Wed Oct 20 21:43:25 2010 |
yuta | Update | CDS | checking whitening filters for MCs and messed up | (Joe, Yuta)
Background:
We found that the damping servo for MC suspensions somehow worked when we turned off the 13Hz Chebyshev filters.
But that does not meet our satisfaction, so we started checking every components.
First of all is the whitening filters.
If we turn on a digital whitening filter(WF), corresponding analog WF should turn off, and vice versa.
Reference:
See DCC #D000210 for the analog circuit of WF. WF has 3Hz zero, 30Hz pole, 100Hz pole. MAX333A bypasses analog WF when supplied +15V(HIGH).
What we did:
1. Compared the transfer function between MC2_SUSPOS_EXC and MC2_ULSEN_IN1 when digital WF on and off. When digital is on/off, analog should be off/on, so there should be difference but couldn't see.
2. We went through the simulink model and found 2 mistakes in the logic. One is the conflict with other optics. Even if we turn on/off digital WF of MC2, it didn't switched analog WF of MC2. Two is the additional bit invert (but it turned out to be our misunderstanding).
3. We (thought we) fixed it, rebuild it, and measured the TF again.(Attachment #1)
[Attached #1]
The red/blue line is when digital WF is on/off. Blue should be bigger(+10dB @ 10Hz according to (analog) WTF) than red, but it was the opposite.
4. To confirm that they are doing the opposite, I checked MAX333A input(pin#1,10,11,20) in "SUS PD Whitening Board" at 1X5-1-8B (which is for MC3) and found that switching is opposite. When I turned off/on the digital WF, MAX333A input was +15V/0V. It should be 0V/+15V.
5. Also, I found that LLSEN digital WF switch switches LRSEN analog WF and vice versa.
[Attached #2]
Transfer functions between MC2_SUSPOS_EXC and MC2_LRSEN_IN1 with;
Red: LR digital off, LL digital on
Blue: LR digital off, LL digital off
Green: LR digital on, LL digital off
As you can see, LL switch is the one which switches LR analog WF now.
Conclusion:
Currently, digital WF on/off corresponds to analog WF on/off.
Also, LL/LR digital WF switch is LR/LL analog WF switch now.
Next work:
- fix the simulink models (or wiring)
- check dewhitening filters
Schematic:
- whitening
MC1 5 PD outputs -> SUS PD Whitening Board(D000210) -> ... (digital WF=3,100:3)
MC2 5 PD outputs -> SUS PD Whitening Board(D000210) -> ... (digital WF=3,100:3)
MC3 5 PD outputs -> SUS PD Whitening Board(D000210) -> ... (digital WF=3,100:3)
D000210 has switches for bypassing analog WT(3,100:3). HIGH to bypass.
- dewhitening
(-) ... -> SOS Dewhitening Board(D000316) -> MC1,3 UL/UR/LR/LL coils
(-) ... -> SOS Dewhitening Board(D000316) -> MC1,2,3 SIDE coils
(SimDW)(InvDW) ... -> LSC Anti-imaging Board(D000186) -> Universal Dewhitening Board(D000183) -> MC2 UL/UR/LR/LL coils
D000316 has switches for bypassing 28Hz elliptic LPF. HIGH to bypass.
D000186 is 7570Hz elliptic LPFs.
D000183 has switches for bypassing dewhitening filter. HIGH to bypass.
See this wiki page for more comprehensive setup. |
Attachment 1: MC2ULSEN.png
|
|
Attachment 2: MC2LRSEN.png
|
|
3750
|
Thu Oct 21 05:51:10 2010 |
kiwamu | Update | Locking | no green beat note | Still I didn't see any beat note signals..
With a help from Suresh, Yuta and Rana, I tried searching for the green beat note by changing the temperature of the X end NPRO.
The noise level after it goes through an RF amplifier (G=23dB) was about -70dBm at 50MHz.
This noise may cover the beat note signal.
I am going to post some details later. |
3752
|
Thu Oct 21 12:15:02 2010 |
rana | Update | PEM | 6.9 Mag EQ in Gulf of California |
Magnitude |
6.9 |
Date-Time |
|
Location |
24.843°N, 109.171°W |
Depth |
10 km (6.2 miles) set by location program |
Region |
GULF OF CALIFORNIA |
Distances |
105 km (65 miles) S of Los Mochis, Sinaloa, Mexico
125 km (75 miles) SW of Guamuchil, Sinaloa, Mexico
140 km (85 miles) NE of La Paz, Baja California Sur, Mexico
1200 km (740 miles) WNW of MEXICO CITY, D.F., Mexico |
Location Uncertainty |
horizontal +/- 6.1 km (3.8 miles); depth fixed by location program |
Parameters |
NST=187, Nph=187, Dmin=843.1 km, Rmss=1.17 sec, Gp=133°,
M-type=teleseismic moment magnitude (Mw), Version=6 |
Source |
|
Event ID |
us2010crbl |
|
3754
|
Thu Oct 21 15:59:28 2010 |
kiwamu | Update | Locking | no green beat note : details | Last night I tried searching for a beat note signal with two different PD trans impedance gains.
Although I didn't find a beat note signal.
- (1. trans impedance gain = 2400)
I started with a trans impedance resistance of R=2.4k, which is 10 times bigger resistance than the original.
The total PD gain should be about 960 [V/W] theoretically if we assume the responsibility of the PD is 0.4 [A/W].
Then I checked the bandwidth of the RFPD using Jenne laser.
The bandwidth was about 30MHz, which is 3 times narrower than the original. And it agrees with our expectation.
As Koji and I mentioned at the last weekly meeting, the cut off frequency of an RFPD follows inverse square root of the trans impedance resistance R.

where C is a capacitance of the photo diode. (See this)
I was expecting the signal level of -50 dBm / rtHz with a 23dB RF amplifier, assuming the line width of the signal is 10kHz.
- (2. trans impedance gain = 240)
I also tried it with the original trans impedance gain (see this entry).
R = 240 [Ohm]
G = 96 [V/W]
BW = 100 [MHz] (I didn't measure it in this time)
expected signal level = -70 dBm/rtHz
Quote: |
Still I didn't see any beat note signals..
|
|
3755
|
Thu Oct 21 18:45:50 2010 |
Koji | Update | PSL | Found the beat at 1064nm | [Koji Suresh]
We found the beat at 1064nm. T(PSL)=26.59deg, T(X-end)=31.15deg.
The X-end laser was moved to the PSL table.
The beating setup was quickly constructed with mode matching based on beam profile measurements by the IR cards.
We used the 1GHz BW PD, Newfocus #1611-FS-AC.
As soon as we swept the Xtal temp of the X-end laser, we found the strong beating. |
3756
|
Thu Oct 21 19:10:39 2010 |
Aidan | Update | PSL | Found the beat at 1064nm |
Quote: |
[Koji Suresh]
We found the beat at 1064nm. T(PSL)=26.59deg, T(X-end)=31.15deg.
The X-end laser was moved to the PSL table.
The beating setup was quickly constructed with mode matching based on beam profile measurements by the IR cards.
We used the 1GHz BW PD, Newfocus #1611-FS-AC.
As soon as we swept the Xtal temp of the X-end laser, we found the strong beating.
|
|
3757
|
Thu Oct 21 21:35:16 2010 |
yuta | Update | CDS | fixed whitening filter switches | (Joe, Yuta)
Summary:
We found some mistakes in whitening filter switches yesterday(see elog #3748).
One was the mis-connection between LL/LR and the second was the bit invert.
So, we fixed it and checked that they are now switching correct by probing MAX333A.
Why we made mistake:
1. LL/LR mis-connection
Simply, Simulink model was wrong.
It occurred because UL/UR/LR/LL order is currently arbitrary and confusing.
For now, we just swap the connection(Attached #1).
In the future, we should fix all orders corresponding to the actual wiring order(UL > LL > UR > LR).
2. Bit invert
It occurred from counter-intuitive output of Contec 32 BO(See Attachment #2(taken from this wiki page)).
If "1," current goes from A to B, so MAX333A input will be 0V, which means analog WF is on.
If "0," no current, so MAX333A input will be +15V, which means analog WF is off(bypassed).
Next work:
- There are no digital WF in SDSEN inputs. So, we have to put them.
- Dewhitening filters are totally messed up now. Switches are wrong, some optics have digital DW but some doesn't, some have 28Hz elliptic LPF.......
We have to grand unify them. |
Attachment 1: LLLR.png
|
|
Attachment 2: Contec23BO.png
|
|
3758
|
Fri Oct 22 00:59:01 2010 |
yuta | Update | CDS | new MEDM screens, new filter banks(whitening and dewhitening related) | (Joe, Yuta)
Summary:
No more discrimination for SIDE!
We had all SIDE filters in SDSEN, so we splitted into SDSEN, SUSSIDE, and SDCOIL just like other DOFs. (Not SIDECOIL. SDCOIL from now on!)
Also, there was a confusion in output filters, so we fixed the filter bank(dewhitening and 28HzELP).
More than that, I found that "13HzCheby" in SUSPOS, SUSPIT, SUSYAW was actually doing ~1.6Hz Chebyshev, so I fixed it. (No wonder we had no damping when turing "13HzCheby" on!)
Our new MEDM screen is Attachment #1.
Input filter design:
Every optics, every OSEM has same analog whitening filter.
So;
digital |
analog |
30,100:3 |
30,100:3 |
ON |
OFF |
OFF |
ON |
Output filter design:
For every SIDE coils and MC1, MC3 coils, they have analog 28Hz elliptic LPFs.
So;
digital |
analog |
28HzELP |
28HzELP |
ON |
OFF |
OFF |
ON |
For MC2(and other main optics) UL/UR/LR/LL coils, they have analog dewhitening filters.
So;
digital |
analog |
InvDW |
SimDW |
DW |
ON |
ON |
OFF |
ON |
OFF |
ON |
InvDW and SimDW were in FM10 and FM9, but I swapped them to make it more consistent.
Now, FM10 switches analog/digital output filiters for every optics.
Let's put 28HzELP in FM9 so that FM9 switches analog/digital for every optics.
Quote: |
Schematic:
- whitening
MC1 5 PD outputs -> SUS PD Whitening Board(D000210) -> ... (digital WF=3,100:3)
MC2 5 PD outputs -> SUS PD Whitening Board(D000210) -> ... (digital WF=3,100:3)
MC3 5 PD outputs -> SUS PD Whitening Board(D000210) -> ... (digital WF=3,100:3)
D000210 has switches for bypassing analog WT(3,100:3). HIGH to bypass.
- dewhitening
(-) ... -> SOS Dewhitening Board(D000316) -> MC1,3 UL/UR/LR/LL coils
(-) ... -> SOS Dewhitening Board(D000316) -> MC1,2,3 SIDE coils
(SimDW)(InvDW) ... -> LSC Anti-imaging Board(D000186) -> Universal Dewhitening Board(D000183) -> MC2 UL/UR/LR/LL coils
D000316 has switches for bypassing 28Hz elliptic LPF. HIGH to bypass.
D000186 is 7570Hz elliptic LPFs.
D000183 has switches for bypassing dewhitening filter. HIGH to bypass.
See this wiki page for more comprehensive setup.
|
Next work:
- Fix Simulink connection and logic so that analog/digital switching go correctly.
- Check if the switching are correct for all channels by measuring transfer functions.
- Currently, we are focusing on MCs, but we have to do the same fixing for other optics, too.
- More than setting the right filter TF, there are switching settings like Zero History, Ramp, Zero Crossing...... We should investigate that.
- Actually, TF of analog DW in D000183 doesn't look like the same with digital SimDW. Why?? |
Attachment 1: HappySideMEDM.png
|
|
3759
|
Fri Oct 22 01:23:13 2010 |
Koji | Update | PSL | Found the beat at 1064nm | [Koji / Suresh]
We worked on the 1064 beating a bit more.
- First of all, FSS and FSS SLOW servo were disabled not to have variating Xtal temp for the PSL.
- The PSL Xtal temp (T_PSL) was scanned from 22deg-45deg while we search the Xtal temp (T_Xend) for the Xend laser to have the beat freq well low (f<30MHz).
The pumping current for each laser was I_PSL = 2.101 [A] and I_Xend = 2.000 [A]
For a certain T_PSL, we found multiple T_Xend because the freq of the laser is not a monotonic function of the Xtal temperature. (see the innolight manual).
T_Xend to give us the beating was categorized in the three sets as shown in the figure. The set on "curve2" is the steadiest one. (Use this!)
The trends were quite linear but the slope was slightly off from the unity.
- T_PSL was scanned to see the trend of the PMC output.
The PMC was sometimes locked to the mode with lower transmission (V_PMCT ~ 3.0V).
When T_PSL ~ 31deg we consistently locked the PMC at higer transmission (V_PMCT ~ 5.3V).
At the moment we decided the operating point of T_PSL = 32.25 deg, V_PMCT = 5.34, where we found the beat at T_Xend=38.28deg.
- We cleaned up the PSL table more than how it was. Returned the tools to their original places.
The X-end laser was shut down and was left on the PSL table.
NEXT:
Kiwamu can move the X-end laser to the Xend and realign it.
Then we should be able to see the green beating quite easily. |
Attachment 1: 101021_beat.pdf
|
|
3760
|
Fri Oct 22 03:37:56 2010 |
Kevin | Update | PSL | Quarter Wave Plate Measurements | [Koji and Kevin]
We measured the reflection from the PBS as a function of half wave plate rotation for five different quarter wave plate rotations. Before the measurement we reduced the laser current to 1 A, locked the PMC, and recorded 1.1 V transmitted through the PMC. During the measurements, the beam was blocked after the faraday isolator. After the measurements, we again locked the PMC and recorded 1.2 V transmitted. The current is now 2.1 A and both the PMC and reference cavities are locked.
I will post the details of the measurement tomorrow. |
3761
|
Fri Oct 22 15:06:43 2010 |
Jenne | Update | SUS | Arts and Crafts! | This afternoon I epoxied the guiderod and wire standoff to the new PRM. I also epoxied the magnets that Suresh picked out to the dumbbell standoffs. We'll let them all cure over the weekend, and then I'll glue the magnets to the optic on ~Monday.
Notes about the epoxy:
Previously, we had been using the "AN-1" epoxy, which is gray, with a clear hardener. Bob recommended we switch to "30-2", which is clear with clear, and has been chosen for use in aLIGO. Both were vacuum approved, but the 30-2 has gone through ~2 months of testing at the OTF (Optics Test Facility?) over in Downs under vacuum, to check the level of outgassing (or really, non-outgassing).
The 30-2 is less viscous than the AN-1, and it takes less glue to do the same job, so we should keep that in mind when applying the epoxy. When I put the glue next to the guiderod and standoff, it got wicked along the length of each rod, which is good. I can't reach the whole length of the rod with my glue applicator because the fixture holding them in place blocks access, so the wicking is pretty handy.
I've also added the updated version of my Status Table for the suspensions. |
Attachment 1: StatusTable.png
|
|
3762
|
Fri Oct 22 16:59:21 2010 |
Jenne | Update | Electronics | Epic Takeover | As the suspension work winds down (we'll be completely done once the ETMs arrive, are suspended, and then are placed in the chambers), I'm going to start working on the RF system.
Step 1: Figure out what Alberto has been up to the last few months.
Step 2: Figure out what still needs doing.
Step 3: Complete all the items listed out in step 2.
Step 4: Make sure it all works.
Right now I'm just starting steps 1 & 2. I've made myself a handy-dandy wiki checklist: RF Checklist. Hopefully all of the bits and pieces that need doing will be put here, and then I can start checking them off. Suggestions and additions to the list are welcome. |
3763
|
Fri Oct 22 18:15:21 2010 |
kiwamu | Update | Locking | Green era: found a green beat note finally | finally we found it !
 |
3764
|
Fri Oct 22 18:22:27 2010 |
Alberto | Update | Electronics | Epic Takeover |
Quote: |
As the suspension work winds down (we'll be completely done once the ETMs arrive, are suspended, and then are placed in the chambers), I'm going to start working on the RF system.
Step 1: Figure out what Alberto has been up to the last few months.
Step 2: Figure out what still needs doing.
Step 3: Complete all the items listed out in step 2.
Step 4: Make sure it all works.
Right now I'm just starting steps 1 & 2. I've made myself a handy-dandy wiki checklist: RF Checklist. Hopefully all of the bits and pieces that need doing will be put here, and then I can start checking them off. Suggestions and additions to the list are welcome.
|
There's also a page dedicated to the progress in the PD upgrade process:
http://lhocds.ligo-wa.caltech.edu:8000/40m/Upgrade_09/RF_System/Upgraded_RF_Photodiodes
There you can find a pdf document with my notes on that. |
3766
|
Fri Oct 22 21:28:27 2010 |
Aidan | Update | Locking | Green era: found a green beat note finally | Nice work!
Quote: |
finally we found it !

|
|
3767
|
Fri Oct 22 23:29:48 2010 |
yuta | Update | CDS | fixed output filter switching | (Joe, Yuta)
Summary:
This evening, we fixed output filter analog/digital switchings in Simulink logic, but they didn't actually switched filters.
That was because what we thought BO0 was BO2 and BO2 was BO0.
We re-wired, re-labeled the cables.
We checked it by using a LED(see "test configuration" in this wiki page).
Also, we checked the Simulink connection by probing MAX333A in SOS Dewhitening Board located at 1X4-8-11B, which switches SIDE coils for all optics.
Next work:
- Currently, FM10 in UL/UR/LR/LL/SDCOIL filters switch analog/digital, but let's make FM9 do the switching.(See schematic in elog #3758)
- Check all the logic by measuring transfer functions(maybe single frequency measurement is enough. we can use tdsdmd etc to automate). |
3768
|
Sat Oct 23 02:25:49 2010 |
Kevin | Update | PSL | Quarter Wave Plate Measurements |
Quote: |
[Koji and Kevin]
We measured the reflection from the PBS as a function of half wave plate rotation for five different quarter wave plate rotations. Before the measurement we reduced the laser current to 1 A, locked the PMC, and recorded 1.1 V transmitted through the PMC. During the measurements, the beam was blocked after the faraday isolator. After the measurements, we again locked the PMC and recorded 1.2 V transmitted. The current is now 2.1 A and both the PMC and reference cavities are locked.
I will post the details of the measurement tomorrow.
|
I measured the reflected power from the PBS as a function of half wave plate rotation for five different quarter wave plate rotations.
The optimum angles that minimize the reflected power are 330° for the quarter wave plate and 268° for the half wave plate.
The following data was taken with 2.102 A laser current and 32.25° C crystal temperature.
For each of five quarter wave plate settings around the optimum value, I measured the reflected power from the PBS with an Ophir power meter. I measured the power as a function of half wave plate angle five times for each angle and averaged these values to calculate the mean and uncertainty for each of these angles. The Ophir started to drift when trying to measure relatively large amounts of power. (With approximately 1W reflected from the PBS, the power reading rapidly increased by several hundred mW.) So I could only take data near the minimum reflection accurately.
The data was fit to P = P0 + P1*sin^2(2pi/180*(t-t0)) with the angle t measured in degrees with the following results:
lambda/4 angle (°) |
t0 (°) |
P0 (mW) |
P1 (mW) |
chi^2/ndf |
V |
318 |
261.56 ± 0.02 |
224.9 ± 0.5 |
2016 ± 5 |
0.98 |
0.900 ± 0.001 |
326 |
266.07 ± 0.01 |
178.5 ± 0.4 |
1998 ± 5 |
16.00 |
0.918 ± 0.001 |
330 |
268.00 ± 0.01 |
168.2 ± 0.3 |
2119 ± 5 |
1.33 |
0.926 ± 0.001 |
334 |
270.07 ± 0.02 |
174.5 ± 0.4 |
2083 ± 5 |
1.53 |
0.923 ± 0.001 |
342 |
273.49 ± 0.02 |
226.8 ± 0.5 |
1966 ± 5 |
1.41 |
0.897 ± 0.001 |
where V is the visibility V = 1- P_max/P_min. These fits are shown in attachment 1. We would like to understand better why we can only reduce the reflected light to ~150 mW. Ideally, we would have V = 1. I will redo these measurements with a different power meter that can measure up to 2 W and take data over a full period of the reflected power. |
Attachment 1: fits.png
|
|
3769
|
Sat Oct 23 03:36:05 2010 |
yuta | Update | CDS | fixed filters for C1SUS, C1RMS, C1MCS | (Joe, Yuta)
Summary:
This Monday, MC suspension damping got something wrong.
We started to check filters and found that digital filters were wrong because of mis-conversion from old filter files to new files.
We converted the file again, and with mutual understanding between Foton and us, we finally got correct filters(I hope!).
What we did:
1. Merged filter files in old /cvs/cds/caltech/chans/ directory into C1SUS.txt(BS,ITMX,ITMY), C1RMS.txt(PRM,SRM), C1MCS.txt(MC123).
2. Rebuilt the RT models in order to get a correct filter file header(they have list of filter modules).
3. Concatenate the header with filter design part which we got from step1.
4. Replaced 'N 2048' with 'N 16384'
It replaces sampling rate of "XXSEN"s.
Basically, step1-4 was the same with what we did last time. We didn't changed the fitler coefficients, so Foton somehow changed the original filter design.
So, this time, we
5. Deleted coefficients like we did on Tuesday (see elog #3774).
But Foton couldn't read the file correctly this time. Foton seemed to be unbeatable.
Even if we replaced the sampling rate, Foton kept saying 2048! (This is maybe because Foton's default value is 2048Hz. Everytime Foton notice some editting in the file, he destroys everything. He hates editting)
The problems were always associated with the sampling rate, so
6. Got back to step4 and undo-ed the replacement.
7. Foton could read it this time, so I changed the sampling rate one by one using Foton GUI.
8. Checked filters using Foton's Bode Plot.(Not for all, but some that had problem before)
9. Splitted SDSEN filters to SDSEN, SUSSIDE, and SDCOIL.
10. Put some missing whitening filters, and 28HzELPs.
BS, PRM, SRM didn't have any 28HzELP for SDCOIL.
ITMX and ITMY SDCOIL had SimDW/InvDW which doesn't make sense(SIDEs don't have analog DW). So, I deleted and replaced with 28HzELP.
F2A issue:
We failed in sending F2A filters to new filter files.
These are a little bit complicated because TO_COIL_X_X filters were named ULPOS,URPOS etc before.
Also, MC3 didn't have any F2As, so maybe we should but the same F2As as MC1/2.
Note that every F2As are different, and TO_COIL matrix have UL,UR,LL,LR order(not same as INMATRIX).
Also, SRM had f2pv instead of F2A!
Next work:
- Check whole filters by actually measuring transfer function between SENs and COILs.
- Damp MC suspentions, and lock MC.
- Measure openloop TF and compare with the designed.
How do you read a Foton filter file:
When you open up a Foton filter file, you see filters like this.
################################################################################
### modulename ###
################################################################################
# SAMPLING modulename samplingrate
# DESIGN modulename n filterdesign
# DESIGN modulename n filterdesign
### ###
modulename n xy z v w filtername gain a1 a2 b1 b2
modulename n xy z v w filtername gain a1 a2 b1 b2
a1 a2 b1 b2
n: filter number
0 for FM1, 1 for FM2, ... , 9 for FM10
x: Input Switching setting
1 Always On
2 Zero History
y: Output Switching setting
1 Immediately
2 Ramp
3 Input Crossing
4 Zero Crossing
z: number of filters cascaded.
v: if y=2, (Ramp Time(sec))*(samplingrate)
if y=3 or 4, Tolerance
w: (Timeout(sec))*(samplingrate)
Note that v and w are changed when sampling rate is changed.
Transfer function will be;
H(1/z)=G*(1+b1/z+b2/z/z)/(1+a1/z+a2/z/z)
z=exp(s/fs)
where fs is the sampling frequency.
Reference:
Kiwamu Izumi: "Notes about Digital Filters," http://tamago.mtk.nao.ac.jp/izumi/green/DigitalFilter.pdf |
3773
|
Sun Oct 24 19:55:50 2010 |
kiwamu | Update | Electronics | lonely RF amplifier on ITMX table | (Rana, Kiwamu)
Last Friday we found a lonely RF amplifier ZHL-3A on the ITMX table.
When we found him we were very sad because he's been setup unacceptably
For example, the signal input was disconnected although a 24V DC was still applied. So he has been making just a heat for a long time.
The power connector was a BNC style which is not official way.
The leg of a decoupling capacitor attached to the DC connector was apparently broken and etc,..
We salvaged him and then cleaned up those cables and the DC power supply.
We don't say like 'don't make a temporary setup', but please clean up them after finishing the work every time. |
3776
|
Mon Oct 25 02:25:21 2010 |
Koji | Update | PSL | Quarter Wave Plate Measurements | Q1. Suppose the laser beam has a certain (i.e. arbitrary) polarization state but contains only TEM00. Also suppose the PSB is perfect (reflect all S and transmit all P). What results do you expect from your expereiment?
Q2. Suppose the above condition but the PBS is not perfect (i.e. reflects most of S but also small leakage of P to the reflection port.) How are the expected results modified?
Q3. In reality, the laser may also contain some thing dirty (e.g. deporarization in the laser Xtal, higher order modes in a certain polarization but different from the TEM00's one, etc). What actually is the cause of 170mW rejection from the PBS? Can we improve the transmitted power through the PBS?
Q4. Why is the visibility for the lambda/4 with 330deg better than the one with 326deg? Yes, as I already explained to Kevin, I suppose it was caused by the lack of the data points in the wider angle ranges.
Quote: |
I measured the reflected power from the PBS as a function of half wave plate rotation for five different quarter wave plate rotations.
The optimum angles that minimize the reflected power are 330° for the quarter wave plate and 268° for the half wave plate.
The following data was taken with 2.102 A laser current and 32.25° C crystal temperature.
For each of five quarter wave plate settings around the optimum value, I measured the reflected power from the PBS with an Ophir power meter. I measured the power as a function of half wave plate angle five times for each angle and averaged these values to calculate the mean and uncertainty for each of these angles. The Ophir started to drift when trying to measure relatively large amounts of power. (With approximately 1W reflected from the PBS, the power reading rapidly increased by several hundred mW.) So I could only take data near the minimum reflection accurately.
The data was fit to P = P0 + P1*sin^2(2pi/180*(t-t0)) with the angle t measured in degrees with the following results:
lambda/4 angle (°) |
t0 (°) |
P0 (mW) |
P1 (mW) |
chi^2/ndf |
V |
318 |
261.56 ± 0.02 |
224.9 ± 0.5 |
2016 ± 5 |
0.98 |
0.900 ± 0.001 |
326 |
266.07 ± 0.01 |
178.5 ± 0.4 |
1998 ± 5 |
16.00 |
0.918 ± 0.001 |
330 |
268.00 ± 0.01 |
168.2 ± 0.3 |
2119 ± 5 |
1.33 |
0.926 ± 0.001 |
334 |
270.07 ± 0.02 |
174.5 ± 0.4 |
2083 ± 5 |
1.53 |
0.923 ± 0.001 |
342 |
273.49 ± 0.02 |
226.8 ± 0.5 |
1966 ± 5 |
1.41 |
0.897 ± 0.001 |
where V is the visibility V = 1- P_max/P_min. These fits are shown in attachment 1. We would like to understand better why we can only reduce the reflected light to ~150 mW. Ideally, we would have V = 1. I will redo these measurements with a different power meter that can measure up to 2 W and take data over a full period of the reflected power. |
|
3777
|
Mon Oct 25 11:39:06 2010 |
Jenne | Update | SUS | Magnets glued to PRM | This morning I glued the magnets to the PRM. Now we wait, and tomorrow afternoon (at the earliest), Suresh and I can balance the PRM. |
Attachment 1: StatusTable.png
|
|
3778
|
Mon Oct 25 20:20:59 2010 |
yuta, Joe | Update | CDS | DAQ channel number etc | Summary:
We split the old SDSEN filters to SDSEN, SDSIDE, SDCOIL last week.
Along with this change, the TP channel number changed unfortunately.
So, we fixed them.
Also, we made FM9 do the output filter analog/digital switching.
What we did:
1. Changed the Simulink logic so that FM9 do the output filter switching, and checked the logic by probing
MAX333A for SDCOILs.
2. After making a new Simulink model and rebuilding, run the following incantation to burt restore filter
settings in /opt/rtcds/caltech/c1/target/c1SYS/c1SYSepics/ (See elog #3706)
sed -i 's/RO \(.*SW[12]R.*\)/\1/' autoBurt.req
3. DAQ channel numbers are listed in C1SYS.ini files in /cvs/cds/rtcds/caltech/c1/chans/daq/.
Channels with # signs are not activated. So, we changed, for example,
#[C1:SUS-MC1_LLSEN_IN1_DAQ]
#acquire=0
#datarate=16384
#datatype=4
#chnnum=10208
to
[C1:SUS-MC1_LLSEN_IN1_DAQ]
acquire=1
datarate=2048
datatype=4
chnnum=10208
for all channels we want to look at.
4. Restarted fb.
Plan:
- measure TFs and see if input and output filter switchings are working correctly
- make a switch that does all filter switching for all 5 OSEMS or 5 COILS
- put optical lever stuff
- fix offset sliders and offset switch
- put F2A filters in TO_COIL matrix (see elog #3769)
- make a nice graphical screen for MCs (like /cvs/cds/caltech/medm/c1/ioo/C1IOO_ModeCleaner.adl)
- write a script that activates DAQ we need
- make a plan
* SYS=SUS, RMS, MCS, etc |
3779
|
Mon Oct 25 23:10:06 2010 |
Koji | Update | IOO | MC is now flashing | [Suresh / Koji]
The MC mirrors are aligned. Now the flashing of the resonances are visible on the MC2 CCD
although the modematching seemed pretty poor.
- The incident power was adjusted to be ~20mW by rotating HWP after the laser source.
The power before the window of the chamber was ~450mW. Where are those missing 1.5W?
- We checked the spot on the last two steering mirrors and the incident beam on MC1.
The beam was too much off from the center of the 1st steering mirror. It was also hitting 1cm north of the MC1.
We adjusted the steering mirrors such that the incident and reflected beams are symmetrically visible at the MC1 tower.
- The MC mirrors are aligned. We first tried to use only MC2 and MC3. And then we used MC1 too as the spot on the MC2 was too high.
- We saw some TEM00 flashes but with many other modes flashing. We checked the beam diameter on the PSL table and on the MC REFL.
The latter one looked twice large as the former one. We concluded the beam is diverging.
- We closed the tank and decided to work on the mode matching tomorrow. |
3781
|
Tue Oct 26 00:45:08 2010 |
rana | Update | PSL | New NPRO Diagnostic Wiring | 
I copy section 6.2 into here to share with you all what the diagnostic capabilities of the new NPRO are. Its not a lot.
We'll need to record a sample of the NPRO output beam on a regular photodiode in order to get a real power monitor. My plan is to use a regular 25-pin Dsub and run it fron the NPRO controller over to the PSL rack and hijack the old MOPA monitoring channels (3113 and 3123 ADCs). |
3782
|
Tue Oct 26 01:53:21 2010 |
Joonho Lee | Update | Electronics | Fuction Generator removed. | Today I worked on how to measure cable impedance directly.
In order to measure the impedance in RF range, I used a function generator which could generate 50MHz signal and was initially connected to the table on the right of the decks.
The reason I am checking the cables is for replacing the cables with impedance of 50 or 52 ohm by those with impedance of 75 ohm.
After I figures out which cable has not proper impedance, I will make new cables and substitute them in order to match the impedance, which would lead to better VIDEO signal.
To test the VIDEO cables, I need a function generator generating signal of frequency 50 MHz.
In the deck on the right of PSL table, there was only one such generator which was connected to the table on the right of the deck.
Therefore, I disconnected it from the cable and took it to the control room to use it because Rana said it was not used.
Then, I tired to find on how to measure the impedance of cable directly but I did not finish yet.
When I finished today works, I put the generator back to the deck but I did not connect to the previous cable which was initially connected to the generator.
Next time, I will finish the practical method of measuring the cable impedance then I will measure the cables with unknown impedance.
Any suggestion would be appreciated. |
3786
|
Tue Oct 26 15:57:10 2010 |
Jenne | Update | SUS | One magnet broken, reglued | [Jenne, Suresh, Thanh (Bram's Grad Student)]
When we removed the grippers from the magnets on the PRM, one of the face magnets broke off. This time, the dumbbell remained glued to the optic, while the magnet came off. (Usually the magnet and dumbbell will stay attached, and both come off together). I had 3 spare magnet-dumbbells, but only one of them was the correct polarization. The strength of the spare magnet was ~128 Gauss, while the other magnets glued to the PRM are all ~180 Gauss. We considered this too large a discrepancy, and so elected to reuse the same magnet as before.
We removed the dumbbell from the optic using acetone. After the epoxy was gently removed, we drag wiped the AR face of the optic (Acetone followed by Iso, as usual), being careful to keep all the solvent away from all the other glue joints. We cleaned off the magnet with acetone (it didn't really have any glue stuck on it...most of the glue was stuck on the dumbbell), and epoxied it to a new dumbbell.
The PRM, as well as the magnet-dumbbell gluing fixture are in the little foil house, waiting for tomorrow's activities. Tomorrow we will re-glue this magnet to the optic, and Thursday we will balance the optic.
This still leaves us right on schedule for giving the PRM to Bob on Friday at lunchtime, so it can bake over the weekend. |
3788
|
Tue Oct 26 17:31:17 2010 |
yuta | Update | CDS | fixed OPLEV stuff and MCL filters | (Joe, Yuta)
Background:
We are currently working on getting rid of "white stuff" in MEDM screens.
Today, we fixed OPLEV stuff, MCL filters, and time stamps.
What we did:
1. Plugged in OPLEV cables to ADC2. (See this wiki page for wiring)
2. Connected ADC2 and OPLEV in Simulink model and fixed MEDM screens for OPLEVs (Attached #1).
3. Put MCL filters for BS,ITMX,ITMY,PRM,SRM.
They don't need them, but just for getting rid of "white stuff."
They are connected to the ground, so the outputs are always 0.
4. Fixed "TIME_STRING"s in MEDM screens so that they show current time correctly.
You only need to put text monitor with channel "C1:FEC-DCU_NODE_ID_TIME_STRING" into master files(DEFAULTNAME things) and run generate_master_screens.py.
It will automatically sets DCU ID correctly!! (Great work, Joe!)
See this wiki page for more info on making MEDM screens.
5. Checked OPLEV for MC2 by pointing a laser pointer to QPD. (For MC2, OPLEV is just a transmission beam position monitor)
Each quadrant looked like they are connected to the right channel numbers.
Plan:
- figure out what C1:SUS-NAME_MODE_SW1 does and fix
- fix Whitening, Dewhitening ON/OFF button in main MEDM screens, so that they switch every channels' filters
- make a new screen for MC (like the old one C1IOO_ModeCleaner.adl)
- create a new mark for new MEDM screens |
Attachment 1: optlevMC2.png
|
|
3789
|
Tue Oct 26 21:27:02 2010 |
Jenne | Update | CDS | fixed OPLEV stuff and MCL filters | Since MC2_TRANS is, in fact, MC2 Transmission, and not an oplev at all (it's not red, and it's not a lever, although it does use a QPD), I propose that the name be changed to something sensical, since calling it an oplev is completely non-sensical. The name change should happen sooner rather than later, to avoid confusion.
Quote: |
(Joe, Yuta)
Background:
Today, we fixed OPLEV stuff, MCL filters, and time stamps.
What we did:
5. Checked OPLEV for MC2 by pointing a laser pointer to QPD. (For MC2, OPLEV is just a transmission beam position monitor)
Each quadrant looked like they are connected to the right channel numbers.
|
|
3791
|
Wed Oct 27 02:26:15 2010 |
kiwamu | Update | IOO | MC locked | (Rana, Koji, Suresh, Yuta, Thanh, Kiwamu)
MC was locked successfully !
Instead of feeding back the signal to the MC length we just injected it to the NPRO pzt with a high voltage (HV) amplifier.
So now we can move on to an in-vac work which needs the main beam to align the stuff.
(mode matching to MC)
Suresh and Thanh (a visitor from ANU) improved the mode matching to the MC.
As written in the entry #3779 the beam after the mode matching lenses were diverging.
It is supposed to converge from 1.7mm radius at the last lens to 1.6 mm radius at the middle point of MC1 and MC2.
They slided the last lens toward the MC to make it more collimated and roughly measured the beam size using a sensor card.
As a fine tuning, they looked at some higher order modes showing up in the MC2 camera, and tried reducing the higher order modes by slightly sliding the last lens.
(assuming the lens position doesn't so much change the alignment)
During the work we removed a steering mirror for green locking because it was on the way of the lens slider.
- - measured optics' distances - -
25.5 cm from 1st lens to the front surface of the EOM
5.5 cm length of the EOM
24.5 cm from the front surface of the EOM to the 2nd lens (concave)
15.5 cm from the 2nd lens to the 1st steering mirror in the zig-zag path
20.5 cm from the steering mirror to the last lens
(preparation for locking)
Rana, Yuta and Koji prepared an old instant amplifier which can produce +/-13V output instead of usual SR560s.
We added an offset (~5V) on the signal to make it within 0-10V which is the input range of the HV amplifier.
If we take SR560, it's probably not sufficiently wide range because they can handle handle only about +/-4V.
We strung a cable from Marconi via the RF stabilizer to the wideband EOM in order to drive the EOM at 24.5MHz.
SInce the EOM doesn't have 50 Ohm input impedance we had to put a 50 Ohm load just before the EOM in order to drive it efficiently.
From a medm screen we set the driving RF amplitude slider (C1:IOO_MCRF_AMPADJ) to 0.0, which provides the maximum RF power on it.
(locking mode cleaner)
At first we unlocked the PMC to see an offset in the error signal without any lights on the MC_REFL PD.
Then we adjusted the offset to zero on the MC servo screen.
At the beginning of the locking the PMC was not stable for some reasons during the MC was locked.
But after increasing the laser power to the MC twice bigger, it looks like the PMC and the MC are quite stable.
|
3792
|
Wed Oct 27 09:02:29 2010 |
steve | Update | Electronics | HP4195A is NOT fixed | www.avalontest.com has fixed the 25MHz oscillation. Contact: Jim Burnham 760-536-0191
Actually, NOT FIXED. |
3794
|
Wed Oct 27 11:34:40 2010 |
josephb | Update | CDS | Modified rc.local for front end machines | What was the problem:
On reboot, c1sus was in a strange state. The epics IOC was running, along with tpman, but there were no loaded front ends.
People had to manually run sudo insmod c1SYSfe.ko.
What was the cause:
Awhile back Alex had commented out the line in the rc.local file which actually loaded the front end modules (i.e. c1x02fe.ko, c1susfe.ko, etc). This was while debugging. He never put it back.
What was the solution:
I uncommented the following line in /diskless/root/etc/rc.local file on the fb machine:
/opt/rtcds/caltech/c1/target/$i/scripts/startupC1rt
Now on reboot, the c1sus machine should start up all the front ends listed in the rtsystab file, located in /diskless/root/etc/ on the fb machine.
|
3795
|
Wed Oct 27 11:52:45 2010 |
josephb | Update | elog | Elog needed to be restarted | I had to restart the elog on Nodus because it was no longer responding. |
3796
|
Wed Oct 27 12:32:53 2010 |
josephb | Update | CDS | fb rebooted to try and fix testpoints | Problem:
Test points were unavailable last night, even after reboots of c1sus and even restarting the daqd process on the frame builder.
Cause:
Its unclear at this time. My guess is flaky fb and mx_stream codes. At the moment, the daqd often requires several restarts as it segfaults within a minute or two of restarting it.
What we did (aka treating the symptoms):
We rebooted the frame builder machine. I also added the daqd and nds processes to the inittab. Now when these die, they will automatically be restarted.
Steps to add to the inittab on fb
0) If not on fb, ssh -X fb
1) cd /etc/
2) sudo vi inittab or sudo emacs init
3) Add a line like: id:runlevels:action:process
The id is a unqiue 2-4 letter and number identifier for the process
Run levels is the run level of linux that it will start at. 345 will cover the normal cases
action is what to do with the process. Respawn makes it run at startup and also restarts it everytime it dies.
process is the command you want to run
See "man inittab" for more details
In this case we added
daq:345:respawn:/opt/rtcds/caltech/c1/target/fb/daqd -c /opt/rtcds/caltech/c1/target/fb/daqdrc > /opt/rtcds/caltech/c1/target/fb/daqd.log
nds:345:respawn:/opt/rtcds/caltech/c1/target/fb/nds pipe > /opt/rtcds/caltech/c1/target/fb/nds.log
4) Save.
5) Run "sudo /sbin/telinit q". This forces init to rexamine the inittab file
daqd and nds will now automatically restart when they die.
Continuing issues:
When the frame builder dies, the mx_stream processes on the front ends die as well. These need to be restarted manually at the moment by using "sudo /etc/restart_streams" while on c1sus.
The framebuilder code shouldn't be this flaky. |
3797
|
Wed Oct 27 15:38:19 2010 |
josephb, yuta | Update | CDS | IO chassis with bad timing was taken back to Downs | Problem:
The front end timing was not working properly for 2 of the IO chassis. They were not being synced to the 1 PPS signal.
This prevented the use of RFM for communication between front ends because time stamps on the transmitted data did not match the cycle on the receiving machine.
Action:
We took one of the incorrectly working chassis over to Downs. Rolf said he would take a look at it tomorrow morning.
Joe will be going over tomorrow morning to talk with Rolf and see what needs to be done to fix it.
|
3798
|
Wed Oct 27 16:15:35 2010 |
Suresh | Update | SUS | Re-glued magnet to the PRM | Thanh and I re-glued the magnet to the PRM following the procedure outlined by Jenne
The PRM in the gluing fixture has been placed in the little foil house and left to cure for a day.
If all goes well the balancing the PRM will be done tomorrow.
|
3799
|
Wed Oct 27 17:06:48 2010 |
josephb | Update | CDS | Moved c1iscey chassis and host interface board to c1ioo | Problem:
Need a working IO chassis connected to c1ioo in order to bring the MC_L into the digital realm, and then via RFM transmit to the c1sus machine.
Attempted Solution:
Move the c1iscey IO chassis to c1ioo while the c1ioo chassis is at downs.
The c1iscey chassis however doesn't seem to be talking to the c1ioo computer. I tried changing the host interface card on the c1ioo chassis. I took out One Stop Systems HIB2-x4-H interface card with serial number 26638 from the c1ioo computer and put in the One Stop Systems HIB2-x4-H with serial number 35242 in from c1iscey into c1ioo. Still didn't work.
All the lights are red on the interface card on the actual chassis and its cooling fan isn't spinning.
Using dmesg on c1ioo shows that it does not see any of the ADC/DAC/BO cards.
Status:
I'm going to wait until tomorrow morning when Rolf gets a chance to look at the c1ioo chassis over at Downs to determine the next step. If we fix the c1ioo chassis, I'm move the c1iscey chassis and its host interface board back to the end. |
3800
|
Wed Oct 27 21:33:42 2010 |
Suresh | Update | Locking | Green from the far end re-obtained |
The mirror which was moved during the mode matching of PSL light to the MC (ref elog #3791) has been repositioned. We once again have the green light from the NPRO on the X (south) arm available on the PSL table.
This light was supposed to be collimated by the two plano convex lenses (f=200mm and f=50mm ref to elog #3771) but it was converging. So I moved the f=50mm lens backwards to make the beam collimated. I checked the beam collimation by introducing an Al coated mirror infront of th PD and diverting the beam temporarily in a free direction. I could then check the collinearity and collimation of both the green beams over a meter. After alignment the mirror was removed and the light is now incident on the PD once again. We can now proceed to look for green beats.
The power from the PSL NPRO was attenuated for the MC locking work of yesterday. It has now been increased to the maximum by rotating the Half Wave Plate (HWP). The power after the PSL is now about 450mW (500mW - 10% picked off for the doubling).
|
3801
|
Wed Oct 27 21:51:29 2010 |
Suresh | Update | PSL | Attenuation of PSL NPRO removed | The laser power was attenuated to 40 mW yesterday for ensuring that the power built up within the MC does not damage the optics.
This however stopped us from the doubling work and besides also reduced the power available for locking the PMC.
Therefore, today the laser attenuation was removed and once again 500mW is available at the exit of the PMC .
However the power sent to the MC has been reduced to 60mW by changing one of the mirrors in the zig-zag to a 33% beam splitter. Though about 450mW is incident on the beam splitter the reflected beam is only about 55mW since the mirror reflectance is specified for P polarised light incident at 45deg while ours is S-polarised incident at less than 45deg. The light transmitted through the beam splitter has been blocked by a beam dump.
|
3802
|
Thu Oct 28 02:01:51 2010 |
Kevin | Update | PSL | Filter for 2W Laser | [Rana and Kevin]
I made a low pass filter for the piezo driver for the 2W laser that is now installed. The filter has a pole at 2.9 Hz. The transfer function is shown in attachment 1.
Attachment 2 shows the outside of the filter with the circuit diagram and attachment 2 shows the inside of the filter. |
Attachment 1: tf.PDF
|
|
Attachment 2: outside.jpg
|
|
Attachment 3: inside.jpg
|
|
3803
|
Thu Oct 28 03:07:53 2010 |
kiwamu | Update | Green Locking | 80MHz VCO for green PLL : a health check | I did a health check for a 80MHz VCO box.
I started taking care with the black VCO box, which has been sitting on the SP table and will be used for converting the green beat signal from frequency to voltage.
The circuit in the box basically consists of three parts: low pass filters (LPFs), a VCO and RF amplifiers.
Today I checked the LPF stage. It looks pretty healthy.
Tomorrow I will check the VCO part, especially I am curious about the VCO range.
(soldering)
Since somebody ( surf students ?) removed some resistors, the VCO was just freely running without being applied any voltage.
I put some resistors back on the circuit board by soldering them.
Now the resistors are placed in the same configuration as the original schematic (link to LIGO DCC) except for the wideband signal path, which has a differential input.
I left the wideband path disconnected from the VCO.
(transfer function measurement)
The LPF part in 'external mod' path contains two stages in series:
one is for cutting off demodulated signals above fc=80MHz and the other one is for PLL servo with pole=1Hz, zero=40Hz.
In order to activate this path I shorted 10th pin of the analog switch: MAX333A.
During the transfer function measurement I injected signals to 'external mod' input and took the output signal from a test point pin TP7.
The plot below shows a fitting result of the measured transfer function of the whole LPF stage. I used liso for the fitting.
The measured filter's shape agreed with the design. (though I haven't checked 80MHz cut off)

|
3804
|
Thu Oct 28 03:21:35 2010 |
Suresh | Update | Locking | Aligned the MC2 transmission photodiode | Yuta and Suresh
The MC2 transmission is seen on the QPD
Once the laser was locked to the cavity, and the PMC was able to follow the laser (ref: elogs by Yuta and Rana today) we had a steady TEMoo mode in the MC. This gave us sufficient transmission through MC2 to be easily visible with an IR viewer and we were able to guide the beam on to the QPD. The beam however seemed to over fill the QPD, a lens (f=180mm) was placed before the beam folding mirror and the spot sized reduced. The the QPD was found to be not fixed to the table and this was also recitified. The signal level is still low: total signal is just about 7 DAQ steps amounting to about 5mV. Tomorrow we plan to work on the alignment of the PSL and MC and thus increase this signal.
A new channel to observe the length variations in the MC.
A long BNC cable was laid from the MC locking electronics next (southwards) to the PSL table to the DAQ cards picking up the signals from the PRM OSEMS. This is to pick up one of the MC locking signals and collect it on a DAQ channel. However as there are no spare DAQ channels currently available one of the PRM OSEM (UL and LL) photodiode channels was unplugged and replaced with the signal from the long BNC cable. For identifying the correct DAQ channel we put in a 2 Vpp 10Hz signal with a function generator into this BNC. Tow signals can be picked up in this fashion and they are available on PRM_LLSEN_IN1 and PRM_ULSEN_IN1. We plan to use this for improving the alignment of the MC tomorrow.
The algorithm for this alignment is that if the beam from the PSL is not centered on the MC1 then tilting MC1 would result in a change in the length of the cavity as seen by the light. Using this as feedback the spot could be precisely centered on the MC1 and then the MC alignment could be completed by moving MC2 and MC3 to reobtain TEM_oo within the cavity. |
3805
|
Thu Oct 28 03:39:58 2010 |
yuta | Update | IOO | got very stable MC locking | (Rana, Suresh, Jenne, Kiwamu, Kevin, Yuta)
Summary:
Last night we locked MC by feeding back the signal to NPRO PZT.
But it was not so stable.
We wanted more gain to lower the seismic motion, but we don't need high gain in high frequency part(>~1kHz) because it may cause something bad(NRPO PZT oscilatting, PMC not able to catch up with the NPRO frequency change, etc).
So, we put DC gain boost today.
It successfully made MC locking stable!
What we did:
1. Lowered the main laser temperature from 32.2° C to 31.8° C.
When we increased the laser temperature, PMC transmission get lower. 32.2° C was on the cliff, so we put it to plateau region.
2. Lowered the gain of PMC servo (2dB instead of 8dB last night), because PMC was oscillating.
We got 5.3V OMC transmission.
3. Made 3Hz pole, 30Hz zero filter and put in NPRO PZT servo loop.
0dB at DC, -20dB at high frequency (see Kevin's elog #3802)
4. Put more gain to NPRO PZT servo loop to compensate -20dB at high frequency.
In a word, we put DC gain boost.
Attachment #1 is the MEDM screen screenshot for MC servo.
5. Aligned the beam into QPD at MC2 trans.
We put lens in front of the QPD.
Now, we can see the actual motion of the beam, and resonance peaks(Attachment #2; not locked at the highest).
(We added 30Hz LPF after each 4 quadrant inputs to reduce noise)
Plan:
- optimize MC suspension alignments
- activate OL DAQ channels
- reduce RFAM
- install tri-mod box
- QPD signal at MC2 should be more high(currently, ~7counts which equals to ~4mV)
- change temperature of X-end laser to get green beat |
Attachment 1: C1IOO_MC_SERVO20101028.png
|
|
Attachment 2: MCrocks!.png
|
|
3806
|
Thu Oct 28 04:23:38 2010 |
rana | Update | IOO | A2L prep | To get the angle to length signal before the c1ioo processor gets going, we need a length signal. We can use either the error signal or the control signal.
I recommend using the control signal since its not puny. The 4-pin LEMO inputs to the OSEM ADC that Suresh has wired are differential so we can, in principle, use either the BNC output of the SERVO plug or the 2-pin LEMO output.
The analog whitening on the OSEM Whitening board should be engaged via the SUS MEDM screen so that we get a good SNR at the A2L dither frequencies.
If the ADC saturates, then we should use a pomona box RC low pass to cut everything off above 100 Hz.
Also, a comment about Yuta's elog: we estimated that the seismic motion was ~1e-7 - 1e-6 meters. The MC linewidth ought to be ~lambda/(2*Finesse) ~ 1e-9.
So, the MC servo as it was was not giving us enough gain (1/f above 50 Hz; UGF ~5-10 kHz) to get the error signal to stay in the linear PDH region. Kevin's filter gave us ~10x more gain at the seismic frequencies (1-3 Hz) of concern. |
3807
|
Thu Oct 28 04:28:50 2010 |
yuta | Update | Green Locking | checked frequency counter SR620 | (Kiwamu, Yuta)
Background:
For green locking, we are planning to feedback frequency differential signal to ETM suspension for the final configuration.
We don't have ETM suspension control system right now, so we are going to feedback the signal to X-end laser frequency for a test.
We have two loops for the servo;
1. coarse locking using frequency counter, feeding back to the laser temperature
2. using VCO, feeding back to the laser PZT
Today, we checked frequency counter SR620 and see how to get the small beat note signal(-48dBm; see elog #3771).
What we did:
1. Using Marconi(RF signal generator), put RF signals to SR620 and see how small signal SR620 can see.
It depends on the frequency. For 80MHz signal, you need more than about -9dBm.
60MHz >-12dBm
70MHz >-10dBm
80MHz >-9dBm
90MHz >-8dBm
100MHz >-7dBm
Since we are going to lock the frequency difference between X-end and PSL to 80MHz, we need at least +40dBm amp before putting the signal into SR620.
RF amplifier ZHL-32A has around +28dBm +28dB gain at 80MHz, so we need 2 of them.
2. Marconi -> ZHL-32A -> ZHL-32A -> SR620 and see how small 80MHz signal SR620 can see.
Around -68dBm. This should be enough.
3. SR620 has "STRIP CHART" output on the rear panel. The output voltage is proportional to the mean frequency of the input.
The output range is 0-8V. So in order to get 4V for 80MHz, set SCALE to 20MHz.
Plan:
- find green beat again and see if SR620 can see it with double ZHL-32A configuration |
3808
|
Thu Oct 28 09:16:17 2010 |
steve | Update | PEM | welding in the vertex area | Fire-smoke sensors in the vertex area #2-31, 2-30 east, 2-32 south/MC2 and 2-37 old control room area are turned off to accommodate the welding
activity of folding crane. These sensors will be reactivated at 3:30pm today.
Stay out of the 40m lab: IFO room till 6 pm today. |
3809
|
Thu Oct 28 11:54:31 2010 |
Koji | Update | Green Locking | checked frequency counter SR620 | ZHL-32A is a high power (well..., actually middle power) amplifier with the max output power of +29dBm (~1W!).
It seems to be overkill.
Your signal is so small so you don't need ZHL-32A, but can use small amp which we may have somewhere in the lab.
And the description:
"RF amplifier ZHL-32A has around +28dBm gain at 80MHz"
The unit is wrong.
Quote: |
(Kiwamu, Yuta)
Background:
For green locking, we are planning to feedback frequency differential signal to ETM suspension for the final configuration.
We don't have ETM suspension control system right now, so we are going to feedback the signal to X-end laser frequency for a test.
We have two loops for the servo;
1. coarse locking using frequency counter, feeding back to the laser temperature
2. using VCO, feeding back to the laser PZT
Today, we checked frequency counter SR620 and see how to get the small beat note signal(-48dBm; see elog #3771).
What we did:
1. Using Marconi(RF signal generator), put RF signals to SR620 and see how small signal SR620 can see.
It depends on the frequency. For 80MHz signal, you need more than about -9dBm.
60MHz >-12dBm
70MHz >-10dBm
80MHz >-9dBm
90MHz >-8dBm
100MHz >-7dBm
Since we are going to lock the frequency difference between X-end and PSL to 80MHz, we need at least +40dBm amp before putting the signal into SR620.
RF amplifier ZHL-32A has around +28dBm gain at 80MHz, so we need 2 of them.
2. Marconi -> ZHL-32A -> ZHL-32A -> SR620 and see how small 80MHz signal SR620 can see.
Around -68dBm. This should be enough.
3. SR620 has "STRIP CHART" output on the rear panel. The output voltage is proportional to the mean frequency of the input.
The output range is 0-8V. So in order to get 4V for 80MHz, set SCALE to 20MHz.
Plan:
- find green beat again and see if SR620 can see it with double ZHL-32A configuration
|
|
3810
|
Thu Oct 28 13:01:34 2010 |
steve | Update | PEM | crane repair guys left for the day |
Quote: |
Fire-smoke sensors in the vertex area #2-31, 2-30 east, 2-32 south/MC2 and 2-37 old control room area are turned off to accommodate the welding
activity of folding crane. These sensors will be reactivated at 3:30pm today.
Stay out of the 40m lab: IFO room till 6 pm today.
|
The new cord wheel was to big. They be back tomorrow?
The lab is open. |
3811
|
Thu Oct 28 16:38:54 2010 |
josephb | Update | CDS | Flaky fb, reverted inittab changes on fb | Problem:
Yuta reported many of the signals being displayed by dataviewer "fuzzier" than normal. And diaggui was not working.
Running "diag -i" reported:
Diagnostics configuration:
awg 21 0 192.168.113.85 822095893 1 192.168.113.85
awg 36 0 192.168.113.85 822095908 1 192.168.113.85
awg 37 0 192.168.113.85 822095909 1 192.168.113.85
tp 21 0 192.168.113.85 822091797 1 192.168.113.85
tp 36 0 192.168.113.85 822091812 1 192.168.113.85
tp 37 0 192.168.113.85 822091813 1 192.168.113.85
This seems to be missing an nds type line between the 3 awgs and the 3 tp lines.
The daqd code (the framebuilder) is being especially flaky today, and I'm starting to see new errors.
[Thu Oct 28 16:13:46 2010] Couldn't open full trend frame file
`/frames/trend/second/9723/C-
T-972342780-60.gwf' for writing; errno 13
epicsThreadOnceOsd epicsMutexLock failed.
epicsThreadOnceOsd epicsMutexLock failed.
epicsThreadOnceOsd epicsMutexLock failed.
epicsThreadOnceOsd epicsMutexLock failed.
epicsThreadOnceOsd epicsMutexLock failed.
Segmentation fault (core dumped)
or
[Thu Oct 28 16:17:06 2010] Couldn't open full frame file
`/frames/full/9723/.C-R-972343024-16.gwf' for writing; errno 13
CA client library tcp receive thread terminating due to a C++ exception
FATAL: exception not rethrown
CA client library tcp receive thread terminating due to a C++ exception
CA client library tcp receive thread terminating due to a C++ exception
FATAL: exception not rethrown
cac: tcp send thread received an unexpected exception - disconnecting
Aborted (core dumped)
What was done today that might have affected it:
A new c1ioo chassis from Downs was connected to c1ioo. I also connected c1ioo to the DAQ network (192.168.114.xxx) which talks to the frame builder.
I started downloading the necessary files to be able to follow Keith's instructions for a standard control / teststand setup in /opt/apps , /opt/rtapps, etc. However, it has not actually been installed yet.
Yuta added additional OL channels to the DAQ config for being recorded.
Attempted Fixes:
I reverted the inittab changes I made in this elog. Didn't help.
I disconnected c1ioo from the DAQ network. Didn't help.
Rebooted the frame builder machine. Didn't help.
I've sent an e-mail to Alex describing the problem to see if he has any idea where we went wrong.
Yuta may try restoring the old DAQ channel choices and see if that makes a difference.
Current Status:
daqd framebuilder code still won't stay up. So no channels at the moment. |
3812
|
Thu Oct 28 19:10:26 2010 |
tara | Update | Electronics | TTFSS for 40m | I keep a set of new TTFSS for 40m in electronic cabinet along the North arm.
The set number is #6. It is working and has not been modified by me.
Other two sets,# 5 and #7, are kept at PSL lab. |
3813
|
Thu Oct 28 20:08:26 2010 |
yuta | Update | Green Locking | revised RF amp cascading | Background:
Yesterday, I said I will use ZHL-32A for amplifying beat note signal, but as Koji pointed out, ZHL-32A is for medium high power.
So I changed my mind to use ZFL-1000LN instead. ZFL-1000LN is a low noise RF amp whose maximum power is 3dBm.
Also, we included a splitter in our consideration.
What I did:
1. Set up a new setup. ZFL-1000LN has a gain of +24dB at 80MHz and splitter ZFRSC-42 has a loss of -6dB. So;
beat note signal -> ZFL-1000LN -> ZFL-1000LN -> ZFRSC-42 => SR620 and VCO
2. Measured frequency-output relation. When the input signal was 80MHz -48dBm, the output was -8.7dBm. For other frequencies;
60MHz -3.3dBm
70MHz -5.7dBm
80MHz -8.7dBm
90MHz -5.5dBm
100MHz -3.5dBm
So, we can see frequency up to >100MHz by SR620 using this setup.
3. Checked harmonics peak levels of the output using an RF spectrum analyzer. When the input signal was 80MHz -48dBm, the height of the peaks were;
80MHz -8.8dBm
160MHz -30dBm
240MHz -42dBm
Other peaks were smaller than the 3rd harmonics. Also, RF PD that detects beat note signal has a cut-off frequency at around 100MHz. So, we don't need to worry about wave transformation for this setup.
Quote: |
ZHL-32A is a high power (well..., actually middle power) amplifier with the max output power of +29dBm (~1W!).
It seems to be overkill.
Your signal is so small so you don't need ZHL-32A, but can use small amp which we may have somewhere in the lab.
And the description:
"RF amplifier ZHL-32A has around +28dBm gain at 80MHz"
The unit is wrong.
|
Yes. I corrected my previous elog. |
|