  11540   Fri Aug 28 20:20:26 2015 ranaUpdatePEMGur interface box cableTo help find out if Steve really melted
the inside of our precious seismometer, lets
hook it up using the handheld seismo wand
  
  11539   Fri Aug 28 20:15:49 2015 ranaUpdateIOOMC2 -> MCL Actuator TF


  
  11538   Fri Aug 28 19:05:53 2015 ranaUpdateIOOIMC TweakWell, green looks better than blue, but
it makes the PCDRIVE go high, which means
its starting to saturate the EOM drive. So
  
  11537   Fri Aug 28 17:04:12 2015 SteveUpdatePEMGur interface box cableAra! Onara suru tsu-mori  fixedNewLongCable.png 
  11536   Fri Aug 28 02:20:35 2015 IgnacioUpdateLSCPRFPMI and MCL FFA day late
but here it is.

Eric
 carmin.pngcarmout.png 
  11535   Fri Aug 28 00:59:55 2015 IgnacioUpdateIOOFinal SISO FF Wiener Filter for MCLThis is my
final
 bode.png2.png3.png4.png6.png 
  11534   Thu Aug 27 04:23:04 2015 ericqUpdateLSCPRFPMI is backGot to a 40 minute lock tonight. All other
locks broke because of me poking something. 

I redid some sensing excitations,
  
  11533   Thu Aug 27 02:09:14 2015 ericqUpdateLSCAUX X Laser Current ChangedI spent some time tonight chasing down
the cause of huge RIN in the X green PDH
transmitted light, which I had started seeing
  
  11532   Thu Aug 27 01:41:41 2015 IgnacioUpdateIOOTriply Improved SISO (T240-X) FF of MCL   1.png2.png4.png5.png3.png 
  11531   Tue Aug 25 16:39:06 2015 SteveUpdatesafetysafety trainingAlessandra Marrocchesi received 40m specific
basic safety training yesteday.
  
  11530   Tue Aug 25 16:33:31 2015 Ignacio, SteveConfigurationPEMSeismometer enclosure copper foil progressSteve ordered
about two weeks ago a roll of 0.5 mm thick
copper foil to be used for the inside of
 picture1.jpgpicture2.jpgIMG_1966.JPGIMG_1969.JPG 
  11529   Tue Aug 25 16:09:54 2015 ericqUpdateIOOIMC TweakA little more information about the IMC
loop tweak...

I increased the overall IMC loop
 imcTweak.pdfCARM_TF.pdf 
  11528   Tue Aug 25 04:15:51 2015 ericqUpdateLSCPRFPMI is backMore PRFPMI locks tonight. Right now, it's
been locked for 22+ minutes, though with
the PRMI still on 3F signals. I think the
  
  11527   Mon Aug 24 16:46:49 2015 ericqUpdateSUSETMY Oplev laser power is fallingRepeated for all optics, ETMY seems like
the only one sharply dropping for now (PRM
is all over the place and hard to gauge,
 OLsum_trend.pdf 
  11526   Mon Aug 24 16:10:07 2015 ericqUpdateSUSETMY Oplev laser power is fallingToday I noticed the box around the ETMY
oplev sum flashing red, as it dipped below
1k. I don't recall seeing this recently,
 ETMYsum_trend.pdf 
  11525   Mon Aug 24 14:05:47 2015 ericqUpdateSUSEricG Investigating L2AThis afternoon, I showed Eric Gustafson
some of the basics of making swept sine measurements
with DTT. We turned off the f2a filters and
  
  11524   Sat Aug 22 15:48:32 2015 KojiSummaryLSCArm locking recoveryAs per Ignacio's request, I restored
the arm locking.

- MC WFS relief
  
  11523   Fri Aug 21 17:15:13 2015 SteveUpdatePEMGur interface box  




  
  11522   Fri Aug 21 08:38:37 2015 SteveUpdateIOOPMC lockedPMC needed to be locked manually.  PMClocked.png 
  11521   Thu Aug 20 18:08:28 2015 IgnacioFrogs40m upgradingFatality. Something broke.

So
 image1.JPG 
  11520   Thu Aug 20 11:31:55 2015 KojiUpdatePEMGur interface box As reported previously, the transfer functions
of the channels look fine. (i.e. All channels
almost identical.)
 IMG_1958.JPG 
  11519   Thu Aug 20 11:09:10 2015 ranaUpdateIOOsome points about seismic FF
When plotting the subtraction performance,
we mainly care about the 0.5 - 10 Hz band,
so we care about the RMS in this band. Don't
  
  11518   Thu Aug 20 02:31:09 2015 ericqUpdateLSCPRFPMI is backPRFPMI
locking has been revived. 

I've
had 6 5min+ locks so far; arm powers usually
 lock.pdf 
  11517   Wed Aug 19 07:58:25 2015 SteveUpdatePEMGur interface box Koji and Steve,

We took transferfunctions of each
channel yesterday. They were 
 GurADCbox.jpg 
  11516   Wed Aug 19 01:45:10 2015 IgnacioUpdateIOODoubly Improved SISO (T240-X) FF of MCLToday I tried
and doubly-improved SISO FF filter on MCL.
This filter has a stronger rolloff than the
 6x 
  11515   Wed Aug 19 00:55:35 2015 IgnacioUpdateLSCLSC-YARM-EXC to LSC-YARM-IN1 TF measurement + error analysisYesterday,
Rana, Jessica and I measured the Transfer
function from LSC-YARM-EXC to LSC-YARM-IN1. 
 master_plot.pngcoherence.png 
  11514   Tue Aug 18 11:16:17 2015 SteveUpdatePEMGur interface box is wonkyThe Guralp ADC  interface box 
D060506 is ready for inspection. It is in
front of 1X1 with open top and running.
 IMG_0009.JPG 
  11513   Tue Aug 18 03:56:09 2015 ericqUpdateLSClocking effortsNow that the updated ALS is stable, and
the PRC angular FF is revived, I've been
working on relocking PRFPMI. While the RMS
 newALSref.pdfxCoh.pdfyCoh.pdf 
  11512   Mon Aug 17 17:48:12 2015 KojiUpdatePEMWasps obliterated maybe...We found the same wasp in the 40m. Megan
found it walking behind Steve desk!
  
  11511   Sun Aug 16 23:26:40 2015 EveUpdateGeneralGaussianity testsI've continued to work on my Gaussianity
tests for S5 L1 data. 

Following the statistical measure
 Gaussianity_noc1.pngGaussianity_noc1.png 
  11510   Sat Aug 15 02:10:35 2015 IgnacioUpdateLSCMCL FF => YARM FFIn my last
post I calculated the different subtractions
(offline) that could be done to YARM alone
 YARM_LIVES.png 
  11509   Fri Aug 14 23:49:34 2015 KojiSummaryGeneralB&K Shaker fixedI fixed a shaker that was claimed to be
broken. I had to cut the rubber membrane
to open the head.
 IMG_1947.JPG 
  11508   Fri Aug 14 21:40:26 2015 IgnacioUpdateLSCQuick static offline subtractions of YARMPlotte below
are the resultant subtractions for YARM using
different witness configurations,
 arms_wiener.png 
  11507   Fri Aug 14 17:20:01 2015 JenneUpdatePEMGur interface box is wonkyIIRC, the Guralp box's 3rd set of channels
do not have all of the modifications that
were made on channels 1 and 2.
  
  11506   Fri Aug 14 12:10:08 2015 KojiUpdatePEMGur interface box is wonkyLet's dismantle the I/F unit from the
rack and connect the cable with the lid open.
We need to trace the signal.
  
  11505   Fri Aug 14 09:07:45 2015 SteveUpdatePEMGur interface box is wonkyAtm1, Before cable swap

Atm2, The long cables were swapped
at the input of the interface box.
 Gur1&2.pnglongCablesSwapped.png 
  11504   Thu Aug 13 23:57:33 2015 IgnacioUpdateLSCYARM coherence plotsI took data
from 1123495750 to 1123498750 GPS
time (Aug 13 at 3AM, 50 mins of data) for
 YARM_psd.pngYARM_GUR1_COH.pngYARM_STS_COH.pngYARM_GUR1_COH.png 
  11503   Thu Aug 13 20:32:07 2015 IgnacioUpdateLSCWorking towards YARM FF The
mode cleaner FF static filtering is by no
means done. More work has to be done
  
  11502   Thu Aug 13 12:06:39 2015 Jessica SummaryIOOBetter predicted subtraction did not work as well OnlineYesterday I adjusted the preweighting of
my IIR fit to the transfer function of MC2,
and also managed to reduce the number of
 6x 
  11501   Wed Aug 12 22:33:36 2015 IgnacioUpdateIOORe-measured MC2 -> MCL TFSince I will need to do transfer function
measurements in order to implement FF for
the arms and the MC2's yaw and pitch
 bode_TF.pngCoherence.png 
  11500   Wed Aug 12 16:48:26 2015 IgnacioUpdateIOOBetter? Nope. MISO WIener (T240-X and T240-Y) FF of MCLLast night, I also worked on a "better"
(an improvement, I thought) of the MISO Wiener
filter (T240-X and T240-Y witnesses) IIR
 6x 
  11499   Wed Aug 12 16:39:46 2015 IgnacioUpdateIOOMISO WIener (T240-X and T240-Y) FF of MCLLast night I performed some MISO FF on
MCL using the T240-X and T240-Y as witnesses.
Here are the results:
 6x 
  11498   Wed Aug 12 14:35:46 2015 ericqUpdateComputer Scripts / ProgramsPDFs in ELOGI've tweaked the ELOG code to allow
uploading of PDFs by drag-and-drop into
the main editor window. Once again we can
 smooth.pdf 
  11497   Wed Aug 12 11:44:13 2015 ericqUpdatePEMGur2 Channels still wonkyIn previous elogs, we saw that the X and
Y spectra out of GUR2 (X end Guralp seismometer)
looked strange (i.e. inconsistent with the
 gur2_centered.png 
  11496   Wed Aug 12 01:32:18 2015 IgnacioUpdateIOOImproved SISO (T240-X) FF of MCLIn my previous elog:11492,
I stated that in order to improve the subtraction
 filter.pngstsx.pngmclonline.pngyarmonline.pngsub.png 
  11495   Tue Aug 11 18:43:42 2015 JessicaUpdateIOOMCL Online SubtractionToday I finished fitting the transfer function
to a vectfit model for seismometers T240_X
and T240_Y, and then used these to filter
 SeisX_bode.pngSeisY_bode.pngMCL_first.pngYarm_first.png 
  11494   Tue Aug 11 16:13:28 2015 EveUpdateGeneralGaussianity testsI’m working on a code to determine
the Gaussianity of a PSD.

 
  
  11493   Tue Aug 11 11:56:36 2015 Ignacio, JessicaUpdatePEMWasps obliterated maybe...The wasp terminator came in today. He obliterated
the known wasp nest.

 IMG_1942.JPGIMG_1943.JPGIMG_1944.JPG 
  11492   Tue Aug 11 11:30:19 2015 IgnacioUpdateIOOSISO (T240-X) FF of MCLLast night we finally got some online subtraction
going. The filter used is described in the
post this eLOG is @eLOG
 MCL.pngMCTRANS.pngYARM.png 
  11491   Tue Aug 11 10:13:32 2015 JessicaUpdateGeneralConductive SMAs seem to work bestAfter testing both the Conductive and Isolated
front panels on the ALS delay line box using
the actual beatbox and comparing this to
 best4.pngisolated4.png 