  12334   Mon Jul 25 17:01:54 2016 SteveUpdateSUSETMX suspension - progressITMY side  : Magnet od 1.9 mm so wire
to magnet gap ~ 0.2-0.3 mm

 
 ITMYside2011.JPG 
  12335   Mon Jul 25 20:37:46 2016 gautamUpdateGeneralETMX suspension - Side OSEM test[Lydia, gautam]

Summary: We
did some preliminary tests to check if at
 osem_test_2016_07_25.png 
  12336   Tue Jul 26 09:56:34 2016 ericqUpdateCDSc1susaux restartedc1susaux (which controls watchdogs and
alignments for all non-ETM optics) was down,
the last BURT was done yesterday around 2PM. 
  
  12340   Tue Jul 26 18:07:57 2016 gautamUpdateGeneralETMX suspension - Further OSEM TestsToday, we did the following:


Once again, inserted all
four face OSEMs till the sensor voltage readouts
 OSEM_spectra.pdf 
  12344   Wed Jul 27 22:42:00 2016 PrafulUpdateElectronicsEM172 AmplifierI recreated Den's microphone amplifier
circuit on a solderless breadboard to test
it and make sure it does what it's supposed
 schematic.png 
  12345   Thu Jul 28 03:22:28 2016 ericqUpdateGeneralETMX magnets re-reglued[gautam, ericq]

Brief summary, some pictures and
such follow in the daytime. 
  
  12346   Thu Jul 28 08:42:36 2016 SteveUpdateSUS wire clampThe clamp is in the machine for milling
off the grooves. It's condition is normal.
The edges needs some 800 sand paper so it
 e12342.jpg 
  12347   Thu Jul 28 13:49:44 2016 gautamUpdateGeneralETMX magnets re-reglued


 ETMX_sideMagnet.JPGETMX_sideMagnet_zoomed.JPG 
  12348   Thu Jul 28 16:43:01 2016 LydiaUpdateGeneralETMX aluminium standoff groove condition 

I took some pictures with the digital
microscope of the aluminum standoffs
 wire_damage_4.jpgwire_damage_zoom.jpgno_wire.jpgcomparison_2.jpg 
  12349   Thu Jul 28 17:48:17 2016 gautamUpdateGeneralETMY LR magnet broke offWhile ETMX magnets were curing, I wanted
to try and suspend ETMY in the endchamber,
put in the OSEMS and see if the magnets aligned
  
  12350   Thu Jul 28 22:43:03 2016 ranaUpdateGeneralETMY UL magnet broke offI'd recommend replacing the wire and
grinding down the clamp to prevent cutting
the wire. Since we have almost never replaced
  
  12351   Fri Jul 29 03:32:30 2016 ericqUpdateGeneralETMX suspended once more[gautam, ericq]

Tonight's progress on ETMX:


All pickle pickers removed
  
  12353   Fri Jul 29 03:59:55 2016 ericqUpdateSUSIn air OSEM diagonalization checkThe question arose whether we can get good
enough data to diagonize our OSEM sensing
matrices in air. 
 BS_OSEM_diag.pdf 
  12354   Fri Jul 29 13:17:34 2016 KojiUpdateGeneralOvenWhile the air bake oven situation is being
improved, how about to buy a cheepo toaster
oven at Target, BestBuy, or anywhere?
  
  12355   Fri Jul 29 15:17:23 2016 SteveUpdateSUSwire and clamping testUnbaked steel music wire from  "Ca
Fine Wire Co" from 24" od spool,
od 0.0017" used. Identical to the one
 43micr242grs.jpg 
  12356   Fri Jul 29 19:37:43 2016 PrafulUpdateElectronicsMic AmplifierI set up a test inverting amplifier circuit
using the LT1677 opamp:

 7x 
  12357   Fri Jul 29 20:12:54 2016 gautamUpdateSUSthird time - no luck[lydia, gautam]

Summary:
  
  12359   Mon Aug 1 14:09:20 2016 SteveUpdateSUSthird time - no luckGautam and Steve,

The clamp's left side was jammed
onto the left guide pin. It was installed
  
  12360   Mon Aug 1 18:50:29 2016 AakashUpdateGeneralAcromag Setup | SURF2016There were many unknown and unsolved problems
with using modbusApp for linux-arm architecture.
So I tried to install the necessary files
 7x 
  12361   Mon Aug 1 20:09:37 2016 ranaUpdateCDSDAFI UpdateI found the DAFI screen as a button inside
of the LSC screen - I think its more logically
found from the sitemap, so I'll move
  
  12362   Wed Aug 3 00:15:39 2016 gautamUpdateSUSETMX suspended[lydia, steve, ericq, gautam]

Summary:
  
  12363   Wed Aug 3 09:26:54 2016 LydiaUpdateSUSETMX suspended: photosHere are the photos we took showing the
magnet positions in the OSEMs, and others
showing the positions of the wire and unglued
 8x 
  12364   Wed Aug 3 10:52:03 2016 SteveUpdateGeneralETMY soaking dishETMY UL epoxy soaking dish. All teflon
in glass.

 
 SD1.jpgSD2.jpgSD3.jpg 
  12366   Wed Aug 3 15:35:19 2016 AakashUpdateGeneralAcromag Setup | SURF2016Lydia helped me to troubleshoot the Accromag
connection problems which I was facing previously. 
If power goes off/turned off manually, the
  
  12367   Wed Aug 3 15:36:57 2016 SteveUpdateSUSNi plated magnets & epoxy orderedNi plated SmCo magnets with specification
of LIGO-C1103521-v2 for SOS ordered from Electron
  
  12368   Wed Aug 3 16:34:59 2016 LydiaUpdateGeneralAcromag Setup | SURF2016Actually, if the power goes off and back
on, the ethernet connection comes back online
after about 5 seconds, or faster if it is
  
  12369   Wed Aug 3 18:53:46 2016 PrafulUpdateElectronicsMic AmplifierI could not get Den's circuit to work
for some reason with microphone input, so
I decided to try to use another circuit I
 simple_amp.pngampTest_03-08-2016_180448.pdfampTest_03-08-2016_180448.txtsimple_amp.png 
  12370   Thu Aug 4 03:14:39 2016 gautamUpdateSUSETMX suspended


  
  12371   Thu Aug 4 10:57:58 2016 ranaUpdateComputer Scripts / ProgramsNODUS update / restarts underwayUsual Ubuntu apt-get upgrades; long delayed
but now happening.
  
  12372   Thu Aug 4 14:21:21 2016 ericq UpdateComputer Scripts / ProgramsWeb things mostly back onlineThe nodus restart caused a bit of downtime.
The apache configuration files were accidentally
deleted the other day, so elog/svn/wikis
  
  12373   Thu Aug 4 15:00:40 2016 ericq UpdateComputer Scripts / ProgramsWeb things mostly back onlineNodus' /export
and /etc
directories are now being backed up at
  
  12374   Thu Aug 4 17:29:17 2016 PrafulUpdateGeneralGuralp CableThe Guralp cable has been pulled and put
in the corner to the left of the water cooler:

 
 13932028_1768980090054533_2094690556_o.jpg 
  12375   Thu Aug 4 17:41:53 2016 KojiUpdateComputer Scripts / ProgramsWeb things mostly back onlineSorry I was writting the elog, but I had
to dive into the chamber (@LHO) before completion.
  
  12377   Fri Aug 5 02:17:10 2016 gautamUpdateSUSETMX Ruby Wire Standoff Ready for GluingPart
1: Rotation of optic


 8x 
  12378   Fri Aug 5 04:43:09 2016 KojiUpdateSUSETMX Ruby Wire Standoff Ready for GluingIf only the LL magnet looks too low, doesn't
this mean that the OSEMs are not arranged
in a square shape?
  
  12379   Fri Aug 5 09:38:12 2016 SteveUpdateSUSSOS sus wire ordered0.0017" OD., 500ft steel music wire
ordered. Pictures of the existing roll are
below. It will be on 8" OD. spool too.
 0.0017.jpg0.0017spec.jpg 
  12380   Fri Aug 5 16:25:08 2016 PrafulUpdateElectronicsMic AmplifierI took the spectrum of an EM172 connected
to my amplifier inside and outside a large
box filled with foam layers:
 mic_box.pngspectra.png 
  12381   Fri Aug 5 18:21:28 2016 gautamUpdateSUSETMX Ruby Wire Standoff GLUED
The latest twist in this apparently
never-ending saga was that even though fine
pitch balancing was achieved, the wire was
 11x 
  12382   Sun Aug 7 14:53:39 2016 ericq UpdateSUSETMX Standoff gluing was successfulI came in to check on ETMX. I freed the
earthquake stops, and found that the OSEMS
were reasonably, but not perfectly, centered.
 ETMX_resonances.pdfETMX_SUSspectra.zip 
  12383   Mon Aug 8 10:35:12 2016 SteveUpdatePEMvent 78 day 38 




 Vent38days.pngvent78d40RgaBg.png 
  12384   Tue Aug 9 00:44:43 2016 gautam UpdateSUSETMY patch-upSummary:

Given that ETMX looks to be in
good shape and the optic and suspension tower
  
  12385   Tue Aug 9 13:53:57 2016 babbottUpdateSEIlong Guralp EX cable repaired on the D-sub sideI checked out the cable that I
took from you, and all of the connections
  
  12386   Tue Aug 9 15:27:57 2016 gautam UpdateSUSETMY patch-upThe pickle
pickers came off nicely and both magnets
seem to be glued on okay. The alignment of
 IMG_2996.JPGIMG_2997.JPG 
  12387   Tue Aug 9 15:50:30 2016 PrafulUpdateGeneralGuralp CableThe Guralp cable has been reconnected and
powered after having the connector changed
out.
  
  12388   Tue Aug 9 16:19:27 2016 SteveUpdateSUStorque driver for wire clampNew
Wiha 28504 torque wrench for SOS wire
clamping. It's range 7.5  - 20 in-lb
  
  12389   Tue Aug 9 19:35:49 2016 ranaUpdateSEIboth Guralp seismometers are functioning and being acquiredAfter
some cable swapping, we now have both Guralp
seismometers running and the times series
 1goodday.png 
  12390   Wed Aug 10 03:08:03 2016 gautam UpdateSUSETMY patch-up[lydia, gautam]

Rana felt it was alright to use
the wire clamp and suspension cage in its
  
  12391   Wed Aug 10 10:32:44 2016 SteveUpdateSEIworking Guralps as they areThe saga has started
here  We have to give credit to
the Boss who fixed it. The seismometers themself
 6x 
  12392   Wed Aug 10 15:34:24 2016 SteveUpdateSUS6 in-lbs torque driver for wire clamp screwThe 7.5 in-lb of Wiha seems at the upper
end of torque range for a 4-40 SS screw

Wiha 28502 ordered with range 5
  
  12393   Wed Aug 10 17:20:16 2016 LydiaUpdateGeneralSeismometer channel names changed[ericq, Lydia]

We
changed the seismometer channel names from,
  
ELOG V3.1.3-