  11072   Thu Feb 26 00:20:54 2015 ericqUpdateLSCModelled effect of relative modulation phaseI'm working on some more modelling
investigations of this whole situation. The
main thing I wanted to do was to look at
 33budget_canceled.png33budget_nominal.png2015-02-ModPhase.zip 
  11071   Wed Feb 25 23:48:57 2015 ranaUpdateLSCsome thoughts
Comparing just the 2 cases with
locking on 33, it seems that the 55 MHz gain
has changed by 14 dB instead of the 10 dB
  
  11070   Wed Feb 25 20:00:39 2015 JenneUpdateLSCDelay line un-installed again - sensing matrix comparisonsI have measured the sensing matrix for
the PRMI at the REFL photodiodes for both
the nominal configuration and the 33MHz cancellation
  
  11069   Wed Feb 25 14:51:13 2015 JenneUpdateLSCDelay line un-installed againAnd now I've removed the delay line,
and am in the process of reverting the demod
phases, etc.
  
  11068   Wed Feb 25 14:35:32 2015 SteveUpdatesafetysafety audit 2015Safety audit went soothly. We thank all
participients.

Correction list:
 safety2015audit.jpg 
  11067   Wed Feb 25 14:18:28 2015 ranaUpdateLSCDelay line re-installed, measurements round 2The Schnupp asymmetry is definitely not
an important parameter (no need for Koji
to explode). It only serves to give us a
  
  11066   Wed Feb 25 12:16:27 2015 KojiUpdateLSCDelay line re-installed, measurements round 2WHAT?
WHAT? WHAT? It's obviously opposite.

If
the reflectivity of the front mirror is fixed
  
  11065   Wed Feb 25 11:01:05 2015 manasaUpdateComputer Scripts / ProgramsNew screen for FOL PID loopCreated a new medm screen C1ALS_FOL_PID.adl for
FOL PID loop control in 
  
  11064   Wed Feb 25 04:59:47 2015 JenneUpdateLSCDelay line re-installed, measurements round 2[Jenne, EricQ, Rana]

We spent this evening measuring
and thinking about our 3f signals, and the
 SensMat_PRMI_24Feb2015.pdf 
  11063   Wed Feb 25 04:21:58 2015 ericqUpdateCDSSome model updatesI changed the suspension library block
to acquire the SUS_[optic]_LSC_OUT channels
at 16k for sensing matrix investigations.
  
  11062   Tue Feb 24 23:39:16 2015 JenneUpdateLSCDelay line re-installed againAbout 5-10 minutes ago I just put in the
modulation cancellation setup according to
the recipe in http://nodus.ligo.caltech.edu:8080/40m/11032
  
  11061   Tue Feb 24 18:54:26 2015 ericqUpdateASCSingle arm QPD ASC stabilityI've lowered the UGFs for the transmission
QPD servos to ~1-2Hz, and made it just an
integrator. I left the arms locked with the
 trend.png 
  11060   Tue Feb 24 01:57:01 2015 JenneUpdateLSCModelled effect of relative modulation phaseI have used Optickle to model the effect
of changing the phase between the 11MHz and
55MHz EOMs.  Also, I have looked at
 ChangeInAngle_PRCLvsMICH.pngREFL33_MICH_SensMat_vsModulationOrder_vsModulationPhase.pngREFL33_SensMat_vsModulationPhase.pngREFL33_PRCL_nominalVsLoweredGamma.pngREFL33_MICH_nominalVsLoweredGamma.png 
  11059   Mon Feb 23 21:57:13 2015 KojiUpdateLSCDelay line installed again (experiment, round 1)Last Wednesday we tried PRMI 3f modulation
cancellation. Under the 3f modulation cancellation,
the PRMI could not be locked
  
  11058   Mon Feb 23 15:04:10 2015 ZachUpdateGeneralElog restartedThe elog crashed while I was creating
an entry to the Cryo log. I restarted it
with the 
  
  11057   Mon Feb 23 08:06:59 2015 ranaUpdateASCQPD frontend code unifiedMight have to also get the OL screens that
go with this new code to see, but the calibrations
don't go into the matrix, but rather
  
  11056   Fri Feb 20 19:09:48 2015 ericqUpdateASCQPD frontend code unifiedI have changed all of the oplevs and transmon
QPDs to use the common ISC QPD library block,
which differs mainly in its divide by zero
  
  11055   Fri Feb 20 14:44:47 2015 ericqUpdateCDSBad CDS behaviorIn an effort to ease the IO load on the
framebuilder, I've cleaned up the DQ
channels being written to disk. The biggest
  
  11054   Fri Feb 20 12:29:01 2015 ericqUpdateCDSAll optics dampedI noticed some diagnostic bits in the c1sus
IOP complaining about user application timing
and FIFO under/overflow (The second and fourth
  
  11053   Fri Feb 20 12:08:10 2015 ericqUpdateCDSBad CDS behaviorI've been able to get all models running.
Most optics are damped, but I'm having
trouble with the ITMs, BS and PRM. 
  
  11052   Thu Feb 19 23:23:52 2015 ChrisUpdateCDSBad CDS behaviorThe frontends have some paths NFS-mounted
from fb. fb is on the ragged edge of being
I/O bound. I'd suggest moving those mounts
  
  11051   Thu Feb 19 15:45:43 2015 ericqUpdateCDSBad CDS behaviorAt about 10AM, the C1LSC frontend stopped
reporting any EPICS information. The arms
were locked at the time, and remained so
  
  11050   Thu Feb 19 04:16:45 2015 JenneUpdateLSCALS fool attempt with PRFPMI[Jenne, EricQ]

We tried several times tonight
to engage the Fool path with the PRFPMI. 
 BEATYcrazy_duringCARMdarmLock_18Feb2015.png19Feb.png 
  11049   Thu Feb 19 04:09:21 2015 ericqUpdateElectronicsSecond QPD Whitening Switch enabledX end QPD has recieved 0.2+0.4 absorptive
ND filters. Y end QPD got one at 0.6. This
appears to have mitigated the saturations
  
  11048   Wed Feb 18 19:06:40 2015 KojiUpdateLSCALS Fool impulse response   latex-image-1.png 
  11047   Wed Feb 18 17:51:40 2015 ericqUpdateLSCALS Fool impulse responseKoji raised a good question about the step
response I wrote about. Namely, if the UGF
of the ALS servo is around 100Hz, we would
 ALSfool_fasterKick.png 
  11046   Wed Feb 18 01:58:51 2015 ericqUpdateLSCALS Fool single arm performance I'm playing around with the lastest
ALS fool feedforward on the Yarm, and I like
what I'm seeing. 
 ALSfool_spectra.pngALSfool_kick.pngALSfool_LoopShapes.pngALSfool_Feb182015.zip 
  11045   Tue Feb 17 19:49:51 2015 KojiUpdateLSCDelay line un-installed againThe modulation setting was reverted.
Demod phase for REFL11/33/55/165 and
AS55 were reverted to the previous numbers
  
  11044   Tue Feb 17 16:44:04 2015 KojiUpdateLSCDelay line installed againFor tonight's experiment, I re-installed
the delay line cable and changed the attenuation
to 10dB for the 55MHz modulation.
  
  11043   Tue Feb 17 16:36:08 2015 JenneUpdateLSCALS fool mathHere is an updated cartoon, with the ALS
sensor explicitly shown as the beatbox times
the closed loop response of the phase tracker
 ALSfool_LoopDiagram_17Feb2015.pngALS_REFL_comboLocking_16Feb2015.nb.zip 
  11042   Tue Feb 17 04:04:32 2015 JenneUpdateLSCALS fool mathI re-did the Mathematica notebook according
to the most current diagram (note to daytime
self: attach .nb file!!!), and found that
 OpenLoopGainComparison_16Feb2015.pngCancellationTFs_DecouplingIsArefl.png 
  11041   Tue Feb 17 00:24:47 2015 rana, jenneUpdateLSCALS Fool filter updated for more cancellationToday
we measured the TFs again and then updated
the filter in the POY -> ALS FF path so
 ALSfoo_150216.png15.png 
  11040   Mon Feb 16 21:52:51 2015 ranaHowToTreasurebig Dataviewer windowsFollowing this
entry, I have made the same change in
the controls account on rossa:
 bigDV.png 
  11039   Mon Feb 16 15:08:26 2015 JenneUpdateLSCALS fool measured decoupling TFDang it, yes. You're right.  I
should have caught that. 

Since Dcpl and Srefl are both zero
  
  11038   Mon Feb 16 03:10:42 2015 KojiUpdateLSCALS fool measured decoupling TFWonkey shape: Looks like a loop supression.
Your http://nodus.ligo.caltech.edu:8080/40m/11016
also suggests it too, doesn't it?
  
  11037   Mon Feb 16 02:49:57 2015 JenneUpdateLSCALS fool measured decoupling TFI have measured very, very carefully the
transfer function from pushing on
MC2 to the Yarm ALS beatnote. 
 ALSfool_measuredActuatorTF_YarmOnly_15Feb2015.pngYARM_actTF_compareActuators.pdf 
  11036   Mon Feb 16 01:45:12 2015 KojiSummaryLSCmodulation depth analysisBased on the measured modulation profiles,
the depth of each modulation was estimated.
Least square sum minimization of the
 modulation_nominal.pdfmodulation_3f_reduced.pdf 
  11035   Mon Feb 16 00:08:44 2015 KojiSummaryLSC[ELOG LIST] 3f modulation cancellationThis KTP crystal has the maximum allowed
RF power of 10W (=32Vpk) and V_pi = 230V.
This corresponds to the maximum allowed
  
  11034   Sun Feb 15 20:55:48 2015 ranaSummaryLSC[ELOG LIST] 3f modulation cancellationI wonder if DRMI can be locked on 3f using
this lower 55 MHz modulation depth. It seems
that PRMI should be unaffected, but that
  
  11033   Sun Feb 15 16:20:44 2015 KojiSummaryLSC[ELOG LIST] 3f modulation cancellationSummary
of the ELOGS

3f
modulation cancellation theory 
  
  11032   Sat Feb 14 22:14:02 2015 KojiSummaryLSC[HOW TO] 3f modulation cancellationWhen I finished my measurements, the modulation
setup was reverted to the conventional one.
If someone wants to use the 3f cancellation
 freq_gen_box.JPGdelay_line.JPGcable_spec.pdf 
  11031   Sat Feb 14 20:37:51 2015 KojiSummaryLSC3f modulation cancellationExperimental
results

- PD response [Attachment
 beat_pd_response.pdfbeat_nominal.pdf3f_reduction.pdfbeat_3f_reduced.pdf 
  11030   Sat Feb 14 20:20:24 2015 JenneUpdateLSCALS fool cartoonThe ALS fool scheme is now diagrammed up
in OmniGraffle, including its new official
icon.  The mathematica notebook has
 ALSfool_LoopDiagram.pngALSfool_LoopDiagram.graffle.zip 
  11029   Sat Feb 14 19:54:04 2015 KojiSummaryLSC3f modulation cancellationOptical
Setup

[Attachment 1]
 beat_setup1.JPGbeat_setup2.JPGelectrical_setup.pdf 
  11028   Sat Feb 14 00:48:13 2015 KojiUpdateLSC3f modulation cancellation[SUCCESS]
The 3f sideband cancellation seemed worked
nicely.
  
  11027   Sat Feb 14 00:42:02 2015 KojiUpdateGeneralRF amplifier for ALSThe RF analyzer was returned to the control
room. There are two beat notes from X/Y confirmed.

I locked the arms and aligned them
  
  11026   Fri Feb 13 19:41:13 2015 ericqUpdateGeneralRF amplifier for ALSAs Koji found one of the spare channels
of the ALS/FOL RF amplifier box nonfunctional
yesterday, I pulled it out to fix it. I found
  
  11025   Fri Feb 13 18:56:44 2015 ranaUpdateElectronicsSecond QPD Whitening Switch enabledDepends on the plots of the whitening I
guess; if its low freq sat, then we lower
the light level with ND filters. If its happening
  
  11024   Fri Feb 13 17:07:51 2015 JenneUpdateElectronicsSecond QPD Whitening Switch enabledI first updated the DCC branches for the
Xend and Yend to reflect the as-built situation
from December 2014, and then I updated the
  
  11023   Fri Feb 13 14:59:13 2015 ericqUpdateElectronicsSecond QPD Whitening Switch enabledWent to zero CARM offset on ALS; transmission
QPDs are still saturating :(

Maybe we need to switch off all
  