donkey alarm 10 minutes ago.
The attached picture is a screen shot of the PMC medm screen.
The value of C1:PSL-PMC_LOCALC ( middle left on the picture ) shows wired characters. It returns "nan" when we do ezcaread.
Joe went to the rack and powered off / on the crate, but it still remains the same. It might be an analog issue (?) |
Attachment 1: PSL-PMC2010-07-22.png
|
|
3269
|
Thu Jul 22 15:59:29 2010 |
Alberto | Update | PSL | PSL front end machine |
Quote: |
It looks like something wrong happened around the PSL front end. One of the PSL channel, C1:PSL-PMC_LOCALC, got crazy.
We found it by the donkey alarm 10 minutes ago.
The attached picture is a screen shot of the PMC medm screen.
The value of C1:PSL-PMC_LOCALC ( middle left on the picture ) shows wired characters. It returns "nan" when we do ezcaread.
Joe went to the rack and powered off / on the crate, but it still remains the same. It might be an analog issue (?)
|
The problem seems to be a software one.
In any case, Kiwamu and I looked at the at the PMC crystal board and demod board, in search of a possible bad connection. We found a weak connection of the RG cable going into the PD input of the demod board. The cable was bent and almost broken.
I replaced the SMA connector of the cable with a new one that I soldered in situ. Then I made sure that the connection was good and didn't have any short due to the soldering. |
3270
|
Thu Jul 22 18:18:54 2010 |
Alberto | Update | PSL | Problem Solved |
Quote: |
Quote: |
It looks like something wrong happened around the PSL front end. One of the PSL channel, C1:PSL-PMC_LOCALC, got crazy.
We found it by the donkey alarm 10 minutes ago.
The attached picture is a screen shot of the PMC medm screen.
The value of C1:PSL-PMC_LOCALC ( middle left on the picture ) shows wired characters. It returns "nan" when we do ezcaread.
Joe went to the rack and powered off / on the crate, but it still remains the same. It might be an analog issue (?)
|
The problem seems to be a software one.
In any case, Kiwamu and I looked at the at the PMC crystal board and demod board, in search of a possible bad connection. We found a weak connection of the RG cable going into the PD input of the demod board. The cable was bent and almost broken.
I replaced the SMA connector of the cable with a new one that I soldered in situ. Then I made sure that the connection was good and didn't have any short due to the soldering.
|
[Alberto, Koji]
By looking at the reference pictures of the rack in the wiki, it turned out that the Sorensen which provides the 10V to the 1Y1 rack was on halt (red light on). It had been like that since 1.30pm today. It might have probably got disabled by a short somewhere or inadvertently by someone working nearby it.
Turning it off and on reset it. The crazy LO calibrated amplitude on the PMC screen got fixed.
Then it was again possible to lock PMC and FSS.
We also had to burtrestore the PSL computer becasue of the several reboots done on it today. |
3271
|
Fri Jul 23 00:13:11 2010 |
rana | Update | PSL | Problem NOT REALLY Solved |
So...who was working around the PSL rack this morning and afternoon? Looks like there was some VCO phase noise work at the bottom of
the rack as well as some disconnecting of the Guralp cables from that rack. Who did which when and who needs to be punished? |