4787
|
Mon Jun 6 16:44:34 2011 |
Koji | Update | Electronics | POY11 tested | The full characterization of POY11 is found
in the PDF.
Resonance at 11.03MHz |
|
4807
|
Fri Jun 10 20:23:56 2011 |
Larisa Thorne | Update | Electronics | Seismometer Box Update/graphs | (continuation of this)
The noise graphs relating total noise |
|
4810
|
Mon Jun 13 16:27:10 2011 |
Jenne | Update | Electronics | Seismometer Box Update/graphs |
|
|
4830
|
Fri Jun 17 00:17:26 2011 |
rana | Configuration | Electronics | 2 RFPDs sent to LLO | Koji and I found 2 RFPD boxes to
send to LLO. We've put them onto Steve's
desk to be overnighted to Valera. |
|
4900
|
Tue Jun 28 15:23:21 2011 |
steve | Update | Electronics | Coilcraft RF-design kits are restocked | Our design kits are full again. They are
waiting for a new brilliant PD design. |
 |
4979
|
Sat Jul 16 18:54:05 2011 |
Ishwita, Manuel | Configuration | Electronics | AA board | We fixed the anti-aliasing board in its
aluminum black box, the box couldn't
be covered entirely because of the outgoing |
  |
5008
|
Wed Jul 20 22:16:27 2011 |
Ishwita, Manuel | Update | Electronics | lying seismometer cable and plugging it | We laid the cable along the cable keeper
from the BACARDI seismometer to the rack
1X6, the excess cable has been coiled under |
|
5022
|
Sun Jul 24 20:36:03 2011 |
haixing | Summary | Electronics | AA filter tolerance analysis | Koji and Haixing,
We did a tolerance analysis to specify
the conner frequency for passive low-pass |
6x |
5023
|
Sun Jul 24 20:47:21 2011 |
rana | Summary | Electronics | AA filter tolerance analysis | This is sort of OK, except the capacitor
connects across the (+) terminals of the
two input opamps, and does not connect to |
|
5024
|
Sun Jul 24 22:19:19 2011 |
haixing | Summary | Electronics | AA filter tolerance analysis |
>> This sort of OK, except
the capacitor connects across the (+) terminals |
|
5034
|
Mon Jul 25 23:43:20 2011 |
Manuel | HowTo | Electronics | Manual for 1201 Low Noise Preamplifier | I found the manual for the Low Noise Preamplifier
Model 1201 at this
link and I attached it. |
 |
5038
|
Tue Jul 26 21:11:40 2011 |
haixing | Summary | Electronics | AA filter tolerance analysis | |
|
5673
|
Sun Oct 16 02:30:00 2011 |
rana | Update | Electronics | Testing REFL165 | Unless the bias feedback circuit has been
tuned for the 1 mm diode, its possible that
you are seeing some C(V) effects. Its easy |
|
5682
|
Mon Oct 17 23:28:32 2011 |
rana | Update | Electronics | StochMon | To get to the bottom of the RFAM mystery,
we've got to resurrect the StochMon to trend
the RFAM after the IMC. |
|
5706
|
Wed Oct 19 18:18:03 2011 |
Suresh | Update | Electronics | StochMon : Filters installed |
|
 |
5734
|
Tue Oct 25 11:48:02 2011 |
Katrin | HowTo | Electronics | solder tiny smd op amps | Yesterday, I had the great pleasure to
solder a tiny 4 x 4 mm op amp with 16 legs
(AD8336). |
|
5748
|
Fri Oct 28 00:53:39 2011 |
rana | Update | Electronics | POP 22/110 Design | The attached PDF shows a possible gain
/ input noise config for the POP 22/110 that
we would use to detect the RF power in the |
|
5764
|
Sun Oct 30 14:08:35 2011 |
rana | Update | Electronics | RFAM monitor in place. ( Uncalibrated ) EPICS troubles |
|
|
5868
|
Fri Nov 11 00:18:53 2011 |
Zach | Update | Electronics | Precision temperature controller | I have made |
|
6265
|
Thu Feb 9 20:01:02 2012 |
rana | Summary | Electronics | Using RF LP filters as dispersion units for the MFD | WE currently use long cables to give
us the dispersion that we want for the MFD.
A cable gives a long delay - both the phase |
|
6290
|
Thu Feb 16 21:13:07 2012 |
Suresh | Update | Electronics | REFL165 repair: PD replaced, DC response checked with a torch light | [Koji, Suresh]
Kiwamu mentioned that REFL165 is
not responding and its DC out seems saturated |
  |
6333
|
Tue Feb 28 16:31:08 2012 |
Suresh | Update | Electronics | REFL165 repair: Characterization | The transfer function and current noise
were measured. The location of the
peak shifts with the amount of incident light |
 |
6337
|
Wed Feb 29 00:22:35 2012 |
Suresh | Update | Electronics | REFL165 repair: Installed on the AS table | 1) The REFL165 has been replaced onto the
AS table.
2) When the PD interface cable is |
|
6339
|
Wed Feb 29 01:14:40 2012 |
Suresh | Update | Electronics | REFL165 repair: Characterization |
|
|
6423
|
Fri Mar 16 06:17:56 2012 |
Suresh | Update | Electronics | REFL165 calibration : measurements |
These are the measurements for estimating
the amplitude of the signal recorded in the |
|
6424
|
Fri Mar 16 10:37:52 2012 |
Jenne | Update | Electronics | Jenne Laser |
|
|
6425
|
Fri Mar 16 16:01:53 2012 |
rana | Update | Electronics | REFL165 calibration : measurements | To characterize the RF V to counts
we need to know the state of the whitening
filter board. Was the filter on or off ? |
|
6428
|
Mon Mar 19 21:25:31 2012 |
Suresh | Update | Electronics | REFL165 calibration : measurements |
|
|
6590
|
Mon Apr 30 22:58:57 2012 |
Jenne | Update | Electronics | 11MHz Marconi set to default after power outage | After a power outage, a Marconi comes back
to it's defaults. It needed to be reset
to the values in elog 5530. |
|
6638
|
Thu May 10 21:13:01 2012 |
Den | Update | Electronics | ADC 3 | ADC 3 INPUT 4 (#3 in the c1pem model if
you count from 0) is bad. It adds DC = ~1
V to the signal as well as noise. I plugged |
|
6857
|
Fri Jun 22 20:00:14 2012 |
Jamie | Omnistructure | Electronics | two RG-405 cables ran from 1X2 rack to control room | [Yaakov, Eric, Jenne, Yuta]
Two of our surfs, Yaakov and Eric,
pulled two unused RG-405/SMA cables that |
|
7372
|
Tue Sep 11 17:17:51 2012 |
Eric Q., Mike J. | Configuration | Electronics | AS beam scan | We conducted a beam scan on the AP table
of the AS beam. We used a lens to focus the
beam onto a power meter, and slowly moved |
|
7377
|
Wed Sep 12 20:08:51 2012 |
ericq | Update | Electronics | AS beam scan |
|
|
7380
|
Thu Sep 13 19:59:43 2012 |
Mike J. | Update | Electronics | AS beam scan | **EDIT:** Mixed up X and Y. Beam is
3.5844 |
|
7386
|
Fri Sep 14 01:35:55 2012 |
Mike J. | Update | Electronics | AS beam scan PLOTS | |
 |
7404
|
Tue Sep 18 22:06:21 2012 |
Mike J. | Update | Electronics | AS beam scan plots and chi-squared | Results of the Razor Blade Beam Scan
The horizontal blade test measured
the beam intensity as a razor blade passed |
  |
7691
|
Thu Nov 8 22:04:43 2012 |
Charles | Update | Electronics | Ethernet Illuminator Control | Configured ethernet controlled power strips
to have static IP addresses: 192.168.113.110,
192.168.113.111 and 192.168.113.112. |
|
7698
|
Mon Nov 12 23:38:50 2012 |
Charles | Update | Electronics | Ethernet Illuminator Control |
|
|
7701
|
Tue Nov 13 00:27:50 2012 |
Jenne | Update | Electronics | Ethernet Illuminator Control |
|
|
7770
|
Fri Nov 30 23:10:36 2012 |
Charles | Update | Electronics | Vertex Illuminators | 3 of the 4 remote controlled illuminators
at the vertex are installed and can now be
turned on via sitemap. There are a total |
|
7873
|
Thu Jan 3 19:19:59 2013 |
rana | HowTo | Electronics | cable racks | Today I found 3 power cables in the
orange Pomona cable tray, put in
so that the cables were |
|
7875
|
Fri Jan 4 13:23:10 2013 |
rana | Update | Electronics | Photodiode transimpedance | You have to correct this transimpedance
ratio by correcting for the different levels
of DC photocurrent in the two devices. |
|
7879
|
Mon Jan 7 19:23:19 2013 |
rana | Update | Electronics | Jamie's 1811 PS from 1998 | |
 |
7881
|
Tue Jan 8 14:07:04 2013 |
Riju | Update | Electronics | Photodiode transimpedance |
|
|
7882
|
Tue Jan 8 15:28:41 2013 |
Riju | Update | Electronics | Photodiode transimpedance |
|
|
7921
|
Sat Jan 19 16:02:28 2013 |
rana | Omnistructure | Electronics | PS cleanup | During our 'Women in Physics' tours today,
we were reminded that there are several bench
power supplies being used as permanent inside. |
|
8008
|
Wed Feb 6 14:51:25 2013 |
Jenne | Update | Electronics | 1 power supply replaced! |
|
|
8011
|
Wed Feb 6 15:11:21 2013 |
Jenne | Update | Electronics | "Temporary" power supply situation | [Jenne, Yuta, Rana, Steve, Manasa]
We have taken stock of the lab
"temporary" power supply situation, |
|
8013
|
Wed Feb 6 15:39:19 2013 |
Steve | Update | Electronics | DC power supplies in cabinets | East arm cabinet E9 and E10
|
|
8016
|
Wed Feb 6 20:00:06 2013 |
Manasa | Update | Electronics | BNC cables piled up at every corner | [Yuta, Steve, Manasa]
There are cables piled up around
the access connector area which have been |
|