IDup Date Author Type Category Subject Text Attachments
  7698   Mon Nov 12 23:38:50 2012 CharlesUpdateElectronicsEthernet Illuminator Control

    

       
            
  
  7699   Tue Nov 13 00:15:08 2012 JenneUpdateIOOInput and Output PZTs turned onI turned on the power supplies for the
output PZTs and pitch and yaw for PZT2. 
This is back to the condition that we had
  
  7700   Tue Nov 13 00:19:51 2012 JenneUpdateCamerasITMX is just fine

    

       
            
  
  7701   Tue Nov 13 00:27:50 2012 JenneUpdateElectronicsEthernet Illuminator Control

    

       
            
  
  7702   Tue Nov 13 00:28:10 2012 AyakaUpdateIOOMC locked, spots centered[Rana, Ayaka, Jenne]
We aligned the REFL beam to the center
of PD.
 spot1112.pngDSC_4902.JPGDSC_4904.JPG 
  7703   Tue Nov 13 08:34:11 2012 SteveSummaryVACpumpdown after 24days at atmosphere 
 
 pd74d1.png24datm.pngpdd1.png 
  7704   Tue Nov 13 11:30:54 2012 ManasaConfiguration40m UpgradingEndtable upgrade for auxiliary green laserI'm set on the mission to get the new bigger endtables setup  ETMX_sch.png 
  7705   Tue Nov 13 16:18:51 2012 ManasaConfiguration40m UpgradingEndtable upgrade for auxiliary green laser : Circularize the 1W NPRO beam profileWith reference to measurements made earlier: elog, 
Beam parameters for Innolight 1W
  
  7706   Tue Nov 13 20:13:07 2012 KojiConfiguration40m UpgradingEndtable upgrade for auxiliary green laser : Circularize the 1W NPRO beam profileThese fs are too short.


  
  7707   Tue Nov 13 21:05:22 2012 AyakaUpdateWienerFilteringAcoustic noise cancellation with MC In order to perform acoustic noise
cancellation with MCL signal, I am trying
to find sweet spots for microphones.
 9x 
  7708   Tue Nov 13 21:05:35 2012 DenUpdateAdaptive Filteringonline and simulationFor a last few days I've been working on
oaf and simulink model to simulate it. First
I did online subtraction from MC when MC_L
 8x 
  7709   Tue Nov 13 22:12:15 2012 JenneUpdateIOOTurn off MCL path before doing MC spot measurement[Jenne, Den]
We have discovered that the MCL loop
squishes the length fluctuations that result
  
  7710   Wed Nov 14 00:56:19 2012 JenneUpdateIOOWFS on, PZT2 ~set, AS camera backWFS
are back on, and working nicely. 
Den and I had seen a problem (which I had
  
  7711   Wed Nov 14 09:01:42 2012 Dumb statement catcherUpdateIOOTurn off MCL path before doing MC spot measurement

    

       
            
  
  7712   Wed Nov 14 13:58:18 2012 JenneUpdateIOOTurn off MCL path before doing MC spot measurement

    

       
            
  
  7713   Wed Nov 14 21:59:09 2012 DenUpdateCDSdaq errorsI tried to add a test point to C1MCS model
and spent next two hours rebooting front-ends,
restarting models and realigning MC.
  
  7714   Thu Nov 15 02:18:24 2012 DenUpdateModern ControlBS oplevI've applied LQR feedback technique to
BS oplev in pitch. I think the most inconvenient
thing in using LQR controller is the amount
 lqr_cost.pngolpit_hg.png 
  7715   Thu Nov 15 03:09:08 2012 JenneUpdateAlignmentNo good progress, many optionsI didn't make any concrete progress today.
AS and REFL cameras are in place, and Manasa
has put ND 0.5 filters on both. I used a
  
  7716   Thu Nov 15 21:52:48 2012 Den, AyakaUpdateGreen Lockingyarm lockedWe aligned and locked Y arm for green:

    installed camera on PSL
to monitor green transmission
    aligned
 yarm.pngetmyf.png 
  7717   Fri Nov 16 00:08:36 2012 KojiUpdateSUSMC2 woesPeople complained about the MC instability:
If
we aligned the MC, it locked nicely for a
  
  7718   Fri Nov 16 03:12:39 2012 Ayaka, DenUpdateGreen Lockingxarm lockedWe aligned and locked xarm for green.
 etmxf.png 
  7719   Fri Nov 16 09:57:57 2012 JenneUpdateGreen Lockingxarm locked

    

       
            
  
  7720   Sat Nov 17 03:30:13 2012 Den, AyakaUpdateAlignmentred in armsWe aligned accurately 00 green in yarm,
changed voltage on PZT2 to see red flashing
at TRY at the normalized level 0.2-0.3. The
  
  7721   Sat Nov 17 18:02:14 2012 DenUpdateAlignmentred in arms

    

       
            
  
  7722   Sat Nov 17 22:50:17 2012 KojiUpdateAlignmentred in armsYou have constraints for the IR beams (i.e.
one PZT and one BS for 8 dofs), so now you
need to align the arms for the input IR beams.
  
  7723   Mon Nov 19 15:12:52 2012 JenneUpdateAlignmentYarm locked IR

    

       
            
  
  7724   Mon Nov 19 15:15:22 2012 JenneUpdateSUSAll oplev gains turned on

    

       
            
  
  7725   Mon Nov 19 15:58:02 2012 SteveUpdateSAFETYsafety trainingCharles Blakemore  our new undergrad
help received 40m specific safety training
today.
  
  7726   Mon Nov 19 20:03:53 2012 ManasaConfiguration40m UpgradingEndtable upgrade for auxiliary green laser : ETMX layout on new tableI have attached the possible layout of
the optics on the new ETMX endtable. More
optics have been added when compared to the
 ETMX_endtable_New_Model.pdf 
  7727   Mon Nov 19 20:17:53 2012 JenneConfiguration40m UpgradingEndtable upgrade for auxiliary green laser : ETMX layout on new table 
For convenience, I would include
a steering mirror in front of the TRX PD. 
  
  7728   Mon Nov 19 22:42:14 2012 KojiConfiguration40m UpgradingEndtable upgrade for auxiliary green laser : ETMX layout on new table
    I don't like the idea to
place the cylindrical lenses right in front
of the laser. 
  
  7729   Mon Nov 19 23:14:31 2012 EvanUpdateCamerasETMYF focusAdjusted focus on ETMYF camera so that
the IR beam is in focus.
  
  7730   Tue Nov 20 02:57:24 2012 Ayaka, Den, KojiUpdateLockingred in armsWe aligned and locked x and y arms.
MCL loop makes arms lock unstable,
adds a lot of noise at frequencies 60-100
 8x 
  7731   Tue Nov 20 11:40:19 2012 ranaConfigurationGreen LockingEnd table upgrade for auxiliary green laser : ETMX layout on new table 
 Mounts:

    No more mounts using the
  
  7732   Tue Nov 20 15:11:22 2012 SteveUpdateGeneralfew more sensing cardsNew  Lumitek IR
Sensor Cards are here. We got 2 pieces
of Q-11-T (2" x 2"), 2 pieces of
  
  7733   Tue Nov 20 15:24:52 2012 AyakaUpdatePEMSeismometers and a microphone I got two seismometers and one microphone
back from Tara.
They are now near the Gurlap under
  
  7734   Tue Nov 20 16:03:37 2012 janoschUpdatePEMSeismometers and a microphone
  
  7735   Tue Nov 20 20:37:51 2012 KojiUpdateIOOMC autolockerLast night I found that the MC autolocker
has not been updated since the chamber was
closed.
  
  7736   Wed Nov 21 01:31:37 2012 Koji, AyakaUpdateLockingalignment on ETMX tableSince the transmission beam on ETMXT camera
seemed to be clipped, we checked the optics
on ETMX table.
 output.nv12.bmp 
  7737   Wed Nov 21 01:36:36 2012 KojiUpdateIOOMCL disabled / WFS clear history restoredAs MCL is disturbing arm locking by injecting
a lot of noise, I have modified 'mcup' 
to disable MCL
  
  7738   Wed Nov 21 21:06:13 2012 AyakaUpdateLSCcalibration of armsMotivation
In order to estimate whether we can
see acoustic coupling in arms or not, we
 7x 
  7739   Sat Nov 24 13:58:07 2012 ranaUpdateLSCcalibration of armsFor the loop diagnosis, its best
to use the method of "IN1/IN2",
rather than manipulate the close loop gain.
  
  7740   Sat Nov 24 22:14:08 2012 KojiUpdateGeneralAP table cleaning upOn Wednesday (21st) night, I checked the
AP table as I wanted to try PRMI locking. 
It was difficult to work with the
 APTable121124.pdfAPTable121124.graffle.zip 
  7741   Sat Nov 24 23:50:35 2012 KojiUpdateIOOMC WFS refusing to workToday I found the IMC was misaligned significantly
by WFS feedback.
Once the feedback was cleared, it
  
  7742   Mon Nov 26 10:06:51 2012 ranaSummaryIOOMC slides from 2002   MCtalk.pdf 
  7743   Mon Nov 26 10:42:06 2012 AyakaUpdateLSCcalibration of armsI uploaded a zip file that contains data
files used for the calibration.
OLTF_x/y.txt: the open loop transfer
 armcalib.zip 
  7744   Mon Nov 26 10:58:20 2012 SteveUpdatePEMair cond maitenance tomorrowAir conditioning maintenance is scheduled
for tomorrow morning till noon.
  
  7745   Mon Nov 26 18:36:17 2012 JenneUpdatePEMBLRMS back

    

       
            
  
  7746   Mon Nov 26 18:56:34 2012 JenneHowToComputersData logging suggestionsWe've been talking for a while about how
we want to store data.  I'm not in love
with keeping it on the elog, although I think
  
  7747   Mon Nov 26 19:27:59 2012 RijuHowTo Testing AG4395A+GPIBRiju, Jenne
We have checked the transfer function
of a bandpass filter using AG4395A network
 tfmag.jpgtfphase.jpg 