  661   Fri Jul 11 23:55:25 2008 alanUpdateLockingRF common mode at zero offset[quote][I]rob, john, yoichi[/I]

[SIZE=3][FONT=Comic Sans MS]Last night we
  
  675   Tue Jul 15 12:44:08 2008 JohnSummaryLockingDRFPMI with DC readoutRob, John

Last night, despite suspect alignment, we
  
  732   Thu Jul 24 03:08:20 2008 robUpdateLocking+f2 DRMI+2ARMS
rob, john, yoichi
  
  848   Mon Aug 18 17:37:14 2008 robUpdateLockingrecovery progress
I removed the beam block after the PSL periscope
and opened the PSL shutter.
 flatlissa.png 
  862   Wed Aug 20 13:23:32 2008 robUpdateLockingDRMI locked
I was able to lock the DRMI this afternoon.
 All the optical levers have been centered.
  
  953   Wed Sep 17 12:58:12 2008 robUpdateLockingbad
Locking was pretty unsuccessful last night.
 All the subparts were locked (ARMs, PRM,
  
  985   Tue Sep 23 13:25:07 2008 robUpdateLockinga bit betterI've been spending time working on the short
DOF loops (PRC,MICH,SRC) in an attempt to
make the 
  
  998   Fri Sep 26 16:08:39 2008 robUpdateLockingsome progressThere's been good progress in locking the
last couple of nights.  A lot of time was
wasted before I found that
  
  1009   Tue Sep 30 13:43:43 2008 robUpdateLockinglast nightSteady progress again in locking again last
night.  Initial acquisition of DRMI+2ARMs
was working well.
  
  1011   Wed Oct 1 00:24:54 2008 ranaUpdateLockinglast nightI had mistakenly left the MC boost off during
my FAST investigations. The script is now
restored.
  
  1014   Wed Oct 1 02:54:03 2008 robUpdateLockingbad
Tried the spring-y side tonight with a discouraging
lack of progress.  There were several locks
  
  1019   Thu Oct 2 02:45:50 2008 robUpdateLockingmarginally betterLocking the DRMI with the +f2 sideband was
marginally better tonight.  I was able to
get it lock stably enough to take transfer
  
  1024   Fri Oct 3 15:57:05 2008 robUpdateLockinglast night, againLast night was basically a repeat of the night
before--marginally better locking with the
DRMI resonating the +f2
  
  1117   Thu Nov 6 10:06:41 2008 steveUpdateLockingarms lock degradationI have been locking the arms in the mornings
lately.
The daily drift of LSC-TRX is ~ 15% and LSC-TRY
 arms.jpg 
  1327   Thu Feb 19 23:50:31 2009 peteUpdateLockingaligned pd's on AP tableYoichi, Peter
While continuing our efforts to lock,
we noticed the procedure failed at a point
  
  1329   Fri Feb 20 03:52:23 2009 YoichiUpdateLockingLocking TonightYoichi, Peter

Tonight, we had a problem with the DD hand
  
  1334   Tue Feb 24 02:23:40 2009 YoichiUpdateLockingLocking - MC board badRob, Yoichi, Alberto, Kiwamu, Kakeru

We found that the OMC alignment feedback
  
  1335   Tue Feb 24 18:42:15 2009 peteUpdateLockingmc board repairPeter, Yoichi
Last night:
  
  1336   Wed Feb 25 03:10:24 2009 YoichiUpdateLockingLocking statusRob, Yoichi, Kakeru, Kiwamu

Tonight, CARM -> MCL hand off was not stable.
  
  1341   Thu Feb 26 19:59:23 2009 YoichiUpdateLockingDaytime lockingOsamu, Yoichi

We tried locking today from about 2PM.
 CM1.pngCM2.pngCM3.png 
  1343   Fri Feb 27 13:49:19 2009 robUpdateLockingthurs nightCould not get past arm power of ~11 or
so.  I was suspicious of the transmon
high-gain/low-gain PD handover, so I ran
  
  1344   Mon Mar 2 03:57:44 2009 YoichiUpdateLockingSunday night lockingTonight's locking started with a boot fest
of the FE computers which were all red when
I came in.
  
  1346   Mon Mar 2 21:16:32 2009 YoichiUpdateLockingLow-gain High-gain PD switching may not be working wellOsamu, Yoichi

This afternoon, I run the locking script
  
  1350   Tue Mar 3 19:26:44 2009 YoichiUpdateLockingLow-gain High-gain PD switching may not be working wellI checked the switching of the QPDX from high
gain to low gain.
Switching happens as expected, but the low
  
  1351   Tue Mar 3 23:59:26 2009 YoichiUpdateLockingLow-gain High-gain PD switching may not be working well[quote]I checked the switching of the QPDX
from high gain to low gain.
Switching happens as expected, but the low
  
  1352   Wed Mar 4 03:37:51 2009 YoichiUpdateLockingLow-gain High-gain PD switching may not be working well[quote]
This sounds like the QPD whitening gain sliders
may be stuck.  The slider twiddling script
  
  1353   Wed Mar 4 03:50:17 2009 YoichiUpdateLockingStill won't go above arm power 10Yoichi, Kentaro

The IFO still loses lock at around arm power
 lockLoss1.pnglockLoss2.pnglockLoss3.png 
  1361   Thu Mar 5 05:07:09 2009 YoichiUpdateLockingWednesday night lockingTonight, I was having a problem with the PO_DC
hand-off.
It fails most of the time.
 lockLoss5.pnglockLoss5-zoom.png 
  1364   Fri Mar 6 05:44:14 2009 YoichiUpdateLockingLocking distracted by the QPD whitenning problem againTonight, I was able to ramp up the arm power
to around 20. Then the DARM loop started
to oscillate and the IFO lost lock in a few
  
  1365   Fri Mar 6 15:23:39 2009 YoichiUpdateLockingLocking distracted by the QPD whitenning problem againBy looking at the time series of DARM signal
at the time of a lock loss, the oscillation
frequency was about 3.5kHz (see the attm1
 lockLoss3.pdflockLoss3-zoom.pdf 
  1382   Tue Mar 10 04:55:41 2009 YoichiUpdateLockingLocking: 3.7kHz large oscillationYoichi, Jenne, Alberto,

As I reported on the last Thursday, there
 CARM_DARM_Spectra.pdfPODC_Spe_AOPath_Engaged.pngPODC_Spe_before_PODC_handoff.pngAOGain3.pngAOGain2.png 
  1388   Wed Mar 11 16:53:48 2009 YoichiUpdateLockingJunks in around kHzRana, Yoichi

Last night, we tried to find out the source
 MC_F-XARM.pdfold-xarm.pdf 
  1390   Wed Mar 11 22:57:48 2009 YoichiUpdateLockingCalibrated XARM error signal spectrumI did a rough calibration of the XARM error
spectrum.
See the attached calibrated spectrum.
 XarmErrorSpeCalibrated.pdf 
  1393   Thu Mar 12 02:18:42 2009 YoichiUpdateLockingMC_I spectra (RF_AM)I took several spectra of MC_I signal (see
attm1).
 MC_I_Spe.pngMC_I-Xarm.png 
  1399   Fri Mar 13 05:16:21 2009 YoichiUpdateLockingLocking updateYoichi, Osamu,

With adjustments of the loop gains during
 AOTF2.pngAOTF2-zoom.png 
  1402   Fri Mar 13 22:07:14 2009 YoichiUpdateLockingCalibrated XARM error signal spectrumOf course I made a mistake.
I put a pole at 1525Hz whereas it should
have been a zero.
 XarmErrorSpeCalibrated.pdf 
  1419   Tue Mar 24 03:05:25 2009 YoichiUpdateLockingLocking tonightMC1 issue:
The MC1 seems to be drifting still. I found
it was off from the SUS drift-mon reference
  
  1426   Wed Mar 25 04:18:28 2009 YoichiUpdateLockingTuesday LockingAfter the new PO_DC PD was installed, I tweaked
several gains to make the locking scripts
work right.
 AO-Loop-p9.png 
  1433   Thu Mar 26 04:27:26 2009 YoichiUpdateLocking3.8kHz peak as a function of the arm powerDuring the power ramp-up, I actuated CARM
using ETMs and measured the transfer functions
to the PO_DC at several arm powers.
 CARM-PODC.pdf 
  1436   Fri Mar 27 02:50:54 2009 YoichiUpdateLockingDD demodulation phase suspiciousI noticed that the gain of PD6_Q (before the
phase rotation) was 0 whereas PD6_I gain
was 15.
  
  1449   Wed Apr 1 15:47:48 2009 YoichiUpdateLocking3.8kHz peak looks like a real optical response of the interferometerYoichi, Peter

To see where the 3.8kHz peak comes from,
 CARM-ETM-EXC.pngAOpath-TFs.png 
  1450   Wed Apr 1 16:14:36 2009 YoichiUpdateLocking3.8kHz peak does not change with SRC offsetYoichi, Peter

We suspected that maybe the 3.8kHz peak is
  
  1454   Fri Apr 3 17:20:05 2009 YoichiUpdateLockingThe 3.8kHz peak seems like the DARM RSE (not 100% sure though)Yoichi, Kentaro,

Last night, we took several measurements
 AO-TF-DARM-OFFSET.pngAO-TF-DARM-DEMOD-PHASE.pngAO-TF-MICH-OFFSET.pngPOX_1I.pngDARM-Loop.pdf 
  1458   Wed Apr 8 02:47:42 2009 YoichiUpdateLockingLocking statusThis is a summary of activities in the last
few nights, although there is not much progress.
 CARM_TFs.pdfDARM_TFs.pdfDARM-CARM-Coupling.pdf 
  1463   Thu Apr 9 12:23:49 2009 peteUpdateLockingtuning ETM common modePete, Yoichi
Last night, we put the IFO in FP
Michelson configuration.  We took transfer
 FPMI-DARM-CARM-ETM-fineScan.pdf 
  1465   Thu Apr 9 23:11:27 2009 robSummaryLockingLaser PM to PO-DC transfer functions at multiple CARM offsetsI've plotted some transfer functions showing
the response at POB DC to laser frequency
(phase) noise.  There are transfer functions
 CARMoffs1.pngCARMoffs2.pngCARMcarpet.png 
  1466   Thu Apr 9 23:20:35 2009 robSummaryLockingLaser PM to REFL-DC transfer functions at multiple CARM offsets


   
        
            
 CARMoffs1_r.pngCARMoffs2_r.pngCARMcarpet_r.png 
  1468   Fri Apr 10 03:10:08 2009 ranaSummaryLockingLaser PM to REFL-DC transfer functions at multiple CARM offsetsI hereby award the previous rainbow transfer
functions the plot innovation of the month
award for its use of optical frequency to
  
  1469   Fri Apr 10 04:54:24 2009 YoichiUpdateLockingREFL_DC for CARMSuggested by Rob and Rana's simulation works,
I tried to use REFL_DC for the CARM error
signal.
 AO-loop-gain-CARM-REFL_DC.png 
  1480   Tue Apr 14 02:59:02 2009 YoichiUpdateLockingPower up until 26Yoichi, Peter,

With careful adjustments of the common mode
  