  15582   Sat Sep 19 18:07:35 2020 KojiUpdateVACTP3 RP failureI came to the campus and Gautam notified
that he just had received the alert from
the vac watchdog.
 vac.pngScreen_Shot_2020-09-19_at_17.52.40.png 
  15584   Sat Sep 19 18:46:48 2020 KojiUpdateGeneralHand soapI supplied a bottle of hand soap. Don't
put water in the bottle to dilute it as it
makes the soap vulnarable for cotamination.
  
  15585   Sat Sep 19 19:14:59 2020 KojiUpdateGeneralITMX released / ETMY UR magnet knocked offThere were two SUSs which didn't look
normal.

- ITMX was easily released by the
 IMG_6218.jpeg 
  15595   Tue Sep 22 16:29:30 2020 KojiUpdateGeneralControl Room AC setting for continuous runningI came to the lab. The control room AC
was off -> Now it is on.

Here is the setting of the AC meant
 P_20200922_161125.jpg 
  15597   Tue Sep 22 23:16:54 2020 KojiUpdateGeneralHEPA InspectionGautam reported that the PSL HEPA stopped
running (ELOG
15592). So I came in today and started
 11x 
  15598   Tue Sep 22 23:17:51 2020 KojiUpdateGeneralHEPA InspectionDimensions / Specs

- HEPA unit dimentions
- HEPA unit manufacturer
 6x 
  15600   Wed Sep 23 10:06:52 2020 KojiUpdateVACTP2 running HOTHere is the timeline. This suggests TP2
backing RP failure.

1st line: TP2 foreline pressure
 Screen_Shot_2020-09-23_at_10.00.43.png 
  15612   Mon Oct 5 00:53:16 2020 KojiUpdateBHDSingle bounce interferometer locked🤘🤘🤘

 
  
  15616   Wed Oct 7 13:06:27 2020 KojiUpdateGeneralPresence in the labTue evening from 4pm~6pm, Koji made a social
distant tour for Anchal. We were present
around the PSL/AS/ETMX tables.
  
  15621   Thu Oct 8 18:40:42 2020 KojiUpdateComputer Scripts / ProgramsFinesse GUIIs it better than Luxor? https://labcit.ligo.caltech.edu/~jharms/luxor.html   
  15625   Wed Oct 14 13:28:04 2020 KojiUpdateCOCITM/ETM spares in DownsThe two ITM spares and two ETM spares are
together stored in the optic storage (B110)
at Downs. c/o Liyuan and GariLynn
 IMG_3073.jpeg 
  15630   Thu Oct 15 20:00:23 2020 KojiSummaryGeneralHEPA AC cord replacementThe AC cord from the PSL HEPA variac
to the junction box was replaced.
Now the HEPA is running at 70%
 P_20201015_200732.jpgP_20201015_200752.jpgP_20201015_202615.jpgP_20201015_204234.jpg 
  15633   Mon Oct 19 15:38:42 2020 KojiUpdateElectronicsLoan: A file binder "40m wiring diagram" I'll bring a file binder "40m
wiring diagram" to home at the next
chance.
  
  15634   Mon Oct 19 15:40:02 2020 KojiUpdatePEMAlaska EQ M7.5Alaska M7.5 20:54UTC https://earthquake.usgs.gov/earthquakes/eventpage/us6000c9hg/executive

I looked at the suspensions. The
  
  15635   Tue Oct 20 20:12:18 2020 KojiSummaryGeneralDJI OSMO Pocket Camera KitI set up an action cam (DJI OSMO Pocket)
and brought it back to the 40m. The kit is
now placed in the control room cabinet together
 20201020200929_IMG_0173.JPG 
  15641   Fri Oct 23 16:41:06 2020 KojiUpdateIOOExcess laser freq noise investigation[Koji, Rana]

We wanted to track down the excess
noise seen in MC_F and other places (see
 plot.pdf 
  15642   Fri Oct 23 19:01:57 2020 KojiSummaryPEMPSL Particle Counter kit removed from the tableThe particle counter on the 40m PSL was
removed. The package was made together with
the OMC lab particle counter (see the packing
 P_20201022_173529.jpgP_20201022_173419.jpg 
  15643   Mon Oct 26 13:35:58 2020 KojiUpdateIOOExcess laser freq noise investigationIn fact, the problem was the grounding
issue (presumably on the IOO racks).
A temporary differential receiver
 6x 
  15646   Wed Oct 28 09:35:00 2020 KojiUpdateCDSRFM errorsI'm starting the model restarts from
remote. Then later I'll show up in the
lab to do more hard resets.
 Screen_Shot_2020-10-28_at_10.06.00.png 
  15649   Wed Oct 28 20:42:53 2020 KojiUpdateGeneralVent Prep
Two arms / BS / PRM / SRM were
aligned. (Attachment 1)
IMC
was aligned by WFS and the WFS offsets were
 10x 
  15655   Mon Nov 2 17:13:19 2020 KojiUpdateGeneralPlan for OMC chamberI believe the mirror next to IM1 is for
the green beams to be delivered to the PSL
table. I think we still want to keep it.
  
  15665   Sun Nov 8 23:24:29 2020 KojiUpdateGeneralVent activities today> I didn't bother to align the green
beams to the arm cavities or re-center the
Oplevs - is this necessary? It is a step
  
  15667   Tue Nov 10 11:31:13 2020 KojiUpdateGeneralPumpdownMain volume pressure as of 11:30AM 2020/11/10  Screen_Shot_2020-11-10_at_11.30.21.png 
  15676   Thu Nov 12 15:40:42 2020 KojiUpdateElectronicsMore systematic noise characterizationYes. The datasheet has a recommendation
circuit with 10uF caps. Companies are careful
to show reproducible, reliably functional
  
  15685   Mon Nov 23 14:52:10 2020 KojiUpdateBHDBHD MMT Optics deliveredBasically, they repeated our specs and
showed the coating performances for HR/AR
for 10deg P and PR/AR for 45deg P. There
  
  15687   Mon Nov 23 23:27:43 2020 KojiSummaryASCQ3000 characterizationLast week and this week I've been working
on the characterization of the Q3000 QPDs.
The QPDs were named 81, 82, 83, and 94.
  
  15702   Thu Dec 3 13:55:52 2020 KojiUpdateASCElectrical LO signal for AS WFSI got a bit confused by your description.

The demod board claims
that the nominal power at each LO port is
  
  15705   Thu Dec 3 21:14:38 2020 KojiUpdateASCElectrical LO signal for AS WFSLet's use RG405 for better shielding.
It is not too stiff. The bending  (just
once) does not break the cable.
  
  15707   Fri Dec 4 03:02:40 2020 KojiUpdateASCElectrical LO signal for AS WFS1. That's true. But we are already
in that regime with the Var attn at 0dB,
aren't we? We can reduce the input to
  
  15708   Fri Dec 4 15:58:22 2020 KojiUpdateIOOWFS2 broken I checked the backplane connection for
IMC WFS2  and found that the
cables for IMC WFS2 and the IMC demod were
  
  15709   Fri Dec 4 19:23:40 2020 KojiUpdateElectronicsAA/AI board testing ongoingI have the setup built for the AA/AI board
testing around the PD testing area. Please
let me leave it like that for a week or so.
  
  15740   Sat Dec 19 02:42:56 2020 KojiUpdate New SMA cables on orderOur favorite (flexible) RF cable is Belden's
1671J (Jacketed solder-soaked coax cable).
It is compatible RG405. I'm not sure
  
  15747   Sun Jan 3 16:26:06 2021 KojiUpdateSUSIMC WFS check (Yet another round of Sat. Box. switcharoo)I wanted to check the functionality of
the IMC WFS. I just turned on the WFS servo
loops as they were. For the past two hours,
 Screen_Shot_2021-01-03_at_17.14.57.png 
  15753   Thu Jan 7 20:07:27 2021 KojiUpdateALSNoisy ALSHow about resurrecting the PSL table green
beat for the X arm to see if the non-fiber
setup shows the same level of the freq noise
  
  15755   Thu Jan 7 23:25:19 2021 KojiUpdateALSNoisy ALSIf the sensing noise level of the end PDH
degraded for some reason, it'd make the
out of loop stability worse without making
  
  15790   Tue Feb 2 18:24:54 2021 KojiUpdateBHDSOS assemblyYou can remove the components of the optical
table enclosure (black ones) and use the
optical table as your working area too.
  
  15791   Tue Feb 2 23:29:35 2021 KojiUpdateCDSCDS crash and CDS/IFO recoveryI worked around the racks and the feedthru
flanges this afternoon and evening. This
inevitably crashed c1lsc real-time process.
  
  15794   Wed Feb 3 18:53:31 2021 KojiUpdateCDSCDS crash and CDS/IFO recoveryReally!? I didn't reboot the machines
between "sudo date" and "rtcds
start c1x0*". I tried rtcds. If it didn't
  
  15801   Wed Feb 10 17:18:03 2021 KojiSummaryBHDSatellite Amplifier Output Offset measurementsTesting the satellite amp i.e. 
PD driver
- To test the noise of the PD transimpedance
  
  15810   Tue Feb 16 15:29:01 2021 KojiUpdateBHDSOS assemblyThe curie temp of SmCo seems about x2 (in
K) of the one for NdFeB. i.e. 600K vs 1000K.
So I believe 177degC = 450K is not an issue.
  
  15813   Wed Feb 17 13:59:43 2021 KojiUpdateSUSCoM on 3"->2" Adapter Ring for SOSNote from today's meeting:

1. Can we adjust the thickness of
  
  15815   Thu Feb 18 03:20:09 2021 KojiSummaryElectronicsCurrent Rack MapFor your planning:  rack_plan.pdf 
  15819   Thu Feb 18 20:20:25 2021 KojiUpdateSUSaLIGO Sat Amp characterizationYeah, it's really inconsistent. You
had 35mA LED drive and the current noise
of the noisy channel was 5e-7 A/rtHz at 1Hz.
  
  15820   Thu Feb 18 20:24:48 2021 KojiSummaryElectronicsA bunch of electronics receivedTodd provided us a bunch of electronics.
I went to Downs to pick them up this afternoon
and checked the contents in the box. Basically,
 IMG_0416.jpeg 
  15824   Fri Feb 19 16:06:01 2021 KojiUpdateSUSCoM Range on 3"->2" Adapter Ring for SOSWe want to move the CoM with the adjustment
range so that the residual deviation is adjusted
by the bottom dumbbell. 0.0003" is well
  
  15826   Fri Feb 19 16:55:26 2021 KojiUpdateSUSCoM Range on 3"->2" Adapter Ring for SOSJordan's screenshot actually shows
that the vertical distance (Y) is 0.0000".
We want to have the vertical distance of
  
  15830   Sat Feb 20 16:46:17 2021 KojiSummaryElectronicsA bunch of electronics receivedWe received currently available sets. We
are supposed to receive more coil drivers
and sat amps, etc. But they are not ready
  
  15835   Tue Feb 23 20:55:19 2021 KojiUpdateSUSOSEM testing for SOSsI can't obtain a consistent view between
the existing drawings/photographs and your
pin assignment. Please review the pin assignment
 SOS_OSEM.pdf 
  15836   Tue Feb 23 23:12:37 2021 KojiSummarySUSSUS invacuum wiringThis is my current understanding
of the in-vacuum wiring:
1. Facts
 SOS_OSEM_cabling.pdf 
  15847   Fri Feb 26 20:20:43 2021 KojiUpdateElectronicsProduction version of the HV coil driver tested with KEPCO HV suppliesThis is very disappointing. Even with KEPCO
linear supply with the improved HV driver
circuit, the noise level is significantly
  