Entry  Wed Aug 11 02:44:37 2010, Zach, Update, elog, elog restarted 
You'll never guess what happened: the elog crashed!
I restarted it with the script. Yay.
Entry  Wed Aug 4 01:44:56 2010, Koji, Configuration, PEM, Accelerometer located on and below the PSL table IMG_2675.jpgIMG_2678.jpgIMG_2682.jpg
[Jenne and Koji] 
In order to characterize the seismic vibration of the PSL table, we put the accelerometers on
and below the PSL table.
    Reply  Tue Aug 10 22:40:55 2010, Koji, Summary, PEM, Accelerometer located on and below the PSL table PEM_100809.pdf
Result of the accelerometer measurement
Introduction

We wanted to characterize the PSL table before the work before its lifting up.
We put a set of three-axis Wilcoxon accelerometers on the ground and another set on the PSL table through the weekend.
Entry  Tue Aug 10 17:21:48 2010, steve, Update, General, PSL optical table is at 30" height P1060586.JPGP1060583.JPGP1060591.JPGP1060593.JPGP1060597.JPG
Mike Gerfen, Koji and Steve,
 
Our old ( 6 x 10 x 1 ) ft Newport optical table will have the same height as the neighboring AP table when this job is finished.
Entry  Tue Aug 10 16:55:38 2010, Jenna, Update, Electronics, c1iovme restarted 
 Alastair and I restarted the c1iovme around the time of my last elog entry (~3:20).
Entry  Tue Aug 10 15:23:35 2010, Jenna, Update, Electronics, Rubidium clock time constant 
[Jenna & Alastair]
We changed the locking time constant on one of the Rubidium clocks using the RbMon software that came with it. We had to use the ancient Dell laptop
latitudeD810 because it has a serial port built in, and we couldn't get the usb->serial adapter to work right with the clock. We tried the usb connector
Entry  Tue Aug 10 05:56:07 2010, Koji, Update, PSL, PSL Table Lifting Preparation IMG_2686.jpgIMG_2691.jpgIMG_2692.jpgIMG_2711.jpg
Work on Aug 9th
Steve, Jenne, Koji, Alberto, Aidan, Jan, Sharmila, Katherine
From
9am to 6pm

    Shutting down the PSL after the 90885
hours of service
    Removals
    
        The accelerometers
Entry  Tue Aug 10 01:05:17 2010, Zach, Update, elog, elog restarted 
 Elog was down. Restarted with the script.
Entry  Mon Aug 9 13:32:02 2010, nancy, Update, IOO, Mode Cleaner ASC 6x
 E-log entry for Friday - will attach more plots to this entry on
wednesday after i am back   to the 40.
 
Started working at some 1030 hrs and recording the Open Loop Tfs for all 6 loops.
The control was not so good,
    Reply  Mon Aug 9 15:54:43 2010, Koji, Update, IOO, Mode Cleaner ASC 
The WFS and QPD servos were working. That was great.
Everything was fine except for the time series plots.
I could not get what story you are telling with the time series.
       Reply  Mon Aug 9 21:50:50 2010, nancy, Update, IOO, Mode Cleaner ASC 


    
        
            Quote:
        
        
            
            The WFS and QPD servos were working. That was great.
Entry  Wed Aug 4 11:10:28 2010, nancy, Update, IOO, Mode Cleaner WFS 
Yesterday, I started twiddling with the Mode Cleaner at about 2 pm.
So the seismic data should be all good before that.
I was using it till about 3.30 am, and then left for the night with locking it and swithcing
    Reply  Wed Aug 4 12:49:42 2010, nancy, Update, IOO, Mode Cleaner WFS 6x


    
        
            Quote:
        
        
            
            Yesterday, I started twiddling with the Mode Cleaner at about 2 pm.
       Reply  Thu Aug 5 01:29:39 2010, nancy, Update, IOO, Mode Cleaner WFS 
 
I calculated the MC1&3 Vs WFS1&2 Output Matrix today from the above measurements with koji's help.
the matrix can be generated from the m file at /cvs/cds/caltech/users/nancy/Align_Matrix/matrix.m
          Reply  Thu Aug 5 11:48:52 2010, nancy, Update, IOO, Mode Cleaner WFS 


    
        
            Quote:
        
        
            
             
             Reply  Thu Aug 5 13:05:53 2010, Koji, Update, IOO, Mode Cleaner WFS 
Upon Nancy's request, I checked the status of the suspensions.
I found that the power strip of the 1Y4 rack was turned off.
Since it has a over current breaker, I don't know whether it happened by someone or over current.
                Reply  Fri Aug 6 10:44:55 2010, nancy, koji, Update, IOO, Mode Cleaner WFS 


    
        
            Quote:
        
        
            
            Upon Nancy's request, I checked the status of the suspensions.
                   Reply  Fri Aug 6 12:33:10 2010, Koji, Update, IOO, Mode Cleaner WFS 


    
        
            
             Yesterday , I put in the Output Matrix, and changed the gain sliders for the 4 WFS loops.
            
        
    

                      Reply  Mon Aug 9 12:46:24 2010, Nancy, Update, IOO, Mode Cleaner WFS 

Entry  Sat Aug 7 21:57:56 2010, rana, Summary, DMF, seisBLRMS restarted Untitled.png
The green xterm on op540m which is running the seisBLRMS DMF got stuck somehow ~3 days ago and lost its NDS connection. I closed the matlab session and
restarted it. Seismic trends are now back online.
Entry  Sat Aug 7 21:09:59 2010, Dmass, Configuration, Computer Scripts / Programs, eLog changes 
I made some changes to the elog on nodus:

    Made a backup of /cvs/cds/caltech/elog/elog-2.7.5/elogd.cfg called elogd.cfg.bk.20100407 in the same directory
    Added a
folder: /cvs/cds/caltech/elog/elog-2.7.5/logbooks/EAGER_Lab
    Restarted the elog daemon via the start-elog-nodus script in the
Entry  Sat Aug 7 11:07:44 2010, Koji, Configuration, PSL, MC kept locked / PMC control gain reduced to +13dB 
Jenne asked us to keep th MC locked and let the seismometers happy through this weekend.
Note that the work at the control room and the desks are no problem as far as you are quiet.
Nancy told Jenne and me that she finished the work and reverted the WFS to the old state at 4:30AM.
Entry  Sat Aug 7 10:47:38 2010, Koji, Summary, elog, elog restarted / source of the trouble eliminated 
Nancy notified me that the elog crashed. It was fixed.

I restarted elog, but it kept crashing. Some of the entries on Aug 6th caused the trouble.
I tried to refresh the pictures in entries 3376, 3377, 3378. Still it kept crashing.
Entry  Thu Aug 5 00:17:41 2010, Koji, Update, PSL, PSL preparation work 
We start the work on the cables at around the PSL table.

Aug 5th 10am-4pm?: (Kiwamu, Alberto, Koji)
    Reply  Thu Aug 5 22:36:11 2010, Koji, Update, PSL, PSL preparation work 
PSL preparation work

Aug 5th 10am-4pm?: (Kiwamu, Alberto, Koji)
       Reply  Fri Aug 6 08:09:15 2010, steve, Update, PSL, PSL preparation work 


    
        
            Quote:
        
        
            
            PSL preparation work
       Reply  Fri Aug 6 20:00:03 2010, Koji, Update, PSL, PSL preparation work PSL_cable_map.pdf
PSL preparation work report
Aug 6th 10am-5pm: (Steve, Jenne, Alberto, Kiwamu, Koji)
- We labeled the cables to be disconnected

    These will be disconnected in order to isolate the PSL table and the frame (housing) from the other part of the lab.
    
Entry  Fri Aug 6 15:50:29 2010, Gopal, Update, Optic Stacks, (Much Better Looking) Displacement-Displacment Transfer Functions MC1_MC3_XTrans.pngMC1_MC3_YTrans.pngMC1_MC3_ZTrans.png
I reran the FDA in COMSOL on the MC1/MC3 Stack and produced the following Displacement-Displacement Transfer Functions:

X-Translational
Drive has a blue background
    Reply  Fri Aug 6 19:46:59 2010, Gopal, Update, Optic Stacks, (Much Better Looking) Displacement-Displacment Transfer Functions 


    
        
            Quote:
        
        
            
            I reran the FDA in COMSOL on the MC1/MC3 Stack and produced the following Displacement-Displacement
Entry  Sat Jul 24 13:13:41 2010, rana, steve, alberto, Update, General, Initial Crane Inspection reveals flaws: wiring and oil 
The guy from KroneCrane (sp?) came today and started the crane inspection on the X End Crane. There were issues with our crane so he's going to resume
on Monday. We turned off the MOPA fur the duration of the inspection.

    None of our cranes have oil in the gearbox and it seems that they never did since they have never been maintained. Sloppy installation
    Reply  Mon Jul 26 10:04:24 2010, steve, Update, General, Initial Crane Inspection reveals flaws: wiring, oil and imbalance 2500.JPGsedrum.JPG


    
        
            Quote:
        
        
            
            The guy from KroneCrane (sp?) came today and started the crane inspection on the X End Crane. There were issues with
       Reply  Wed Jul 28 16:42:21 2010, steve, Update, General, Crane gear box cheched P1060503.JPGP1060511.JPGP1060515.JPGP1060517.JPG


    
        
            Quote:
        
        
            
            
            
    
          Reply  Fri Aug 6 17:47:36 2010, steve, Summary, General, cranes load tested at 1998 lbs 0.9ton.PDFP1060523.JPGP1060532.JPGP1060541.JPG


    
        
            Quote:
        
        
            
            
            
  
Entry  Thu Aug 5 16:18:37 2010, Alastair, Update, General, clocks still not locked fully - hut removed 
The rubidium clocks are still not quite locked together, though it is clear that the beat frequency has dropped a lot since yesterday.
I checked on the clocks and the 1pps sync light is on.  The clocks are really hot again though despite the gap left at the bottom of the igloo. 
The side of the clocks were hot enough to not be touchable.
    Reply  Fri Aug 6 16:20:08 2010, rana, Update, General, clocks still not locked fully - phase shifted Untitled.pdf
 I tried to get the clocks to be closer to 90 deg for the relative phase by adding some cable length to one of the lines, but they are still wandering
too much. We need to use the serial interface and up the gain in their 1PPS locking loops.
Entry  Fri Aug 6 12:22:04 2010, Jenne, Update, IOO, MC data taking over the weekend 
Nancy has the Mode Cleaner for her work for the night, and is going to leave the MC happy, locked, autolocker on, WFS enabled, the works, and write down
in the elog the time that she's finished. After that, I'm taking MC/seismic data all weekend long.  During the weekend, if at all possible, please don't go into the IFO room, especially near the Mode Cleaner.  If you do need to go into
    Reply  Fri Aug 6 12:44:29 2010, Koji, Update, IOO, MC data taking over the weekend 
Question:
Do you like to keep the WFS turned on? 
Entry  Thu Aug 5 17:59:23 2010, Koji, Update, PSL, Cable removal from the control room IMG_2684.jpg
[Alberto, Kiwamu, and Koji]
We removed the BNC cables from the control room. 
The work was as hard as the one I had when I swept a 300m tunnel with a vacuum...
Entry  Wed Aug 4 20:58:22 2010, Gopal, Update, WIKI-40M Update, 7.28.10 - 8.4.10 Weekly Update 
Summary of this week's activities:
7/28:    Finished Y-Translational 4-Stack Analysis
"Tapered Cantilever" COMSOL tutorial
Entry  Wed Aug 4 20:15:07 2010, Jenne, Update, PEM, Yay! Guralps work again! 
After much hassle, the Guralp cable from the ADC Out of the breakout box to the ADC is fixed, and everything is plugged in and working again.  The seismometers are back in their regular positions at the ends of
the MC, ready for some excellent seis/MC combo data. 
Entry  Wed Aug 4 19:50:58 2010, rana, Configuration, Electronics, Rubidium clocks too hot: hut removed beat.png
Alastair found that the foam hut that he and Jan put on top of the Rb clocks to temperature stabilize them was too good of an insulator. The Rb boxes
had gotten very hot and became internally unlocked as seen on the front panel.
After we let them cool down with the box off, I turned them back on. After several minutes the 'Locked' light came back on. Some minutes after
Entry  Tue Aug 3 11:50:16 2010, Jenne, Update, PEM, Seismometer Problem Tracked down 
After some cable swapping this morning, I have determined which cable is bad.  It's the Gur1 cable between the seismometer and the breakout
box.  This is a milspec -> 37pin d-sub cable.  I'll pull out the cable and have a look at it after lunch.
    Reply  Wed Aug 4 03:01:56 2010, Jenne, Update, PEM, Catastrophic Cable Failures 


    
        
            Quote:
        
        
            
            After some cable swapping this morning, I have determined which cable is bad.  It's the Gur1 cable between
Entry  Sat Jul 31 17:37:36 2010, josephb, Update, CDS, Cables needed for CDS test 
Last Thursday, Kiwamu and I went through the cabling necessary for a full damping test of the vertex optics controled by the sus subsytem, i.e. BS, ITMX,
ITMY, PRM, SRM.  The sus IO chassis is sitting in the middle of the 1X4 rack.  The c1sus computer is the top 1U computer in that rack.
ADCs:
The hardest part is placing the 2x D-sub connectors to scsi on the lemo break out boxes connected to the 110Bs.  The breakout
    Reply  Tue Aug 3 11:17:10 2010, kiwamu, Update, CDS, Diagrams for Cables needed for CDS test sus_wiring.pngsus_new_wiring.pngsus_missing_stuff.png
Current Wiring Setup for the Suspension Controls
Entry  Tue Aug 3 03:15:06 2010, nancy, Update, IOO, MC back to locked mode 
I turned the WFS gain to 0.02 back, and the MC is locked, the data for the seismic motion might be meaningful nowforth.
Entry  Mon Aug 2 18:02:46 2010, Jenne, Update, PEM, Seismometer Update: Still not good, but perhpas getting better 
Today's seismometer diagnosis activities are still underway, this is just an update (since I did some reboots):
Problem 1: X and Z channels on both seismometers were flipped.  I unplugged an X cable (East/West on the cable labels) and the Z channel (vert)
would go to floating ADC zero.  Rana suggested that the ADCs sometimes have random channel hopping, and that a reboot of the c0dcu1 computer which
    Reply  Tue Aug 3 01:51:01 2010, Jenne, Update, PEM, Seismometer Update: Still not good, but perhaps getting better 


    
        
            Quote:
        
        
            
            Today's seismometer diagnosis activities are still underway, this is just an update (since I did some reboots):
Entry  Mon Aug 2 21:52:57 2010, Koji, Update, IOO, MC is running at the full power 
[Nancy and Koji]
We restored the full power operation of the MC.
Restoration of the suspensions
Entry  Mon Aug 2 17:12:28 2010, Koji, Update, General, Lab schedule for the week of Aug. 2 PSL_work_schedule.pdf
Aug
2 Mon - 5 Thu WFS work (Nancy)
Entry  Fri Jul 30 09:55:34 2010, steve, Configuration, VAC, slow pumpdown has started slowpd5hr.jpg
Bob and Steve closed BS chamber with the help of the manual Genie lift and the pump down started. The PSL shutter was closed and manual block was placed
in the beam path. High voltage power supplies were checked to be off.
Pumping speed ~ 1 Torr/min was achieved at  1/8 of a turn opened roughing valve RV1
    Reply  Fri Jul 30 17:54:55 2010, steve, HowTo, VAC, How to stop and start slow pumpdown 


    
        
            Quote:
        
        
            
            Bob and Steve closed BS chamber with the help of the manual Genie lift and the pump down started. The PSL shutter
       Reply  Fri Jul 30 22:07:42 2010, Koji, HowTo, VAC, How to stop and start slow pumpdown 
Kiwamu and Koji
The rough pumping was stopped at 230mmTorr. We are going to resume the pumping tomorrow morning.

    RV1 was closed with torque wheel.
    V3 was closed from MEDM
    RP3 wss stopped from MEDM
    The quick
          Reply  Sat Jul 31 10:12:05 2010, kiwamu, Update, VAC, Re: How to stop and start slow pumpdown 
I resumed the pumping down. It started from 9:55 am.
             Reply  Sat Jul 31 14:59:33 2010, kiwamu, Update, VAC, Re: How to stop and start slow pumpdown 
I stopped the pumping at 14:50 pm because I was going back home. I did the same procedure as Koji
wrote down (see here).
The P1 pressure reached 32 Torr.
    Reply  Mon Aug 2 10:26:15 2010, steve, Configuration, VAC, very slow pumpdown completed veryslowpd.jpg


    
        
            Quote:
        
        
            
            Bob and Steve closed BS chamber with the help of the manual Genie lift and the pump down started. The PSL shutter was
Entry  Sun Aug 1 21:40:27 2010, rana, Summary, PSL, FSS: SLOWDC response Untitled.png
I bet you thought that the NPRO slow actuator response could be well represented by a pole at ~0.1 Hz? Well, that's just what they want you to believe.
I attach the response measured in FSS-FAST (with no feedback to the SLOW actuator) when the SLOW is given a step. As you may remember from
kindergarten, the response of a single pole low pass should just be an exponential. Clearly, there's more here than 1 pole.
Entry  Sun Aug 1 21:04:45 2010, rana, Summary, Computer Scripts / Programs, MC Autolocker fixed 
Someone had left a typo in the MC autolocker script recently while trying to set the lock threshold to 0.09. As a result, the autlocker wouldn't run.
I repaired it, made a few readability improvements, and checked in the new version to the SVN. If you make script changes, check them in. If
you think its too minor of a change for a SVN checkin, don't do it at all.
Entry  Sat Jul 31 22:35:01 2010, Koji, Update, VAC, Vac-P1 still 1.2mtorr 
I resumed the pumping from 19:00.
Now the valve RV1 is full open. But the pumping is really slow as we are using only one RP.
After 3hrs of pumping, P1 reached 1.2mmtorr but still we need 2hrs of pumping...
    Reply  Sun Aug 1 18:50:35 2010, Koji, Update, VAC, P1 = 0.48 torr 
I resumed the pumping today. Now the pressure is 0.48 torr. The pump was stopped.


    
        
            Quote:
        
        
            
Entry  Sat Jul 31 04:03:11 2010, Gopal, Summary, Optic Stacks, Complete Displacement Translational Transfer Function Matrix Progress_Report_2.png
Over the past 36 hours, I've run full-fledged FDAs on KALLO.
The transfer functions for translational drives and responses are neatly described by the attached transfer
function "matrix."
Entry  Fri Jul 30 17:52:12 2010, Koji, Configuration, General, Jacket 
Jenne and I need our jackets. We removed them from the Rb clock.
Thanks for making them warm. Probably a Scotland sweater would fit.
    Reply  Fri Jul 30 19:20:15 2010, Alastair, Configuration, General, Jacket 


    
        
            Quote:
        
        
            
            Jenne and I need our jackets. We removed them from the Rb clock.
Entry  Fri Jul 30 17:24:22 2010, Jenne, Update, PEM, Guralps back in the 40m 
I brought the GUR2 seismometer back from Bridge so I can get some more MC/Seismic data during the next week while we're pumped down, before we start
doing things to the PSL table.  Both of the Gur Seismometers are connected back up to the breakout box as of ~3:27pm today.  Alastair still has
the handheld controller thing (which I use for mass centering, on occasion), since he'll want the seismometer back in a week or two when I'm done with
Entry  Thu Jul 29 12:31:09 2010, Koji, Summary, General, Lab Schedule PSL_work_schedule.pdf
July 
29 Thu BS chamber work: Move cable towers / green steering mirrors / (2 TTs with TT charactrization) / Put the heavy door by 5PM.
30 Fri Pumping down
    Reply  Fri Jul 30 14:53:09 2010, steve, Update, General, PSL table leg upgrade psllegs.PDF


    
        
            Quote:
        
        
            
            July 
Entry  Fri Jul 30 12:46:38 2010, Koji, Configuration, 40m Upgrading, BS chamber status BS_layout_Jul_29_2010.png
 
Entry  Wed Apr 21 10:09:23 2010, kiwamu, Update, Green Locking, waist positon of Gaussian beam in PPKTP crystals focal_positin_edit.png
Theoretically the waist position of a Gaussian beam (1064) in our PPKTP crystal differs by ~6.7 mm from that of the incident Gaussian beam.
So far I have neglected such position change of the beam waist in optical layouts because it is tiny compared with the entire optical path.
But from the point of view of practical experiments, it is better to think about it.
    Reply  Tue Apr 27 14:18:53 2010, kiwamu, Update, Green Locking, waist positon of Gaussian beam in PPKTP crystals mode_in_PPKTP.png
The mode profile of Gaussian beams in our PPKTP crystals was calculated.
I confirmed that the Rayleigh range of the incoming beam (1064 nm) and that of the outgoing beam (532 nm)
is the same.
       Reply  Thu Jul 29 21:13:39 2010, Dmass, Update, Green Locking, waist positon of Gaussian beam in PPKTP crystals 


    
        
            Quote:
        
        
            
            The mode profile of Gaussian beams in our PPKTP crystals was calculated.
          Reply  Thu Jul 29 22:58:25 2010, kiwamu, Update, Green Locking, waist positon of Gaussian beam in PPKTP crystals 
- As you said, I just calculated the waist position in the crystal because the speed of light changes in a medium and eventually the waist position also
changes.
- Yes, I did. Once you get a beam with the right waist size, you just put your crystal at the waist position with the offset.
             Reply  Fri Jul 30 00:02:15 2010, Dmass, Update, Green Locking, waist positon of Gaussian beam in PPKTP crystals 


    
        
            Quote:
        
        
            
            - As you said, I just calculated the waist position in the crystal because the speed of light changes in a medium and
                Reply  Fri Jul 30 09:51:58 2010, kiwamu, Update, Green Locking, Re: waist positon of Gaussian beam in PPKTP crystals efficiency_waist_edit.pngPPKTPmode.pngPSL_doubling.pngmode_in_PPKTP.png
Okay, I guess I understand what you want to know. I did the following steps.
1. calculated the conversion efficiency as a function of the waist size. I found w~50um was the optimum waist.

Entry  Thu Jul 29 22:08:32 2010, Alberto, Update, SUS, More optics installed on the BS table 
[Koji, Steve, Kiwamu, Alberto]
- This afternoon we installed a few new optics on the BS table: GR_PBS, GRY_SM2, GRY_SM1.
- We pulled up the cables so that we had more freedom to move one of the cable towers farther South.
    Reply  Fri Jul 30 02:54:04 2010, Koji, Update, 40m Upgrading, More optics installed on the BS table 
July 29 Thu [Steve, Alberto, Kiwamu, Koji]
We placed some optics in the BS chamber.
The chambers are ready to be pumped down on Friday once the heavy door is placed.
Entry  Thu Jul 29 20:43:32 2010, Gopal, Summary, Optic Stacks, Modeling Tips and Tilts XTrans_XDisp.pngXTrans_XDisp.pngYTrans_XDisp.pngZTrans_XDisp.png
I have discovered a method of completely characterizing
the 6x6 response of all six types (x-,y-, and z- translational/rotational) of oscillatory disturbances at the base of the stack. 
 
Entry  Thu Jul 29 15:35:21 2010, josephb, kiwamu, Update, CDS, Working out ADC/DAC/BO wiring 
We are currently using the SUS wiring diagram found on Ben Abbott's page (link here) to determine
the ADC/DAC/BO channel numbers for each individual optics inputs and outputs.  Basically it involves tracking the paths back from the Pentek's, XY220,
and IC110Bs to a point where we can identify it as a Coil UL or a PD whitening filter control or whatever it might be.
    Reply  Thu Jul 29 17:12:48 2010, josephb, kiwamu, Update, CDS, Re:Working out ADC/DAC/BO wiring DSC_2302.png
We have installed  4 BO boards, 3 DAC boards and 1 ADC board for new C1SUS.
They are on
the 1Y5 rack. 
 
 
Entry  Thu Jul 29 17:11:16 2010, Gopal, Update, COMSOL Tips, Including Gravity in COMSOL 
[Gopal, Jan]
For the past couple of days, Jan and I have been discussing a major issue in
COMSOL involving modeling both oscillatory and non-oscillatory forces simultaneously while using FDA. It turns out that he and I had run into the
Entry  Thu Jul 29 12:31:24 2010, josephb, Update, CDS, Working DAC, working IOP - next up SUS Simple_Green_Control.png
Ok, after a few minutes of talking to Alex, I got the correct "GUI syntax" through my head, and we now have a simple working green end control
which in fact puts signals out through the DAC.
Note to self, do not put any additional filters or controls in the IOP module.  Basically just change the master block with GDS numbers, DCU_ID
Entry  Thu Jul 29 12:13:28 2010, kiwamu, Summary, CDS, near term plan current_setup_v2.png
 [Joe and Kiwamu]
Entry  Thu Jul 29 11:33:38 2010, kiwamu, Update, CDS, PCI5565 driver for RFM 
 Yesterday I installed a PCI-5565 driver on new C1SUS in order to test the RFM.
Since the
RFM on the new CDS is not working, we had to test it by using some softwares.

Entry  Mon Jul 26 20:12:18 2010, kiwamu, Update, SUS, OSEMs on PRM DSC_2279.JPG
 [Alberto and Kiwamu]
We installed the OSEMs to the new PRM.
As I wrote down on the elog (
    Reply  Tue Jul 27 21:38:21 2010, kiwamu, Update, SUS, installation of in-vac optics DSC_2284.JPGDSC_2291.JPG
[Alberto and Kiwamu] 
We put the PRM back approximately on the right place.
Also we installed the pick off mirrors and the PZT mirror.
       Reply  Wed Jul 28 17:32:37 2010, kiwamu, Update, SUS, installation of in-vac optics 
[Jenne, Koji and Kiwamu]
We have installed two steering mirrors and the green periscope.
Also we took the tip-tilt out from the chamber for characterization
          Reply  Thu Jul 29 10:39:43 2010, kiwamu, Update, SUS, Re: installation of in-vac optics 
I updated the
last entry about the in-vac work (see here)
Entry  Wed Jul 28 18:24:57 2010, Jenne, Update, Green Locking, 2 Green Periscopes have mirrors, aligned 
[Koji, Jenne, Kiwamu]
This is to describe the work that went on in the Cleanroom today.  Kiwamu's entry will detail the tidbits that happened in the chamber.
We engraved the periscope mounts with the mirror info for the mirrors which were placed in the periscope.  We also engraved the barrels of
Entry  Wed Jul 28 14:54:56 2010, steve, Update, SAFETY, evacuation drill at the 40m P1060520.JPG
Head count at the evacuation drill today. I checked  alarms and flashers  at room 104,102,101, 103, 105 and 107. They were really loud and
bright. 
Entry  Wed Jul 28 14:34:29 2010, channa, Update, Computers, installation on allegra 
I have done the following on allegra and rosalba:
[root@allegra caltech]# yum install glade2
On rosalba the matplotlib was out of date with respect to allegra.  I have no idea how the version 0.98 on allegra got there, but I left it. 
Entry  Wed Jul 28 13:06:47 2010, Razib, Update, Phase Camera,  mean_100_snaps.pngstd_100_snaps.pngphase_cam_calc.pdf
Attached are some calculation that I did previously for the phasecamera setup. This shows the nature of the beat signal that we are measuring.
I am also trying to characterize the noise source of the camera also. Following images shows the mean dark noise (with no light on the camera)
and the standard deviation for 100 snaps at an exposure time of 500 µs.
ELOG V3.1.3-