5767
|
Mon Oct 31 08:55:19 2011 |
steve | Update | VAC | pressure plot at day 53 |
Quote: |
I was lucky to notice that the nitrogen supply line to the vacuum valves was leaking. Closed ALL valves. Open supply line to atm. Fixed leak.
This was done fast so the pumps did not have to be shut down. Pressurized supply line and open valves to
"Vac Normal" condition in the right sequence.
|
|
Attachment 1: pd71-m-d53.png
|
|
5766
|
Sun Oct 30 23:03:37 2011 |
SUS_Diagonalizer | Update | SUS | SUS input matrix diagonalization complete (EXAMPLE) |
New SUS input matrix diagonalization complete. Matrices have been written to the frontend. Summaries for each optic can be viewed below.
(THIS IS AN EXAMPLE---no new matrices have been written.) |
Attachment 1: MC1.png
|
|
Attachment 2: MC2.png
|
|
Attachment 3: MC3.png
|
|
Attachment 4: ETMX.png
|
|
Attachment 5: ETMY.png
|
|
Attachment 6: ITMX.png
|
|
Attachment 7: ITMY.png
|
|
Attachment 8: PRM.png
|
|
Attachment 9: SRM.png
|
|
Attachment 10: BS.png
|
|
5765
|
Sun Oct 30 23:01:51 2011 |
Zach | Update | SUS | diagAllSUS -- automated input matrix generator |
I finally got around to wrapping up the SUS input matrix diagonalizer. The files I have added to ...scripts/SUS/peakFit are:
- kickAll: Restores all SUS angle biases using /cvs/cds/rtcds/caltech/c1/medm/c1ifo/cmd/C1IFO_OPTICrestore.cmd XXXX &, then runs 'freeswing all'. Finally, writes an elog entry with the time when the optics were kicked and saves the gps time to 'kickAll.time'.
- diagAllSUS.m: An M-file that calls all the other individual M-files needed for the matrix generation. What it does:
- Reads kick time from kickAll.time
- Runs getSensors.m to get time series from all optics' sensors
- Runs makeSUSSpectra.m to generate spectra from time series
- Runs findPeaks.m to fit spectra for peak frequencies
- Loops through all optics and runs, in sequence:
- inmat = findMatrix(XXXX) to generate the matrix for each optic
- writeSUSinmat(XXXX,inmat) to write that matrix to the frontend
- diagAllSUS: A wrapper for diagAllSUS.m. It also writes an elog entry and attaches the pre and post spectra demonstrating the diagonalization. The entry following this one is an example.
The following lines should eventually be added to the controls@nodus crontab:
0 8 * * 0 /cvs/cds/rtcds/caltech/c1/scripts/SUS/peakFit/kickAll
0 18 * * 0 /cvs/cds/rtcds/caltech/c1/scripts/SUS/peakFit/diagAllSUS
(i.e., do the kick at 8am on Sundays and run the diagonalization 10 hrs later at 6pm). They can be done on a different machine but then my elog commands will need to be modified.
Before we add them, we should check that they do, in fact, work. We can do this sometime while I'm at the 40m this week.
|
5764
|
Sun Oct 30 14:08:35 2011 |
rana | Update | Electronics | RFAM monitor in place. ( Uncalibrated ) EPICS troubles |
Quote: |
{Suresh, Jamie, Mirko]
We adapted the Stochmon box to include LP filters at 1.8Hz behind the RMS parts.
Then measured the RMS signals for different RF signal levels at 11.0.65, 29.5, 55.325MHz provided by a RF freq. generator.
As you can see in the data below the suppression of the BP filters of neighboring frequencies is only 35-35dB in power (see also manufacturer specs).
We therefor want to substract crosstalk, by calculating it out. We decided to use C-code in CDS. No computer crashing this time :)
|
This is neat idea, but it seems like it would be easier to just add another set of rf BP filters inside of the StochMon box. Luckily, Steve was thinking ahead and ordered extra filters. |
5763
|
Sat Oct 29 22:57:03 2011 |
Mirko | Update | LSC | AM modulation due to non-optimal SB frequency |
[Kiwamu, Mirko]
Non-optimal 11MHz SB frequency causes PM to be transformed into AM.
m_AM / m_PM = 4039 * 1kHz / df , with df beeing the amount the SB freq. is off.
Someone might want to double check ths. |
Attachment 1: IMC.pdf
|
|
5762
|
Sat Oct 29 05:50:44 2011 |
kiwamu | Update | LSC | MC suspensions misaligned to avid railing for PZTs |
I have shifted the alignment of the MC suspensions such that the PZT won't rail.
Since I didn't care of the spot positions on the MC mirrors, currently they are terribly off from the centers.
After the shift, I realigned the Stochmon PD again.
The attachments below shows the alignment of MC and PZTs before shifting the just for a record

Quote from #5754
|
Tomorrow night I will intentionally introduce offsets in the MC suspensions to avoid the railing.
The goal will be a scan of the incident beam while measuring the recycling gain.
|
|
5761
|
Sat Oct 29 02:35:39 2011 |
Suresh | Update | IOO | WFS_MASTER screen and lockin screens fixed |
I have fixed the WFS_MASTER screen and several of the subscreens such as the MCASS and MC_WFS_LKIN.
Since MC_WFS_LKIN uses six demodulators and single oscillator I could not use the automatically built Lockin screens.
I built one using the compact filter banks mentioned earlier
The phases in the WFSlockins have yet tp be set. |
5760
|
Fri Oct 28 20:39:19 2011 |
MIrko | Update | LSC | RFAM monitor in place. ( Uncalibrated ) EPICS troubles |
{Suresh, Jamie, Mirko]
We adapted the Stochmon box to include LP filters at 1.8Hz behind the RMS parts.
Then measured the RMS signals for different RF signal levels at 11.0.65, 29.5, 55.325MHz provided by a RF freq. generator.
As you can see in the data below the suppression of the BP filters of neighboring frequencies is only 35-35dB in power (see also manufacturer specs).
We therefor want to substract crosstalk, by calculating it out. We decided to use C-code in CDS. No computer crashing this time :)
We however ran into the problem that the RMS signal channels are acquired by the slow (EPICS) maschine c1iool0. Channels are (C1:IOO-RFAMPD_33MHZ , -"-133MHZ, -"-166MHZ) and we could not access those in the CDS c1ioo model. Using the EpicsIn block we got an CA.Exception stating that the variable was hosted on multiple servers. We then tried to use the EzcaRead to access the variables. Got an compile error, about the compiler not beeing able to connect all parts. It seems that the EzcaRead left behind a "ghost" part in the model (something with M1:SYS-FOO_BAR which is the default naming of the EzcaRead block) even after we deleted that block. We toyed around with the /opt/rtcds/caltech/c1/chans/daq/C1EDU_IOO.ini and /cvs/cds/caltech/target/c1iool0/ioo.db files. We tried to uncomment the "old" (33,133,166) channels there to get rid of the conflict, but that didn't work.
We want to write the outputs to C1:IOO-RFAMPD_11MHZ , -"-29MHZ, -"-55MHZ EPICS channels.
We had to get the model back from the svn to get it running again. |
Attachment 1: MC_DC11MHz.m
|
Pwr=[-60,-55,-50,-45,-40,-35,-30,-25,-20,-10,-5,0,5,10]';
Voltage11=[2.12,2.10,2.03,1.93,1.83,1.71,1.59,1.47,1.35,1.10,0.97,0.85,0.71,0.61]';
Voltage11=spline(Pwr,Voltage11,linspace(-60,10,15));
PwrSmooth=linspace(-60,10,15);
Voltage29=[2.14,2.14,2.14,2.14,2.14,2.14,2.13,2.12,2.09,1.94,1.84,1.73,1.61,1.49];
Voltage29=spline(Pwr,Voltage29,linspace(-60,10,15));
Voltage55=[2.16,2.16,2.16,2.16,2.16,2.16,2.16,2.15,2.14,2.13,2.10,2.04,1.94,1.83];
... 5 more lines ...
|
Attachment 2: MC_DC29MHz.m
|
Pwr=[-60,-55,-50,-45,-40,-35,-30,-25,-20,-15,-10,-5,0,5,10]';
Voltage11=[2.16,2.16,2.16,2.16,2.15,2.16,2.15,2.13,2.10,2.03,1.93,1.81,1.70,1.58,1.46]';
Voltage29=[2.12,2.10,2.03,1.93,1.83,1.70,1.59,1.47,1.34,1.22,1.09,0.97,0.84,0.71,0.61]';
Voltage55=[2.16,2.15,2.16,2.16,2.15,2.15,2.14,2.10,2.0,1.97,1.97,1.77,1.65,1.50,1.37]';
%% Example 55MHz inj.
Voltage11=[2.00];
... 21 more lines ...
|
Attachment 3: MC_DC55MHz.m
|
Pwr=[-60,-55,-50,-45,-40,-35,-30,-25,-20,-15,-10,-5,0,5,10]';
Voltage11=[2.16,2.16,2.16,2.16,2.16,2.15,2.15,2.15,2.15,2.15,2.12,2.09,2.00,1.89,1.78]';
Voltage29=[2.14,2.14,2.14,2.13,2.14,2.13,2.11,2.06,1.98,1.88,1.76,1.64,1.52,1.40,1.27]';
Voltage55=[2.14,2.11,2.05,1.96,1.85,1.73,1.61,1.48,1.36,1.23,1.10,0.98,0.84,0.71,0.61]';
plot(Pwr,Voltage55)
%%
... 17 more lines ...
|
Attachment 4: RFAMPD.c
|
double x;
double y;
double z;
double temp1;
double temp2;
double Corrx;
double Corry;
double Corrz;
... 49 more lines ...
|
5759
|
Fri Oct 28 18:33:59 2011 |
steve | Update | VAC | leaking nitrogen line fixed |
I was lucky to notice that the nitrogen supply line to the vacuum valves was leaking. Closed ALL valves. Open supply line to atm. Fixed leak.
This was done fast so the pumps did not have to be shut down. Pressurized supply line and open valves to
"Vac Normal" condition in the right sequence. |
Attachment 1: allvalvesclosed.png
|
|
5758
|
Fri Oct 28 15:45:52 2011 |
Mirko | Update | Computers | Nifty screen generator |
Quote: |
Suresh showed me a cool script that Mirko made, but didn't elog about.
You tell the script what filter banks you want, and it creates a screen for each with a bunch of different filter module display formats. Then you can copy the format you like into the actual screen you're modifying.
Currently PEM, LSC and IOO (and maybe others?) have "fmX" folders inside their medm/c1.../master folders. For each subsystem, we need to copy this folder, and modify the generic .adl file so that it puts in the correct subsystem letters. Once this is done, you can just run the generateFMscreens.py after putting in your filter bank names.
|
Wasn"t me. |
5757
|
Fri Oct 28 15:33:06 2011 |
Jenne | Update | Computers | Nifty screen generator |
Suresh showed me a cool script that Mirko made, but didn't elog about.
You tell the script what filter banks you want, and it creates a screen for each with a bunch of different filter module display formats. Then you can copy the format you like into the actual screen you're modifying.
Currently PEM, LSC and IOO (and maybe others?) have "fmX" folders inside their medm/c1.../master folders. For each subsystem, we need to copy this folder, and modify the generic .adl file so that it puts in the correct subsystem letters. Once this is done, you can just run the generateFMscreens.py after putting in your filter bank names. |
5756
|
Fri Oct 28 14:56:02 2011 |
Jenne | Update | CDS | CSS/BOY installed on pianosa |
Quote: |
I've installed Control System Studio (CSS) on pianosa, from the version 3.0.2 Red Hat binary zip. It should be available as "css" from the command line.
CSS is a new MEDM replacement. It's output is .opi files, instead of .adl files. It's supposed to include some sort of converter, but I didn't play with it enough to figure it out.
Please play around with it and let me know if there are any issues.
links:
|
So far I've only given it about half an hour of my time, but it is *really* frustrating so far. There don't seem to be any instructions on how to tell it what our channels are / how to link CSS to our EPICS databases. Or, the instructions that are there say "do it!", but they neglect to mention how... Also, there exists (maybe?) an ADL->BOY converter, but I can't find any buttons to click, or how to import an .adl, or what I'm supposed to do. Also, it's not clear how to get to the editor to start making screens from scratch.
It looks like it has lots of nifty indicators and buttons, but I would have felt better if I had been able to do anything.
Another thing that is going to be a problem: the Shell Command button that we use all over the place in our MEDM screens is not supported by this program. It's listed in the "limitations" of the ADL2BOY converter. This may kill the CSS program immediately. Jamie: did Rolf/anyone mention a game plan for this? It's super nice to be able to run scripts from the screens.
Moral of the story: I'm annoyed, and going to continue making my OAF screens in MEDM for now. |
5755
|
Fri Oct 28 12:47:38 2011 |
jamie | Update | CDS | CSS/BOY installed on pianosa |
I've installed Control System Studio (CSS) on pianosa, from the version 3.0.2 Red Hat binary zip. It should be available as "css" from the command line.
CSS is a new MEDM replacement. It's output is .opi files, instead of .adl files. It's supposed to include some sort of converter, but I didn't play with it enough to figure it out.
Please play around with it and let me know if there are any issues.
links:
|
5754
|
Fri Oct 28 05:17:13 2011 |
kiwamu | Update | LSC | locking activity : PZT1 is still railing |
Status update on the LSC activity:
To see how good/bad the beam pointing is, I locked the Y arm with POY11.
Then I ran the ASS servo to automatically correct the alignment of the ITMY and ETMY suspensions and also the beam pointing.
The result is that the PZT1_X is still railing to the negative side.
Due to it the transmitted light from the Y arm is about 0.6 or so which is supposed to be 1 if the beam pointing is perfect.
The EPICS value of PZT1_X is at the minimum of -10 and the ASS servo tried to push it more negative side.
Tomorrow night I will intentionally introduce offsets in the MC suspensions to avoid the railing.
The goal will be a scan of the incident beam while measuring the recycling gain. |
5753
|
Fri Oct 28 04:57:00 2011 |
kiwamu | Update | LSC | POX11 demod board broken |
The POX11 demodulation board is broken. It needs to be fixed in the daytime tomorrow.
It only outputs the Q signal and nothing is coming out from the I output.
(Some stuff checked)
[OK] ADCs
[OK] Whitening filters looked fine. Their gains were controllable from EPICS.
[OK] Connection between the POX11 RFPD and demodulator box
[OK] The Q signal showed the PDH signal of the X arm with an amplitude of about 200 counts, which almost is the same as that of POY11.
[NOT GOOD] The I signal from the demod board
[OK] Outputs cables, which send the I and Q signals from the demod board to WFs are fine.
(Test on the POX11 demod board)
As usual, a test signal whose frequency is shifted by a little bit from that of LO was injected to the RF input of the board to see if the circuit is working.
The I signal didn't show up and there were no signals even in the monitor LEMO output.
Something is wrong in the I signal demodulation path on the circuit board.
Here is an actual time series of the I and Q signals in dataviewer. The I signal outputs just junk while the Q showed a nice sine curve.
 |
5752
|
Fri Oct 28 03:42:50 2011 |
kiwamu | Update | LSC | ITMX table needs to be refined |
(POX)
The POX beam had been 80% clipped at a black glass beam dump of the POX11 RFPD.
I steered the first mirror in the POX path to fix the clipping. Then the beam was realigned onto the RFPD.
However the beam is still very close to the black glass, because the incident angle to the second mirror is not 45 deg .
We need to refine the arrangement of the POX11 optics a bit more so that the beam will never be clipped at the black glass.
(POP)
The POP optics also need to be rearranged to accommodate one more RFPD.
Additionally Rana, Suresh and I discussed the possible solutions of POP22/110 and decided to install a usual PD (PDA10A or similar) instead of a custom-made.
So a plan for the POP detectors will be something like this:
+ design an optical layout.
+ buy a 2 inch lens whose focul length is long enough ( #5743 )
+ rearrange the optics and install POP22/110
+ lay down a long SMA cable which sends the RF signal from POP22/110 to the LSC rack.
+ install a power splitter just before the demod board so that the signal is split into the 22MHz demoad board and 110MHz demod board.
=> make sure we have a right splitter for it.
+ install a band pass filter after the power splitter in each path.
=> A 22MHz band pass filter is already in hand. Do we have 110MHz band pass filter somewhere in the lab ?
The picture here shows the latest configuration on the ITMX table.

Quote from #5743 |
RF photo diodes POP55 and POX11 are installed. The beams are aligned to the photo diodes.
|
|
5751
|
Fri Oct 28 03:12:37 2011 |
Suresh | Update | IOO | MC2 realigned to align MC to PSL |
Around 6PM on the 27th, I found that the C1:IOO-MC_RFPD_DCMON had risen to about 2.5V. I checked the trend of MC2 sensors and found that between 2PM and 6PM, MC2 had drifted in a strange way. And also that the alignment had grown worse over several days. I also noticed that the spot on the MC2F camera had shifted to the left.
I attempted to correct the alignment (decrease the C1:IOO-MC_RFPD_DCMON to ~0.5V ) by just moving the MC2 and succeeded!! So it is quite likely that most of the slow MC drift is arising due to MC2 table drift.

I decided to try and close the MC2_TRANS QPD to MC2 loops separately to see if MC alignment becomes stable But several screens needed to be fixed before we could try anything. So I fixed C1IOO_WFS_MASTER, C1IOO_WFS_INMATRIX and ...OUTMATRIX screens. Deleted the older ones to avoid confusion.
In this process I noticed that the directory of $screens$/c1mcs/master contains copies of older C1SUS_MC1 , 2. and 3 screens which look very similar to the new autogenerated screens. Some of the links in the WFS screens were pointing to the old screens. I have redirected the links and I will delete these in a couple of days, if no one objects.
|
5750
|
Fri Oct 28 02:41:18 2011 |
Suresh | Metaphysics | elog | elog unresponsive: restarted |
Elog did not respond despite running the /cvs/cds/caltech/elog/start-elog.csh script two times.
It worked the after the third restart.
|
5749
|
Fri Oct 28 01:13:17 2011 |
kiwamu | Update | SUS | ITMX oplev : iris fully opened |
I found that the sum of the ITMX oplev signals had gone down to zero yesterday.
I checked the ITMX table and found two iris on the He-Ne laser path were blocking the beam on their apertures.
I guess this is because we were working around there for installation of POP/POX and may have touched some of the oplev optics.
Then I fully opened the apertures of those two iris and the sum went back to nominal of 600 counts. |
5748
|
Fri Oct 28 00:53:39 2011 |
rana | Update | Electronics | POP 22/110 Design |
The attached PDF shows a possible gain / input noise config for the POP 22/110 that we would use to detect the RF power in the DRMI. Design is in the SVN.
If Kiwamu/Jenne say that this has good enough sensing noise for the lock triggering than we will build it. This is using a 2mm diode.
If we can get away with 1 mm, we might as well use a PDA10CF for now. |
Attachment 1: poy22110.pdf
|
|
5747
|
Thu Oct 27 18:00:38 2011 |
kiwamu | Summary | LSC | Offsets in LSC signals due to the RFAMs : Optickle simulation |
The amount of offsets in the LSC signals due to the RFAMs have been estimated by an Optickle simulation.
The next step is to think about what kind of effects we get from the RFAMs and estimate how much they will degrade the performance.
(Motivation)
We have been having relatively big RFAM sidebands ( #5616), which generally introduce unwanted offsets in any of the LSC demodulated signals.
The motivation was that we wanted to estimate how much offsets we've been having due to the RFAMs.
The extreme goal is to answer the question : 'How big RFAMs do we allow for operation of the interferometer?'.
Depending on the answer we may need to actively control the RFAMs as already planed ( #5686).
Since the response of the interferometer is too complicated for analytic works, so a numerical simulation is used.
(Results : Offsets in LSC error signals)



Figure: Offsets in unit of meter in all the LSC demodulated signals. Y-axis is the amount of the offsets and the X-axis represents each signal port.
In each signal port, the signals are classified by color.
(1) Offsets in the PRCL signal. (2) Offsets in the MICH signal. (3) Offsets in the SRCL signal.
Roughly the signals showed offsets at a 0.1 nm level.
The numerical error was found to be about 10-10 nm by running the same simulation without the AM sidebands.
Here is a summary of the amount of the offsets:
|
offsets [nm] (1f signal port) |
offsets [nm] (3f signal port) |
biggest offsets [nm] (signal port) |
PRCL |
0.3 (REFL11) |
0.2 (REFL33) |
1 (REFL55) |
MICH |
0.00009 (AS55) |
0.8 (REFL33) |
7 (POP11) |
SRCL |
0.1 (REFL55) |
0.1 (REFL165) |
40 (POX11) |
In the SRCL simulation REFL11I, REFL11Q, POP11I, POP11Q and POX11I didn't show any zero crossing points within 100 nm range around the resonance.
It is because that the SRCL doesn't do anything for the 11MHz sidebands. So it is the right behavior.
However POX11 was somewhat sensitive to the SRCL motion and showed a funny signal with a big offset.
(Simulation setup)
I applied the current PM/AM ratio according to the measurement ( #5616, #5519)
The modulation indices used in the simulation are :
+ PM index in 11MHz = 0.17
+ PM index in 55MHz = 0.14
+ AM index in 11MHz = 0.17 / 200 = 8.5x10-4
+ AM index in 55MHz = 0.14 / 200 = 7.0x10-4
Note that the phases of the AM and PM sidebands are the same.
For clarity, I also note the definition of PM/AM ratio as well as how the first order upper sideband looks like.


The optical parameters are all at ideal length although we may want to check the results with more realistic parameters:
+ No arm cavities
+ PRCL length = 6.75380
+ SRCL length = 5.39915
+ Schnupp asymmetry = 3.42 cm
+ loss in each optic = 50 ppm
+ PRCL = resonant for 11 and 55MHz
+ MICH = dark fringe
+ SRCL = resonant for 55 MHz
The matlab script will be uploaded to the cvs server.
Quote from #5686 |
8. In parallel to those actions, figure out how much offsets each LSC error signal will have due to the current amount of the RFAMs.
=> Optickle simulations.
|
|
5746
|
Thu Oct 27 16:09:37 2011 |
kiwamu | Update | LSC | RF distribution box : two more 55MHz available |
The diagram of the RF distribution box has been updated according to the modification ( #5744).
Both pdf and graffle files are available on the 40m svn : https://nodus.ligo.caltech.edu:30889/svn/trunk/suresh/40m_RF_upgrade/
Here shows the latest version of the diagram.

Quote from #5744 |
I will update the diagram of the RF distribution box (#4342) tomorrow.
|
|
5745
|
Thu Oct 27 03:32:45 2011 |
Koji | Summary | IOO | RFAM monitor progress |
[Suresh, Mirko, Koji]
A cable from the stochmon box to the cross connect for the EPICS ADCs is installed.
The power supply and the signal outputs are concentrated in a single DSub 9pin connector
that is newly attached on the box.
The connection from the stochmon side of the cable and the EPICS value was confirmed.
The calibration of them looks fine.
To do:
- Once the stochmon box is completed we can immediately test it.
- The EPICS channel names are still as they were. We need to update the database file of c1iool0, the chans file for the slow channel.
The pinout is as following
-------------
| 1 2 3 4 5 | Female / Inside View
\ 6 7 8 9 /
\---------/
1 - 11MHz Signal
2 - 30MHz Signal
3 - 55MHz Signal
4 - NC
5 - +5V supply
6 - 11MHz Return
7 - 30MHz Return
8 - 55MHz Return
9 - Supply ground |
5744
|
Wed Oct 26 23:03:03 2011 |
kiwamu | Update | LSC | RF distribution box : two more 55MHz available |
The RF distribution box has been modified so that it generates two more 55 MHz LO signals.
After the modification I put the box back in place.
Then I checked the MICH and YARM locking quickly as a working test of the distribution box and it is working fine so far.
I will update the diagram of the RF distribution box (#4342) tomorrow.
(Motivation)
Since we newly installed POP55 (#5743) an LO signal was needed for the demodulation.
However the RF distribution box didn't have any extra LO outputs.
Therefore we had to make a modification on the RF distribution box so that we can have a 55MHz LO signal for POP55.
Eventually I made two more 55MHz outputs including one spare.
(Modification)
The box actually had two extra output SMAs which had been just feed-thru connectors on the front panel without any signals going through.
In the box the modules consist of two categories; the 11MHz system and 55MHz system. I modified only a part of the 55MHz system.
The modification was done in this way:
* split two branches of 55MHz into four branches by installing two new power splitters (ZMSC-2-1).
* made and installed some SMA cables whose length were adjusted to be nicely fit in the box.
* readjustment of the RF levels to 0-2 dBm at the outputs by replacing some attenuators.
* checked the signals if all of them were happily coming out or not.
Also I found that the POX11 and POY11 demod boards were connected to the whitening filters in a wrong way.
The I and Q signals were in a wrong order. So I corrected them so that the upper inputs in the whitening filter is always the I signal.
(RF levels on 55MHz LO outputs)
Since the demod board requires a certain level of the RF signal as as LO, the LO signals have to be 0-2 dBm.
Here are the RF level in each 55MHz output after the adjustment of the level.
AS55 = 0.76 dBm
REFL55 = 0.76 dBm
POP55 = 0.79 dBm
spare = 0.83 dBm
Those numbers were measured by an oscilloscope and the oscilloscope was configured to measure the rms with the input impedance matched to 50Ohm.
In the measurement I used the actual input seed 55MHz signal from the RF generation box to drive the distribution box. |
5743
|
Wed Oct 26 20:30:47 2011 |
Katrin | Update | LSC | POX11 and POP55 installed |
[Katrin,Jenne]
RF photo diodes POP55 and POX11 are installed. The beams are aligned to the photo diodes.
PD |
DC out dark |
DC out bright |
light power |
calculated DC output |
|
POX11 |
0.1mV |
1.3mV |
0.09mW |
3mV |
POP55 |
35mV |
55mV |
3 to 4 µW |
25mV |
I used 0.7 A/W for the response and 50V/A for POP55 according to elog page #4576.
To install the third RF photo diode we need to order a plano-convex lens with a focal length of 750 or
maybe even better 1000 |
5742
|
Wed Oct 26 11:35:08 2011 |
Katrin | Update | Green Locking | YARM PDH box |

From time to time the 20 dB jump in the transfer function still occurs. The new AD8336 op amp did not change that issue. I am sure that the op amp was broken,
because the amplitude of the sine did not change when I turned the gain knob.
The above two curves were measured with different input amplitude of the sine from the spectrum analyzer. Nothing changed in between except that there was no
jump when Kiwamu was around. Very strange. Testing the electronic board led to no clue what is happening.
For now, I will just use the PDH box as it is, but one should keep this odd behaviour in mind. |
5741
|
Wed Oct 26 10:07:16 2011 |
steve | Update | General | no electricity ....... will be recheduled |
Quote: |
Subject: Electricity Interruption- 11/19/11
From: "PP Service Center" <PPService@caltech.edu>
To: "DeLaRosa, Dario" <Dario.DeLaRosa@caltech.edu>,
...snip... "Sullivan, Joan O." <sully@caltech.edu> |
CALIFORNIA INSTITUTE OF TECHNOLOGY
FACILITIES MANAGEMENT
UTILITY & SERVICE INTERRUPTION
**PLEASE POST**
Building: Central Engineering Services (C.E.S.)
LIGO Gravitational Physics building adjacent to C.E.S.
Safety Storage adjacent to CES
Steele House
Keck Lab
Date: November 19, 2011
Time: 8:00 a.m. To 9:00 a.m.
Interruption: Electricity
Contact: Mike Anchondo ext. 4999 Tom Brennan 4984
*This interruption is required for maintenance of high voltage switchgear in Campus Sub Station.
|
CANCELLED, WILL BE RESCHEDULED* |
5740
|
Tue Oct 25 21:49:13 2011 |
Den | Update | Adaptive Filtering | Adaptive filter witness and EP SNR |
Quote |
Coherence of seismometers to MCL:
STS1 is located at the vertex. x-axis along the x arm.
GUR1 is located at the IMC MC2 mirror. Same orientation.

=> 1. Only the x-direction has good coherence (to be expected)
2. Only good coherence at 1.5-4Hz (huh?)
So probably other noise sources are dominating. Let's look into noise projections. Remember IMC autoalignment is off.
A quick adaptive filter run with only the GUR1 and STS1 witnesses applied only to MCL didn't really do anything. Some more thought needs to be invested into the AA and shaping filters.
|
Indeed, only GUR1_X is reasonable. Static Wiener filtering (length = 2500) of MCL with witness channels GUR_1_X, GUR_1_Y, GUR_1_Z proves your measurements.
We need to callibrate seimometers. I think that now we see velocity, not displacement. It might be useful to amplify the seimometer singal before ADC to make sure that our signal is not ADC noise. |
Attachment 1: gur1_x.jpg
|
|
Attachment 2: gur1_y.jpg
|
|
Attachment 3: gur1_z.jpg
|
|
5739
|
Tue Oct 25 21:23:17 2011 |
Dmass | Bureaucracy | elog | Elog Restarted |
Elog went nonreponsive. SSH'ed into nodus to run restart script. Elog came back ~15 minutes later. |
5738
|
Tue Oct 25 20:04:40 2011 |
Mirko | Update | Adaptive Filtering | Adaptive filter witness and EP SNR |
We currently have the code running for all DOFs using all witness channels. By default nothing is applied. C-Code parameters can be changed via the respective EPICS variables. Sanity checks in the C-Code make sure the code doesn't crash when nothing / zeros are fed to the code. Let's look into applying FF to one DOF only as a starting point. We start with MCL.
Remember there are two possible signals to look into MC-F and MC-Servo. See page 5695 http://nodus.ligo.caltech.edu:8080/40m/5695
Dark noise: MC-F over MC-Servo which is unconnected in this measurement:

=> At least 20dB SNR. ADC noise should not be an issue. Of course more is always better.
Coherence of seismometers to MCL:
STS1 is located at the vertex. x-axis along the x arm.
GUR1 is located at the IMC MC2 mirror. Same orientation.

=> 1. Only the x-direction has good coherence (to be expected)
2. Only good coherence at 1.5-4Hz (huh?)
So probably other noise sources are dominating. Let's look into noise projections. Remember IMC autoalignment is off.
A quick adaptive filter run with only the GUR1 and STS1 witnesses applied only to MCL didn't really do anything. Some more thought needs to be invested into the AA and shaping filters. |
5737
|
Tue Oct 25 18:50:22 2011 |
Suresh | Update | IOO | C1IOO model modified to include new WFS lockin structure |
Some small fixes to the c1ioo model.
1) I edited the WFS lockin modules to make use of new library part called demod.
2) c1ioo model has been compiled and restarted.
3) fb was restarted at Tue Oct 25 18:43:55 PDT 2011
Quote: |
I forgot to mention another change I made to the C1IOO model.
The location of the WFS global switch has been shifted. It now cuts off signals just before the WFS servo filters.
I have also added some test points just before the switch so that we can monitor the WFS sensor signals even if the switch is off.
Quote:
|
A while back we faced the problem that when we use several lockins to excite the MC degrees of freedom, their relative phase was not known. The solution suggested was to use one oscillator and several demodulators.
I have now modified the C1IOO.mdl so that this can be implemented. Previously we were using the MC_ASS lockins for WFS work. I have not separated the WFS and MC_ASS structures.
Other jobs to be done in this context are:
1) The medm screens associated with WFS lockins need to be updated with new channel names.
2) The scripts associated with both MC_ASS decentering and WFS ouput matrix determination have to be updated with the new channel names.
3) I also deleted all mdem screens in the $screens$/c1ioo/ directory after copying them to $screens$/c1ioo/bak/. After installing the new c1ioo model $screens$/c1ioo model now contains just the automatically created screens. All other user made screens should to into $screens$/c1ioo/master/ directory
This is a pic of the new c1ioo model:

|
|
|
5736
|
Tue Oct 25 18:09:44 2011 |
jamie | Update | CDS | New DEMOD part |
I forgot to elog (bad Jamie) that I broke out the demodulator from the LOCKIN module to make a new DEMOD part:

The LOCKIN part now references this part, and the demodulator can now be used independently. The 'LO SIN in' and 'LO COS in' should receive their input from the SIN and COS outputs of the OSCILLATOR part. |
5735
|
Tue Oct 25 16:24:58 2011 |
Suresh | Update | IOO | C1IOO model modified to include new WFS lockin structure |
I forgot to mention another change I made to the C1IOO model.
The location of the WFS global switch and the WFS_GAIN have been shifted. The switch now cuts off signals just before the WFS servo filters.
I have also added some test points just before the switch and the so that we can monitor the WFS error signals which would be unaffected even if the WFS_GAIN is changed..
Quote:
|
A while back we faced the problem that when we use several lockins to excite the MC degrees of freedom, their relative phase was not known. The solution suggested was to use one oscillator and several demodulators.
I have now modified the C1IOO.mdl so that this can be implemented. Previously we were using the MC_ASS lockins for WFS work. I have not separated the WFS and MC_ASS structures.
Other jobs to be done in this context are:
1) The medm screens associated with WFS lockins need to be updated with new channel names.
2) The scripts associated with both MC_ASS decentering and WFS ouput matrix determination have to be updated with the new channel names.
3) I also deleted all mdem screens in the $screens$/c1ioo/ directory after copying them to $screens$/c1ioo/bak/. After installing the new c1ioo model $screens$/c1ioo model now contains just the automatically created screens. All other user made screens should to into $screens$/c1ioo/master/ directory
This is a pic of the new c1ioo model:

|
|
5734
|
Tue Oct 25 11:48:02 2011 |
Katrin | HowTo | Electronics | solder tiny smd op amps |
Yesterday, I had the great pleasure to solder a tiny 4 x 4 mm op amp with 16 legs (AD8336).
I figured out that the best and fastest way to do it is
- to put solder with the soldering iron on every contact of the electronic board (top side)
- heat the bottom side of the electronic board with a heat gun
- use a needle to test if the solder is melted
- if it is melted place the op amp on the electronic board
- apply some vertical force on the op amp for proper contact and heat for 1 to 2 more minutes
- done
|
5733
|
Tue Oct 25 01:19:17 2011 |
Suresh | Update | Computers | fb restarted and c1ioo model committed to svn |
When I installed the new model I restarted the fb between 1 and 1:30 AM PDT Oct 25, 2011 |
5732
|
Tue Oct 25 01:14:15 2011 |
Suresh | Update | IOO | C1IOO model modified to include new WFS lockin structure |
A while back we faced the problem that when we use several lockins to excite the MC degrees of freedom, their relative phase was not known. The solution suggested was to use one oscillator and several demodulators.
I have now modified the C1IOO.mdl so that this can be implemented. Previously we were using the MC_ASS lockins for WFS work. I have now separated the WFS and MC_ASS structures.
Other jobs to be done in this context are:
1) The medm screens associated with WFS lockins need to be updated with new channel names.
2) The scripts associated with both MC_ASS decentering and WFS ouput matrix determination have to be updated with the new channel names.
3) I also deleted all medm screens in the $screens$/c1ioo/ directory after copying them to $screens$/c1ioo/bak/. After installing the new c1ioo model $screens$/c1ioo directory now contains just the automatically created screens. All other user made screens should go into $screens$/c1ioo/master/ directory
This is a pic of the new c1ioo model:

|
5731
|
Mon Oct 24 20:00:21 2011 |
Mirko | Update | CDS | Tiny little scripts |
Located in /opt/rtcds/caltech/c1/userapps/release/cds/common/scripts
Script 1: Diagreset.sh
Hits the diag reset buttons on the models on c1lsc and c1sus computers.
Script 2: Burtrest.sh
Restores the burt files from "today" 4am. Use a text editor if you want to change the times. |
5730
|
Mon Oct 24 19:48:16 2011 |
Mirko | Update | Adaptive Filtering | Filter execution time |
Toyed around some more with the adaptive filters.
Execution time:
nTaps Downsampling factor Execution time average / max in ca. 3 min [us], (480 us available)
1000 16 110 / 150
2000 16 280 / 340
3000 16 380 / 470
4000 16 Over limit
Now we are running with Downsampling 32, 4000 Taps => max 410us execution time.
I tried to desynchronize the downsampled operations of the filters of the different DOFs. That however increased execution time by about 10%. So I undid that.
|
5729
|
Mon Oct 24 17:23:14 2011 |
SUS_Diagonalizer | Update | SUS | Optics kicked |
This is a cron-elog test. No optics have been kicked. |
5728
|
Mon Oct 24 14:26:01 2011 |
steve | Update | General | no electricity on Nov. 19 |
Subject: Electricity Interruption- 11/19/11
From: "PP Service Center" <PPService@caltech.edu>
To: "DeLaRosa, Dario" <Dario.DeLaRosa@caltech.edu>,
...snip... "Sullivan, Joan O." <sully@caltech.edu> |
CALIFORNIA INSTITUTE OF TECHNOLOGY
FACILITIES MANAGEMENT
UTILITY & SERVICE INTERRUPTION
**PLEASE POST**
Building: Central Engineering Services (C.E.S.)
LIGO Gravitational Physics building adjacent to C.E.S.
Safety Storage adjacent to CES
Steele House
Keck Lab
Date: November 19, 2011
Time: 8:00 a.m. To 9:00 a.m.
Interruption: Electricity
Contact: Mike Anchondo ext. 4999 Tom Brennan 4984
*This interruption is required for maintenance of high voltage switchgear in Campus Sub Station.
|
5727
|
Fri Oct 21 18:20:54 2011 |
Mirko | Update | CDS | First OAF version running |
[Jenne, Jamie, Mirko]
We got the first version of the oaf code based on Matt"s code running!! :-)
Produces already data for e.g. MICH DOF. But don"t trust that. It's only 10 taps long and delay is not adjusted. |
5726
|
Fri Oct 21 16:59:14 2011 |
Katrin | Update | Green Locking | YARM PDH box broken |
I could not improve the locking. So, I checked the transfer function of the PDH box again. The transfer function looks okay if the gain knob is <=2.0.
If the gain knob is >2.0 the 20dB step appears in the transfer function (see elog page 5713). This step is shifted to higher frequencies if the gain is
increased. The PZT drive out was not saturated at any time. Yesterday, I checked the electronic circuit with a gain of 2.0. Thus, I couldn't find the broken
gain amplifier (AD8336). The amplifier is ordered in will arrive on Monday. |
5725
|
Fri Oct 21 16:06:12 2011 |
Suresh | Update | Computer Scripts / Programs | MC input matrices empty again |
The MC suspensions were not damping and the reason was traced to the empty imput matrices in the suspension controls. This has been an issue in the past as well when the sus machine is rebooted some of the burt restore does not populate these matrices.
I ran the burtgooey and restored c1mcs.snap file. from a couple of hours ago.
|
5724
|
Fri Oct 21 15:49:35 2011 |
Suresh | Update | IOO | PMC input alignment improved |
The image on the PMCR camera was quite assymetric and PMC output was at 80% .... upon improving the alignment I managed to push it up to 87%
|
5721
|
Fri Oct 21 11:02:47 2011 |
steve | Update | PSL | PSL laser turned ON |
Quote: |
In order to move the emergency shut off switch in room 103 I had to turn off the 2 W Innolight laser. This job will take an hour.
|
It is back on. |
Attachment 1: laserofon.png
|
|
5720
|
Fri Oct 21 10:32:14 2011 |
steve | Update | PSL | PSL laser turned off |
In order to move the emergency shut off switch in room 103 I had to turn off the 2 W Innolight laser. This job will take an hour.
|
5719
|
Fri Oct 21 09:08:33 2011 |
steve | Update | SUS | MC2 Misalignment |
Thinks to do before the NEXT realignment:
B, tie 4 ancher bolts on table legs to the floor
C, tie 4 dog clamps between table and chamber
D, check the locked position of the 4 x 4 positioning screws
E, check bellow protecting 4 tubes are not shorting
A, here is the concrete slab cut
It reminds me to check the IFO vacuum envelope dog clamps on the chambers to floor with torque wrench.
|
Attachment 1: mc2.jpg
|
|
5718
|
Fri Oct 21 02:57:38 2011 |
Suresh | Summary | SUS | MC2 Misaligned 2:27PM on Wednesday : cause traced |
Quote: |
Quote: |
There looks some activity at around MC2 on Wednesday afternoon.
It caused the misalignment of MC2. Misalignment was not found in MC1/3.
It seems that the incident beam on the MC was aligned in the evening.
This increased the MC transmission but it is vibible that the spot on MC2 is shifted from the center.
We need an action on this issue tomorrow in the daytime.
|
I am working on fixing this. You might some strange stuff going on in the control room screens. Pls ignore it till I am done.
|
While chatting with Jenne I learnt that some substantial amount of work had taken place yesterday around the MC2 chamber. This was associated with the relocating of seismometers. ref elog.
I reiterate what is well known for quite sometime: MC2 table is not well isolated from the ground. And we should not approach this chamber unless absolutely necessary. I have blocked off the area around it which we should avoid. It is a serious waste of time and effort to realign the MC each time the MC2 table decides to settle into a new position.
Steve tells me that the mild-steel frame supporting the chamber+MC2_table sits with two legs on one concrete slab while the other two legs sit on another one. The frame is also quite weak without sufficient gussets or cross connects. The next time we have a major shutdown we must replace this frame with a more sturdy one which sits on one slab (preferably the one on which the rest of the MC sits).
Till we improve this mounting, I suggest that we avoid that area as much as possible.
|
5717
|
Fri Oct 21 02:36:44 2011 |
Suresh | Summary | SUS | MC2 Misaligned 2:27PM on Wednesday : MC Realigned |
Quote: |
Quote: |
There looks some activity at around MC2 on Wednesday afternoon.
It caused the misalignment of MC2. Misalignment was not found in MC1/3.
It seems that the incident beam on the MC was aligned in the evening.
This increased the MC transmission but it is vibible that the spot on MC2 is shifted from the center.
We need an action on this issue tomorrow in the daytime.
|
I am working on fixing this. You might some strange stuff going on in the control room screens. Pls ignore it till I am done.
|
I have realigned the MC by recentering the spots on all the MC optics. The current spot positions (in mm) are:
MC1P MC2P MC3P MC1Y MC2Y MC3Y
0.2245 0.3364 -0.2801 -1.8891 0.1631 -1.744
Initially the lockins 2 and 5 showed very small outputs. This was traced to the fact that we have recently switched on a 28Hz ELP filter module in the MC2 ASC filter bank which introduces an extra phase of about 75deg.. See this elog.
When the MC ASS lockins were initially setup, the phase was set with this filter module switched off. Since quite some time has passed since the last calibration of these phases, I readjusted the phases to minimise the Q_OUTPUT and I also adjusted the GAINs in the SIG filter banks of all the six lockins so that their I_OUT's drop by the calibration value of -2.65 when an offset of 0.1 is introduced into the MC suspension output matrices. Two short scripts in the $scripts$/ASS/ directory help in setting and removing these offsets. They are called MCxoffsetOn and MCxoffsetOff. They have to be edited appropriately to address each DoF of the MC.
The $scripts$/ASS/mcassUp script., which sets up everything to make the MC spot decentering measurement, has been edited to set these new phases and gains. The old settings have been commented out.
I then centered the spots on the WFS sensors and the MC_TRANS QPD. We are now ready to make the MC WFS output matrix transfer coef measurement again, but this time with the WFS loops closed.
|
5716
|
Thu Oct 20 18:57:35 2011 |
Suresh | Summary | SUS | MC2 Misaligned 2:27PM on Wednesday |
Quote: |
There looks some activity at around MC2 on Wednesday afternoon.
It caused the misalignment of MC2. Misalignment was not found in MC1/3.
It seems that the incident beam on the MC was aligned in the evening.
This increased the MC transmission but it is vibible that the spot on MC2 is shifted from the center.
We need an action on this issue tomorrow in the daytime.
|
I am working on fixing this. You might some strange stuff going on in the control room screens. Pls ignore it till I am done.
|