40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 102 of 341  Not logged in ELOG logo
IDup Date Author Type Category Subject
  5087   Mon Aug 1 23:29:24 2011 Manuel, IshwitaUpdateWienerFilteringGetting Data by matlab

We tried to acquire data from the seismometers and the mode cleaner using the Matlab function

datalist = NDS2_GetData({'C1:PEM-SEIS_GUR1_X_IN1_DQ'}, 996258376 , 10, CONFIG.nds.C)

and encountered the following error

Warning: daq_request_data failed
 
??? Error using ==> NDS2_GetData
Fatal Error getting channel data.

The same error was obtained with the following other channels

C1:PEM-SEIS_GUR2_X_IN1_DQ

C1:PEM-SEIS_STS_1_X_IN1_DQ

But we are able to get data from channel

C1:LSC-MC_OUT_DQ

for the same gps time.

We checked with Dataviewer that the data are saved (we viewed data of last 24h) for every channel.

  5088   Tue Aug 2 02:20:09 2011 kiwamuUpdateGreen LockingY arm green beam axis : done

I succeeded in locking the green light to the Y arm cavity, but it wasn't so robust. Something is unhealthy in the electronics.

I am leaving the Y green system as it is because I already can see a plenty of the green light flashing in the BS chamber.

So just a flashing of the green light is good enough for the in-vac-work.

DONE.

Quote from #5084

The next step is to refine some electronics in the PDH loops to get the green light locked to the Y arm cavity. 

 

  5089   Tue Aug 2 02:35:23 2011 kiwamuUpdateGeneralpreparation of the vent : status and plan

The vent will take place on Wednesday.

Plan for Tuesday :

  (Morning) Preparation of necessary items for the low power MC (Steve / Jamie)

  (Daytime) Measurement of the MC spot positions (Suresh)

  (Daytime) Arm length measurement (Jenne)

  (Nighttime) Locking of the low power MC (Kiwamu / Volunteers)

 

Plan for Wednesday :

  (Early morning) Final checks on the beam axis, all alignments and green light (Steve / Kiwamu / Volunteers )

  (Morning) Start the vent (Steve)

  (daytime-nighttime) Taking care of the Air/Nitrogen cylinders (Everybody !!)

 

Status of the vent preparation :

 

  (not yet) Low power MC

  (ongoing) Measurement of the arm lengths

  (ongoing) Measurement of the MC spot positions

  (80% done) Estimation of the tolerance of the arm length (#5076)

  (done) Alignment of the Y green beam (#5084)

  (done) Preparation of beam dumps (#5047)

  (done) Health check of shadow sensors and the OSEM damping gain adjustment (#5061)

  (done) Alignment of the incident beam axis (#5073)

  (done) Loss measurement of the arm cavities (#5077)

  5090   Tue Aug 2 10:53:03 2011 JenneOmnistructureSAFETYRegular door out of service. Use Control Room Door only!!

The hazardous waste people are moving chemicals around outside our door, and have roped off our regular front door. 

Please go around, and use the control room door to enter and exit.  It is currently unlocked, although I'll lock up when I leave for LIGOX.

  5091   Tue Aug 2 11:02:52 2011 JenneUpdateSUSPRM Watchdog tripped

I found PRM watchdog tripped.  It's all better now.

  5092   Tue Aug 2 11:52:44 2011 kiwamuSummaryGeneralSchedule

I have updated the 40m public calender.

Main change :

  + The vent starts from 3rd of August

  + Keiko and Anamaria (LSU) come from 13th of August

  5093   Tue Aug 2 15:41:06 2011 kiwamuUpdateIOOMeasurement of MC spot positions : done

[Suresh / Kiwamu]

 The measurement of the spot positions on the MC mirrors are DONE.

Surprisingly the spot positions are not so different from the ones measured on May.

    Feb 26 2011      May 08 2011 (New !) Aug 2 2011
MC1 pit [mm]   1.6   1.9  1.93
MC2 pit [mm]   6.4   9.0 9.03
MC3 pit [mm]   1.4   2.0 2.01
MC1 yaw [mm]   -1.5   -1.7 -1.72
MC2 yaw [mm]   1.0   0.2 0.178
MC3 yaw [mm]   -1.3   -1.9 -1.87

 

(some notes)
We used Valera's script senseMCdecenter to estimate the spot positions ( see his entry).

It returns so many EPICS error messages and sometime some measured values were missing. So we had to throw away some of the measurements.

Anyways we gave the resultant ASCII file to Valera's matlab file sensmcass.m to get the actual amount of off-centering in milli-meter.

The attached file is the resultant plot from his matlab code.

Attachment 1: MCdecenter.png
MCdecenter.png
  5094   Tue Aug 2 16:43:23 2011 jamieUpdateCDSNDS2 server on mafalda restarted for access to new channels

In order to get access to new DQ channels from the NDS2 server, the NDS2 server needs to be told about the new channels and restarted.  The procedure is as follows:

ssh mafalda
cd /users/jzweizig/nds2-mafalda
./build_channel_history
./install_channel_list
pkill nds2
# wait a few seconds for the process to quit and release the server port
./start_nds2

This procedure needs to be run every time new _DQ channels are added.

We need to set this up as a proper service, so the restart procedure is more elegant.

An additional comment from John Z.:

    The --end-gps parameter in ./build_channel_history seems to be causeing
    some trouble. It should work without this parameter, but there is a
    directory with a gps time of 1297900000 (evidently a test for GPS1G)
    that might screw up the channel list generation. So, it appears that
    the end time requires a time for which data already exists. this
    wouldn't seem to be a big deal, but it means that it has to be modified
    by hand before running. I haven't fixed this yet, but I think that I
    can probably pick out the most recent frame and use that as an end-time
    point. I'll see if I can make that work...

  5095   Tue Aug 2 16:55:21 2011 ranaUpdateABSLArm length measurement : cavity kick technique

Quote:

I made some attempts to measure the current length of the arm cavities by using the mass-kicking technique.

 Why not just scan the Green laser to measure the arm lengths instead? The FSR of the arm is ~3.75 MHz and so all you have to do is lock the arm green and then sweep the PZT until the resonance is found at 3.75 MHz.

L.png

  5096   Tue Aug 2 17:40:04 2011 JennyUpdatePSLReducing beam intensity incident on photodiode

I am using a PDA255 photodiode to measure the power outputted by the NPRO beam on the PSL table. (I'm going to then use a network analyzer to measure the amplitude response of the PZT to being driven at a range of frequencies. I'll detect the variation in in response to changing the driving frequency using this PDA255.)

The PDA255 has an active area of 0.8mm^2 and a maximum intensity for which the response is linear of 10mW/cm^2. This means that a beam I focus on the PD must have a power less than 0.08 mW (and even less if the spot size is smaller than the window size).

I used a power meter to measure the beam power and found it was 0.381 mW.

The second polarizing beam splitter in the setup transmits most of the beam power, but reflects 0.04 mW (according to the power meter). I'm going to place the photodiode there in the path of the reflected beam.

  5097   Tue Aug 2 19:05:13 2011 kiwamuUpdateIOOpeparation for low power MC : HWP+PBS installed

[Steve / Kiwamu]

  An attenuator, consisting of two HWPs and a PBS, has been installed on the PSL table for the MC low power state.

Those items allow us to reduce the amount of the incident power going into the MC.

We haven't decreased the power yet because we still have to measure the arm lengths.

After we finish the measurement we will go to the low power state.

We have adjusted the polarization after the last HWP using another PBS. Now it is S-polarizing beam.

After the installation of the attenuator the beam axis has changed although we were immediately able to lock the MC with TEM00 mode.

I touched two steering mirrors on the PSL table to get the transmitted power of MC higher. At the moment the transmitted power in MC_TRANS is at about 30000 cnts.

 

The attached picture is the setup of the attenuator on the PSL table.

Attachment 1: attenuator.png
attenuator.png
  5098   Tue Aug 2 23:02:48 2011 NicoleUpdateSUSFixed Accelerometer


 

The EM shaker was broken (the input terminals (banana inputs) had snapped off. To fix this, I have mounted two banana input mounting posts to a metal mount that Steve attached to the shaker.

shakerposts.jpgclipstoamplifier.jpg

However, because bananas do not provide a secure connection (they easily fall out), I have made special wires to connect the banana inputs of the shaker to the mounted banana inputs of the amplifier I am using (along with the sine generating function of the HP 3563A signal analyzer). Upon Koji's suggestion, I have made C-shaped clips to attach to the banana post mounts. These clips are made from insulated ring terminals.

 newclips.jpg

Today I tested the  shaker and it works! WOOT! I currently have the shaker attached to the horizontal sliding platform (without the TT suspension on it).

Using a 750mHz signal from the HP 3563A with an amplitude of 500 mV amplified to 0.75V, I have gotten the shaker to displace the platform (without the TT suspension on it) 1 mm.

  5099   Tue Aug 2 23:14:21 2011 NicoleUpdateSUSCorrected Screw Alignment in TT Suspension Base

The TT suspension base was not able to be securely mounted to the optical table (i.e. mounted with 4 screws)

because the spacing between the screw holes in the base did not have the correct spacing for mounting on a table with a 1 inch pitch.

newholes.jpg

We have carefully removed the suspension from the problematic base. PLEASE BE VERY CAREFUL AROUND THE TABLE NEXT TO THE MC-2 CHAMBER! THE TT SUSPENSION IS RESTING THERE WITHOUT THE BASE! I will reattach it to the base tomorrow morning when I am less tired and more careful!

We measured the base to be about 4.882" x 3.774". The screw hole spacing is about 3.775" and 2.710" respectively. I have changed the diameter of the screw holes from 0.26" to 0.315" and have been able to successfully mount the suspension to the 1 inch pitch table next to the MC-2 chamber.

 

mountedmirrorflash.jpg

Now that the TT suspension can be mounted, I am going to be aligning a 670nm LED laser and balancing the mirror on the TT tomorrow morning. I will be using a beam blocker but please still be careful.

  5100   Wed Aug 3 01:30:04 2011 JenneUpdateLSCAbsolute length of Xarm and Yarm measured

So far, this is just preliminary, because I haven't done full error analysis to determine the error on my measurements.  That will hopefully be done by tomorrow afternoon (so before we start taking off doors).

I find that the length of the Xarm is:  37.5918 meters.

I find that the length of the Yarm is:   37.5425 meters.

I used the mass-kicking technique, as summarized by Kiwamu, and fully described by Alberto.  More words / description to follow with the full error analysis.

  5101   Wed Aug 3 02:20:33 2011 KojiUpdateLSCREFL165 PD

REFL165 PD has been made from the old 166MHz PD.
As the required inductance was ~10nF level, the stray inductance of the circuit pattern was significant.
So, I am not so confident with the circuit functionality before the optical transfer function test.

I will test REFL33 and REFL165 with the Jenne laser to see how they work.

P8031393.JPG

P8031391.JPG

Attachment 1: REFL165_schematic_110802_KA.pdf
REFL165_schematic_110802_KA.pdf
  5102   Wed Aug 3 02:28:08 2011 Manuel, IshwitaUpdateWienerFilteringWiener Filtering in X-arm

Wiener Filtering was applied on the data collected from the X-arm during the time: GPS time-996380715 (Aug 02, 2011. 21:25:00. PDT) to GPS time-996382215 (Aug 02, 2011. 21:50:00. PDT) for a duration of 1500 seconds. During this time the X-arm was locked, we checked it by acquiring data from channel C1:LSC-TRX_OUT_DQ .

The seismometers were near the beam splitter (guralp2) and near MC2 (guralp1).

Target data was obtained from channel C1:LSC-XARM_IN1_DQ.

Schermata-6.png

Schermata-7.png

Following graphs were obtained after applying the Wiener filter:

 

      1.Seismic data acquired from Guralp1 (X and Y) and Guralp2 (X and Y)                              2.Seismic data acquired from Guralp2 X                                                              3.Seismic data acquired from Guralp2 Y 

WFgur1X1Y2X2YN20000srate2048.pngWFgur2XN20000srate2048.pngWFgur2YN20000srate2048.png

These graphs were obtained with srate = 2048 (sample rate) and N = 20000 (order of the filter).

Graph 1 is the best because the black (residual) line is below the red (target) line for low frequencies since we used seismic data from 4 channels. Graph 3 is the worst because we used seismic data from only one Y channel (Y axis of Guralp2) that is less related with the X-arm mirrors' motion since they are oriented orthogonally.

  5103   Wed Aug 3 04:29:48 2011 kiwamuUpdateIOOpeparation for low power MC : HWP+PBS installed

[Suresh / Kiwamu]

 The incident beam power going into MC was decreased down to 20 mW by rotating the HWP that we set yesterday.

A 10% beam splitter which was sitting before MCREFL_PD was replaced by a perfect reflector so that all the power goes into the PD.

And we confirmed that MC can be still locked by increasing C1IOO-MC_REFL_GAIN. Some modifications in the Autolocker script need to be done later.

Also we opened the aperture of the MC2F camera to clearly see the low power beam spot.

 

WE ARE READY FOR THE VENT !!

Power after the EOM = 1.27 W

Power after the HWPs and PBS = 20.2 mW

Power on MCREFL = 20 mW (MC unlocked)

MCREFL_DC = 0.66 V (with MC locked)

 

Quote:

After we finish the measurement we will go to the low power state.

 

  5104   Wed Aug 3 09:40:13 2011 steveUpdateVACvent

The vacuum system is coming up to atm. The vent was started with slow N2 flow. The PSL output is blocked 2 places on the table.

Attachment 1: pzt2011Aug3.png
pzt2011Aug3.png
  5105   Wed Aug 3 11:34:59 2011 kiwamuUpdateVACRe: vent

An important rule during the in-vac work :

   Do not change the incident axis of the X and Y green beams.

Because of the air flows and unusual pressure in the chambers, the DC alignment of the suspensions become less reliable in terms of the beam pointing reference.

The green beams will be considered as references during the vent.

Quote:

The vacuum system is coming up to atm. The vent was started with slow N2 flow.

  5106   Wed Aug 3 12:24:08 2011 Manuel, IshwitaUpdateWienerFilteringWeekly summary

Last Friday (Jul 29) we reinstalled the blue breakout box, and changed the names of the C1:PEM channels. Elog Reference

We continued the work on the simulation ad applied wiener filter on the simulated ground motion, but the result is unsatisfactory, yet. We will post reasonable results soon.

We did wiener filtering for the first time on real data from the Xarm while it was locked. Elog Reference

  5107   Wed Aug 3 12:27:01 2011 NicoleSummarySUSWeekly Summary

This week I have determined the linear region for my photosensor. I have determined the linear region to be -14.32 V/cm in the region 0.4cm 0.75 cm.

In order to obtain this voltage plot, I used a 287K resistor to set the max voltage output for the photodiodes. This calibration was obtained using a small rectangular standing mirror (not the TT testing mirror that Steve has ordered for me).

calibrationplot.jpg

I have also been working on the second half of the photosensor circuit (to power the LED and read out voltages for the second photosensor head). I have assembled the constant-current section of the circuit and need to do the voltage-output section of the circuit. I also need to finish assembling the second photosensor head and cables.

 

I submitted my Second Progress Report on Tuesday.

 

I have attached the mirror to the TT suspension. We are using 0.006 diameter tungsten wire to suspend the mirror. I am currently working on balancing the mirror.

 

This morning, I realized that the current set-up of the horizontal shaker does not allow for the TT to be securely mounted. I was going to change the drill holes in the horizontal slider base (1 inch pitch). Jamie has suggested that it is better to make a pair of holes in the base larger. The circled holes are the ones that will be expanded to a 0.26" diameter so that I can mount the mirror securely to the horizontal slider base. There is a concern that a bit of the TT suspension base will hang over the edge of the horizontal sliding plate. We are not sure if this will cause problems with shaking the mirror evenly. Suggestions/advice are appreciated.

newholestobe.JPG

Attachment 1: calibrationplot.jpg
calibrationplot.jpg
Attachment 2: calibrationplot.jpg
calibrationplot.jpg
  5108   Wed Aug 3 12:37:57 2011 KojiSummarySUSWeekly Summary

I vote for making an adapter plate between the sliding plate and the bottom base.

Quote:

This morning, I realized that the current set-up of the horizontal shaker does not allow for the TT to be securely mounted. I was going to change the drill holes in the horizontal slider base (1 inch pitch). Jamie has suggested that it is better to make a pair of holes in the base larger. The circled holes are the ones that will be expanded to a 0.26" diameter so that I can mount the mirror securely to the horizontal slider base. There is a concern that a bit of the TT suspension base will hang over the edge of the horizontal sliding plate. We are not sure if this will cause problems with shaking the mirror evenly. Suggestions/advice are appreciated.

 

  5109   Wed Aug 3 14:20:49 2011 steveUpdateSUSITMX and PRM damping restored

ITMX watchdog tripped around 5 Torr  and the PRM around 450 Torr of this vent. They were restored than.

We are at 580 Torr now.

  5110   Wed Aug 3 14:40:32 2011 kiwamuSummaryGeneralminutes of 40m meeting : plan tomorrow

Tomorrow's main goal is : let the both X and Y green light come out from the chambers.

 

Plan of the in-vac work for tomorrow :

    - Removal of the access connector and the BS north door, starting from 9:00 AM. (requires 6 people)

   - If necessary, align ITMs and ETMs to get the green light nicely flashing / locked.

   - Take pictures of the BS and IOO table before installing / repositioning some optics.

   - Repositioning of the green periscope in the BS chamber to let the Y green light go through it.

   - Steer some green mirrors on the IOO table to let the Y green light come out from the chamber.

   - Steer some green mirrors on the BS table to let the X green light come out from the chamber.

   - Put some beam traps on the BS table

   - Leveling of the BS table. (Do we need to level the IOO table ? it will change the spot positions on the MC mirrors somewhat)

   - Take pictures again.

   - Extra jobs : if we still have some more times, lock MC and check the beam clearance at the Faraday. Also check some possible beam clippings for the IR beam.

   - Close the chamber with the light doors.

   - Softball game at 6:30 PM.

  5111   Wed Aug 3 16:18:30 2011 steveUpdateVACvent is almost finished

P1 at 750 Torr,  Vent valve VV1 is left open to atm. It will reach equilibrium in an hour. The PSL power is not turned down yet. The beam path is blocked at two places inside the enclosure.

The low power 20 mW  input  MC can be locked tonight.

 

Next day's - Thursday notes:

We  reached 760 Torr of atm in 8 hrs at  1.7 Torr/ min with 5 cylinders of  instrument grade air.

Used 500 PSI of nitrogen to get to 20 Torr at first.  This flow  and the first cylinder of N2  has to be slower next time.

 

We did the following:

1, oplevs were centered on locked caveties 2 days ago

2, just before vent we rechecked arms  pointing, they wre still locking

3,eloged strain gages  and turned HVs off at PZT-steering   at 1Y3 rack and OMC-PZT ps at 1X2 aux rack. They were actually off.

4,,checked jam nuts 6 places x 3

5, turned oplev servos off

6, eloged SUS summery page

7, checked particle counts at ITMY: 1 micron 300,  0.5 micron 3K,  0.3 micron  30K particles / cft-min. All three cranes were wiped 2 days ago.

8, blocked laser into IFO: main 1064 in 1.25W were blocked by manual 2 blocks on PSL table. The PSL output shutter was left open to be triggered by interlock  at P1 3 mTorr. It did!

    Aux NPRO-AP and green light  input at both ends were blocked.

Attachment 1: vent8h.jpg
vent8h.jpg
  5112   Wed Aug 3 22:22:47 2011 KojiUpdateLSCREFL165 PD

This REFL165 was good in terms of RF, but I forgot to make the DC path functioning.

I will try some ideas to fix this tomorrow.

  5113   Wed Aug 3 22:31:38 2011 KojiUpdateLSCREFL33 PD

REFL33 is ready for the installation

Characterization results of REFL33 is found in the PDF attachment.

Resonance at 33.18MHz
Q of 6.0, transimpedance 2.14kOhm
shotnoise intercept current = 0.52mA (i.e. current noise of 13pA/rtHz)

Notch at 10.97MHz
Q of 22.34, transimpedance 16.2 Ohm

Notch at 55.60MHz
Q of 42.45, transimpedance 33.5 Ohm

 

Attachment 1: REFL33_test_110801_KA.pdf
REFL33_test_110801_KA.pdf REFL33_test_110801_KA.pdf REFL33_test_110801_KA.pdf REFL33_test_110801_KA.pdf
  5114   Thu Aug 4 00:04:52 2011 JennyUpdatePSLNetwork analyzer and PD set up to measure amplitude response of PZT

Today I placed the PDA255 photodiode on the PSL table to catch the small amount of beam power reflected by the second polarizing beam splitter in my setup. I plugged the PD output to the oscilloscope to measure the voltage output and positioned the PD such that the voltage output was maximized. At best I was able to achieve a 300 mV DC output voltage from the PD, (which seems a bit low, as the PD is specified to go from 0 to 5 V and the specifications say that the response becomes nonlinear after 10 mW/cm^2 and my beam has an intensity of approximately 5 mw/cm^2. I would therefore expect to get more beam power but after over an hour of maneuvering, 300 mV was the highest voltage output I could get).

I am planning, tomorrow afternoon, to take a measurement of the amplitude response of the PZT driving the NPRO laser. I moved the 4395 spectrum/network analyzer to near the PSL table and connected the RF output to an RF splitter. I fed one output of that into the PZT and the other output into the R port on the network analyzer. I fed the PD output into the A port. I plan to measure A/R as a function of driving frequency, sweeping from 10 Hz to 30 mHz.

I also worked to improve the mode matching of the NPRO beam coming from the AP table to the reference cavity. I drove the temperature of the NPRO at 0.100 Hz with an amplitude of 0.300 V, which Koji told me corresponds to a 1GHz change in the laser frequency. The transmission from the cavity is being monitored by a camera connected to a TV monitor, and also by a PD connected to an oscilloscope. I then repositioned the second lens in my mode matching setup in an attempt to increase the transmission peaks from the zeroth order spacial mode and decrease the transmission peaks from higher order modes. I may have improved the mode matching slightly but I was unable to improve it significantly.

  5115   Thu Aug 4 01:49:08 2011 SureshUpdateIOOMC is locked

I measured the power transmitted from the PSL to the MC. It is 19mW.

The MC is now locked.  The MC Autolocker script cannot be used now since the tigger conditions are not met.  It has been disabled on the C1IOO-LOCK_MC screen.  The boost switch also is set to zero.  Increasing the boost results in MC unlocking. 

The C1:IOO-MC_RFPC_DCMON was going from 1.4 (MC Unlocked) to 0.66 (MC_locked).  I thought we ought to have a factor of ten drop in this since under high power conditions we used to have a drop of about 5.6 to 0.6.   So I adjusted the zig-zag at the end of the PSL table to improve the alignment.  It now goes from 1.4 to 0.13 when the MC is locked.   The lock is also much more stable now.  It still does not tolerate any boost though.

I checked to make sure that the beam centering on MC_REFL PD is optimal since I touched the zig-zag.  The RFPD output is now 0.7V (MC unlocked).  This matches well with the fact that we used to have 3.5V on it with the MC unlocked.  And we have cut the down the power incident on this by a factor of 5.  Because 1W -> 20mW at the PSL table  and 10% BS -> 100% Y1-...

 

 

  5116   Thu Aug 4 08:25:51 2011 steveUpdateSUSsus at atm

ITMX and PRM moved alot.  BS and   ITMY just a little based on oplev reference.

Attachment 1: SUSatm.png
SUSatm.png
Attachment 2: oplevsaftervent.png
oplevsaftervent.png
  5117   Thu Aug 4 09:42:19 2011 KojiUpdateABSLABSL Laser shutter closed

The shutter of the ABSL laser is closed for the vent work.

  5118   Thu Aug 4 11:18:12 2011 ranaUpdateSUSsus at atm

Remember that the Oplevs are not good references because of the temperature sensitivity. The week long trend shows lots of 24 hour fluctuations.

  5119   Thu Aug 4 20:05:23 2011 NicoleUpdateSUSNew Horizontal Sliding Base Mount for TT suspension testing

In order to more-securely mount the TT supsion to the horizontal sliding base, I have made a sub-mounting plate (upon Koji's suggestion) to go in between the horizontal sliding base and the TT suspension base. I made many mistakes in this once-pristine aluminum board. I learned that using a ruler is not good enough for determining where to make holes. Upon Koji's suggestion, I have completed the mounting plate by first making a full-scale diagram on Solid Works, printing it out, and then using the diagram to determine where to make my punch holes. Thank you also to Manuel for helping me drill and to Suresh for teaching me how to use the taps!

topplateview.jpgplatform.jpg

I have been able to successfully mount the plate to the horizontal sliding platform. The TT suspension base is mounted to the front of the mounting plate (there are counter-sink screws at the front connecting the platform to the slider so that the screw heads do not obstruct the TT base). I have been able to successfully mount the TT suspension base to the mounting plate. I have also reattached the TT suspension frame to its original base (the one that I modified so that the TT could be mounted to a 1 inch pitch surface). Currently, the TT suspension is mounted to the optical table I have been working on (next to the MC-2 chamber). I am working on balancing the mirror. I am going to balance the mirror using a 670nm LED laser.

Below is a picture of the laser and the laser block I am using. After I took this photo, I have mounted the laser and the block to the optical table next to the MC-2 chamber.

 

lasersetup.jpg

I have already leveled the laser and I will plan to work on balancing the mirror tomorrow morning (my hands were shaking a lot this afternoon/evening, so I think it would be best to wait until the morning when I will be more careful). I am now going to work on the second half of my photosensor circuit box and second sensor head.

 

Please do not touch the 670nm laser on the optical table next to the MC-2 chamber! It has been leveled. Please also be careful around the optical table, since the TT suspension is mounted to the table!

  5120   Fri Aug 5 02:00:09 2011 kiwamuSummaryGeneralsummary of today's invac work

[Jamie, Jenne, Suresh, Steve, Koji, Kiwamu]

We got two green beams coming out from the chambers !

 

Summary of today's invac work :

    - removed the access connector and the BS north door

    - realigned the X and Y arm to the green beams.

    - installed a HWP on the ETMY table to rotate the polarization of the green beam to P.

    - repositioned the first periscope on the BS table.

    - repositioned the second periscope on the IOO table.

    - steered some green mirrors on the IOO and OMC chamber to let the Y green beam come out to the PSL table.

    - installed a PBS in front of the first periscope to spatially overwrap two green beams.

    - adjusted the incident angle of the PBS to maximize the power of the Y green beam, which is transmitted through it.

   - steer two mirrors on the BS table to align the X green beam

   - installed two beam dumps, one is near the PBS to eliminate a ghost in the X green beam, and the other is on the back side of IPPOS/ANG pick off window.

   - closed the doors.

 

Some Notes:

 When steering the final green mirror on the OMC table, accidentally we changed the alignment of the MC incident mirror.

So the alignment of the incident beam going into MC has changed, and we haven't re-aligned it yet.

 During we were installing the PBS on the BS table, we found that the allowable incident angle for the Y beam is about ~ 55 deg, which maximizes the amount of the transmitted Y green.

Since the PBS had been considered to be 45 deg incident in our optical layout, this required several modifications in the green mirrors.

To have a clear X green beam path going into the PBS, we had to slide the PBS and periscope to the West.

The periscope is now sitting on the very edge of the BS table, and in fact ~ 20% of the bottom plate of the periscope is already sticking out.

Also since 30% of the area of the PBS's post is on a hole, which is somehow for the stack, we had to use three dog clamps instead of a folk clamps to make the contact tight.

  5121   Fri Aug 5 04:03:16 2011 KojiUpdateLSCREFL165 PD

REFL165 PD was made and tested. The characterization results are in the PDF file.

Resonance at 166.12MHz
Q of 7.3, transimpedance 667Ohm (Series Resistance = Z/Q2 = 2.5Ohm)
shotnoise intercept current = 4.3mA (i.e. current noise of 36pA/rtHz) 


As the circuit pattern had ~10nH level strain inductance, some technique was needed.

  • The diode was pushed in so as to reduce the lengths of the legs as short as possible.
  • The inductor for the resonant circuit has been located as close to the photodiode as possible
  • The other side of the inductor was needed to be bypassed by a large (0.1uF) capacitor, as the original circuit pattern (D1-L5-C33//R22) was too skinny and long.
  • C32 is also moved next to the diode.
  • The path of the photo current circuit was made thicker by Cu tapes.

Now the size of the loop for the resonant circuit is comparable with the size of SOIC-8 opamp.
(Left-Top corner of the photo)

This improved the resonant gain by factor of 8.5dB at the test with TEST INPUT. (Analyzer photo)

There is no tunable component.
The resonant freq was adjusted by a parallel inductance (270nH) to the main inductor (15nH).

P8041394.JPG

P8041395.JPG

Attachment 3: REFL165_test_110804_KA.pdf
REFL165_test_110804_KA.pdf REFL165_test_110804_KA.pdf REFL165_test_110804_KA.pdf REFL165_test_110804_KA.pdf
  5122   Fri Aug 5 08:08:42 2011 kiwamuSummaryGeneralplan today

Today's main mission is : adjustment of the arm length

 

   + Open the ETMX(Y) door, starting from 9:00 AM

   + Secure the ETMX(Y) test mass by tightening the earthquake stops.

   + Move the ETMX(Y) suspension closer to the door side

   + Inspect the OSEMs and take pictures before and after touching the OSEMs.

   + Level the table

   + Adjust the OSEM positions

   + Move the ETMX(Y) suspension to have designed X(Y)arm length

   + Level the table again

   + Align the ETMX(Y) such that the green beam resonate

  5123   Fri Aug 5 13:51:51 2011 steveUpdateSUScross coupling

We need a plan how to minimize cross coupling in the OSEMs now

  5124   Fri Aug 5 14:03:12 2011 steveUpdateGeneralmalfunctioning crane at ETMY

The horizontal trolley drive stopped working  at the east end this morning. It is working intermittently. In the worst case we can take the door off with the manual -Genie- lift.

I'm working with Konecrane to solve  the wormgear drive problem.

  5125   Fri Aug 5 15:11:24 2011 kiwamuUpdateSUSRe: cross coupling

There is a page on the 40m wiki explaining the procedure.

  http://blue.ligo-wa.caltech.edu:8000/40m/OSEM_adjustment_proceedure

 

Additionally there are several old elogs about the cross-coupling minimization, which can be useful for us:

[1] "SUS ranking from measured data", iLog by Osamy (Aug.22 2005)

[2] "TF from position to sensors for ETMX ", iLog by Osamu (Aug.25 2005)

[3] "Further OSEM tweaking", iLog by Rana (Oct.3 2006)

Quote from #5123

We need a plan how to minimize cross coupling in the OSEMs now

 

  5126   Fri Aug 5 18:29:35 2011 JennyUpdatePSLNetwork analyzer and PD set up to measure amplitude response of PZT

Quote:

Today I placed the PDA255 photodiode on the PSL table to catch the small amount of beam power reflected by the second polarizing beam splitter in my setup. I plugged the PD output to the oscilloscope to measure the voltage output and positioned the PD such that the voltage output was maximized. At best I was able to achieve a 300 mV DC output voltage from the PD, (which seems a bit low, as the PD is specified to go from 0 to 5 V and the specifications say that the response becomes nonlinear after 10 mW/cm^2 and my beam has an intensity of approximately 5 mw/cm^2. I would therefore expect to get more beam power but after over an hour of maneuvering, 300 mV was the highest voltage output I could get).

I am planning, tomorrow afternoon, to take a measurement of the amplitude response of the PZT driving the NPRO laser. I moved the 4395 spectrum/network analyzer to near the PSL table and connected the RF output to an RF splitter. I fed one output of that into the PZT and the other output into the R port on the network analyzer. I fed the PD output into the A port. I plan to measure A/R as a function of driving frequency, sweeping from 10 Hz to 30 mHz.

I also worked to improve the mode matching of the NPRO beam coming from the AP table to the reference cavity. I drove the temperature of the NPRO at 0.100 Hz with an amplitude of 0.300 V, which Koji told me corresponds to a 1GHz change in the laser frequency. The transmission from the cavity is being monitored by a camera connected to a TV monitor, and also by a PD connected to an oscilloscope. I then repositioned the second lens in my mode matching setup in an attempt to increase the transmission peaks from the zeroth order spacial mode and decrease the transmission peaks from higher order modes. I may have improved the mode matching slightly but I was unable to improve it significantly.

The ABSL beam had been blocked so that it wouldn't enter the interferometer. I moved the block so that the beam I've been using is unblocked by the beam going to the interferometer is still blocked.

I positioned a fast lens (f=28.7mm) a little over an inch in front of the PDA255 in order to decrease the spot size incident on the PD. I adjusted the rotation angle of the half wave plate to maximize the transmitted power through the PBS to the cavity and minimize the power reflected to my PD. I then adjusted the lens potion to fix the beam on the PD. The voltage output of the PD is now 150mW, but I have the ability to increase the incident power by rotating the wave plate slightly.

Now all I need is to set up the network analyzer again to record the amplitude response to modulating the PZT from 10 Hz to 30 MHz, reduce the input voltage into the analyzer using a DC block.

  5127   Fri Aug 5 20:37:34 2011 jamieSummaryGeneralSummary of today's in-vacuum work

[Jamie, Suresh, Jenne, Koji, Kiwamu]

After this morning's hiccup with the east end crane, we decided to go ahead with work on ETMX.

Took pictures of the OSEM assemblies, we laid down rails to mark expected new position of the suspension base.

Removed two steering mirrors and a windmill that were on the table but where not being used at all.

Clamped the test mass and moved the suspension to the edge of the table so that we could more easily work on repositioning the OSEMs.  Then leveled the table and released the TM.

Rotated each OSEM so that the parallel LED/PD holder plates were oriented in the vertical direction.  We did this in the hopes that this orientation would minimize SD -> POS coupling.

For each OSEM, we moved it through it's full range, as read out by the C1:SUS-ETMX_{UL,UR,LL,LR,SD}PDMon channels, and attempted to adjust the positions so that the read out was in the center of the range (the measured ranges, mid values, and ultimate positions will be noted in a follow-up post).  Once we were satisfied that all the OSEMs were in good positions, we photographed them all (pictures also to follow).

Re-clamped the TM and moved it into it's final position, using the rails as reference and a ruler to measure as precisely as possible :

ETMX position change: -0.2056 m = -20.56 cm = -8.09 in (away from vertex)

Rebalanced the table.

Repositioned the mirror for the ETMX face camera.

Released TM clamps.

Rechecked OSEM centering.

Unblocked the green beam, only to find that it was displaced horizontally on the test mass about half an inch to the west (-y).  Koji determined that this was because the green beam is incident on the TM at an angle due to the TM wedge.  This presents a problem, since the green beam can no longer be used as a reference for the arm cavity.  After some discussion we decided to go with the TM position as is, and to realign the green beam to the new position and relock the green beam to the new cavity.  We should be able to use the spot position of the green beam exiting the vacuum at the PSL table as the new reference.  If the green X beam exiting at the PSL table is severely displaced, we may decide to go back in and move ETMX to tweak the cavity alignment.

At this point we decided that we were done for the day.  Before closing up, we put a piece of foil with a hole in it in front of the the TM face, to use as an alignment aperture when Kiwamu does the green alignment.

Kiwamu will work on the green alignment over the weekend.  Assuming everything works out, we'll try the same procedure on ETMY on Monday.

  5128   Fri Aug 5 20:44:26 2011 jamieMetaphysicsTreasureFilm crew here Monday morning

Just a reminder that a film crew will be here Monday morning, filming Christian Ott for some Discovery channel show.

They are slated to be here from 8am to 12:30pm or so.  They will take a couple of shots inside the lab, and the rest of the filming should be of Christian in the control room (which they will "clean up" and fit with "sexy lighting").  I will try to be here the whole time to oversee everything.

  5129   Fri Aug 5 22:54:29 2011 kiwamuMetaphysicsTreasureCrane crew here Monday morning

Also, according to Steve, there will be some crane guys for fixing the Y end crane issue (#5124) Monday morning.

Quote from #5128

 a film crew will be here Monday morning. They are slated to be here from 8am to 12:30pm or so. 

  5130   Sat Aug 6 03:10:05 2011 SureshSummaryGeneralSummary of today's in-vacuum work

The table below gives the OSEM positions as seen on the slow chanels C1:SUS-ETMX_{UL,UR,LL,LR,SD}PDMon

ETMX_OSEMs.png

Note that the side OSEM has the fast channel (OUTPUT) available and we used that to locate it.

When we began work the OSEMs were photographed so that we have a record of their locations till now.  It was difficult to get accurate estimate of the magnet offset inside the OSEM we could not see the screen on the camera while clicking.  We then took some pictures after finishing the work. These are given below

 

          Before_osem_adj.JPG         After_osem_adjustment.JPG

 

The picture of the left is from before OSEMs were moved. It can be seen that OSEMs are rotated to make sure that the magnets avoid touching the teflon sheets which hold the shadow sensors.  The picture on the right shows the positions of the OSEMs after we adjusted their positions.  This time we kept the teflon sheets vertical as shown to minimise the coupling between the Side and Axial directions.

We needed to reposition them once again after we moved the tower to the center of the table.

Pictures with more detail will be posted to the wiki later.

 

 

 

 

  5131   Sat Aug 6 13:38:02 2011 ranaSummaryGeneralSummary of today's in-vacuum work

This OSEM placement is just the OPPOSITE of what the proper placement is.

Usually, we want to put them in so that the LED beam is vertical. This makes the OSEM immune to the optic's vertical mode.

The orientation with the horizontal LED beam makes the immunity to the side mode better, but may spoil the vertical.

In reality, neither of these assumptions is quite right. The LED beam doesn't come out straight. That's why Osamu and I found that we have to put in some custom orientations.

Also, the magnet gluings relative to the OSEM bracket centers are not perfectly aligned. So...I am saying that the OSEMs have to be oriented empirically to reduce the couplings which we want to reduce.

 

  5132   Sun Aug 7 11:10:54 2011 SureshSummaryGeneralOSEM positioning

Quote:

This OSEM placement is just the OPPOSITE of what the proper placement is.

Usually, we want to put them in so that the LED beam is vertical. This makes the OSEM immune to the optic's vertical mode.

The orientation with the horizontal LED beam makes the immunity to the side mode better, but may spoil the vertical.

In reality, neither of these assumptions is quite right. The LED beam doesn't come out straight. That's why Osamu and I found that we have to put in some custom orientations.

Also, the magnet gluings relative to the OSEM bracket centers are not perfectly aligned. So...I am saying that the OSEMs have to be oriented empirically to reduce the couplings which we want to reduce.

 

 Kiwamu suggested that since the side resonance is at a lower frequency than the bounce (~17Hz)  we ought to worry about the side more than the bounce.  If this is okay we can reposition the OSEMs to minimise this coupling. 

More over, in the current position, the OSEM s will not sense the side motion!!  So we definitely need to reposition them.  Sorry! I was being a spatz. 

  5133   Sun Aug 7 14:01:58 2011 kiwamuUpdateGreen LockingX green beam re-aligned

[Jenne / Kiwamu]

 The X green beam has been realigned to compensate the effect of the ETMX repositioning.

After the alignment we became able to lock the 00 mode with the X green beam.

 

For the alignment:

  spot position on the ETMX mirror = within ~ 1 cm. This number is strictly constrained by a homemade aluminum iris that Jamie put last Friday.

  spot position on the ITMX mirror = unknown, but looks pretty good on the CCD camera.

  spot position on the PSL table = ~ 1 mm downward from what it used to be. The horizontal alignment is perfect.

Conclusions :

  The X green beam again became a reference of the beam axis.

  The ETMX suspension tower is in a good place.

Quote from #5127

Kiwamu will work on the green alignment over the weekend.  Assuming everything works out, we'll try the same procedure on ETMY on Monday.

 

  5134   Sun Aug 7 14:11:53 2011 JenneUpdatePEMParticle counts through the roof

[Jenne, Kiwamu]

While Kiwamu was finalizing the X green alignment, I started to prepare to remove the ETMY door, and begin checking out its OSEMs, etc, so we could start moving it to it's new place, and figure out why it's been wonky for a while.  I ran the particle counter, and we have a factor of ~5 more particles than normal.  Kiwamu and I agreed not to open ETMY.  Since we had briefly opened the IOO and Output Optics chambers to check the X green's position on the PSL table, we immediately shut those doors.  They were probably open for ~15 minutes or so.  (Yes Steve, we should have checked before opening any doors, but at least we remembered to check at all, and the doors were only open for a few minutes rather than for a few hours.)

I attach a 24hrs trend of the particle counts, for reference.  It looks like it's been a little high for a while, but today it's really dirty in the air.

Attachment 1: ParticleCount_High_7Aug2011.png
ParticleCount_High_7Aug2011.png
  5135   Sun Aug 7 18:41:12 2011 JenneUpdateGreen LockingX green beam re-aligned

Quote:

The ETMY suspension tower is in a good place.

Quote from #5127

Kiwamu will work on the green alignment over the weekend.  Assuming everything works out, we'll try the same procedure on ETMY on Monday.

 

 You meant ETMX, right?  ETMY still hasn't been touched.

KI : sorry, I meant ETMX. I fixed the entry.

  5136   Mon Aug 8 00:12:58 2011 ranaUpdateCDSdiagonalization of MC input matrix

I've finally completed the SUS/peakFit/ scripts which find the new input matrix for the SUS. MC1, MC2, MC3, and ITMX have been matrix'd.

I tried to do the BS, but it came out with very funny matrix elements. Also the BS is missing its DAQ channels again (JAMIE !) so we can't diagnose it with the free swinging method.

To continue, we have to get some good data and try this again. Right now there are some weird issues with a lot of the optics. I've also set the damping gains for the optics with the new matrices.

 Ex.

new_matrix = findMatrix('ITMX')

writeSUSinmat('ITMX', new_matrix)

writeSUSinmat.m

this script writes the values to the MEDM input SUS matrix. To do the writing, I used the low level 'caput' command instead of ezcawrite since the ezca libraries are getting deprecated.

caput doesn't really have good diagnostics, so I use matlab to check the return status and then display to the terminal. You can just rerun it if it gives you an error.

 

 

A coupled of normalization notes:

1) The POS/PIT/YAW rows are scaled so that the mean of abs(FACE elements) = 1. Previously, I had the max element = 1.

2) The SIDE row is scaled so that the SIDE element = +1.

3) I then normalized the ROWS according to the geometrical factors that Jamie has calculated and almost put into the elog.

 

All these scripts have been added to the SVN. I've removed the large binary data files from the directory though. You can just rsync them in to your laptop if you want to run this stuff remotely.

ELOG V3.1.3-