40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 205 of 341  Not logged in ELOG logo
IDdown Date Author Type Category Subject Text Attachments
  6862   Sun Jun 24 00:10:45 2012 yutaUpdateGreen Lockingcurrent beat electronicsI moved amplifiers for beat PD at PSL table
to 1X2 rack. Current beat setup from PD to
ADC is shown below. Setup for X beat and
 beatsetup20120623.png 
  6861   Sat Jun 23 19:57:22 2012 yutaSummaryComputersc1ioo is downI tried to restart c1ioo becuase I can't
live without him.
  
  6860   Sat Jun 23 18:44:15 2012 steveUpdateGeneralpower surge has no effect on the labI was notified by CIT Utilities that there
was a power surge or short power outage this
after noon.
  
  6859   Sat Jun 23 02:29:18 2012 yutaUpdateGreen LockingX arm mode scan resultsX
arm finesse is 416 +/- 6, mode-matching ratio
is 91.2 +/- 0.3%
 XarmScan20120623.pngfine8FSRscanXarm.png 
  6858   Fri Jun 22 20:58:15 2012 JenneUpdateGreen LockingCalibrated POX spectra - Xarm stabilized by ALS[Yuta, Jenne, Koji]
We stabilized the Xarm using the
ALS and took a spectrum of POX as our out
 POX11_I_ERR_calib_residualCavityMotion_better.pdf 
  6857   Fri Jun 22 20:00:14 2012 JamieOmnistructureElectronicstwo RG-405 cables ran from 1X2 rack to control room[Yaakov, Eric, Jenne, Yuta]
Two of our surfs, Yaakov and Eric,
pulled two unused RG-405/SMA cables that
  
  6856   Fri Jun 22 19:52:47 2012 JamieUpdateComputersottavia reconfigured as CDS workstationottavia has been reinstalled with Ubuntu
10.04 LTS, and has been configured as a CDS
workstation.
 workstation-setup.sh 
  6855   Fri Jun 22 17:51:04 2012 JenneUpdateCamerasGreen Trans camera repositioning attempt[Yuta, Jenne]
We tried to find a different place,
not in the main green transmitted beam path,
  
  6854   Fri Jun 22 13:37:17 2012 JenneUpdateComputersfb lost connection...Perhaps related to the fact that Jamie
is copying a lot of stuff over the network
to back up Ottavia before converting her
  
  6853   Fri Jun 22 10:52:18 2012 yutaUpdateGreen Locking2 arm ALS - Success!!!!Answers to questions from Koji.
  
  6852   Fri Jun 22 03:37:42 2012 KojiUpdateGreen Locking2 arm ALS - Success!!!!Are
these correct?
1. It is a nice work.
  
  6851   Fri Jun 22 02:21:57 2012 JenneUpdateGreen Locking2 arm ALS - Success!!!![Yuta, Jenne]
We
 P6210140.AVIspeedy_gonzales-5257.jpg 
  6850   Thu Jun 21 20:07:18 2012 JamieUpdateGreen LockingImproved beatbox returnsI've reinstalled the beatbox in the 1X2
rack.  This improved version has the
X and Y arm channels stuffed, but just one
  
  6849   Thu Jun 21 15:36:51 2012 yutaUpdateLockingX arm alignmentI aligned X arm so that the beam spot comes
roughly on the center.
  
  6848   Thu Jun 21 15:00:55 2012 steveUpdateSAFETYSURFs 2012 safety trainingMasha, Eric, Yaakov, Liz and Sasha received
40m specific basic safety training.
 surfs2012.JPG 
  6847   Thu Jun 21 12:56:49 2012 yutaUpdateLockingETMX 1064 trans camera

    

       
            
 Xarm_EndTableLayout_NewTransCamera.png 
  6846   Thu Jun 21 12:13:35 2012 JenneUpdatePEMflow bench must be running all times

    

       
            
  
  6845   Thu Jun 21 09:10:08 2012 steveUpdatePEMflow bench must be running all timesThe south end flow bench HEPA filter should
be run all times. You can turn it off for
a measurement or two but remember we are
  
  6844   Thu Jun 21 09:01:18 2012 steveUpdateSTACISthe resurrection of STACIS -surfThere is a small wood cabinet under the
south end flow bench, labeled STACIS.
Unit is complete with extension cards
 IMG_1364.JPG 
  6843   Thu Jun 21 02:57:16 2012 JenneBureaucracyGeneral40m Action ItemsPRIORITY
ITEMS:
  
  6842   Thu Jun 21 01:58:29 2012 JenneUpdateGreen LockingXgreen preparations[Yuta, Koji, Jenne]
Lots
of small things happened tonight, in preparation
  
  6841   Wed Jun 20 18:43:57 2012 yutaUpdateLSCcalibrated POX error signal[Jenne, Yuta]

We did the same calibration for POX.
 LSCxarmTF_usingITMX.pngLSCxarm_HAover1plusG.pngPOXerrorcalibration.png 
  6840   Wed Jun 20 18:09:23 2012 yutaUpdateLockingboth arms aligned, ITMX oplev centered[Jenne, Yuta]

We aligned FPMI. I also centered ITMX
 FPMIalignment2010620.png 
  6839   Wed Jun 20 18:00:10 2012 ranaSummaryPSLsummaries Nice PSL summaries from LHO:
https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=3187
  
  6838   Wed Jun 20 16:37:11 2012 yutaUpdateLockingETMX 1064 trans camera[Jenne, Yuta]

We made ETMXT camera working.
  
  6837   Wed Jun 20 01:02:20 2012 JamieUpdateGreen LockingRF amp removed from X arm ALS setupI very badly forgot to log about this in
the crush of surfs.
I removed Koji's proto-beatbox RF
  
  6836   Wed Jun 20 00:02:16 2012 yutaUpdateGreen Lockingslower scan using phase tracking ALSFor those of you who want to see plots
from slower scan.
 YarmScan20120619.png 
  6835   Wed Jun 20 00:01:04 2012 JenneUpdateLSCcalibrated POY error signal[Yuta, Jenne]
We have measured the out of loop
residual motion of the Yarm while locked
 LSC_POY_11_I_ERR_calib_19June2012.pdfPOY_calib_19June2012_FiltBankSettings.png 
  6834   Tue Jun 19 23:36:19 2012 yutaUpdateLSCcalibrated POY error signal[Jenne, Yuta]

We calibrated POY error signal(C1:LSC-POY11_I_ERR).
 LSCyarmTF_usingITMY.pngLSCyarm_HAover1plusG.pngPOYerrorcalibration.png 
  6833   Tue Jun 19 20:26:50 2012 JenneHowToLockingSummer PlanJenne and Yuta's
Summer Plan
  
  6832   Mon Jun 18 23:54:31 2012 yutaUpdateGreen Lockingphase tracker for ALS[Koji, Jenne, Yuta]

Summary:
 newIQLOCK.pngYarmScan20120618.pngfine8FSRscan.png 
  6831   Mon Jun 18 23:38:39 2012 JenneUpdateLSCAdded LSC channels to framesSince the .ini files get overwritten every
time a model is compiled now, we need to
put all channels we want saved to frames
  
  6830   Mon Jun 18 17:28:03 2012 yutaSummaryComputersbugs in CDS_PARTS/simLinkParts/FcnFcn module in CDS_PARTS is used to include
a user defined function in a model.
We should be able to use this by entering
 model_sindegbug.pngdv_sindegbug.png 
  6829   Mon Jun 18 16:23:59 2012 JenneUpdateLockingLSC trigger updateThe LSC triggers for the individual filter
modules in a filter bank now works. 
This is handy so that boosts can come on
  
  6828   Mon Jun 18 02:31:43 2012 yutaSummaryGreen Lockinganalysis of mode scan dataI analyzed mode scan data from last week.
Mode matching ratio for Y arm is 86.7
 coarse5FSRscan.pngfine1FSRscan.pngHOMRFSB.pngcoarse5FSRscanHOMRFSB.png 
  6827   Sat Jun 16 19:32:11 2012 yutaUpdateGreen LockingY arm length using 5FSR scanI know!
But I think there's some error (~
10% ?) in calibrating the beatbox. In elog
  
  6826   Sat Jun 16 18:51:44 2012 KojiUpdateGreen LockingY arm length using 5FSR scan

    

       
            
  
  6825   Sat Jun 16 18:17:00 2012 yutaUpdateGreen LockingY arm length using 5FSR scanCalibrating error signal to beat
frequency;
  I injected 0 dBm RF sine wave
 ALS-BEATY_COARSE_I_IN1_DQ.pngCalibYarmScan20120614_2.png 
  6824   Sat Jun 16 13:01:17 2012 yutaUpdateGreen Lockingscanned Y arm for 5FSR

    

       
            
 YarmScanDV.png 
  6823   Sat Jun 16 12:03:41 2012 ZachUpdateGreen Lockingscanned Y arm for 5FSRIs that time stamp really correct?
I wanted to look at the signal closely to
see if I could get any feeling for why it
  
  6822   Sat Jun 16 01:03:21 2012 yutaUpdateGreen Lockingused longer delay line for mode scan[Mengyao, Yuta]

Last night, I used 1.5 m delay line
 YarmScan20120615.png 
  6821   Fri Jun 15 13:33:39 2012 yutaUpdateGreen LockingADC noise contribution to ALSADC noise is not a limiting noise source
in a current ALS setup.
 YarmALSnoise20120615.png 
  6820   Fri Jun 15 01:53:05 2012 KojiUpdateGreen Lockingscanned Y arm for 5FSRInteresting.
It seems for me that there is a dependence
of the noisiness as the beat frequency is
  
  6819   Fri Jun 15 00:50:54 2012 yutaUpdateGreen Lockingscanned Y arm for 5FSRI scanned Y arm for 5FSR (below).
I could done this after I put a whitening
filter.
 YarmScan20120614_2.png 
  6818   Thu Jun 14 21:37:37 2012 yutaUpdateGreen Lockingsucceeded in 1FSR mode scan[Jenne, Yuta]

We couldn't scan the Y arm for 1FSR
 YarmScan20120614.png 
  6817   Thu Jun 14 04:53:39 2012 yutaSummaryGreen Lockingdesigning ALS loop for mode scan[[Requirement]]
 Arm cavity FWHM for IR is 
 ALSloop.pngALSopenloop.pngyarmlength.png 
  6816   Thu Jun 14 01:36:34 2012 yutaUpdateGreen Lockingcan't scan Y arm for 1FSR[Jenne, Koji, Yuta]

We tried to scan of the Y arm but
 BEATY_COARSE_LoopOnOff.png 
  6815   Wed Jun 13 17:39:13 2012 yutaUpdateGreen Lockingcalibrating the beatbox[Jenne, Yuta]

We put 0 dBm sine wave to the RF input
 ALS-BEATY_COARSE_I_IN1_DQ.pngALS-BEATY_COARSE_Q_IN1_DQ.pngALS-BEATY_FINE_I_IN1_DQ.pngALS-BEATY_FINE_Q_IN1_DQ.png 
  6814   Wed Jun 13 11:19:05 2012 steveUpdateSUSIDC receptacles clampedThe MC_ IDC 64 pin cables from sat. amplifiers
to junction-interface-board towards 
whitening - dewhitening at the back of rack
 IMG_1341.JPG 
  6813   Wed Jun 13 11:10:56 2012 ranaUpdateGreen Lockingmy first modescan (sort of)You can easily calculate whether or not
the coarse readout will work by thinking
about the scan resolution you need given
  
ELOG V3.1.3-