  7952   Tue Jan 29 10:59:37 2013 lazy personFrogsGeneralbetter plan 
 I propose we work around this
problem with giant
  
  7951   Tue Jan 29 10:50:02 2013 JenneUpdateLockingVideo of PRM-flat test cavity 
I think Den accidentally edited and
overwrote my entry, rather than replying,
  
  7950   Mon Jan 28 21:36:44 2013 tall guyFrogsGeneralsmall people on noticeIf I catch anyone putting small booties
into the large bootie bin, I will make said
person eat small booties.
  
  7949   Mon Jan 28 21:32:38 2013 jamieUpdateAlignmenttweaking of alignment into half PRC[Koji, Jamie]
We tweaked up the alignment of the
half PRC a bit.  Koji started by looking
  
  7948   Mon Jan 28 19:15:14 2013 ManasaUpdateScatteringScattering setup [Jan, Manasa]
We are trying to get some scattering
measurements in the Y-arm cavity. We have
  
  7947   Mon Jan 28 19:07:45 2013 ManasaUpdateGeneralSN6 Laseroptik mirror - Tranmittance measurementsI repeated the measurements using NPRO
instead of Crystalaser. I am attaching optical
layouts for these measurements for future
 sn6_trans0128.pngtrans_layout.png 
  7946   Mon Jan 28 17:59:02 2013 RijuUpdate Photodiode transimpedanceSummary: Measurement and plot of
shot-noise-intercept-current for PDA255.
 shotnoiseinterceptpda255.pdf 
  7945   Mon Jan 28 17:01:19 2013 DenUpdateLockingVideo of PRM-flat test cavityWhat mode will you get if lock the cavity
PRM - ITMY/ITMX/TEST MIRROR without PR2,
PR3 and BS?
  
  7944   Fri Jan 25 08:10:42 2013 SteveUpdateGeneralLaseroptik mirrors Here are the German plots. Unfortunately
they are not logarithmic. 
 Ed: Proprietary data removed.
  
  7943   Thu Jan 24 16:34:56 2013 ranaUpdateGeneralLaseroptik mirror - SN6  I have two questions:
1) Are we sure that the T measurement
is not being compromised by some systematic?
  
  7942   Thu Jan 24 16:31:46 2013 SteveUpdatePEMbuilding exterier wall paintedThe wood exteier walls, gutters and doors
were painted at CES-Annex building #69
 IMG_1880.JPG 
  7941   Thu Jan 24 16:23:24 2013 KojiUpdateGeneralLaseroptik mirror - SN6The mirror T is completely out of spec.
We should find or request the data sheet
of the mirror.
  
  7940   Thu Jan 24 15:16:50 2013 ManasaUpdateGeneralLaseroptik mirror - SN6I repeated the transmittance measurements
of Laseroptik SN6 at 1064nm. The rotation
stage could only resolve 2 deg rotation (We
 sn6_trans1.png 
  7939   Thu Jan 24 14:40:09 2013 SteveUpdateVACRGA backgroung RGA background with VM2 open to Maglev
at day 37
 
 Screenshot.png 
  7938   Thu Jan 24 11:40:55 2013 JenneUpdateSUSETMX damping restored

    

       
            
  
  7937   Thu Jan 24 11:31:45 2013 SteveUpdateSUSETMX damping restoredETMX sus damping restored.  It is
still noisy
  
  7936   Thu Jan 24 08:28:35 2013 SteveUpdateVACwhen to pumpdown ?Reminder: I will be on vacation next week.
We would have to put the access connector
in tonight if I'm pumping down Friday, tomorrow.
  
  7935   Wed Jan 23 22:02:25 2013 ManasaUpdateGeneralLaseroptik mirror - SN6

    

       
            
  
  7934   Wed Jan 23 20:46:46 2013 Zen MasterUpdateLockingPR-flat cavity status - locks!

    

       
            
 yinYang.png 
  7933   Wed Jan 23 20:27:05 2013 RijuUpdate Photodiode transimpedanceToday I have repeated the expt for shot
noise intercept current. Koji found that
the Spectrum analyzer is going to saturation,
  
  7932   Wed Jan 23 20:24:05 2013 KojiUpdateGeneralLaseroptik mirror - SN6Got confused (even after I talked with
Manasa).
The plot shows the number ~0.01 or
  
  7931   Wed Jan 23 19:05:16 2013 JenneUpdateLockingPR-flat cavity status - locks!Status update
I (with help from Q) have redone
the POP path on the ITMX table.  1"
  
  7930   Wed Jan 23 18:16:11 2013 ManasaUpdateGeneralLaseroptik mirror - SN6 I repeated the transmittance measurements
of LaserOptik SN6 @1064nm.
Transmittance for s-polarization 
 sn6_trans.png 
  7929   Wed Jan 23 11:43:19 2013 RijuUpdate Photodiode transimpedance

    

       
            
 pda255shotnoiseintercept1.pdf 
  7928   Tue Jan 22 19:53:01 2013 JenneUpdateLockingPR-flat cavity status - not lockedThe PR-flat cavity is flashing, although
not locked.  I am too hungry to continue
right now.
  
  7927   Tue Jan 22 19:51:52 2013 KojiUpdate Photodiode transimpedance- The data should be plotted in a log-log
scale.
- The data points were only taken
  
  7926   Tue Jan 22 17:29:29 2013 RijuUpdate Photodiode transimpedanceRiju
Summary:  I am stuck with the
measurement of shot-noise-intercept-current
 pda255shotnoiseintercept.pdf 
  7925   Tue Jan 22 11:09:58 2013 SteveUpdateTip - TiltSS mirror holder orderedKoji's design of the SS  2" mirror
holder with flexure spring optic retainer 
like Polaris-K1 has been ordered. We are
 D1201574_v1_Tip-Tilt_Mirror_Holder.SLDDRW 
  7924   Tue Jan 22 09:18:50 2013 SteveUpdateTip - Tilttip tilt bases orderedThe corrected drawing base for tip tilt
with coils are going to the shop. The will
be back by the end of the week.
 TTbc.PDF 
  7923   Tue Jan 22 09:10:19 2013 SteveUpdatePEMbig foot is dirtyPlease wipe, clean car wheels and wear
booties entering the 40m lab.
Obviously this person has no idea
 IMG_1873.JPGIMG_1874.JPG 
  7922   Sat Jan 19 18:23:31 2013 ranaUpdateComputersAll front ends but c1lsc are downAfter sshing into several machines and
doing 'sudo shutdown -r now', some of them
came back and ran their processes.
 FE.png 
  7921   Sat Jan 19 16:02:28 2013 ranaOmnistructureElectronicsPS cleanupDuring our 'Women in Physics' tours today,
we were reminded that there are several bench
power supplies being used as permanent inside.
  
  7920   Sat Jan 19 15:05:37 2013 JenneUpdateComputersAll front ends but c1lsc are downMessage I get from dmesg of c1sus's IOP:
[   44.372986] c1x02: Triggered
the ADC
 upside_down_cat-t2.jpg 
  7919   Fri Jan 18 15:08:13 2013 jamieUpdateAlignmentalignment of temporary half PRC[jenne, jamie]
Jenne and I got the half PRC flashing. 
We could see flashes in the PRM and PR2 face
  
  7918   Fri Jan 18 12:08:08 2013 KojiUpdateLockingDust?No


  
  7917   Fri Jan 18 09:54:18 2013 JenneUpdateLockingPRM - Flat mirror cavity

    

       
            
  
  7916   Fri Jan 18 00:41:34 2013 JenneUpdateLockingDust?I was thinking tonight about more possible
reasons that our PRC sucks, and I wonder
if dust on the BS could create the problem.
  
  7915   Thu Jan 17 19:33:53 2013 ManasaUpdateAlignmentPRM oplev We had to work on redesigning the
oplev layout in BSC when I found that the
positions of the mirrors were clipping IPPOS
  
  7914   Thu Jan 17 17:04:49 2013 SteveUpdateVACvac monitor screen changedThe COVAC_MONITOR.adl was changed. Ion
pump labeled as disconnected means: ion pump
controller power turned off and ion pump
 screenmod.png 
  7913   Thu Jan 17 15:48:21 2013 JenneUpdateLockingPRM - Flat mirror cavity 
 2" G&H mirror is installed
on a DLC mount just in front of the BS. 
  
  7912   Thu Jan 17 11:01:19 2013 JenneUpdateAlignmentYesterday's alignment work[Jamie, Jenne, Manasa]
Yesterday's goal was to get the input
beam centered on the PRM, the BS and ETMY
  
  7911   Thu Jan 17 01:27:54 2013 Tara(?)SummaryIOONoise budget for MC

    

       
            
  
  7910   Thu Jan 17 00:17:31 2013 JenneUpdateLockingExpected PRC gains

    

       
            
  
  7909   Wed Jan 16 20:27:16 2013 ranaUpdateLockingExpected PRC gainsWhy would we use such a bad optic in our
recycling cavity? Is 1.5% the spec for these
mirrors? Is this the requirement that Kiwamu
  
  7908   Wed Jan 16 19:08:51 2013 KojiSummaryIOONoise budget for MCI missed the point. 
Do you mean that we can measure the coating
  
  7907   Wed Jan 16 18:58:08 2013 RijuUpdate Photodiode transimpedanceToday I have taken the reading for shot
noise intercept current for the PDA255 -
MC transmission RF PD. To do that I have
  
  7906   Wed Jan 16 18:52:49 2013 JenneUpdateLockingPRM - Flat mirror cavity planGame plan:
* Put 2" G&H mirror into
BS chamber, in front of BS.
  
  7905   Wed Jan 16 18:08:06 2013 JenneUpdateLockingExpected PRC gainsI was calculating the power recycling gains
we expect for different versions of the PRC,
and I am a little concerned that we aren't
  
  7904   Wed Jan 16 10:57:37 2013 taraSummaryIOONoise budget for MCI calculated thermal noise in mode cleaner
(MC) mirrors and compared it with the measured
MC noise. Thermal noise won't be a significant
 mc_nb_TN.pngmc_nb_TN.pngmc_nb_TN.figMC_nb.m.zipmc_nb_TN_2013_01_18.png 
  7903   Wed Jan 16 08:29:45 2013 SteveUpdateSUSPRM damping restoredPRM oplev servo turned off.  OLPIT
servo gain 0.15 and OLYAW  -0.3 set
to ZERO.  PRM damping restored
  