Mon Oct 1 19:21:03 2012 |
rana | Configuration | PEM | changed PEM DQ channels | Changed the list of channels to be written to frames from having the IN1 suffix to OUT. Now we can load the calibration of the channel into the filter module and the DQ channel will be calibrated.
We should do this wherever possible so that our channels will have real calibrations associated with them.
SEIS_GUR1_X_OUT 256
SEIS_GUR1_Y_OUT 256
SEIS_GUR1_Z_OUT 256
SEIS_GUR2_X_OUT 256
SEIS_GUR2_Y_OUT 256
SEIS_GUR2_Z_OUT 256
SEIS_STS_1_X_OUT 256
SEIS_STS_1_Y_OUT 256
SEIS_STS_1_Z_OUT 256
SEIS_STS_2_X_OUT 256
SEIS_STS_2_Y_OUT 256
SEIS_STS_2_Z_OUT 256
SEIS_STS_3_X_OUT 256
SEIS_STS_3_Y_OUT 256
SEIS_STS_3_Z_OUT 256
MIC_1_OUT 2048
MIC_2_OUT 2048
MIC_3_OUT 2048
MIC_4_OUT 2048
MIC_5_OUT 2048
MIC_6_OUT 2048
ACC_MC1_X_OUT 2048
ACC_MC1_Y_OUT 2048
ACC_MC1_Z_OUT 2048
ACC_MC2_X_OUT 2048
ACC_MC2_Y_OUT 2048
ACC_MC2_Z_OUT 2048
XARM_DIFFERENTIAL_MOTION_IN1 256
XARM_DIFFERENTIAL_MOTION_OUT 256
YARM_DIFFERENTIAL_MOTION_IN1 256
YARM_DIFFERENTIAL_MOTION_OUT 256
Next we should up the rate at which the model runs up to 16 kHz so that we can record the microphones at 16 kHz. FM radio has information up to 20 kHz. AM radio goes up to ~8 kHz. We should be at least as modern as AM radio. How do we make the change? How do we make sure the FOTON file stays OK?
I have made some changes to the daily summary file to compensate. New files is /users/public_html/40m-summary/share/c1_summary_page.ini. |
7458
|
Mon Oct 1 17:03:01 2012 |
steve | Update | SUS | ETMY oplev relayed |
Quote: |
Quote: |
Quote: |
The typical sign of a dying gas laser is that it glows for a few minutes only. The power supplies are fine.
Two new JDS - Uniphase 1103P lasers ( NT64-104 ) arriving on Monday, May 21
|
Yesterday I swapped in new He/Ne laser with output power 3.5 mW The return spot on qpd is large ~6mm in diameter and 20,500 counts
The spot size reduction require similar layout as ETMX oplev.
|
The oplev path is relayed and the spot size on the qpd is reduced. I still have to clean up and replace "Miki Mouse" lens holder.
There was no IP-ANG coming out of the chamber at this time!
|
I did relayed the oplev path with new f 500 mm lens |
Attachment 1: IMG_1673oplevETMY.jpg
|
|
7457
|
Mon Oct 1 16:05:01 2012 |
jamie | Update | CDS | mx stream restart required on all front ends | For some reason the frame builder and mx stream processes on ALL front ends were down. I restarted the frame builder and all the mx_stream processes and everything seems to be back to normal. Unclear what caused this. The CDS guys are aware of the issue with the mx_stream stability and are working on it. |
7456
|
Mon Oct 1 13:11:43 2012 |
Jenne | Update | IOO | PZT inspection elogs | I'll come back to the PZTs later, but I want to write down all the elogs I have found so far that look relevant.
http://nodus.ligo.caltech.edu:8080/40m/699
http://nodus.ligo.caltech.edu:8080/40m/5368
nodus.ligo.caltech.edu:8080/40m/5431 |
7455
|
Mon Oct 1 11:08:25 2012 |
Jenne | Update | IOO | MC WFS adjustment |
Quote: |
Found a bug in the IOO screen: All of the 6 WFS signal indicators is liked to the same info (C1:IOO-MC1_PIT_OUTPUT).
Fix this, Jenne! Baaaaagghhhhh! 
|
My bad. As it turns out, you can't copy and paste between MEDM instances. It is now fixed. |
7454
|
Mon Oct 1 09:02:17 2012 |
Steve | Update | VAC | pumpdown reached vacuum normal |
Quote: |
The new cold cathode gauge CC1 is in place. We were at atmosphere for 28 days ......more later
|
cc1 = 2.3e-5 Torr at day 6 vacuum normal |
Attachment 1: pd73md6.png
|
|
Attachment 2: pd73md6b.png
|
|
7453
|
Mon Oct 1 07:25:48 2012 |
Steve | Update | SUS | MC2 & ETMX sus damping restored | |
7452
|
Fri Sep 28 21:17:41 2012 |
Koji | Update | IOO | MC WFS adjustment | MC WFS was fixed. Now it is running constantly with the autolocker.
Found a bug in the IOO screen: All of the 6 WFS signal indicators is liked to the same info (C1:IOO-MC1_PIT_OUTPUT).
Fix this, Jenne! Baaaaagghhhhh! 
What I did:
1. C1:IOO-MC_RFPD_DCMON indicator was saturating. "HOPR" of this entry was set to 5 by running the following command:
ezcawrite C1:IOO-MC_RFPD_DCMON.HOPR 5
2. Scan MC2 spot position by using /opt/rtcds/caltech/c1/scripts/MC/moveMC2 scripts.
or the adjustment, C1:SUS-MC2_ASCPIT_EXC and C1:SUS-MC2_ASCYAW_EXC were excited with 300cnt at 12Hz and 10Hz, respectively.
The corresponding peaks (i.e. ANgle to length coupling) in C1:IOO-MC_F were monitored on DTT and adjusted so that the peaks are approximately nulled.
3. moveMC2 scripts are not perect to keep the maximum of the transmission. So, the alignment was adjusted with MC1 and MC3.
4. Repeated 2 and 3 until the alignment converges.
5. Once I got satisfied with the MC2 spot position, I went to the MC2 table and aligned the steering mirror before the QPD.
6. As these actions above moves the REFL beam, I went to the MC REFL path and adjusted the MC REFL PD position and the MC WFS spot positions.
7. Checked if the alignment is still good. The MC REFL is 0.50~0.51. Pretty good.
8. Run /opt/rtcds/caltech/c1/scripts/MC/WFS/WFS_FilterBank_offsets to register the current WFS offset etc.
9. At this point, MC WFS started working fine. I also confirmed the autolocker worked with this setting.
--------
Checked how the things are going in the morning. There were several unlocks. But the autolocker and WFS kept the cavity lcoked again.
Very good.
Some power fluctuation of ~1% is observed in the MC trans. I checked the PMC trans and found it is also fluctuating by 1% in a coherent way.
So I judge the WFS itself is fine. (See attached)
|
Attachment 1: MC_WFS_12h.png
|
|
7451
|
Fri Sep 28 07:49:30 2012 |
Steve | Update | 40m Upgrading | New ETMX layout - Version 1.0 | Where is IP-ANG ? It is good practice to use two mirrors at launching and detecting the beam, so you can walk it - precisely adjust it.
The window can be replaced at ~$1,500 ea. 10 weeks as optical quality BK7 with dual AR |
7450
|
Thu Sep 27 20:05:57 2012 |
Koji | Update | 40m Upgrading | New ETMX layout - Version 1.0 | 1. I wonder how the mode profiling/matching was considered in the new layout.
I can see the distances between the components and lenses are largely different from the old ones.
This is OK if you plan to go through a new mode matching solution with new lenses.
But it takes
a certain amount of time.
Note that we don't care the distance after the last lens as the Rayleigh range there is supposed
to be long enough to allow this kind of change.
2. The huge frustration of the green alignment in the old setup was caused by the 3D beam steering
at the last two 2" mirrors. i.e. the beam elevation on the table does not match with the beam elevation of the cavity.
In order to avoid this, I suggest you to use three 45 deg 2" mirrors instead of two. In fact these mirrors are supposed to be used at 45deg incidence!
3. The incident green beam and the transmitted IR beam should share a same path as they
share a same cavity mode.
This means that you should use a harmonic separator for the transmitted light pick-off.
4. Use the harmonic separator for the fiber path too. Get the mirror spec from Jamie.
5. Since the optical window on the chamber has a wedge angle, the beam paths are not straightforward.
The cavity beams can't be moved as they are constrained by the arm cavity.
Probably there is almost no freedom to move even for the oplev beams.
It would be safe just to follow the old positions and angles on the window.
Make sure the beam on the drawing is realistic. The angles of the oplev beams in the old setup look strange.
Is there a possibility to replace the optical window so that it has an AR for 532 and 1064 at least???
6. I wonder if the rejected beam by the Farady have a realistic angle or not. Check it with the old setup.
It is definitely better to have a steering mirrror and a lens before the refl PD.
7. The IR QPD and trans PD are intended to be used for the low and high power detection.
I forgot which is which. So check the range of them and think about the power distribution.
8. We should have separated CCDs for IR Trans and Green Refl.
We had a terrible ghost green beam on the IR trans CCD.
Thus, think about the amount of ghost reflection and consider filtering if necessary. |
7449
|
Thu Sep 27 19:03:06 2012 |
Manasa | Update | 40m Upgrading | New ETMX layout - Version 1.0 | I am working towards redesigning the endtables. I've attached the first version of the layout. As per Steve's comment I've tried to leave a 2" empty space on all sides of the table. It still has to be updated with the whole 40m layout to be more precise about the pickoff and the ingoing beam directions. |
Attachment 1: ETMX_OldNew.png
|
|
7448
|
Thu Sep 27 17:00:41 2012 |
Eric Gustafson | Update | LSC | 40 meter photodiode frequency response measurement system installation | Mike and I installed all of the telescopes and launching hardware for REFL11, REFL33, REFL55, AS55, MCRef, POX11 and POP55. On Monday afternoon Steve will work with us on the fiber routing. Steve is buying some protective covers for the fibers.
|
7447
|
Thu Sep 27 16:26:11 2012 |
Steve | Update | LSC | RF fibre protection in cable trays |
Quote: |
Jenne, Mike and I installed all of the post holders we could today including: REFL11, REFL33, REFL55, AS55, MCRef, POX11 and POP55. We did not install AS110, POY or REFL165 because there are interferences that will require moving stuff around. We also did not mount POP22 because it is a peely wally ThorLabs PD that will be replaced by a strong, straight and right thinking LIGO PD in the fullness of time. We did move it out of the way however which is no more than it deserves. Next step this afternoon Mike and I will install all of the telescopes and launching hardware. Then with the help of Steve we will begin routing the fibers. The 1x16 splitter module will be here by next Monday, the laser by the following Friday and then we will light up the fibers.
|
I'm proposing split loom tubing that would run in the cable tray to protect the fibers inside of it. This tubing diameter in the cable tray can be 1.5-2" and out of the tray 0.75"
|
7446
|
Thu Sep 27 15:21:04 2012 |
Steve | Update | SUS | TT height adaptor & PTICH damping |
Quote: |
[Koji, Steve, Den]
TT alignment is fine, yaw damping is satisfactory, pitch damping is slow. We might want to add magnets to the mirror and attach blades to the frame for pitch edge current damping.
We are moving towards electronics testing.
|
Atm1, TT 1.5" high adaptor base will be back from the shop in 10 days.
Atm2, There is no PITCH damping, YAW edie current damping works well at 0.5 mm gap
Atm3, Adjustable Al -disc that contains a small magnet is purely designed.
We have to come up with a solution to have damping in PITCH |
Attachment 1: 09271201.PDF
|
|
Attachment 2: IMG_1656.JPG
|
|
Attachment 3: IMG_1654.JPG
|
|
7445
|
Thu Sep 27 13:05:55 2012 |
Eric Gustafson | Update | LSC | 40 meter photodiode frequency response measurement system installation | Jenne, Mike and I installed all of the post holders we could today including: REFL11, REFL33, REFL55, AS55, MCRef, POX11 and POP55. We did not install AS110, POY or REFL165 because there are interferences that will require moving stuff around. We also did not mount POP22 because it is a peely wally ThorLabs PD that will be replaced by a strong, straight and right thinking LIGO PD in the fullness of time. We did move it out of the way however which is no more than it deserves. Next step this afternoon Mike and I will install all of the telescopes and launching hardware. Then with the help of Steve we will begin routing the fibers. The splitter module will be here by next Monday, the laser by the following Friday and then we will light up the fibers.
|
7444
|
Wed Sep 26 23:55:14 2012 |
Jenne | Update | IOO | WFS centered | Since the MC spots are good, I put the beam back on WFS 1 and WFS 2.
Also, I changed the indicators on the LockMC screen to reflect the change in elog 7289, where we added another on/off switch for the WFS so that the ASS could be on, but the WFS off. For the last month, the WFS could be disabled, but the MC screen's indicators would suggest that we were pushing very significantly on all 3 MC mirrors. Now the MC screen reflects reality a little better.
I also uncommented the WFS lines in the mcup script. Den had commented them out, but didn't elog about it! C'mon Den, please elog stuff!!!! (He confessed out loud the other day, but it still wasn't in the elog).
I'm leaving the WFS loops disabled (even though the MC autolocker tries to turn them on, I have them manually disabled using the extra on/off switch) since they're unstable. I'm in the process of figuring out what's wrong. So far, the WFS improve the MC alignment for a minute or two, and then they totally misalign the MC. This is a work in progress. |
7443
|
Wed Sep 26 17:09:15 2012 |
Den | Update | SUS | TT | [Koji, Steve, Den]
TT alignment is fine, yaw damping is satisfactory, pitch damping is slow. We might want to add magnets to the mirror and attach blades to the frame for pitch edge current damping.
We are moving towards electronics testing. |
7442
|
Wed Sep 26 16:59:30 2012 |
jamie | Update | IOO | IPANG ND filter installed |
Quote: |
[Whomever took away this ND filter without elogging it was BAD!!! (Jamie, when we first found IPANG coming out of the vacuum during this vent, we moved some of the mirrors on the out-of-vac table in the IPANG path. Was the ND filter removed at that time? Or has it been out for much longer, and we never noticed because IPANG wasn't coming nicely out of the vacuum / was clipping on the oplev lens?)
|
I do not remember removing anything from that setup. We just moved some mirrors and lenses around |
7441
|
Wed Sep 26 11:03:04 2012 |
Steve | Update | VAC | pumpdown completed | The new cold cathode gauge CC1 is in place. We were at atmosphere for 28 days ......more later
|
Attachment 1: 40mpd73md1.png
|
|
Attachment 2: 28datm.png
|
|
7440
|
Wed Sep 26 01:10:34 2012 |
Jenne | Update | IOO | PZT2 not working?!?! MC back to normal | [Jenne, Evan, Den]
MC REFL beam is back on the PD, and the mode cleaner locks. It looks like we're a little high on the MC Refl camera, but the MC spots were measured, and don't look like they changed from Friday (or maybe Monday?), the last time they were measured. We took this to be acceptable MC alignment, and did not touch the PSL table's pointing.
The laser power reduction optics were removed, and placed out of the way on the PSL table (where do they belong?). PSL-POS and PSL-ANG aren't quite perfectly centered, but a beam dump had been in the way of that path, so I don't know if they drifted bad, or if it was a sudden thing. The beam is still hitting the QPDs though. After removing the beam power reducing optics, we recentered the MC REFL beam on the REFL PD, still not touching any PSL alignment. MC mirrors were aligned (Den did this work while I showed Evan around, so I don't know by how much), and MC Trans was maximized (really MC Refl was minimized, making sure that the unlocked MC Refl was the usual 4.something units on the EPICS readback.
We turned on the PZT high voltage supplies for the output steering PZTs and for the input steering PZTs. We left the OMC locking PZT supplies off, since we're still not using the OMC. Sadly, the beam coming out of the AS port looks clipped somewhere. [SELF: attach the videocapture shot when you get to work tomorrow] We tried moving PZT2's sliders, but nothing happened!!! I can move BS and the ITMs to get the beam mostly unclipped, but I really need to be able to move the PZTs, or at least one of them. IPPOS and IPANG beams are hitting the QPDs (although they're not centered perfectly), so the PZTs came back mostly to the same positions, but not exactly. Evan and I sat next to the input steering PZT controllers in 1Y3, and moved the sliders around. For most of the range, nothing changes on the LCD screen for either PZT2 pitch or yaw. Yaw can make 2 small steps near the far negative side of the slider, but nothing happens for most of the slider. Pitch really doesn't do anything for any part of the slider. We ensured that the LED labeled "CL ON" was not illuminated, next to the button labeled "closed loop", for all 4 controllers (PZT1 and 2, pitch and yaw). Sad!! I don't know if the LCD screen on the front panel of the PZT controllers is a readback of signal supplied to the PZTs, or of the strain gauges. I really hope it's the controller that's not working, rather than the PZTs themselves. The PZTs were fine before we vented, and Koji and I did our centering of the PZT range check during the vent, so they were fine then. What happened??? All PZT high voltage supplies were off during the pump-down. I turned them off yesterday, and Evan and I turned them back on tonight around 9:30pm or 10pm. What else could make them bad?
Without being able to move PZT2, just using BS and / or ITMs, I was unable to completely make the beam look nice on the AS camera. I came close, but it still seems a little bit funny, and I had to move the suspended optics quite a bit to find that place. This is not good. |
7439
|
Tue Sep 25 22:40:55 2012 |
Jenne | Update | IOO | IPANG ND filter installed | [Jenne, Evan Hall]
Both IPPOS and IPANG beams are (after turning on the input and output PZTs) hitting their QPDs. However IPANG was saturating. We went down to take a look, and we had ~2.8mW incident on the QPD. There was an ND filter sitting unmounted, next to the diode, and an empty fork directly in front of the diode. Since IPPOS also has an ND filter in front, we stuck this ND filter back in. Now we are no longer saturating.
We're not hitting (yet) the center of these 2 PDs, but we're at least hitting the diodes, so it shouldn't be too hard to steer the input PZTs.
Whomever took away this ND filter without elogging it was BAD!!! (Jamie, when we first found IPANG coming out of the vacuum during this vent, we moved some of the mirrors on the out-of-vac table in the IPANG path. Was the ND filter removed at that time? Or has it been out for much longer, and we never noticed because IPANG wasn't coming nicely out of the vacuum / was clipping on the oplev lens?)
|
7438
|
Tue Sep 25 17:21:01 2012 |
Steve | Update | VAC | pumpdown completed | IFO P1= 1mTorr, CC1 = 5e-4 Torr. The IFO is ready for work. Be free to open the shutter and turn on HV
Jenne will double check the MC -REFL path for 1W power. There is a manual block on the PSL table.
Atm1, Pirani gauge at 9 h 10 min
Atm2, Pump configuration of pd#73 at 9h 35 min
|
Attachment 1: pd73.png
|
|
Attachment 2: pd73b.png
|
|
7437
|
Tue Sep 25 14:29:07 2012 |
Jenne | Update | General | POP in-vac optics aligned, POY beam coming out of vac |
Quote: |
What was the reasoning / resolution of the POP forward/backward beam? Are we going to have the right beam for DRMI locking?
|
From Koji's email to me:
"With the backward beam you can see the returning beam even when the PRM is misaligned. That's the only difference. Once the PRM is aligned both beams have the same information."
So, we should be fine. |
7436
|
Mon Sep 24 23:45:39 2012 |
rana | Update | General | POP in-vac optics aligned, POY beam coming out of vac | What was the reasoning / resolution of the POP forward/backward beam? Are we going to have the right beam for DRMI locking? |
7435
|
Mon Sep 24 20:28:13 2012 |
rana | Summary | elog | restarted elog: was unresponsive | |
7434
|
Mon Sep 24 17:47:05 2012 |
Steve | Update | VAC | ready to pump | The IFO will be at atmosphere overnight. The annuloses are pumped down. The access connector jam nuts are tightened to 45 ft/lbs
Actual pumpdown will start tomorrow morning. |
7433
|
Mon Sep 24 17:03:39 2012 |
Jenne | Update | PEM | 1X1 rack power |
Quote: |
I've installed Guralp readout box back and it turned out that it does not work with voltage provided from the rack (+13.76 0 -14.94). +/-12 voltage regulators inside the box convert it to -0.9 0 -12. I've connected the box to +/-15 DC voltage supply to measure seismic motion at the ETMY table. Readout box works fine with +/- 15.
|
I'm not sure what the problem is here. Den and I looked at it for a few minutes, before I went back to helping with putting doors on. The Sorensons are not supplying the rack power for 1X1. There are some flat cables which go from the fuses on the side of the rack up to the cable tray, and go elsewhere. Den is going to continue looking into this, but I think it's a moderately high priority, since lots of things should be getting served by that same power. |
7432
|
Mon Sep 24 16:59:31 2012 |
Jenne | Update | VAC | Doors on, ready to pump | [The 40m Family]
The access connector and all heavy doors are back on.
Jamie put the regular viton EQ stops back on PRM, since he had to adjust the distance between the EQ stops and the PRM anyway. Jamie also waved an IR card near the IPANG steering mirrors in ETMY, but it was not possible to take a good photo. Jamie certifies that the beam is centered on both of those 2 optics.
I have centered IPPOS and IPANG QPDs.
All oplevs need a little realignment, especially ETMY, which had it's lens removed (Rana has a Wall of Shame photo of this, which is why it was removed by him). Steve will look into this tomorrow, after he starts pumping.
I have turned off all PZT high voltage supplies for in-vac PZTs: The input PZTs, the output PZTs, and the OMC PZTs (which weren't on, but I confirmed they were off).
I have also prepared the 3 low-power items for high power: MC refl's path was changed back to regular BS, AS camera was moved to its nominal position, and IPPOS has its ND filters back. MC refl and the AS camera will need to be realigned once we're actually at high power tomorrow afternoon.
Long vent, but good work everyone. |
7431
|
Mon Sep 24 10:27:50 2012 |
Steve | Omnistructure | General | do not leave stuff on optical table tops | The SP table was found open this morning. Please, do not make optics dirty!
I cleaned up the tops of the SP table.
Stop storing your junk, boxes, laptops, etc. on the optical tables. This includes the big SP table. Please move all of that junk into racks or shelves, etc. |
7430
|
Sun Sep 23 22:40:48 2012 |
Den | Update | Modern Control | MC_L locking | I've applied LQR approach to MC_L locking. Results show that LQR does not make MC_F signal smaller below 0.3 Hz in contrast with classical locking. This might indicate that in this frequency range we see sensing noise as LQR was provided with state-space model of MC only so it tries to reduce displacement noise. It is also possible that state-space model is not accurate enough.
|
Attachment 1: LQR_MCL.pdf
|
|
7429
|
Sat Sep 22 01:03:30 2012 |
Den | Update | PEM | ETMY table | I've installed Guralp readout box back and it turned out that it does not work with voltage provided from the rack (+13.76 0 -14.94). +/-12 voltage regulators inside the box convert it to -0.9 0 -12. I've connected the box to +/-15 DC voltage supply to measure seismic motion at the ETMY table. Readout box works fine with +/- 15.
Seismic noise on the ETMY table measured to be a few times higher then on the floor in horizontal direction in the frequency range 50 - 200 Hz. Attached are compared spectrums of X, Y and Z motions. |
Attachment 1: X.pdf
|
|
Attachment 2: Y.pdf
|
|
Attachment 3: Z.pdf
|
|
7428
|
Fri Sep 21 22:43:51 2012 |
Den | Update | SUS | TT alignment | I did TT alignment using red laser and QPD.

I had a problem aligning TT with frame number SN-035 as some screws are damaged so all what I could move were 2 blades on the sides of the mirrors. But this was not enough to align pitch and yaw simultaneously. It is possible to align pitch only, but then I got a huge yaw angle (~0.05-0.1 rad). The only option I had was to make a reasonable alignment in yaw and then suspend several washes on the screw on the bottom of the mirror to align pitch.
Attached are flag positions inside coils. 1 - SN034, 2 - SN012, 3 - SN006, 4 - SN035. For each TT there are 4 pictures with flag 1-4: UL, LL, UR, LR |
Attachment 3: 1.pdf
|
|
Attachment 4: 2.pdf
|
|
Attachment 5: 3.pdf
|
|
Attachment 6: 4.pdf
|
|
7427
|
Fri Sep 21 22:25:44 2012 |
Mike J. | Update | General | POX, POY, PR2 pics | Unaltered PR2 images, with IR card, without card, and steering mirror:
  
Unaltered POX and POY images:
 
The POX images only needed a major brightness reduction and increased contrast to view:
 
The POY images needed their intensity histograms shifted slightly right and made left-tailed:
  
|
7426
|
Fri Sep 21 20:48:24 2012 |
Jenne | Update | General | POP in-vac optics aligned, POY beam coming out of vac | Getting POP:
We put a green laser pointer at ~4 inches on the POX table, and steered it using a mirror on the POX table to hit the center of the last in-vac mirror that POP sees. I then steered that mirror so we were hitting the center of the other POP in-vac steering mirror, and hitting the same spot as the main IR beam. It is easy to hold an IR card in front of PR2 and see the IR and green beams simultaneously. I aligned both of the POP in-vac steering mirrors such that the green beam is co-aligned with the IR beam at PR2, as well as as far as I could reach toward the face of PRM from the ITMX door.
Note: The drawings by Koji have the POP "forward" beam (transmission through PR2 of the beam from PRM to PR2) dumped, while the POP "backward" beam (transmission through PR2 of the beam from PR3 to PR2) leaving the vacuum. I aligned the steering mirrors such that the 'forward' beam would come out, although no dump is in place to dump the other beam. I can't think of a reason why we care one way or the other, but I feel like Koji has perhaps mentioned something in the past. I need to figure this out before we put doors on.
Getting POY:
Like yesterday with POX, we used the Watec with the aperture fully open to look at the POY pickoff, while I held the IR card in front of the mirror, to confirm that the beam was ~on the center of the optic. Then we took the lens off the camera, and made sure that the POY beam hit the CCD on the POY table.
To do list for Monday: While we are putting the heavy doors on, someone needs to wave an IR card in front of the IPANG steering mirrors in the ETMY chamber, while someone else takes a photo / still snapshot with the Watec. Also, Manasa wanted to retake in-vac photos of at least the ITMY chamber, since SR2 was moved a very slight amount. Also, also, someone tall needs to put the regular EQ stops on the PRM face (we have the old spring ones in there now).
Before pumpdown, we also need to get the IPANG beam centered on the PD. The beam is cleanly coming out of the vacuum and hitting the first out of vac steering mirror, I just haven't centered it onto the QPD.
Barring any other thoughts that people have of things that *must* be done before we pump down, I think we're ready to start putting heavy doors on the chambers on Monday.
Other thoughts, for next vent: We need to re-look at the ITMY table. POY's pickoff is just too close to the main beam. Is it possible to move the AS steering mirrors and get POY from the BS table? VENT CZAR: please put looking at this on the next vent to-do list. |
7425
|
Fri Sep 21 12:12:56 2012 |
Manasa | Configuration | PSL | AOM installation | {Jan, Manasa}
We installed the AOM driver back on the PSL table this morning. To calibrate the AOM RF output we connected a 1V dc to the modulation input of the driver and we are convinced with the setup.
Before we direct the rf signal to the AOM, in order to check its diffraction efficiency, we would like to setup an rf PD at the AOM output. We think we have place for a filter and PD after the AOM (replacing a beam dump) and would like to confirm the position before we actually install them. The layout is the picture below showing sweet spots for the new pd to sit. If you think it may disturb the system in any way, let us know!

|
7424
|
Thu Sep 20 22:52:38 2012 |
Den | Update | MachineLearning | Feedback controller |
Quote: |
I have uploaded to my directory a directory neural_plant. The most important file is reference_plant.c, which compiles with the command
|
We would appreciate some plots. Learning curves of recurrent NN working as a plant are interesting. For harmonic oscillator your RNN should not contain any hidden layers - only 1 input and 1 output node and 2 delays at each of them. Activation function should be linear. If your code is correct, this configuration will match oscillator perfectly. The question is how much time does it take to adapt.
Does FANN support regularization? I think this will make your controller more stable. Try to use more advanced algorithms then gradient descent for adaptation. They will increase convergence speed. For example, look at fminunc function at Matlab. |
7423
|
Thu Sep 20 20:07:38 2012 |
Jenne | Update | General | REFL, IPANG are coming out of the vac |
Quote: |
[Jenne, Unni, Manasa]
I touched some in-vac steering mirrors, so we have REFL and IPANG coming out of the vacuum, not clipping. IPPOS was done yesterday. I re-checked a few optics in the AS path that were hard to see yesterday while the plastic light access connector was in place, and AS still looks good.
Except for POX, POY, POP, and putting the regular EQ stops back on PRM, I think we're done with the in-vac stuff.
|
[Rana, Jenne, Manasa]
POX is coming out of the vacuum. We'll do POY tomorrow. We were able to hold the Watec outside the chamber and focus it on the pickoff mirror, and make sure it was roughly centered. Then we took the lens off the camera, put the camera in the POX beam path, and I steered the pickoff mirror until we were hitting the camera. POY will be done the same way.
POP is more challenging, since the transmission of the G&H mirrors is so low. We're not able to see a beam on an IR card held in the POP beam path. I had thought of removing PR2, getting the beam out, then putting PR2 back (using the same dog clamping some alignment markers technique that we use for the test masses), but the G&H mirrors have a 2 degree wedge, so this won't work. It would be fine for pitch, since the arrow is on the side of the optic, but it wouldn't be correct for yaw.
Maybe we should do something similar to what Suresh et. al. did when they set POP up originally - I think they put a green laser pointer on the POX table, and aligned it such that they were hitting the correct spot on PR2 and PRM (correct = the same as the IR spot, which should be the center of the optics). If we can do that with the POP in-vac steering mirrors, then we're fine, and POP should come out when we're back to high power.
All video capture snapshots of tonights pictures are on the pianosa desktop. |
7422
|
Thu Sep 20 19:56:05 2012 |
Jenne | Update | General | Pickoffs are hard to see |
Quote: |
Quote: |
My hope is that the DRMI flashes will be bright enough to see on the PO beams. IF we get 10 mW through the Faraday, you should get some buildup when the carrier resonates in the DRMI.
If the recycling gain is 10 and the pickoff fraction is 100 ppm you ought to get ~10 uW on PO. How much of the recycling cavity power gets out of POP?
|
[Manasa, Jenne]
We think this math is wrong.
If we have P mW through the Faraday, PRM's transmission is 5.5%, BS transmission is 50%, Recycling gain is ~10, pickoff fraction is ~100ppm, we have:
P mW * 5.5e-2 * 0.5 * 10 * 100e-6 = P * 2750e-8 mW = P * 2.7e-5 mW.
So, if P=10 (10mW through the Faraday), we should have 2.7e-4 mW = 2.7e-7 W = 0.27 microwatts = not so many watts.
If P = 100 (100mW through the Faraday), we should have 2.7 microwatts. Still, not so many watts.
We have the Watec pointed at POY right now, DRMI is flashing, I'm waving the IR card in front of the mirror, and Manasa isn't able to see anything on the monitor. The power into the vacuum is 100mW (we just measured and adjusted it), so even if we were getting a full 100mW through the Faraday, it would be hard to see. If we're assuming we get ~half the power through the Faraday, then we should only have 1 microwatt
|
We can't mathdo  |
7421
|
Thu Sep 20 17:05:26 2012 |
Jenne | Update | General | REFL, IPANG are coming out of the vac | [Jenne, Unni, Manasa]
I touched some in-vac steering mirrors, so we have REFL and IPANG coming out of the vacuum, not clipping. IPPOS was done yesterday. I re-checked a few optics in the AS path that were hard to see yesterday while the plastic light access connector was in place, and AS still looks good.
Except for POX, POY, POP, and putting the regular EQ stops back on PRM, I think we're done with the in-vac stuff. |
7420
|
Thu Sep 20 14:55:06 2012 |
Jenne | Update | General | Pickoffs are hard to see |
Quote: |
My hope is that the DRMI flashes will be bright enough to see on the PO beams. IF we get 10 mW through the Faraday, you should get some buildup when the carrier resonates in the DRMI.
If the recycling gain is 10 and the pickoff fraction is 100 ppm you ought to get ~10 uW on PO. How much of the recycling cavity power gets out of POP?
|
[Manasa, Jenne]
We think this math is wrong.
If we have P mW through the Faraday, PRM's transmission is 5.5%, BS transmission is 50%, Recycling gain is ~10, pickoff fraction is ~100ppm, we have:
P mW * 5.5e-2 * 0.5 * 10 * 100e-6 = P * 2750e-8 mW = P * 2.7e-5 mW.
So, if P=10 (10mW through the Faraday), we should have 2.7e-4 mW = 2.7e-7 W = 0.27 microwatts = not so many watts.
If P = 100 (100mW through the Faraday), we should have 2.7 microwatts. Still, not so many watts.
We have the Watec pointed at POY right now, DRMI is flashing, I'm waving the IR card in front of the mirror, and Manasa isn't able to see anything on the monitor. The power into the vacuum is 100mW (we just measured and adjusted it), so even if we were getting a full 100mW through the Faraday, it would be hard to see. If we're assuming we get ~half the power through the Faraday, then we should only have 1 microwatt |
7419
|
Thu Sep 20 11:39:40 2012 |
Manasa | Update | General | DRMI aligned again, but with good arms |
QPD at IPPOS has been centered by removing the filter at the QPD.
So we need to remember to check back on AS camera path and the IPPOS as well in addition to the usual MCrefl path.
P.S. We would be happy to have a new laptop in the lab to replace "Belladonna"! |
7418
|
Thu Sep 20 08:50:14 2012 |
Masha | Update | MachineLearning | Machine Learning Update | Hi everyone,
I've been working a bit on neural network code for a controller, and thus far I have code that creates a reference plant neural network. This is necessary to perform a gradient-descent learning algorithm on a controller neural network (one that reads an error signal and outputs actuation force). Because the error signal is read after the previous output of the controller neural network has passed through the plant, in order to calculate the gradient, either the inverse of the plant needs to be calculated, or the plant can be simulated through a neural network, and the error signal can thus be back-propagated through the plant neural network to find the gradient with respect to the output (as opposed to with respect to the plant), and thus back-propagated through the controller network in order to learn.
I have uploaded to my directory a directory neural_plant. The most important file is reference_plant.c, which compiles with the command
gcc reference_plant.c -o reference_plant -lfann -lm
The code runs on a file called reference_plant.data, which consists of a series of delayed inputs i_1, i_2, i_3 ... i_{n - 1} of the plant signals and then an output that is i_{n}, the subsequent signal. Parallel streams may also be used, if more than one signal is to be read. The top of the file must contain the number of total training packets (input-output groups), followed by the number of inputs, and the number of outputs. reference_plant.c also has constant variables which specify the number of hidden neurons, which can be changed.
All of this code runs on the FANN library. If the code doesn't seem to be compiling, then it means the library might have to be downloaded and built from source.
Thus far, I have created my own plant in simulink (the driven, damped harmonic oscillator, as before), and obtained results of 0.0002929292 training MSE after 5 epochs (subsequently lowered to 0.000) and 0.000 training error. This, however, is due to the fact that my plant is overly simple, and seems only to need 3 time-delayed plant signals, rather 31 to specify it (since all motion is second-order).
It should be fairly easy to use interferometer signals as input to this plant by just reading some signals and parsing them into time-delayed groups. (I tried this over the summer with my previous code, and it seemed to work, although I haven't accessed any of the channels to obtain data lately).
In terms of LIGO stuff this week, I'm going to be finishing up (writing) my final report, but please let me know if you have any comments or concerns.
Thanks! |
7417
|
Thu Sep 20 08:34:46 2012 |
Manasa | Update | General | DRMI aligned again, but with good arms |
Quote: |
[Jenne, Manasa]
Using the alignment of the PZTs and BS from pre-dinner, where the beam was hitting the center of both ETMs, we aligned the DRMI. The beam was off on the SRM in yaw by ~half a beam diameter, so I undid Koji's movement of SR2 from a week ago. I loosened the SR2 dog clamps, touched it gently on the base to do a little bit of angle, then re-clamped it. Once again, Steve's new brass centering target was awesome, since it was on the SRM while I was moving SR2.
We approximately recentered the beam on the AS camera, although it didn't need much once we got the beam out of the vacuum, by centering it on all of the output AS path mirrors.
We also got IPPOS out of the vacuum. Manasa was in the process of centering the QPD when the laptop died from too long being unplugged, so we leave that for tomorrow.
Left to do:
REFL path. REFL is not coming out of the vacuum, and with the light access connector I can't reach any of the REFL steering mirrors, since they're in the center of the IOO table.
IPANG. Should be easy.
POP, POX, POY. Need to the the camera-on-a-stick back down to the corner (from ETMY) and point it at the pickoff mirrors to ensure that beam is getting out of the vacuum.
|
Steve!! The light access connector got more ripped during the work last night ...we've just taped it back. We might need to figure out a better way to do this than just cutting through the cover. |
7416
|
Thu Sep 20 01:29:04 2012 |
Manasa | Configuration | PSL | AOM installation |
Quote: |
Mannasa and Unni and I looked at the RF driver for the AOM. It was fine.
With the ALC input left unconnected, with the power supply set to +28V, it was drawing 0.56 A.
By adjusting the modulation input we were able to get 1.1 Vrms into the scope (terminated at 50 Ohms) after going through 2 10dB attenuators. 11 Vrms into 50 Ohms is 33.8 dBm ~ 2W.
The RF power trimpot on the front of the driver is now adjusted so that -0.31 to 0.69 V takes the driver output from off to 2W output at 80 MHz.
The previous distorted signal that Jan and Manasa saw was at a level of ~100 mVrms, which is ~0.5 mW of power. At this tiny drive level, the internal amplifier is not linear and is mostly putting out a signal at ~160 MHz.
We checked by putting a square wave into the modulation input that the RF power from the driver would indeed shut off with a time scale of ~20 ns. Manasa will add a picture to this entry. We are ready now to calibrate the transmitted power of the AOM v. the modulation input voltage and then to measure the step time of the AOM.
Remember: do NOT believe the spec sheet of whatever PD you are using. All commercial PDs are slower than they advertise. In order to measure a <1 us step time you must use a PD with a >50 MHz 'bandwidth'.
|
|
Attachment 1: TEK00000.PNG
|
|
7415
|
Thu Sep 20 01:28:14 2012 |
Jenne | Update | General | DRMI aligned again, but with good arms | [Jenne, Manasa]
Using the alignment of the PZTs and BS from pre-dinner, where the beam was hitting the center of both ETMs, we aligned the DRMI. The beam was off on the SRM in yaw by ~half a beam diameter, so I undid Koji's movement of SR2 from a week ago. I loosened the SR2 dog clamps, touched it gently on the base to do a little bit of angle, then re-clamped it. Once again, Steve's new brass centering target was awesome, since it was on the SRM while I was moving SR2.
We approximately recentered the beam on the AS camera, although it didn't need much once we got the beam out of the vacuum, by centering it on all of the output AS path mirrors.
We also got IPPOS out of the vacuum. Manasa was in the process of centering the QPD when the laptop died from too long being unplugged, so we leave that for tomorrow.
Left to do:
REFL path. REFL is not coming out of the vacuum, and with the light access connector I can't reach any of the REFL steering mirrors, since they're in the center of the IOO table.
IPANG. Should be easy.
POP, POX, POY. Need to the the camera-on-a-stick back down to the corner (from ETMY) and point it at the pickoff mirrors to ensure that beam is getting out of the vacuum. |
7414
|
Wed Sep 19 23:17:25 2012 |
rana | Configuration | PSL | AOM installation | Mannasa and Unni and I looked at the RF driver for the AOM. It was fine.
With the ALC input left unconnected, with the power supply set to +28V, it was drawing 0.56 A.
By adjusting the modulation input we were able to get 1.1 Vrms into the scope (terminated at 50 Ohms) after going through 2 10dB attenuators. 11 Vrms into 50 Ohms is 33.8 dBm ~ 2W.
The RF power trimpot on the front of the driver is now adjusted so that -0.31 to 0.69 V takes the driver output from off to 2W output at 80 MHz.
The previous distorted signal that Jan and Manasa saw was at a level of ~100 mVrms, which is ~0.5 mW of power. At this tiny drive level, the internal amplifier is not linear and is mostly putting out a signal at ~160 MHz.
We checked by putting a square wave into the modulation input that the RF power from the driver would indeed shut off with a time scale of ~20 ns. Manasa will add a picture to this entry. We are ready now to calibrate the transmitted power of the AOM v. the modulation input voltage and then to measure the step time of the AOM.
Remember: do NOT believe the spec sheet of whatever PD you are using. All commercial PDs are slower than they advertise. In order to measure a <1 us step time you must use a PD with a >50 MHz 'bandwidth'. |
7413
|
Wed Sep 19 19:38:37 2012 |
Jenne | Update | General | Spot centered on BS, ETMY, ETMX | [Unni, Manasa, Jenne]
It turned out that the beam was a teeny bit high in the corner, so we touched PZT1 and PZT2 knobs to translate the beam down a bit.
Now the beam is centered on the BS (using the 45 degree non-iris target), centered on ETMY (using Steve's latest target, which worked perfectly), and then BS was aligned a tiny bit (really, it didn't need much) to get the beam centered on ETMX.
After dinner I'll align ITMX and ITMY such that their beams retroreflect and I get MICH fringes. I'll also align SRM and PRM to retroreflect. Check no clipping on AS path, get REFL path out, center IPPOS and IPANG, check POX, POY and POP. Then, I think we might be almost done. |
7412
|
Wed Sep 19 17:48:47 2012 |
Den | Update | Modern Control | ETMX | Time domain control using LQR technique is now applied to ETMX sus position. The plan was to do it for oplevs, I'll do it after the vent.
The cost function for state space variables was determined by TF 900 / (s + 30)^2. There was no penulty imposed for velocity, only for position. We can try that configuration as well.

|
7411
|
Wed Sep 19 15:41:27 2012 |
Manasa | Configuration | PSL | AOM installation |
AOM driver has been removed from the PSL table for testing. However the AOM is still inside; so there should be no problems with the alignment. |
7410
|
Wed Sep 19 13:12:48 2012 |
Jenne | Update | IOO | Power into vacuum increased to 75mW | The power buildup in the MC is ~400, so 100mW of incident power would give about 40W circulating in the mode cleaner.
Rana points out that the ATF had a 35W beam running around the table in air, with a much smaller spot size than our MC has, so 40W should be totally fine in terms of coating damage.
I have therefore increased the power into the vacuum envelope to ~75mW. The MC REFL PD should be totally fine up to ~100mW, so 75mW is plenty low. The MC transmission is now a little over 1000 counts. I have changed the low power mcup script to not bring the VCO gain all the way up to 31dB anymore. Now it seems happy with a VCO gain of 15dB (which is the same as normal power). |
|