ID Date Author Type Categoryup Subject Text Attachments
  14910   Sun Sep 29 15:58:19 2019 gautamUpdateLSCPOX locking attemptSummary:

There is no visible PDH error signal
on the POX11 channels. As a result, I am
 POXlockAttempt.png 
  14915   Mon Sep 30 14:16:43 2019 gautamUpdateLSCPOX PD checkout - solvedI confirmed that there is light incident
on the POX photodiode. So the problem must
lie downstream in the demod / whitening /
  
  14920   Tue Oct 1 21:19:51 2019 gautamUpdateLSCPRMI locked on carrierSummary:

The
 PRMIlocked.png 
  14924   Wed Oct 2 11:52:16 2019 gautamUpdateLSCPRMI Oplev loop checkoutI measured the OLTF of both the PRM Oplev
loops. Nothing odd sticks out as odd to me
in this measurement - there seems to be ~40
 PRM_OLTF.pdf 
  14926   Wed Oct 2 23:15:02 2019 gautamUpdateLSCFPMI lockingSummary:

I was able to lock the FPMI. The
lock was quite stable. However,
 FPMIlocked.png 
  14928   Thu Oct 3 11:01:18 2019 ranaUpdateLSC(PR)FPMI lockingwonder if its possible to do variable finesse
locking

Gabriele mentioned that Virgo used
  
  14933   Thu Oct 3 19:40:18 2019 gautamUpdateLSCPOX/POY imbalanceSummary:

There is an imbalance between
the POX and POY detector outputs reported
  
  14954   Tue Oct 8 18:35:09 2019 gautamUpdateLSCLocking prepIn preparation for some locking work tonight,
I did the following at the POP in air table
with the PRMI locked on carrier:
  
  14960   Wed Oct 9 18:15:26 2019 gautamUpdateLSCPRMI 3f lockingAfter making sure the beams were hitting
the 3f photodiodes on the "AP"
table, I was able to lock the PRMI with the
 PRMI_1f.png 
  14961   Wed Oct 9 22:02:58 2019 gautamUpdateLSCREFL55 whitening issueThis
problem has re-surfaced. Is this indicative
of some problem with the on-board VGA? Even
  
  14962   Thu Oct 10 01:12:56 2019 gautamUpdateLSCLocking studiesSummary:


ALS control of arms in
the CARM/DARM basis seems pretty robust -
 PRFPMIstudies.png 
  14963   Thu Oct 10 22:11:53 2019 gautamUpdateLSCTrans QPD checkout
I removed the flip-mount that was
installed on the EY in-air table for the
mode-spectroscopy project (see 
 EY_table_20191010.jpegdarkNoise.pdf 
  14973   Wed Oct 16 11:42:17 2019 gautamUpdateLSCPoor separation of PRCL/MICH in 3f signalsSummary:

There is poor separation
of the PRCL and MICH length error signals
 sensMat.pdf 
  14974   Thu Oct 17 11:19:28 2019 gautamUpdateLSCLocking activity last night
Tuning the MICH-->PRM
output matrix element


Locked
the PRMI with the carrier field resonant
 MICH2PRCLnulling.pdfconsolidatedSensingMatrices.pdf.zip 
  14983   Tue Oct 22 00:52:27 2019 gautamUpdateLSCLocking updates
Transition of arms from POX/POY
to CARM/DARM was much smoother today - a
change was made at the EX PDH setup, see
 PRMI3f_ALS_Oct21sensMat.pdf 
  14996   Tue Oct 29 01:24:45 2019 gautamUpdateLSCMore locking updatesSummary:


The two arm lengths can
be controlled reliably in the CARM/DARM basis
 POP22_feature.pngPRMI3f_ALS_Oct21sensMat.pdf 
  14997   Tue Oct 29 15:13:19 2019 gautamUpdateLSCMore locking updatesI looked at some signals for a 10 second
period when the PRMI was locked with at some
CARM offset, just before the PRMI lost lock,
 PRMI_lockloss.png 
  14998   Tue Oct 29 17:40:48 2019 gautamUpdateLSCMore locking updatesI set up a photodiode (PDA10CF) in the
IFO REFL beampath and the Agilent NA is sitting
on the east side of the PSL enclosure. This
 broadSweep.pdfzoomSweep.pdf 
  14999   Wed Oct 30 01:27:00 2019 gautamUpdateLSCMore locking updatesTried a bunch of things tonight.


Modified the "ELP300"
filter module in the MICH filter bank - this
 POP_ASS.png 
  15000   Wed Oct 30 11:53:41 2019 gautamUpdateLSCMICH loop shape tuningI changed the shape of the low pass filter
to reduce high frequency sensor noise injection
into the MICH control signal. The loop stability
 MICH_ELP.pdfMICH_ELP_TFs.pdf 
  15001   Wed Oct 30 17:08:40 2019 gautamUpdateLSCPOP22 investigationThe POP beam coming out of the vacuum chamber
is split by a 50/50 BS and half is diverted
to the POP22/POP110/POPDC photodiode (Thorlabs
 POP22anomaly.pdf 
  15009   Mon Nov 4 15:29:47 2019 gautamUpdateLSCPOP signal pathThere are many versions of the POP22 signal
path I found on the elog, e.g. this
thread. But what I saw at the LSC rack
 POPchain.pdf 
  15010   Mon Nov 4 16:06:58 2019 gautamUpdateLSCPOP optical pathI did some re-alignment of the POP beam
on the IX in air table. Here are the details:


 IMG_8027.JPGlensRealignment.jpgPOPrealigned.pngPOPdither.pngPRMfixed.png 
  15012   Tue Nov 5 11:52:27 2019 gautamUpdateLSCLocking notesSummary:

I am still unable to achieve arm
powers greater than TRX/TRY ~10 while keeping
 PRMI3f_ALS_Nov4sensMat.pdf 
  15014   Wed Nov 6 02:08:48 2019 gautamUpdateLSCLocking updatesSummary:

There seems to be stronger-than-expected
coupling between CARM and the 3f sensors. 
  
  15015   Wed Nov 6 17:05:45 2019 gautamUpdateLSCCARM calibrationSummary:

A coarse calibration of the CARM
error point (when on ALS control) is 
 CARMscan.pdfCARMcalib.pdfscan.hdf5.zip 
  15016   Wed Nov 6 17:45:34 2019 gautamUpdateLSC~Here is a comparison of the response of
various DoFs in our various RFPD sensors
for two different CARM offsets. Even in the
  
  15026   Thu Nov 14 23:56:18 2019 ranaUpdateLSCoff the bad Violin filtersWe turned off many excessive violin mode
bandstop filters in the LSC.

Due to some feedforward work by
 out.pdf 
  15028   Fri Nov 15 11:58:12 2019 gautamUpdateLSCoff the bad Violin filtersThe clue was that the loop X arm POX loop looked
to have low (<3dB)) gain margin around
600 Hz (and again at 700 Hz). 
 violinFix.pdfnewViolinConfig.png 
  15029   Fri Nov 15 12:08:04 2019 gautamUpdateLSCPOPDC whitening boardThe DC port of the Bias-Tee is routed to
(a modified
version of) the iLIGO whitening board.
  
  15034   Mon Nov 18 21:04:38 2019 gautamUpdateLSCLocking - some ideasSome ideas Koji suggested:


Try approaching the CARM
offset zero point "from the other side"
  
  15037   Wed Nov 20 01:07:18 2019 gautamUpdateLSCLocking - progressSummary:


CARM offset was reduced
to 0 with the PRMI locked.
TRY
 PRFPMI.png 
  15038   Wed Nov 20 12:14:17 2019 gautamUpdateLSCLocking - progressI took a look at the TRX/TRY RIN reported
in the single arm POX/POY lock, and compared
the performance of the two available PDs
 TRX_TRY_comparison.pdf 
  15039   Wed Nov 20 17:20:24 2019 YehonathanUpdateLSCQPD Investigation{Gautam, Yehonathan}

In search of the source of discrepancy
between the QPD readings in the X and Y arms,
  
  15040   Wed Nov 20 17:52:00 2019 gautamUpdateLSCQPD MEDM screen updateKoji and I had noticed that there was some
discrepancy between the switchable gain stages
of the EX and EY QPDs. Sadly, there was no
 newLookQPD.pngTRX_TRY_comparison.pdfIMG_8186.JPG 
  15041   Wed Nov 20 21:29:28 2019 gautamUpdateLSCPRG ~13After the QPD fix, both arms report consistent
buildup - see Attachment #1. The peak values
touch ~250, corresponding to a PRG of ~13.
 PRG13.pdf 
  15042   Thu Nov 21 12:46:22 2019 gautamUpdateLSCCM board studyIn preparation for trying out some high-bandwidth
Y arm cavity locking using the CM board,
I hooked up the POY11_Q_Mon channel of the
 YARM_CMslow.pdf 
  15043   Thu Nov 21 13:14:33 2019 KojiUpdateLSCCM board studyOne of the differences between the direct
POY and the CM_SLOW POY is the presence of
the CM Servo gain stages. So this might mean
  
  15044   Thu Nov 21 19:08:58 2019 gautamUpdateLSCHigh BW lock of Y arm length to PSL frequencySummary:

The Y arm cavity length was locked
to the PSL frequency with ~26kHz
 highBW_POY.pdfCM_UGF.pdfIN2_ACcoupling.pdf 
  15045   Fri Nov 22 00:54:14 2019 gautamUpdateLSClocking notes[KA, GV]

There was no shaking (that disturbed
the locking) tonight!
 ALSnoiseIncrease.pdf 
  15046   Mon Nov 25 19:11:22 2019 gautamUpdateLSCALS noise re-lookI re-checked the ALS noise in the following
configurations:


PRM is misaligned.
Michelson
 ALSnoise.pdf 
  15049   Tue Nov 26 17:07:41 2019 gautamUpdateLSCPOX / POY calibrationSummary:

Since we are using the POX and
POY photodiodes as out-of-loop sensors for
 POX_POY_sensorNoise.pdfALSnoise_20191125.pdf 
  15050   Tue Nov 26 18:16:08 2019 ranaUpdateLSCPOX / POY calibration...maybe the opto-mechanical
CARM plant
  
  15051   Wed Nov 27 12:16:52 2019 gautamUpdateLSCITMX and ITMY OSEMs with low and high circulating powerSummary:

The ITMX OSEMs report elevated
noise in the 10-100 Hz band when we have
 ITMXshadowSensors.pdfITMYshadowSensors.pdf 
  15052   Wed Nov 27 13:14:02 2019 ranaUpdateLSCITMX and ITMY OSEMs with low and high circulating powerif the RP don't fit

u must acquit

sweep the laser amplitude
  
  15053   Wed Nov 27 16:10:29 2019 gautamUpdateLSCAOM reconnectedi reconnected the AOM driver to the AOM
in the main beam path (it was hijacked for
the AOM in the AUX laser path for Anjali's
  
  15056   Wed Nov 27 23:24:01 2019 gautamUpdateLSCNo shaking but no inspiration eitherSummary:

I tried the following minor changes
to the locking procedure to see if there
  
  15057   Sun Dec 1 13:38:48 2019 ranaUpdateLSCNo shaking but no inspiration eitheryes, reasonable   
  15062   Tue Dec 3 00:03:57 2019 gautamUpdateLSCGreen ALS also shows elevated noise with high arm buildupSummary:  ALSnoiseIncrease_greenBeat.pdf 
  15073   Wed Dec 4 19:54:27 2019 gautamUpdateLSCA look to the pastTrawling
through some past elogs, I saw that the ALS
noise increase as a function of CARM offset
 DRFPMI_2016March.pdf 