  12654   Thu Dec 1 08:02:57 2016 SteveUpdateLSCglitching ITMY_UL_LL 

 
 ITMY_UL_LL.png 
  12673   Thu Dec 8 07:56:05 2016 SteveUpdatePEMEQ6.5m Northen CANo damage. ITMY is glitching, so it has
not been damped.

 
 eq6.5FerndaleCA.png16d_glitching-_trend.pngEQ6.5_&4.7mFerndaleCa.png 
  12674   Thu Dec 8 10:13:43 2016 SteveUpdateLSCglitching ITMY_UL has a history 

 
 glitching__ITMY-UL_2007.png 
  12681   Thu Dec 22 09:37:20 2016 SteveUpdateVACRGA scan at day 63Valve configuration: vacuum normal

RGA head temp: 43.5 C

Vac envelope temp: 23 C
 pd80-d63.pngpd80-580Hz-d63.png 
  12687   Thu Dec 29 10:24:56 2016 SteveUpdatePEMEQ5.7mHawthorn NVSus damping restored.

 
 eq5.7HawthorneNV.png 
  12696   Mon Jan 9 09:18:47 2017 SteveUpdatePEMpower glitchThere was a power glitch last night around
1:15am

The vacuum was not effected.
 MondayMorning.png 
  12697   Mon Jan 9 16:12:30 2017 SteveUpdateGeneralOptical Layout in DCCCaltech Facilities promissed to email the
40m facility drawings in Cad format.

I organized the old of optical
 drawings_on_paper.jpg 
  12698   Tue Jan 10 14:24:09 2017 SteveUpdateVACRGA scan at day 82Valve configuration: vacuum normal

Vacuum envelope: 23C

Rga head: 44C
 pd80VNd82.png 
  12699   Tue Jan 10 16:20:11 2017 SteveUpdateCDSpower glitch......Raid is rebuildingJamie started the fm40m Raid rebuilding.
It has been beeping since the power outage.

Summary pages have no reading since
 rebuilding_in_progress.png 
  12705   Thu Jan 12 10:14:49 2017 SteveHowTosafetyclosing a doorThe door was not locked this morning.

Please do not use this door if
you can not close it!
 odc.jpghowtocloseadoor.jpg 
  12707   Thu Jan 12 13:45:58 2017 SteveHowTosafetyclosing a doorIt was requested this morning.




  
  12711   Fri Jan 13 10:53:03 2017 SteveUpdatePEMdoors are fixedControl room to outside door was
realigned. 

  
  12712   Fri Jan 13 14:18:28 2017 SteveUpdatePEMair condition fixedThe old control room AC  has been
stick in heating mode for
about 2 months. It's thermostate
 PEM_120d.png 
  12737   Thu Jan 19 08:25:12 2017 SteveUpdateSUSMC1 SUS electronics investigation


 MC1_MC3_ITMY_ETMX_sensors.pngsensors_UL.png 
  12743   Fri Jan 20 14:42:12 2017 SteveUpdatePEMCaltech weather stationWe should be able to connect to this station   
  12744   Fri Jan 20 17:12:37 2017 SteveUpdatePEMEM172 mic is hooked up in the PSLGautam and Steve,

It is hanging in the midle of the
PSL enclosure. Wired to 1X1 to get plus and
 EM172c.jpg 
  12745   Mon Jan 23 10:24:01 2017 SteveUpdateSUSMC1 SUS electronics investigationTwo day plot of glitching suspentions:
MC3, ITMY and ETMX
 3glitchingSUS.png 
  12747   Mon Jan 23 17:24:26 2017 SteveUpdateOptical LeversETMY Oplev HeNe is running hotETMY He/Ne 1103P  body temp is 
~45 C The laser was seated loosely 
in the V-mount with black rubber padding.
 ETMY_oplev__lasers.png 
  12760   Fri Jan 27 14:50:04 2017 SteveUpdateSUS wire standoffs updateThe 3 pieces of Sapphire
v-groove test cuts are back. They look
good. The suspension wire 0.0017" (
  
  12788   Thu Feb 2 12:17:48 2017 SteveUpdateGeneral USB microscopeThis
AmScope microscope would have 3.5x-180x
magnification, calibratable measurement function,
  
  12802   Mon Feb 6 10:05:28 2017 SteveUpdateSUSclamped cablesThe bottom 5 cable connections from Sat-Amp
to Whittering Filter at 1X5 were clamped
today.
 clamped.jpg 
  12808   Tue Feb 7 16:23:49 2017 SteveUpdateGeneralpower interruption tomorrow                                                                                                                                  
received this note: at 4:11pm Tuesday, Feb
7, 2017
  
  12813   Thu Feb 9 08:03:08 2017 SteveUpdateGeneral USB microscope orderedhttp://www.amscope.com/3-5x-180x-boom-stand-trinocular-zoom-stereo-microscope-with-144-led-ring-light-and-10mp-camera.html
will be ordered today.
  
  12843   Tue Feb 21 17:05:14 2017 SteveUpdateGeneralProjector lamp replacedThis bulb was
blown out on Feb 4, 2017 after 2 months
of operation.
 blownup.jpg 
  12848   Thu Feb 23 14:50:26 2017 SteveUpdateGeneral USB microscope returnedThe microscope shipped back to the vendor
for credit yesterday.

  
  12853   Mon Feb 27 15:33:10 2017 SteveUpdateVACRGA scan at day 130Valve configuration: vacuum normal

Vacuum envelope: 23.5 C

RGA head: 46.6 C
 rgascan@130d.png 
  12858   Wed Mar 1 08:28:04 2017 SteveUpdateOptical Leversoplev laser RINGautam and Steve,

New JDSU 1103P HeNe oplev laser
RIN was measured on the SP table with cover
 RIN_1103P.pngRIN_HeNe.png 
  12860   Wed Mar 1 17:25:28 2017 SteveUpdateLSCMCREFL condition picturesGautam and Steve,

 

Our MCREFL rfpd
 IMG_3646.JPGmcRefl_1.jpgmcRefl_3.jpgmcRefl_5.jpg 
  12863   Thu Mar 2 13:59:04 2017 SteveUpdateOptical Leversoplev laser RINThe laser got much better at low frequency
as it warmed up. This laser is almost as
good as the electronics?
 1103P@2d.png 
  12866   Fri Mar 3 17:24:21 2017 SteveUpdateOptical Leversoplev laser RINCorrected oplev laser RIN plot at
day 3

RXA:
 3march17.pdf 
  12868   Mon Mar 6 09:12:59 2017 SteveUpdatesafety crane inspection 2017All 3 cranes inspected by professional
Fred Goodbar of Konecranes and load tested
with 450 lbs at max reach on Friday, March
 inspection_2017.jpg 
  12871   Mon Mar 6 16:32:36 2017 SteveUpdateGeneralold NPRO 16 years old Lightwave NPRO M126-1064-700,
sn 415 power output is tripping continously
to zero.
  
  12872   Tue Mar 7 15:17:19 2017 SteveBureaucracyGeneralproperty tagProperty tag found.  property_tag.jpg 
  12873   Wed Mar 8 15:28:37 2017 SteveUpdateOptical Leversoplev laser RINGautam and Steve,




 RIN.jpgRIN_1103P_rotated.pdf 
  12876   Thu Mar 9 17:26:43 2017 SteveUpdateGeneralattempted ETMY picture takingI removed the video monitoring can and
replaced it with Olympus SP-570UZ camera.
It has no IR blocker. The OSEM light are
  
  12881   Fri Mar 10 18:00:22 2017 SteveUpdateGeneralattempted ETMY picture takingYour technique works Koji  P3100044.JPGP3090024.JPG 
  12884   Mon Mar 13 16:47:41 2017 SteveUpdatePEMY-arm air conditon belts replacedThe east end AC unit is arching over and
running rough at CES. Called for mechanic.......

Both belts were replaced and the
  
  12885   Tue Mar 14 09:08:11 2017 SteveUpdateOptical LeversETMX HeNe is deadETMX oplev laser is dead. It will be replaced
this after noon. Sus damping recovered.
 this_morning.png 
  12886   Tue Mar 14 10:40:30 2017 SteveUpdateSUSOSEM filters are inWe have 50 pieces in the clean cabinet.  filters.jpg 
  12888   Tue Mar 14 15:05:18 2017 SteveUpdateOptical Levershistory of ETMX oplev laser 




 ETMX_1103P_3y.png 
  12889   Thu Mar 16 08:22:16 2017 SteveUpdateSUSETMX damping Finally I see what kicks the sus damping
off

 laser_power_glitch.png 
  12890   Fri Mar 17 10:47:16 2017 SteveUpdateOptical Levers ETMX oplev laser replacedJDSU 1103P. sn T8070866, made March 2007,
output power 2.7 mW,  on pd 17,750 counts,

GV
  
  12892   Fri Mar 17 15:30:39 2017 SteveSummarySUSoplev laser summary updated   
     March  17, 
2017        
 oplev_sums.png 
  12908   Mon Mar 27 15:36:27 2017 SteveUpdatePEMparticle counter inside of PSL enclousure 




 HEPA_flow_effects.png 
  12909   Mon Mar 27 16:01:55 2017 SteveUpdatePEMX arm AC set to 68F 

The X arm air conditioner was not
regulating properly. The arm temp was warmer
  
  12913   Tue Mar 28 16:47:40 2017 SteveUpdatesafetyProjector bulb is out againThree replacement
bulbs ordered

Rana can discribe
  
  12915   Wed Mar 29 09:24:28 2017 SteveUpdateloresummery pagesThe summery pages are working at a slow
motion speed. It's response time 12 minutes.
  
  12917   Wed Mar 29 16:38:00 2017 SteveOmnistructureTreasuresus fiber illluminated   fiber.jpg 
  12921   Fri Mar 31 10:16:07 2017 SteveUpdatesafetylaser safety glasses annual inspectionLaser safety glasses cleaned in "
Dawn Ultra " mild soap - water solution
and measured for 1064 nm transmission at
 march2017.jpg2017march.jpg 
  12922   Fri Mar 31 16:10:33 2017 SteveUpdateTreasureLes Guthman Les Guthman interviews  gradstudent
Graig.

Main laser emergency shut off was
  
ELOG V3.1.3-