  10986   Sat Feb 7 13:34:11 2015 KojiSummaryPSLISS AOM driver checkI wanted to check the status of the ISS.
The AOM driver response was measured on Friday
night.
 AOM_drive.pdfAOM_response.pdf 
  11000   Wed Feb 11 03:41:12 2015 KojiUpdateLSCPRC error signal RF spectraAs the measurements have been done under
feedback control, the lower RF peak height
does not necessary mean
  
  11005   Wed Feb 11 18:11:46 2015 KojiSummaryLSC3f modulation cancellation33MHz sidebands can be elliminated by careful
choice of the modulation depths and the relative
phase between the modulation signals.
  
  11012   Thu Feb 12 11:59:58 2015 KojiUpdateLSCNew Locking Paradigm - Loop-gebraThe goals are:

- When the REFL path is
dead (e.g. S_REFL = 0), the system goes back
  
  11019   Thu Feb 12 23:47:45 2015 KojiUpdateLSC3f modulation cancellation- I built another beat setup on the PSL
table at the South East side of the table.
- The main beam is not touched, no
  
  11027   Sat Feb 14 00:42:02 2015 KojiUpdateGeneralRF amplifier for ALSThe RF analyzer was returned to the control
room. There are two beat notes from X/Y confirmed.

I locked the arms and aligned them
  
  11028   Sat Feb 14 00:48:13 2015 KojiUpdateLSC3f modulation cancellation[SUCCESS]
The 3f sideband cancellation seemed worked
nicely.
  
  11029   Sat Feb 14 19:54:04 2015 KojiSummaryLSC3f modulation cancellationOptical
Setup

[Attachment 1]
 beat_setup1.JPGbeat_setup2.JPGelectrical_setup.pdf 
  11031   Sat Feb 14 20:37:51 2015 KojiSummaryLSC3f modulation cancellationExperimental
results

- PD response [Attachment
 beat_pd_response.pdfbeat_nominal.pdf3f_reduction.pdfbeat_3f_reduced.pdf 
  11032   Sat Feb 14 22:14:02 2015 KojiSummaryLSC[HOW TO] 3f modulation cancellationWhen I finished my measurements, the modulation
setup was reverted to the conventional one.
If someone wants to use the 3f cancellation
 freq_gen_box.JPGdelay_line.JPGcable_spec.pdf 
  11033   Sun Feb 15 16:20:44 2015 KojiSummaryLSC[ELOG LIST] 3f modulation cancellationSummary
of the ELOGS

3f
modulation cancellation theory 
  
  11035   Mon Feb 16 00:08:44 2015 KojiSummaryLSC[ELOG LIST] 3f modulation cancellationThis KTP crystal has the maximum allowed
RF power of 10W (=32Vpk) and V_pi = 230V.
This corresponds to the maximum allowed
  
  11036   Mon Feb 16 01:45:12 2015 KojiSummaryLSCmodulation depth analysisBased on the measured modulation profiles,
the depth of each modulation was estimated.
Least square sum minimization of the
 modulation_nominal.pdfmodulation_3f_reduced.pdf 
  11038   Mon Feb 16 03:10:42 2015 KojiUpdateLSCALS fool measured decoupling TFWonkey shape: Looks like a loop supression.
Your http://nodus.ligo.caltech.edu:8080/40m/11016
also suggests it too, doesn't it?
  
  11044   Tue Feb 17 16:44:04 2015 KojiUpdateLSCDelay line installed againFor tonight's experiment, I re-installed
the delay line cable and changed the attenuation
to 10dB for the 55MHz modulation.
  
  11045   Tue Feb 17 19:49:51 2015 KojiUpdateLSCDelay line un-installed againThe modulation setting was reverted.
Demod phase for REFL11/33/55/165 and
AS55 were reverted to the previous numbers
  
  11048   Wed Feb 18 19:06:40 2015 KojiUpdateLSCALS Fool impulse response   latex-image-1.png 
  11059   Mon Feb 23 21:57:13 2015 KojiUpdateLSCDelay line installed again (experiment, round 1)Last Wednesday we tried PRMI 3f modulation
cancellation. Under the 3f modulation cancellation,
the PRMI could not be locked
  
  11066   Wed Feb 25 12:16:27 2015 KojiUpdateLSCDelay line re-installed, measurements round 2WHAT?
WHAT? WHAT? It's obviously opposite.

If
the reflectivity of the front mirror is fixed
  
  11074   Thu Feb 26 01:53:35 2015 KojiUpdateLSCModelled effect of relative modulation phaseOk... This is what I was afraid of, and
it seems true.
i.e. the relation ship of the modulations
  
  11117   Sun Mar 8 00:05:37 2015 KojiConfigurationLSCCARM and DARM on RF signals!!!!!!!!!!!!!!!!!!!!Exciting! How long was it?   
  11129   Tue Mar 10 19:59:13 2015 KojiFrogsCamerasMessage from the IFO   IMG_1825.JPG 
  11134   Wed Mar 11 19:15:03 2015 KojiSummaryLSCROUGH calibration of the darm spectrum during the full PRFPMI lockI made very rough calibration of
the DARM spectra before and after the transition
for the second lock on Mar 8.
 150308_DARM_ROUGH_CALIB.pdf 
  11135   Wed Mar 11 19:48:25 2015 KojiUpdateGeneralFOL troubleshootingThere is a frequency counter code written
by the summer student.
The code needed some cleaning up.
  
  11137   Thu Mar 12 11:57:38 2015 KojiUpdateGeneralFOL troubleshootingBTW, during this trouble shoot, we looked
at the IR beatnote spectrum between the Xend
and the PSL.
  
  11151   Fri Mar 20 13:29:33 2015 KojiUpdateIOOWaking up the IFOIf the optics moved such amount, could
you check the PD alignment once the optics
are aligned?
  
  11171   Wed Mar 25 18:27:34 2015 KojiSummaryGeneralSome maintainance- I found that the cable for the AS55 LO
signal had the shielding 90% broken. It was
fixed.
  
  11173   Wed Mar 25 18:48:11 2015 KojiSummaryLSC55MHz demodulators inspection[Koji Den EricG]

We inspected the {REFL, AS, POP}55
demodulators.
  
  11174   Wed Mar 25 21:44:20 2015 KojiUpdateLSCIFO recovery / PRFPMI locking activity[Koji, Den]

- Aligned the arms with
ASS. It had alot of offset accumulated. We
  
  11179   Fri Mar 27 14:47:57 2015 KojiUpdateLSCals->pdh transition, prcl on 1f, alignmentJenne and I interviewd Den this afternoon
to make the things clear

- His "duty cycle" is
 CARM_transition.pdf 
  11180   Fri Mar 27 20:32:17 2015 KojiSummaryLSCLocking activity- Adjutsed the IMC WFS operating point.
The IMC refl is 0.42-0.43.

- The arms are aligned with ASS
  
  11183   Tue Mar 31 03:02:44 2015 KojiUpdateLSCSome locksAssuming the carrier mode in PRC is stable
and the SB is the one moving, can we just
use the POP DC QPD to control PRM?
  
  11245   Fri Apr 24 21:31:20 2015 KojiSummaryCDSautomatic mxstream resettingWe were too much annoyed by frequent stall
of mxstream. We'll update the RCG when
time comes (not too much future).
  
  11248   Sat Apr 25 03:32:45 2015 KojiUpdateASCBroken Xass?I spent a day to fix the XARM ASS, but
no real result. If the input of the 6th DOF
servo is turned off, the other error signals
  
  11312   Tue May 19 17:03:34 2015 KojiUpdateCDSMXstream restart script working (beta)AutoMX is resetting mx_stream every 5 minutes.
Basically everytime AutoMX is called,
it resets mx_stream. Is the mx_stream
  
  11323   Sun May 24 14:45:02 2015 KojiHowToGeneralHow to launch StripTools at specified locationsLLO Operator Tips:

koji.arai@cr9:/opt/rtcds/userapps/trunk/asc/l1/scripts/initial_alignment$
cat autostart_strips.sh 
  
  11332   Thu May 28 17:00:04 2015 KojiUpdateIOOFSS SLOW not engaged: is this intentional?I found that FSS SLOW servo is not engaged.
Is this intentional test to keep the NPRO
temp constant?
  
  11334   Thu May 28 21:10:46 2015 KojiUpdateGreen LockingALS-X noise huntingI have been looking at the X-end ALS setup.
I was playing with the control bandwidth
to see the effect to the phase tracker output
 150528_ALS.pdf 
  11337   Fri May 29 12:49:53 2015 KojiUpdateComputer Scripts / ProgramsChiara Backup HiccupIn fact, the file access is supposed to
be WAY faster now than in the RAID case.

As
  
  11338   Fri May 29 15:12:39 2015 KojiUpdateComputer Scripts / ProgramsChiara Backup HiccupActual data  backup_hours.pdf 
  11353   Thu Jun 11 19:40:59 2015 KojiUpdateVACVacuum comp. rebootedThe serial connections to the vacuum gauges
were recovered by rebooting c1vac1 and c1vac2.

Steve claimed that the vacuum screen
  
  11386   Wed Jul 1 09:33:31 2015 KojiUpdateGeneralShutters closed, watch dogs disabled for the RCG upgradeI closed the PSL/GREEN shutters and shut
off the LSC feedback/SUS watch dogs at 9AM
PDT, to allow Jamie to start his disruptive
  
  11394   Tue Jul 7 23:26:19 2015 KojiUpdateCDSAttempt to list CDS issuesAs Jamie succeded to realize somewhat workable
condition of the 40m CDS, I tried to list
the obvious CDS issues so that we can attack
 burt.log.zip 
  11399   Thu Jul 9 16:39:03 2015 KojiConfigurationGeneralHow to set up your own summary page environment on the LDG clusterHere is the summary of my investigation
how to set up my own "summary page"
environment on the LDG (LIGO Data Grid) cluster.
  
  11461   Wed Jul 29 21:40:39 2015 KojiSummaryCDSStatus of the frame data syncingThe trend data hasn't been
synced with LDAS since Jul 27 5AM local.

40m:
  
  11465   Thu Jul 30 11:47:54 2015 KojiSummaryCDSStatus of the frame data syncingToday it was synced at 5AM but
that was all.

40m:
  
  11466   Thu Jul 30 13:34:52 2015 KojiUpdatePEMY sesimostation is back onPlease check the spectra. If something
is wrong, please swap the cables between
X and Y in order to see if the cable is still
  
  11475   Sat Aug 1 20:46:29 2015 KojiUpdatePEMX seismo station short cable removedOMG   
  11506   Fri Aug 14 12:10:08 2015 KojiUpdatePEMGur interface box is wonkyLet's dismantle the I/F unit from the
rack and connect the cable with the lid open.
We need to trace the signal.
  
  11509   Fri Aug 14 23:49:34 2015 KojiSummaryGeneralB&K Shaker fixedI fixed a shaker that was claimed to be
broken. I had to cut the rubber membrane
to open the head.
 IMG_1947.JPG 