40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 192 of 357  Not logged in ELOG logo
ID Date Author Typeup Category Subject
  7715   Thu Nov 15 03:09:08 2012 JenneUpdateAlignmentNo good progress, many options
I didn't make any concrete progress today. AS and REFL cameras are in place, and Manasa has put ND 0.5 filters on both. I used a 
camera to look at the back of the Faraday, and aligned PRM such that it was retroreflecting, and then tried to align ITMY to have once 
fringes with the PRM at that place. I failed in this, since the AS beam on the AS table was starting to dall off the first mirror on the table. 
I then restored all the suspensions to where they were before I started touching them today. 

I moved ETMY face camera so that it is looking at the front of the black glass, but it's hard to tell where the beam is.  I was thinking 
about setting up a temporary camera to look at the back of ITMY to help guide PZT steering, but I haven't done this yet. 

Koji and I then talked about the several different options I have for references, and how many different knobs I  can turn. I'm sleeping 
on it for now, and hopefully I'll have more insight on what to do tomorrow. 
  7716   Thu Nov 15 21:52:48 2012 Den, AyakaUpdateGreen Lockingyarm locked

We aligned and locked Y arm for green:

  • installed camera on PSL to monitor green transmission
  • aligned green path on the ETMY table to see the beam on the PSL camera
  • misaligned ETMY and aligned ITMY to see reflected beam on REFL PD
  • installed green transmission PD on PSL
  • aligned ETMY and locked YARM to 00 mode

I've switched error channel cable to output monitor. Whitening filter is need for scattering measurements.

   etmyf.png  yarm.png

  7717   Fri Nov 16 00:08:36 2012 KojiUpdateSUSMC2 woes

People complained about the MC instability: If we aligned the MC, it locked nicely for a while.
Then suddenly you find that it got totally misaligned with the order of 0.2 with the alignment slider.
This misalignment usually happens for MC2, but it happend on MC3 once.

Surprisingly to me, this instability happened even without MCL and WFS, not only once but at least three times.
This suggests that the suspensions are the cause of the trouble.

I played with the MC2 suspension for a while in the afternoon. It seems that it has a hysteresis (or say, bistablity).
And the nominal alignment of MC2 seems close to the point where the transition happens. (Dunno why)

I further played with MC2 and found that a step of POS actuation by +/-10000 induces this transition go and back.
When the POS kick is in the negative direction (-10000), the MC2 seems to return to the preferrable
position. Therefore, I applied DC position force of -5000 to pull the mirror a bit from the nominal position.

I let the MC locked for ~4hours without MCL and WFS, it kept good alignment and the lock was stable
except for unlocks because of the activties by Den and Ayaka.

All of them has been done without previous monitor data as the tools were available.

The MC2 situation is not conclusive but we should check how we can prevent the bistable transition
by restricting MCL/WFS.

  7718   Fri Nov 16 03:12:39 2012 Ayaka, DenUpdateGreen Lockingxarm locked

We aligned and locked xarm for green.

etmxf.png

  7719   Fri Nov 16 09:57:57 2012 JenneUpdateGreen Lockingxarm locked

Quote:

We aligned and locked xarm for green.

 

 That's really, really awesome!

 

  7720   Sat Nov 17 03:30:13 2012 Den, AyakaUpdateAlignmentred in arms

We aligned accurately 00 green in yarm, changed voltage on PZT2 to see red flashing at TRY at the normalized level 0.2-0.3. The plan was to lock yarm using POY11 and green from other side, maximize red TRY by adjusting PZT2. But POY11 does not go out of the vacuum, so we adjusted TRY by flashing. 2 DOFs of PZT2 is not enough to match 4 DOFs of red beam so we adjusted both PZT2 and cavity mirrors. TRY flashing is 0.5-0.6 and green is still locking to 00 though its transmission is not maximized. We'll fix it later by adjusting input green beam.

Next we wanted to get red beam on TRX PD. Beam steering was done by BS only. We misaligned BS in pitch and excited BS angle motion by 1000 counts. We could see red beam moving on the wall of ETMX chamber. We moved it to ETMX mirror frame, estimated position of the mirror center and moved BS to this position. The beam should be approximately in the middle. For now we can not see red beam on the camera at ETMX table, more work is needed.

  7721   Sat Nov 17 18:02:14 2012 DenUpdateAlignmentred in arms

Quote:

POY11 does not go out of the vacuum

 It does but slighty low and does not get on mirrors. We need to change optic mounts to adjust the height. Red is flashing in yarm at 00 and 10 modes. TRY is ~0.4-0.5.

I've adjusted BS angle, camera and TRX PD at ETMX table so I can see red flashing at 03 mode while green is locked to 00 and its transmission is maximized. I thought that by adjusting BS angle, I will be able to align red to 00 not disturbing green, but this was not the case. Maximum TRX I could get was 0.1. I've adjusted POX to get into PD and I can see PDH signal though I can't lock as cavity is still misaligned for red.

  7722   Sat Nov 17 22:50:17 2012 KojiUpdateAlignmentred in arms

You have constraints for the IR beams (i.e. one PZT and one BS for 8 dofs), so now you need to align the arms for the input IR beams.
The PZT and BS should be aligned so that you have the beam spots as center as possible with the above restrictions.

Then realign end greens for the given arm alignment. You can replace the mounts if necessary to align the end green.
Even if you lose the coarse alignment of the green, realignment is not difficult as you know now

Quote:

Quote:

POY11 does not go out of the vacuum

 It does but slighty low and does not get on mirrors. We need to change optic mounts to adjust the height. Red is flashing in yarm at 00 and 10 modes. TRY is ~0.4-0.5.

I've adjusted BS angle, camera and TRX PD at ETMX table so I can see red flashing at 03 mode while green is locked to 00 and its transmission is maximized. I thought that by adjusting BS angle, I will be able to align red to 00 not disturbing green, but this was not the case. Maximum TRX I could get was 0.1. I've adjusted POX to get into PD and I can see PDH signal though I can't lock as cavity is still misaligned for red.

 

  7723   Mon Nov 19 15:12:52 2012 JenneUpdateAlignmentYarm locked IR

Quote:

Quote:

POY11 does not go out of the vacuum

 It does but slighty low and does not get on mirrors. We need to change optic mounts to adjust the height. Red is flashing in yarm at 00 and 10 modes. TRY is ~0.4-0.5.

I've adjusted BS angle, camera and TRX PD at ETMX table so I can see red flashing at 03 mode while green is locked to 00 and its transmission is maximized. I thought that by adjusting BS angle, I will be able to align red to 00 not disturbing green, but this was not the case. Maximum TRX I could get was 0.1. I've adjusted POX to get into PD and I can see PDH signal though I can't lock as cavity is still misaligned for red.

 [Ayaka, Jenne]

We put the POY beam onto the POY PD.  The Yarm is currently locked on IR with ~0.65 transmission.

 

  7724   Mon Nov 19 15:15:22 2012 JenneUpdateSUSAll oplev gains turned on

Quote:

Oplev values that were changed to zero:

PRM P=0.15, Y=-0.3

SRM P=-2.0, Y=2.0

BS P=0.2, Y=-0.2

ITMY P=2.1, Y=-2.0

ITMX P=1.0, Y=-0.5

ETMX P=-0.2, Y=-0.2

ETMY P=0.5, Y=0.6

Also, PRCL was changed in the LSC input matrix from REFL33I to AS55I, since there is no REFL beam out of the IFO :(

 Ayaka and I restored all of the oplev gains to these values.  The exception is ETMY, which has both gains negative.  I am unsure if this is a transcription error on my part, or if something physical has changed.  The layout of the ETMY oplev was modified (since Rana took out the offending lens) but that shouldn't affect the sign of the gains.

  7725   Mon Nov 19 15:58:02 2012 SteveUpdateSAFETYsafety training

Charles Blakemore  our new undergrad help received 40m specific safety training today.

  7729   Mon Nov 19 23:14:31 2012 EvanUpdateCamerasETMYF focus

Adjusted focus on ETMYF camera so that the IR beam is in focus.

  7730   Tue Nov 20 02:57:24 2012 Ayaka, Den, KojiUpdateLockingred in arms

We aligned and locked x and y arms.

MCL loop makes arms lock unstable, adds a lot of noise at frequencies 60-100 Hz. We'll fix it.

At some point we were not able to lock because of ADC overflows of PO signals. They happened if whitening filters were enabled. So we reduced the gain of POX whitening filters down to 36 dB and POY - to 39 dB. Now cavities can be locked with whitening filters.

Also we changed the pedestal of the lens in the beam path to the POX because the beam was too high.

 

arms.png    arms_psd.png

etmxf.png  etmyf.png

itmxf.png     itmyf.png

etmxt.png    etmyt.png

  7732   Tue Nov 20 15:11:22 2012 SteveUpdateGeneralfew more sensing cards

New  Lumitek IR Sensor Cards are here. We got 2 pieces of Q-11-T (2" x 2"), 2 pieces of Q-11-T (0.75" x 0.75")  and one Q-11 (4" x 5")

  7733   Tue Nov 20 15:24:52 2012 AyakaUpdatePEMSeismometers and a microphone

 I got two seismometers and one microphone back from Tara.

They are now near the Gurlap under the MC.

  7734   Tue Nov 20 16:03:37 2012 janoschUpdatePEMSeismometers and a microphone

Quote:

 I got two seismometers and one microphone back from Tara.

They are now near the Gurlap under the MC.

 If anybody wants a fancy single-axis seismometer for a while (GS-13), then please let me know.

  7735   Tue Nov 20 20:37:51 2012 KojiUpdateIOOMC autolocker

Last night I found that the MC autolocker has not been updated since the chamber was closed.
i.e. The low power version had been used.

I logged into op340m and modified crontab via "crontab -e" so that the normal power version is spawned.

  7736   Wed Nov 21 01:31:37 2012 Koji, AyakaUpdateLockingalignment on ETMX table

Since the transmission beam on ETMXT camera seemed to be clipped, we checked the optics on ETMX table.

We aligned the lens so that it is orthogonal to the beam, then the beam shape looks fine.

output.nv12.bmp

Also we removed some an-used optics which were used for fiber input.

  7737   Wed Nov 21 01:36:36 2012 KojiUpdateIOOMCL disabled / WFS clear history restored

As MCL is disturbing arm locking by injecting a lot of noise, I have modified 'mcup'  to disable MCL

As MC WFS keeps failing to start up when it is locked, the lines in 'mcwfsoff' to clear WFS filter history were restored.

  7738   Wed Nov 21 21:06:13 2012 AyakaUpdateLSCcalibration of arms

Motivation

In order to estimate whether we can see acoustic coupling in arms or not, we have to calibrate signals to phase noise.

Method

I used the same method as Yuta and Jenne did (6834).
I switched from ETM locking to ITM locking since only ITM actuators are calibrated (5583), and measured the open loop transfer function and the transfer function from ITM excitation to POX/POY error signal. Then I can estimate the calibration value H [counts/m] from POY/POX error signal to displacement.

Results

Yarm; H = 9.51 x 1011 counts/m
  OL_y.pngerr_exc_y.pngPOY_disp_y.png

Xarm; H = 6.68 x 1011 counts/m
OL_x.pngerr_exc_x.pngPOX_disp_x.png

Phase noise in arms:
XY_phase.png
blue; Xarm, green; Yarm

 

Next Step

I will calibrate the acoustic signal and see if it is reasonable that we can see the acoustic coupling signal in the arms.
But I guess it is difficult. Actually I have not seen coherence between ETM feedback signals and acoustic sounds yet. (I measured acoustic noise near POX and in PSL table.)

If I find that it is hopeless, I will create some sounds and try to measure transfer function from acoustic sound to arm cavity signals.
I am interested in how the transfer function calculated by wiener filtering is different from the measured transfer function.

 

Note

I found that we do not have enough phase margin. This is why the arm locking is not so stable.

  7739   Sat Nov 24 13:58:07 2012 ranaUpdateLSCcalibration of arms

For the loop diagnosis, its best to use the method of "IN1/IN2", rather than manipulate the close loop gain. In this way, you can directly plot the swept sine measurement from DTT as the open loop gain.

Also, for reporting calibration, we should all try to record the current settings better. Anything that may change the loop gain should be recorded along with the Bode plot and the DATA must be posted to the elog - no more of just posting plots.

We need to know, e.g.

  1. what is the power in the arms?
  2. are the LSC whitening filters on?
  3. are the SUS dewhitening filters on?
  4. What normalization is being used in the LSC?
  5. What digital filters are on in the X/YARM loop filter bank?

Resistance is feudal.

  7740   Sat Nov 24 22:14:08 2012 KojiUpdateGeneralAP table cleaning up

On Wednesday (21st) night, I checked the AP table as I wanted to try PRMI locking. 
It was difficult to work with the table as there were so many unnecessary components on it.
Also the beams went through complicated paths as they have funny angles. 

So I decided to clean up of IMC REFl WFS, IFO REFL, and IFO AS paths.
I found that the AS beam was highly astigmatic as the beam went through a (too-much-) tilted lens.
I made several blocked optical paths for REFL and AS for future extension of the detection system.

The current status of the table was uploaded below.

The optical spectrum analyzers and the aux NPRO were left untouched but they should be moved
somewhere (either on the table or outside) which does not disturb the other optical paths.
 


After the cleaning, I started locking PRMI. I could lock PRMI stably. But I could not figure out how
the intra-cavity mode looked like as I did not have the POP camera. The power recycling
gain was not quantitatively evaluated as I did not have POP and I wasn't sure how the beam was aligned at POX/POY.


We need to know:

- Quantitative evaluation of the beam shape in the PRC

- Quantitative evaluation of the power recycling gain

Some obvious things to be fixed

- The POX whitening filters seem not switching. This issue should be checked at the circuit module itself and at the BIO.

- The POX beam is not well focused on the PD. This was particularly clear when PRMI was locked with carrier.

- The POP beam is going nowhere. We need POP55 and POP CCD for diagnoses.

I haven't checked ITMY table.

Attachment 1: APTable121124.pdf
APTable121124.pdf
Attachment 2: APTable121124.graffle.zip
  7741   Sat Nov 24 23:50:35 2012 KojiUpdateIOOMC WFS refusing to work

Today I found the IMC was misaligned significantly by WFS feedback.
Once the feedback was cleared, it locks with nice visibility.
But WFS misaligns it again as soon as the intergrators are engaged.
I checked the beam on the table, but found nothing really wrong.
The offsets of the error signals were nulled at the input filter modules of the WFS segments.
They did not fix the problem.

The instability started about 48hour ago, that means my work on the AP table did not 
made immediate trouble. But it does not mean anything.

For now, the WFS outputs are off. More work is needed to find what's wrong.

 

  7743   Mon Nov 26 10:42:06 2012 AyakaUpdateLSCcalibration of arms

I uploaded a zip file that contains data files used for the calibration.

OLTF_x/y.txt: the open loop transfer function (measured by IN1/IN2 in arm servo filter bank).
coh_x/y.txt: coherence of OLTF. I used the data where coherence > 0.98.
ext_err_x/y.txt: the transfer function from ITM excitation signal to POX/POY error signal.
coh_x2/y2.txt: coherence of ext_err. I used the data where coherence > 0.98.

The LSC whitening filter was off because the xarm was unlocked when the POX Q whitening filter was turned on. (We have to study what was wrong.)
The SUS whitening filters were on.
The all digital filters except +6dB filter were on.

Attachment 1: armcalib.zip
  7744   Mon Nov 26 10:58:20 2012 SteveUpdatePEMair cond maitenance tomorrow

Air conditioning maintenance is scheduled for tomorrow morning till noon.

  7745   Mon Nov 26 18:36:17 2012 JenneUpdatePEMBLRMS back

Quote:

 I got two seismometers and one microphone back from Tara.

They are now near the Gurlap under the MC.

 I have finally plugged GUR1 back in....it is down at ETMY for now, since that's where the cable was.  BLRMS are back up on the projector.

  7748   Mon Nov 26 23:45:52 2012 AyakaUpdateIOOTuning MC_L

[Rana, Jamie, Ayaka]

We could not lock the arms with MC_L loop on. Therefore we measured the change in YARM error signal when the MC_L is turned on.

POYerr_MCL.jpg

(data; POYerr_MCF.xml in zip file)

Green line; POY error signal when MCL loop was on and the YARM loop gain (0.5) was so high that the saturated control signal made funny peak around 250 Hz.
Blue line; POY error signal when MCL loop was off and the YARM loop gain was low (0.2).
Pink line; POY error signal when MCL loop was on (the gain was -300) and the YARM loop gain was low (0.2).
Red line; POY error signal when MCL loop was on, another low pass filter (2nd order, cut off frequency of 55Hz) was added to MCL loop and the YARM loop gain was low (0.2).

We also changed the filter trigger in order to lock YARM. The FM7 and 8 trigger was turned off. It means that spectrum above was took with FM2,3,4,5,6,9,10 on. Whitening filters were also on.

MCL control signal makes the arm spectrum bad because the MCL control signal moves MC2 mirror additionally and adds extra frequency noise.
Ideally, error signal should be the same at higher frequencies and go down at the lower frequencies when the MCL loop is on because MCL signal should suppress the seismic noise.

Before we added the LPF, MCF/MCL loop cross over (which was taken with the template /users/Templates/MC/MCL-MCF_xover-2012-8-23.xml) is below;

MCL-MCFxover.jpg

(MCL-MCF_xover.xml in zip file)

After the LPF is added, the cross over has been changed as below;
MCL-MCFxover2.jpg

(MCL-MCF_xover2.xml in zip file)

For now, I will just turn off the MCL loop for the acoustic noise experiments.

Attachment 4: 121126.tar.gz
  7750   Tue Nov 27 00:45:20 2012 jamieUpdateIOOMC_L and laser frequency noise spectra

I grabbed the a plot of the iLIGO PSL frequency noise spectrum from the Rana manifesto:

laser_noise.pdf

Rana's contention is that this spectrum (red trace) is roughly the same as for our NPRO.

From the jenne/mevans/pepper/rana paper Active noise cancellation in a suspended interferometer I pulled a plot of the calibrated MC_L noise spectrum:

MCL_noise.pdf

The green line on this plot is a rough estimate of where the above laser frequency noise would fall on this plot.  The conversion is:

    L / f  =  10 m / 2.8e14 Hz = 3.5e-14 m/Hz

which at 10 Hz is roughly 1.5e-11 m.  This puts the crossover somewhere between 1 and 10 Hz.

  7751   Tue Nov 27 01:03:42 2012 AyakaUpdateWienerFilteringSound on PSL

 Last Thursday, I put the speaker and my laptop in the PSL table, and make triangular wave sound with the basic frequency of 40Hz, and Gaussian distributed sound.
(I create the sounds from my laptop using the software 'NHC Tone Generator' because I could not find the connector from BNC to speaker plug.)
And I measured the acoustic coupling in MCF signal. The all the 6 microphones were set in PSL table around PMC and PSL output optics. 

The performance of the offline noise cancellation with wiener filter is below.
(The target signal is MCF and the witness signals are 6 microphones.)

  • With Gaussian sound (Sorry for wrong labeling 'XARM' and no calibration)
    gauss_psdcoh_mcf.png
  • With 40Hz Triangular sound (Sorry for no calibration again)
    tri40_psdcoh_mcf.png

I can see some effects on MCF due to the sound on PSL table. Though I can subtract some acoustic signal and there are no coherence between MCF signal and mic signals, still some acoustic noise remains.
This is maybe because of some non-linearity effects or maybe because we have other effective places for acoustic coupling measurement. More investigations are needed.

Also, I compared the wiener filter and the transfer function from microphones signal to MCF signal. They should be the same ideally.
gauss_filters_mcf.pngGauss_estimatefilters_mcf.png

(Left: Wiener filter, Right: Transfer function estimated by the spectrum. They are measured when the Gaussian sound is on.)

These are different especially lower frequencies than 50 Hz. The wiener filter is bigger at lower frequencies. I guess this adds extra noise on the MCF signal. (see the 1st figure.)
The wiener filter can be improved by filterings. But if so, I want to know how can we determine the filters. It is interesting if we have some algorithms to determine the filters and taps and so on.
The more investigations are also needed.

  7752   Tue Nov 27 03:26:00 2012 JenneUpdateGeneralPOP beam on POP55, POP camera

Quote:

Some obvious things to be fixed

-  We need POP55 and POP CCD for diagnoses.

 Done.  The beam is also going vaguely in the direction of POP110, but I can't see the beam, so it's tricky. 

Order of operations:

1. Find POP on the table, place iris so I wouldn't forget.  Find beam by putting big IR card where I think beam should be, look at IR card with IR viewer.

2. Move and re-clamp 2" lens so beam is on center of lens.

3. Move and re-clamp 1st (2") mirror so that beam is on center of mirror. 

4. Remove BS-33% so that all the beam goes to POP55, steer that 1st mirror so beam is on POP55's little mirror.  Align little mirror so beam is centered on POP55 (as seen by looking at PD with viewer, finding "edges" of PD, going back to center).

5. Put BS-33% back in place.  The reflected portion of this beam is not possible to see using card+viewer technique.

6. Remove BS-50% that reflects half of this beam to POP110.  Find beam reflected from BS-33% by waving POP camera around.  Steer BS-33 until beam goes back in the direction that the camera used to be mounted.  Adjust camera mount and BS-33 so that beam is on camera.

7. Put BS-50% back in place.  Steer it around with voltmeter on PD to see if beam ever hits PD.  Unsuccessful. Give up, since we have POP55, and POP camera.

8. Make a youtube video: POP, AS, REFL, ITMXF (all on Quad3) - PRMI coarsely aligned, no IFO parts locked.  MICH was locked earlier, but not during video time.

  7753   Tue Nov 27 09:44:09 2012 SteveUpdateVACpumpdown at 15 days

 It took 15 days to reach 1e-5 Torr

The RGA was turned on yesterday. The rga scan plotting was not working. Den wrote a Python script to log and plot because the Mathlab had trouble receiving data from our old c0raga.

It turned out that the Joe's log&plot system was working today. Thanks Den and Joe.

Here is the first rga scan.

 

Attachment 1: pd74m15d.png
pd74m15d.png
Attachment 2: pd74m15d_rga1d.png
pd74m15d_rga1d.png
  7754   Tue Nov 27 11:27:22 2012 SteveUpdatePEMparticle count is going up

Jeff is still working on the filter banks.
Attachment 1: 5yparticles.png
5yparticles.png
Attachment 2: particle_6m_6m.png
particle_6m_6m.png
Attachment 3: 1y_particlec.png
1y_particlec.png
Attachment 4: 10y_particle_c.png
10y_particle_c.png
  7755   Tue Nov 27 17:45:42 2012 SteveUpdateGeneralspare optics of AP table moved to cabinet

Spare optics from the AP table were moved to glass cabinet in the east arm. I'm not sure this is the right place. We'll see what everybody thinks.

There were two UNMARKED optics. Shame on you! No pencil marks on the optics either. These optics were shipped to the FBI for finger tip analysis.

 

 

Attachment 1: IMG_1819.JPG
IMG_1819.JPG
  7756   Tue Nov 27 19:06:16 2012 RijuUpdate Testing AG4395A+GPIB

 I ve tested another bandpass filter today with similar set-up. This time I took the data with corrected reference level. To set this reference-level the filter was disconnected and the cable was connected "thru" according to the instructions provided in the manual of AG4395A at http://cp.literature.agilent.com/litweb/pdf/04395-90040.pdf, page 3-10. The transfer functions are as follows 

Attachment 1: tfmag1.jpg
tfmag1.jpg
Attachment 2: tfphase1.jpg
tfphase1.jpg
  7759   Wed Nov 28 23:18:35 2012 CharlesUpdatePEMDecreased RMS in Seismometers

The attached plots display RMS noise from various accelerometers and seismometers over the past 90 days. One can see how after the reinstallation of the seismometers in November, RMS from the GUR1Z and GUR1X channels decreases by a factor of about 100 from data in August. Additionally, the RMS over the course of the last 90 days has notably decreased in all instruments. In many cases, the RMS is only the result of inherent electronics noise, rather than from a signal.

Attachment 1: 8-31_11-29_PEM-RMS.jpg
8-31_11-29_PEM-RMS.jpg
  7760   Wed Nov 28 23:55:13 2012 AyakaUpdateWienerFilteringSound on PSL

 I have been searching for the way we can subtract signal better since I could see the acoustic coupling signal remains in the target signal even though there are no coherence between them.

I changed the training time which is used to decide wiener filter.
I have total 10 minutes data, and the wiener filter was decided using the whole data before.
tri40_psdcoh_mcf_varioust.pnggauss_psdcoh_mcf_varioust.png

(Right: the performance with the data when the triangular sound was created. Left: the performance with the data when the gaussian sound was created.)

I found that the acoustic signal can be fully subtracted above 40 Hz when the training time is short. This means the transfer functions between the acoustic signals and MCF signal change.
However, if the wiener filter is decided with short-time training, the performances at lower frequencies get worse. This is because wiener filter do not have enough low-frequency information.

So, I would like to find the way to combine the short-time training merit and long-time training merit. It should be useful to subtract the broad-band coupling noise.

  7761   Thu Nov 29 00:15:13 2012 Den, KojiUpdateIOOMC WFS work

Quote:

The instability started about 48hour ago, that means my work on the AP table did not 
made immediate trouble. But it does not mean anything.

For now, the WFS outputs are off. More work is needed to find what's wrong.

 

 The problem was caused by low reflectivity of the mirror that splits MC reflected beam into two: first goes to trash, second - to WFS. Power before the mirror was 100mW, reflected beam that goes to WFS was 0.3mW. Using dataviewer we learnt that the beam intensity was ~5 times more in the past.

This happened because the mirror position was adjusted a few days ago. Its reflection depends on the angle of incidence and amount of light to WFS was significantly reduced. We could either increase the angle of incidence or use two mirrors with high reflectivity instead of this with high transmission.

We've chosen the second variant not to confuse anyone in future with non-45 degrees angles. We are now using one mirror with reflectivity 98% to direct most power to the trash while other 2% are directed using the second mirror to WFS path. We now have 0.7 mW on WFS1 and 1.3 mW on WFS2.

Then we adjusted WFS 

  • blocked the beam and run scripts/MC/WFS/WFS_FilterBank_offsets to calculate offsets in the WFS servo
  • aligned MC and centered beams on WFS 1 and 2
  • provided excitation to MC1 at 5 Hz (400 counts) and adjusted I&Q phase rotation
  • adjusted the gain and changed it in MC autolocker (reduced from 0.25 to 0.15 as we now have more power of WFS as before)

We were able to close the loops. The phase margin is too low though, we need to improve feedback filters.

DSC_4942.JPG

Attachment 1: wfs_fb.pdf
wfs_fb.pdf wfs_fb.pdf wfs_fb.pdf wfs_fb.pdf wfs_fb.pdf wfs_fb.pdf
  7762   Thu Nov 29 02:43:48 2012 AyakaUpdatePEMDecreased RMS in Seismometers

Quote:

The attached plots display RMS noise from various accelerometers and seismometers over the past 90 days. One can see how after the reinstallation of the seismometers in November, RMS from the GUR1Z and GUR1X channels decreases by a factor of about 100 from data in August. Additionally, the RMS over the course of the last 90 days has notably decreased in all instruments. In many cases, the RMS is only the result of inherent electronics noise, rather than from a signal.

The Image is replaced

[Den, Ayaka]

We found that seismometer was working and the calibration in the filter banks should have been wrong.
We turned off the all FM2 filter in RMS filter banks.

We also installed STS seismometer. It is under the BS. Now we have spectrum of three seismometers.
GUR1Xfilterbank.pngseismometers1129.pdf


 

RA: the above plot is kind of unreadable and useless. Please replace with something legible and put in some words about why there is a wrong filter, what exactly it is, etc., etc. etc. And why would you leave in a filter which is not supposed to be on? We might as well leave a few secretly broken chairs in the control room...

Attachment 2: seismometers1129.jpg
seismometers1129.jpg
  7763   Thu Nov 29 09:58:06 2012 DenUpdatePEMDecreased RMS in Seismometers

Quote:

[Den, Ayaka]

We found that seismometer was working and the calibration in the filter banks should have been wrong.
We turned off the all FM2 filter in RMS filter banks.
 

We also installed STS seismometer. It is under the BS. Now we have spectrum of three seismometers;
 

RA: the above plot is kind of unreadable and useless. Please replace with something legible and put in some words about why there is a wrong filter, what exactly it is, etc., etc. etc. And why would you leave in a filter which is not supposed to be on? We might as well leave a few secretly broken chairs in the control room...

 First of all, STS-2 is in the end of X arm, GUR2 is under BS, GUR1 is in the end of Y arm.

BLRMS were small because we applied calibration from counts to um/s two times. In the past we had calibration in the RMS BP filter bank (vel2vel = FM2). Now we have calibration in the seismometer input filter bank so we can save calibrated _OUT channels.

  7764   Fri Nov 30 02:40:44 2012 DenUpdateAdaptive FilteringYARM

I've applied FIR adaptive filter to YARM control. Feedback signal of the closed loop was used as adaptive filter error signal and OAF OUT -> IN transfer function I assumed to be flat because of the loop high gain at low frequencies. At 100 Hz deviation was 5 dB so I've ignored it.

I've added a filter bank YARM_OAF to C1LSC model to account for downsampling from 16 kHz to 2 kHz and put low-pass filter inside.

I've used GUR 1&2 XYZ channels as witnesses. Bandpass filters 0.4-10 Hz we applied to each of them. Error signal was filters using the same bandpass filter and 16 Hz 40 dB Q=10 notch filter. As an AI filter I used 32 Hz butterworth 4 order low-pass filter. Consequently, AI, bandpass and notch filters were added to adaptive path of witness signals.

I've used an FIR filter with 4000 taps, downsampling = 16, delay = 1, tau = 0, mu = 0.01 - 0.1. Convergence time was ~3 mins.

yarm_oaf.png

  7766   Fri Nov 30 11:38:35 2012 SteveUpdatePSLfoil removed from enclosure

Aluminum foil replaced by sheet metal on Enclosure and AP table.

Attachment 1: IMG_1822.JPG
IMG_1822.JPG
  7767   Fri Nov 30 11:49:24 2012 KojiUpdateAdaptive FilteringYARM

This is interesting. I suppose you are acting on the ETMY.
Can you construct the compensation filter with actuation on the MC length?
Also can you see how the X arm is stabilized?

This may stabilize or even unstabilize the MC length, but we don't care as the MC locking is easy.

If we can help to reduce the arm motion with the MCL feedforward trained with an arm sometime before,
this means the lock acquisition will become easier. And this may still be compatible with the ALS.

Why did you notched out the 16Hz peak? It is the dominant component for the RMS and we want to eliminate it.

  7769   Fri Nov 30 22:11:50 2012 DenUpdateAdaptive FilteringARMS

Quote:

This is interesting. I suppose you are acting on the ETMY.
Can you construct the compensation filter with actuation on the MC length?
Also can you see how the X arm is stabilized?

This may stabilize or even unstabilize the MC length, but we don't care as the MC locking is easy.

If we can help to reduce the arm motion with the MCL feedforward trained with an arm sometime before,
this means the lock acquisition will become easier. And this may still be compatible with the ALS.

Why did you notched out the 16Hz peak? It is the dominant component for the RMS and we want to eliminate it.

 I actuate on ETMY for YARM and ETMX for XARM. For now I did adaptive filtering for both arms at the same time. I used the same parameters for xarm as for yarm.

I've notched 16 Hz resonance because it has high Q and I need to think more how to subtract it using FIR filter or apply IIR.

I'll try MC stabilazation method.

Attachment 1: arms_oaf.pdf
arms_oaf.pdf
  7770   Fri Nov 30 23:10:36 2012 CharlesUpdateElectronicsVertex Illuminators

 3 of the 4 remote controlled illuminators at the vertex are installed and can now be turned on via sitemap. There are a total of 15 controls for "Illum", but only the 3 labeled with MC, BS-PRM and ITMY-SRM are functional.

  7771   Sat Dec 1 00:13:16 2012 DenUpdateAdaptive FilteringARMS and MC

Quote:

 

 I actuate on ETMY for YARM and ETMX for XARM. For now I did adaptive filtering for both arms at the same time. I used the same parameters for xarm as for yarm.

I've notched 16 Hz resonance because it has high Q and I need to think more how to subtract it using FIR filter or apply IIR.

I'll try MC stabilazation method.

 Adaptive filtering was applied to MC and X,Y arms at the same time. I used a very aggressive (8 order) butterworth filter at 6 Hz as an AI filter for MC not to inject noise to ARMS as was done before

Mu for MC was 0.2, downsample = 16, delay = 1. I was able to subtract 1 Hz. Stack subraction is not that good as for arms but this is because I used only one seismometer for MC that is under the BS. I might install accelerometers under MC2.

EDIT, JCD, 18Feb2013:  Den remembers using mu for the arms in the range of 0.01 to 0.1, although using 0.1 will give extra noise.  He said he usually starts with something small, then ramps it up to 0.04, and after it has converged brings it back down to 0.01.

Attachment 1: arms_mcl_oaf.pdf
arms_mcl_oaf.pdf
  7772   Sat Dec 1 00:24:37 2012 Den, AyakaUpdateAlignmentBS chamber

Today at 11:13 AM the stack of invacuum BS table was kicked and IFO misaligned. We adjusted PZT2 voltage by ~20 V in yaw such that IPPOS was restored. Then we could lock arms.

push.png

  7773   Sat Dec 1 13:56:38 2012 ranaUpdateAlignmentBS chamber

 

 Whoever was working around the BS chamber at 11 AM on Friday should admit it now and take the punishment.

For those of you who like to do work on the interferometer without reporting it in the elog because you think that what you did doesn't affect anything, this is your example of how our time can be wasted by such laziness.

  7774   Sat Dec 1 16:58:14 2012 ManasaUpdateWIKI-40M UpdateOptical tables

I have updated the wiki with the layout of the out-of-vac optical tables: Updated optical tables

I used the new camera to take pictures.

Lesson learnt after the update:

To use the new canon to take better pictures of optics tables; set the camera to manual mode; no flash and iso at around 800 or higher if you can hold the camera still for that long. The autofocus works beautifully...so you will not need any minor tweaking of lens to take pictures. 

  7775   Sun Dec 2 00:37:49 2012 DenUpdateSUSTT cable problem

 This week I've got all TT stuff baked and today was testing eddy current damping and electronics.

In the beginning everything was good: ring magnets fit mirror holder holes and their interaction with actuation magnets is strong enough to keep damping magnets in the wholes. I've put the frame horizontally and kicked it, magnets were still in the whole. Brackets also fit to the TT frame.  

DSC_4945.JPG     DSC_4946.JPG

I've tested eddy current dumping during ring down measurements, it was strong enough.

DSC_4947.JPG    DSC_4948.JPG

Then I started to test electronics. I've provided signal to TT1 channels and could see it in the clean room. But then things went terrible. I just could not connect TT cables to OSEMS, there is not enough space in the OSEM for the connector to plug in.

DSC_4949.JPG     DSC_4952.JPG

Connector should be machines to be more narrow. There is actually no reason for a connector to have this shape. I think it was designed to fit perfectly the OSEM frame but turned out to be ~0.5 mm wider then it should be.

  7777   Mon Dec 3 16:37:09 2012 SteveUpdatePEMEM 172 microphones ordered

Quote:

I've put EM 172 microphones inside Steve's isolation box to measure their noise. I've attached mics to each other and aligned them using the tape.

At low frequencies (below 1 Hz) the noise is limited by ADC as there is a 10 Hz high-pass filter inside mic readout box.

ADC noise is measured by splitting the signal from 1 mic into 2 ADC channels.

em172.png

 BT EM172 microphones  are ordered.

ELOG V3.1.3-