40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 255 of 348  Not logged in ELOG logo
ID Date Author Type Categoryup Subject
  7542   Sun Oct 14 17:26:03 2012 DenUpdatePEMETMX table

Quote:

 I've measured ETMX table motion compared to ground motion using accelerometers. Data and settings in the xml file are at the svn directory 40m_seismic/etmx

 High frequency (>60 Hz) resonances that are present at the ETMX motion spectrum seem to be understandable. Amplification ETMX/GROUND of a factor of 2 at 1 Hz is interesting. I've monitored ACC DQ channels for a few hours and noticed that usually spectrum looks like in the previous elog. But every ~40 min ETMX motion is much higher then ground motion at low frequencies (<5 Hz). I wonder if this a reaction of a table to outside disturbances or accelerometer issue.

 

Attachment 1: etmx.pdf
etmx.pdf etmx.pdf etmx.pdf etmx.pdf
  7543   Sun Oct 14 20:51:20 2012 DenUpdatePEMAA board

Quote:

 

 But every ~40 min ETMX motion is much higher then ground motion at low frequencies (<5 Hz). I wonder if this a reaction of a table to outside disturbances or accelerometer issue. 

 This could come from AA board, its range is +/- 2.5 V, RMS of the ETMX table motion is a few times higher then ground motion, so ETMX accelerometer signal was corrupted.

acc.png

As this small AA range has already caused problems before, I decided to increase it. I've looked through the board scheme and found that all its differential line receives and output amplifiers have absolute maximum range of 40V. We used KEPKO power supply for this board with a voltage range up to 6 V. So I've replaced it with a BK PRECISION power supply and set it to +/- 15 V. Now AA board range is 7.5 V.

DSC_4747.JPG 

I'll leave accelerometers near ETMX table. It's interesting to measure table motion in the morning when trucks drive by.

  7546   Mon Oct 15 10:14:22 2012 DenUpdatePEMAA board

Quote:

 

 I'll leave accelerometers near ETMX table. It's interesting to measure table motion in the morning when trucks drive by.

 That low frequency effect was due to AA board, now it is gone.

  7562   Tue Oct 16 21:57:47 2012 DenUpdatePEMaccelerometers

All accelerometers are now at the table behind 1X4, cables are near readout box.

  7567   Wed Oct 17 09:01:37 2012 SteveUpdatePEMants on the PSL table

Quote:

We observed one or two ants climbing over PMC optics without booties and safety glasses.

The floor was mopped with strong Bayer Home Pest Control solution in the Vertex area.

Do not work inside the 40m lab if you are sensitive to chemicals!

 

 The floor is mopped again with strong PEST CONTROL SOLUTION in water in the Vertex area.

Do not plane to work in the IFO-room till noon if you are sensitive to chemicals!

  7626   Thu Oct 25 21:02:34 2012 DenUpdatePEM1x7 dc power

 We now stop using bench DC power supplies for microphone preamp and PEM AA board. DC power is wired from 1x5 rack suppliers. I've installed a beam to mount fuse houses in the 1x7 as we did not have one.

DSC_4779.JPG

  7645   Wed Oct 31 14:31:34 2012 SteveUpdatePEMhigh particle count

High particle count confirmed with #2 counter

Attachment 1: highparticlec.png
highparticlec.png
  7650   Wed Oct 31 22:56:41 2012 DenUpdatePEMacoustic noise

Microphone preamp box had a low-pass filter at 2kHz, Ayaka changed it to 20 kHz by replacing 100pF capacitor with a 10pF.

We've measured frequency response of the box. Signal from the microphone was split into two. One path went to the box, while another was amplified by the gain 20 (and bandpass filter 1Hz - 300kHz) and sent to spectrum analyzer. Coherence and frequency response were measured using box output and amplified input. Low-pass filter in the box does not limit our sensitivity.

Acoustic noise significantly decreases at frequencies higher then 2kHz. So we need to modify the circuit by adding whitening filter.

pre_after_mic.png  freq_resp.png

I've plugged in PMC length channel into PEM board CH15 through and amplifier (gain=200) that is AC coupled to avoid ~2.5 DC V coming from PMC servo.  I measured coherence with microphone that was located ~30 cm higher. Measurements show contribution of acoustic noise to PMC length in the frequency range 20-50 Hz. In this range PMC length / MC length coherence is ~0.5.

Acoustic noise couples to PMC length in a non-stationary way. 5 minutes after the first measurement I already see much higher contribution. This was already discussed here. I've made C1:X02-MADC3_TP_CH15 a DQ channel at 64kHz. This a fast PMC length channel.

Next step will be to use several microphones located around PMC for acoustic noise cancellation.

DSC_4792.JPG    DSC_4793.JPG

Attachment 3: pmc.pdf
pmc.pdf pmc.pdf
Attachment 4: pmc_high.pdf
pmc_high.pdf
  7680   Wed Nov 7 11:43:29 2012 SteveUpdatePEMhigh particle count ALART

The BS camber is open only. We should close ASAP

Outside air quality is 1.7- 2.2  million particles  / cf min of 0.5 micron

Attachment 1: highpartcount.png
highpartcount.png
  7681   Wed Nov 7 14:59:35 2012 SteveUpdatePEMhigh particle count ALART & oplevs

Quote:

The BS camber is open only. We should close ASAP

Outside air quality is 1.7- 2.2  million particles  / cf min of 0.5 micron

 Air is still bad and the chambers are closed. Before lunch  Jamie repointed the PRM oplev. Manasa and I reset oplevs: BS and ITMX.

ETMX and ETMY are fine.

SRM and ITMY oplevs needs more work.

Attachment 1: oplevs.png
oplevs.png
  7733   Tue Nov 20 15:24:52 2012 AyakaUpdatePEMSeismometers and a microphone

 I got two seismometers and one microphone back from Tara.

They are now near the Gurlap under the MC.

  7734   Tue Nov 20 16:03:37 2012 janoschUpdatePEMSeismometers and a microphone

Quote:

 I got two seismometers and one microphone back from Tara.

They are now near the Gurlap under the MC.

 If anybody wants a fancy single-axis seismometer for a while (GS-13), then please let me know.

  7744   Mon Nov 26 10:58:20 2012 SteveUpdatePEMair cond maitenance tomorrow

Air conditioning maintenance is scheduled for tomorrow morning till noon.

  7745   Mon Nov 26 18:36:17 2012 JenneUpdatePEMBLRMS back

Quote:

 I got two seismometers and one microphone back from Tara.

They are now near the Gurlap under the MC.

 I have finally plugged GUR1 back in....it is down at ETMY for now, since that's where the cable was.  BLRMS are back up on the projector.

  7754   Tue Nov 27 11:27:22 2012 SteveUpdatePEMparticle count is going up

Jeff is still working on the filter banks.
Attachment 1: 5yparticles.png
5yparticles.png
Attachment 2: particle_6m_6m.png
particle_6m_6m.png
Attachment 3: 1y_particlec.png
1y_particlec.png
Attachment 4: 10y_particle_c.png
10y_particle_c.png
  7759   Wed Nov 28 23:18:35 2012 CharlesUpdatePEMDecreased RMS in Seismometers

The attached plots display RMS noise from various accelerometers and seismometers over the past 90 days. One can see how after the reinstallation of the seismometers in November, RMS from the GUR1Z and GUR1X channels decreases by a factor of about 100 from data in August. Additionally, the RMS over the course of the last 90 days has notably decreased in all instruments. In many cases, the RMS is only the result of inherent electronics noise, rather than from a signal.

Attachment 1: 8-31_11-29_PEM-RMS.jpg
8-31_11-29_PEM-RMS.jpg
  7762   Thu Nov 29 02:43:48 2012 AyakaUpdatePEMDecreased RMS in Seismometers

Quote:

The attached plots display RMS noise from various accelerometers and seismometers over the past 90 days. One can see how after the reinstallation of the seismometers in November, RMS from the GUR1Z and GUR1X channels decreases by a factor of about 100 from data in August. Additionally, the RMS over the course of the last 90 days has notably decreased in all instruments. In many cases, the RMS is only the result of inherent electronics noise, rather than from a signal.

The Image is replaced

[Den, Ayaka]

We found that seismometer was working and the calibration in the filter banks should have been wrong.
We turned off the all FM2 filter in RMS filter banks.

We also installed STS seismometer. It is under the BS. Now we have spectrum of three seismometers.
GUR1Xfilterbank.pngseismometers1129.pdf


 

RA: the above plot is kind of unreadable and useless. Please replace with something legible and put in some words about why there is a wrong filter, what exactly it is, etc., etc. etc. And why would you leave in a filter which is not supposed to be on? We might as well leave a few secretly broken chairs in the control room...

Attachment 2: seismometers1129.jpg
seismometers1129.jpg
  7763   Thu Nov 29 09:58:06 2012 DenUpdatePEMDecreased RMS in Seismometers

Quote:

[Den, Ayaka]

We found that seismometer was working and the calibration in the filter banks should have been wrong.
We turned off the all FM2 filter in RMS filter banks.
 

We also installed STS seismometer. It is under the BS. Now we have spectrum of three seismometers;
 

RA: the above plot is kind of unreadable and useless. Please replace with something legible and put in some words about why there is a wrong filter, what exactly it is, etc., etc. etc. And why would you leave in a filter which is not supposed to be on? We might as well leave a few secretly broken chairs in the control room...

 First of all, STS-2 is in the end of X arm, GUR2 is under BS, GUR1 is in the end of Y arm.

BLRMS were small because we applied calibration from counts to um/s two times. In the past we had calibration in the RMS BP filter bank (vel2vel = FM2). Now we have calibration in the seismometer input filter bank so we can save calibrated _OUT channels.

  7777   Mon Dec 3 16:37:09 2012 SteveUpdatePEMEM 172 microphones ordered

Quote:

I've put EM 172 microphones inside Steve's isolation box to measure their noise. I've attached mics to each other and aligned them using the tape.

At low frequencies (below 1 Hz) the noise is limited by ADC as there is a 10 Hz high-pass filter inside mic readout box.

ADC noise is measured by splitting the signal from 1 mic into 2 ADC channels.

em172.png

 BT EM172 microphones  are ordered.

  7780   Tue Dec 4 10:44:36 2012 SteveUpdatePEMshort power outage

 

CALIFORNIA INSTITUTE OF TECHNOLOGY

                 FACILITIES MANAGEMENT

 

**PLEASE POST**

 

 

Building:         Campus

 

Date:             Thursday October 04,2012

 

This morning at 2:17 a.m. much of the City of Pasadena including our Campus experienced a electric power sag of short duration, approximately 1/10 of a second. The cause was a fault on one of Pasadena’s 17KV circuits. Some sensitive equipment have been impacted.

                 

Contact:          Mike Anchondo x-4999

  7807   Tue Dec 11 08:53:52 2012 SteveHowToPEMcables needs care

How NOT to:

The janitor can not clean in areas like this. He may only steps on these cables accidentally as he dust wiping our chambers.

Attachment 1: IMG_1839.JPG
IMG_1839.JPG
  7809   Tue Dec 11 10:09:04 2012 AyakaHowToPEMcables needs care

Quote:

How NOT to:

The janitor can not clean in areas like this. He may only steps on these cables accidentally as he dust wiping our chambers.

 Sorry for the mess. I fixed it.

  7819   Thu Dec 13 01:49:53 2012 ranaUpdatePEMseismometers

Quote:

I've installed Guralp readout box back and it turned out that it does not work with voltage provided from the rack (+13.76 0 -14.94).  +/-12 voltage regulators inside the box convert it to -0.9 0 -12. I've connected the box to +/-15 DC voltage supply to measure seismic motion at the ETMY table. Readout box works fine with +/- 15.

Seismic noise on the ETMY table measured to be a few times higher then on the floor in horizontal direction in the frequency range 50 - 200 Hz. Attached are compared spectrums of X, Y and Z motions.

 Looking at the PEM BLRMS, I noticed that the GUR1Z channel had a much reduced microseism compared to the GUR1X. Looking at the BLRMS screens everything seems ON, although its a mess (too many filters in the banks, etc. - clean this up, PEM people).

 Looking at the Z channel in DTT, I see that the Z spectra looks double high pass filtered below ~1 Hz.     Needs some attention in the daytime.

From Den and Ayaka's elog entry from Nov 29, its clear that this problem is there at that time. It seems that the seismometer was not even hooked up before then. Perhaps Tara returned the seismometer around Thanksgiving and then someone here hooked it up but neglected to log this work? If so, please make an elog now describing the installation of this sensor at the 40m and log any future work which takes place at the 40m lab even if you think it is inconsequential.

Attachment 1: a.pdf
a.pdf
  7825   Thu Dec 13 21:21:34 2012 DenUpdatePEMseismometers

Quote:

 

 Looking at the PEM BLRMS, I noticed that the GUR1Z channel had a much reduced microseism compared to the GUR1X. Looking at the BLRMS screens everything seems ON, although its a mess (too many filters in the banks, etc. - clean this up, PEM people).

 Looking at the Z channel in DTT, I see that the Z spectra looks double high pass filtered below ~1 Hz.     Needs some attention in the daytime.

From Den and Ayaka's elog entry from Nov 29, its clear that this problem is there at that time. It seems that the seismometer was not even hooked up before then. Perhaps Tara returned the seismometer around Thanksgiving and then someone here hooked it up but neglected to log this work? If so, please make an elog now describing the installation of this sensor at the 40m and log any future work which takes place at the 40m lab even if you think it is inconsequential.

 Yesterday I wanted to recenter Guralps. I turned them off, understood that would be able to center them because we do not have power cable to Guralp box from Tara yet and turned them back on.

I've switched Guralp cables and spectrums are fine now.

Attachment 1: gur.pdf
gur.pdf
Attachment 2: gur_fix.pdf
gur_fix.pdf
  7831   Fri Dec 14 09:05:04 2012 steveUpdatePEMTT after eq 6.1

 

 SRM, ITMY and MC3 moved some what, but how are the TT? It may worth looking at them before we vent.

Attachment 1: eq6.1.png
eq6.1.png
  7841   Mon Dec 17 19:47:15 2012 ranaUpdatePEMSeismic StripTool config updated

I have updated the Seismic Striptool display which is plotted on the wall in the control room. Please take a look and make comments. We should finalize it and not change it anymore.

By having an unchanging display, we can get used to small changes in the seismic environment which disrupt our locking.

  1. y-scale is now linear; the log-scale was suppressing the factor of 2-3 variations which are important to us.
  2. Just as the rainbow does, the colors now go from red to purple to represent the noise from 0.1 - 30 Hz: the red traces are 0.1-0.3 Hz, the green/yellow traces are 0.3-3 Hz, and the blue/purple traces are 3-30 Hz.
  3. This is just showing GUR1. Let's try to keep this seismometer working so that we can have some long term record of the seismicity here. This means don't click off the buttons, disconnect the sensor, reboot the machine, etc. When you do do these things, elog them.
Attachment 1: SeismicRainbow.png
SeismicRainbow.png
  7923   Tue Jan 22 09:10:19 2013 SteveUpdatePEMbig foot is dirty

Please wipe, clean car wheels and wear booties entering the 40m lab.

Obviously this person has no idea about our clean room rules.

Attachment 1: IMG_1873.JPG
IMG_1873.JPG
Attachment 2: IMG_1874.JPG
IMG_1874.JPG
  7942   Thu Jan 24 16:31:46 2013 SteveUpdatePEMbuilding exterier wall painted

The wood exteier walls, gutters and doors were painted at CES-Annex building #69

Attachment 1: IMG_1880.JPG
IMG_1880.JPG
  7954   Tue Jan 29 14:34:42 2013 JenneUpdatePEMSecret Seis filters

 

The BLRMS have been bad again, since the computer crash of last week.  Finally getting around to looking into it, I discovered that there are filter banks that have the microns/sec calibration filters, which are not accessible from the sitemap.  I have added links to them for GUR1 and GUR2.  We need to make the PEM/BLRMS screens macro-expansion-y, so that I don't have to change each screen individually.

Anyhow, the BLRMS are back.

 

  7999   Tue Feb 5 09:08:11 2013 SteveUpdatePEM exterier doors to be painted

Quote:

The wood exteier walls, gutters and doors were painted at CES-Annex building #69

 The east and south end of the 40m emergency exit doors are sealed- tapped off temporarily.  They will be painted on the out side only.  This job will be done by tomorrow noon

 Do not open chamber if you smell the paint !

Attachment 1: IMG_0055.JPG
IMG_0055.JPG
  8001   Tue Feb 5 10:18:54 2013 SteveUpdatePEMhigh particle count ALART

Quote:

Quote:

The BS camber is open only. We should close ASAP

Outside air quality is 1.7- 2.2  million particles  / cf min of 0.5 micron

 Air is still bad and the chambers are closed. Before lunch  Jamie repointed the PRM oplev. Manasa and I reset oplevs: BS and ITMX.

ETMX and ETMY are fine.

SRM and ITMY oplevs needs more work.

 The bad outside air quality is pushing up the inside counts.

The outside air is 5 million counts / cf min for 0.3 micron and 2 million counts / cf min for 0.5 micron particles

Do not open chamber over 10,000 counts / cf min of 0.5 micron

Attachment 1: badair.png
badair.png
Attachment 2: 100d_airq.png
100d_airq.png
  8010   Wed Feb 6 15:10:22 2013 SteveUpdatePEMnew safety signs on exterier doors

Quote:

Quote:

The wood exteier walls, gutters and doors were painted at CES-Annex building #69

 The east and south end of the 40m emergency exit doors are sealed- tapped off temporarily.  They will be painted on the out side only.  This job will be done by tomorrow noon

 Do not open chamber if you smell the paint !

 The east and south end of the 40m emergency exit doors received new safety signs.

Attachment 1: IMG_0068.JPG
IMG_0068.JPG
  8202   Thu Feb 28 14:51:43 2013 JenneUpdatePEM"Rock Monster" causing low frequency noise

Manasa and I have been wondering why the low frequency seismic noise has been different in the last few days/weeks.  I went over to visit the Rock Monster, a.k.a. the Earth Surface Dynamics Laboratory next door, and the grad student turned it up for me (increased water flow significantly) for a few minutes.  Then I came back to look at the BLRMS, and you can see the low frequency increase a few minutes ago, when he turned it up on high.  He said that they run it on low almost all the time, 9am-5pm ish, and on high for an hour or so at a time periodically during the day.

RockMonsterOn_ForFewMinutes.png

  8219   Mon Mar 4 11:30:47 2013 SteveUpdatePEMair cond problem

 

 The air cond is out of order at the area covered by racks 1 X 1  through 1 X 7

The arm X and Y AC units are working.

Attachment 1: 40dAC.png
40dAC.png
  8270   Mon Mar 11 16:29:29 2013 SteveUpdatePEMproposed seismometer locations

Granite base 20" x 20" x 5"  locations are on the CES side of our IFO arms:  as shown ETMY_ south-west, ETMX_north-east, ITMX_south-east . No height limitation. This side of the tube has no traffic.

SS cover McMaster#  41815T4  (H) SS container cov

Attachment 1: ETMY_sw.jpg
ETMY_sw.jpg
Attachment 2: ITMX_se.jpg
ITMX_se.jpg
Attachment 3: ETMX_ne.jpg
ETMX_ne.jpg
  8272   Mon Mar 11 19:01:21 2013 JenneUpdatePEMproposed seismometer locations

This is my interpretation of where Steve is proposing to place the seismometers (he wrote ITMX southwest, but I'm pretty sure from the photo he means southeast).

I think his point is that these locations are on the less-used side of the beam tube, so they will not be in the way.  Also, they are not underneath the tube, so we will not have any problems putting the covers on/taking them off.

 

 

SeismometerProposedPositions_11March2013.pdf

  8314   Wed Mar 20 08:59:24 2013 SteveUpdatePEMair cond maitenance is today

 Outside air quality at 8:30am  1.1 million of 0.5 micron particles cf / min with MET counter #2

AC unit output measured  directly at the outlet of east end  6,000  and south end 3,000 of 0.5 micron particles cf / min with counter #2

Lab measurement at the top of IOOC 20K of 0.5 micron particles cf / min with MET #1

HEPA filters: 2 mobile units  at the middle of the east arm at speed 100%, PSL enclosure 20% and south end flow bench are running.

The PSL, south flow bench and mobile unit (Envirco  sn69406003 ) measured zero counts of 0.5 & 0.3 micron particles

Envirco HEPA filter unit sn 69406001 measured 50 particles of 0.5micron and 320 of 0.3 micron particles cf / min with counter #2

This filter will be replaced.

                   

NOTE: inside air quality peaks we can not control. It depends on the weather and our house keeping.

When your work is particle sensitive keep your eyes on the particle counter.

 

 

 

 

Attachment 1: 100dPEM.png
100dPEM.png
  8456   Mon Apr 15 16:10:52 2013 DenUpdatePEMseismometer isolation kit

 We got granite bases today from the manufacturer. We plan to set them up on Wednesday, 8 am. Please note, there will be an installation mess at Xend, Yend and corner during ~4 hours. Let us know if you have any objections to do this at this particular time.

Installation locations are specified in elog 8270, scheme attached is valid except for Xend. Instrument will be installed on the place of nitrogen containers.

(  next to the wall at corner sout-east of the south end )

  8461   Thu Apr 18 15:08:14 2013 SteveUpdatePEMseismometer isolation kit in place

Quote:

 We got granite bases today from the manufacturer. We plan to set them up on Wednesday, 8 am. Please note, there will be an installation mess at Xend, Yend and corner during ~4 hours. Let us know if you have any objections to do this at this particular time.

Installation locations are specified in elog 8270, scheme attached is valid except for Xend. Instrument will be installed on the place of nitrogen containers.

(  next to the wall at corner sout-east of the south end )

 The carpenter shop finished the installation of the 3 granite bases.Rapid Set Cement All high strength non-shrink grout was used.

 Compressive strength  3000 PSI at 1 hour and 9000 PSI at day 28 The janitor is still cleaning up after them at the south end.

The  soft silicon gas kits are working well with the SS can.  Den is making  the adaptor plate drawing for the feedthrough.

Attachment 1: grouting1.jpg
grouting1.jpg
Attachment 2: grouting2.jpg
grouting2.jpg
Attachment 3: grouting3.jpg
grouting3.jpg
Attachment 4: grouting4.png
grouting4.png
  8466   Fri Apr 19 15:19:25 2013 JamieUpdatePEMTrilliums moved from bench to concrete

I moved the two Trillium seismometers that Den left on the electronics bench out onto the new concrete blocks in the lab that will be their final resting places.  I moved one onto the slab at the vertex and the other to the slab at the Y end.  I left them both locked and just sitting on the concrete.

The pile of readout electronics that were sitting next to them I moved on to the yellow foam box half way down the MC tube, between the MC tube and the X arm tube.  This is obviously not a good place to store them, but I couldn't think of a better place to put them for the moment.

  8477   Tue Apr 23 16:17:45 2013 DenUpdatePEMseismometer isolation kit in place

Quote:

 

 The carpenter shop finished the installation of the 3 granite bases.Rapid Set Cement All high strength non-shrink grout was used.

 Compressive strength  3000 PSI at 1 hour and 9000 PSI at day 28 The janitor is still cleaning up after them at the south end.

The  soft silicon gas kits are working well with the SS can.  Den is making  the adaptor plate drawing for the feedthrough.

 To put everything in one place I add a final drawing of the base to this elog.

 Next time we continue with wiring and putting temperature and pressure sensors inside the box. Connector support plate drawing is attached. We'll have sensors inside the kit with STS-2 or Trillium as their connector is small enough (19 pin vs 26 pin for Guralps) that we can put an additional 4 pin lemo connecor (2 pins for each sensor). I think EGG.0B.304.CLL is good for this application. Temperature and pressure sensor we can by from omega.

Attachment 1: Base.pdf.pdf
Base.pdf.pdf
Attachment 2: ConnectorPlate.pdf
ConnectorPlate.pdf
  8526   Fri May 3 08:55:55 2013 SteveUpdatePEM3.2 M earthquake
Attachment 1: 3.2eqChannelIland.png
3.2eqChannelIland.png
Attachment 2: 3.2eq.png
3.2eq.png
  8558   Thu May 9 02:47:23 2013 JenneUpdatePEMT240 at corner station - cabling thoughts

Something that I want to look at is the coherence between seismic motion and PRM motion.  Since Den has been working on the fancy new seismometer installations, I got caught up for the day with getting the new corner seismometer station set up with a T240.  Later, Rana pointed out that we already have a Guralp sitting underneath the POX table, and that will give us a good first look at the coherence.  However, I'm still going to write down all the cable thoughts that I had today:

The cables that came with the electronics that we have (from Vladimir and tilt meter -land) are not long enough to go from the seismometer to 1X7, which is where I'd like to put the readout box (since the acquisition electronics are in that rack). I want to make a long cable that is 19pin MilSpec on one end, and 25pin Dsub on the other.  This will eliminate the creative connector type changes that happen in the existing setup.  However, before making the cable, I had to figure out what pins go to what.  So.

25pin Dsub          19pin MilSpec

1                   P

2                   N

3                   E

4                   No conn

5                   D

6                   R and V

7                   H

8                   J

9                   No connection

10                  T

11                  F

12                  L

13                  No conn

14                  B

15                  A

16                  R and V

17                  No conn

18                  C

19                  G

20                  G

21                  K

22                  U

23                  No conn

24                  S

25                  M 

I am not sure why R and V are shorted to each other, but this connection is happening on the little PCB MilSpec->ribbon changer, right at the MilSpec side.  I need to glance at the manual to see if these are both ground (or something similar), or if these pins should be separate. Also, I'm not sure why 19 and 20 are shorted together.  I can't find (yet) where the short is happening. This is also something that I want to check before making the cable.

Den had one Female 19 pin MilSpec connector, meant for connecting to a T240, but the cable strain relief pieces of the connector have 'walked off'.  I can't find them, and after a solid search of the control room, the electronics bench, and the place inside where all of Den's connectors were stored, I gave up and ordered 2 more.  If we do find the missing bits for this connector, we can use it for the 2nd T240 setup, since we'll need 2 of these per seismometer. If anyone sees mysterious camo-green metal pieces that could go with a MilSpec connector, please let me know.

  8567   Mon May 13 23:05:51 2013 JenneUpdatePEMGUR1 masses recentered

[Evan, Jenne]

Evan brought the Guralp handheld readout paddle and cables back from the ATF (Zach is using GUR2 and one of the T240s for gyro stuff this week), and we recentered the GUR1 masses.  N/S and Vert were okay (within 0.1 V), but E/W was at -0.5 V, so we set it at zero.  We then plugged the Guralp back in, and turned on the readout box.

There isn't much of a change on the BLRMS on the wall, so it's possible that we weren't actually having any trouble anyway. 

  8634   Thu May 23 20:58:49 2013 JenneUpdatePEMPRM, ITMX optics tripped, restored

M5.7 - 11km WNW of Greenville, California

Time
Location
40.190°N 121.064°W
Depth
0.0km
  8752   Wed Jun 26 01:30:31 2013 ranaSummaryPEMVariation in 10-30 Hz seismic RMS

For quite a while (no one knows how long), we've seen fluctuations in the 10-30 Hz seismic motion. This shows up as the purple trace on the seismic BLRMS on the wall projector.

The second plot shows that this is not only a periodic increase in the usual 29.5 Hz HVAC peak, but also an anomolous 32.2 Hz peak. Probably some malfunctioning machinery - maybe in the 40m or maybe on the roof.

Attachment 1: gur1z_rms.pdf
gur1z_rms.pdf
Attachment 2: gur1z.pdf
gur1z.pdf
  8764   Thu Jun 27 15:50:03 2013 JenneUpdatePEMBLRMS are going crazy

The BLRMS are totally crazy today!  I'm not sure what the story is, since it's been this way all day (so it's not an earthquake, because things eventually settle down after EQs).  It doesn't seem like anything is up with the seismometer, since the regular raw seismic time series and spectrum don't look particularly different from normal.  I'm not sure what's going on, but it's only in the mid-frequency BLRMS (30mHz to 1Hz).

Here are some 2 day plots:

 

WeirdBLRMSincrease_27June2013_rawSeis.png

WeirdBLRMSincrease_27June2013_Gur1xBLRMS.png

WeirdBLRMSincrease_27June2013_lowFreqBLRMS.png

  8765   Thu Jun 27 15:58:27 2013 SteveUpdatePEMPSL enclosure particle counts

 

 Particle counts were measured inside the PSL enclosure at 40VAC variac setting.

Directly under HEPAs, on the east half of the optical table : 0.3 micron    20  particles / cu ft,       0.5 micron     0  p / cu ft,      0.7 micron        0 p / cu /ft

Away from the HEPAs, on the west edge of the op table:     0.3 micron   960  particles / cu ft,      0.5 micron    50 p / cu ft,       0.7 micron      20 p / cu ft

We may want to increase the RPM a little bit.

 

Checked our counters against recently calibrated Met One GT-526:

#1 is right on, #2 measured 25% less (it will go for calibration)

 

Flow bench at the south end :   zero particle for all 3 sizes

 

IOO chamber particle logging  measures  0.5 and 1 micron.  Chamber opening limit is set to 10,000 particles max of 0.5 micron

At 10,000 p of 0.5 micron means ~100,000 - 180,000 p of 0.3 micron

We may have to lower the limit.

  8773   Thu Jun 27 21:45:48 2013 ranaUpdatePEMBLRMS are going crazy

Its an increase in the microseismic peak. Don't know what its due to though.

Attachment 1: useism.pdf
useism.pdf
  8856   Tue Jul 16 13:48:26 2013 SteveUpdatePEMcranes cleaned

Keven and Steve,

The 3 cranes tested and  wiped off as preparation for upcoming vent.

ELOG V3.1.3-