ID Date Authorup Type Category Subject Text Attachments
  13172   Tue Aug 8 17:44:11 2017 SteveUpdateVACunintended pump downPumpdown stopped for over night  at
~  1 Torr

The roughing line disconnected. 
 stopped_pumping.png 
  13175   Wed Aug 9 11:59:51 2017 SteveUpdateVACunintended pump down at vacuum normalpd80b has reached Vac Normal. IFO pressure
0.5 mTorr

We need our vauum channels back
 pdc.pngpd80b@vacnormal.png 
  13182   Thu Aug 10 09:31:57 2017 SteveUpdateSUSITMX sensor voltageThere must be some bad connection




 9daysITMX.pngvacGlitchITMX.png 
  13186   Thu Aug 10 15:45:34 2017 SteveUpdateVACaccidental vent to 17 TorrFinally we got the cold cathode gauge working.
IFO pressure 7e-6 Torr-it, vacuum normal
valve configuration with all 4 ion pump gate
 Mag_UPS.jpgUPSbatteryPack.jpgclosing_V1.jpg 
  13208   Tue Aug 15 08:22:16 2017 SteveUpdateGeneralprojector light bulb replacedBL-FS300C-PH-LE was
replaced after 2,904 hrs  It did
not explode this time.  The 4 monts
  
  13213   Wed Aug 16 14:57:01 2017 SteveUpdatePSLref Cavity heating blanket power supplyThe last entry I found relating to ref
cavity was
2011 Aug 19
 refCavPSheat.jpgrefCavTempSens.jpgrefTempSen_1X1.jpg 
  13223   Thu Aug 17 08:42:27 2017 SteveUpdatePSLPSL HEPAThe PSL HEPA was running noisy at 100V  
The bearing is wearing out. I turned it down
to 30V It is quiet there.
  
  13231   Mon Aug 21 09:08:54 2017 SteveUpdateSUStiny glitchesThey are synchronised tiny glitches. They
are not mechanical.
 
 glitches.png 
  13245   Wed Aug 23 10:11:46 2017 SteveUpdateVACvacuum valve specificationsThe V1
gate valve specs installed at 40m wiki
page. VAT model number 10846-UE44-0007 Our
  
  13260   Mon Aug 28 10:28:07 2017 SteveUpdateVACRGA scan 20 day after 17 TorrThe RGA was turned on 7 days ago. 
It's 46 C now.   The X-arm room
tem ~20 C
 20d_after_17_torr.png 
  13268   Tue Aug 29 15:35:19 2017 SteveUpdateVACvacuum pump specifications & manualsRP1 and RP3 roughing pump manual of
Leybold D30A oily rotory pump

Fore pump of
 1759489_995814a9-55c1-4fa1-b333-f446d272c1cc.jpg 
  13284   Fri Sep 1 08:25:08 2017 SteveUpdateSUSMC1 glitchingMC1, MC2 and MC3 damping turned off to
see glitching action at 9:57am

 MC1glitching.pngMC1kicks.png 
  13302   Fri Sep 8 07:54:04 2017 SteveUpdatePEMM8.1 eqNothing tripped. No obvious damage.  M8.1.png 
  13303   Fri Sep 8 10:22:30 2017 SteveUpdatePEMtemp sensor updateThe weight of SS can with copper liner 
is 12.2 kg

Is 1 Amp for the heating jacket
  
  13305   Mon Sep 11 09:47:53 2017 SteveUpdateGeneralWIMA caps refilledInstock WIMA caps refilled to a minimum
50 pieces each.
 WIMA.png 
  13308   Mon Sep 11 15:58:02 2017 SteveUpdateGeneralNPRO for repairThis NPRO has a tripping power output******

 
 NPRO_tripping.jpg 
  13309   Mon Sep 11 16:46:09 2017 SteveUpdatePEMearthquakes


 M5.4eq.png 
  13321   Tue Sep 19 11:05:26 2017 SteveUpdateVACRGA scan at day 334  

 
 RGAscan334d.png 
  13322   Tue Sep 19 14:00:41 2017 SteveUpdatePEMearthquakesNo sus tripped. Seimometers do not see
the 5.3M  ?

 7.1MX5.3J.png 
  13346   Fri Sep 29 11:16:52 2017 SteveUpdateALSY End table correctedThe first Faraday isolater rejected beam
path from the NPRO is fixed.

 
 ETMYf1.jpg 
  13366   Fri Oct 6 17:08:09 2017 SteveUpdateALSX End table beam traps correctedThere are no more double sided tape on
this table.

 
 c1.jpgc2.jpgc3.jpgc4.jpg 
  13371   Wed Oct 11 10:29:43 2017 SteveUpdateElectronicsSR560 noise levelGautam and Steve,

All 3 show the same noise level
~80 nV / rt Hz at 1 kHz as shown. Batteries
 sr560.jpgsr560noise.jpg 
  13377   Thu Oct 12 07:56:33 2017 SteveHowToCamerasETMX GigE side view at 50 deg of IR scattering Telescope front lens to wall distance
25 cm,  GigE camera lenght 6 cm and
cat6 cable 2cm
 Image__2017-10-11__15-29-52_15k400g.pngImage__2017-10-12__15-50-18wipedRefocud2.pngcamCan16cm.jpg 
  13397   Mon Oct 23 09:17:41 2017 SteveUpdatePEMants alartDo not leave organic trash or food boxes
in the 40m to attrack ants !
 ants.jpgants2.jpg 
  13399   Tue Oct 24 16:43:11 2017 SteveUpdateCDSslow machine bootfest[ Gautam , Steve ]

c1susaux
  
  13402   Fri Oct 27 09:34:20 2017 SteveUpdatePEMearthquakesLompoc 4.3M and 3.7M Avalon

 
 recentEQs.png 
  13403   Fri Oct 27 10:14:11 2017 SteveUpdateVACRGA scan at day 372Valve configuration: Vacuum normal

Note: Tp2 running at 75Krpm 0.25A
26C has a  load high pitch sound today.
 RGA_scan_d372.png 
  13406   Mon Oct 30 08:08:06 2017 SteveUpdatesafetysafety trainingUdit Kahndelwal received 40m specific basic
safety traning on Friday, Oct. 27
  
  13407   Mon Oct 30 10:09:41 2017 SteveUpdateVACTP2 failed IFO pressure 1.2e-5 Torr at 9:30am




 tp2failed.pngifo_1.0E-5_Torrit.pngtp2failed2dago.png4days.png 
  13409   Mon Nov 6 09:09:43 2017 SteveUpdateVACsetting up new TP2 turboOur new Agilent Technology TwisTorr
84FS AG rack controller ( English Manual
pages 195-297 )  RS232/485, product
  
  13423   Fri Nov 10 08:52:21 2017 SteveUpdateVACTP3 drypump replacedPSL shutter closed at 6e-6 Torr-it    

   The foreline pressure
of the drypump is 850 mTorr at 8,446 hrs
  
  13424   Fri Nov 10 13:46:26 2017 SteveUpdatePEMM3.1 local earthquakeBOOM
SHAKA-LAKA
 3.1M_local_EQ.png 
  13426   Tue Nov 14 08:54:37 2017 SteveUpdateIOOMC1 glitching   MC1_glitching.png 
  13454   Sun Nov 26 19:38:40 2017 SteveUpdateVACTP3 drypump replaced againThe TP3 foreline pressure was 4.8 Torr,
50K rpm 0.54A and 31C........Maglev rotation
normal 560 Hz.......    IFO pressure
 after_replacement.png 
  13459   Thu Nov 30 10:38:39 2017 SteveUpdateVACannuloses are not pumpedAnnuloses are not pumped for 30 days, since
TP2 failed.  IFO pressure 7e-6 Torr
it,  Rga 2.6e-6 Torr
 AnnulosesNotPumped.png 
  13475   Thu Dec 14 08:59:17 2017 SteveUpdateGeneralwe are here   8_days.png 
  13492   Tue Dec 26 17:24:24 2017 SteveUpdateGeneralpower outageThere was a power outage.

The IFO pressure is 12.8 mTorr-it
and it is not pumped. V1 is still closed.
 powerOutage.png 
  13495   Tue Jan 2 15:43:35 2018 SteveUpdateVACpumpdown after power outage 




 pumpdown_from_44_mTorr.png 
  13499   Wed Jan 3 15:13:55 2018 SteveUpdateGeneralprojector light bulb replacedBulb 
is replaced.

  
  13521   Wed Jan 10 09:49:28 2018 SteveUpdatePEMthe rat is backFive mechcanical traps set inside of boxes.
Red-white warning tape on top of each.

  
  13526   Wed Jan 10 16:27:02 2018 SteveConfigurationSEIload cell for weight measurementWe could use similar
load cells   to make the actual
weight measurement on the Stacis legs. This
 stacis3LoadCells.png 
  13530   Thu Jan 11 09:57:17 2018 SteveUpdateDAQacromag at ETMXGood going Johannes!




 Acromg_in_action.png 
  13532   Thu Jan 11 14:47:11 2018 SteveUpdatePSLshelf work for tomorrowI have just received the scheduling of
the PSL self work for tomorrow. Gautam and
I agreed that if it is needed I will shut
  
  13538   Fri Jan 12 10:26:24 2018 SteveUpdatePSL PSL shelf work scheduleMeasurements for good fit were made. The
new shelf will be installed on next Tuesday
at 2pm
  
  13550   Tue Jan 16 16:18:47 2018 SteveUpdatePSLnew PSL shelf in place[ Johannes, Rana, Mark and Steve ]

On the second trial the shelf was
installed. Plastic cover removed. South end
 DSC00020.JPG 
  13570   Tue Jan 23 16:02:05 2018 SteveConfigurationSEIload cells1500 and 2000 lbs load cells arrived from
MIT to measure the vertical loads on each
leg.
 stacis3LoadCells.pngDSC00025.JPGDSC00026.JPG 
  13581   Thu Jan 25 08:27:25 2018 SteveUpdatePEMearthquakesM4 local earthquake at 10:10 UTC   
There is no sign of damage.

....here is an other one.........M5.8
 M4_local_eq.pngM4.pngM5.8Ferndale_CA.png 
  13588   Tue Jan 30 10:22:20 2018 SteveUpdateVACRGA scan at day 175pd80b rga scan at 175 day.  IFO pressure
7.3e-6 Torr-it

Condition: vacuum normal, annuloses
 RGApd80-b-468-175d.pngRgaOn20hrs.png 
  13592   Wed Jan 31 15:46:05 2018 SteveUpdatesafetycrane inspectionAnnual crane inspection with load tests
is scheduled for Monday, Feb 5, 2018 from
8 to 11:30am
  
  13615   Wed Feb 7 15:50:42 2018 SteveUpdateVACIFO pressure monitoringHornet cold cathode gauge analoge output 
[ DSub9 pin 3 and 7 ] are wired to go ETMX
Acromag. It was reading 4.9V at 7.8e-6 Torr
 CC_analog.png 