  6431   Tue Mar 20 17:50:44 2012 SureshUpdateComputersBeam Scan machine fixed

There was something wrong with the Beam Scan PC.  The  mouse and screen were not responding and the PC was asking for drivers for any new hardware that we plugged in.  We called in the services of Junaid and co. since we do not have a Win98 Second Edition installation disk in the lab.   Junaid came with the disk, we changed the screen and the mouse and installed everything. 

We tried to get the network going on the PC so that we could update stuff easily over the net.  This didnt succeed. For now, we still have to depend on a Win98se CD to get drivers if any new hardware is connected to this machine. 

For future reference, some notes:

1)  We will get a copy of Win98SE for the lab from Junaid

2) We have to use a USB mouse from Dell. We have several spares of this. The drivers for these are present in the machine. 

 

 

The Beam Scan is working okay now.  We will proceed with the beam profile measurements.  

  6432   Wed Mar 21 10:31:28 2012 steveUpdateSAFETYAC power to ETMY turned off

ETMY sus damping was disabled. Green locking laser and associated electronics turned off. Computers and power supplies turned off at rack 1Y4

The electricians picking up ac power from 1Y4 manual disconnect box and installing conduit line to ISCT-ETMY east end optical table.

There will be no more daisy chaining this way. 

  6433   Wed Mar 21 11:32:47 2012 steveUpdateSAFETYAC power to ETMX turned off

Quote:

ETMY sus damping was disabled. Green locking laser and associated electronics turned off. Computers and power supplies turned off at rack 1Y4

The electricians picking up ac power from 1Y4 manual disconnect box and installing conduit line to ISCT-ETMY east end optical table.

There will be no more daisy chaining this way. 

 The power is back on at ETMY . c1iscey has not been restarted.

Now I'm turning ac power off at ETMX for the same job to be done.

  6434   Wed Mar 21 19:12:27 2012 steveUpdateComputersAC power back on both ends

Quote:

Quote:

ETMY sus damping was disabled. Green locking laser and associated electronics turned off. Computers and power supplies turned off at rack 1Y4

The electricians picking up ac power from 1Y4 manual disconnect box and installing conduit line to ISCT-ETMY east end optical table.

There will be no more daisy chaining this way. 

 The power is back on at ETMY . c1iscey has not been restarted.

Now I'm turning ac power off at ETMX for the same job to be done.

 The power was turned back on at 4pm It took some time for Suresh to restart the computers. We have damping but things are not perfect yet. Auto BURTH did not work well.

  6435   Thu Mar 22 08:14:21 2012 steveUpdateGreen Lockingdichroic mirror quotes with large wedge

Quote:

Dichroic mirror quotes are in the wiki.

ATF is pricy.

We got a good price from Laseroptik, but the wedges are 5 arcminutes. The fused silica grade is 0F, meaning the  homogeneity is 5 ppm instead of 1ppm.  I requested an other large wedge quote on the substrates.We may have to get substrates from somebody else and ship it to Germany

MLT quote is outrageously high

REO is not interested in this low volume job.

 

 The Laseroptik quote is here.The 2 degrees wedge cost is $40 on each optics!  See wiki

  6436   Thu Mar 22 16:45:06 2012 kiwamuUpdateCDSc1scx and c1scy not properly running

It seems that neither c1scx nor c1scy is working properly as their ADC counts are showing digital-zeros.

However the IOPs, c1gcx and c1gcy look running fine, and also the IOPs seem successfully recognizing the ADCs according to dmesg.

Also there is one more confusing fact : c1scx and c1scy are synchronizing to the timing signal somehow.

I restarted the c1scx front end model to see if this helps, but unfortunately it didn't work.

As this is not the top priority concern for now, I am leaving them as they are now with the watchgods off.

(I may try hardware rebooting them in this evening)

Quote from #6434

The power was turned back on at 4pm It took some time for Suresh to restart the computers. We have damping but things are not perfect yet. Auto BURTH did not work well.

 

  6437   Thu Mar 22 17:35:59 2012 kiwamuUpdateLockingmode profiles of the POP and POX beams : not bright enough

I tried to measure the beam profiles at the POP and POX ports as Koji mentioned in his entry (#6421).

However it turned out that the beam powers were too small to be measured with our beam scan at those ports.

So I will move on to measurements at the REFL port as Rana suggested because the laser power is much larger than that of POP and POX.

(If the data of the POP and POX beam profiles turn out to be very necessary, we will do the razor blade technique with a more sensitive photo diode)

Quote from #6421

More precise analysis can be done with quantitative analysis of those two spots with Beamscan. This could happen tomorrow.

 

  6438   Thu Mar 22 17:41:15 2012 sureshUpdateCDSc1scx and c1scy not properly running

Quote:

It seems that neither c1scx nor c1scy is working properly as their ADC counts are showing digital-zeros.

Quote from #6434

The power was turned back on at 4pm It took some time for Suresh to restart the computers. We have damping but things are not perfect yet. Auto BURTH did not work well.

 When Steve and I restarted the c1iscex and c1iscey computers after the power shutdown, the models within them did not start-up automatically.  I had to start them manually from a terminal in the control room. 

I also tried rebooting the FB a couple of times.  Did not make any difference.

Manually starting the c1x05, c1scy and c1x01, c1scx models (with the Burt Restore button ON) did not resolve the issue of zeros in the epics screens.  though it did re-establish timing. 

  6439   Thu Mar 22 23:43:56 2012 KojiUpdateCDSc1scx and c1scy not properly running

Did you guys checked if the simplant switch is set to "REAL WORLD" mode?

Edit by KI:

Bingo ! The input signals were bypassed to the simplant. I switched the simplant settings to REAL WORLD and now both end suspensions are working fine.

  6440   Fri Mar 23 01:59:59 2012 kiwamuUpdateIOOREFL beam currently unavilable

[Suresh / Kiwamu]

Currently the REFL beam is bypassed by additional mirrors and blocked by a razor blade dump.

Therefore the signals associated with the REFL ports (e.g. REFL11, REFLDC and etc.) are unavailable.

Just be aware of it.

  6441   Fri Mar 23 05:10:46 2012 SureshUpdateIOOBeam Profile measurements: Errors too large to yield good fits.

[Kiwamu, Suresh]

   Today we attempted to measure the beam profile of the REFL beam under two conditions:

              (a) with PRM aligned and ITMs misaligned 

              (b) with PRM misaligned and ITMs aligned

 

The raw data is shown below.  In each of the above conditions we measured in both the vertical (v) and horizontal (h) directions.   The measurements in the vertical direction were better than the ones in the horizontal direction because the optics had a horizontal oscillation which gave larger errors in measurement.

rawdataplot.png

 

Looking at the general trend of these lines it is clear that modes are not matched since the beam reflected by the PRM has a different divergence than that reflected from ITMs.  The beam is also astigmatic as the vertical and horizontal directions have different divergences.

I could find beam parameters only for the Blue line above (Profile in the vertical direction while PRM was aligned).   The fit is quite sensitive to the data points close to the waist, so we need to make better (lower St.Dev.) measurements near the AP table closer to the beam waist.  The intensity with only one ITM aligned is too low and also contributes to the errors.   The beam size is close to 6mm in the horizontal direction, this coupled with yaw oscillations give large errors in this measurement.

Here is the only reliable fit that could be obtained, which is for the prompt reflection from the PRM in the vertical direction

PRM-to-REFL-profile.pdf

The fit function I used is  Beam Dia = Waist { Sqrt [ 1+  ((z + z0)/zr)^2). The fit parameters we get for this data are

z0 = 7.7 m 

Waist = 2.4 mm

zr = 6.9 m

 

Will make another attempt later today...

 

 

 

 

  6443   Mon Mar 26 12:50:24 2012 ZachUpdateelogrestarted with script

On the plus side, it was the first time I've had to do it in a while..

  6444   Mon Mar 26 15:15:16 2012 kiwamuUpdateIOOexpected beam profile of PRM reflection

I have estimated how the mode profile of the PRM reflection should be, as shown in the plot blow.

A conclusion here is :

   we should be able to constrain the PRM curvature situation if measurements are precise and accurate enough with a level of less than ~ 100 um

 

In the calculation two cases are considered :

      (1) PRM has the correct curvature of  +122 m. This is shown as solid curves in the plot.

      (2) PRM has a wrong curvature of - 122 m (mirror is flipped) This is shown as dashed curves in the plot. 

expected_edit.png

The plot above shows beam radii of the PRM reflections for vertical and horizontal profiles in each case.
The x-axis is distance from PRM in meter and the y-axis is the beam radii in mm.
As for the initial beam parameter, I used the measured values (see the wiki), which are that of after the beam exits from the mode matching telescope and before it goes to PRM.
 
(1) If PRM has the correct curvature, the reflection after it passes MMT1 will have ~ 1.6 mm beam radii.
This is intuitively correct because the beam profiles should match to that of the MC exiting beam  (see the wiki), which has waist size of 1.5 - 1.6 mm if everything is perfect.
(2) When PRM is flipped, the beam starts converging at the beginning as PRM act as a convex mirror, resulting in smaller beam sizes after it comes out from the telescope.
Roughly speaking the waist sizes will be different by ~ 5 mm between those two cases, so our measurement should be more precise and accurate than this number.

Note:

 I have omitted the effect from the PRM thickness. Therefore PRM is dealt as just a curved reflector with RoC of +/- 122 m in the calculation.

 

  6445   Mon Mar 26 16:25:44 2012 kiwamuUpdateIOOexpected v.s. measured beam profile of PRM reflection

[Suresh / Kiwamu]

 We did the 2nd round of the PRM reflection mode scan on Friday.

It seems that the PRM curvature maybe correct if we look at the vertical mode, however but the horizontal mode doesn't seem to agree with any of the expected lines.

In order to increase the reliability of the measurement, we need to confirm the beam profile of the incident beam by looking at the IP-POS beam.

Right now Suresh and Keiko are mode-scanning the IP-POS beam.

 

 


The plot below shows both the expected beam profiles (see the detail in #6444) and the actual data. 

PRMreflection.png

This plot is the same as one shown in the previous entry (#6444) with newly added actual data.
The errorbar in each data point is the standard deviation obtained by 100 times of averaging.
In this plot I made the error bars 10 times bigger in order to let them visible in the plot, so the actual deviation is much lesser than they appear.
 

(Discussion)

 The vertical profile (shown in red) seems to be close to the curve for the correct PRM case.
However the horizontal profile has a bigger waist size of about  2 mm.
While measuring the waist size Suresh and I have noticed that the rotational angle of the scan head affects the measurement by 10% or so.
Of course in each data point we tried making the incident beam normal to the scan head by rotating the scan head.
But this 10% is not big enough to explain the discrepancy in the horizontal mode.
There are some possible scenario which can distort the beam shape in the horizontal direction:
  • Clipping at some optics. (Since the beam shape looked very Gaussian, the amount of the clipping could be very slight ?)
  • Astigmatism at some optics. (Possibly in the telescope ?)

(Some distances)

DSC_4001_small.jpg
 

(Some notes)

We did the following things prior to the measurement.

  • Put a boost filter in the PRM_OLYAW to suppress the beam jitter below 1 Hz.
  • Checked the MC WFS servo loop although it looked healthy.

Quote from #6444

I have estimated how the mode profile of the PRM reflection should be, as shown in the plot blow.

A conclusion here is :

   we should be able to constrain the PRM curvature situation if measurements are precise and accurate enough with a level of less than ~ 100 um 

 

  6446   Mon Mar 26 18:04:43 2012 kiwamuUpdateIOOmode scan at the REFL port

For those who are interested in the actual data, I attache the actual data in zip file together with a python plot code.

The distance was set such that the 1st steering mirror (the one at the very left in the previous cartoon diagram #6445 ) in the REFL path is positioned at zero.


mode_profile.png

 

- - - Fitting results (chi-square fitting done by gnuplot):

All values are in unit of meter

# PRM (v) (Last tree points are excluded as the beam were clipped at the aperture of the beam scan)

w0          = 0.0015114        +/- 2.192e-05    (1.451%)
z0           = -4.46073         +/- 0.05605      (1.256%)

# PRM (h)

w0           = 0.00212796       +/- 1.287e-05    (0.6049%)
z0           = -2.53079         +/- 0.1688       (6.668%)

# ITM (v) (Last two points are excluded as the beam were clipped at the aperture of the beam scan)

w0           = 0.00190696       +/- 4.964e-05    (2.603%)
z0           = -8.09908         +/- 0.1525       (1.882%)

# ITM (h)

w0           = 0.0032539        +/- 4.602e-05    (1.414%)
z0           = -1.89484         +/- 1.524        (80.42%)

 

Attachment 2: REFLmodescan.zip
  6447   Mon Mar 26 23:47:54 2012 SureshUpdateIOOBeam Profile measurement: IPPOS beam

[Keiko, Suresh]

   Keiko and I measured the IPPOS beam profile.  The fit parameters  are :

  Vertical Horizontal
Waist (mm) 2.77 2.48
Rayleigh length (m) 23.5m 15.87
Waist location (m) 0.81 m 1.85

BeamProfile_IPPOS.png

 

The data files are attached. 

Attachment 2: BeamProfile_IPPOS.pdf
BeamProfile_IPPOS.pdf
Attachment 3: BeamProfileData_IPPOS.xlsx
  6448   Tue Mar 27 02:05:40 2012 SureshUpdateIOOBeam Profile measurement: IPPOS beam

Quote:

[Keiko, Suresh]

  Vertical Horizontal
Waist (mm) 2.77 2.48
Rayleigh length (m) 23.5m 15.87
Waist location (m) 0.81 m 1.85

BeamProfile_IPPOS.png

 

 

If we assume the nominal wavelength of the IR light to be 1064nm and constrain the Rayleigh length to be zr = (pi w0^2)/lambda we obtain the following fit parameters: (these are compared with the beam profile measurements of June/18/2010 available in the wiki )

  Vertical

Vertical 

06/18/2010

Horizontal

Horizontal

06/18/2010

waist (radius) (mm) 2.77 2.81 2.47 2.91
Rayleigh length (m) (computed) 22.62   18.14  
Waist location w.r.t. to MM2 * 3.37 5.36 4.15 1.96

* I updated the waist waist location coz because I missed-out the distance in air from the vacuum port to the optic on the IPPOS table.

 

BeamProfile_IPPOS.png

 

 

Attachment 1: BeamProfile_IPPOS.png
BeamProfile_IPPOS.png
  6449   Tue Mar 27 02:18:31 2012 keikoUpdateIOOBeam Profile measurement: IPPOS beam

Keiko, Rana, Suresh

Related to the beam profile of IPPOS today, we tried to measure the beam size at the ETMY point in order to estimate the input beam mode. We measured the beam size hitting at the suspension frame by a camera image, with two situations to see two "z" for beam profile.

(1) Input beam is slightly misaligned and the injected beam hits the end mirror frame. Assuming z=0 at the input mirror, this should be z=40m.

(2) Input beam hits the centre of the end mirror, and ITMY is slightly misaligned and the beam hits the end mirror frame after the one-round trip. Assuming z=0 at the ITM, this position should be z=120m.

text9149.png

The injected beam at the end point and the one round trip ligt at the end point should be the same size, if the input mode matches to the cavity mode. You can see if your injected light is good for the cavity or not. We compared and assumed the above two beam sizes by looking at the photo of the beam spot.

(1) first_cap.png (2) second_cap.png

 Assuming the zoom factor difference by the part below the beam (shown with allow in the photos. Arbitrary unit.), the beam in (2) is smaller than expected (roughly 40%?).

However this is a very rough estimation of the beam sizes! It is difficult to assume the beam size shown on the photos! It looks smaller only because the power of (2) is smaller than of (1). I don't think we can say anything from this rough estimation. One may be able to estimate better with CCD camera instead of this normal camera. 

 

Attachment 1: text9149.png
text9149.png
  6450   Tue Mar 27 02:46:28 2012 kiwamuUpdateIOOREFL beam available

The dump and some temporary mirrors were removed and now the REFL beam is available again.

I locked PRMI with REFL signals, it locked as usual.

Quote from #6440

Currently the REFL beam is bypassed by additional mirrors and blocked by a razor blade dump.

  6451   Tue Mar 27 11:54:18 2012 JenneUpdateIOOBeam Profile measurement: IPPOS beam

I'd like to know what we expect for these numbers.  I've added to Kiwamu's drawing some more distances, so I can calculate the expected beam size at the IPPOS port.

IPPOS_distances.pdf

 

Quote:

Quote:
 

  Vertical

Vertical 

06/18/2010

Horizontal

Horizontal

06/18/2010

waist (radius) (mm) 2.77 2.81 2.47 2.91
Rayleigh length (m) (computed) 22.62   18.14  
Waist location w.r.t. to MM2 * 3.37 5.36 4.15 1.96

 

  6452   Tue Mar 27 16:06:59 2012 keikoUpdateIOOBeam Profile measurement: IPPOS beam

I changed the ETMY CCD camera angle so that we can see the suspension frame in order to repeat the same thing as yesterday. The ETMY camera is not looking at the beam or mirror right now.

  6453   Tue Mar 27 17:19:02 2012 KojiUpdateIOOBeam Profile measurement: IPPOS beam

The mode looks quite terrible in the plot, but in reality it is an illusion of the plot.

The actual TEM00 content in this beam is ~99.7%


mode_coupling.pdf

Based on the above document, you can calculate power coupling between two elliptic gaussian modes.

Give the parameters of one beam as

zRy1 = pi*(2.77e-3)^2 / 1064e-9
z1y = 3.37
zRx1 = pi*(2.47e-3)^2 / 1064e-9
z1x = 4.15

Then, assume a round beam for the other.

zRx2 = zRy2 = zR
z2x = z2y = z0

Then run the optimizer to find the maximum for the quantity |C|^2

|C|^2 max: 0.9967
zR = 20.19
z0 = 3.804

  6454   Tue Mar 27 17:38:03 2012 JenneUpdateIOOAnother possibility / thought

I'm meditating over the mode matching from the mode cleaner to the ITMs, and I had another thought:

Have we changed the pointing of the MC significantly enough that we are no longer on the center of the MMT mirrors?  To be this significant, we would probably also have had to scoot the Faraday a bit too, since it's skinny like a straw.  It looks like our measurements of the input beam have been the following:

MC waist, 21 May 2010

After MMT2, 18 June 2010 (a few days before this, we flipped the MMT2 mount to 'perfect' the mode matching up to 99.3%, so I don't think the MMT has moved since then.)

After MMT2, 26 March 2012

There's a big o' ~2 year gap between our measurements, and we've been in and out of the vacuum a few times since then.  I'll flip through the elog, but does anyone have any memory of us moving the Faraday after June 2010?  When was the last time we made sure that we were at least close to the center of the MMT mirrors? 

  6455   Tue Mar 27 17:52:08 2012 KojiUpdateIOOAnother possibility / thought

It is quite likely that we touched the Faraday in Nov 2010.

In this entry http://nodus.ligo.caltech.edu:8080/40m/3874 I wrote that I removed the MCT optics in the chamber.
This is the pickoff between the IMC and the Faraday. This causes the beam shift. Therefore, the Faraday had
to be moved.

There were intensive in-chamber activities from Nov to Dec 2010. I am sure that almost everytime we went into
the chamber, we checked the spot position on the MMT mirrors as well as the TT and PZT mirrors.

Does the miscentering of the spots on the MMT mirrors cause the mode matching significantly changed?

  6456   Tue Mar 27 18:03:46 2012 JenneUpdateIOOBeam Profile measurement: IPPOS beam

This is wrong!  See following elog for corrected plot (and explanation)

I'm not done meditating on what's going on, but here's what I've got right now:

BeamProp_usingMeas2012MMTnumbersLowRes.png

This is a beam profile, using the distances from the combined Kiwamu / Jenne sketch earlier today. 

0 meters along the horizontal axis is meters from the Mode Cleaner waist. (Yes, I was bad and forgot to label it.  Get over it.)

The pink and green dots to the left of the plot are the MC fitted waist measurements that we made in May 2010.

The pink and green dots in the ~center of the plot are the fitted waist measurements that Suresh and Keiko took yesterday, of the IPPOS path, so after the MMT.

The black dot is where we would like our non-astigmatic beam to be.  This is the calculated waist size of the cavity mode, using the new ~37.76m distance, after we moved the ETMs to their current positions.  The black dot indicates 3.036 mm at the ITM (averaged between the BS-ITMX and BS-ITMY distances).

The moral of the story that I'm getting from this plot:  something funny is going on.

The distances Kiwamu quoted on the sketch are very close to the ones that I used for designing and checking the MMT in the first place, which were based off of measurements using rulers etc to measure distances.  Steve said he looked at photos today, and agreed that Kiwamu's numbers looked reasonable also.

Something that we haven't done lately is measure the position of each optic from every other optic, along the beam path.  I propose we come up with a clever way to put a target on top of / next to mirrors, and then a way to hold the laser distance measure-er at an optic, so that we can go thorough systematically and measure the actual path that our beam is seeing.  Maybe this is too much work, and not worth it, but it would make me happier.  In my head, these 'fixtures' are just small pieces of cleaned aluminum, one that can sit on a mirror mount, and one which we can use to align the laser ruler to approximately the front of an optic.  Nothing fancy.

  6457   Tue Mar 27 21:20:32 2012 JenneUpdateIOOBeam Profile measurement: IPPOS beam

Quote:

The moral of the story that I'm getting from this plot:  something funny is going on.

 Yup, something funny was going on.  Nic's MM code that I used, "a la mode", calls for the focal length of the optics, whereas the code that I wrote and used for ages called for the radius of curvature.  f = R/2.  Fixing that factor of 2 I get something more like:

BeamProp_usingMeas2012MMTnumbersLowRes.png

This is obviously much better, in that the beam profile goes through (within some error) both of the measured sets of points - the MC waist measured in May 2010, and after the MMT measured yesterday.

So, what does it all mean?  That I'm not sure of yet.

  6459   Tue Mar 27 23:37:35 2012 SureshUpdateIOOBeam Profile measurement: IPPOS beam

Quote:

Quote:

The moral of the story that I'm getting from this plot:  something funny is going on.

 Yup, something funny was going on.  Nic's MM code that I used, "a la mode", calls for the focal length of the optics, whereas the code that I wrote and used for ages called for the radius of curvature.  f = R/2.  Fixing that factor of 2 I get something more like:

.....

So, what does it all mean?  That I'm not sure of yet.

 In an attempt to estimate the errors on the fit parameters I upgraded my Mathematica code to use the function 'NonlinearModelFit', which allows us to define weights and also reports the errors on the fit parameters.   The plots have been upgraded to show the error bars and residuals.

Beam-Profile_IPPOS_wError.png

 

The parameters determined are given below and compared to the earlier measurements of 06/18/2010

Vertical Profile:

Parameter Estimate Standard Error 95% Confidence interval 06/18/2010 measurement
Waist (mm) 2.768 0.005 2.757 -- 2.779 2.81
Waist location from MMT2 (m) 5.85 0.12 5.625-- 6.093 5.36

 

 

Horizontal Profile:

Parameter Estimate Standard Error 95% Confidence Interval 06/18.2010 measurement
Waist (mm) 2.476 0.009 2.455 -- 2.496 2.91
Waist Location from MMT2 (m) 4.935 0.145 4.645 -- 5.225 1.96

 

There is a significant change in the beam waist location (as compared to previous report) because I corrected a mistake in the sign convention that I was using in measuring the distances to the waist from the zero-reference.
 

  6460   Wed Mar 28 01:17:29 2012 JenneUpdateIOOBeam Profile measurement: IPPOS beam

As I was a little dissatisfied with the inaccuracy in the distance numbers in Kiwamu's sketch, I went back to the 18 Dec 2010 table layout drawing for more accurate numbers.  These are now included in this round of plots.

Also, I include astigmatism due to the incident angles on MMT1 (~3.5 deg) and MMT2 (~1 deg).

First plot, IPPOS path, using the recent (fixed) measurements from Suresh to fix the beam width.  Note that the old 2010 measurements of the MC waist are consistent with this measurement.

Second plot, Main IFO path all the way to the ITM (average) position, using the 2010 MC waist measurements to fix the beam width.

Third plot, Main IFO path all the way to the ITM position, but with PRM flipped (negative RoC), using the 2010 MC measurement to fix beam width.

With the PRM correctly oriented (2nd plot), I get beam waists of (x = 2.529 mm, y = 2.646 mm), which corresponds to a mode matching to the arm cavity of (eta = 97.43%, PRM correct).

With the PRM flipped (3rd plot), I get beam waists of (x = 3.176 mm, y = 3.3 mm), which corresponds to a mode matching to the arm cavity of (eta = 99.55%, PRM flipped).

 

First plot:

BetterDistances_IPPOSLowRes.png

Second plot (this is how the MMT was designed to be, before the ETMs were moved, which made the ideal waist larger):

BetterDistances_MainIFO_PRMnormalLowRes.png

Third plot:

BetterDistances_MainIFO_PRMflippedLowRes.png

For both the 2nd and 3rd plot, we can't look at the post-MMT waist measurements, since that distance on the plots is after the PRM, which is a curved optic.  So the fact that the post-MMT measurements match the correct-PRM plot better than the flipped-PRM plot can't be taken to be meaningful.

Moral of the story:  I'm not sure how to interpret any of this to tell us if the PRM is flipped or not, since the measurements are all of the beam profile before the beam sees the PRM.  We'd have to measure the profile after the PRM somehow in order to get that information.  We have okay but not great mode matching to the arm if the PRM is correct, but I don't know that we readjusted the MMT after we moved the ETMs.  I don't remember recalculating any optimal telescope lengths after the arm length change.  If we need better mode matching, I can do that calculation, although given how much space we don't have, it would be hard in practice to move the MMT mirrors by much at all.

  6461   Wed Mar 28 18:26:47 2012 JenneUpdateIOOBeam Profile measurement: IPPOS beam

More calculations....

Game Plan:  Using MC waist measured beam as the starting point of beam propagation, move MMT mirrors around until the beam profile fits with the IPPOS measurements from Monday.

Plot 1:  Allow MMT mirrors to move as much as they want to.  Note that the Y-beam goes through the IPPOS measured point. (This implies we put the MMT in the wrong place by ~half a meter.  Unlikely)

Plot 2: Using MMT locations from plot 1, what does the beam look like at the ITMs? 

Plot 3:  Allow MMT mirrors to move as much as 2cm.  Note that the Y-beam doesn't quite go through IPPOS measured point.

Plot 4: Using MMT locations from plot 3, what does the beam look like at the ITMs?

For all of the above, I was optimizing the propagation of the Y-beam profile.  Since the X-beam profile measurement is so different, if I want to optimize to X and let the MMT mirrors move as much as they want, MMT1 ends up inside the MC.  Unlikely.  So I'm just looking at Y for now, and maybe Suresh or someone needs to rethink the error bars on their measurements or just remeasure.

Plot 1:

OptimizedMMT_halfMeterChange_IPPOSviewLowRes.png

Plot 2:

OptimizedMMT_halfMeterChange_IFOviewLowRes.png

Plot 3:

OptimizedMMT_pt2cMeterChange_IPPOSviewLowRes.png

Plot 4:

OptimizedMMT_pt2cMeterChange_IFOviewLowRes.png

  6462   Wed Mar 28 20:54:02 2012 JenneUpdateIOOBeam Profile measurement: IPPOS beam

I fitzed by hand with the numbers for the incident angles on MMT1 and MMT2, and then let the code optimize the position of MMT1 and MMT2.

Here I have set the incident angle for MMT1 = 25deg, and MMT2 = 12deg (should be 3.5deg, 1deg by design).  The length of the telescope doesn't want to change by more than 7mm, but the position of the telescope wants to change by 1.3meters.  Is it possible that the distances on the Monday IPPOS measurements aren't actually correct?

MMT_moved_by_1pt3meters_MMT1th_25_MMT2th_12_LowRes.png

Attachment 1: MMT_moved_by_1pt3meters_MMT1th_25_MMT2th_12_LowRes.png
MMT_moved_by_1pt3meters_MMT1th_25_MMT2th_12_LowRes.png
  6464   Thu Mar 29 11:29:27 2012 keikoUpdateLSCPOP22/POP110 amplifires

Yesterday I and Kiwamu connected two amplifiers (mini-circuit, ZFL-1000LNB+) for POP22/110. Dataviewer can see some signals. I'll test the signal levels and freq components before the rack just in case. [Kiwamu, Keiko]

  6466   Thu Mar 29 18:42:11 2012 keikoUpdateLSCPOP22/POP110 amplifires

Adding two amplifiers on POP22/110, I checked the signals going to the dmod board of 22 and 110.

The signal flows: Photodetector of POP --> Amp1 --> Amp2 --> RF splotter --> bandpass filter for 22MHz / 110MHz --> 22MHz / 110MHz demod board.

 

 

 

 Here is the picture of RF spectrum just after the bandpass filter of 22MHz going to the 22MHz demod board. The signal peak at 22MHz is about -40dBm. There is a structure slightly lower than 22MHz.

P3290004.JPG

The below is the RF spectrum for 110MHz branch. The peak at 110MHz is about -15dBm. The peak on the left of 110MHz is 66MHz peak.

P3290005.JPG

 

Quote:

Yesterday I and Kiwamu connected two amplifiers (mini-circuit, ZFL-1000LNB+) for POP22/110. Dataviewer can see some signals. I'll test the signal levels and freq components before the rack just in case. [Kiwamu, Keiko]

 

  6469   Fri Mar 30 08:58:37 2012 steveUpdateSAFETYSafety glasses checked

Bob cleaned all safety glasses in 10 % Liquinox soap in water solution first.  The  transmittance of glasses were checked at 100 mW 1064 nm S polarization, beam diameter  0.5 mm at 0-20 deg incident.

Coherent FieldMate power meter measured T= < 1 mW of all glasses.

Attachment 1: safety_glasses_checked.jpg
safety_glasses_checked.jpg
  6470   Fri Mar 30 09:37:13 2012 steveUpdateSAFETYsafety audit 2012 CORRECTIONS

Quote:

Correction list by visiting safety committee, Haick Issaian is not shown:

1,  update laser, crane operator list and post it

2,  check fire extinguishers monthly, date and initials must be on the tags

3,  move drinking water tower so it does not block fire extinguisher

4,  post updated crane doc at cranes

5,  post present phone lists at IFO room phones

6,  emergency laser shutoff at the south end must be mounted without C-clamp

7,  use heavy cable tie to insure position of  mag-fan on cabinet top

 

Additional to do list:

a,  safety glasses to be cleaned

b, let the electrical shop fix Rack-AC power to optical tables at the ends

c, measure transmission of  laser safety glasses

d, update  IFO outside door info signs

e, update laser inventory and post it

f,  schedule annual crane inspection and renew maintenance contract

g, PSL enclosure  inner shelf needs a good clean up so it is earthquake safe

 

 Completed with the exception of d and g

  6471   Fri Mar 30 10:20:51 2012 kiwamuUpdateLSClocking last night

I was trying to make the DRMI lock more robust.

Increasing the gains of the oplev on SRM helped a lot, but the lock is still not solid enough for measurements.

According to some line injection tests, the SRCL and MICH signals show up in AS55Q with almost the same amplitudes.

I tried to diagonalize the input matrix (particularly MICH-SRCL in AS55) based on the result of the line injection tests, but I ran out the time.

Work continues.

  6472   Fri Mar 30 10:24:14 2012 steveUpdateSUSnew OSEM locking plunger

Quote:

Quote:

Our existing 300 series SS plungers from McMastercar #8476A43 are silver plated as Atm2 shows.

Problems:  1, they become magnetized after years being close to the magnets

                     2, they oxidize by time so it is hard to turn them

                    

I looked around to replace them.

Titanium body, nose and beryllium copper spring. None magnetic for UHV enviorment.

Can be made in 7 weeks at an UNREASONABLE $169.00 ea at quantity of 50

 In order to get a better price from Vlier's Tom Chen I changed Ti body back to SS304L-siver plated and music wire spring. The price is still ~$120 ea. at quantity 50

I will talk to Mike G about modifying the  McMaster plunger with a hex nut.

 Conclusion: There is no need to silver plate the SS plunger at the torque level we hold the OSEMs

Mike Gerfen will be done with 50 pieces by April 4 and he will give them to Bob for cleaning. They should be cleaned and baked in a jig so the springs would be compressed for better venting.

Attachment 1: new_OSEM_locker_screw.jpg
new_OSEM_locker_screw.jpg
Attachment 2: IMG_0109.JPG
IMG_0109.JPG
Attachment 3: IMG_0111.JPG
IMG_0111.JPG
  6473   Fri Mar 30 17:37:09 2012 steveUpdateGeneralcutting green welding glass for beam dumps

Quote:

Schott, green welding glass, shade 14, 3 mm thick  was measured in the beam path of 1.2W, S polarization of 1064nm at ~1 mm diameter size as MC reflected path.

Absorption 95%, R 5% at incident angle 25-50 degrees. It looks like the perfect material for beam trap.

 

 The CIT Chemistry Glass Shop cuts turned out to be sloppy using diamond disc blade  cutter.

East coast Precision Glass & Optics  offered scribed cut and polished side. Their quote price was high and time consuming.

The GLASS HOUSE  shop in Pasadena 626 / 796-9151 on Walnut did a good and cheap job. Oscar the cutter will finish the rest of the cutting by next Friday, April 6

He used scribed cutting technic and his 1" x  0.5"  pieces are good. Bob will pick up them up.

Attachment 1: IMG_0096.JPG
IMG_0096.JPG
  6474   Sat Mar 31 08:01:07 2012 kiwamuUpdateLSCDRMI measurement

I have measured the sensing matrix of the DRMI although the lock still doesn't stay for a long time.

As for the noise budget, it looks very tough as there are more glitches than that in the PRMI.

In this weekend I will take some more trials in the DRMI lock until I am satisfied.

  6475   Mon Apr 2 18:24:34 2012 keikoUpdateLSCRAM simulation for Full ifo

 I extended my RAM script from DRMI (3DoF) to the full IFO (5DoF).

Again, it calculates the operation point offsets for each DoF from the opt model with RAM. Then the position offsets are added to the model, and calculates the LSC matrix. RAM level is assumed as 0.1% of the PM modulation level, as usual, and lossless for a simple model.

 

 

Original matrix without RAM:

REFL f1 : 1.000000    0.000000    0.000008    -0.000005    0.000003 

  AS f2 : 0.000001    1.000000    0.000005    -0.003523    -0.000001 

 POP f1 : -3956.958708    -0.000183    1.000000    0.019064    0.000055 

 POP f2 : -32.766392    -0.154433    -0.072624    1.000000    0.024289 

 POP f2 : 922.415913    -0.006625    1.488912    0.042962    1.000000 

 

(MICH and SRCL uses the same sensor, with optimised demodulation phase for each DoF.) 

Operation position offsets are:

PRCL   -3.9125e-11 m

SRCL    9.1250e-12 m

CARM  5.0000e-15 m  

and no position offsets for DARM and MICH (because they are differential sensor and not affected by RAM offsets).

 

Resulting matrix with RAM + RAM offsets, normalised by the original matrix:  

REFL f1 : 0.001663    -0.000000    0.003519    0.000005    -0.000003 

  AS f2 : 0.000004    0.514424    0.000004    -0.001676    -0.000001 

 POP f1 : 7.140984    -0.001205    15.051807    0.019254    0.000417 

 POP f2 : 0.029112    -0.319792    0.042583    1.000460    0.024298 

 POP f2 : -0.310318    -0.014385    -1.761519    0.043005    0.999819 

 

As you can see in the second matrix, the CARM and DARM rows are completely destroyed by the RAM offsets! The signals are half reduced in the DARM case, so the mixture between DARM and MICH are about 50% degraded.

 I also would like to extend this script to use the DC readout, but don't know how to calculate the postion offset for AS_DC because the error signal is not zero-crossing for AS_DC anymore. Do you have any suggestions for me?

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

  6476   Mon Apr 2 18:58:32 2012 JenneUpdateIOOBeam Profile measurement: IPPOS beam

Suresh noted that I never wrote down the waist positions of the beam propagated through the MMT (based on where we think it is from ruler-based measurements).  This uses the MC waist measured beam as the starting point, and just propagates through the MMT, out to the IPPOS port.

Yq:

  Properties:
                    q: 2.2488 +23.8777i
               lambda: 1.0640e-06
            waistSize: 0.0028  (at z = 13.2676 meters)
               waistZ: 2.2488    (relative to z = 13.2676 meters)
      divergenceAngle: 1.1910e-04
    radiusOfCurvature: 255.7849
            beamWidth: 0.0029
        rayleighRange: 23.8777

So, to sum up, the vertical beam waist is 2.8438 mm at 11.0188 meters from the MC waist.

 

Xq:

  Properties:
                    q: 5.1953 +24.7342i
               lambda: 1.0640e-06
            waistSize: 0.0029   (at z = 13.2676 meters)
               waistZ: 5.1953    (relative to z = 13.2676 meters)
      divergenceAngle: 1.1702e-04
    radiusOfCurvature: 122.9525
            beamWidth: 0.0030
        rayleighRange: 24.7342

So, to sum up, the horizontal beam waist is 2.8943 mm at 8.0723 meters from the MC waist.

In pictorial form,

NonOptimized_propagatedWaistPlotted_LowRes.png

  6477   Mon Apr 2 23:06:38 2012 SureshUpdateIOOBeam Profile measurement: IPPOS beam: Mystery deepens

Quote:

I fitzed by hand with the numbers for the incident angles on MMT1 and MMT2, and then let the code optimize the position of MMT1 and MMT2.

Here I have set the incident angle for MMT1 = 25deg, and MMT2 = 12deg (should be 3.5deg, 1deg by design).  The length of the telescope doesn't want to change by more than 7mm, but the position of the telescope wants to change by 1.3meters.     Is it possible that the distances on the Monday IPPOS measurements aren't actually correct?

 

I am trying to track down possible errors in our measurements. 

So as a first step I am recomputing the IPPOS waist location with respect to the MC waist, using the same optical layout diagram as the one used by Jenne in her calculations.  Pic of Jenne's lab notebook showing location of optics is attached.  

IPPOS: measurement elog 6459: Vertical Std.Error Horizontal Std.Error
Waist  2.768 mm 5 microns 2.476 mm 10 microns
Waist location from MC waist  12.411 m  17 mm  9.572 m 54 mm

Std Dev of residuals from fit function

  37 microns   54 microns

 Let us compare it with the old measurement of the IPPOS beam from June/18/2010.

IPPOS: measurement June 18th 2010 Vertical Std.Error Horizontal Std.Error
Waist  2. 812mm 8 microns 2.909 mm 20 microns
Waist location from MC waist  9.265 m  224 mm  5.869 m 415 mm

Std Dev of residuals from fit function

  ~ 25 microns   ~25 microns

Note that there is a discrepancy of about 3.2 m in the waist location for the vertical profile and about 3.5 m for the horizontal profile between these two measurements. 

Let us compare these measurements with what is expected from calculations.  Jenne uses the known parameters of MC waist and the locations of the MMT optics to compute the parameters for the IPPOS beam:

IPPOS: Jenne's Calculations elog 6476:

Vertical Std.Error Horizontal Std.Error
Waist  2.844 mm   2.894 mm  
Waist location from MC waist  11.019 m   8.072 m  

 As the 2010 measurements are reported wrt to MMT2 and calculations are wrt MCwaist, I have used the distance between the MCwaist to MMT2 = 3.910 m to shift the reference from MMT2 to MC waist. Refer to the attached diagram from Jenne's notes for this MMT2 <--> MC waist distance.

There is a discrepancy of 1.5 meters between the calculations and recently measured waist location.  The discrepancy with the 18Jun2010 measurement is much larger, about 3 meters in both v and h.

Are such variations to be expected between two successive measurements?  I looked at another case where we have two measurements of a beam to see what to expect.

I looked at the REFL (Reflection from PRM) case, where we repeated a measurement, to see how much variation could happen in w0 and zr, between repeated measurements. This was a particularly bad case as our first attempt had problems due to OL servo loop oscillations in the PRM suspension damping.  We fixed that later and measurement 2 has smaller residuals. And I think we are doing okay in IPPOS case as seen by the reduced scatter of the residuals.

 These are the fits from the REFL beam measurement 1

REFL: Reflection from PRM: measurement 1 Vertical Error Horizontal Error
Waist 1.662 mm 4 microns 2.185 mm 4 microns
Waist location from MMT2 after reflection at PRM 1.781 m 17 mm 4.620 m 53 mm
Std.Dev. of residuals from fit function   61 microns   98 microns

 I have also recomputed the fits to the data from REFL beam measurement 2.  They match the earlier fits reported by kiwamu in his elog 6446

REFL: Reflection from PRM: measurement 2 Vertical Error Horizontal Error
Waist 1.511 mm 3 microns 2.128 mm 3 microns
Waist location from MMT2 after reflection at PRM 1.281 m 9 mm 3.211 m 37 mm
Std. Dev of residuals from fit function   58 microns   61 microns

 

Note that between these two measurements the beam waist location has shifted by 0.5 m for the vertical and about 1.3 m for the horizontal cases.  So variations of 1.5 m in the waist locations are possible if we are not careful.  But this is a particularly extreme example, I think we are doing better now and the measurement is unlikely to change significantly if we repeat it.

Some notes:

Fits for IPPOS and both REFL measurements 1 and 2 are attached. 

The zero reference for the z axis of the IPPOS beam plot is at a distance of 6.719 m from MC waist for a beam propagating towards the IPPOS QPD.

The zero reference for the z axis of the REFL beam plots is at a distance of 5.741 m from the MMT2 in the direction of a beam reflected by PRM and propagating towards the REFL port.

 

Attachment 1: 40mOpticsLocations.pdf
40mOpticsLocations.pdf
Attachment 2: Beam-Profile_IPPOS_wError.pdf
Beam-Profile_IPPOS_wError.pdf
Attachment 3: Beam-Profile_PRM_1_wError.pdf
Beam-Profile_PRM_1_wError.pdf
  6478   Tue Apr 3 01:52:15 2012 ZachUpdateLSCRAM simulation for Full ifo

Quote:

 I also would like to extend this script to use the DC readout, but don't know how to calculate the postion offset for AS_DC because the error signal is not zero-crossing for AS_DC anymore. Do you have any suggestions for me?

 I don't think I understand the question. AS_DC should not have a zero crossing, correct?

  6479   Tue Apr 3 12:42:19 2012 Mike J.UpdateComputersHysteresis Model

Here's my first hysteresis model in Simulink. It's based on the equation y=Amplitude*sin(frequency*t+phase)+(hysteresis/frequency2) as a solution to y''+frequency2*y+hysteresis=0. All values in the model are variables that should be manipulated through the model workspace or external code.

Attachment 1: hysteresis1.mdl
Model {
  Name			  "hysteresis1"
  Version		  7.6
  MdlSubVersion		  0
  GraphicalInterface {
    NumRootInports	    0
    NumRootOutports	    0
    ParameterArgumentNames  ""
    ComputedModelVersion    "1.9"
    NumModelReferences	    0
... 734 more lines ...
  6480   Tue Apr 3 14:11:33 2012 keikoUpdateLSCRAM simulation for Full ifo

Quote:

Quote:

 I also would like to extend this script to use the DC readout, but don't know how to calculate the postion offset for AS_DC because the error signal is not zero-crossing for AS_DC anymore. Do you have any suggestions for me?

 I don't think I understand the question. AS_DC should not have a zero crossing, correct?

 That's right. I calculate the offset of the operation point (when you have RAM) from the zero-crossing point of the PDH signals. I don't know how to do that for AS_DC, because it doesn't cross zero anymore anytime.

  6481   Tue Apr 3 14:17:18 2012 keikoUpdateLSCRAM simulation for Full ifo

I add a flow-chart drawing what the scripts do and how the scripts calculate the LSC matrix.

flowchart.png

 

(1) First, you calculate the LSC matrix WITHOUT RAM or anything, just for a reference. This is the first matrix shown in the quoted post.

(2) The script calculates the LSC matrix with RAM. Also, the heterodyne signals for all 5 DoF are calculated. The signals have offsets due to the RAM effect. The operating position offsets are saved for the next round.

(3) The script calculates the LSC matrix again, with RAM plus the offset of the operation points. The matrix is shown in the last part of the quoted post.

 

Now I am going to check (A) LSC matrices (matrix 2, the second matrix of above chart) with different RAM levels (B) Are pos-offsets degrade the CARM and DARM so much (See, the quated result below), is that true?

Quote:

Original matrix without RAM:

REFL f1 : 1.000000    0.000000    0.000008    -0.000005    0.000003 

  AS f2 : 0.000001    1.000000    0.000005    -0.003523    -0.000001 

 POP f1 : -3956.958708    -0.000183    1.000000    0.019064    0.000055 

 POP f2 : -32.766392    -0.154433    -0.072624    1.000000    0.024289 

 POP f2 : 922.415913    -0.006625    1.488912    0.042962    1.000000 

 

(MICH and SRCL uses the same sensor, with optimised demodulation phase for each DoF.) 

Operation position offsets are:

PRCL   -3.9125e-11 m

SRCL    9.1250e-12 m

CARM  5.0000e-15 m  

and no position offsets for DARM and MICH (because they are differential sensor and not affected by RAM offsets).

 

Resulting matrix with RAM + RAM offsets, normalised by the original matrix:  

REFL f1 : 0.001663    -0.000000    0.003519    0.000005    -0.000003 

  AS f2 : 0.000004    0.514424    0.000004    -0.001676    -0.000001 

 POP f1 : 7.140984    -0.001205    15.051807    0.019254    0.000417 

 POP f2 : 0.029112    -0.319792    0.042583    1.000460    0.024298 

 POP f2 : -0.310318    -0.014385    -1.761519    0.043005    0.999819 

 

 

 

  6482   Tue Apr 3 15:50:58 2012 keikoUpdateLSCRAM simulation for Full ifo

Oops, Yesterday's results for DARM was wrong!

I got more convincing results now. 

 

> (B) Are pos-offsets degrade the CARM and DARM so much (See, the quoted result below), is that true? 

 

Here is the new results. It does change CARM a lot, but not DARM:
 
Matrix1 (normalised so that the diagonals are 1):
REFL f1 : 1.000000    0.000000    0.000008    -0.000005    0.000003 
  AS f2  : 0.000001    1.000000    0.000005    -0.003523    -0.000001 
 POP f1 : -3956.958708    -0.000183    1.000000    0.019064    0.000055 
 POP f2 : -32.766392    -0.154433    -0.072624    1.000000    0.024289 
 POP f2 : 922.415913    -0.006625    1.488912    0.042962    1.000000 
(=Matrix 2)
 
Position offsets:
only CARM, 4.6e-16 (this number changed because I increased the resolution of the calculation)
 
Matrix3 (normalised by matrix 1):
REFL f1 : 0.039780    -0.000000    0.003656    0.000005    -0.000003 
  AS f2  : 0.000008    1.000017    0.000005    -0.003499    -0.000001 
 POP f1 : 159.146819    -0.000138    15.605155    0.019393    0.000055 
 POP f2 : 1.277223    -0.154415    0.047344    1.000008    0.024289 
 POP f2 : -35.422498    -0.006633    -1.886454    0.042963    1.000000 

 

  • CARM got a small position offset which degrades CARM signal 2 orders of mag (still the biggest signal in the sensor, though).
  • DARM was not so bad, and probably the change of the DoF mixture is mostly not changed.
  • Matrices don't change only with 1e-4 RAM. It changes with position offsets.
  • I'll see how the matrix changes without position offsets but only with RAM effects, changing RAM levels.
  • Again, above is C1 configuration, 1e-4 RAM level of PM level.

 

 

Quote:

I add a flow-chart drawing what the scripts do and how the scripts calculate the LSC matrix.

flowchart.png

 

(1) First, you calculate the LSC matrix WITHOUT RAM or anything, just for a reference. This is the first matrix shown in the quoted post.

(2) The script calculates the LSC matrix with RAM. Also, the PDH signals for all 5 DoF are calculated. The PDH signals have offsets due to the RAM effect. The operating position offsets are saved for the next round.

(3) The script calculates the LSC matrix again, with RAM plus the offset of the operation points. The matrix is shown in the last part of the quoted post.

 

Now I am going to check (A) LSC matrices (matrix 2, the second matrix of above chart) with different RAM levels (B) Are pos-offsets degrade the CARM and DARM so much (See, the quated result below), is that true? 

 

  6483   Tue Apr 3 22:50:37 2012 keikoUpdateLSCRAM simulation for Full ifo

Koji and Jamie suggested me to include the coupling between DoFs when I calculate the last matrix. So far, I just add all the pos-offsets of 5 DoFs and re-calculate the matrix again. However, once I add one DoF pos-offset, it could already change the LSC matrix therefore different pos-offset to the other four DoF, we must iterate this process until we get the equilibrium pos-offsets for 5 DoFs.

I also noticed an error in the optical configuration file. AM mod levels were smaller than that supposed to be because of the hald power going through the AM-EOMs in the MZI paths. Also I have put PM-Mods in the MZT path which gives the smaller mod indexes. So, smaller mod levels were applied both for PM and AM. As PM-AM ratio is still kept in this, so the matrices were not very wrong, I assume. I'll modify that and post the results again.

  6485   Wed Apr 4 21:43:16 2012 Mike J.UpdateComputersBetter Hysteresis Model

A better hysteresis model based on the simple harmonic oscillator equation. Useless variables have been removed and output can now be saved to workspace for plotting. The model is at "/users/mjenson/matlab/SHO_hyst.mdl".

Attachment 1: SHO_hyst.png
SHO_hyst.png
  6486   Wed Apr 4 23:57:35 2012 keikoUpdateLSCRAM simulation for Full ifo

 I'm still wondering whether iteration version or simple version is closer approximation to the real situation. Sorry for few explanations here. I will try to present those on Friday.

 

Anyway, here is the results for both:

%*.*.*. Original matrix w/o RAM .*.*.*

REFL f1 : 1.000000        0.000000    -0.000003    -0.000005    0.000007 

  AS f2 : 0.000002        1.000000    0.000009    -0.003522    -0.000002 

 POP f1 : -3954.521443    -0.000965    1.000000    0.019081    -0.000152 

 POP f2 : -32.770726    -0.154433    -0.072594    1.000000    0.024284 

 POP f2 : 922.393978    -0.006608    1.488319    0.042948    1.000000 

 

*** Iteration *** 

%*.*.*. Resulting matrix w/ RAM .*.*.*

REFL f1 : 0.039125    -0.000000    0.003665       0.000005    -0.000007 

  AS f2 : 0.000010    1.000431    0.000009       -0.003500    -0.000002 

 POP f1 : 156.420221    -0.000246    15.586838    0.019406    -0.000154 

 POP f2 : 1.255806    -0.154275    0.047313       1.000008    0.024285 

 POP f2 : -34.814720    -0.006600    -1.884850    0.042950    1.000000 

Offsets converged to:

PRCL =  2.1e-15, MICH = 1.1e-17, SRCL = -3.8e-15, CARM = 2.2e-16, DARM = 0 

(POP CARMs became so much smaller compared with the other matrix below, because the offsets are added al of 5 DoFsl at once here.)

 

*** no iteration, offsets added for each DoF separately ***

REFL f1 : 0.020611        -0.000000    0.003600    0.000005    -0.000007 

AS f2   : 0.000002        1.000000    0.000009    -0.003522    -0.000002 

POP f1  : 1842.776419    -0.000198    21.533358    0.019404    -0.000132 

POP f2  : -32.700639    -0.153095    -0.072481    0.999995    0.024360 

 POP f2 : 922.393862    -0.006435    1.488298    0.042949    0.999982 

Added offsets:

PRCL =  7.5e-15, MICH = 6.25e-16, SRCL = -1.4e-14, CARM = 4.5e-16, DARM = 0

* So far, I used to add all the offsets at once. This time I add CARM and get the CARM row, add PRCL and get the PRCL row... and so on.

Quote:

Koji and Jamie suggested me to include the coupling between DoFs when I calculate the last matrix. So far, I just add all the pos-offsets of 5 DoFs and re-calculate the matrix again. However, once I add one DoF pos-offset, it could already change the LSC matrix therefore different pos-offset to the other four DoF, we must iterate this process until we get the equilibrium pos-offsets for 5 DoFs.

I also noticed an error in the optical configuration file. AM mod levels were smaller than that supposed to be because of the hald power going through the AM-EOMs in the MZI paths. Also I have put PM-Mods in the MZT path which gives the smaller mod indexes. So, smaller mod levels were applied both for PM and AM. As PM-AM ratio is still kept in this, so the matrices were not very wrong, I assume. I'll modify that and post the results again.

 

  6487   Thu Apr 5 01:07:08 2012 Mike J.UpdateComputersHysteresis Plots

Here are the hysteresis plots from the most recent model, which uses a modified harmonic oscillator equation y''=-(Frequency)2*y-Hysteresis.  The hysteresis constant seems to change both the amplitude and equilibrium point of the pendulums, which is akin to changing the length of a pendulum without changing the frequency. This does not make sense. Perhaps the hysteresis value should be moved to the "spring" constant for the pendula and not restricted to a position-biasing value.

SHO_hyst_plot.png