ID |
Date |
Author |
Type |
Category |
Subject |
12539
|
Fri Oct 7 20:25:14 2016 |
Koji | Update | CDS | Power-cycled c1psl and c1iool0 |
Found the MC autolocker kept failing, It turned out that c1iool0 and c1psl went bad and did not accept the epics commands.
Went to the rack and power cycled them. Burt resotred with the snapshot files at 5:07 today.
The PMC lock was restored, IMC was locked, WFS turned on, and WFS output offloaded to the bias sliders.
The PMC seemed highly misaligned, but I didn't bother myself to touch it this time. |
12540
|
Fri Oct 7 20:56:15 2016 |
Koji | Update | SUS | Output matrix diagonalization |
I wanted to see what is the reason to have such large coupling between pitch and yaw motions.
The first test was to check orthogonality of the bias sliders. It was done by monitoring the suspension motion using the green beam.
The Y arm cavity was aligned to the green. The damping of ITMY was all turned off except for SD.
Then ITMY was misaligned by the bias sliders. The ITMY face CCD view shows that the beam is reasonably orthogonally responding to the pitch and yaw sliders.
I also confirmed that the OPLEV signals also showed reasonablly orthogonal responce to the pitch and yaw misalignment.
=> My intuition was that the coils (including the gain balance) are OK for a first approximation.
Then, I started to excite the resonant modes. I agree that it is difficult to excite a pure picth motion with the resonance.
So I wanted to see how the mixing is frequency dependent.
The transfer functions between ITMY_ASCPIT/YAW_EXC to ITMY_OPLEV_PERROR/YERROR were measured.
The attached PDFs basically shows that the transfer functions are basically orthogonal (i.e. pitch exc goes to pitch, yaw exc goes to yaw) except at the resonant frequency.
I think the problem is that the two modes are almost degenerate but not completely. This elog shows that the resonant freq of the ITMY modes are particularly close compared to the other suspensions.
If they are completely degenerate, the motion just obeys our excitation. However, they are slightly split. Therefore, we suffer from the coupled modes of P and Y at the resonant freq.
However, the mirror motion obeys the exitation at the off resonance as these two modes are similar enough.
This means that the problem exists only at the resonant frequencies. If the damping servos have 1/f slope around the resonant freqs (that's the usual case), the antiresonance due to the mode coupling does not cause servo instability thank to the sufficient phase margin.
In conclusion, unfortunately we can't diagnalize the sensors and actuators using the natural modes because our assumption of the mode purity is not valid.
We can leave the pitch/yaw modes undiagnalized or just believe the oplevs as a relatively reliable reference of pitch and yaw and set the output matrix accordingly.
The figures will be rotated later. |
12555
|
Wed Oct 12 22:51:26 2016 |
Koji | Update | General | Clipping could not be improved by the PZTs - NEED ANOTHER VENT |
[Gautam Koji]
We engaged the HV driver to the output port PZTs, hoping to mitigate the AS port clipping. Basically, the range of the PZT is not enough to make the beam look clean. Also, our observation suggested there are possible multiple clipping in the chamber. We need another vent to make the things clearly right. Eric came in the lab and preparing the IFO for it.
1. Before the test, the test masses have been aligned with the dither servo.
2. We looked at the beam shape on the AS camera with a single bounce beam. We confirmed that the beam is hard-clipped at the upper and left sides of the beam on the video display. This clipping is not happening outside of the chamber.
3. We brought an HV power supply to the short OMC rack. There is a power supply cable with two spades. The red and black wires are +150V and GND respectively.
4. The voltage of +/-10V was applied on each of the four PZT drive inputs. We found that the motion of the beam on the camera is tiny and in any case, we could not improve the beam shape.
5. We wondered that if we are observing ANY improvement of the clipping. For this purpose, we aligned AS110 sensor every time we gave the misalignment with the PZTs. Basically, we are at the alignment to have the best power we can get. We thought this was weird.
6. Then we moved the AS port spot with the ITMX. We could clearly make the spot more round. However, this reduced the power at the AS port reduced by ~15%. When the beam was further clipped, the power went down again. Basically, the initial alignment gave us the max power we could get. As the max power was given with the clipped beam, we get confused and feel safer to check the situation with the chambers open.
During this investigation, we moved the AS port opitcs and the AS camera. So they are not too precise reference of the alignment. The PZT HV setup has been removed. |
12558
|
Thu Oct 13 14:49:57 2016 |
Koji | Configuration | PEM | XLR(F)-XLR(M) cable took from the fibox to the Blue microphone |
[Gautam Koji]
XLR(F)-XLR(M) cable for the blue microphone is missing. Steve ordered one.
We found one in the fibox setup. As we don't use it during the vent, we use this cable for the microphone.
Once we get the new one, it will go to the fibox setup.
|
12580
|
Tue Oct 25 18:07:28 2016 |
Koji | Update | General | PRFPMI locked, arms loss improved |
Great to hear that we have the PRG of ~16 now!
Is this 150ppm an avg loss per mirror, or per arm? |
12584
|
Thu Oct 27 13:48:20 2016 |
Koji | Update | General | PRFPMI locked, arms loss improved |
It is also difficult to have a high arm transmission without having high PRG.
What about to plot the arm trans and the REFL DC power in a timeseries?
Or even in a correlation plot (X: Arm Trans or PRG vs Y: REFL Reflectivity)
This tells you an approximate location of the critical coupling, and allows you to calibrate the PRG, hopefully. |
12641
|
Sat Nov 26 19:16:28 2016 |
Koji | Update | IOO | IMC WFS Demod board measurement & analysis |
[Rana, Koji]
1. The response of the IMC WFS board was measured. The LO signal with 0.3Vpp@29.5MHz on 50Ohm was supplied from DS345. I've confirmed that this signal is enough to trigger the comparator chip right next to the LO input. The RF signal with 0.1Vpp on the 50Ohm input impedance was provided from another DS345 to CH1 with a frequency offset of 20Hz~10kHz. Two DS345s were synced by the 10MHz RFreference at the rear of the units. The resulting low frequency signal from the 1st AF stage (AD797) and the 2nd AF stage (OP284) were checked.
Attachment 1 shows the measured and modelled response of the demodulator with various frequency offsets. The value shows the signal transfer (i.e. the output amplitude normalized by the input amplitude) from the input to the outputs of the 1st and 2nd stages. According to the datasheet, the demodulator chip provides a single pole cutoff of 340kHz with the 33nF caps between AP/AN and VP. The first stage is a broadband amplifier, but there is a passive LPF (fc=~1kHz). The second stage also provides the 2nd order LPF at fc~1kHz too. The measurement and the model show good agreement.
2. The output noise levels of the 1st and 2nd stages were meausred and compared with the noise model by LISO.
Attachment 2 shows the input referred noise of the demodulator circuit. The output noise is basically limited by the noise of the first stage. The noise of the 2nd stage make the significant contribution only above the cut off freq of the circuit (~1kHz). And the model supports this fact. The 6.65kOhm of the passive filter and the input current noise of AD797 cause the large (>30nV/rtHz) noise contribution below 100Hz. This completely spoils the low noiseness (~1nV/rtHz) of AD797. At lower frequency like 0.1Hz other component comes up above the modelled noise level.
3. Rana and I had a discussion about the modification of the circuit. Attachment 4 shows the possible improvement of the demod circuit and the 1st stage preamplifier. The demodulator chip can have a cut off by the attached capacitor. We will replace the 33nF caps with 1uF and the cut off will be pushed down to ~10kHz. Then the passive LPF will be removed. We don't need "rodeo horse" AD797 for this circuit, but op27 is just fine instead. The gain of the 1st stage can be increased from 9 to 21. This should give us >x10 improvement of the noise contribution from the demodualtor (Attachment 3). We also can replace some of the important resistors with the thin film low noise resistors. |
12645
|
Tue Nov 29 17:45:06 2016 |
Koji | Update | IOO | IMC WFS Demod board measurement & analysis |
Summary: The demodulator input noise level was improved by a factor of more than 2. This was not as much as we expected from the preamp noise improvement, but is something. If this looks OK, I will implement this modification to all the 16 channels.
The modification shown in Attachment 1 has actually been applied to a channel.
- The two 1.5uF capacitors between VP and AN/AP were added. This decreases the bandwidth of the demodulator down to 7.4kHz
- The offset trimming circuit was disabled. i.e. Pin18 of AD831 was grounded.
- The passive low pass at the demodulator output was removed. (R18, C34)
- The stage1 (preamp) chip was changed from AD797 to OP27.
- The gain of the preamp stage was changed from 9 to 21. Also the thin film resistors are used.
Attachment 2 shows the measured and expected output signal transfer of the demodulator. The actual behavior of the demodulator is as expected, and we still keep the over all LPF feature of 3rd order with fc=~1kHz.
Attachment 3 shows the improvement of the noise level with the signal reffered to the demodulator input. The improvement by a factor >2 was observed all over the frequency range. However, this noise level could not be explained by the preamp noise level. Actually this noise below 1kHz is present at the output of the demodulator. (Surprisingly, or as usual, the noise level of the previous preamp configuration was just right at the noise level of the demodulator below 100Hz.) The removal of the offset trimmer circuit contributed to the noise improvement below 0.3Hz.
|
12661
|
Fri Dec 2 18:02:37 2016 |
Koji | Update | IOO | IMC WFS Demod board measurement & analysis |
ELOG of the Wednesday work.
It turned out that the IMC WFS demod boards have the PCB board that has a different pattern for each of 8ch.
In addition, AD831 has a quite narrow leg pitch with legs that are not easily accessible.
Because of these, we (Koji and Rana) decided to leave the demodulator chip untouched.
I have plugged in the board with the WFS2-Q1 channel modified in order to check the significance of the modification.
WFS performance before the modification
Attachment 1 shows the PSD of WFS2-I1_OUT calibrated to be referred to the demodulator output. (i.e. Measured PSDs (cnt/rtHz) were divided by 8.9*2^16/20)
There are three curves: One is the output with the MC locked (WFS servos not engaged). The second is the PSD with the PSL beam blocked (i.e. dark noise). The third is the electronics noise with the RF input terminated and the nominal LO supplied.
This tells us that the measured PSD was dominated by the demodulator noise in the dark condition. And the WFS signal was also dominated by the demod noise below 0.1Hz and above 20Hz. There are annoying features at 0.7, 1.4, 2.1, ... Hz. They basically impose these noise peaks on the stabilized mirror motion.
WFS performance after the modification
Attachment 2 shows the PSD of WFS2-Q1_OUT calibrated to be referred to the demodulator output. (i.e. Measured PSDs (cnt/rtHz) were divided by 21.4*2^16/20)
There are three same curves as the other plot. In addition to these, the PSD of WFS2-I1_OUT with the MC locked is also shown as a red curve for comparison.
This figure tells us that the measured PSD below 20Hz was dominated by the demodulator noise in the dark condition. And the WFS signal is no longer dominated by the electronics noise. However, there still are the peaks at the harmonics of 0.7, 1.4, 2.1, ... Hz. I need further inspection of the FWS demod and whtening boards to track down the cause of these peaks. |
12662
|
Sat Dec 3 13:27:35 2016 |
Koji | Update | IOO | IMC WFS Demod board measurement & analysis |
ELOG of the work on Thursday
Gautam suggested looking at the preamplifier noise by shorting the input to the first stage. I thought it was a great idea.
To my surprise, the noise of the 2nd stage was really high compared to the model. I proceeded to investigate what was wrong.
It turned out that the resistors used in this sallen-key LPF were thick film resistors. I swapped them with thin film resistors and this gave the huge improvement of the preamplifier noise in the low frequency band.
Attachment 1 shows the summary of the results. Previously the input referred noise of the preamp was the curve in red. We the resistors replaced, it became the curve in magenta, which is pretty close to the expected noise level by LISO model above 3Hz (dashed curves). Unfortunately, the output of the unit with the demodulator connected showed no improvement (blue vs green), because the output is still limited by the demodulator noise. There were harmonic noise peaks at n x 10Hz before the resistor replacement. I wonder if this modification also removed the harmonic noise seen in the CDS signals. I will check this next week.
Attachment 2 shows the current schematic diagram of the demodulator board. The Q of the sallen key filter was adjusted by the gain to have 0.7 (butter worth). We can adjust the Q by the ratio of the capacitance. We can short 3.83K and remove 6.65K next to it. And use 22nF and 47nF for the capacitors at the positive input and the feedback, respectively. This reduces the number of the resistors. |
12668
|
Tue Dec 6 13:37:02 2016 |
Koji | Update | IOO | IMC WFS Demod board measurement & analysis |
I have implemented the modification to the demod boards (Attachment 1).
Now, I am looking at the noise in the whitening board. Attachment 2 shows the comparison of the error signal with the input of the whitening filter shorted and with the 50ohm terminator on the demodulator board. The message is that the whitening filter dominates the noise below 3Hz.
I am looking at the schematic of the whitening board D990196-B. It has an VGA AD602 at the input. I could not find the gain setting for this chip.
If the gain input is fixed at 0V, AD602 has the gain of 10dB. The later stages are the filters. I presume they have the thick film resistors.
Then they may also cause the noise. Not sure which is the case yet.
Also it seems that 0.7Hz noise is still present. We can say that this is coming from the demod board but not on the work bench but in the eurocard crate. |
12669
|
Tue Dec 6 16:47:40 2016 |
Koji | Update | IOO | IMC WFS whitening filter investigation |
The whitening board saids it is Rev B, but the actual component values are more like Rev. C.
The input stage (AD602) has an input resistor of 909 Ohm.
This is causing a big attenuation of the signal (x1/10) because the input impedance of AD602 is not high. And this screws up the logarithm of the gain.
I don't think this is a right approach. |
12670
|
Tue Dec 6 17:54:08 2016 |
Koji | Update | IOO | IMC WFS whitening filter investigation |
The input resistor 909Ohm of AD602 was shorted. I've confirmed that the gain (= attenuation by voltage division) was increased by a factor of 10.
This modification was done for WFS2-I1 and WFS2-Q1. Also the thick film resistors for the WFS2-I1 channel was all replaced with thin film resistors.
Attachment 1 shows the comparison of the noise levels. The curves were all calibrated referred to the response of the original whitening filter configuration.
(i.e. measurement done after the gain change was compensated by the factor of 10.)
Now the AF chain is not limited by the noise in the whitening filter board. (Brown)
In fact, this noise level was completely identical between I1 and Q1. Therefore, I don't think we need this resistor replacement for the whitening filter board.
We can observe the improvement of the overall noise level below 10Hz. (Comparison between green and red/blue)
As the signal level goes up, the noise above 100Hz was also improved.
Now we need to take care of the n x 0.7Hz feature which is in the demod board...
|
12671
|
Tue Dec 6 22:41:49 2016 |
Koji | Update | IOO | IMC WFS whitening filter investigation |
I have implemented the same modification (shorting the input resistor of AD602) to the two whitening boards. |
12676
|
Tue Dec 13 17:26:42 2016 |
Koji | Update | IOO | IMC WFS whitening filter investigation |
Rana pointed out that this modification (removal of 900Ohm) leave the input impedance as low as 100Ohm.
As OP284 can drive up to 10mA, the input can span only +/-1V with some nonlinearity.
Rather than reinstalling the 900Ohms, Rana will investigate the old-days fix for the whitening filter that may involve the removal of AD602s.
Until the solution is supplied, the IMC WFS project is suspended. |
12679
|
Mon Dec 19 22:05:09 2016 |
Koji | Summary | IOO | PMC, IMC aligned. The ringdown PD/Lens removed. |
PMC and IMC were aligned on Friday (16th) and Today (19th).
The PD and lens for the ringdown experiment were removed as they were blocking the WFS. |
12680
|
Wed Dec 21 21:03:06 2016 |
Koji | Summary | IOO | IMC WFS tuning |
- Updated the circuit diagrams:
IMC WFS Demodulator Board, Rev. 40m https://dcc.ligo.org/LIGO-D1600503
IMC WFS Whitening Board, Rev. 40m https://dcc.ligo.org/LIGO-D1600504
- Measured the noise levels of the whitening board, demodboard, and nominal free running WFS signals.
- IMC WFS demod phases for 8ch adjusted
Injected an IMC PDH error point offset (@1kHz, 10mV, 10dB gain) and adjusted the phase to have no signal in the Q phase signals.
- The WFS2 PITCH/YAW matrix was fixed
It was found that the WFS heads were rotated by 45 deg (->OK) in CW and CCW for WFS1 and 2, respectively (oh!), while the input matrices were identical! This made the pitch and yaw swapped for WFS2. (See attachment)
- Measured the TFs MC1/2/3 P/Y actuation to the error signals |
12682
|
Thu Dec 22 18:39:09 2016 |
Koji | Summary | IOO | IMC WFS tuning |
Noise analysis of the WFS error signals.
Attachment 1: All error signals compared with the noise contribution measured with the RF inputs or the whitening inputs terminated.
Attachment 2: Same plot for all the 16 channels. The first plot (WFS1 I1) shows the comparison of the current noise contributions and the original noise level measured with the RF terminated with the gain adjusted along with the circuit modification for the fair comparison. This plot is telling us that the electronics noise was really close to the error signal.
I wonder if we have the calibration of the IMC suspensions somewhere so that I can convert these plots in to rad/sqrtHz...? |
12683
|
Fri Dec 23 20:53:44 2016 |
Koji | Summary | IOO | IMC WFS tuning |
WFS1 / WFS2 demod phases and WFS signal matrix |
12684
|
Fri Dec 23 21:05:56 2016 |
Koji | Summary | IOO | IMC WFS tuning |
Signal transfer function measurements
C1:SUS-MC*_ASCPIT_EXC channels were excited for swept sine measurements.
The TFs to WFS1-I1~4, Q1~4, WFS1/2_PIT/YAW, MC2TRANS_PIT/YAW signals were recorded.
The MC1 and MC3 actuation seems to have ~30Hz elliptic LPF somewhere in the electronics chain.
This effect was compensated by subtracting the approximated time delay of 0.022sec.
The TFs were devided by freq^2 to make the response flat and averaged between 7Hz to 15Hz.
The results have been summarized in Attachment 3&4.
Attachment 4 has the signal sensing matrix. Note that this matrix was measured with the input gain of 0.1.
Input matrix for diagonalizing the actuation/sensor response
Pitch

e.g. To produce pure WFS1P reaction, => -1.59 MC1P + 0.962 MC2P + 0.425 MC3P
Yaw

|
12685
|
Sun Dec 25 14:39:59 2016 |
Koji | Summary | IOO | IMC WFS tuning |
Now, the output matrices in the previous entry were implemented.
The WFS servo loops have been engaged for several hours.
So far the REFL and TRANS look straight. Let's see how it goes. |
12686
|
Mon Dec 26 12:45:31 2016 |
Koji | Summary | IOO | IMC WFS tuning |
It didn't go crazy at least for the past 24hours. |
12689
|
Thu Dec 29 16:52:51 2016 |
Koji | Summary | IOO | IMC WFS tuning |
Koji responding to Rana
> For the rough calibration below 10 Hz, we can use the SUS OSEM cal: the SUSPIT and SUSYAW error signals are in units of micro-radians.
I can believe the calibration for the individual OSEMs. But the input matrix looked pretty random, and I was not sure how it was normalized.
If we accept errors by a factor of 2~3, I can just naively believe the calibration factors.
> If the RF signals at the demod input are low enough, we can consider either increasing the light power on the WFS or increasing the IMC mod. depth.
The demod chip has the conversion factor of about the unity. We increased the gains of the AF stages in the demod and whitening boards. However, we only have the RMS of 1~20 counts. This means that we have really small RF signals. We should check what's happening at the RF outputs of the WFS units. Do we have any attenuators in the RF chain? Can we skip them without making the WFS units unstable? |
12715
|
Fri Jan 13 21:41:23 2017 |
Koji | Update | CDS | DC errors |
I think I fixed the DC error issue
1. I added the leap second (leapsecond ?) entry for 2016/12/31, 23:60:00 UTC to daqdrc
[OLD]
set gps_leaps = 820108813 914803214 1119744016;
[NEW]
set gps_leaps = 820108813 914803214 1119744016 1167264018;
2. Restarted FB and all realtime models
Now I don't see any RED light. |
12756
|
Wed Jan 25 17:30:03 2017 |
Koji | Update | IMC | 29.5 MHz modulation depth measurement plan |
I'm afraid that the bidirectional coupler, designed to be 50ohm in/out, disturbs the resonant circuit designed for the EOM which is almost purely capacitive.
One possible way could be to measure the transfer function using the active FET probe from the triple resonant input to the output with the EOM attached.
Another way: How about to measure the reflection before the resonant circuit? Then, of course, there is the triple resonant interface circuit between the power combiner and the EOM. This case, we will see how much power is consumed in EOM and the resonant circuit. Then we can use the previous measurement to see the conversion factor between the power consumption to the modulation depth. Kiwamu may give us his measurement. |
12757
|
Wed Jan 25 18:18:08 2017 |
Koji | Summary | IOO | MCL / MCF / Calibration |
jiSome notes on the FSS configuration: ELOG 10321 |
12761
|
Fri Jan 27 15:36:17 2017 |
Koji | Update | SUS | wire standoffs update |
Very nice! I got excited.
- Don't you ask Calum and co to check the groove size with their microscopes? Just give the samples and the wire.
- Do we want to make a simple "guitar" setup to measure the vibration Qs with Al piece, glass prism, ungrooved Sapphire, this grooved sapphire, grooved ruby, etc?
|
12783
|
Wed Feb 1 11:51:19 2017 |
Koji | Update | IMC | RF AM stabilization box pulled out |
For a comparison: OMC ELOG 238 |
12805
|
Mon Feb 6 18:20:08 2017 |
Koji | Update | IMC | MCL Feedback - simulink model updated |
It is more accurate to model the physical frequency noises at various places.
cf. See also 40m ALS paper or Shigeo Nagano PDH thesis on https://wiki-40m.ligo.caltech.edu/40m_Library
- The output 4 should be "Laser frequency"
- Seismic path should be excluded from the summing node
- The output after the PMC: "Laser frequency after the PMC"
- "Laser frequency after the PMC" is compared (diffed) with the output 1 "mirror motion in Hz"
- The comparator output goes to the cav pole, the PD, and the PDH gain: This is the output named "PDH Error"
- Tap a new path from "Laser frequency after the PMC" and multiply with the cav pole (C_IMC)
- Tap a new path from "Mirror motion" and multiply with the cavity high pass (s C_IMC/omega)
- Add these two: This is the output named "Frequency noise transmitted by IMC" |
12821
|
Fri Feb 10 19:32:15 2017 |
Koji | Update | IMC | IMC Demod board |
The input impedance of the mixer is not constant. As the diode switches, it changes dynamically. Because of this, the waveform of the LO at the mixer input (i.e. the amplifier output) is not sinusoidal. Some of the power goes away to harmonic frequencies. Also, your active probe is calibrated to measure the power across the exact 50Ohm load, which is not in this case. The real confirmation can be done by swapping the mixer with a 50Ohm resistor. But it is too much. Just confirm the power BEFORE the amp is fine. +/-1dB does not change the mixer function much.
Instead, we should measure
- Orthogonality
- Gain imbalance
of the I/Q output. This can be checked by supplying an RF signal that is 100~1kHz away from the LO frequency and observe I&Q outputs. |
12837
|
Fri Feb 17 20:04:43 2017 |
Koji | Update | General | Projector not functional / Zita partially working |
Koji, Gautam, Johannes
We quickly checked the situation of the projector in the control room.
- We found that the proejctor was indicating "lamp error".
==> Steve, could you remove the projector from the ceiling and check if it still does not work?
If it still does not work, send it back to the vender. It should be covered by the previous service.
- Zita seemed happy with the DVI output. We tried the dual display configration and VGA and DVI are active right now.
The DVI output (from RADEON something video card) is somewhat strange. We probably need to look into the video display situation. |
12846
|
Thu Feb 23 09:32:20 2017 |
Koji | Update | SUS | wire standoffs update |
Kyle took high quality images of the three sapphire prisms using the microscope @Downs. He analyzed the images to see the radius of the groove.
They all look sufficiently sharp for a 46um steel wire. Thumbs up.
I am curious to see how the wire Q is with grooved sapphires, ungrooved sapphires, grooved ruby, grooved aluminum stand off, and so on. |
12875
|
Thu Mar 9 15:25:12 2017 |
Koji | Update | General | IMC/XYarms aligned/locked |
As per Steve's request, I've checked the alignment of the IMC and the arms. These three cavities are locked and aligned. |
12877
|
Thu Mar 9 20:11:04 2017 |
Koji | Update | General | attempted ETMY picture taking |
The attached is the ETMY image with the single arm locked. This was the best I could do. Here is the recipe
- Turn on SP570UZ
- Switch to "M" mode (Manual aperture and exposure)
- Set the aperture to be the widest (smallest F number) and the exposure to be maximum (15 second).
- Switch to AF mode by the lens side switch
- Use the lens dial to adjust the zoom until the OSEMs fill the central 1/3 box (i.e. 1/9 area of the field of view). If you zoom more, you can't focus the spot later.
- Use menu button to switch to ISO1600 (You are now capable to see the beam spot)
- Switch to MF mode by the lens side switch
- Use the lens dial to adjust the focus to have the sharpest image of the spot. This can be achieved at the focal distance of ~1m
- Use menu button to switch back to ISO64
- Push the shutter (I didn't use it, but you should be able to use 2sec timer)
|
12912
|
Mon Mar 27 22:40:44 2017 |
Koji | Summary | IOO | MCL / MCF / Calibration |
In http://nodus.ligo.caltech.edu:8080/40m/11793 I posted the calibrated PMC free-running displcament with/without IMC locked. Unfortunately, this measurement was done with a part of the IMC electronics not perfect (https://nodus.ligo.caltech.edu:8081/40m/11794). I did the same measurement after the fix, but there is no low freq data http://nodus.ligo.caltech.edu:8080/40m/11795.
Assuming we have the similar error signal leve in the low freq band as The entry 11793, the IMC is considered to be noisier than the PMC between 0.8 and 4Hz. But we should do the same measurement with the current electronics.
The PMC calibration can be found in this entry http://nodus.ligo.caltech.edu:8080/40m/11780 |
12966
|
Wed May 3 16:46:18 2017 |
Koji | Configuration | Computers | catastrophic multiple monitor failures |
- Is there any machine that can handle 4K? I have one 4K LCD for no use.
- I also can donate one 24" Dell |
12967
|
Wed May 3 16:47:45 2017 |
Koji | Update | General | PI pzt inventory check |
I also have a functional one on my desk, which has one of the wires repaired.
Quote: |
One is broken, two are ready to steer green and 3 available in un known condition
|
|
12992
|
Mon May 15 19:21:04 2017 |
Koji | Update | Computer Scripts / Programs | FSSslow / MCautolocker restarted |
It seems that FSS slow servo stopped working.
I found that megatron was restarted (by Rana, to finish an apt-get upgrade) on ~18:47 PDT today.
controls@megatron|~> last -5
controls pts/0 192.168.113.216 Mon May 15 19:15 still logged in
controls pts/0 192.168.113.216 Mon May 15 19:14 - 19:15 (00:01)
reboot system boot 3.2.0-126-generi Mon May 15 18:50 - 19:19 (00:29)
controls pts/0 192.168.113.200 Mon May 15 18:43 - down (00:04)
controls pts/0 192.168.113.200 Mon May 15 15:25 - 17:38 (02:12)
FSSslow / MCautolocker were restarted on megatron.
|
13032
|
Fri Jun 2 00:54:08 2017 |
Koji | Update | ASS | Xarm ASS restoration work |
While Gautam is working the restoration of Yarm ASS, I worked on Xarm.
Basically, I have changed the oscillator freqs and amps so as to have linear signals to the misalignment of the mirrors.
Also reduced the complexity of the input/output matrices to avoid any confusion.
Now the ITM dither takes care of the ITM alignment, and the ETM dither takes care of the ETM alignment.
The cavity alignment servos (4dofs) are running fine although the control band widths are still low (<0.1Hz).
The ETM spot positions should be controlled by the BS alignment, but it seems that these loops have suspicion about the signal quality.
While Gautam wa stouching the input TTs, we occasionally saw anomalously high transmission of the arm cavities (~1.2).
We decided to use this beam as this could have indicated partial clipping of the beam somewhere in the input optics chain.
Then the arm cavity was aligned to have reasonably high transmission for the green beam. i.e. Use the green power mon PD as a part of the alignment reference.
This resulted very stable transmission of both the IR and green beams. We liked them. We decide to use this a reference beam at least for now.
Attachment1: GTRX image at the end of the work.
Attachment2: ASSX screen shot
Attachment3: ASSX servo screen shot
Attachment4: Green ASX servo screen shot
Attachment 5: Screen shot of the ASS X strip tool
Attachment 6: Screen shot of the ASS X input matrix
Attachment 7: Screen shot of the ASS X output matrix |
13081
|
Mon Jun 26 22:01:08 2017 |
Koji | Update | General | c1iscex timing troubles |
I tried a couple of things, but no fundamental improvement of the missing LED light on the timing board.
- The power supply cable to the timing board at c1iscex indicated +12.3V
- I swapped the timing fiber to the new one (orange) in the digital cabinet. It didn't help.
- I swapped the opto-electronic I/F for the timing fiber with the Y-end one. The X-end one worked at Y-end, and Y-end one didn't work at X-end.
- I suspected the timing board itself -> I brought a "spare" timing board from the digital cabinet and tried to swap the board. This didn't help.
Some ideas:
- Bring the X-end fiber to C1SUS or C1IOO to see if the fiber is OK or not.
- We checked the opto-electronic I/F is OK
- Try to swap the IO chassis with the Y-end one.
- If this helps, swap the timing board only to see this is the problem or not. |
13094
|
Sat Jul 1 14:27:00 2017 |
Koji | Update | General | DRMI re-locked |
Basically we use the arm cavities as the reference of the beam alignment. The incident beam is aligned such that the ITMY angle dither is minimized (at least at the dither freq).
This means that we have no capability to adjust the spot poisitions on the PRM, SRM, BS, ITMX optics.
We are still able to minimize A2L by adding intentional asymmetry to the coil actuators. |
13110
|
Mon Jul 10 22:07:35 2017 |
Koji | Update | General | All FEs down |
I think this is the boot disk failure. I put the spare 2.5 inch disk into the slot #1. The OK indicator of the disk became solid green almost immediately, and it was recognized on the BIOS in the boot section as "Hard Disk". On the contrary, the original disk in the slot #0 has the "OK" indicator kept flashing and the BIOS can't find the harddisk.
|
13112
|
Tue Jul 11 15:12:57 2017 |
Koji | Update | General | All FEs down |
If we have a SATA/USB adapter, we can test if the disk is still responding or not. If it is still responding, can we probably salvage the files?
Chiara used to have a 2.5" disk that is connected via USB3. As far as I know, we have remote and local backup scripts running (TBC), we can borrow the USB/SATA interface from Chiara.
If the disk is completely gone, we need to rebuilt the disk according to Jamie, and I don't know how to do it. (Don't we have any spare copy?) |
13154
|
Mon Jul 31 20:35:42 2017 |
Koji | Summary | Computers | Chiara backup situation summary |
Summary
- CDS Shared files system: backed up
- Chiara system itself: not backed up
controls@chiara|~> df -m
Filesystem 1M-blocks Used Available Use% Mounted on
/dev/sda1 450420 11039 416501 3% /
udev 15543 1 15543 1% /dev
tmpfs 3111 1 3110 1% /run
none 5 0 5 0% /run/lock
none 15554 1 15554 1% /run/shm
/dev/sdb1 2064245 1718929 240459 88% /home/cds
/dev/sdd1 1877792 1426378 356028 81% /media/fb9bba0d-7024-41a6-9d29-b14e631a2628
/dev/sdc1 1877764 1686420 95960 95% /media/40mBackup
/dev/sda1 : System boot disk
/dev/sdb1 : main cds disk file system 2TB partition of 3TB disk (1TB vacant)
/dev/sdc1 : Daily backup of /dev/sdb1 via a cron job (/opt/rtcds/caltech/c1/scripts/backup/localbackup)
/dev/sdd1 : 2014 snap shot of cds. Not actively used. USB
https://nodus.ligo.caltech.edu:8081/40m/11640
|
13159
|
Wed Aug 2 14:47:20 2017 |
Koji | Summary | Computers | Chiara backup situation summary |
I further made the burt snapshot directories compressed along with ELOG 11640. This freed up additional ~130GB. This will eventually help to give more space to the local backup (/dev/sdc1)
controls@chiara|~> df -m
Filesystem 1M-blocks Used Available Use% Mounted on
/dev/sda1 450420 11039 416501 3% /
udev 15543 1 15543 1% /dev
tmpfs 3111 1 3110 1% /run
none 5 0 5 0% /run/lock
none 15554 1 15554 1% /run/shm
/dev/sdb1 2064245 1581871 377517 81% /home/cds
/dev/sdd1 1877792 1426378 356028 81% /media/fb9bba0d-7024-41a6-9d29-b14e631a2628
/dev/sdc1 1877764 1698489 83891 96% /media/40mBackup
|
13170
|
Mon Aug 7 22:50:57 2017 |
Koji | Update | General | New wifi router for the GC network installed |
I have replaced the old 11n wifi router (CISCO / Linksys) for the GC network with a new one with 11ac technology.
The new one is a 3band wifi router. Thus it has one 2.4GHz (11n) SSID and two 5GHz (11ac) SSIDs. All these have been set to be hidden. Just come to the 40m and find the necessary info for the connection.
Note that the user id / password for the admin tool have been changed from the default values. |
13201
|
Sun Aug 13 01:35:09 2017 |
Koji | Update | SUS | Glitches stay on MC1 |
We used to have similar suspension excursion at ETMX. This was the motivation to replace the stand-offs from Al ones to ruby ones. Did the replacement solve the issue at ETMX? |
13216
|
Wed Aug 16 17:14:02 2017 |
Koji | Update | CDS | front-end/DAQ network down for kernel upgrade, and timing errors |
What's the current backup situation? |
13218
|
Wed Aug 16 18:06:01 2017 |
Koji | Update | CDS | front-end/DAQ network down for kernel upgrade, and timing errors |
We also need to copy chiara's root. What is the best way to get the full image of the root FS?
We may need to restore these root images to a different disk with a different capacity.
Is the dump command good for this? |
13251
|
Thu Aug 24 18:51:57 2017 |
Koji | Summary | LSC | First cavity length reconstruction with a neural network |
Phenomenal! |