  11693   Thu Oct 15 10:59:12 2015 KojiUpdateLSCDRFPMI Locked for 20 secGreat
job!
Many thanks Eric, Gautam, and all
  
  11710   Fri Oct 23 19:27:19 2015 KojiUpdateCDSFrequency counting - workable setup preparedIt is a nice scan. I'm still thinking
about the equivalence of the moving average
and the FIR low pass that I have mentioned
  
  11720   Wed Oct 28 14:07:44 2015 KojiUpdateIOOMC WFS Offsets updateMC WFS offsets were updated to have a better
operating point.
 
  
  11725   Mon Nov 2 17:39:01 2015 KojiFrogsGeneralDRFPMI celebrationやったー!Yatta!  yatta.jpg 
  11747   Tue Nov 10 11:40:03 2015 KojiUpdateLSCUpdated interpretation of peaksWhat is the uncertainty of your RoC estimation?

One measurement of the ETMY ROC
was 57.6m, but we trust another measured
  
  11748   Tue Nov 10 11:41:56 2015 KojiUpdateLSCUpdated interpretation of peaksFYI: I've also reported the similar
mod depths of

11M: 0.194
  
  11753   Wed Nov 11 22:11:15 2015 KojiUpdateSUSPSL/IOO maintenance, TM SUS check upPMC

- Before any of the following work,
I went to the PSL table and aligned the PMC.
 TM_SUS.pdfTM_SUS_SD.pdf 
  11754   Wed Nov 11 22:50:39 2015 KojiConfigurationCDSSlow machine time&dateI was gazing at the log file for Autolocker
script (/opt/rtcds/caltech/c1/scripts/MC/logs/AutoLocker.log
)
  
  11763   Fri Nov 13 22:32:54 2015 KojiSummaryPSLPMC LO degraded, usual ERA-5 replacement, LO recovered[Yutaro, Koji]

We found that the PMC LO level
was fluctuating in a strage way (it was not
 PMC_LO.pngIMG_2093.JPGIMG_2091.JPGIMG_2095.JPGIMG_2096.JPG 
  11764   Sat Nov 14 00:52:25 2015 KojiSummaryCDSInvestion on EPICS freeze[Yutaro, Koji]

Recently "EPICS Freeze"
is so frequent and the normal work on the
  
  11765   Sun Nov 15 22:43:48 2015 KojiSummaryPSLPMC LO degraded, usual ERA-5 replacement, LO recoveredI think the IMC locking was somewhat improved.
Still it is not solid as long time before.

Before the PMC fix (attachment
 C1-MULTI_E6875C_TIMESERIES-1131408017-86400.pngC1-MULTI_E6875C_TIMESERIES-1131580817-86400.png 
  11768   Mon Nov 16 19:05:59 2015 KojiSummaryPSLPMC servo circuit review, follow up measurementsPMC follow up measurements have been done.
The servo circuit was reviewed.

Now the PMC, IMC, X/Y arms are
  
  11769   Mon Nov 16 21:32:49 2015 KojiSummaryPSLPMC servo circuit review, follow up measurementsThe result of the precise inspection
for the PMC servo board for the 40m was done.

The record, including the
  
  11773   Tue Nov 17 15:49:23 2015 KojiConfigurationIOOMC Autolocker modified/opt/rtcds/caltech/c1/scripts/MC/AutoLockMC.csh 
was modified last night.

1. Autolocker sometimes forget
  
  11775   Tue Nov 17 16:21:10 2015 KojiSummaryPSLPMC servo circuit review, follow up measurementsI'm still analyzing the open loop TF
data. Here I report some nominal settings
of the PMC servo
  
  11780   Wed Nov 18 16:51:58 2015 KojiSummaryPSLPMC servo calibrationSummary

The PMC servo error (MIX OUT MON
on the panel) and actuation (HV OUT MON)
 PMC_err_cal.pdf 
  11781   Wed Nov 18 16:53:17 2015 KojiSummaryPSLPMC PZT capacitanceThe PMC PZT capacitance was measured.
- Turn off the HV supplies. Disconnect
HV OUT cable.
  
  11782   Wed Nov 18 17:09:22 2015 KojiSummaryPSLPMC Servo analysisSummary

The PMC servo was analysed. OLTF
was measured and modeled by ZPK (Attachment
 PMC_OLTF.pdfPMC_DSP.pdf 
  11791   Thu Nov 19 17:06:57 2015 KojiConfigurationCDSDisabled auto-launching RT processes upon FE bootingWe want to startup the RT processes one
by one on boot of FE machines.

Therefore /diskless/root/etc/rc.local
  
  11793   Fri Nov 20 15:44:12 2015 KojiSummaryPSLAdded 17.5kHz LPF to the PMC servoAs a final tune of the PMC servo, I've
added 1nF cap at the error signal amplification
stage. The diagram has been updated and uploaded
 PMC_OLTF.pdfPMC_noise_comparison.pdf 
  11794   Sat Nov 21 00:45:30 2015 KojiUpdateIOOIMC fixBased on the observation of the PMC error
signal, I started measuring the IMC OLTF.
Immediately, it
 PDRFsignal_cable.JPGQsignal_cable1.JPGQsignal_cable2.JPGQsignal_cable3.JPG 
  11795   Sat Nov 21 00:46:33 2015 KojiUpdateIOOIMC OLTFHere is the comparison before and after
the fix.

Before the work, the UGF was ~40kHz.
 IMC_OLTF.pdfPMC_noise_comparison.pdf 
  11797   Sun Nov 22 12:07:09 2015 KojiUpdateIOOIMC fixTO DO

Done (Nov 23, 2015) -
Check if the attenuator is still there in
  
  11798   Sun Nov 22 12:12:17 2015 KojiUpdateIOOIMC OLTFWell. I thought a bit more and now I think
it is likely that this is just the servo
bump as you can see in the closed-loop TF.
  
  11800   Mon Nov 23 20:32:43 2015 KojiUpdateLSCFrequency source fixed, IMC LO level adjustedThe frequency source was fixed. The IMC
LO level was adjusted.

IMC is locked => OLTF measured
  
  11801   Mon Nov 23 21:48:49 2015 KojiUpdateLSCFrequency source fixed, IMC LO level adjustedThe trouble we had: the 29.5 MHz
source had an output of 6 dBm instead
of 13 dBm.
 IMG_2136.JPGIMG_2129.JPGIMG_2133.JPG 
  11802   Mon Nov 23 22:12:10 2015 KojiUpdateIOOLO level check for the IMC demod boardIn order to check the proper LO level,
the IMC demod board was checked. As
a short summary, -8dBm is the proper input
 IMG_2137.JPGIMG_2138.JPG 
  11804   Tue Nov 24 01:14:23 2015 KojiUpdateLSCALSY recoveredSorry, I completely forgot to turn the
Marconi on...
  
  11808   Wed Nov 25 10:07:10 2015 KojiUpdateIOOLO level check for the IMC demod boardI didn't finish making the DCC entry
for this module yet.

But the attenuators are
  
  11811   Wed Nov 25 16:46:10 2015 KojiUpdateIOOLO level check for the IMC demod boardAwwww. I found that the demod board has
the power splitter (PSCJ-2-1) with one output
unterminated.
 D990511-40m_151123.pdf 
  11817   Thu Nov 26 19:39:27 2015 KojiUpdateLSCCurrent state of the frequency source, and possible improvementUploaded on T1000461
too.
 RF_Frequency_Source.pdf 
  11822   Sun Nov 29 12:32:26 2015 KojiUpdateLSCCurrent state of the frequency source, and possible improvementI'm not claiming we need to modify
the frequency source immediately as we are
not limited by the oscillator amplitude or
  
  11826   Mon Nov 30 15:17:57 2015 KojiUpdateLSCLO level check for the LSC RF distribution boxT1000461
tells us that the nominal LO input is 2dBm
although we don't know what's the
  
  11829   Mon Nov 30 18:27:30 2015 KojiUpdateLSCstrange behavior of ASDCIt wasn't fully mentioned in ELOG
11814.
We checked the PD first and this behavior
  
  11834   Tue Dec 1 17:26:14 2015 KojiUpdateGeneralMegatron maitenenceSLOWDC servo was dead. I followed EricQ's
instruction.
  
  11835   Tue Dec 1 20:20:16 2015 KojiUpdateGeneralMegatron maitenenceMC Autolocker got stack somewhere. I had
to go to megatron and kill MC Autolocker.

init relaunched the autolocker
  
  11852   Sat Dec 5 21:28:33 2015 KojiHowToCamerasWhen image capture does not work...Do we need "make" everytime?
Do you mean just running "modprobe"
didn't work?
  
  11867   Wed Dec 9 18:45:58 2015 KojiSummaryGeneralNetwork Topology Check[Eric Q, Gautam, Koji]

We went through the network connections
to produce the mapping of the instruments.
 Network_topology_9Dec2015.xlsxNetwork_topology_9Dec2015.pdf 
  11873   Fri Dec 11 13:28:36 2015 KojiUpdateLSCPower recycling gain estimation from arm loss measurementCan I ask you to make a plot of the power
recycling gain as a function of the average
arm loss, indicating the current loss value?
  
  11876   Fri Dec 11 23:12:09 2015 KojiSummaryCOCLoss map measurement documentYutaro left detailed slides for his loss
map measurement

https://dcc.ligo.org/LIGO-G1501547
  
  11920   Thu Jan 7 19:04:25 2016 KojiUpdateLSCAUX X Freq Noise measuredThe next step is to compare this data with
the same measurement with the PSL and the
AUX laser on the PSL table (or the end Y
  
  11923   Fri Jan 8 22:37:17 2016 KojiUpdateGeneralNew WiFi routerI configured a new wifi bridge for a GPIB
Instruments.

The some facts are described on
  
  11941   Thu Jan 21 00:02:11 2016 KojiUpdateLSCHopeful signsThat's a good news. Only quantitative
analysis will tell us if it is true or not.

Also we still want to analyze the
  
  11964   Sat Jan 30 09:56:24 2016 KojiUpdateGreen LockingInnolight laser is 10 years oldIt is strange that there is no difference
between with and without NE, isn't it?
  
  11968   Mon Feb 1 15:43:18 2016 KojiUpdateGreen LockingInnolight laser is 10 years oldThis is the same one as what you got from
Steve. But you can find full pages.

https://wiki-40m.ligo.caltech.edu/PSL/NPRO
  
  11971   Tue Feb 2 18:54:02 2016 KojiUpdateGreen LockingLightwave NPRO moved from PSL table to SP tablejiIn fact, it is one of the most difficult
type mode profiling to measure a beam directly
out from a laser source.
  
  11974   Thu Feb 4 09:16:46 2016 KojiUpdateGreen LockingLightwave NPRO moved from PSL table to SP tableI don't think the discrepancy is a
serious issue as long as the mode is clean.
The mode is determined by the NPRO crystal
  
  11989   Fri Feb 12 19:07:52 2016 KojiUpdateGreen LockingLaser swap - green light recovered but no flashes in the arm800e-6 / 0.225^2 = 0.016

=> 1.6%/W

I thought Kiwamu had roughtly 2%/W.
  
  11991   Mon Feb 15 13:09:33 2016 KojiUpdateGeneralSomething has gone wrong - was there a power outage?Looks like that's the case. LIGO GC
also sent an e-mail that there was a popwer
glitch.
  
  12000   Fri Feb 19 15:12:38 2016 KojiSummaryPEMGuralp Health CheckI measured the guralp raw outputs
and the TFs using the handheld unit and an
FFT analyzer.
 Guralp_Raw_PSD.pdflow_vel_comparison.pdfGuralp_TF.pdf 