  4653   Fri May 6 15:42:55 2011 valeraMetaphysicsIOOInput mode cleaner length and 11 MHz modulation frequency

 After Kiwamu set the REFL11 phases in the PRMI configuration (maximized PRM->REFL11I reesponse) I tried to measure the MC length and the 11 MHz frequency missmatch by modulating the 11 MHz frequency and measuring the PM to AM conversion after the MC using the REFL11Q signal. The modulation appears in the REFL11Q with a good snr but the amplitude does not seem to go through a clear minimum as the 11 MHz goes through the MC resonance.

We could not relock the PRMI during the day so I resorted to a weaker method - measuring the amplitude of the 11 MHz sideband in the MC reflection (RF PD mon output on the demod board) with a RF spectrum analyzer. The minimum frequency on the IFR is 11.065650 MHz while the nominal setting was 11.065000 MHz. The sensitivity of this method is about 50 Hz.

  15567   Thu Sep 10 15:43:22 2020 JonUpdateBHDInput noise spectra for A+ BHD modeling

As promised some time ago, I've obtained input noise spectra from the sites calibrated to physical units. They are located in a new subdirectory of the BHD repo: A+/input_noises. I've heavily annotated the notebook that generates them (input_noises.ipynb) with aLOG references, to make it transparent what filters, calibrations, etc. were applied and when the data were taken. Each noise term is stored as a separate HDF5 file, which are all tracked via git LFS.

So far there are measurements of the following sources:

  • L1 SRCL
  • H1 SRCL
  • L1 DHARD PIT
  • L1 DSOFT PIT
  • L1 CSOFT PIT
  • L1 CHARD PIT

These can be used, for example, to make more realistic Hang's bilinear noise modeling [ELOG 15503] and Yehonathan's Monte Carlo simulations [ELOG 15539]. Let me know if there are other specific noises of interest and I will try to acquire them. It's a bit time-consuming to search out individual channel calibrations, so I will have to add them on a case-by-case basis.

  15181   Fri Jan 31 16:04:30 2020 gautamUpdateIOOInput pointing drift

One factor which hampers locking efforts is the apparent drift of the input beam into the IFO. Over timescales of ~1 hour, I have noticed that the spot on the AS camera drifts significantly (~1 spot size) in pitch. The IPPOS QPD bears out this observation, see Attachment #1. The IMC WFS control signals do not show a correlated drift, hence my claim that the TTs are to blame.

I am able to correct this misalignment by moving TT1 in pitch (see Attachment #2, which shows some signals from a ~1 hour PRMI lock, during which time the pointing drifted, and I corrected it by moving TT1 pitch). Assuming the problem is purely TT1 pitch drifting, this corresponds to 3mm / 6m ~500urad of shift in 1 hour - seems very large. The fact that the drift is only present in pitch and doesn't really show up in yaw makes me think the problem is likely mechanical (unless the voltage to the top two coils is drifting relative to the bottom, but no LR drift, which would be very coincidental). At the moment, this is just an annoyance, but it'd be good for this problem to be fixed.

I haven't yet figured out how to make ndscope export these plots to SVG preserving the dark color theme, hence the weird light axes...

Attachment 1: IPdrift.pdf
IPdrift.pdf
Attachment 2: IPdrift_PRMI.pdf
IPdrift_PRMI.pdf
  15841   Wed Feb 24 12:29:18 2021 gautamUpdateGeneralInput pointing recovered

While working at the LSC rack, I lost the input pointing into the IFO (the TT wiring situation is apparently very fragile, and this observation supports the hypothesis that the drifting TTs are symptomatic of some electronics issue). After careful beam walking, it was recovered and the dither alignment system was used to maximize TRX/TRY once again. No lasting damage done. If I can figure out what the pin-mapping is for the TT coils in vacuum, I'm inclined to try installing the two HAM-A coil drivers to control the TTs. Does anyone know where I can find said pin-out? The wiki page links seem broken and there isn't a schematic available there...

Ok it should be possible to back it out from the BOSEM pin out, and the mapping of the in-vacuum quadrupus cable, though careful accounting of mirroring will have to be done... The HAM-A coil driver actually already has a 15 pin output like the iLIGO coil drivers that are currently in use, but the pin mapping is different so we can't just replace the unit. On the bright side, this will clear up 6U of rack space in 1Y2. In fact, we can also consider hooking up the shadow sensor part of the BOSEMs if we plan to install 2 HAM-A coil drivers + 1 Dual satellite amplifier combo (I'm not sure if this number of spares is available in what we ordered from Todd).

  3114   Thu Jun 24 11:16:32 2010 Sharmila, Rana and KiwamuHowToVACInspection of the BS (sorry, no sounds)
  3093   Mon Jun 21 14:21:34 2010 Jenne, KiwamuUpdatePhotosInspection of Magnets for the TTs

Some pictures of "magnet inspection" from Picasa.

The coating of some magnets are chipped...

  3095   Mon Jun 21 20:11:21 2010 KojiUpdatePhotosInspection of Magnets for the TTs

Were these magnets chipped before the Ni plating?

RA: Yes, it looks like this is the case. We also smashed some of the magnets against a metal surface and saw that a black grime was left. We should hold the magnets with a clean teflon clamp to measure the Gauss. Then we have to wipe the magnets before installing. I share Jenne's concern about the press-fit damaging the plating and so we need to consider using using glue or the ole magnetic attachment method. We should not rely on the structural integrity of the magnets at all.

  15865   Thu Mar 4 23:57:35 2021 KojiSummaryElectronicsInspection of the new custom dsub cables

I made the inspection of the new custom DSub cables (came from Texas).

The shelled version gives us some chance to inspect/modify the internal connections. (good)
The wires are well insulated. The conductors are wrapped with the foils and then everything is in the braid tube shield. The braid is soldered on one of the connectors. (Attachment  3/4 shows the soldering of the conductor by intentionally removing one of the insulations).

It wasn't clear that if the conductors are twisted or not (probably not).

Attachment 1: 20210304235251_IMG_0527.jpg
20210304235251_IMG_0527.jpg
Attachment 2: 20210304235302_IMG_0528.jpg
20210304235302_IMG_0528.jpg
Attachment 3: 20210304235339_IMG_0529.jpg
20210304235339_IMG_0529.jpg
Attachment 4: 20210305000050_IMG_0530.jpg
20210305000050_IMG_0530.jpg
Attachment 5: 20210305000610_IMG_0531.jpg
20210305000610_IMG_0531.jpg
Attachment 6: 20210305000615_IMG_0532.jpg
20210305000615_IMG_0532.jpg
  14175   Wed Aug 22 00:22:05 2018 KojiSummaryElectronicsInspection of the possible dual backplane interfaces for Acromag DAQ

[Johannes, Koji]

We went around the LSC, PSL, IOO, and SUS racks to check how many dual backplane interfaces will be required.

Euro card modules are connected to the backplane with two DIN 41612 connectors (as you know). The backplane connectors provide DC supplies and GND connections.
In addition, they are also used for the input and output connections with the fast and slow machines.

According to the past inspection by Johannes, most of the modules just use the upper DIN41612 connector (called P1). But there are some modules exhibited the possibility of the additional use of the other connector (P2).

Tuesday afternoon Johannes and I made the list of the modules with the possible dual use. And I took a time to check the modules with DCC, Jay's schematics, and the visual inspection of the actual modules.

LSC Rack

  • Common mode servo (D040180 Rev B)
    • Schematic source D040180 Rev B D1500308
    • Assesment: Both P1 and P2 are to be connected to Acromag, but there are only a few channels on P2
    • P1: 1A-32A Digital In
    • P2: 1A-3A Analog Out (D32/33/34, SLOW MON and spare?)
            9A Digital Out for D35 (Limitter)
            10A-15A Spare
            16A Digital In (Latch Enable/Disable)
            25A, 25C  Differential Analog in (Differential offset input, indicated as "BIAS") 
  • PD Interface (D990543 Rev B)
    • Schematic source D990543 RevB
    • Assesment: No connection necessary. We don't monitor/control anything of any LSC PDs from Acromag.

PSL Rack

  • Generic DAQ Interface (D990155) - This is a DAC interface.
    • Schematic source: Jay's page D990155 Rev.B All the lines between P2 and P3 are connected.
    • Assesment: Only P2 is to be connected to Acromag.
    • P1 DAC mon -> not necessary
    • P2 A1-A16, Connected to DAC in P2-P3
  • PMC Servo
    • Schematic source: LIGO DCC D980352
    • Assesment: Only P1 (1A-9A) is to be connected to Acromag. (Just one DSub is sufficient)
    • P1 1A-9A
  • Crystal Ref (D980353)
    • Schematic source: LIGO DCC D980353
    • Assesment: Only P1 (1A-4A) is to be connected to Acromag. (Just one DSub is sufficient)
    • P1 1A-4A
  • TTFSS REV A
    • Schematic source: PNot found
    • Assesment: Probably Only P1 is sufficient. We need to analyze the board to figure out the channel assignment.

IOO Rack

  • PD Interface (D990543 Rev B)
    • Schematic source D990543 RevB
    • Assesment: Only P1 connection is sufficient.
  • Generic DAQ Interface (D990155)
    • Assesment: Remove the module. We already have the same module in PSL Rack. This is redundant.
  • Common mode servo (D040180 Rev B)
    • See above
  • Pentek Generic Input Board D020432
    • Schematic source Jay's page D020432-A
    • Assesment: No connection. There is no signal on the backplane.

SUS Rack

  • SUS Dewhitening
    • Schematic source: Jay's page D000316-A
    • Assesment: No connection.
    • We can omit Mon CHs.
    • Bypass/Inputs are already connected to the fast channels.

 

  14177   Wed Aug 22 12:22:27 2018 ranaSummaryElectronicsInspection of the possible dual backplane interfaces for Acromag DAQ

I think we don't need to keep Crystal Ref: we can change this into a regular Wenzel box with no outside control or monitoring.

Quote:

 

  • Crystal Ref (D980353)
    • Schematic source: LIGO DCC D980353
    • Assesment: Only P1 (1A-4A) is to be connected to Acromag. (Just one DSub is sufficient)
    • P1 1A-4A

 

  5672   Sat Oct 15 17:06:20 2011 KojiUpdateLSCInstallation REFL165

REFL165 was installed on the AP table last night.

Meanwhile I found the DC power level at the REFL PDs were 0.8~1.2V if the PRM is aligned and the IFO is not locked.
This corresponds to 16~24mA (20~30mW). This is too big.

The HWP of the REFL path were adjusted so that we have 6~10mA (8~12mW) on each PDs.

  4002   Wed Dec 1 02:39:00 2010 SureshUpdateSUSInstallation of ETMU07 as ETMX

[Kiwamu, Jenne, Koji, Suresh]

The following steps in this process were completed.

1)  Secured the current ETMX (Old ETMY) with the earth quake stops.

2) Removed the OSEMs and noted the Sl no. of each and its position

3) Placed four clamps to mark the location of the current ETMX tower (Old ETMY's position on the table)

4) Moved the ETMX (Old ETMY) tower to the clean table flow bench.  In the process the tower had to be tilted during removal because it was too tall to pass upright through the vacuum chamber port.  It was scary but nothing went wrong.

5) Koji calculated the location of the new ETMX and told us that it should be placed on the north end of the table.

6) Moved the OSEM cables, counter balancing weights and  the 'chopper' out of the way.  Had to move some of the clamps securing the cables.

7) Moved the ETMU07 tower from the clean room to the ETMX table

8) Positioned the OSEMs as they were placed in the earlier tower and adjusted their position to the middle of the range of their shadow sensors.  The four OSEMs on the face did not give us any trouble and were positioned as required.  But the side OSEM could not be put in place.  The magnet on the left side, which we are constrained to use since the tower is not designed to hold an OSEM on the right side, seems a little too low (by about a mm) and does not interrupt the light beam in the shadow sensor.  The possible causes are

   a) the optic is rotated.  To check this we need to take the tower back to the clean room and check the location of the optic with the traveling microscope.  If indeed it is rotated, this is easy to correct.

   b) the magnet is not located at the correct place on the optic.  This can also be checked on the clean room optical bench but the solution available immediately is to hold the OSEM askew to accommodate the magnet location.  If time permits the magnet position can be corrected.

We have postponed the testing of the ETMU07 tower to 1st of Nov Dec.

 

  10261   Wed Jul 23 11:15:54 2014 AkhilUpdateElectronicsInstallation of FCs in the 40m

 As a part of installation of two(X-ARM and Y-ARM) frequency counters in the 40m, I have tested their performance when using them both on a single Raspberry Pi. The timing plots are attached. There are almost no timing issues in this configuration and it can be said that there is no harm using both of the FCs on the same platform.

We will be installing the FC box inside the lab and carry out few tests with RF mon beat note inputs.

Attachment 1: Timingwith2FCs.png
Timingwith2FCs.png
  6090   Thu Dec 8 16:54:05 2011 JenneUpdateRF SystemInstallation of new demod box

So far:

* New demod box has been placed in the LSC rack.

* An extra set of +\- 24V has been prepared at the terminal block where all the Crates get their power on the LSC rack.  The grounds were all connected up, but the hot wires weren't.  So there were extra spots available, but none that were pre-wired with my voltages.

     * To do this I turned off all the power supplies in the short rack, since I decided to be a live chicken rather than a dead duck.  After hooking up the new terminal block slots, I turned on the power supplies.

 * Make a power cable to go from the terminal block to the box

Still to do:

 * Connect the spare 55MHz LO and the POP (or POX or POY) unused 11MHz LO to the back of the box

 * Move the PD inputs to the new demod box rather than the borrowed AS11 and POP55 boards

 * Plug the I & Q outs for each freq into some spare ADC channels - must first make sure we have 4 open channels.

 * See if it works.

  15500   Fri Jul 24 15:40:59 2020 JordanUpdateVACInstallation of two new UPS units

I installed the Tripp Lite SMX1000RT2U and Tripp Lite Smart1000LCD at the bottom of the 1x8 electronics rack. These are plugged in to power, and are ready for testing. All other cables (serial, usb, etc.) have been left on the table next to the 1x8 rack.

Attachment 1: UPS.jpg
UPS.jpg
  16826   Tue May 3 14:02:09 2022 AnchalSummaryBHDInstalled POP path in ITMX Chamber

[Anchal, JC]

I installed POP_SM4 and POP_SM5 in the ITMX chamber in the nominal positions. This must have affected the ITMX Oplev because I could see that one of the ITMX oplev beam was going through POP_SM5. It needs to be changed in order to follow the original plan. However, since POP_SM5 is a 1064 line mirror, it is transparent to the opleve beam, so maybe we can just use the ITMX oplev in the current fashion.

Next steps:

  • Get flashing back on the XARM.
  • Try to get the correct phase angle in POX11 so that we can lock XARM with IR too.
  • Inspect ITMX Oplev. The quadrant sum is low so maybe it needs adjustment in the in-ar table.
  • Check if ITMX oplev path needs to be adjusted inside the chamber.
  15916   Fri Mar 12 18:10:01 2021 AnchalSummaryComputer Scripts / ProgramsInstalled cds-workstation on allegra

allegra had fresh Debian 10 installed on it already. I installed cds-workstation packages (with the help of Erik von Reis). I checked that command line caget, caput etc were working. I'll see if medm and other things are working next time we visit the lab.

  10905   Thu Jan 15 18:06:34 2015 JenneUpdateComputer Scripts / ProgramsInstalled kerberos on Rossa

I have installed kerberos on Rossa, so that I don't have to type my name and password every time I do an svn checkin, since I'm making some modifications and want to be sure that everything is checked in before and afterwards. 

I ran sudo apt-get install krb5-user.  I didn't put in a default_realm when it prompted me to during install, so I went into the /etc/krb5.conf file and changed the default_realm line to read default_realm = LIGO.ORG

Now we can use kinit, but we must (as usual) remember to kdestroy our credentials when we're done.

As a reminder, to use:

> kinit albert.einstein

Password for albert.einstein@LIGO.ORG: (type your pw here)

When you're finished, run

> kdestroy

The end.

  10906   Thu Jan 15 18:10:19 2015 jamieUpdateComputer Scripts / ProgramsInstalled kerberos on Rossa
Quote:

I have installed kerberos on Rossa, so that I don't have to type my name and password every time I do an svn checkin, since I'm making some modifications and want to be sure that everything is checked in before and afterwards. 

I ran sudo apt-get install krb5-user.  I didn't put in a default_realm when it prompted me to during install, so I went into the /etc/krb5.conf file and changed the default_realm line to read default_realm = LIGO.ORG

Now we can use kinit, but we must (as usual) remember to kdestroy our credentials when we're done.

As a reminder, to use:

> kinit albert.einstein

Password for albert.einstein@LIGO.ORG: (type your pw here)

When you're finished, run

> kdestroy

The end.

WARNING: since the workstations are all shared user, if you forget to kdestroy the next user can commit under your user ID.  It might be good to set the timeout to be something much shorter than 24 hours, like maybe 1, or 2.

  10907   Thu Jan 15 18:30:18 2015 JenneUpdateComputer Scripts / ProgramsInstalled kerberos on Rossa

 

Quote:

WARNING: since the workstations are all shared user, if you forget to kdestroy the next user can commit under your user ID.  It might be good to set the timeout to be something much shorter than 24 hours, like maybe 1, or 2.

Good call.  I added a line ticket_lifetime = 3600, which should make it destroy the credentials after an hour.

  4511   Mon Apr 11 19:09:59 2011 SureshUpdateRF SystemInstalled low pass filters in the demod boards

 

As part of the RF system upgrade some of the demod boards in the lab were modfied.  The filter U5 (see the circuit schematic) was replaced. These changes are tabulated below.

 

Filters installed in the demod boards
Serial number Old name of the card New name of the card Filter installed Remarks
107 POY33 REFL33 SCLF-33+ R14=50Ohm
118 AP133, ASDD133 REFL55 SCLF-65  
114 PO199 REFL165 SCLF-190 R14=50Ohm
120 PO133 POP110 SCLF-135  
123 SP133 POP55 SCLF-65+ AT1 removed, R14=50Ohm
122 SP199, REFLDD199 AS165 SCLF-190  
121 SP166, REFL16 POP11 SCLF-10.7  
116 AP199 199 MHz POP165 SCLF-190  
126 AS166 33.3 MHz POX11 SCLF-10.7  
119 POX 33.3 MHz POY11 SCLF-10.7  
021 24.5 MHz (LLO) REFL11 SCLF-10.7  
020 24.5 MHz SCLF-45 POP22 SCLF-21.4  
022 24.5 MHz SCLF-45 AS11 with amp SCLF-10.7  
029 24.5          SCLF-f5 AS55 with amp SCLF-65  

 

Next, I and Q phase has to be checked for orthogonality. And noise levels of the cards have to measured.

 

 

 

  4514   Mon Apr 11 23:35:02 2011 ranaUpdateRF SystemInstalled low pass filters in the demod boards

I am a little concerned about using these low pass filters so close to the band edge. Recall that there is no on-board preamp for the RF input to the mixer.

So, if the input impedance of the filters is not 50 Ohms, we will get some unwanted reflections and sensitivity to cable length.

I think its worth while to check the impedance or S-parameters of these things with the LO activated to find out if we need to remove them or not.

  10890   Tue Jan 13 03:47:46 2015 ericqUpdateCDSInstalled new ethernet driver on Chiara

Chiara threw another network hissy fit. Dmesg was spammed with a bunch of messages like eth0: link up appearing rapidly. 

Some googling indicated that this error message in conjuction with the very ethernet board and driver that Chiara had in use could be solved by updating with an appropriate driver from the manufacturer.

In essence, I followed steps 1-7 from here: http://ubuntuforums.org/showthread.php?t=1661489

So far, so good. We'll keep an eye out to see how it works...

  4238   Wed Feb 2 09:56:55 2011 KojiSummaryGreen LockingInstalled the freq divider and Rana's PFD

- The freq divider and Rana's PFD were hooked up to the ADCs. (Attachment 1)
(I leave the analog PFD not explained in this entry.)
For this purpose, the VCO feedback signal has been disconnected and the beat signal was moved from the VCO loop to the analog PFD.

The output level of the splitter was +12dBm and was too high for the freq divider.
So, I had to stupidly add an attenuator of 10dB before the box.

- Gain of the digital PFD LPF

The LPF of the digital PFD had the gain of -4096 to let the output signal indicate the direct frequency reading.

The gain has been changed to -67.108864
such that the output shows the direct reading of the beat freq in the unit of MHz

-4096*2^14/10^6 = -67.108864

 

- Attachment 2 shows the acquired beat note through the freq divider.
The blue is the beat note between "green locked" and "IR locked only to MC" (i.e. MC vs XARM)
The red is the beat note with the both beam locked to the arm

The freq divider is a bit flaky in some freq region as the divided output sometimes shows freq jumps or the captured at a freq.
I still don't know why it happens. We have to check why this happens.

Attachment 1: freq_divider_installation.pdf
freq_divider_installation.pdf
Attachment 2: 110201_freq_divider_output.pdf
110201_freq_divider_output.pdf
  3650   Tue Oct 5 13:59:17 2010 Leo SingerConfigurationComputersInstalled x264-devel on Allegra

 I installed the package x264-devel on allegra.martian.  This package provides headers and libraries for the popular h264 video codec.  I am going to use this in the GStreamer streaming media server on Allegra.

  4806   Fri Jun 10 18:49:40 2011 kiwamuUpdateIOOIntensity Noise after the MC

Last night the relative intensity noise (RIN) of the beam after MC was measured.

It looks like the RIN is dominated by the motion of the MC mirrors, possibly the angular motions because we don't have any angular stabilization servos.

Suresh will estimate the contribution from the MC mirrors' angular motion to the RIN in order to see if this plot makes sense.

 

(RIN)

 The spectrum below 30 Hz seems to be dominated by the motion of the MC suspensions because it very resembles the spectra of those.

Above 30 Hz the spectrum becomes somehow flat, which I don't know why at the moment.

A rough estimation of the shot noise gave me a level of 10^{-9} RIN/sqrtHz, which is way below the measured spectrum.

RIN_afterMC.png

 

(Setup)

 All of the suspended mirrors were intentionally misaligned except for the MC mirrors and PRM to avoid interference from the other optics.

In this setup it allows us to measure the intensity noise of the laser which is transmitted from MC.

The beam transmitted from MC is reflected by PRM and finally enters into the REFL11 RFPD.

The DC signal from the RFPD was acquired at C1:LSC-REFL_DC_DQ as the laser intensity.

As well as the RIN measurement I took a spectrum when the beam is blocked by a mechanical shutter on the PSL table.

This data contains the dark noise from REFL11 and circuit noise from the whitening, AA board and ADC. It is drawn in black in the plot.

The cut off at 15 Hz is possibly due to the digital unwhitening (two poles at 15 Hz and two zeros at 150 Hz) filter to correct the analog whitening filter.

  8224   Mon Mar 4 19:38:21 2013 CharlesUpdateGeneralIntensity Stabilization and Control Systems

 I have been studying Jamies master's thesis concerning intensity stabilization of a solid-state laser (the 1064 nm specifically) to the ~10^-9 level, as well as relevant supporting material. I have also been reading about general control systems, photodiodes and acousto-optic modulators to help facilitate work on the ISS. 

Now that Altium has been properly installed, I have also begun familiarizing myself with the program and general libraries of boards and devices that have already been modeled with the program.

  4392   Wed Mar 9 18:17:11 2011 kiwamuUpdateGreen LockingIntensity noise coupling

Here is a new plot for the differential noise measurement. I plot a noise contribution from the intensity noise (brown curve).

If we believe this data, the differential noise is NOT dominated by the intensity noise of the PSL.

diff_noise.png

 


(intensity noise coupling measurement)

 Here is a plot for the transfer functions (TFs) from the intensity noise DCPD to the beat signal.

IN_TF.png

   In principle these TFs tell us how much intensity noise are contributed into the differential noise.

When I measured the spectra shown above, the frequency offset of the beatnote was at about 8 MHz from the zero cross point.

Keeping the same lock, I measured the transfer function (red curve) by using the swept sine technique on DTT. The setup for this measurement is depicted on the last entry (#4389).

Then I made the spectra above by multiplying the intensity spectrum by this TF.

  Later I measured another transfer function when the beatnote was at about 2 MHz from the zero cross point.

According to this measurement, our MFD gets insensitive to the intensity noise as the beat offset goes close to the zero cross point. This is consistent with what we expected.

  4397   Thu Mar 10 14:06:54 2011 kiwamuUpdateGreen LockingIntensity noise limits the beatnote sensitivity

We are limited by the intensity noise of the X arm transmitted green light.

Since the intensity noise from the PSL wasn't big enough to explain the differential noise (#4392), so this time I measured the noise contribution from the X arm transmitted light.
diff_noise_Mar8.png

 


(coupling measurement)

  IN_TF_complete.png

  I performed the same intensity noise coupling measurement, but this time between the DC signal of the beatnote RFPD and the beatnote signal.

 While measuring it, I excited the intensity of the PSL laser by using the same AOM like I did yesterday. This AM cause the observable intensity noise on the beatnote RFPD.

With the excited AM, we can pretend to have an excited AM on the green transmitted light from the X arm, of course assuming the intensity noise coupling from the PSL is less.

  4398   Thu Mar 10 14:22:58 2011 kiwamuUpdateGreen LockingIntensity noise limits the beatnote sensitivity

The next steps we should do are :

    - to investigate a cause of the intensity fluctuation
          * end green laser
          * suspensions' angular motions
          * doublecheck the RIN contribution if it's from the PSL or the X arm in the beatnote RFPD to make sure the RIN is dominated by the X arm transmitted light
  
    - to think about how to make the system insensitive to the intensity noise
          - bring the beat frequency to the zero cross point of the MFDs ?
          - PLL ?

Quote:

We are limited by the intensity noise of the X arm transmitted green light.

  4399   Thu Mar 10 14:29:05 2011 KojiUpdateGreen LockingIntensity noise limits the beatnote sensitivity

We can modify the freq divider circuit to make it a comparator.

Quote:

The next steps we should do are :

    - to investigate a cause of the intensity fluctuation
          * end green laser
          * suspensions' angular motions
          * doublecheck the RIN contribution if it's from the PSL or the X arm in the beatnote RFPD to make sure the RIN is dominated by the X arm transmitted light
  
    - to think about how to make the system insensitive to the intensity noise
          - bring the beat frequency to the zero cross point of the MFDs ?
          - PLL ?

Quote:

We are limited by the intensity noise of the X arm transmitted green light.

 

  4400   Thu Mar 10 14:30:53 2011 ranaUpdateGreen LockingIntensity noise limits the beatnote sensitivity

There are 3 standard techniques to reduce this effect:

1) Stabilize the end laser by sensing the green light coming into the PSL before recombination and feeding back with SR560 (this is the only one that you should try at first).

2) Moving to the center of the MFD fringe via ETM steps.

3) Auto-alignment of the beam to the arm.

  4383   Tue Mar 8 06:29:06 2011 kiwamuUpdateGreen LockingIntensity noise setup

[Jenne, Chris, Kiwamu]

 A photo diode and an AOM driver have been newly setup on the PSL table to measure the intensity noise coupling to the beat note signal.

We tried taking a transfer function from the PD to the beat, but the SNR wasn't sufficient on the PD. So we didn't get reasonable data.

 

(what we did)

  - put a DCPD after the doubling crystal on the PSL table. The PD is sitting after the Y1 mirror, which has been used for picking off the undesired IR beam.

  - installed the AOM driver (the AOM itself had been already in place)

  - injected some signals onto the AOM to see if we can see an intensity fluctuation on the PD as well as the beat signal

 

(intensity noise)

  In order to have better SNR for the intensity measurement, we put an AC coupled SR560 with the gain of 100 just before the ADCs.

When a single frequency signal was applied from a Stanford Research's function generator to the AOM, we could clearly see a peak at the doubled frequency of the injected signal.

Also a peak at the same frequency was found on the beat note signal as well.

But when random noise was injected from the same function generator, the random noise looked below the ADC noise.

Jenne adjusted the output voltage from the PD to about 1 V to avoid a saturation in the analog path, but later we realized that the ADC counts was marely ~ 20 counts.

So we will check the ADC tomorrow. Hopefully we will get a good SNR.

  4401   Thu Mar 10 16:00:53 2011 Aidan, JoeUpdateGreen LockingIntensity stabilization loop using beatnote DC.

Aidan: Joe and I have added a channel that takes the DC output from the vertex beatnote PD and sends it, via RFM, to a DAC at the ETMX end. Immediately before the output is a filter C1GCX_AMP_CTRL. The output of the DAC is connected to the CURRENT LASER DIODE modulation input on the back of the Innolight driver. This will modulate the current by 0.1A/V input.

We should be able to modulate the green laser on the end now and stabilize the intensity of the amplitude on the beatnote PD at the vertex. (In this configuration, the ampltiude noise of the PSL laser will be injected onto the end laser - we may want to revisit that).

Joe's comments on model change:

I added a RFM connection at the output of the C1:GCV-XARM_BEAT_DC filter in the c1gcv model.  The RFM connection is called: C1:GCV-SCX_ETMX_AMP_CTRL.

This RFM connection goes to the c1scx model and into Kiwamu's GCX box, which uses top_names.  There's a filter inside called AMP_CTRL, so the full filter name is C1:GCX-AMP_CTRL.  The output then goes to the 7th DAC output.

Photos:

  1. NPRO CURRENT CTRL plugged into DAC channel 7
  2. You can actually see it's channel 7 in this image
  3. The other end plugged into the back of the Innolight driver
  4. Schematic of the setup
  5. Updated C1ALS_OVERVIEW MEDM screen (I don't know why the field in the background turned orange - maybe it's coming into a long dry summer?)

 

Quote:

There are 3 standard techniques to reduce this effect:

1) Stabilize the end laser by sensing the green light coming into the PSL before recombination and feeding back with SR560 (this is the only one that you should try at first).

 

The reason that I chose this PD is that, apparently, the green light coming from the cavity is clipped when it is picked off for its DC PD.

Attachment 1: P1000313.jpg
P1000313.jpg
Attachment 2: P1000314.jpg
P1000314.jpg
Attachment 3: P1000315.jpg
P1000315.jpg
Attachment 4: GREEN_ISS_LOOP.pdf
GREEN_ISS_LOOP.pdf
Attachment 5: Screenshot-C1ALS_OVERVIEW.adl.png
Screenshot-C1ALS_OVERVIEW.adl.png
  4403   Thu Mar 10 21:45:34 2011 ranaUpdateGreen LockingIntensity stabilization loop using beatnote DC.

Ridiculous and hacky. Digital stabilization removed as well as the old "leave a pile of equipment on a stool" strategy.

We used a a BNC cable to send a pickoff of the beam before the recombination to the end via an SR560.

  10020   Tue Jun 10 12:49:46 2014 AkhilUpdateComputer Scripts / ProgramsInterfacing UFC-6000 with Raspberry Pi completed

 Goal:

 To interface the Mini Circuits RF Frequency Counter(FC) Model UFC-6000 with Raspberry Pi on Linux platform. Also to create a User friendly interface, to control the FC with command lines.

 

Highlights of the Code(script attached):

The  code enables the user to communicate and control different parameters of the UFC like:

1)Frequency Range Selection( for the device to read different frequencies, AutoRange is set by default).

2)Sampling Time (The time intervals for which the data will be retrieved)

3)Read Device Status(Whether the device is reading data or not).

 

Description of the Code:

HID USB Interfacing by sending byte Values.

 

 1)Read The Freq or Range 

Reading the Freq is done by reading the 1st and 2nd LCD of the Frequency counter.

1st line containing Range information, 2nd line is the Frequency result

the code should be send is 2

1st byte: 2

The returned 64 byte array is as follows:

1st byte: 2

2nd byte to Byte17 the ascii value of 16 characters of the 1st LCD line

Byte18 to Byte33  the ascii value of 16 characters of the 2nd LCD line

 

2) Set the Range  

By default Freq Counter is in "AutoRange" mode.

To set the range manually send the code 4

1st byte: 4

2nd byte: the range value. can be any legal range value. for auto range  need to be 255.

the 64 byte array is:

1st byte: 4

 

3)Set the Sample Time   

By default Freq Counter Sample Time is 1 sec.

you can set the sample time from 0.1 sec and up in step of 0.1 sec.To set the Sample Time send the code 3

1st byte: 3

2nd byte: the sample value in sec double 10.

for example: to set the sample time  to 0.4 sec 2nd byte need to be: 4

the 64 byte array is:

1st byte: 3

 

These bytes can be changed by changing the values of buffer[0] and buffer[1]  in function /*Send Report to the device*/ in the main program.

The data is written into a .txt file(example attached) and the user can  control the recording of data. The frequency data can now be made to talk to EPICS through slow channels.

The data from the .txt file can be used for error analysis at different sampling periods.

 

Results:

The interface of the FC with the Pi is now complete.

 

Plan:

 Make this FC talk to EPICS through slow channels.

 

 

 

 

 

 

Attachment 1: Interfacing_Script.zip
  2504   Mon Jan 11 16:59:14 2010 AlbertoUpdateABSLInterferometer Busy

I'm currently running a measurement on the PRC.

Please don't interfere with the IFO until it is done. Talk with Alberto if you've been intending to work inside the lab.

Thank you.

  6595   Thu May 3 13:07:05 2012 KojiBureaucracyGeneralInterferometer check-up session

From 2PM, Jenne/Den/Koji. Finished at 10:30PM

  • General
    • More LCD screens! Wider screens!
      • Allegra - only can handle single head right now, but should prepare for the eventual upgrade of that videocard/whole computer and get 2 monitors.
      • Every workstation should have 2 24" monitors.  We currently have 2 good ones (one on Rossa, one on Pianosa), so we need 6.
    • Remove Control room shelf currently holding OSA 'scopes
    • Adjust both TV shelves so that they're at the same height, ~6" higher than the current left shelf.
      This gives space for stacking the 24" monitors for each computer.
    • Audio system for the signals!!!! Even a crappy one!
           /cvs/cds/ligo/centos5.5/cds/project/gds-2.14.2/Monitors/rockIFO
           /cvs/cds/gds/gds/Monitors/rockIFO
           (really /cvs/cds, not /opt)

  • IOO
    • PZT or Picomotor mounts for PSL/ALS beams
    • ALS on the both arm simultaneously / common/diff ALS scripts
  • SUS
    • PRM OPLEV servos are sometimes not stable. (fixed, JCD/DM/KA, 5/2/2012)
    • ETMX oplev - replace 1064nm lens with 632nm lens (KBC037, -100)
  • IFO
    • It seems that there is an occasional common-mode power transient in the arm transmissions.
      -> Track down the cause and fix.
    • Fix arm ASS even on CentOS
    • Drift of the green incident axis -> Assess the amount of the drift / replace the mount
    • Calibration of POP22 / AS110
    • PMC/IMC/ARM characterization (loss, finesse, reflectivity, etc)
  • CDS
    • Capture OSA signals in CDS (the 'scope TDS1001B has a USB port in the back for connecting to the computer)
    • Transmon (arms) for high and low power
    • POX11 whitening is not toggling the analog whitening???
    • CentOS machines cannot open simulink models properly (get "Bad Link"s everywhere, so you can't do anything useful).
    • Dataviewer and DTT can't connect to framebuilder from CentIOS machines
  • Electronics
    • Actuator noise level characterization (coil driver response in V/m & coil driver noise level V/rtHz)
    • Beat box
    • I/Q DFD
    • Improvement of POP22/110/AS110 RF circuits.
  • MEDM
    • OL alert in the watch dog screen (this is awesome!) should have small text saying "OL" (done, JCD, 5May2012)
    • Complete 40m overview screen - everything should be clickable with pseudo 3D icons
    • BETTER SUSPENSION SCREEN!!!
    • Script to generate a MEDM tree
    • Resurrect MEDM snapshots
       
    • The LSC screen has some "white" boxes -> investigate and fix them (done, KA, 5/2/2012)
    • Make the LSC control screen (compact) in the vertical arrangement (done, KA, 5/2/2012)
    • The signals of the watch dog screen should go red if any of the WDs are not enabled.
      =>Copy the logic of the signals in LSC screen.
      (done, KA, 5/2/2012)
    • Remove c1gfd from the cds status screen (done, KA, 5/2/2012)
    • Add "BURT" status indicators on the cds status screen (done, KA, 5/2/2012)
  • SCRIPT
    • Locking scripts with integrator triggers / disabling when unlocked
    • Daily diagnosis of the MC spot positions (there must be something already...)
    • Daily/occasional adjustment of the incident axis on the MC
    • Suspension "misalign/restore" script is too wild => make a new script (Done, JCD, 7May2012)
    • IFO_CONFIGURE scripts still do a burt restore, so step the optics.  Need to remove optic alignment from config .snap files, use reg restore script instead.
    • OPLEV/OSEM trending script before the IFO work for diagnosis.
    • Auto-locker for PMC/Arm/etc
  • Video
    • If each video screen has a caption, that would be great
    • GUI interface of "videoswitch" Mike!
    • Mouse Jiggler for Zita (called Swinput?)
  • Ubuntu
    • burttoday is not aliased in ubuntu. burttoday:      aliased to cd /opt/rtcds/caltech/c1/burt/autoburt/today/
    • ASS doesn't run on Ubuntu!
      ezcawrite: error while loading shared libraries: libca.so: cannot open shared object file: No such file or directory
      ezcaswitch: error while loading shared libraries: libca.so: cannot open shared object file: No such file or directory
       
  2271   Sun Nov 15 18:42:10 2009 AlbertoUpdateLockingInterferometer fully locked for 3331 seconds

This afternoon, I tried to lock the interferometer again after a few days.

After a couple of failed attempts, and relocks of the MZ, the interferometer stayed locked continuously for about 50 minutes, with arm power of about 92.

I just wanted to check the status of the interferometer so I didn't do any particular measurement in the meantime.

  10613   Wed Oct 15 20:10:29 2014 ericqUpdateLSCInterim DARM Signal

I've done some preliminary modeling to see if there is a good candidate for an IR DARM control signal that is available before the AS55 sign flip. From a DC sweep point of view, ASDC/(TRX+TRY) may be a candidate for further exploration. 


As a reminder, both Finesse and MIST predict a sign flip in the AS55 Q control signal for DARM in the PRFPMI configuration, at a CARM offset of around 118pm.

dcAS55_DARM.pdfAS55Flip.pdf

The CARM offset where this sign flip occurs isn't too far off of where we're currently losing lock, so we have not had the opportunity to switch DARM control off of ALS and over to the quieter IR RF signal of AS55. 


Here are simulated DC DARM sweep plots of our current PRFPMI configuration, with a whole bunch of potential signals that struck me. 

Although the units of most traces are arbitrary in each plot (to fit on the same scale), each plot uses the same arbitrary units (if that makes any sense) so slopes and ratios of values can be read off. 

dcDARMSweep-300.pdfdcDARMSweep-120.pdfdcDARMSweep-50.pdfdcDARMSweep-0.pdf

In the 300 and 120pm plot, you can see that the zero crossing of AS55 is at some considerable DARM offset, and normalizing by TRX doesn't change much about that. "Hold on a second," I hear you say. "Your first plots said that the sign flip happens at around 120pm, so why does the AS55 profile still look bad at 50pm?!" My guess is that, probably due to a combination of Schnupp and arm length asymmetry, CARM offsets move where the peak power is in the DARM coordinate. This picture makes what I mean more clear, perhaps:

2dSweep.pdf

Thus, once we're on the other side of the sign flip, I'm confident that we can use AS55 Q without much problem. 


Now, back to thoughts about an interim signal:

ASDC by itself does not really have the kind of behavior we want; but the power out of AS as a fraction of the ARM power (i.e. ASDC/TRX in the plot) seems to have a rational shape, that is not too unlike what the REFLDC CARM profile looks like.

Why not use POPDC or REFLDC? Well, at the CARM offsets we're currently at, POPDC is dominated by the PRC resonating sidebands, and REFLDC has barely begun to decline, and at lower CARM offsets, they each flatten out before the peak of the little ASDC hill, and so don't do much to improve the shape. Meanwhile, ASDC/TRX has a smooth response to points within some fraction of the DARM line width in all of the plots. 

Thus, as was discussed at today's meeting, I feel it may be possible to lock DARM on ASDC/(TRX+TRY) with some offset, until AS55 becomes feasible.

(In practice, I figure we would divide by the sum of the powers, to reduce the influence of the DARM component of just TRX; we don't want to have DARM/DARM in the error signal for DARM)

Two caveats are:

  • The slope of this signal actually looks more quadratic than linear. Is this ok/manageable?
  • I have not yet made any investigation into the frequency dependent behavior of this thing. Transmission in the denominator will have the CARM pole in it, might get complicated. 

[Code and plots live in /svn/trunk/modeling/PRFPMI_radpressure]

 

  10618   Thu Oct 16 16:21:42 2014 ericqUpdateLSCInterim DARM Signal

I've added (TRX-TRY)/(TRX+TRY) to the DC DARM sweep plots, and it looks like an even better candidate. The slope is closer to linear, and it has a zero crossing within ~10pm of the true DARM zero across the different CARM offsets, so we might not even need to use an intentional DARM offset. 

dcDARMSweep-300.pdfdcDARMSweep-120.pdfdcDARMSweep-50.pdfdcDARMSweep-0.pdf

 

  14980   Mon Oct 21 11:44:19 2019 gautamUpdatesafetyInterlock reconnected to Innolight controller

We also took this opportunity to re-connect the interlock to the Innolight controller (after it was disconnected for diagnosing the mysterious NPRO self-shutdowns). The diode pump current was dialled down to 0, the interlock wires reconnected, and then the diode current was ramped back up to the nominal 2.1 A. The fan to cool the unit remains mounted in a flaky way as we couldn't locate the frame Chub had made for a more secure mounting solution. 

It seems like the pointing of the beam out of the laser head varies somewhat after the startup - I had to adjust the pointing into the PMC a couple of times by ~1 full turn of the Polaris mount screws, but the IMC has been locked (mostly) for the last ~16 hours.

Quote:

I've checked the state of the laser interlock switch and everything looked normal.

  17307   Wed Nov 23 17:21:44 2022 JCUpdateVACInterlocks may have been tripped due to N2 pressure loss

[Paco, JC]

While changing out one of the N2 tanks today, one of the fitting stripped. This caused a major loss of pressure. I replaced one fitting then realized there was a second leak around the area of the gauge. Paco and I changed this and everything should be back up and running. Thhe interlocks may have been tripped  within the last 2 hours.

 

Attachment 1: Screen_Shot_2022-11-23_at_5.22.57_PM.png
Screen_Shot_2022-11-23_at_5.22.57_PM.png
Attachment 2: Screen_Shot_2022-11-23_at_5.23.09_PM.png
Screen_Shot_2022-11-23_at_5.23.09_PM.png
  4625   Wed May 4 13:51:51 2011 valeraConfiguration Intermittent MC3 UL PD signal

The attached plot shows the 30 day trend of the MC3 UL PD signal. The signal dropped to zero at some point but now it is close to the level it was a few weeks ago. There still could be a problem with the cable.

The rest of the MC1,2,3 PD signals looked ok.

Attachment 1: mc3ulpdmon.pdf
mc3ulpdmon.pdf
  2598   Fri Feb 12 14:19:28 2010 rana, steveHowToloreInternational Fax

Steve showed me how to send an international fax today:

  1. Load paper.
  2. Dial:   011 - (country code) - number
  3. Press START (either the black or color option)
  4. wait for the screaming fax noise
  5. Done

 

  6560   Tue Apr 24 14:30:08 2012 JamieUpdateCDSIntroducing: rtcds, a utility for interacting with the CDS RTS/RCG

The new rtcds utility

I have written a new utility for interacting with the CDS RTS/RCG system: "rtcds".  It should be available on all workstations and front-end machines, but certain commands are restricted to run on certain front end machines (build, start, stop, etc.).  Here's the help:

controls@c1lsc ~ 0$ rtcds help
Usage: rtcds <command> [arg]

Available commands:

  build|make SYS      build model
  install SYS         install model
  start SYS|all       start model
  restart SYS|all     restart running model
  stop|kill SYS|all   stop running model
  list                list all models for current host

controls@c1lsc ~ 0$ 

Please use this new utility from now on when interacting with RTS.

I'll be improving and expanding it as we go.   Please let me know if you encounter any problems.

 

  3193   Mon Jul 12 11:20:56 2010 Gopal HowToCOMSOL TipsIntrusions (Negative Extrusions)

For the sake of future users, I have decided to periodically add tips and tricks in using COMSOL that I have figured out, most probably after hours of circuitous efforts. They will always be listed under the new COMSOL Tips category.

Today's topic: Intrusions

COMSOL has a very user-friendly interface for taking objects from 2D to 3D using the "extrusion" feature. But suppose one wants to design an object which contains screw holes or some other indentation. I've found that creating "punctures" in COMSOL is either impossible or very complicated.

Instead, COMSOL encourages users to always "add" to the object. In other words, one must form the lowest level first, then build layers sequentially on top using new work plane and boolean difference operators. This will probably be a bit clearer with an example:

1) First, create the planar projection in a work plane:

Screen_shot_2010-07-12_at_10.51.22_AM.png

2) Extrude the first layer only in the regular fashion:

Screen_shot_2010-07-12_at_10.51.28_AM.png

 3) Add a new work plane which is offset in the z-direction to the deepest point of the intrusion.

Screen_shot_2010-07-12_at_10.52.08_AM.png

 4) Now, create the shape of the intrusion in this new work plane.

Screen_shot_2010-07-12_at_10.53.53_AM.png

5) Use the Boolean "Difference" to let COMSOL know that, on this plane, the object has a hole.

 Screen_shot_2010-07-12_at_10.54.36_AM.png

 6) Extrude once more from the second work plane to complete the intrusion.

Screen_shot_2010-07-12_at_10.55.36_AM.png

  3194   Mon Jul 12 12:16:50 2010 DmassHowToCOMSOL TipsIntrusions (Negative Extrusions)

 An entry on the 40m wiki page might serve you better, and be easier to sift through once all is said and done

  2615   Fri Feb 19 02:38:32 2010 KojiConfigurationoplevsIntsant green oplevs for ITMs shooting from the ends

I set up instant green oplevs for ITMs.

A green laser pointer has been set on each end table. It illuminates the ITM center. The beam goea through the ETM substrate.
The reflected green beam returns to the ETM if the ITMs are aligned. Even though the reflected beam to the end is too big, this can
be a rough reference for each ITM.

Note: The green laser pointer at the ETMX were borrowed from Frank. We must return it to him when we finish the work.

  2083   Mon Oct 12 18:37:55 2009 ZachUpdatePSLInventory

--Apologies for the late post--

I was at the PSL table taking an inventory of the components for a while after Koji, Steve, and Kiwamu were there. I set the HEPAs back to 20% when I left (assuming that they were turned up when the compartment was opened).

ELOG V3.1.3-