  9595   Tue Feb 4 01:02:03 2014 KojiUpdateGreen LockingETMX green powerManasa, Steve: Please revisit the Xend
oven temperature again.
 Screenshot-Untitled_Window.png 
  9621   Mon Feb 10 22:21:55 2014 manasaUpdateGreen LockingX and Y arm green tunedY arm green: Nothing
much was disturbed. I touched the steering
mirrors and brought 
  
  9624   Tue Feb 11 21:22:02 2014 manasaUpdateGreen LockingALS X and Y arm restoredThe X and
Y arms were locked successfully using ALS
  
  9625   Tue Feb 11 22:17:06 2014 KojiUpdateGreen LockingALS X and Y arm restoredNice restoration. We eventually want to
make transition of the servo part from ALS
to LSC model for the further handing off
  
  9633   Thu Feb 13 16:48:33 2014 SteveUpdateGreen LockingX and Y arm green tuned details

    

       
            
  
  9665   Mon Feb 24 17:21:42 2014 SteveUpdateGreen Lockinggreen fiber status today

    

       
            
 6x 
  9721   Tue Mar 11 19:38:26 2014 manasaUpdateGreen LockingALS Slow servo settings

    

       
            
  
  9727   Fri Mar 14 10:31:10 2014 jamieUpdateGreen LockingALS Slow servo settings

    

       
            
  
  9837   Mon Apr 21 23:33:57 2014 ranaSummaryGreen LockingHP 8591E reads low by 140 Hz out of 10 MHzTo
check the basolute frequency stability of
the old monochrome HP 8591E RF Spectrum analyzer
  
  9937   Fri May 9 11:23:11 2014 steveUpdateGreen Lockingdecreased X green light power Green light power decreased from 3 mW to
1 mW at the ETMX-ISCT shutter. More later.
  
  9939   Fri May 9 21:18:51 2014 KojiUpdateGreen LockingReverted X green light power It is actually very tricky to measure the
green power at the output of the doubling
crystal as the IR often leaks into the measurement.
  
  9990   Fri May 23 11:58:28 2014 manasaUpdateGreen LockingY arm green alignment tunedThe Y arm green transmission had come down
to 0.3 and the green steering mirrors on
the Y end table required some minor alignment
  
  10396   Thu Aug 14 22:58:59 2014 rana, jenneSummaryGreen LockingALS DIFF tuning We've been having trouble tuning
the ALS DIFF matrix. Trying to see if the
MC2 EXC can be cancelled in ALS DARM by adjusting
 sweep.pdf 
  10407   Mon Aug 18 18:33:57 2014 ericqUpdateGreen LockingYarm Green PDHSo far today, I've been working with the
Y-end green PDH locking. Using a SR560 to
roll off the AG4395A output to take a loop
 yLoop.pdfcrap.pdf 
  10408   Tue Aug 19 01:01:36 2014 Jenne, RanaUpdateGreen LockingYarm Green PDH[ Rana, Jenne]
We remeasured the Yend PDH box.
When we first started, the green
 YPDH_OLG.pdfYPDH_noise.pdf 
  10409   Tue Aug 19 18:32:40 2014 ericqUpdateGreen LockingYarm Green PDHHeading to dinner, going to come back for
more green fun, but here's a quick update:
Xarm Peak-to-Peak of the PDH signal
  
  10411   Tue Aug 19 23:11:15 2014 JenneUpdateGreen LockingYarm Green PDH 
 Here is a plot of last night's
data with both the control and the error
 ErrAndCtrlSpectra_VoltsPerRtHz.png 
  10412   Wed Aug 20 02:38:41 2014 JenneUpdateGreen LockingYarm Green PDH - requirement

    

       
            
 NoiseConsideration.pdf 
  10413   Wed Aug 20 04:09:21 2014 ericqUpdateGreen LockingXarm Green PDHI've made
a whole bunch of measurements on the Xarm
green situation.
TL;DRs:

 Xbodes.pdfXspectra.pdfscopeSweep.jpg 
  10415   Wed Aug 20 16:10:43 2014 ericqUpdateGreen LockingXarm Green PDHA MIST simulation tells me that the green
pdh horn-to-horn displacement is about 1.2nm,
or ~18kHz. I used this, along with the scope
 XspectraCombined.pdf 
  10416   Wed Aug 20 18:05:18 2014 JenneUpdateGreen LockingYarm Green PDH - requirement 
 I calibrated the control signal
from Volts to Hz using the rough PZT calibration
 YPDH_noise.pdf 
  10417   Wed Aug 20 21:09:16 2014 ericqUpdateGreen LockingXarm Green PDHI remeasured all of the noise spectra again
today, making sure the input attenuation
was as low as it could safely be. I also
 scopeSweep.jpgXspectra.pdfYspectra.pdfXYcomp.pdf 
  10418   Thu Aug 21 02:42:17 2014 rana, ericqConfigurationGreen LockingGain changes on Green Y PDH[rana, ericq]
We spent time trying to relieve the
Yend green PDH of it troubles. 
 TEK00002.PNGTEK00003.PNG 
  10420   Thu Aug 21 19:04:52 2014 ericqConfigurationGreen LockingGain changes on Green Y PDHI found that the barrel of one the BNC
to BNC connectors used for getting the output
of the PDH servo box to the laser controller
 TEK00005.PNG 
  10430   Tue Aug 26 23:16:49 2014 ericqConfigurationGreen LockingGain changes on Green Y PDH Yesterday I measured the spectra
and OLTF of the Y-Arm green PDH, after the
LO touch-up and 60Hz hunt from last week.
 Ytfs.pdfpdhComparison.pdfnewY.zip 
  10433   Wed Aug 27 18:03:47 2014 ericqUpdateGreen LockinguPDH Box Checkup Quick post of plots and data; I'll
fill in more detail tonight. 
TL;DR: I pulled both green PDH boxes
 uPDH_X.JPGuPDH_Y.JPGupdhTFs.pdfupdhNoises.pdfuPDH_Aug27.zip 
  10437   Thu Aug 28 17:34:20 2014 ericqUpdateGreen LockinguPDH Box CheckupI had noticed in the past, that the digital
control signal monitor for the X end would
saturate well before the ADC should saturate
 TEK00006.PNG 
  10438   Fri Aug 29 17:28:07 2014 ranaUpdateGreen LockinguPDH Box Checkup 
  
  10439   Fri Aug 29 23:53:35 2014 ericqUpdateGreen LockinguPDH Box CheckupI had pulled out both X and Y servo boxes
for inspection, put the Y box back, soldered
in a missing op amp power capacitor on the
  
  10441   Tue Sep 2 20:02:06 2014 ericqUpdateGreen LockinguPDH Box CheckupI narrowed down the saturation point in
the X green PDH box to the preamp inside
the AD8336, but there is still no clear answer
  
  10443   Wed Sep 3 00:17:22 2014 ranaUpdateGreen LockinguPDH Box CheckupWhat monitor point is being plotted here?
Or is it a scope probe output?
If this saturation is in the uPDH-X
  
  10447   Wed Sep 3 20:38:40 2014 ericqUpdateGreen LockinguPDH Box CheckupThe traces were from the front panel output
BNCs, but the VGA preamp exhibited this asymmetric
saturation at its output.  
  
  10457   Fri Sep 5 04:07:44 2014 ericqUpdateGreen LockingX end uPDH Box ReplacedJust a quick note, plots and data will
come tomorrow:
I grabbed an unused uPDH board from
  
  10460   Fri Sep 5 16:13:00 2014 ericqUpdateGreen Lockingnew X Green PDH measurements I measured the noise spectra and
loop TF of the green PDH with the newly stuffed
board. Unfortunately, I never took the noise
 6x 
  10463   Sat Sep 6 01:48:54 2014 ranaUpdateGreen Lockingnew X Green PDH measurements

    

       
            
  
  10464   Sat Sep 6 14:49:12 2014 ericqUpdateGreen Lockingnew X Green PDH measurementsI used a modulation depth of 0.3, which,
if I recall correctly, is what we aimed for
on the Y-arm when we adjusted the LO signal
 xGComparison.pdfxGspectra.pdf 
  10465   Sun Sep 7 17:06:38 2014 ranaUpdateGreen Lockingnew X Green PDH measurements600 Hz seems ~OK. From the measured reflectivities
for 532
nm, the green Finesse = 108. So
  
  10471   Mon Sep 8 14:14:13 2014 JenneUpdateGreen Lockingnew Y Green PDH measurementsThese are plots and notes from last week's
PDH adventures. 

For the PDH servo box re-design,
 6x 
  10666   Tue Nov 4 14:46:00 2014 manasaUpdateGreen LockingMissing beatnotesSummary: Cannot find beatnotes
between the arms and PSL.
I wanted to measure the ALS out of
  
  11334   Thu May 28 21:10:46 2015 KojiUpdateGreen LockingALS-X noise huntingI have been looking at the X-end ALS setup.
I was playing with the control bandwidth
to see the effect to the phase tracker output
 150528_ALS.pdf 
  11361   Mon Jun 15 22:36:40 2015 rana, kojiUpdateGreen LockingBeatBox Assay: not looking goodBecause the ALS beatbox schematic is out-of-date
and misleading, we pulled the box to photograph
the current implementation and figure out
  
  11363   Fri Jun 19 01:24:26 2015 rana, kojiUpdateGreen LockingBeatBox Assay: not looking goodWe had decided a few days ago, to bypass
the IF part of the BeatBox board and put
some of the Haixing Maglev generic filter
  
  11364   Fri Jun 19 01:55:35 2015 ericqUpdateGreen LockingBeatBox Assay: not looking good


  
  11604   Wed Sep 16 03:37:06 2015 KojiSummaryGreen LockingWorkable delay line setup prepared[Koji Gautam]

The variable delay line has been
setup for practical use. The hardware and
 6x 
  11715   Mon Oct 26 19:10:59 2015 gautamUpdateGreen LockingAUX PDH loop characterizationI began my attempts to characterize the
PDH loops at the X end today. My goal was
to make the following measurements:
  
  11865   Tue Dec 8 23:24:08 2015 gautamUpdateGreen LockingY end laser (Lightwave) PZT calibrationSummary:

I
measured the PZT actuator gain for the Lightwave
 Ycalib_8Dec.pdf 
  11877   Sun Dec 13 21:55:28 2015 gautamUpdateGreen LockingY end laser (Lightwave) PZT calibrationSummary:

After
the discussions at the Wednesday meeting,
 IMG_5972.JPGServoY_TF_13Dec2015.pdfDatanCode.zipPSD_916Hz.pdf 
  11879   Mon Dec 14 16:27:11 2015 gautamUpdateGreen LockingY-end AUX PDH noise breakdownSummary:

I've
attached the results from my measurements
 YAUX_NB_Dec2015.pdfPDH_errSig_Calib.pdf 
  11887   Wed Dec 16 18:34:40 2015 gautamUpdateGreen LockingGreen beat channels temporarily set up as IR beat channelsSince there
are a few hours to go before the locking
efforts tonight, I've temporarily borrowed
  
  11888   Wed Dec 16 23:15:28 2015 ericqUpdateGreen LockingGreen beat channels temporarily set up as IR beat channelsWith the IR beats going to the nominal
ALS channels as Gautam left them, we're
able to measure the free running frequency
 2015-12-16_AUXfreerunning.pdf 