40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 114 of 355  Not logged in ELOG logo
ID Date Author Type Category Subjectup
  16780   Thu Apr 14 18:34:51 2022 PacoSummaryBHDITMY Oplev reinstalled (Re: 2 in oplev mirrors incompatible with LMR2V)

[Paco, Yehonathan]

We installed ITMYOL1 and ITMYOL2 on the ITMY chamber. We aligned the ITMY OpLev beam and closed the loop successfully, we then had a second round of YARM aligment, where we brought the Y peak transmission up from 0.04 counts to 0.09 counts (up by a factor of two). We still couldn't close the YARM loop but we have a better alignment.

  5436   Fri Sep 16 16:34:54 2011 PaulUpdateSUSITMY SRM oplev telescope plan

I've calculated a suitable collimating telescope for the ITMY/SRM oplev laser, based on the specs for the soon-to-arrive 2mW laser (model 1122/P) available here: http://www.jdsu.com/ProductLiterature/hnlh1100_ds_cl_ae.pdf

Based on the fact that the 'beam size' value and 'divergence angle' value quoted don't match up, I am assuming that the beam radius value of 315um is _not_ the waist size value, but rather the beam size at the output coupler. From the divergence angle I calculated a 155um waist, (zR = 12cm). This gives the quoted beam size of about 316um at a distance of 8.5" away from the waist. This makes me think that the output coupler is curved and the waist is at the back of the laser, or at least 8.5" from the output coupler.

The collimating telescope gives a waist of size 1142um (zR=6.47m) at a distance of 1.427m away from the original laser waist, using the following lens combo:

 

L1 f=-0.15 @ 0.301m

L2 f=0.3 @ 0.409m

 

This should be fine to get a small enough spot size (1-2mm) on the QPDs.

 

  13830   Thu May 10 11:38:19 2018 gautamUpdateGeneralITMY UL

Looking at Steve's plot, I was reminded of the ITMY UL OSEM issue. The numbers don't make sense to me though - 300um of DC shift in UL with negligible shifts in the other coils should have made a much bigger DC shift in the Oplev spot position.

  975   Mon Sep 22 12:06:58 2008 robUpdateSUSITMY UL OSEM


Last week I found the ITMY UL OSEM dead. I went around and checked the connections on the various flat ribbon cables
in the suspension control chain; pushing hard on the rack end of the long cable that goes from the sus electronics rack to the
ITMY sat amplifier fixed the problem. It's been fine since then.

NB: A visual inspection of the cable connection would not have revealed a problem. You just can't trust those flat
ribbon connectors with the hook latches.
  2024   Tue Sep 29 23:43:46 2009 robUpdateSUSITMY UL OSEM

We had a redo of elog entry 975 tonight.  The noisy OSEM was fixed by jiggling the rack end of the long cable.  Don't know exactly where--I also poked around the OSEM PD interface board.

In the attached PDF the reference trace is the noisy one.

  12638   Wed Nov 23 16:21:02 2016 gautamUpdateLSCITMY UL glitches are back

 

Quote:

As an aside, we have noticed in the last couple of months glitchy behaviour in the ITMY UL shadow sensor PD output - qualitatively, these were similar to what was seen in the PRM sat. box, and since I was able to get that working again, I did a similar analysis on the ITMY sat. box today with the help of Ben's tester box. However, I found nothing obviously wrong, as I did for the PRM sat. box. Looking back at the trend, the glitchy behaviour seems to have stopped some days ago, the UL channel has been well behaved over the last week. Not sure what has changed, but we should keep an eye on this...

I've noticed that the glitchy behaviour in ITMY UL shadow sensor readback is back - as mentioned above, I looked at the Sat. Box and could not find anything wrong with it, perhaps I'll plug the tester box in over the Thanksgiving weekend and see if the glitches persist...

  12643   Mon Nov 28 10:27:13 2016 gautamUpdateSUSITMY UL glitches are back

I left the tester box plugged in from Thursday night to Sunday afternoon, and in this period, the glitches still appeared in (and only in) the UL channel.

So yesterday evening, I pulled the Sat. Box. out and checked the DC voltages at various points in the circuit using a DMM, including the output of the high current buffer that supplies the drive current to the shadow sensor LEDs. When we had similar behaviour in the PRM box, this kind of analysis immediately identified the faulty component as the high current buffer IC (LM6321M) in the bad channel, but everything seems in order for the ITMY box. 

I then checked the Satellite Amplifier Termination Board, which basically just adds 100ohm series resistors to the output of the PD readout, and all the resistors seem fine, the piece of insulating material affixed to the bottom of this board is also intact. I then used the SR785 in AC coupled mode to look at the high frequency spectrum at the same points I checked the DC voltages with the DMM (namely the drive voltage to the LEDs, and the PD readout voltages on the PCB as well as on the pins of the connector on the outside of the box after the termination board (leading to the DAQ), and nothing sticks out here in the UL channel either. Of course it could be that the glitches are intermittent, and during my tests they just weren't there...

I am hesitant to start pulling out ICs and replacing them without any obvious signs of failure from them, but I am out of debugging ideas...


One possibility is that the problem lies upstream of the Sat. Box - perhaps the UL channel in the Suspension PD Whitening and Interface Board is faulty. To test, I have now hooked up ITMY Sat. Box. + tester box to the signal chain of ETMY. If I can get the other tester box back from Ben, I will plug in the ETMY sat. box. + tester to the ITMY signal chain. This should tell us something...

  12644   Tue Nov 29 11:07:37 2016 SteveUpdateLSCITMY UL glitches are back

400 days plot. Satelite amp ITMY has been swapped with ETMY

Unlabeled sat.amps are labeled. This plot only makes sense if you know the Cuh-Razy sat amp locations.

  16920   Wed Jun 15 17:03:17 2022 yutaUpdateSUSITMY ULCOIL issue solved, loose connection in sat amp box

[Anchal, Yuta]

We fixed the issue of ITMY ULCOIL not driving ITMY by replacing one of the 64pin ribbon cable in the satellite amplifier box.
We thought the coil driver and the sat amp box are OK by checking the voltage change at the output of the sat amp box by giving an offset to UL coil driver, but it was not giving a current change, probably due to too much contact resistance in the cables.
It was sneaky because it was not completely disconnected.

All the coils for our suspensions are now working!

What we did:
 - Using breakout boards, the output current of sat amp box was measured using FLUKE multimeter. It turned out that UL is not giving measurable current. We also confirmed that UR coil driver can drive UL by re-directing the current from UR coil driver to UL. This means that the UL magnet was not de-magnetized!
 - Measured the coil resistance from at the coil driver output and found that UL coil seen from there has too high resistance which cannot be measured with the multimeter, whereas UR coil was measured to be ~30 Ohms.
 - Went back to the feedthru and measured the resistance of UL coil. Upto the output of the Satellite Amp Terimator, the resistance was measured to be ~16 Ohms, but not at the input of the Satellite Amp Terimator (Attachment #1,2).
 - It turned out that #16 pin of 64pin ribbon cable in between the Satellite Amp Terimator (LIGO-D990021) and the Satellite Amp board (LIGO-D961289) at the Satellite Amp Terimator side was not good (Attachment #3).
 - Replaced the cable and confirmed that ULCOIL can kick ITMY (Attachment #4).
 - C1:SUS-ITMY_TO_COIL matrix was reverted to default values.

Next:
 - We might have to re-commission Yarm ASS again since pitch-yaw coupling have changed. -> EDIT: Checked that it works (except for ITM PIT L), including offloading offsets (writeASS_offsets.py), 18:30 local.
 - Now that LO1 LLCOIL issue is solved and LO2 stuck is solved, we should do the free swing test again to identify the resonant frequencies.
 - OSEM sensor diagonalization (input matrix), coil balancing (and F2A)

  16896   Tue Jun 7 17:26:21 2022 yutaUpdateSUSITMY ULCOIL mystery not solved

[Paco, Yuta]

We investigated the ITMY ULCOIL issue (40m/16873).
ULSEN is sensing the optic motion but ULCOIL cannot move the optic.
We confirmed that the coil input is there upto satellite amplifier output.
We also checked that ULCOIL have 3.3 mH and 16 Ohms, which are consistent with other coils.
Mystery remains...
We need to investigate ITMY ULCOIL in the next vent.

What we did:
 - Checked again that C1:SUS-ITMY_ULCOIL_OFFSET does not kick ITMY using OSEM sensor signals and oplev signals. ULSEN moves when ITMY is kicked by other coils.
 - Checked that kick gives voltage changes at coil driver and satellite amplifier output. We unplugged J1 DB25 cable from the feedthru flange and checked the signals sent to coil with oscilloscope.
 - Measured inductance (using BK PRECISION LCR meter) and resistance (using Fluke) of coils for ITMY. Below is the result. UL coil seems to be consistent with other coils. (It seems like BK PRECISION one wil give wrong resistance if the dial is set to the resistance value which is too low compared with the one you want to measure. If you want to measure 16Ω, set the dial to larger than 20Ω, not 2Ω)

Feedthru connector: ITMY1
Pin 3-15 / R = 16.3Ω / L = 3.32 mH (UL)
Pin 7-19 / R = 16.4Ω / L = 3.30 mH (UR)
Pin11-23 / R = 16.2Ω / L = 3.31 mH (LL)

Feedthru connector: ITMY2
Pin 3-15 / N/A
Pin 7-19 / R = 16.3Ω / L = 3.30 mH (SD)
Pin11-23 / R = 16.4Ω / L = 3.33 mH (LR)


Discussions:
 - UL is the only short OSEM in ITMY OSEMs.
 - ITMY have dumbells for magnets.
 - If UL magnet is off, ULSEN would not work. Something not magnetic is working for shadow sensing for UL? Dumbells?
 - ULSEN just sensing some coupling from other OSEMs?

  16903   Wed Jun 8 18:16:20 2022 yutaUpdateSUSITMY ULCOIL mystery: Coil driver swap test

To see if the ULCOIL channel of the ITMY coil driver is working or not, I swapped ITMY coil driver and ITMX coil driver by swapping DB15 cable (see Attachment #2).

With this swap, I confirmed that ITMX can be kicked with C1:SUS-ITMY_ULCOIL_OFFSET, but ITMY cannot be kicked with C1:SUS-ITMX_ULCOIL_OFFSET (see Attachment #1).

This means that the issue is not the in-air electronics.
Mystery remains again...
We need to investigate ITMY ULCOIL in the next vent.


I revereted the swap and confirmed that damping loops work fine again.

  16904   Thu Jun 9 23:08:39 2022 ranaUpdateSUSITMY ULCOIL mystery: Coil driver swap test

what was the result of the inductance measurement? should be ~3.3 mH as measured from the flannge or cable that goes to the flange from sat amp.

 

  16905   Fri Jun 10 13:02:14 2022 yutaUpdateSUSITMY ULCOIL mystery: Coil driver swap test

ITMY ULCOIL was measured to have ~3.3 mH as measured from the flange. RTFE 40m/16896 .
 

Quote:

what was the result of the inductance measurement? should be ~3.3 mH as measured from the flannge or cable that goes to the flange from sat amp.

 

 

  16908   Fri Jun 10 15:04:23 2022 ranaUpdateSUSITMY ULCOIL mystery: Coil driver swap test

Its good that the inductance test passed. This means that the coil is OK. How does the inspection photo look? This is the one you guys took of the ITM OSEM that shows the position of the magnet w.r.t. the coil. Also, how does the free swinging spectra look? Either one of these might indicate a broken magnet, or a sticky EQ stop.

  16910   Fri Jun 10 21:10:01 2022 yutaUpdateSUSITMY ULCOIL mystery: Coil driver swap test

We checked the photos we have, but we didn't have the photos which show ULCOIL situation clearly.

Free swing of ITMY (and others) will be done this weekend to see the OSEM spectra and resonant frequencies.

  17346   Thu Dec 8 16:21:40 2022 PacoSummaryCalibrationITMY actuation strength cal with 5 lines

[Anchal, Paco]

After debugging the hardware, on gpstime 1354422834 we turned on 5 cal lines on ITMY to test the ALS calibration for the single arm along with our error estimates.

Note: the YARM IR lock lasted > 8 hours, but the GRY transmission dropped twice during the evening and hopped back up, so the phase tracker jumped a couple of times.


Configuration

YAUX laser was locked to the YARM through the analog PDH servo (UGF ~ 2 kHz), YARM was locked to the PSL with POY11 (UGF ~ 200 Hz), and the ALS phase tracker was set to output the beat frequency noise in Hz. HEPA was left on during this measurement. The oscillators were similar to previous instances: gains of 70@211.1Hz, 100@313.31Hz, 100@315.17Hz, 300@575.17Hz and 15@30.92Hz with appropriate notches on ETMY to avoid POY11 loop supression.


Analysis

For YARM, the high bandwidth YAUX laser loop with transfer function G ensures that the relative laser frequency fluctuations correlate with the relative length fluctuations as:

\frac{\delta \nu}{\nu} = \frac{G}{1 - G} {\frac{\delta L}{L}}

Then, getting the magnitude of the YARM displacement at calibration frequencies is possible by knowing the arm cavity length, open loop gain, and absolute frequency (wavelength). The relative calibration error on the magnitude of the displacement is

\frac{\Delta \lvert \delta L\lvert}{\lvert \delta L \lvert} = \left[ \left(\frac{\Delta {L}}{L}\right)^2+ \left(\frac{\Delta {\lambda}}{\lambda}\right)^2 + \left(\frac{\Delta {{\delta \nu}}}{{\delta \nu}}\right)^2 + \left(\frac{\Delta \lvert G \lvert}{\lvert G \lvert(\lvert G \lvert - 1)}\right)^2 \right]^{1/2}

including the relative uncertainties in the YARM length, wavelength, and open loop gain. Interestingly, the loop gain term weighs proportionally less as G increases, so even if G = 100 (10), its relative error contribution would be < 1%. To estimate our total error, we assume the wavelength and YARM length are 1064.1(5) and 37.79(1), and add the frequency dependent values for G with 10% error. Finally, we use the rms ASD to estimate the relative error from the beatnote fluctuation measurement.

The measurement was done similar to other instances, taking the 'C1:ALS-BEATY_FINE_PHASE_OUT_HZ_DQ' timeseries (sampled at 16 kHz) and demodulating at the calibration frequencies above to get the mean YAUX laser frequency fluctuation and its uncertainty from the demodulated rms ASD.


Results

Attachment #1 shows the raw timeseries, Attachment #2 shows the spectra around the cal lines, Attachment #3 shows the demodulated timeseries, Attachment #4 shows the final result for the 5 lines, including the tallied errors as detailed above.

ITMY actuation = 4.92(11) nm / count / f^2


Discussion

We compared our results from Attachment #4 against a MICH referenced ITMY actuation calibration found here; which Yuta guess-timated a 10% uncertainty (gray shaded band in Attachment #4). An important correction came for the 575 Hz line, not just because the YAUX OLG is small but because a violin filter on ITMY LSC output has a 1.4475 gain bump. In fact we collected any additional digital gains from the ITMY output filters:

Output filter digital gains at cal lines (from foton)
30.92 Hz 211.1 Hz 313.31 Hz 315.31 Hz 575.17 Hz
1.00007 1.0034 1.0101 1.01017 1.4475

Next

  • Consider moving the 575 Hz line to avoid additional digital gain, but try to remain at high frequency.
    • Maybe here we can use the resonant gain MOKU filters that Radhika is designing.
  • Setup a live loop gain calibration to reduce the uncertainty for the high frequency cal line.
    • We can also just grab GTRY (transmission) as a proxy for optical gain and use for budgeting.
  • Work on setting up constraints for error mitigation based on allan deviation of the beatnote and PDH nonlinearity.
  • Move this to FPMI or some other lock configuration
  14627   Mon May 20 22:06:07 2019 gautamUpdateSUSITMY also kicked

For good measure:

The following optics were kicked:
ITMY
Mon May 20 22:05:01 PDT 2019
1242450319
  8172   Tue Feb 26 16:13:18 2013 BrettUpdateSUSITMY and ETMY mysterious loop gain difference of 2.5

While doing initial measurements for the new global damping infrastructure I discovered that the ETMY loop between the OSEM actuation and the OSEM sensors has a gain that is 2.5 times greater than the ITMY.  The result is that to get the same damping on both, the damping gain on the ETMY must be 2.5 times less than the ITMY. I do not know where this is coming from, but I could not find any obvious differences between the MEDM matrices and gains.

I uploaded a screenshot of measured transfer functions of the damped ITMY and ETMY sus's. Notice that the ETMY measurement is 2.5 times higher than the ITMY. The peak also has a lower Q, despite having the same damping filters running because of this mysterious gain difference. Lowering the damping gain of the ETMY loop by this 2.5 factor results in similar Q's.

  5422   Thu Sep 15 18:24:54 2011 PaulUpdateSUSITMY and SRM Oplev current status - comparison with ITMY

Just to find out where we are currently, I plotted the ITMY and SRM oplev spectra along with the ETMY oplev spectra. ETMY seems to be very good, so comparing with this seemed useful, so we know how much we have to improve by. The SRM power spectrum appears to be around 2 orders of magnitude higher than ETMY over pretty much the whole measurement band. The ITMY power spectrum is not so bad as the SRM above about 60Hz. Next thing to do is to check the dark noise level for the ITMY and SRM QPDs.

  5423   Thu Sep 15 18:31:27 2011 PaulUpdateSUSITMY and SRM Oplev current status - comparison with ITMY

Quote:

Just to find out where we are currently, I plotted the ITMY and SRM oplev spectra along with the ETMY oplev spectra. ETMY seems to be very good, so comparing with this seemed useful, so we know how much we have to improve by. The SRM power spectrum appears to be around 2 orders of magnitude higher than ETMY over pretty much the whole measurement band. The ITMY power spectrum is not so bad as the SRM above about 60Hz. Next thing to do is to check the dark noise level for the ITMY and SRM QPDs.

 The title of this post should of course have been " ... - comparison with ETMY" not " ... - comparison with ITMY"

  5418   Thu Sep 15 16:45:59 2011 PaulUpdateSUSITMY and SRM Oplev status

Today I worked on getting the ITMY and SRM oplevs back in working order. I aligned the SRM path back onto the QPD. I put excitations on the ITMY and SRM in pitch and yaw and observed the beam at the QPDs to check for clipping. They looked clean from clipping.

 
Measurements of the beam power at various points:
 
Straight after the laser - 7.54mW
After the BS in the SRM path - 1.59mW
After the BS in the ITMY path - 3.24mW
Incident on the SRM QPD - 0.03mW
Incident on the ITMY QPD - 0.25mW
 
Counts registered from the QPD sum channels:
 
SRM QPD SUM dark count - 1140
SRM QPD SUM bright count - 3250
 
ITMY QPD SUM dark count - 150
ITMY QDP SUM bright count - 12680
 
The power incident on the SRM QPD seems very low with respect to the ITMY QPD. Is the SRM mirror coating not very reflective for the He-Ne laser?There are some back reflections from lenses, which we should be careful of to avoid scattering.
  5501   Wed Sep 21 16:31:28 2011 PaulUpdateSUSITMY and SRM actuator response functions

 I divided the open loop transfer functions by the filter response and the sensor responses (previously measured calibration factors) to leave just the actuator responses. I've attached the actuator responses plotted in radians/count and phase over frequency.

Next step: fit the actuator response with poles and zeros.

EDIT: I divided by the wrong filter function earlier - the plots there now are divided by the correct filter function

  5510   Thu Sep 22 00:00:10 2011 PaulUpdateSUSITMY and SRM actuator response functions - complex fitting results

Here are the results of the complex fitting. The residuals are bigger this time, but still probably small enough to be ok(?), with the possible exception of ITMY PITCH (due again I think to the data points straddling the resonance).

ITMY YAW actuator response complex fit

-- Fit completed after 282 iterations--

 Started with: Gain = 3e-05,
 Q factor = 5,
 Pole frequency = 0.6776,
 Fit results:  Gain = 1.14673e-06,
 Q factor = 12.9471,
 Pole frequency = 0.766531
 Residual (normalised against the sum of input datapoints) = 0.0688174
 
ITMY PITCH actuator response complex fit
-- Fit completed after 191 iterations--
 Started with: Gain = 3e-05,
 Q factor = 5,
 Pole frequency = 0.6776,
 Fit results:  Gain = 1.25105e-06,
 Q factor = 3.88981,
 Pole frequency = 0.706744
 Residual (normalised against the sum of input datapoints) = 0.144165
 
SRM YAW actuator response complex fit
-- Fit completed after 246 iterations--
 Started with: Gain = 3e-05,
 Q factor = 5,
 Pole frequency = 0.6776,
 Fit results:  Gain = 3.34137e-06,
 Q factor = 9.6875,
 Pole frequency = 0.854913
 Residual (normalised against the sum of input datapoints) = 0.0153646
 
SRM PITCH actuator response complex fit
-- Fit completed after 266 iterations--
 Started with: Gain = 3e-05,
 Q factor = 5,
 Pole frequency = 0.6776,
 Fit results:  Gain = 7.97529e-06,
 Q factor = 7.63888,
 Pole frequency = 0.568227
 Residual (normalised against the sum of input datapoints) = 0.0319653
  5507   Wed Sep 21 23:05:16 2011 PaulUpdateSUSITMY and SRM actuator response functions - fitting results

 I used an fminsearch function to fit the SRM and ITMY actuator response magnitudes. The testfunction was just that for a single second order pole, but it gave what I consider to be good fits for the following reasons:

*for 3 of the 4 fits the residuals were less than 0.5% of the summed input data points. The worst one (ITMY pitch) was about 2.7%, which I think is due to the resonance happening to be right in the middle of two data points.

*the tolerance of 1 part in 10^9 was reached quickly from not very finely tuned starting points.

The test function was: G=abs(Gp./(1+1i.*f./fp./Qp-(f./fp).^2)), where G(f) is the actuator response magnitude, Gp is the pole gain, fp is the pole frequency, and Qp is the pole Q factor.

In the end I just fitted the response magnitude. I was initially fitting the complex response function, but ran into problems which I think were cased by overall phase offsets between the data and test function. Can I canvass for opinion if fitting the magnitude is OK, or should I try again fitting the phase too?

Anyway, here are the results of the fits, and I've attached plots of each too (each one in linear and log y axis because each on its own might be misleading for fits):

EDIT - I added more points to the otherwise sparse looking fitted curves

 

ITMY PITCH actuator response fit

-- Fit completed after 190 iterations--

 Started with: Gain = 3e-06,

 Q factor = 5,

 Pole frequency = 1,

 Fit results:  Gain = 1.32047e-06,

 Q factor = 4.34542,

 Pole frequency = 0.676676

 Residual (normalised against the sum of input datapoints) = 0.0268321

 

ITMY YAW actuator response fit

-- Fit completed after 156 iterations--

 Started with: Gain = 3e-06,

 Q factor = 5,

 Pole frequency = 1,

 Fit results:  Gain = 1.14456e-06,

 Q factor = 8.49875,

 Pole frequency = 0.730028

 Residual (normalised against the sum of input datapoints) = 0.00468077

 

SRM PITCH actuator response fit

 -- Fit completed after 192 iterations--

 Started with: Gain = 3e-06,

 Q factor = 5,

 Pole frequency = 1,

 Fit results:  Gain = 7.94675e-06,

 Q factor = 7.16458,

 Pole frequency = 0.57313

 Residual (normalised against the sum of input datapoints) = 0.00301265

 

SRM YAW actuator response fit

 -- Fit completed after 156 iterations--

 Started with: Gain = 3e-06,

 Q factor = 5,

 Pole frequency = 1,

 Fit results:  Gain = 3.34179e-06,

 Q factor = 9.57601,

 Pole frequency = 0.855322

 Residual (normalised against the sum of input datapoints) = 0.000840468

  5499   Wed Sep 21 14:44:25 2011 PaulUpdateSUSITMY and SRM open loop transfer functions

 

 Here are the open loop transfer functions for ITMY and SRM. The various settings for the OLTFs were as follows:

Oplev filter used for all OLTFs: 300^2:0

Gains for oplev servos (for each OLTF only the 1 servo for the measured TF was on. They are all set back to 0 now):

SRM yaw gain = 1

SRM pitch gain = -1

ITMY yaw gain = -1

ITMY pitch gain = 1

measurement band = 0.2Hz to 200Hz

points = 33

swept sine magnitude envelope: amp = 2 for f > 60Hz, amp = 0.1 for f < 60Hz

Measurement points were from e.g. C1-SUS-ITMY-OLPIT-IN2 to C1-SUS-ITMY-OLPIT-IN1 to give a TF of -(loop gain).

Next step is to divide this through by the sensor reponse (i.e. the calibration factor measured earlier) and the filter response to get just the actuator response. 

 

  5457   Mon Sep 19 12:23:30 2011 PaulUpdateSUSITMY and SRM oplev beam size reduced + next steps

I replaced the lenses that were there with a -150mm lens followed by a +250mm lens. This gave a significantly reduced beam size at the QPDs. With the beam analyzer up and running it should be possible to optimize this later this afternoon. Next I will remove the SRM QPD from the path and make measurements of the beam spot position movement and corresponding OSEM values for different DC mirror offsets. I will then repeat the process for ITMY.

  5487   Tue Sep 20 18:03:45 2011 PaulUpdateSUSITMY and SRM oplev calibrations - measured and estimated

The measured calibration factors for the oplevs are as follows:

 
SRM pitch: 666urad per count on channel C1-SUS-SRM-OLPIT-INMON
SRM yaw: 557urad per count on channel C1-SUS-SRM-OLYAW-INMON
 
ITMY pitch: 470urad per count on channel C1-SUS-ITMY-OLPIT-INMON
ITMY yaw: 491urad per count on channel C1-SUS-ITMY-OLYAW-INMON
 
Since I'm going to calibrate all the other oplevs with the rougher technique of estimating the angle from the OSEM signals directly, I thought I would check the result of such an estimation for the oplevs I have calibrated already. My method was as follows:
 
dA = change in angle
dx = change in OSEM flag position
dV = change in OSEM PD voltage
dC = change in OSEM counts
D = optic diameter
L = distance between OSEMs = D/sqrt(2)-0.002m = 0.052m
dV/dx = OSEMs volts per meter flag position change = 1700 V/m
dC/dV = OSEM counts per volt = 2^16/40 = 65536/40 counts/V
 
counts per radian = dC/dA = dV/dx  x   dC/dV   x  1/L = 1700*65536/40/0.052 = 5.3564x10^7 counts/rad
 
radians per count = dA/dC = 1.867x10^-8, or 0.019 urad/count
 
This is around a factor of 1000 smaller than what I measured earlier, reported in entry 5468. I guess this might be an issue with the whitening filter on the OSEMs, but my initial feeling was that this was only a factor of a few. If anyone can see a big obvious mistake in my above calculations please let me know!
 
 
  5488   Tue Sep 20 19:00:49 2011 PaulUpdateSUSITMY and SRM oplev calibrations - measured and estimated

 

Kiwamu noticed that the 1/L in the counts per radian should have just been L, which accounts for most of the discrepancy. We checked the input filters on the OSEMs, and they have 10dB of gain at DC. Accounting for this, estimates on the order of 20urad/count, which is much more reasonable!

  5494   Wed Sep 21 00:37:01 2011 ranaUpdateSUSITMY and SRM oplev calibrations - measured and estimated

I found that some of the Optical Lever Servos were ON today and injecting nonsense into the interferometer optics. I have set all of the gains = 0 to save us more headaches.

Please leave them OFF until we review the servo and noise characterization results in the elog.

  5496   Wed Sep 21 09:10:15 2011 PaulUpdateSUSITMY and SRM oplev calibrations - measured and estimated

Quote:

I found that some of the Optical Lever Servos were ON today and injecting nonsense into the interferometer optics. I have set all of the gains = 0 to save us more headaches.

Please leave them OFF until we review the servo and noise characterization results in the elog.

 I had previously set the gains to zero, see the first line of my entry on Monday 5468. I should have the servo and noise characterisation done today for these oplevs today, so we can review it soon.

  17618   Sat Jun 3 19:56:59 2023 PacoUpdateCalibrationITMY calibration with ALS and swept line

ITMY actuator calibrated with higher resolution using ALS

Today I came briefly in the lab and restarted all models because the beams were missing in the BHD camera; implying the ASS model controlling TTs has tripped into a weird state. The problem got fixed after that and I recovered the IFO alignment more or less. Thus I began another quick calibration run --


I wrote a script to put calibration line and update the LSC YARM notch filter using python-foton so as to take an ALS calibration sweep of the YARM test mass actuator responses. For a given sweep frequency array, the script loops over the frequency, updates the notchSensMat (LSC-YARM FM10), loads the coefficients, and ramps up the SENSMAT OSC gain and frequency to inject the noise. The YAUX laser is locked to the YARM and YARM is locked to PSL during this measurement. Then, the real-time demodulated BEAT_Y signals are averaged for ~ 3 seconds (just wanted a quick scan, not a very accurate or precise one) and the result is plotted in Attachment #1. Notice that I have included the phase tracker loop correction to account for the high frequency calibration line inferred response (1 + i f/2000)

I then used foton in gui mode and exported the violin filter transfer function (FM1-4 and FM6 for ITMY) magnitude and inverted its effect to get the "dc actuation strength" of ITMY.

Discussion

  • The value is consistent with our recently poor actuation due to c1sus - DAC-0 failure, at ~ 3 nm /count /f^2. I look forward to another measurement after the DAC-0 has been replaced / fixed.
  • No more "high frequency" lift as seen before?
  • There is still some residual frequency dependence near the phase tracker BW; maybe not just a simple integrator?

The idea of having the response shape is so I can then use the 5 lines only to get the response to high accuracy.

  17621   Mon Jun 5 18:42:13 2023 PacoUpdateCalibrationITMY calibration with ALS and swept line

ITMY actuator calibrated with higher resolution using ALS after c1sus-DAC0 replacement

I repeated the previous measurement (with fewer points) and the result is summarized in Attachment #1 -- just as a quick confirmation that the actuation strength (POS) is restored to its nominal ~ 5 nm / ct / f^2 value.

 (plot updated Tue Jun 6 10:34:27 2023 without violin filters and higher resolution)

  17624   Wed Jun 7 15:51:47 2023 KojiUpdateCalibrationITMY calibration with ALS and swept line

Can you make this flat by compensating the gain bump due to the UGF of the phase tracker?

 

  4878   Fri Jun 24 10:38:01 2011 steveUpdateCamerasITMY camera gets fixed

ITMY gets new Tamron M118FM50 that has improved close focusing. It is a small fixed focal length camera so the video tube cover can be put on.

The Watec LCL-902K 1/2" ccd camera was losing it power supply voltage because of bad connection. It was replaced.

  5351   Wed Sep 7 00:01:23 2011 SureshUpdateIOOITMY chamber ready for heavy doors

[Jenne, Suresh]

We did the following things in the ITMY chamber today:

1) We tried to get the ITMY stuck again by adjusting the coil gains so that it goes into the orientation where it used to get stuck.  We (reassuringly) failed to get it stuck again.  This, as we came to know later, is because kiwamu had rotated the side OSEM such that the optic does not get stuck . However the OSEM beam is at about 30 deg to the vertical and the SD is sensitive to POS motion now resulting in the poorer separation of modes as noted by Jenne earlier (5439)

2) We checked the earthquake stops and repositioned two at the bottom (towards the AR side of the optic)  which we had backed out earlier.

3) We took pics of all the OSEMS.

4) Checked to see if there are any stray beams with an IR card.  There were none.

5) I obtained the max values of the OSEMS by misaligning the optic with the coil offsets.  These values are in good agreement with those on the wiki

OSEM     UL     UR     LR     LL      SD

Max      1.80    1.53   1.68   1.96    2.10

Current  0.97   0.79    0.83   0.97   1.02

 

We can close the heavy doors tomorrow morning.

  16868   Fri May 20 20:03:48 2022 PacoUpdateBHDITMY chamber work finished - LO and AS overlapped

[Paco, Anchal, Yuta]

Today, in short we:

  • Recovered alignment of arm cavities, PRC (only ITMX aligned), and then altogether with SRM and PRM aligned to maximize all DCPD levels (AS, POP, REFL, TRX, TRY), but SRC was not flashing and the SRM yaw alignment slider was around its max value, so after recording beam positions on cameras Anchal went into the BS chamber and helped steer the SRC alignment using a combination of SRM, SR2 and AS1. After this every beam was nominally aligned except for LO and AS, which remained to be mode matched.
  • Mode matched LO3-LO4 by hand -- cheeky -- from the ITMY chamber, the final separation between these two mirrors grew by almost 3 inches with respect to the design (!!!) but the LO and AS beams came out nicely. The canonical path used for the steering was LO path, and then we overlapped the beams with the help of a gige basler camera and a couple of DCPDs (Thorlabs).
  • Yuta and Paco started running final checks in preparation for Monday (pumpdown). We aligned the IFO, but noted that using Restore/Misalign sometimes results in hysteresis.. so it is not very reliable for fine alignment modes. Then we optimized DC levels, centered all oplevs, and tweaked Green input alignment on XARM and YARM. The XARM was maximized, but in YARM we could still not get high TEM-00 flashing ...
    • Unfortunately, we discovered a slight clipping of the GTRY beam through PR3 which could mean the current alignment (pointing) is not hitting PR3 center optimally.
  • Attached are the screenshot of current aligned state after the work tonight, with oplevs centered, and the OSEM sensor values.

  16899   Tue Jun 7 19:40:45 2022 AnchalUpdateSUSITMY changed output matrix to disable use of UL coil

Since UL coil actuation is lost, we modified the output matrix of ITMY to use only UR, LR and LL face coils for POS, PIT and YAW actuation. The output matrix was changed to following:

  POS PIT YAW SIDE
UL 0 0 0 0
UR 1 1 0 0
LL 1 0 1 0
LR 0 -1 -1 0
SIDE 0 0 0 1

 

 

 

 

 

After this change, the damping was still working as good as before. I took PIT to POS/PIT/YAW and YAW to POS/PIT/YAW coupling measurements by exciting C1:SUS-ITMY_ASCPIT[YAW]_EXC and seeing effect at C1:SUS-ITMY_SUS[POS/PIT/YAW]_IN1 when the damping loops were off. Attached are the results. We were able to reduce PIT to YAW and YAW to PIT coupling by 10 dB by this simple change in output matrix. More coil balancing or off-diagonal termsmight help more and should be attempted if required. The coupling to POS did not change much.

Note that attachment 1 shows transfer functions from excitation point to the DOF sensing inputs while attachment two looks at ratio of C1:SUS-ITMY_SUS[POS/PIT]_IN1 to C1:SUS-ITMY_SUSYAW_IN1 which is the actual quantity of interest. I didn't repeat the PIT measurement due to lack of time.

Also note that all such measurements are being recorded in our new measurements git repo. We'll populate this repo with diaggui template+data files as we do measurements.

  5562   Wed Sep 28 07:36:41 2011 steveUpdateSUSITMY damping restored

ITMY suspention damping restored

  1188   Mon Dec 8 17:50:21 2008 YoichiUpdateSUSITMY drift
The suspension drift monitor shows that the ITMY alignment was shifted after the earthquake.
Looks like only the UL sensor had a step at the earthquake (see the attachment 1).
So it is probably an electronics problem.
I pushed in the cable between the rack and the ITMY satellite amplifier, but no change observed.
Actually, the ITMY-UL sensor looks like it has been dead before the earthquake.
The second attachment shows a long-term trend of the UL sensor.
The sensor output had been around zero since Nov. 17th.
When I disabled the output of the UL sensor, the sus-drift-mon fields turned green.
So I think the drift-mon's reference values are wrong, and currently the ITMY is in a good alignment.

I also attached the free-swing measurements of the ITMY taken on Aug. 18th and today.
There is no notable change in the resonant frequencies.
  16563   Mon Jan 10 15:45:55 2022 PacoUpdateElectronicsITMY feedthroughs and in-vac cables installed - part I

The ITMY 10" flange with 10 DSUB-25 feedthroughs has been installed with the cables connected at the in-vac side.  This is the first of two flanges, and includes 5 cables ordered vertically in stacks of 3 & 2 for [[OMC-DCPDs, OMC-QPDs, OMC-PZTs/Pico]] and [[SRM1, SRM2]] respectively from right to left. During installation, two 12-point silver plated bolts were stripped, so Chub had to replace them.

  16569   Tue Jan 11 10:23:18 2022 PacoUpdateElectronicsITMY feedthroughs and in-vac cables installed - part II

[Paco, Chub]

The ITMY 10" flange with 4 DSUB-25 feedthroughs has been installed with the cables connected at the in-vac side. This is the second of two flanges, and includes 4 cables ordered vertically in stacks of 2 & 2 for [[AS1-1, AS1-2, AS4-1, AS4-2]] respectively. No major incidents during this one, except maybe a note that all the bolts were extremely dirty and covered with gunk, so we gave a quick swipe with wet cloths before reinstalling them.

  12456   Wed Aug 31 18:07:43 2016 JohannesUpdateSUSITMY free swinging

[Lydia, Gautam, Koji, Johannes]

Summary of things done today:

  • Rebalanced ITMY table
    • After waiting until today to see if the table would relax into a level position, engaged the earthquake stops for SRM and moved the large counterweight by ~4 inches. The table is now level to within ~0.1 mrad in direction of the access port
    • Since the relaxing seems to take some time, we will open ITMX and ETMX chamber tomorrow and level the tables with additional weights, so the springs can get used to 'levelness' again
  • Cleaned ITMY, SRM and SR2 optics
    • Koji drag-wiped all three optics and cleaned the table in general where accessible. He was able to remove the sliver discussed in elog https://nodus.ligo.caltech.edu:8081/40m/12455
    • We measured the particle count in the chamber and found it to be 4000 for 0.3 microns and 660 for 0.5 microns.
  • We pulled out stops on ETMY ITMY and roughly centered the OSEMs half-way, using photos of the previous OSEM rotation as a reference point for their orientation. We foudn that the green beam is hitting ITMY almost centered and that the reflection doesn't seem to steer off too much, but were not yet able to see any returned light on the ETMY cameras.

Unless we get lucky and get the green light to flash in the cavity by playing with the mirror alignment, we will open the ETMY chamber tomorrow. On one hand we can look for the reflected green light in the chamber, or alternatively the IR beam transmitted by ITMY. This way we can obtain estimates for the OSEM biasing and perform the final centering of the OSEMs. We will then also address the bounce mode minimization in ITMY and check if the previous orientations still hold.

  14506   Mon Apr 1 22:33:00 2019 gautamUpdateCDSITMY freed

While Anjali is working on the 1um MZ setup, the pesky ITMY was liberated from the OSEMs. The "algorithm" :

  • Apply a large (-30000 cts) offset to the side coil using the fast system.
  • Approach the zero of the YAW DoF from -2.00V, PIT from +10V (you'll have to jiggle the offsets until the optic is free swinging, and then step the bias down by 0.1). At this point I had the damping off.
  • Once the PIT bias slider reaches -4V, I engaged all damping loops, and brought the optic to its nominal bias position under damping. 

While doing this work, I noticed several errors corresponding to EPICS channel conflicts. Turns out the c1susaux2 EPICS server was left running, and the MEDM screens (and possibly several scripts) were confused. There has to be some other way of testing the new crate, on an isolated network or something - please do not leave the modbus service running as it potentially interferes with normal IFO operation. For good measure, I stopped the process and shut down the machine since I saw nothing in the elog about any running tests.

Quote:

ITMY became stuck during this process

  14508   Tue Apr 2 15:02:53 2019 JonUpdateCDSITMY freed

blushI renamed all channels on c1susaux2 from "C1:SUS-..." to "C1:SUS2-..." to avoid contention. When the new system is ready to install, those channel names can be reverted with a quick search-and-replace edit.

Quote:

While doing this work, I noticed several errors corresponding to EPICS channel conflicts. Turns out the c1susaux2 EPICS server was left running, and the MEDM screens (and possibly several scripts) were confused. There has to be some other way of testing the new crate, on an isolated network or something - please do not leave the modbus service running as it potentially interferes with normal IFO operation. For good measure, I stopped the process and shut down the machine since I saw nothing in the elog about any running tests.

  14443   Fri Feb 8 02:00:34 2019 gautamUpdateSUSITMY has tendency of getting stuck

As it turns out, now ITMY has a tendency to get stuck. I found it MUCH more difficult to release the optic using the bias jiggling technique, it took me ~ 2 hours. Best to avoid c1susaux reboots, and if it has to be done, take precautions that were listed for ITMX - better yet, let's swap out the new Acromag chassis ASAP. I will do the arm locking tests tomorrow.

  14623   Mon May 20 11:33:46 2019 gautamUpdateSUSITMY inspection

With Chub providing illumination via the camera viewport, I was able to take photos of ITMY this morning. All the magnets look well clear of the OSEMs, with the possible exception of UR. I will adjust the position of this OSEM slightly. To test if this fix is effective, I will then cycle the bias voltage to the ITM between 0 and the maximum allowed, and check if the optic gets stuck.

  12455   Tue Aug 30 20:26:36 2016 gautamUpdateSUSITMY installed

[gautam, johannes, lydia]

Today we installed ITMY into position in the chamber.

  • First, we took the F.C coat off both faces
  • A stream of ionized nitrogen was used during the peeling process. We took as much care as possible not to blow towards the SRM. 
  • F.C. films came off smoothly. But when we looked at a picture we took prior to putting the optic in place, it looks like there may be a sliver of F.C. left on the optic. There are also a few specks of dust visible on the HR face, but well away from the clear aperture (see Attachment #1). Do we want to use isopropanol + optical tissue to try and remove these?
  • After F.C removal, we moved the optic into place against its stops. Returned OSEM connector tower to approximately its original place as it was moved to facilitate shifting the ITM to the edge of the table. 
  • I cleaned up the tangled mess of OSEM connector wires. On the ITMY tower, the OSEM cables have been tied using pieces of thin copper wire so as to avoid the wires straying into the beam path. Checked that wires are in grooves on both sides.
  • Unfortunately we were not able to start on setting up a cavity today, because when we checked the leveling of the ITM, we found that it was significantly not level. This is probably because the ITM was at the edge of the table. The cage is rather heavy and the location it was put in had a large lever arm. In any case, the table is slowly relaxing back to their usual state, Steve recommended we leave it overnight.
  • Other issues:
    • the UL sensor on ITMY also seemed to show some evidence of glitchy behaviour. Looking in the Satellite box, I didn't see any obvious probelms like I did for the ETMY box (for which I am not even sure if I did a legitimate fix anyways). I guess we have to keep observing and think about doing something about this if it really is problematic.
    • SRM barrel is pretty dusty. So is SR3. Do we want to clean these? If so how? F.C. or isoprop drag wipe?

We did some quick checks with the green beam and the IR beam. With the help of the custom Iris for the suspension towers, we gauged that both beams are pretty close to the center of the test mass. So we are in a not unreasonable place to start trying to align the beam. Of course we didn't check if the beam makes it to the ETM today.


The SRM OSEM sensor problem seems to have been resolved by moving the ITM back to its place as we suspected. The values are converging, but not to their pre-vent values (attachment #2). We can adjust these if necessary I guess... Or perhaps this fixes itself once the table returns to its neutral position. This remains to be monitored.


In the never-ending B-R mode reduction saga - we found what we think is an acceptable configuration now. Spectrum attached (Attachment #3). The top two OSEMs are now nearly 90 degrees rotated, while the bottom two are nearly horizontal. Anyways I guess we just have to trust the spectra. I should also point out that the spectra change rather significantly from measurement to measurement. But I think this is good enough to push ahead, unless anyone thinks otherwise?

  12720   Sat Jan 14 22:39:30 2017 ranaSummarySUSITMY is drifting ?

https://nodus.ligo.caltech.edu:30889/detcharsummary/day/20170114/sus/susdrift/

ITMY is not like the others. Real or just OSEM madness?

  5458   Mon Sep 19 13:13:10 2011 PaulUpdateSUSITMY oplev available for use: SRM not for the moment

 I've got the bench set up for the measurement of the beam spot change with DC SRM alignment offsets. The ITMY oplev is aligned and fine to use, but the SRM one isn't until further notice (probably a couple of hours).

  5326   Tue Aug 30 14:44:06 2011 kiwamuUpdateSUSITMY released without opening chambers

The ITMY mirror was released. The OSEM readouts became healthy.

 

To see what is going on, I changed the PIT DC bias slider on ITMY from 0.8 to -1 or so, and then the optic started showing a free swinging behavior.

If there were no responses to the DC bias, I was going to let people to open the chamber to look at it closer, but fortunately it released the optic.

Then I brought the slider back to 0.8, and it looked still free swinging. Possibly the optic had been stacked on some of the OSEMS as Jamie expected.

Quote from #5320

ITMY, which is supposed to be fully free-swinging at the moment, is displaying the tell-tale signs of  being stuck to one of it's OSEMs. 

Do we have a procedure for remotely getting it unstuck?  If not, we need to open up ITMYC and unstick it before we pump.

 

ELOG V3.1.3-