    Reply  Mon Oct 5 00:53:16 2020, Koji, Update, BHD, Single bounce interferometer locked 
🤘🤘🤘

 
Entry  Wed Oct 7 13:06:27 2020, Koji, Update, General, Presence in the lab 
Tue evening from 4pm~6pm, Koji made a social distant tour for Anchal. We were present around the PSL/AS/ETMX tables.
    Reply  Thu Oct 8 18:40:42 2020, Koji, Update, Computer Scripts / Programs, Finesse GUI 
Is it better than Luxor? https://labcit.ligo.caltech.edu/~jharms/luxor.html
Entry  Wed Oct 14 13:28:04 2020, Koji, Update, COC, ITM/ETM spares in Downs IMG_3073.jpeg
The two ITM spares and two ETM spares are together stored in the optic storage (B110) at Downs. c/o Liyuan and GariLynn
Entry  Thu Oct 15 20:00:23 2020, Koji, Summary, General, HEPA AC cord replacement P_20201015_200732.jpgP_20201015_200752.jpgP_20201015_202615.jpgP_20201015_204234.jpg
The AC cord from the PSL HEPA variac to the junction box was replaced.
Now the HEPA is running at 70%


Showed up at the 40m at 7pm
Entry  Mon Oct 19 15:38:42 2020, Koji, Update, Electronics, Loan: A file binder "40m wiring diagram"  
I'll bring a file binder "40m wiring diagram" to home at the next chance.
There is another one on the shelf in the control room.

(I thought I put it in my bag, but it looks like that I left it somewhere around the fax area)
Entry  Mon Oct 19 15:40:02 2020, Koji, Update, PEM, Alaska EQ M7.5 
Alaska M7.5 20:54UTC https://earthquake.usgs.gov/earthquakes/eventpage/us6000c9hg/executive

I looked at the suspensions. The watchdogs have not been tripped.

IMC was locked but continually shaken. (and occasional unlock)
Entry  Tue Oct 20 20:12:18 2020, Koji, Summary, General, DJI OSMO Pocket Camera Kit 20201020200929_IMG_0173.JPG
I set up an action cam (DJI OSMO Pocket) and brought it back to the 40m. The kit is now placed in the control room cabinet together with the Canon DSLR.

I might have left the USBC chaging cable at home this time. Will bring it back next time.-> The cable was returned to the kit on Oct 23rd.
Entry  Fri Oct 23 16:41:06 2020, Koji, Update, IOO, Excess laser freq noise investigation plot.pdf
[Koji, Rana]

We wanted to track down the excess noise seen in MC_F and other places (see the previous
report by Gautam)
Entry  Fri Oct 23 19:01:57 2020, Koji, Summary, PEM, PSL Particle Counter kit removed from the table P_20201022_173529.jpgP_20201022_173419.jpg
The particle counter on the 40m PSL was removed. The package was made together with the OMC lab particle counter (see the packing list below).

The kit was picked up by Radhika for a python code to read out the numbers.
    Reply  Mon Oct 26 13:35:58 2020, Koji, Update, IOO, Excess laser freq noise investigation 6x
In fact, the problem was the grounding issue (presumably on the IOO racks).
A temporary differential receiver at the TTFSS side was built using an SR560 and a few ponoma cables. This removed the structures ~850Hz.


The MC Servo Output was disconnected from the TTFSS box and monitored with SR785. The 850Hz structure was kept visible no matter what
    Reply  Wed Oct 28 09:35:00 2020, Koji, Update, CDS, RFM errors Screen_Shot_2020-10-28_at_10.06.00.png
I'm starting the model restarts from remote. Then later I'll show up in the lab to do more hard resets.
==> It seems that the RFM errors are gone. Here
are the steps.
Entry  Wed Oct 28 20:42:53 2020, Koji, Update, General, Vent Prep 10x

Two arms / BS / PRM / SRM were aligned. (Attachment 1)
IMC was aligned by WFS and the WFS offsets were offloaded.
Suspension
Status Snapshot (Attachment 2)
Oplevs are aligned (Attachment 3)
Xarm green was aligned in the
daytime. Xarm green refl DC (C1:ALS-X_REFL_DC_OUTPUT) was 620 (aligned) ~1300 (drifted). When unlocked, it was 3750.
Yarm green: I saw no flash.
    Reply  Mon Nov 2 17:13:19 2020, Koji, Update, General, Plan for OMC chamber 
I believe the mirror next to IM1 is for the green beams to be delivered to the PSL table. I think we still want to keep it. Otherwise, the plan looks
fine.
    Reply  Sun Nov 8 23:24:29 2020, Koji, Update, General, Vent activities today 
> I didn't bother to align the green beams to the arm cavities or re-center the Oplevs - is this necessary? It is a step in the pre-close up checklist,
so maybe we should do it... The green transmission does reach the PSL table...

I don't think so. The beam is reaching the PSL, so we have no motivation to change the green alignment. Regarding the oplev, the green refl
    Reply  Tue Nov 10 11:31:13 2020, Koji, Update, General, Pumpdown Screen_Shot_2020-11-10_at_11.30.21.png
Main volume pressure as of 11:30AM 2020/11/10
    Reply  Thu Nov 12 15:40:42 2020, Koji, Update, Electronics, More systematic noise characterization 
Yes. The datasheet has a recommendation circuit with 10uF caps. Companies are careful to show reproducible, reliably functional circuit examples on datasheets.
So, if the caps are there you should try to replicate the design.




Quote:


    Reply  Mon Nov 23 14:52:10 2020, Koji, Update, BHD, BHD MMT Optics delivered 
Basically, they repeated our specs and showed the coating performances for HR/AR for 10deg P and PR/AR for 45deg P. There is no RoC measurement by the
vendor.
Nevertheless, their RoC (paper) specs should be compared with our request.
Entry  Mon Nov 23 23:27:43 2020, Koji, Summary, ASC, Q3000 characterization 
Last week and this week I've been working on the characterization of the Q3000 QPDs. The QPDs were named 81, 82, 83, and 94.


Dark current [OMC LAB ELOG 402]: All the segments looked similar and acceptable
except for the seg1 of #82. It has a smaller reverse breakdown voltage (~6V) but even this is an acceptable level.
Impedance [OMC
    Reply  Thu Dec 3 13:55:52 2020, Koji, Update, ASC, Electrical LO signal for AS WFS 
I got a bit confused by your description.

The demod board claims that the nominal power at each LO port is 10dBm. So we want to give at least 16dBm to the (external?)
4way power splitter, but we only have 15dBm. As you said, the actual LO power reaching the FET mixier (PE4140) is the level of ~20dBm. But you said the
    Reply  Thu Dec 3 21:14:38 2020, Koji, Update, ASC, Electrical LO signal for AS WFS 
Let's use RG405 for better shielding. It is not too stiff. The bending  (just once) does not break the cable.

Are you going to full replacement of the 55MHz system? Or just remove the 7dBm and then implement the proposed modification for the 55MHz line?
    Reply  Fri Dec 4 03:02:40 2020, Koji, Update, ASC, Electrical LO signal for AS WFS 
1. That's true. But we are already in that regime with the Var attn at 0dB, aren't we? We can reduce the input to the amp by 1-2dBm sacrificing
the EOM out by that amount (we can compensate this for the demo out by removing the 1dB attn).

2. Not 100% sure but one possible explanation is that we wanted to keep the Marconi output large (or as large as possible) to keep the SNR between
    Reply  Fri Dec 4 15:58:22 2020, Koji, Update, IOO, WFS2 broken  
I checked the backplane connection for IMC WFS2  and found that the cables for IMC WFS2 and the IMC demod were
swapped during my IMC noise hunting activities. I reverted it just now.

But we need to check if this damaged anything such as the WFS2 head, the WFS2 demod, etc, once the IMC locking is back.
Entry  Fri Dec 4 19:23:40 2020, Koji, Update, Electronics, AA/AI board testing ongoing 
I have the setup built for the AA/AI board testing around the PD testing area. Please let me leave it like that for a week or so.

12/4 TF Tested 5 PCBs
12/6 TF Tested 19 PCBs (12min/PCB) - found 1 failure (S2001479 CH1) -> Fixed 12/11
    Reply  Sat Dec 19 02:42:56 2020, Koji, Update, , New SMA cables on order 
Our favorite (flexible) RF cable is Belden's 1671J (Jacketed solder-soaked coax cable). It is compatible RG405. I'm not sure if there is off-the-shelf
SMA cables with 1671J.

 
    Reply  Sun Jan 3 16:26:06 2021, Koji, Update, SUS, IMC WFS check (Yet another round of Sat. Box. switcharoo) Screen_Shot_2021-01-03_at_17.14.57.png
I wanted to check the functionality of the IMC WFS. I just turned on the WFS servo loops as they were. For the past two hours, they didn't run away.
The servo has been left turned on. I don't think there is no reason to keep it turned off.
    Reply  Thu Jan 7 20:07:27 2021, Koji, Update, ALS, Noisy ALS 
How about resurrecting the PSL table green beat for the X arm to see if the non-fiber setup shows the same level of the freq noise (e.g. the PDH locking
became super noisy due to misalignment etc).
    Reply  Thu Jan 7 23:25:19 2021, Koji, Update, ALS, Noisy ALS 
If the sensing noise level of the end PDH degraded for some reason, it'd make the out of loop stability worse without making the end pdh error level
degraded.
It's just speculation.
    Reply  Tue Feb 2 18:24:54 2021, Koji, Update, BHD, SOS assembly 
You can remove the components of the optical table enclosure (black ones) and use the optical table as your working area too.

 
Entry  Tue Feb 2 23:29:35 2021, Koji, Update, CDS, CDS crash and CDS/IFO recovery 
I worked around the racks and the feedthru flanges this afternoon and evening. This inevitably crashed c1lsc real-time process.
Rebooting c1lsc caused multiple crashes (as usual) and I had to hard reboot c1lsc/c1sus/c1ioo
This made the "DC" indicator of the IOPs for these hosts **RED**.
    Reply  Wed Feb 3 18:53:31 2021, Koji, Update, CDS, CDS crash and CDS/IFO recovery 
Really!? I didn't reboot the machines between "sudo date" and "rtcds start c1x0*". I tried rtcds. If it didn't work, it used
date. Then tried rtcds. (repeat) The time was not synched at all wrt the time zones and also the time. There were 1~3 sec offset besides the TZ problem.

 
    Reply  Wed Feb 10 17:18:03 2021, Koji, Summary, BHD, Satellite Amplifier Output Offset measurements 
Testing the satellite amp i.e.  PD driver
- To test the noise of the PD transimpedance amps: Leave the PD input open (do not short the terminal goes to the PD)
- To test the current noise of the LED drivers: Short the output with an appropriate Rs to have the nominal current.
    Reply  Tue Feb 16 15:29:01 2021, Koji, Update, BHD, SOS assembly 
The curie temp of SmCo seems about x2 (in K) of the one for NdFeB. i.e. 600K vs 1000K. So I believe 177degC = 450K is not an issue. Just make sure the
curie temp, referring the specific property for the magnets from this company. (You already know the company from the procurement doc). It'd be great
if you upload the doc on the 40m wiki.
    Reply  Wed Feb 17 13:59:43 2021, Koji, Update, SUS, CoM on 3"->2" Adapter Ring for SOS 
Note from today's meeting:

1. Can we adjust the thickness of the cylindrical hole for the mirror to move the COM in the plane of the wires. (We should be able to do that)
Entry  Thu Feb 18 03:20:09 2021, Koji, Summary, Electronics, Current Rack Map rack_plan.pdf
For your planning:
    Reply  Thu Feb 18 20:20:25 2021, Koji, Update, SUS, aLIGO Sat Amp characterization 
Yeah, it's really inconsistent. You had 35mA LED drive and the current noise of the noisy channel was 5e-7 A/rtHz at 1Hz. The RIN is 1.4e-5 /rtHz.
The approx. received photocurrent is 30uA as we discussed today and this should make the noise around 4e-10 A/rtHz at 1Hz. However, the readout noise level
is better than this level. (well below 1e-10 A/rtHz)
Entry  Thu Feb 18 20:24:48 2021, Koji, Summary, Electronics, A bunch of electronics received IMG_0416.jpeg
Todd provided us a bunch of electronics. I went to Downs to pick them up this afternoon and checked the contents in the box. Basically, the boxes are
pretty comprehensive to produce the following chassis


8 HAM-A coil driver chassis
7 16bit Anti-Aliasing chassis
4 18bit Anti-Imaging chassis
5 16bit Anti-Imaging
    Reply  Fri Feb 19 16:06:01 2021, Koji, Update, SUS, CoM Range on 3"->2" Adapter Ring for SOS 
We want to move the CoM with the adjustment range so that the residual deviation is adjusted by the bottom dumbbell. 0.0003" is well within the
range and good enough.

 
    Reply  Fri Feb 19 16:55:26 2021, Koji, Update, SUS, CoM Range on 3"->2" Adapter Ring for SOS 
Jordan's screenshot actually shows that the vertical distance (Y) is 0.0000". We want to have the vertical distance of CoM from the wire clamping
point to be 0.9mm in the nominal SOS design (this might need to be adjusted to have a similar pitch resonant freq for the different inertia of moment).
Let's say it is ~mm ish.
    Reply  Sat Feb 20 16:46:17 2021, Koji, Summary, Electronics, A bunch of electronics received 
We received currently available sets. We are supposed to receive more coil drivers and sat amps, etc. But they are not ready yet.

 
    Reply  Tue Feb 23 20:55:19 2021, Koji, Update, SUS, OSEM testing for SOSs SOS_OSEM.pdf
I can't obtain a consistent view between the existing drawings/photographs and your pin assignment. Please review the pin assignment again to check
if yours is correct.
Entry  Tue Feb 23 23:12:37 2021, Koji, Summary, SUS, SUS invacuum wiring SOS_OSEM_cabling.pdf
This is my current understanding of the in-vacuum wiring:
1. Facts


We have the in-air cable pinout. And Gautam recently made a prototype of D2100014 custom cable, and it worked as expected.
The
    Reply  Fri Feb 26 20:20:43 2021, Koji, Update, Electronics, Production version of the HV coil driver tested with KEPCO HV supplies 
This is very disappointing. Even with KEPCO linear supply with the improved HV driver circuit, the noise level is significantly higher than the 20kOhm
R thermal noise.

What is special with the HP supplies? Can you replace KEPCOs with the HP supply, one by one to specify which one is making the noise bad?
    Reply  Thu Mar 4 15:48:26 2021, Koji, Summary, PEM, Watchdog tripped, Optics damped back 
EQs seen on Summary pages
https://nodus.ligo.caltech.edu:30889/detcharsummary/day/20210304/pem/seismic_blrms/
    Reply  Thu Mar 4 23:16:08 2021, Koji, Update, LSC, REFL55 demod board rework D990511-00_REFL55.pdfP_20210304_215602.jpgP_20210304_222400.jpgP_20210304_222412.jpg20210304234400_IMG_0526.jpg
A new hybrid splitter (DQS-10-100) was installed. As the amplification of the final stage is sufficient for the input level of 3dBm, I have bypassed
the input amplification (Attachment 1). One of the mixer was desoldered to check the power level. With a 1dB ATTN, the output of the last
ERA-5 was +17.8dBm (Attachment 2). (The mixer was resoldered.)
Entry  Thu Mar 4 23:57:35 2021, Koji, Summary, Electronics, Inspection of the new custom dsub cables 6x
I made the inspection of the new custom DSub cables (came from Texas).

The shelled version gives us some chance to inspect/modify the internal connections. (good)
The wires are well insulated. The conductors are wrapped with the foils and then everything is in the braid tube shield. The braid is soldered on
    Reply  Fri Mar 5 00:53:09 2021, Koji, Summary, Electronics, A bunch of electronics received 
Received additional front/rear panels. Updated the original entry and Wiki [Link]

 
    Reply  Fri Mar 5 15:31:23 2021, Koji, Update, LSC, REFL55 demod board rework 
Missed to note: The IF test was done at TP7 and TP6 using pomona clips i.e. brefore the preamp.

 
    Reply  Fri Mar 5 15:32:53 2021, Koji, Summary, Electronics, A bunch of electronics received 
The parts will be ordered by Koji The components for the additional BIO I/F have been ordered.
Entry  Tue Mar 9 12:41:29 2021, Koji, Summary, Electronics, Investigation on the invacuum Dsub cables SOS_OSEM_cabling.pdf
I believe the aLIGO style invac dsub cables and the conventional 40m ones are incompatible.
While the aLIGO spec is that Pin1 (in-vac) is connected to the shield, Pin13 (in-vac) is the one for the conventional cable. I still have to
check if Pin13 is really connected to the shield, but we had trouble before for the IO TTs https://nodus.ligo.caltech.edu:8081/40m/7864.