5485
|
Tue Sep 20 16:45:09 2011 |
Jenne | Update | SUS | SUS diag stuff... just so I remember what I'm doing |
Has the Q been checked? Still in progress...
Optic |
POS |
PIT |
YAW |
SIDE |
ITMX |
done |
done |
done |
done |
ITMY |
done |
done |
fine?? |
done |
ETMX |
done |
done |
done |
done |
ETMY |
done |
done |
done |
done |
BS |
done |
done |
done |
done |
PRM |
done |
done |
done |
done |
SRM |
done |
done |
done |
done |
MC1 |
|
|
|
|
MC2 |
|
|
|
|
MC3 |
|
|
|
|
So, update as of 6:17pm: I have tuned the damping gains for all IFO optics. Everything is good, except for ITMY Yaw. It's probably fine, the optic damps okay, but it doesn't look like a nice clean ringdown. I haven't taken the time to go back and look at it again.
I have to go to a dinner, but later (probably in the morning, so I don't disturb evening locking) I'll check the MC Qs. |
5486
|
Tue Sep 20 17:45:30 2011 |
kiwamu | Update | CDS | daqd is restarting by hisself ? |
[Jenne / Kiwamu]
Fb was sick. Dataviewer and Fourier Tools didn't work for a while.
After 10 minutes later they became healthy again. No idea what exactly was going on.
One thing we found was that : during the sickness of fb, it looks like daqd was restarting by hisself. Is this normal ??
Here is the bottom sentences of restart.log. Apparently daqd was rebooting although we didn't command to do so.
daqd_start Tue Sep 20 02:41:17 PDT 2011
daqd_start Tue Sep 20 13:18:12 PDT 2011
daqd_start Tue Sep 20 17:33:00 PDT 2011
|
5487
|
Tue Sep 20 18:03:45 2011 |
Paul | Update | SUS | ITMY and SRM oplev calibrations - measured and estimated |
The measured calibration factors for the oplevs are as follows:
SRM pitch: 666urad per count on channel C1-SUS-SRM-OLPIT-INMON
SRM yaw: 557urad per count on channel C1-SUS-SRM-OLYAW-INMON
ITMY pitch: 470urad per count on channel C1-SUS-ITMY-OLPIT-INMON
ITMY yaw: 491urad per count on channel C1-SUS-ITMY-OLYAW-INMON
Since I'm going to calibrate all the other oplevs with the rougher technique of estimating the angle from the OSEM signals directly, I thought I would check the result of such an estimation for the oplevs I have calibrated already. My method was as follows:
dA = change in angle
dx = change in OSEM flag position
dV = change in OSEM PD voltage
dC = change in OSEM counts
D = optic diameter
L = distance between OSEMs = D/sqrt(2)-0.002m = 0.052m
dV/dx = OSEMs volts per meter flag position change = 1700 V/m
dC/dV = OSEM counts per volt = 2^16/40 = 65536/40 counts/V
counts per radian = dC/dA = dV/dx x dC/dV x 1/L = 1700*65536/40/0.052 = 5.3564x10^7 counts/rad
radians per count = dA/dC = 1.867x10^-8, or 0.019 urad/count
This is around a factor of 1000 smaller than what I measured earlier, reported in entry 5468. I guess this might be an issue with the whitening filter on the OSEMs, but my initial feeling was that this was only a factor of a few. If anyone can see a big obvious mistake in my above calculations please let me know!
|
5488
|
Tue Sep 20 19:00:49 2011 |
Paul | Update | SUS | ITMY and SRM oplev calibrations - measured and estimated |
Kiwamu noticed that the 1/L in the counts per radian should have just been L, which accounts for most of the discrepancy. We checked the input filters on the OSEMs, and they have 10dB of gain at DC. Accounting for this, estimates on the order of 20urad/count, which is much more reasonable! |
5490
|
Tue Sep 20 21:13:39 2011 |
Suresh | Update | IOO | MC aligned and PSL beam into MC readjusted |
This morning after Kiwamu maximised the PSL beam coupling into the MC we noticed that the MC2 face camera showed the spot position had moved away from the center by about a diameter. So I checked the beam spot positions with MCASS and indeed found that the spot on MC2 had moved to about 6mm away from the center in yaw and about 3mm in pitch. I adjusted the MC2 (and only MC2) to recenter the spots on all the three mirrors. The new spot positions are given below
spot positions in mm (MC1,2,3 pit MC1,2,3 yaw):
1.3337 -0.2660 0.6641 -1.0973 0.0468 -1.7130
The PSL beam into MC has been readjusted for maximal coupling into MC.
|
5491
|
Tue Sep 20 23:01:37 2011 |
Keiko | Update | IOO | AM modulation mistery |
Keiko, Suresh
AM modulations are still there ... the mechanical design for the stages, RF cables, and connections are not good and affecting the alignment.
I write the activity in the time series this time - Because we suspect the slight EOM misalignment to the beam produces the unwanted AM sidebands, we tried to align the EOM as much as possible. First I aligned the EOM tilt aligner so that the maximum power goes through. I found that about 5% power was dumped by EOM. After adjusting the alignment, the AM modulation seemed be much better and stable, however, it came up after about 20 mins. They grew up up to about -40dBm, while the noise floor is -60 dBm (when AM is minimised, with DC power of 8V by PDA225 photodetector).
We changed the EOM stage (below the tilt aligner) from a small plate to a large plate, so that the EOM base can be more stable. The EOM stands on the pile of several black plate. There was a gap below the tilt aligner because of a small plate. So we swapped the small plate to large plate to eliminate the springly gap. However it didn't make any difference - it is the current status and there is still AM modulations right now.
During above activities, we leaned that the main cause of the EOM misalignment may be the RF cables and the resonator box connected to the EOM. They are connected to the EOM by an SMA adaptor, not any soft cables. It is very likely applying some torc force to the EOM box. The resonator box is almost hunging from the EOM case and just your slight touch changes EOM alinment quite a bit and AM mod becomes large.
I will replace the SMA connector between the resonator box and EOM to be a soft cable, so that the box doesn't hung from EOM tomorrow. Also, I will measure the AM mod depth so that we compare with the PM mod depth.
Quote: |
Keiko, Anamaria
We started to investigate the AM modulation mistery again. Checking just after the EOM, there are AM modulation about -45dBm. Even if we adjust the HWP just before the EOM, AM components grow up in 5 mins. This is the same situation as before. Only the difference from before is that we don't have PBS and HWP between the EOM and the monitor PD. So we have a simpler setup this time.
We will try to align the pockells cell alignment tomorrow daytime, as it may be a problem when the crystal and the beam are not well parallel. This adjustment has been done before and it didn't improve AM level at that time.
|
|
5493
|
Wed Sep 21 00:34:29 2011 |
rana | Update | SUS | SUS diag stuff... just so I remember what I'm doing |
ETMX was ringing up when it was mis-aligned for Y arm locking. I restored the input matrix to something more diagonal and its now damping again. Needs more work before we can use the calculated matrix. |
5494
|
Wed Sep 21 00:37:01 2011 |
rana | Update | SUS | ITMY and SRM oplev calibrations - measured and estimated |
I found that some of the Optical Lever Servos were ON today and injecting nonsense into the interferometer optics. I have set all of the gains = 0 to save us more headaches.
Please leave them OFF until we review the servo and noise characterization results in the elog. |
5496
|
Wed Sep 21 09:10:15 2011 |
Paul | Update | SUS | ITMY and SRM oplev calibrations - measured and estimated |
Quote: |
I found that some of the Optical Lever Servos were ON today and injecting nonsense into the interferometer optics. I have set all of the gains = 0 to save us more headaches.
Please leave them OFF until we review the servo and noise characterization results in the elog.
|
I had previously set the gains to zero, see the first line of my entry on Monday 5468. I should have the servo and noise characterisation done today for these oplevs today, so we can review it soon. |
5497
|
Wed Sep 21 11:35:07 2011 |
steve | Update | VAC | RGA scan |
RGA scan with maglev pumping speed at day 14 of the pump down.
The larger inserted box contains the tuning parameters of the SRS 200 amu RGA |
Attachment 1: rgascanpd71m14d.jpg
|
|
5499
|
Wed Sep 21 14:44:25 2011 |
Paul | Update | SUS | ITMY and SRM open loop transfer functions |
Here are the open loop transfer functions for ITMY and SRM. The various settings for the OLTFs were as follows:
Oplev filter used for all OLTFs: 300^2:0
Gains for oplev servos (for each OLTF only the 1 servo for the measured TF was on. They are all set back to 0 now):
SRM yaw gain = 1
SRM pitch gain = -1
ITMY yaw gain = -1
ITMY pitch gain = 1
measurement band = 0.2Hz to 200Hz
points = 33
swept sine magnitude envelope: amp = 2 for f > 60Hz, amp = 0.1 for f < 60Hz
Measurement points were from e.g. C1-SUS-ITMY-OLPIT-IN2 to C1-SUS-ITMY-OLPIT-IN1 to give a TF of -(loop gain).
Next step is to divide this through by the sensor reponse (i.e. the calibration factor measured earlier) and the filter response to get just the actuator response.
|
Attachment 1: ITMY_SRM_oplev_OLTFs.png
|
|
5500
|
Wed Sep 21 16:22:14 2011 |
rana | Update | SUS | Summary screen |
The SUS SUMMARY screen is now fully activated. You should keep it open at all times as a diagnostic of the suspensions.
No matter how cool you think you are, you are probably doing something bad when trying to lock, measure any loop gains, set matrices, etc. Use the screen.
This is the link to the automatic snapshot of the SUS SUMMARY screen. You can use it to check the Suspensions status with your jPhone.
Auto SUS SUMMARY Snapshot
When the values go yellow its near the bad level. When its red, it means the optic is misaligned or not damped or has the wrong gain, etc.
So don't ignore it Steve! If you think the thresholds are set too low then change them to the appropriate level with the scripts is SUS/ |
5501
|
Wed Sep 21 16:31:28 2011 |
Paul | Update | SUS | ITMY and SRM actuator response functions |
I divided the open loop transfer functions by the filter response and the sensor responses (previously measured calibration factors) to leave just the actuator responses. I've attached the actuator responses plotted in radians/count and phase over frequency.
Next step: fit the actuator response with poles and zeros.
EDIT: I divided by the wrong filter function earlier - the plots there now are divided by the correct filter function |
Attachment 1: ITMY_PITCH_actuator_response.png
|
|
Attachment 2: ITMY_YAW_actuator_response.png
|
|
Attachment 3: SRM_PITCH_actuator_response.png
|
|
Attachment 4: SRM_YAW_actuator_response.png
|
|
5502
|
Wed Sep 21 16:44:18 2011 |
Keiko | Update | IOO | AM modulation mistery |
AM modulation depths are found to be 50 times smaller than PM modulation depths.
m(AM,f1) ~ m(AM, f2) = 0.003 while m(PM, f1)=0.17 and m(PM, f2)=0.19.
Measured values;
* DC power = 5.2V which is assumed to be 0.74mW according to the PDA255 manual.
*AM_f1 and AM_f2 power = -55.9 dBm = 2.5 * 10^(-9) W.

AM f2 power is assumed to be the similar value of f1. I can't measure f2 (55MHz) level properly because the PD (PDA255) is 50MHz bandwidth. From the (P_SB/P_CR) = (m/2) ^2 relation where P_SB and P_CR are the sideband and carrier power, respectively, I estimated the rough the AM modulation depths. Although DC power include the AM SB powers, I assumed that SB powers are enough small and the DC power can be considered as the carrier power, P_CR. The resulting modulation depth is about 0.003.
On the other hand, from the OSA, today's PM mod depths are 0.17 and 0.19 for f1 and f2, respectively. Please note that these numbers contains (small) AM sidebands components too. Comparing with the PM and AM sideband depths, AM sidebands seems to be enough small.
Quote: |
Keiko, Suresh
AM modulations are still there ... the mechanical design for the stages, RF cables, and connections are not good and affecting the alignment.
|
|
Attachment 1: P9210138.JPG
|
|
5503
|
Wed Sep 21 17:42:35 2011 |
rana | Update | IOO | AM modulation misery |
I'd like to see some details about how to determine that the ratio of 1:50 is small enough for AM:PM.
* What have people achieved in past according to the elogs© of the measurements?
* What do we expect the effect of 1:50 to be? How much offset does this make in the MICH/PRC/SRC loops? How much offset is too much?
Recall that we are using frontal modulation with a rather small Schnupp Asymmetry...  |
5504
|
Wed Sep 21 18:53:03 2011 |
Keiko | Update | IOO | AM modulation misery |
The signal offset due to the AM modulation is estimated by a simulation for PRCL for now. Please see the result below.
Too see how bad or good the AM modulation with 1/50 modulation depths of PM, I ran a simulation. For example I looked at PRCL sweep signal for each channel. I tried the three AM modulation depths, (1) m_AM=0 & m_PM = 0.17 (2) m_AM = 0.003 & m_PM = 0.17 which is the current modulation situation (3) m_AM = 0.17 & m_PM = 0.17 in which AM is the same modulation depth as PM. For the current status of (2), there are offsets on signals up to 0.002 while the maximum signal amplitude is 0.15. I can't tell how bad it is.... Any suggestions?
(1) m_AM=0 & m_PM = 0.17. There is no offset in the signals.

(2) m_AM = 0.003 & m_PM = 0.17. There are offsets on signals up to 0.002 while the maximum signal amplitude is 0.15.

(3) m_AM = 0.17 & m_PM = 0.17. There are offsets on signals up to 0.1 while the maximum signal amplitude is 0.2.

I will look at MICH and SRCL in the same way.
Quote: |
I'd like to see some details about how to determine that the ratio of 1:50 is small enough for AM:PM.
* What have people achieved in past according to the elogs© of the measurements?
* What do we expect the effect of 1:50 to be? How much offset does this make in the MICH/PRC/SRC loops? How much offset is too much?
Recall that we are using frontal modulation with a rather small Schnupp Asymmetry... 
|
|
5505
|
Wed Sep 21 19:20:41 2011 |
Suresh | Update | IOO | PSL beam into MC was off in Pitch. Readjusted. |
I found the PSL beam into the MC off in pitch by large amount. I readusted the PSL beam for optimal coupling.
The beam had shifted on the WFS as well. So I recentered the DC signal on the WFS with the MC unlocked. However both the DC and RF signals on the WFS shift when we lock the MC. This ought to indicate sub-optimal coupling of PSL into MC. But instead, if we were to reduce these offsets on the WFS by adjusting the MC axis it leads to higher reflected power from the MC.
The current plan is to retain these RF offsets and lock the WFS with a DC offset in the servo filters. |
5506
|
Wed Sep 21 21:13:35 2011 |
rana | Update | IOO | AM modulation misery |
How about changing the x-axis of all these plots into meters or picometers and tell us how wide the PRC resonance is? (something similar to the arm cavity linewidth expression)
Also, there's the question of the relative AM/PM phase. I think you have to try out both I & Q in the sim. I think we expect Q to be the most effected by AM. |
5507
|
Wed Sep 21 23:05:16 2011 |
Paul | Update | SUS | ITMY and SRM actuator response functions - fitting results |
I used an fminsearch function to fit the SRM and ITMY actuator response magnitudes. The testfunction was just that for a single second order pole, but it gave what I consider to be good fits for the following reasons:
*for 3 of the 4 fits the residuals were less than 0.5% of the summed input data points. The worst one (ITMY pitch) was about 2.7%, which I think is due to the resonance happening to be right in the middle of two data points.
*the tolerance of 1 part in 10^9 was reached quickly from not very finely tuned starting points.
The test function was: G=abs(Gp./(1+1i.*f./fp./Qp-(f./fp).^2)), where G(f) is the actuator response magnitude, Gp is the pole gain, fp is the pole frequency, and Qp is the pole Q factor.
In the end I just fitted the response magnitude. I was initially fitting the complex response function, but ran into problems which I think were cased by overall phase offsets between the data and test function. Can I canvass for opinion if fitting the magnitude is OK, or should I try again fitting the phase too?
Anyway, here are the results of the fits, and I've attached plots of each too (each one in linear and log y axis because each on its own might be misleading for fits):
EDIT - I added more points to the otherwise sparse looking fitted curves
ITMY PITCH actuator response fit
-- Fit completed after 190 iterations--
Started with: Gain = 3e-06,
Q factor = 5,
Pole frequency = 1,
Fit results: Gain = 1.32047e-06,
Q factor = 4.34542,
Pole frequency = 0.676676
Residual (normalised against the sum of input datapoints) = 0.0268321
ITMY YAW actuator response fit
-- Fit completed after 156 iterations--
Started with: Gain = 3e-06,
Q factor = 5,
Pole frequency = 1,
Fit results: Gain = 1.14456e-06,
Q factor = 8.49875,
Pole frequency = 0.730028
Residual (normalised against the sum of input datapoints) = 0.00468077
SRM PITCH actuator response fit
-- Fit completed after 192 iterations--
Started with: Gain = 3e-06,
Q factor = 5,
Pole frequency = 1,
Fit results: Gain = 7.94675e-06,
Q factor = 7.16458,
Pole frequency = 0.57313
Residual (normalised against the sum of input datapoints) = 0.00301265
SRM YAW actuator response fit
-- Fit completed after 156 iterations--
Started with: Gain = 3e-06,
Q factor = 5,
Pole frequency = 1,
Fit results: Gain = 3.34179e-06,
Q factor = 9.57601,
Pole frequency = 0.855322
Residual (normalised against the sum of input datapoints) = 0.000840468 |
Attachment 1: ITMY_PITCH_actuator_response_FIT.png
|
|
Attachment 2: ITMY_YAW_actuator_response_FIT.png
|
|
Attachment 3: SRM_PITCH_actuator_response_FIT.png
|
|
Attachment 4: SRM_YAW_actuator_response_FIT.png
|
|
5508
|
Wed Sep 21 23:25:51 2011 |
kiwamu | Update | SUS | Re:ITMY and SRM actuator response functions - fitting results |
Did you take the 180 deg shift into your account ?
Since your measurement was done when the loop was closed, there must be an additional 180 deg phase shift (in other words, minus sign).
Quote from #5507 |
In the end I just fitted the response magnitude. I was initially fitting the complex response function, but ran into problems which I think were cased by overall phase offsets between the data and test function. Can I canvass for opinion if fitting the magnitude is OK, or should I try again fitting the phase too?
|
|
5509
|
Wed Sep 21 23:44:45 2011 |
Paul | Update | SUS | Re:ITMY and SRM actuator response functions - fitting results |
Quote: |
Did you take the 180 deg shift into your account ?
Since your measurement was done when the loop was closed, there must be an additional 180 deg phase shift (in other words, minus sign).
Quote from #5507 |
In the end I just fitted the response magnitude. I was initially fitting the complex response function, but ran into problems which I think were cased by overall phase offsets between the data and test function. Can I canvass for opinion if fitting the magnitude is OK, or should I try again fitting the phase too?
|
|
I thought I had, but apparently not, and I'd made another error or two in the complex version of my fitting routine. I've fixed them now, thanks! I'll put up the new fitting results tomorrow morning. |
5510
|
Thu Sep 22 00:00:10 2011 |
Paul | Update | SUS | ITMY and SRM actuator response functions - complex fitting results |
Here are the results of the complex fitting. The residuals are bigger this time, but still probably small enough to be ok(?), with the possible exception of ITMY PITCH (due again I think to the data points straddling the resonance).
ITMY YAW actuator response complex fit
-- Fit completed after 282 iterations--
Started with: Gain = 3e-05,
Q factor = 5,
Pole frequency = 0.6776,
Fit results: Gain = 1.14673e-06,
Q factor = 12.9471,
Pole frequency = 0.766531
Residual (normalised against the sum of input datapoints) = 0.0688174
ITMY PITCH actuator response complex fit
-- Fit completed after 191 iterations--
Started with: Gain = 3e-05,
Q factor = 5,
Pole frequency = 0.6776,
Fit results: Gain = 1.25105e-06,
Q factor = 3.88981,
Pole frequency = 0.706744
Residual (normalised against the sum of input datapoints) = 0.144165
SRM YAW actuator response complex fit
-- Fit completed after 246 iterations--
Started with: Gain = 3e-05,
Q factor = 5,
Pole frequency = 0.6776,
Fit results: Gain = 3.34137e-06,
Q factor = 9.6875,
Pole frequency = 0.854913
Residual (normalised against the sum of input datapoints) = 0.0153646
SRM PITCH actuator response complex fit
-- Fit completed after 266 iterations--
Started with: Gain = 3e-05,
Q factor = 5,
Pole frequency = 0.6776,
Fit results: Gain = 7.97529e-06,
Q factor = 7.63888,
Pole frequency = 0.568227
Residual (normalised against the sum of input datapoints) = 0.0319653 |
Attachment 1: ITMY_PITCH_actuator_response_complex_FIT.png
|
|
Attachment 2: ITMY_YAW_actuator_response_complex_FIT.png
|
|
Attachment 3: SRM_PITCH_actuator_response_complex_FIT.png
|
|
Attachment 4: SRM_YAW_actuator_response_complex_FIT.png
|
|
5511
|
Thu Sep 22 01:05:28 2011 |
Katrin | Update | Green Locking | New modulation frequency (Y arm) |
[Kiwamu / Katrin]
On Wednesday, the green light was locked to the Y arm cavity.
Modulation frequency was changed from 279kHz to 178875Hz. The amplitude was changed from 10Vpp to 0.01Vpp to achieve a modulation index of 0.38. The modulation frequency was changed to minimize AM. With the new modulation frequency the laser light could still locked to the cavity.
The signal of the LO and the photodiode are multiplied by a ZAD-8 mini circuit mixer (Level 7). This mixer requires LO input is +7dBm = 1.4Vpp. Thus we put a 36dB attenuator between the LO and the PZT at the laser. PDH error signal shows lots of peaks that are most likely higher order sidebands. Thus, the next step is to work on the low-pass filter. However the SNR of the error signal has improved with the new modulation frequency. With the old mod. frequency the PDH signal was 4mVpp and the noise floor was 2mVpp.
Phase between the photodiode signal and LO is shifted by about 10 degrees. Step two is to work on a phase shifter.
|
5512
|
Thu Sep 22 01:45:41 2011 |
Keiko | Update | LSC | Locking status update |
Keiko, Anamaria
Tonight we want to measure the LSC matrix for PRMI and compare the simulation posted last night (#5495).
First. we locked MICH and PRCL, and measured the OLT to see how good the locking is. The following rough swept sine plots are the OLTs for MICH and PRCL. The gain setting was -10 and 0.5 for MICH and PRCL, respectively. Integrators were off. Looking at the measured plots, MICH has about 300 Hz UGF, when the gain is -20, and PRCL has about 300 HZ UGF, too, when the gain is 0.8.


As these lokings seemed good, so we tried the LSC matrix code written by Anamaria. However it is not working well at this point. When the script add excitations to the exc channels, they kick the optics too much and the lockings are too much disturbed...
Also, we have been trying to lock PRC with the SB resonant, it doesn't work. Looking at the simulated REFL11I (PRCL) signal (you can see it in #5495 too), the CR and SB resonances have the opposite signs... But minus gain never works for PRCL. It only excites the mirror rather than locking. |
5513
|
Thu Sep 22 04:49:14 2011 |
Anamaria | Update | LSC | Locking status update - Some Scripts, No Louck |
The scripts I wrote can be found in /users/anamaria/scripts/sensemat/
]There are two of them:
- one that sets all the switches, gains, frequencies, etc, then cycles through the various RFPDs I and Q into the LOCKIN signal, so as to see the sensing matrix.
- the second one is a matlab script that takes the crappy file tdsavg outputs and makes it into a cute mag/phase matrix.
They're quite primitive at this point, I've forgotten a lot of tcsh... may improve later. But could be useful later to someone else at least.
I don't think it's particularly the fault of the script that we can't measure the sensing matrix. We can slam on the excitation by hand, and it holds for a little while. I set a wait time for lock to adjust, and most times it just oscillates a bit for a few seconds. Also, the script turns on the excitation and it's done, the rest is just measurement, then turns it off at the end. So during the script, there's not much to deal with, except keeping the lowpass filters quiet when switching the signal to demod; but that doesn't go anywhere, so it definitely doesn't disturb the ifo. Turns out pressing the RSET clear history button needs a 2 to make it happen.
I think I might prefer to set the excitation to run, and then do the old retrieve-data-later-nds-matlab thing. I do not trust these measurements without coherence and a bit of variance study, given instabilities.
Point is... Even on carrier, the PRC lock is not stable by any means. Can barely turn on low freq boosts, every other lock. Until we fix the lock stability issue, there's not much to measure I guess.
Unfortunately, I don't know how to make that happen. Before we leave on Friday we could do a few sanity checks such as measuring the noise of the RFPDs vs ADC+whitening, which I may have said I would do; and perhaps setting up a couple OSAs, one on REFL, one on AS, to make sure we know what the sidebands are doing. Both of which Rana suggested at some point.
(There used to be a quote here from Keiko here but I got mad when it reformated my entire log to be one cluster- hence the look) |
5514
|
Thu Sep 22 10:43:50 2011 |
Paul | Update | SUS | Power spectrum with different filter gains |
I thought it might be informative before trying to optimise the filter design to see how the current one performs with different gain settings. I've plotted the power spectra for ITMY yaw with filter gains of 0, 1, 2, 3 and 4.
All of the higher gains seem to perform better than the 0 gain, so can I deduce from this that so far the oplev control loop isn't adding excess noise at these frequencies? |
Attachment 1: ITMY_YAW_closed_vs_open_noise.pdf
|
|
5515
|
Thu Sep 22 11:49:05 2011 |
kiwamu | Update | LSC | some LSC scripts don't run on pianosa |
Found some LSC scripts didn't run on pianosa. Particularly all the scripts on the C1:IFO_CONFIGURE screen don't run.
They need to be fixed. |
5516
|
Thu Sep 22 11:50:37 2011 |
Koji | Update | LSC | Locking status update |
Both loops basically have no phase margins. i.e. unstable. How can you lock PRMI with these servos?
Quote: |
The following rough swept sine plots are the OLTs for MICH and PRCL. The gain setting was -10 and 0.5 for MICH and PRCL, respectively. Integrators were off. Looking at the measured plots, MICH has about 300 Hz UGF, when the gain is -20, and PRCL has about 300 HZ UGF, too, when the gain is 0.8.
|
|
5517
|
Thu Sep 22 13:45:17 2011 |
Paul | Update | SUS | ETMX actuator response fits |
Fitting results:
Pitch
-- Fit completed after 305 iterations--
Started with: Gain = 3e-05,
Q factor = 5,
Pole frequency = 0.6776,
Fit results: Gain = 1.85497e-06,
Q factor = 23.7233,
Pole frequency = 0.956686
Residual (normalised against the sum of input datapoints) = 0.0202483
Yaw
-- Fit completed after 334 iterations--
Started with: Gain = 3e-05,
Q factor = 5,
Pole frequency = 0.6776,
Fit results: Gain = 2.518e-06,
Q factor = 7.21618,
Pole frequency = 0.853559
Residual (normalised against the sum of input datapoints) = 0.0570132 |
Attachment 1: ETMX_PITCH_actuator_response_complex_fit.png
|
|
Attachment 2: ETMX_YAW_actuator_response_complex_fit.png
|
|
5518
|
Thu Sep 22 13:56:56 2011 |
kiwamu | Update | ASC | C1ASS : status update |
The output matrix in the C1ASS servo were coarsely readjusted and the servos seemed working.
However it is difficult to say the servo is very good or so-so,
because the ETMY suspension moves a lot and hence the cavity eigen axis moves a lot too. 
(to do)
+ optimization of the ETMY oplevs and OSEM damping.
+ evaluation of the performance of the C1ASS with a good damping.
(Background)
Since we have installed the new mid-HV amplifier for the PZT1 mirror ( #5450) it changed the response of the PZT1 (gains from EPICS to the actual angles).
Therefore the C1ASS output matrix needs be adapted to the new PZT1 response.
(What I have done)
What I was measuring was a coupling from each PZT mirror to both beam angle and beam position by looking at the output from the LOCKINs.
So this measurement eventually gives us a nicely diagonalized output matrix by inverting the coupling.
However the measurement turned out to be difficult because the ETMY moved too much.
In fact the cavity eigen axis also moves and the fluctuation was larger than the intentionally introduced beam angle/translation offsets, which are for the coupling measurement.
Instead of measuring the couplings, I put some numbers into the matrix based on a guess.
Since the PZT1 HV amp became weaker than that of PZT2, the elements in the output matrix should be amplified by some number.
Right now the PZT1 amp can drive the mirror in a range of -5 -30 V with EPICS range of +/-10 counts, and for PZT2 it is about 0 -150V with EPICS range of +/-5 counts.
So the difference of the responses in unit of V/counts is about 8.5.
The PZT1 elements in the matrix were multiplied by this number and I became able to close the servos.
|
+ Modification of C1ASS (Kiwamu)
|
|
5519
|
Thu Sep 22 15:53:37 2011 |
Mirko | Update | LSC | RF modulation depth measurement again |
Toyed around with the modematching some more today.
The outermost glass elements of the OSA are about 28cm apart.
With the OSA beeing a confocal cavity that should mean that the ROC of every mirror is 28cm on the cavity side. If the input surface is flat we need a 28cm focusing lens for good MM. If it's not we shouldn't need any MM.
Tried a f=250mm lens on different positions first. Got at best about 570mV (PD gain=10) in transmission of the OSA.
Then tried a f=1000mm lens. Best transmission 1.22V (7.2% transmission). SB were (PD gain =100) 11MHz: 87.2mV (m=0.17) , 55MHz: 59.2mV (m=0.14)
Lost the position while toying around. Left it then at 1.0V transmisison at 15:15 local time. Let's see how much it drifts. SBs for this were 11MHz: 52.8mV (m=0.17), 55MHz: 73.8mV (m=0.14)
[Ed by KA: If the carrier transmission was really 1.22V and 1.0V the modulation depths calculated are inconsistent with the measurement.]
Spacing between carrier 11MHz and 55MHz SBs seems consistent, and leads to a FSR measurement of 1.5GHz, also fine.
Update: After 90mins no change in carrier transmitted power. Next morning: Carrier transmission down by 10%.



|
5520
|
Thu Sep 22 17:29:42 2011 |
Keiko | Update | IOO | AM modulation mistery |
AM modulation will add offset on SRCL signal as well as PRCL signal. About 2% of the signal amplitude with the current AM level. MICH will not be affected very much.
From #5504, as for the AM modulation I checked the MICH and SRCL signals in addition to the last post for PRCL, to see the AM modulation effect on those signals. On the last post, PRCL (REFL11I) was found to have 0.002 while the maximum signal amplitude is 0.15 we use . Here, I did the same simulation for MICH and SRCL.
As a result, MICH signals are not affected very much. The AM modulation slightly changes signal slopes, but doesn't add offsets apparently. SRCL is affected more, for REFL signals. All the REFL channels get about 0.0015 offsets while the signal ampliture varies up to 0.002. AS55I (currently used for SRCL) has 1e-7 offset for 6e-6 amplitude signal (in the last figure) - which is the same offset ratio comparing with the amplitude in the PRCL case -
(1) MICH signals at AS port with AM m=0

(2) MICH signals at AS port with AM m=0.003

(3) SRCL signals at AS/REFL port with AM m=0
(3) SRCL signals at AS/REFL port with AM m=0.003


Quote: |
How about changing the x-axis of all these plots into meters or picometers and tell us how wide the PRC resonance is? (something similar to the arm cavity linewidth expression)
Also, there's the question of the relative AM/PM phase. I think you have to try out both I & Q in the sim. I think we expect Q to be the most effected by AM.
|
|
5521
|
Thu Sep 22 17:48:20 2011 |
kiwamu | Update | SUS | bad oplev on ETMY |
It turned out the oplev controls on ETMY were just bad.
It looks like the whitening filters have been OFF and because of that the resultant open-loop was not crossing the unity gain.
I will check the whitening filters.
(open-loop transfer function)
The blue dots are the measured data points and the green curve is the fit.
Apparently the open-loop doesn't go above the unity gain, so the oplev had been doing nothing.
If we try to increase the overall gain it will oscillate because of the phase delay of more than 180 deg around 3 Hz.
The red curve is the expected one with the whitening filters (WFs) properly engaged.
Note that WF are supposed to have two zeros at 1 Hz and two poles at 10 Hz.

Quote from #5518 |
(to do)
+ optimization of the ETMY oplevs and OSEM damping.
|
|
5523
|
Thu Sep 22 20:12:54 2011 |
kiwamu | Update | SUS | ETMY oplev whitening engaged |
The whitening filters for the ETMY oplevs are back. 
The whitening board had been in the rack but the ADC was connected directly to the oplev interface board without going through the whitening board.
In fact the interface board and the whitening board had been already connected. So the ADC was making a shortcut.
I disconnected the ADC from the interface board and plugged it to the output of the whitening board.
Here is an example of the new open-loop transfer function with the whitening filters.

Note :
before the measurement I increased the control gain by an arbitrary number to obtain gain of more than 1 around 1 Hz.
Quote from #5521 |
I will check the whitening filters.
|
|
5524
|
Thu Sep 22 22:53:06 2011 |
Suresh | Update | Computer Scripts / Programs | Activated DAQ channels in C1IOO model and restared fb |
To look at the WFS servo signals I was using test points in the servo filter banks. This is not recommended for regular operation since acquiring the testpoint data at 16k loads the fb. Instead, I ran the daqconfig script from the scripts directory and activated the IN1_DQ, IN2_DQ and OUT_DQ channels in all the six servo filter banks (at 2048 Hz sampling rate) and then restarted the fb. However the c1ioo Sun machine stopped responding after this. Koji and I went in to see what was going on and the machine was not reponding to a keyboard plugged directly into the machine. The screen display showed no reponse to our key press. So we did a hardware reboot with the tiny switch in front of the machine. It came up okay and all the c1ioo models were back in action.
I then checked with the dataviewer to make sure that I can see the trends on the newly activated DQ channels. They were all fine. |
5525
|
Thu Sep 22 22:55:01 2011 |
Anamaria | Update | LSC | POX channel = POY PD connected + Bad Rack |
Keiko, Anamaria
We decided we needed a DC channel to sense the gain in the PRC, so we set to align POY55. It took a while because the beam was very weak, and it comes in upwards, so we used a couple of mirrors to bring to a reasonable flat level, and put it on the PD. Then we went to read the DC out and we got 1.3V stationary! Nonsense. We also realized there is no LO for this PD, or any other 55MHz PD, aside from REFL55. Oh well, we only wanted the DC for now. POY55 is aligned (decently).
Koji told me to try swapping the power cable, so I unplugged it at the rack and plugged it in another power card. And it worked! I then moved the DC out (back of rack) to follow the front, and it turns out POY55 diode is read on the POXDC channel. I plugged and unplugged it in disbelief, but it is what it is. At least we have a readout on the power level in PRC.
I attach a picture of the power cards for the LSC RFPDs, with the 3 I found to be bad, and showing current config. I had to move REFL11 and POY55 from their assigned spot.
The two on the lower left are bad in the sense that they put an offset on the PD and make the DC readout be 1.3V for no reason (when working, for example, POY55 read 60mV). The one on the lower right I had trouble with some time ago, it made the PD not read any voltage at all (when working it would read at least 100mV). Beyond that I have not investigated what is up, since I could find working plugins. |
Attachment 1: RFPDpowerRack2.pdf
|
|
5526
|
Thu Sep 22 23:02:15 2011 |
Suresh | Update | IOO | no light on WFS2. Realigned input onto both WFS |
Rana noticed that the sum on WFS2 was about 10 times smaller than that on WFS1. Though the beam appeared centered on the DC QPD screens it was not really true. When I went and checked the actual beam position it was landing on the metal enclosure of the WFS2 sensor and scattering back on to the diode.
I also checked the power levels of light landing on the sensors It was about 0.25mW in both cases. This needs further investigation since the power split at the beam spitter is like 0.25mW onto WFS1 and 0.45 towards WFS2. The lost 0,20 mW has to be traced and we have to be sure that it is not scattered around on the table.
|
5527
|
Thu Sep 22 23:10:07 2011 |
Suresh | Update | IOO | proposed modifications to the C1IOO model |
Rana advised that we put in a lockin-output matrix which will allow us to excite any combination of MC mirrors so that we can excite pure translations or rotations of the MC beam axis. This would require us to direct a lockin output into all the three mirrors simultaneously with a +1 or -1 as needed in the matrix.. |
5529
|
Fri Sep 23 16:25:01 2011 |
steve | Update | VAC | vac rack UPS battery replaced |
APC Smart -UPS 2200 model: SUA2200RM2U batteries were replaced by compatible RBC43, 8x 12V5A |
Attachment 1: P1080252.JPG
|
|
Attachment 2: P1080254.JPG
|
|
5530
|
Fri Sep 23 16:56:07 2011 |
Mirko | Update | LSC | Desired MC modulation frequency measurement, tuning of modulation frequency |
[ Mirko, Koji, Suresh ]
Looked into the modulation frequency that should pass the input MC. With a locked MC looked at the RF output of the PD in refl of the MC. Looked at the beat between 11MHz and 29.5MHz. Minimizing it by fine-tuning the 11MHz freq. ( which means maximizing the 11MHz transmission).
SB freq. [MHz] Beat power [dBm]
11.065650 -75
11.065770 -80 (diving into spec. analyzer instrument noise)
11.066060 -80 (surfacing out of spec. analyzer instrument noise)
Set the freq. to the middle of the last two points: 11.065910MHz at 16:26.
ToDo: How big a problem is the AM? |
5531
|
Fri Sep 23 17:31:14 2011 |
Katrin | Update | Green Locking | Stray light reduction (Y) |
I inserted several beam blocks and iris on the Y arm Green table. There was/is lots of stray light because a lot of the mirrors are not under an angle of incident of 45°. Some stray light is left since couldn't find an appropriate beam block/dump due to lack of space on the table.
|
5532
|
Fri Sep 23 17:57:34 2011 |
Paul | Update | SUS | Oplev filter optimization for 2 poles and 2 zeros |
I have made a function to optimise the overall gain, pole frequencies and zero frequencies for the oplev filter. The script will optimize any user defined number of poles and zeros in order to minimise the RMS motion below a certain cut off frequency (in this case 20Hz). The overall gain is adjusted so that each trial filter shape always has a UGF of 10 Hz.
I have a attached a plot showing the power spectrum and RMS curves for the optimization result for 2 zeros and 2 poles, optimized to give a minimal RMS below 20Hz.
I have also attached a plot showing the loop gain and the filter transfer function.
The noise spectrum shows that the optimised filter gives a better noise performance below 10Hz, but a servo oscillation at the UGF of 10 Hz means it injects a lot of motion around this frequency. Should I consider some more aggressive way to force the script to keep a decent phase margin?
The fminsearch results show that the 'optimized' solution is two resonant peaks:
-- Optimisation completed after 571 iterations--
Started with:
Pole 1 frequency = 1 Hz
Pole 2 frequency = 2 Hz
Zero 1 frequency = 0.1 Hz
Zero 2 frequency = 5 Hz
Overall gain = 1
Finished with:
Pole 1 frequency = 0.0497181 Hz
Pole 2 frequency = 2.01809 Hz
Zero 1 frequency = 0.0497181 Hz
Zero 2 frequency = 2.01809 Hz
Overall gain = 71970.1
Initial RMS below 10 Hz = 5.90134e-06
Remaining RMS below 10 Hz = 8.42898e-07
|
Attachment 1: optimised2p2z_v1.png
|
|
Attachment 2: optimised2p2z_v1_TFs.png
|
|
5533
|
Fri Sep 23 18:00:54 2011 |
Suresh | Update | IOO | PSL beam realigned to MC |
I noticed that the beam centering on the WFS had changed over night and the MC_TRANS_SUM was about 40k counts. When well aligned this SUM is around 50-55k counts. So PSL coupling into MC was suboptimal. It was not clear whether the MC shifted or the PSL beam shifted. So I looked at the PSL ANG and POS QPDs.

The plots above show the gradual drift of the PSL beam in vertical direction during the last 8hrs or so. But the last bit shows the adjustments I had to make to reobtain optimal alignment. And these adjustments are not undoing the drift! This would indicate that the MC axis has also shifted during the same time period. |
5534
|
Sat Sep 24 01:21:11 2011 |
kiwamu | Update | SUS | damping test |
As a suspension test I am leaving all of the suspensions restored and damped with OSEMS but without oplevs |
5535
|
Sat Sep 24 01:38:14 2011 |
kiwamu | Update | CDS | c1scx and c1x01 restarted |
[Koji / Kiwamu]
The c1scx and c1x01 realtime processes became frozen. We restarted them around 1:30 by sshing and running the kill/start scripts. |
5536
|
Sat Sep 24 01:51:02 2011 |
rana | Update | SUS | Oplev filter optimization for 2 poles and 2 zeros |
Quote: |
I have made a function to optimise the overall gain, pole frequencies and zero frequencies for the oplev filter. The script will optimize any user defined number of poles and zeros in order to minimise the RMS motion below a certain cut off frequency (in this case 20Hz). The overall gain is adjusted so that each trial filter shape always has a UGF of 10 Hz.
|
I think this is a nice start. Its clear that we don't want to use this feedback law, but the technique can be tweaked to do what we want by just tweaking our cost function.
Let's move the scripts into the SUS/ scripts area and then start putting in weights that do what we want:
1) Limit the gain peaking at the upper UGF to 6 dB.
2) Minimum phase margin of 45 deg.
3) Minimum gain margin of 10 dB.
4) Lower UGF = 0.1 Hz / Upper UGF = 10 Hz.
5) Assume a A2L coupling of 0.003 m/rad and constrain the injected noise at the test mass to be less than the seismic + thermal level.
6) Looser noise contraint above 50 Hz for the non TM loops. |
5537
|
Sat Sep 24 02:09:43 2011 |
kiwamu | Update | SUS | Re:Oplev filter optimization for 2 poles and 2 zeros |
Good work for the oplev noise simulations. Here are some comments/questions:
(A) The noise looks suppressed but the open-loop transfer function doesn't look so good, because it doesn't have sufficient phase margins at the UGFs (0.01 and 10 Hz).
I guess it is better to have a phase margin detector in your code so that the code automatically rejects a bad phase margin case.
Actually since the number of data points are finite, the rms detector in the simulation can not always find a sharp loop oscillation.
(B) The resultant poles and zeros seem canceling each other but the filter still has a structure. Is something wrong ?
Quote from #5332 |
Pole 1 frequency = 0.0497181 Hz
Pole 2 frequency = 2.01809 Hz
Zero 1 frequency = 0.0497181 Hz
Zero 2 frequency = 2.01809 Hz
|
|
5538
|
Sat Sep 24 09:55:42 2011 |
Keiko | Update | IOO | AM modulation mistery |
From the night day before yesterday (Sep 22nd, Thursday night. Sorry for my late update), there are more AM modulations than I measured in the previous post. It is changing a lot, indeed! Looking at the REFL11 I and Q signals on the dataviewer, the signal offset were huge, even after "LSCoffset" script. Probably the modulation index of AM was same order of PM at that time. The level of AM mod index is changing a lot depending on the EOM alingment which is not very stable, and also on the environment such as temperature .
To reduce AM modulations, here I note some suggestions you may want to try :
* Change the SAM connectors between RF resonator and EOM to be a soft but short connector, so that the resonator box doesn't hung from the EOM.
* Change the RF resonator base to be stable posts. Now several black plates are piled to make one base.
* Install a temperature shield
* Also probably you want to change the BNC connector on the RF resonator to be SMA.
* Be careful of the EOM yaw alignment. Pitch seemed to be less sensitive in producing AM than yaw alignment.
Quote: |
AM modulation will add offset on SRCL signal as well as PRCL signal. About 2% of the signal amplitude with the current AM level. MICH will not be affected very much.
From #5504, as for the AM modulation I checked the MICH and SRCL signals in addition to the last post for PRCL, to see the AM modulation effect on those signals. On the last post, PRCL (REFL11I) was found to have 0.002 while the maximum signal amplitude is 0.15 we use . Here, I did the same simulation for MICH and SRCL.
As a result, MICH signals are not affected very much. The AM modulation slightly changes signal slopes, but doesn't add offsets apparently. SRCL is affected more, for REFL signals. All the REFL channels get about 0.0015 offsets while the signal ampliture varies up to 0.002. AS55I (currently used for SRCL) has 1e-7 offset for 6e-6 amplitude signal (in the last figure) - which is the same offset ratio comparing with the amplitude in the PRCL case -
|
|
5539
|
Sat Sep 24 17:12:54 2011 |
Koji | Update | LSC | Realignment of REFL / Some 3f PRMI locking / Recycling Gain |
[Koji Suresh]
Activity on Friday evening
- The REFL path has been thoroughly aligned
As I did not like the REFL spot misaligned on the REFL CCD, we went to the AP table.
Many optics had the spots not on the middle of the optic, including the PBS whose post was not fixed on the post holder.
We aligned the optical paths, the RF PDs, and the CCD. The alignment of the PD required the use of the IR viewer.
One should not trust the DC output as a reference of the PD alignment as it is not enough sensitive to the clipping.
We aligned the optical paths again after the reasonable alignment of PRM is established with the interferometer.
"Next time when you see REFL spot is not at the center of the camera, think what is moved!"
- The REFL165 PD is disconnected from the power supply
I found that the REFL165 PD is producing 7.5V output at the DC monitor no matter how the beam is blocked.
As I could not recover this issue by swapping the power connector at the LSC rack, I disconnected the cable
at the RFL165 PD side. I need to go through the PD power supply circuit next week.
- PRMI alignment policy of the night
Kiwamu has aligned Y-arm some time ago (Thursday evening?). I decided not to touch ITMY.
So the Michelson is aligned by ITMX, PRC is aligned by PRM.
- Michelson locking
The short Michelson was locked with AS55Q and the MICH filter. We could use the gain of +/-20 for locking,
and could increase it up to ~+/-250. At the max gain, the all three integrators and the two resonant gains
could be activated. The sign depends on which fringe you want at the AS port (bright or dark).
In this condition, the output of the POXDC channel (which is actually the POY DC out -- c.f. This entry)
is used to determine the internal power. It was ~70cnt.
- PRMI locking
Then the PRMI was locked. There was some confusion of the gains because of the limitters at the servo filters
(which yielded the locking with 1bit outputs no matter how much the gains were....)
After all, I decided to use REFL33I for the PRCL for the test. The PRCL gain was -0.3~-1.0 for the carrier lock, but
was highly dependent on the alignment. i.e. if accidentally hit the high power recycling gain, it oscillated easily
and the lock was lost. Probably this was the first 3f locking at the 40m in the current optical config, if
Kiwamu did not do that secretly. The SB lock was also obtained by flipping the sign of the PRCL servo.
The difficulty we had was the instability when the recycling gain became big. We were monitoring the POXDC
(i.e.DCout of the POY PD). When this exceeds 5000, many glitches appears in the LSC signals and disturbs the lock.
This was not the fringes from neither the arms nor the SRC.
The observed POY DC with the carrier resonant PRMI was 5000~8000vcnt (momentary). |
5540
|
Sat Sep 24 17:45:56 2011 |
Paul | Update | SUS | Re:Oplev filter optimization for 2 poles and 2 zeros |
Quote: |
(B) The resultant poles and zeros seem canceling each other but the filter still has a structure. Is something wrong ?
Quote from #5332 |
Pole 1 frequency = 0.0497181 Hz
Pole 2 frequency = 2.01809 Hz
Zero 1 frequency = 0.0497181 Hz
Zero 2 frequency = 2.01809 Hz
|
|
Ah yes, well noticed. I think I have tracked this down to just a bug in printing of fitting results: It was just printing the pole results for the zeros too. The results for the same fit now read:
Finished with:
Pole 1 frequency = 0.0497181 Hz
Pole 2 frequency = 2.01809 Hz
Zero 1 frequency = 0.0972455 Hz
Zero 2 frequency = 6.50126 Hz
Overall gain = 71970.1
EDIT: sorry, I forgot that when you write a reply, the author is still by default the person you are replying to unless you change it!
|