40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 101 of 344  Not logged in ELOG logo
ID Date Author Typeup Category Subject
  2251   Thu Nov 12 11:19:10 2009 KojiUpdatePSLAbandoned Frequency Generator

Last night there was an activity for a calibratuon work, which I helped. I can take care of the FG.

Quote:

This morning I found a frequency generator connected to something on the PSL table sitting on the blue step next to the sliding doors.

Is anyone using it? Has it been forgotten there? If that's the case, can the interested person please take care of removing it?

 

  2252   Thu Nov 12 11:34:38 2009 AlbertoUpdateABSLWorking on the AP table

Quote:

Yes it did.

For long time, the crystal temperature C1:PSL-126MOPA_LTMP was 43~46deg. Now it is 34deg. Try ~10deg lower temperature.

Quote:

I wonder if the all the tinkering on the PSL laser done recently to revive the power have changed the PSL NPRO temperature and so its frequency. That could also explain why the beat doesn't show up at the same temperature of the NPRO as I used to operate it. Although I scanned the NPRO temperature +/- 2 deg and didn't see the beat.

 

 Beat found at 30MHz with auxiliary NPRO temperature of 37.19 degrees, vs. ~48 deg as it used to be.

The beat is small (-70dBm). PLL alignment has to be improved.

  2253   Thu Nov 12 12:50:35 2009 AlbertoUpdateComputersStochMon calibrated channels added to the data trend

I added the StochMon calibrated channels to the data trend by including the following channel names in the C0EDCU.ini file:

[C1:IOO-RFAMPD_33MHZ_CAL]
[C1:IOO-RFAMPD_133MHZ_CAL]
[C1:IOO-RFAMPD_166MHZ_CAL]
[C1:IOO-RFAMPD_199MHZ_CAL]

Before saving the changes I committed C0EDCU.ini to the svn.

Then I restarted the frame builder so now the new channels can be monitored and trended.

  2254   Thu Nov 12 12:51:45 2009 AlbertoUpdateABSLWorking on the AP table

Quote:

I've opened the AP table and I'm working on it.

Also auxiliary NPRO turned on and mechanical shutter opened.

AP table and aux NPRO shutter just closed.

  2255   Thu Nov 12 15:40:27 2009 josephb, koji, peterUpdateComputersETMY and Megatron test take 1

We connected megatron to the IO chassis which in turn was plugged into the rest of the ITMY setup.  We had manually turned the watchdogs off before we touched anything, to ensure we didn't accidently drive the optic.  The connections seem to go smoothly.

However, on reboot of megatron with the IO chassis powered up, we were unable to actually start the code.  (The subsystem has been renamed from SAS to TST, short for test).  While starttst claimed to start the IOC Server, we couldn't find the process running, nor did the medm screens associated with it work.

As a sanity test, we tried running mdp, Peter's plant model, but even that didn't actually run.  Although it also gave an odd error we hadn't seen before:

"epicsThreadOnceOsd epicsMutexLock failed."

Running startmdp a second time didn't give the error message, but still no running code.  The mdp medm screens remained white.

We turned the IO chassis off and rebooted megatron, but we're still having the same problem.

 

Things to try tomorrow:

1) Try disconnecting megatron completely from the IO chassis and get it to a state identical to that of last night, when the mdp and mdc did run.

2) Confirm the .mdl files are still valid, and try rebuilding them

  2256   Thu Nov 12 16:13:05 2009 AlbertoUpdatePSLMC Trans Offset

On Rana's suggestion I checked the MC transmission QPD (C1:IOO-MC_TRANS_SUM). I found that the readout is almost zero when the MC is unlocked.

I unlocked the Mode Cleaner turning off the LSC control and disabling the autolocker. The QPD reads 0.014. It seems that there is no offset.

I also checked with the IR card around the photodetector and I didn't see any stray beam.

  2257   Thu Nov 12 16:53:59 2009 AlbertoUpdateABSLWorking on the AP table

Quote:

Quote:

Yes it did.

For long time, the crystal temperature C1:PSL-126MOPA_LTMP was 43~46deg. Now it is 34deg. Try ~10deg lower temperature.

Quote:

I wonder if the all the tinkering on the PSL laser done recently to revive the power have changed the PSL NPRO temperature and so its frequency. That could also explain why the beat doesn't show up at the same temperature of the NPRO as I used to operate it. Although I scanned the NPRO temperature +/- 2 deg and didn't see the beat.

 

 Beat found at 30MHz with auxiliary NPRO temperature of 37.19 degrees, vs. ~48 deg as it used to be.

The beat is small (-70dBm). PLL alignment has to be improved.

 PLL alignment improved. Beat amplitude = -10dBm. Good enough.

DC readouts at the PLL photodiode:

V_NPRO = -4.44V

V_PSL = -3.76V

The NPRO beam is attenuated by a N.D.=1 attenuator just before going to the photodiode.

Something strange happened at the last. Right before -10dBm, the amplitude of the beat was about -33dBm. Then I was checking the two interfering beams with the IR card and saw that they overlapped quite well. I then turned my head back to the spectrum analyzer and suddenly the beat was at -10dBm. Not only, but a bunch of new peaks had appeared on the spectrum. Either I inadvertently hit the PD moving it to a better position or something else happened.

Like if someone was making some other modulation on the beam or the modulation depth of the PSL's sidebands had gone up.

  2258   Thu Nov 12 17:15:43 2009 KojiUpdatePSLMC Trans Offset

OK. I have been keeping my eyes on the MC transmission. In deed, the MC trans has been well kept at around 7.7 since the last PSL work.
Even it was over the 8 today!

Quote:

On Rana's suggestion I checked the MC transmission QPD (C1:IOO-MC_TRANS_SUM). I found that the readout is almost zero when the MC is unlocked.

I unlocked the Mode Cleaner turning off the LSC control and disabling the autolocker. The QPD reads 0.014. It seems that there is no offset.

I also checked with the IR card around the photodetector and I didn't see any stray beam.

 

  2260   Thu Nov 12 17:42:04 2009 KojiUpdatePSLMC Trans Offset

PC_DRIVE is also improving after the last PSL work!

Quote:

OK. I have been keeping my eyes on the MC transmission. In deed, the MC trans has been well kept at around 7.7 since the last PSL work.
Even it was over the 8 today!

 

  2261   Thu Nov 12 18:10:27 2009 AlbertoUpdateABSLPLL Locked

I locked the PLL and made some first measuremtns of the spectrum of the error signal. I'll post them later.

I closed the shutter of the NPRO.

  2262   Fri Nov 13 03:38:47 2009 kiwamuUpdateElectronicsmulti-resonant EOM --- impedance of LC tank circuit ----

I have measured the impedance of the LC tank circuit which I referred on my last entry.

The configuration of the circuit is exactly the same as that time.

In order to observe the impedance, by using Koji's technique I injected a RF signal into the output of the resonant circuit.

In addition I left the input opened, therefore the measured impedance does not include the effect of the transformer.

 

- - - - - - - - - - - - results

The measured impedance is attached below; "LCtank_impedance.png"

The peak around 50MHz is the main resonance and it has impedance of ~1500 [Ohm], which should go to infinity in the ideal case (no losses).

In fact the impedance looked from the input of the circuit gets reduced by 1/n^2, where "n" is the turn ratio of the transformer.

By putting the n=4, the input impedance of the circuit should be 93 [Ohm]. This is a moderate value we can easily perform impedance-matching by some technique.

I also fitted the data with a standard model of equivalent circuit (see attachment 2).

In the figure.2 red component and red letter represents the design. All the other black stuff are parasites.

But right now I have no idea the fitted value is reasonable or not.

For the next I should check the input impedance again by the direct way; putting the signal into the input.

 

 

 

  2263   Fri Nov 13 05:03:09 2009 kiwamuUpdateElectronicsmulti-resonant EOM --- input impedance of LC tank ----

I measured the input impedance of the LC tank circuit with the transformer. The result is attached.

It looks interesting because the input impedance is almost dominated

by the primary coil of the transformer with inductance of 75nH (see attachment 1).

The impedance at the resonance is ~100 [Ohm], I think this number is quite reasonable because I expected that as 93 [Ohm]

 

Note that the input impedance can be derived as follower;

(input impedance) = L1 + Z/n^2.

Where L1 is the inductance of the primary coil, Z is the load in the secondary loop and n is the turn ratio.

 

I think now I am getting ready to enter the next phase \(^o^)/

  2264   Fri Nov 13 09:47:18 2009 josephbUpdateComputersMegatron status lights lit

Megatron's top fan, rear ps, and temperature front panel lights were all lit amber this morning.  I checked the service manual, found at :

http://docs.sun.com/app/docs/prod/sf.x4600m2?l=en&a=view

According to the manual, this means a front fan failed, a voltage event occured, and we hit a high temperature threshold.  However, there were no failure light on any of the individual front fans (which should have been the case given the front panel fan light).  The lights remained on after I shutdown megatron.  After unplugging, waiting 30 seconds, and replugging the power cords in, the lights went off and stayed off.  Megatron seems to come up fine.

I unplugged the IO chassis from megatron, rebooted, and tried to start Peter's plant model.  However, it still prints that its starting, but really doesn't.  One thing I forgot to mention in the previous elog on the matter, is that on the local monitor it prints "shm_open(): No such file or directory" every time we try to start one of these programs.

  2266   Fri Nov 13 10:28:03 2009 josephb, alexUpdateComputersMegatron is back to its old self

I called Alex this morning and explained the problems with megatron.

Turns out when he had been setting up megatron, he thought a startup script file, rc.local was missing in the /etc directory.  So he created it.  However, the rc.local file in the /etc directory is normally just a link to the /etc/rc.d/rc.local file.  So on startup (basically when we rebooted the machine yesterday), it was running an incorrect startup script file.  The real rc.local includes line:

/usr/bin/setup_shmem.rtl mdp mdc&

Hence the errors we were getting with shm_open().  We changed the file into a soft link, and resourced the rc.local script and mdp started right up.  So we're back to where we were 2 nights ago (although we do have an RFM card in hand).

Update:  The tst module wouldn't start, but after talking to Alex again, it seems that I need to add the module tst to the /usr/bin/setup_shmem.rtl mdp mdc& line in order for it to have a shared memory location setup for it.  I have edited the file (/etc/rc.d/rc.local), adding tst at the end of the line.  On reboot and running starttst, the code actually loads, although for the moment, I'm still getting blank white blocks on the medm screens.

  2267   Fri Nov 13 14:04:27 2009 josephb, kojiUpdateComputersUpdated wiki with RCG instructions/tips

I've placed some notes pertaining to what Koji and I have learned today about getting the RCG code working on the 40m wiki at:

http://lhocds.ligo-wa.caltech.edu:8000/40m/Notes_on_getting_the_CDS_Realtime_Code_Generator_working

We're still trying to fix the tst system, as the moment its reporting an invalid number of daq channels and during daq initialization it fails.  (This from the /cvs/cds/caltech/target/c1tst/log.txt file). Note: This problem is only on megatron and separated from the conventional DAQ system of the 40m.

cpu clock 2800014
Warning, could open `/rtl_mem_tst' read/write (errno=0)
configured to use 2 cards
Initializing PCI Modules
2 PCI cards found
***************************************************************************
1 ADC cards found
        ADC 0 is a GSC_16AI64SSA module
                Channels = 64
                Firmware Rev = 512

***************************************************************************
1 DAC cards found
        DAC 0 is a GSC_16AO16 module
                Channels = 16
                Filters = None
                Output Type = Differential
                Firmware Rev = 3

***************************************************************************
0 DIO cards found
***************************************************************************
0 IIRO-8 Isolated DIO cards found
***************************************************************************
0 IIRO-16 Isolated DIO cards found
***************************************************************************
0 Contec 32ch PCIe DO cards found
0 DO cards found
***************************************************************************
0 RFM cards found
***************************************************************************
Initializing space for daqLib buffers
Initializing Network
Found 1 frameBuilders on network
Waiting for EPICS BURT at 0.000000 and 0 ns 0x3c40c004
BURT Restore = 1
Waiting for Network connect to FB - 10
Reconn status = 0 1
Reconn Check = 0 1
Initialized servo control parameters.
DAQ Ex Min/Max = 1 32
DAQ Tp Min/Max = 10001 10094
DAQ XTp Min/Max = 10094 10144
DAQ buffer 0 is at 0x8819a020
DAQ buffer 1 is at 0x8839a020
daqLib DCU_ID = 10
DAQ DATA INFO is at 0x3e40f0a0
Invalid num daq chans = 0
DAQ init failed -- exiting

  2268   Fri Nov 13 15:01:07 2009 JenneUpdateComputersUpdated wiki with RCG instructions/tips

Quote:

I've placed some notes pertaining to what Koji and I have learned today about getting the RCG code working on the 40m wiki at:

http://lhocds.ligo-wa.caltech.edu:8000/40m/Notes_on_getting_the_CDS_Realtime_Code_Generator_working

We're still trying to fix the tst system, as the moment its reporting an invalid number of daq channels and during daq initialization it fails.  (This from the /cvs/cds/caltech/target/c1tst/log.txt file).

 Dmass tells me that you have to record at least one channel.  ie at least one channel in your .ini file must be set to acquire, otherwise the DAQ will flip out.  It seems to be unhappy when you're not asking it to do things.

  2269   Fri Nov 13 22:01:54 2009 KojiUpdateComputersUpdated wiki with RCG instructions/tips

I continued on the STAND ALONE debugging of the megatron codes.

- I succeeded to run c1aaa with ADC/DAC. (c1aaa is a play ground for debugging.)

  The trick was "copy DAC block from sam.mdl to aaa.mdl".
  I don't understand why this works. But it worked.
  I still have the problem of the matrices. Their medm screens are always blank. Needs more works.

- Also I don't understand why I can not run the build of c1tst when I copy the working aaa.mdl to tst.mdl.

- The problem Joe reported: "# of channels to be daqed" was solved by

make uninstall-daq-aaa
make install-daq-aaa

  This command is also useful.

daqconfig

- Now I am in the stable development loop with those commands

killaaa
make uninstall-daq-aaa
make aaa
make install-aaa
make install-daq-aaa
make install-screens-aaa
startaaa

  I have made "go_build" script under /home/controls/cds/advLigo

usage:
./go_build aaa

- Note for myself: frequently visited directories

/home/controls/cds/advLigo/src/epics/simLink (for model)
/home/controls/cds/advLigo
(to build)
/cvs/cds/caltech/target/c1aaa
(realtime code log)
/cvs/cds/caltech/target/c1aaaepics (ioc log)
/cvs/cds/caltech/medm/c1/aaa (medm screens)
/cvs/cds/caltech/chans
(filter coeffs)
/cvs/cds/caltech/chans/daq (daq settings)

  2270   Sat Nov 14 06:46:48 2009 KojiUpdateComputersUpdated wiki with RCG instructions/tips

I am still working on the c1aaa code. Now it seems that C1AAA is working reasonably (...so far).

1) At a certain point I wanted clean up the system status. I have visited /etc/rc.local to add c1aaa for realtime to non-realtime task

before:
/usr/bin/setup_shmem.rtl mdp mdc tst&
after:
/usr/bin/setup_shmem.rtl mdp mdc tst aaa&

   I rebooted the system several times.

sudo /sbin/reboot

2) I found that gabage medm screens accumulated in ~/cds/advLigo/build/aaaepics/medm after many trials with several simulink models.
This directory is always copied to /cvs/cds/caltech/medm/c1/aaa at every make install-screens-aaa
This caused very confusing MEDM screens in the medm dir like C1AAA_ETMX_IN_MATRX.adl (NOT ETMY!)

I did

cd ~/cds/advLigo
make clean-aaa

to refresh aaaepics dir. The current development procedure is

killaaa
make clean-aaa
make uninstall-daq-aaa
make aaa
make install-aaa
make install-daq-aaa
make install-screens-aaa
startaaa

3) Sometimes startaaa does not start the task properly. If the task does not work, don't abandon.
Try restart the task. This may help. 

killaaa
(deep breathing several times)
startaaa

What to do next:

- MEDM works

* make more convenient custom MEDM screens so that we can easily access to the filters and switches
* retrofit the conventional SUS MEDM to the new system

- once again put/confirm the filter coeffs and the matrix elements

- configure DAQ setting so that we can observe suspension motion by dataviewer / dtt

- connect the suspension to megatron again

- test the control loop

  2271   Sun Nov 15 18:42:10 2009 AlbertoUpdateLockingInterferometer fully locked for 3331 seconds

This afternoon, I tried to lock the interferometer again after a few days.

After a couple of failed attempts, and relocks of the MZ, the interferometer stayed locked continuously for about 50 minutes, with arm power of about 92.

I just wanted to check the status of the interferometer so I didn't do any particular measurement in the meantime.

  2272   Mon Nov 16 09:37:41 2009 steveUpdatePEMconstruction is getting noisier

 

ATM1: Seismometers are saturating, suspensions are OK

Atm2 : activity next door,  diesel Backhoe and diesel concrete cutter are running

Atm3: CES exhaust fans output  is pick up by 40M-Annex-North AC unit  intake. The 40m office room has some diesel smell...........see # 2377

  2273   Mon Nov 16 15:13:25 2009 josephbUpdateComputersezcaread updated to Yoichi style ezcawrite

In order to get the gige camera code running robustly here at the 40m, I created a "Yoichi style" ezcaread, which is now the default, while the original ezcaread is located in ezcaread.bin.  This tries 5 times before failing out of a read attempt.

  2274   Mon Nov 16 15:18:10 2009 haixingUpdateSUSStable magnetic levitation without eddy-current damping

By including a differentiator from 10 Hz to 50 Hz, we increase the phase margin and the resulting

magnetic levitation system is stable even without the help of eddy-current damping.

The new block diagram for the system is the following:

 block_diagram_new.png

Here the eddy-current damping component is removed and we add an additional differential

circuit with an operational amplifier OP27G.

In addition, we place the Hall sensor below the magnet to minimize the coupling between

the coil and the Hall sensor.

The resulting levitation system is shown by the figure below:

mag_levi_new.jpg

  2277   Mon Nov 16 17:35:59 2009 kiwamuUpdateLSCOMC-LSC timing get synchronized !

An interesting thing was happened in the OMC-LSC timing clock.

Right now the clock of the OMC and the LSC are completely synchronized.

 The trend data is shown below. At the first two measurements (Oct.27 and Nov.1),  LSC had constant retarded time of 3Ts (~92usec.).

The last measurement, on Nov.15, number of shifts goes to zero, this means there are no retarded time.

Also the variance between the two signal gets zero, so I conclude the OMC and the LSC are now completely synchronized.

The measurement on Nov.8 is somehow meaningless, I guess the measurement did not run correctly by an influence from megatron(?)

 

OMC-LSC.png

 

  2278   Tue Nov 17 00:42:12 2009 KojiUpdateComputersUpdated wiki with RCG instructions/tips

Dmass, Joe, Koji


A puzzle has been solved: Dmass gave us a great tip

"The RGC code does not work unless the name of the mdl file (simulink model) matches to the model name "

The model name is written in the second line. This is automatically modified if the mdl file is saved from simulink.
But we copied the model by using "cp" command. This prevent from the TST model working!

megatron:simLink>head tst.mdl
Model {
  Name                    "tst"
  Version                 7.3
  MdlSubVersion           0

...
...
...

This explained why the AAA model worked when the DAC block has been copied from the other model.
This was not because of the ADC block but the saving model fixed the model name mismatch!


Now our current working model is "C1TST". Most of the functionalities have been implemented now:

  • The simulink model has been modified so that some of the functionalities can be accomodated, such as LSC/ASC PIT/ASC YAW.
  • Some filter names are fixed so as to inherit the previous naming conventions.
  • The SUS-ETMY epics screen was modified to fit to the new channel names, the filter topologies, and the matrices.
  • The chans file was constructed so that the conventional filter coefficients are inherited.
  • All of the gains, filter SWs, matrix elements have been set accordingly to the current ETMY settings.
  • burt snapshot has been taken: /cvs/cds/caltech/target/c1tstepics/controls_1091117_024223_0.snap
    burtrb -f /cvs/cds/caltech/target/c1tstepics/autoBurt.req -o controls_1091117_024223_0.snap -l /tmp/controls_1091117_024215_0.read.log -v

What to do next:

  • Revisit Oplev model so that it accomodates a power normalization functionality.
  • ETMY QPD model is also missing!
  • Clean up mdl file using subsystem grouping
  • Check consistency of the whitening/dewhitening switches.
  • Connect ADC/DAC to megatron
  • Test of the controllability
  • BTW, what is happened to BIO?
  • Implementation of the RFM card

Directories and the files:

  • The .mdl file is backed up as
    /home/controls/cds/advLigo/src/epics/simLink/tst.mdl.20091116_2100

  • The default screens built by "make" is installed in
    /cvs/cds/caltech/medm/c1/tst/
    They are continuously overridden by the further building of the models.

  • The custom-built medm screens are stored in
    /cvs/cds/caltech/medm/c1/tst/CustomAdls/

    The backup is
    /cvs/cds/caltech/medm/c1/tst/CustomAdls/CustomAdls.111609_2300/

  • The custom-built chans file is
    /cvs/cds/caltech/chans/C1TST.txt

    The backup is
    /cvs/cds/caltech/chans/C1TST.111609

  • burt snap shot file
    /cvs/cds/caltech/target/c1tstepics/controls_1091117_024223_0.snap
  2279   Tue Nov 17 10:09:57 2009 josephbUpdateEnvironmentFumes

The smell of diesel is particularly bad this morning.  Its concentrated enough to be causing me a headache.  I'm heading off to Millikan and will be working remotely on Megatron.

  2284   Tue Nov 17 21:09:17 2009 Jenne, ranaUpdateGeneralLittle Thorlabs Photodiode

[Rana, Jenne]

We opened up the little Thorlabs battery operated PD to see what was inside.  Rana took some pictures, and I drew a schematic (attached).  It's just a diode, biased with a battery (albeit a crazy 22.5V battery).

---------------
Comment by KA: PD is Hamamatsu S1223-01 Si PIN diode.


What a crazy battery. The main point is that it looks like this can be used for reasonable purposes: uses a load resistor on the BNC connector and you can use some pre-amp (e.g. Busby box or SR560) to have a low noise PD readout. You can also use the SR560 in its A-B mode as an 'opamp'. Ground the A input and the use a pole at 1 Hz and make the Output go into the B input through some large series resistor. The BNC from the PD gets Teed into the B input as well.

Then this becomes a transimpedance circuit readout of the diode, using the current noise of the SR560 as the limit.

  2287   Tue Nov 17 21:21:30 2009 robUpdateSUSETMY UL OSEM

Had been disconnected for about two weeks.  I found a partially seated 4-pin LEMO cable coming from the OSEM PD interface board. 

  2289   Wed Nov 18 01:12:15 2009 ranaUpdatePEMseismometers were not saturating during Halloween weekend
  2292   Wed Nov 18 14:55:59 2009 kiwamuUpdateElectronicsmulti-resonant EOM --- circuit design ----

The circuit design of multi-resonant EOM have proceeded.

By using numerical method, I found the some best choice of the parameters (capacitors and inductors).

In fact there are 6 parameters (Lp, L1, L2, Cp, C1, C2) in the circuit to be determined.

whole_circuit.png

In general the less parameter gives the less calculation time with performing the numerical analysis. Of course it looks 6 parameters are little bit large number.

In order to reduce the arbitrary parameters, I put 4 boundary conditions.

Each boundary conditions fixed resonant peaks and valleys; first peak=11MHz, third peak=55MHz, first valley=19MHz, second valley=44MHz.

designed.png

So now the remaining arbitrary parameters successfully get reduced to 2. Only we have to do is optimize the second peak as it to be 29.5MHz.

Then I take C1 and C2 as free parameters seeing how the second peak agree with 29.5MHz by changing the value of the C1 and C2.

mont.png

the red color represents the good agreement with 29.5MHz, in contrast blue contour represents the bad.

 You can see some best choice along the yellow belt. Now what we should do is to examine some of that and to select one of those.

  2294   Wed Nov 18 16:58:36 2009 kiwamuUpdateElectronicsmulti-resonant EOM --- EOM characterization ---

In designing the whole circuit it is better to know the characteristic of the EOM.

I made impedance measurement with the EOM (New Focus model 4064) and I found it has capacitance of 10pF.

This is good agreement with the data sheet which says "5-10pF".

The measured plot is attached below. For comparison there also plotted "open" and "10pF mica".

In the interested band( from 1MHz to 100MHz), EOM looks just a capacitor.

But indeed it has lead inductance of 12nH, resistance of 0.74[Ohm], and parasitic capacitance of 5.5pF.

In some case we have to take account of those parasites in designing.

EOM_impedance.png

 

  2295   Wed Nov 18 22:38:17 2009 KojiUpdateElectronicsmulti-resonant EOM --- EOM characterization ---

How can I get those values from the figure?

Quote:

But indeed it has lead inductance of 12nH, resistance of 0.74[Ohm], and parasitic capacitance of 5.5pF. 

 

  2297   Thu Nov 19 09:25:19 2009 steveUpdateMOPAwater was added to the laser chiller

I added ~500 cc of distilled water to the laser chiller yesterday.

  2298   Thu Nov 19 09:48:54 2009 steveUpdatePEMconstruction effect

8 days plot: Thurdsay, Friday, Sat and Sun without construction

  2299   Thu Nov 19 09:55:41 2009 josephbUpdateComputersTrying to get testpoints on megatron

This is a continuation from last night, where Peter, Koji, and I were trying to get test point channels working on megatron and with the TST module.

Things we noticed last night:

We could run starttst, and ./daqd -c daqdrc, which allowed us to get some channels in dataviewer.  The default 1k channel selection works, but none of the testpoints do. 

However, awgtpman -s tst does appear in the processes running list.

The error we get from dataviewer is:

Server error 861: unable to create thread
Server error 23328: unknown error
datasrv: DataWriteRealtime failed: daq_send: Illegal seek

Going to DTT, it starts with no errors in this configuration.  Initially it listed both MDC and TST channels.  However, as a test, I moved the tpchn_C4.par , tpchn_M4.par and tpchn_M5.par files to the directory backup, in /cvs/cds/caltech/target/gds/param.  This caused only the TST channels to show up (which is what we want when not running the mdc module.

We had changed the daqdrc file in /cvs/cds/caltech/target/fb, several times to get to this state.  According to the directions in the RCG manual written by Rolf, we're supposed to "set cit_40m=1" in the daqdrc file, but it was commented out.  However, when we uncommented it, it started causing errors on dtt startup, so we put it back.  We also tried adding lines:

set dcu_rate 13 = 16384;
set dcu_rate 14 = 16384;

But this didn't seem to help.  The reason we did this is we noticed dcuid = 13 and dcuid = 14 in the /cvs/cds/caltech/target/gds/param/tpchn_C1.par file.  We also edited the testpoint.par file so that it correctly corresponded to the tst module, and not the mdc and mdp modules.  We basically set:

[C-node1]
hostname=192.168.1.2
system=tst

in that file, and commented everything else out.

At this point, given all the things we've changed, I'm going to try a rebuild of the tst and daq and see if that solves things.

 

  2300   Thu Nov 19 10:19:04 2009 josephbUpdateComputersMegatron tst status

I did a full make clean and make uninstall-daq-tst, then rebuilt it.  I copied a good version of filters to C1TST.txt in /cvs/cds/caltech/chans/ as well as a good copy of screens to /cvs/cds/caltech/medm/c1/tst/.

Test points still appear to be broken.  Although for a single measurement in dtt, I was somehow able to start, although the output in the results page didn't seem to have any actual data in the plots, so I'm not sure what happened there - after that it just said unable to select test points.  It now says that when starting up as well.  The tst channels are the only ones showing up.  However, the 1k channels seem to have disappeared from Data Viewer, and now only 16k channels are selectable, but they don't actually work.  I'm not actually sure where the 1k channels were coming from earlier now that I think about it.  They were listed like C1:TST-ETMY-SENSOR_UL and so forth.

RA: Koji and I added the SENSOR channels by hand to the .ini file last night so that we could have data stored in the frames ala c1susvme1, etc.

  2307   Fri Nov 20 11:32:48 2009 HaixingUpdateSUSMagnetic levitation

I added an integrator to increase the gain at low frequencies (below 5 Hz). In addition, I increased

the band of the differentiator. The schematics for both integrator and differentiator are the following:

IntDiff.PNG

The magnetic is stably levitated.

S8007340.jpg

I turned off the light to get rid of 60Hz noise on the photodiode. I tried to measured the

open-loop transfer function of this setup, but somehow the SR560 is always saturate

when I injected the signal from SR785, which produces some weird results at

low-frequencies.

 

In addition, I found out that when the light is turned on, the levitation

can be stable even when I inverted the sign of the control loop. The control signal

on the osciloscope is the following:

S8007333.jpg

 

This oscillator is around 120 Hz, which should be  the harmonics of 60 Hz from light pollution.

I am not sure exactly why it is stable when the control-loop sign is flipped. This could

be similar to the Pauli trap in the iron trap, because the coil not only provides a force

but also provides the rigidity. The sign of such rigidity depends on the sign of the control

current. If such oscillating rigidity changes at a frequency much higher than the response

frequency of the magnet, it will stablize  the system simply by significantly increasing

the inertial of the magnet.More investigations are essential to completely understand it.

 

For information about Pauli trap, one can look at the wikipedia:

http://en.wikipedia.org/wiki/Quadrupole_ion_trap

 

 

 

  2310   Fri Nov 20 17:44:38 2009 JenneUpdateAdaptive FilteringSome svn shenanigans

[Sanjit, Jenne]

Sanjit and I are trying to put names to some signals which exist in SimuLink land, but which don't (yet) exist in EPICS land.  The deelio is that for each of the chosen SEIS signals in the ASS_TOP_PEM screen, the signal is split.  One part of the signal is used to decide how the adaptive filter should look, and the other part is actually used when doing the on-line subtraction.  Previously only the part of the signal which is used to decide on the Adaptive Filter could be seen on the screens, and had names. 

Before touching anything on the Simulink ASS.mdl, I did an svn check in, which put things at revision 36639. 

To try to make the desired signals exist, I put cdsFilt boxes (to create filter modules for each of these signals), and gave each of them a name (kind of like the Neverending Story....once they have a name, they'll exist).  My new names are C1:ASS-TOP_PEM_#_APPLY, which correspond to the previously-existing C1:ASS-TOP_PEM_#_ADPT (these are the ones that are along the top of the ASS_TOP_PEM matrix screen).  This version of the simulink model was checked in, and the svn is now at revision 36640.

We then did some "make clean", "make ass" and "make install-ass" action, and burt restored c1assepics, but nothing seems to be happening.  The screen doesn't have white boxes all over the place, and we didn't get any errors when we did the makes, and I'm sure we burt restored correctly (made sure the ASS GDS screen had a 1 in the lower left box etc), but all the values on the screen are still zero.  

When we ran the ass front end in terminal on the c1ass machine, we did see an error: "Invalid chan num found 2 = 30624" "DAQ init failed -- exiting".  I think this means that we need to have told some file somewhere that I was going to be adding 8 new channels. (maybe an .ini file?) Hopefully the Joe & Peter team can help us out with this, since they've been doing this kind of thing for the new system.

Moral of the story is, the new (non-working) simulink file has been svn checked in as revision 36640, and we're reverting to revision 36639, which was before I touched anything today.

  2311   Mon Nov 23 00:46:09 2009 rana, robUpdatePSLISS RIN: Its too high by 10x

This plot shows the RIN as measured by the ISS. Its ~2 x 10^-7, whereas its supposed to be more like 3 x 10^-8.

The ISS has DC coupled RIN channels (with a _F suffix) and AC coupled RIN channels (with a _FW suffix). By using a swept sine, Rob determined that the AC coupled channels have an AC coupling pole at ~80 Hz. The attached plot uses this and then has the overall gain adjusted to match with the _F channels below 10 Hz.

The _F channels can be converted directly into RIN by just dividing the spectra by the mean value of the time series. The dark offset of these channels is small and so this only introduces a ~5-10% calibration error.

Question #1: Why is the RIN so bad? According to the MEDM screen, the photocurrent on the MON/SENS PDs is 1.9/1.3 mA. That's sort of low, but should still allow us to get 5x10^-8 in RIN.

Question #2: Does it make an effect on the current DC Readout work? IF so, should we try to fix up the ISS in a temporary way? Since the in-loop and out-of-loop detectors are completely coherent, all of the noise is likely just unsuppressed noise from the laser. We are unable to increase the gain because of the high frequency noise from the NPRO.

 

Let's remember to replace this ISS with a new one that can drive an AOM. Need a volunteer to get us a new ISS.

 

  2312   Mon Nov 23 10:11:03 2009 steveUpdatePEMlong term temp fluctuation of the 40m lab

Quote:

This first plot shows the RC temperature channels' performance from 40 days ago, before we disabled the MINCO PID controller. Although RCTEMP is supposed to be the out of loop sensor, what we really care about is the cavity length and so I've plotted the SLOW. To get the SLOW on the same scale, I've multiplied the channel by 10 and then adjusted the offset to get it on the same scale.

 The second plot shows a period after that where there is no temperature control of the can at all. Same gain scaling has been applied to SLOW as above, so that instead of the usual 1 GHz/V this plot shows it in 0.1 GHz/V.

The third plot shows it after the new PID was setup.

Summary: Even though the PID loop has more gain, the true limit to the daily fluctuations in the cavity temperature and the laser frequency are due to the in-loop sensors measuring the wrong thing. i.e. the out-of-loop temperature is too different from the in-loop sensor. This can possibly be cured with better foam and better placement of the temperature sensors. Its possible that we're now just limited by the temperature gradients on the can.

 Here is a 7 years plot of  of the 40m temperature variations.

  2313   Mon Nov 23 11:03:00 2009 steveUpdateSUSjackhammer special well under control

Quote:

I've changed the watchdog rampdown script so it brings the SUS watchdogs to 220, instead of the 150 it previously targeted.  This is to make tripping less likely with the jackhammering going on next door.  I've also turned off all the oplev damping.

 Saturday's special event of braking up the large concrete pieces in CES bay was un event full.

  2315   Mon Nov 23 17:53:08 2009 JenneUpdateComputers40m frame builder backup acting funny

As part of the fb40m restart procedure (Sanjit and I were restarting it to add some new channels so they can be read by the OAF model), I checked up on how the backup has been going.  Unfortunately the answer is: not well.

Alan imparted to me all the wisdom of frame builder backups on September 28th of this year.  Except for the first 2 days of something having gone wrong (which was fixed at that time), the backup script hasn't thrown any errors, and thus hasn't sent any whiny emails to me.  This is seen by opening up /caltech/scripts/backup/rsync.backup.cumlog , and noticing that  after October 1, 2009, all of the 'errorcodes' have been zero, i.e. no error (as opposed to 'errorcode 2' when the backup fails).  

However, when you ssh to the backup server to see what .gwf files exist, the last one is at gps time 941803200, which is Nov 9 2009, 11:59:45 UTC.  So, I'm not sure why no errors have been thrown, but also no backups have happened. Looking at the rsync.backup.log file, it says 'Host Key Verification Failed'.  This seems like something which isn't changing the errcode, but should be, so that it can send me an email when things aren't up to snuff.  On Nov 10th (the first day the backup didn't do any backing-up), there was a lot of Megatron action, and some adding of StochMon channels.  If the fb was restarted for either of these things, and the backup script wasn't started, then it should have had an error, and sent me an email.  Since any time the frame builder's backup script hasn't been started properly it should send an email, I'm going to go ahead and blame whoever wrote the scripts, rather than the Joe/Pete/Alberto team.

Since our new raid disk is ~28 days of local storage, we won't have lost anything on the backup server as long as the backup works tonight (or sometime in the next few days), because the backup is an rsync, so it copies anything which it hasn't already copied.  Since the fb got restarted just now, hopefully whatever funny business (maybe with the .agent files???) will be gone, and the backup will work properly. 

I'll check in with the frame builder again tomorrow, to make sure that it's all good.

  2316   Mon Nov 23 19:36:28 2009 JenneUpdateAdaptive FilteringHow to add ASS channels, so that they're saved to frames

[Jenne, Sanjit]

We would like several channels from the OAF/ASS screen to be saved to frames, so that we can use the channels for our OAF model.  In theory, this should involve uncommenting the desired channels in the .ini file (.../caltech/chans/daq/C1ASS.ini), and restart the frame builder.  Since this .ini file was generated a long time ago, and things have been changed since then, the chnnums in the .ini file and the corresponding .par file don't match up.  We need to go through the .par file (/cvs/cds/gds/param/tpchn_c3.par), and look up the chnnums for our channels, and copy those numbers into the .ini file.  Figuring out what was going on involved many fb40m restarts, but on the last one of the night, I restarted the backup script, so it should (hopefully) run tonight, and get all of the frames that we've been missing.

Notes to self: 

*  When adding channels to other front ends, the end of the process is to click the blue button on the C0DAQ_DETAIL screen next to your computer.  C1ASS isn't on that screen.  Instead, in the C1ASS_GDS screen, click DAQ Reload.

*  The channel names for the Test Points and the .ini files must be different.  That's why there's a '_2048' suffix at the end of every channel in our .ini file.

*  tpchn_C1 is all of the old-style system test points.  tpchn_C2 is the C1OMC, and tpchn_C3 is for the C1ASS testpoints.

*  When uncommenting channels in the C1ASS.ini file, make sure acquire is set to 1 for every channel we want saved.  The default in this .ini file is set to acquire = 0.

  2317   Mon Nov 23 21:30:29 2009 JenneUpdateLSCMeasured MC length

With Koji's help, I measured the length of the Mode Cleaner.

The new modulation frequencies (as quoted on the Marconi front panels) are: 

165.980580 MHz

 33.196116 MHz

132.784464 MHz

199.176696 MHz

The Frequency Counter readback is 165980584.101 Hz (a 4Hz difference).  All of the Marconi's front-panel frequencies read ###.##### MHz Ext, and the Frequency standard has it's "locked" light illuminated, and the 1pps input light blinking, so I think everything is still nicely locked to the frequency standard, and the frequency standard is locked to the GPS.

While changing the marconi's, I accidentally touched the MC's 29.5 MHz marconi.  It is set back to the nominal value (according to Kiwamu's rack photos) of 29.485MHz.  But the phase might be sketchy, although hopefully this doesn't matter since we don't do a double demodulation with it.

I also ran the scripts in the wiki page: How To/Diagonalize DRMI Length Control to set the DD Phases.

 

 

  2318   Mon Nov 23 21:36:38 2009 KojiUpdateIOOAligned PMC/RC

I aligned the beam goes to PMC. It increased the MC Trans from 8.25 to 8.30.

I also aligned the beam goes to RC.
When I touched the FSS box (wrong: this was the VCO driver) that was close to one of the steering mirror, suddenly the RC trans increased.
It is now 9.8. I am afraid that it gets saturated. I could not reproduce the phenomenon. This could be caused by a bad contact?
Note that I didn't see there is any loose optic.

  2319   Tue Nov 24 08:00:16 2009 ranaUpdateLSCMeasured MC length

I propose that from now on, we indicate in the elog what frequencies we're referring to. In this case, I guess its the front panel readback and not the frequency counter -- what is the frequency counter readback? And is everything still locked to the 10 MHz from the GPS locked Rubidium clock?

Plus, what FSS Box? The TTFSS servo box? Or the VCO driver? As far as I know, the RC trans PD doesn't go through the FSS boxes, and so its a real change. I guess that a bad contact in the FSS could have made a huge locking offset.

 

  2320   Tue Nov 24 10:36:21 2009 KojiUpdateLSCMeasured MC length

What I meant was the VCO driver, not the FSS box.

As for the frequency, all written numbers were the Marconi displays.
The number on the frequency counter was also recorded, and so will be added to the previous entry shortly... 

Quote:

I propose that from now on, we indicate in the elog what frequencies we're referring to. In this case, I guess its the front panel readback and not the frequency counter -- what is the frequency counter readback? And is everything still locked to the 10 MHz from the GPS locked Rubidium clock?

Plus, what FSS Box? The TTFSS servo box? Or the VCO driver? As far as I know, the RC trans PD doesn't go through the FSS boxes, and so its a real change. I guess that a bad contact in the FSS could have made a huge locking offset.

 

  2321   Tue Nov 24 14:33:22 2009 AlbertoUpdateABSLworking on the AP table

I'm working on the AP table. I also opened the auxiliary NPRO shutter. The auxiliary beam is on its path on the AP table and PSL table.

  2322   Tue Nov 24 16:06:45 2009 JenneUpdateComputers40m frame builder backup acting funny

Quote:

As part of the fb40m restart procedure (Sanjit and I were restarting it to add some new channels so they can be read by the OAF model), I checked up on how the backup has been going.  Unfortunately the answer is: not well.

I'll check in with the frame builder again tomorrow, to make sure that it's all good.

 All is well again in the world of backups.  We are now up to date as of ~midnight last night. 

  2324   Tue Nov 24 19:16:02 2009 AlbertoUpdateABSLworking on the AP table

Quote:

I'm working on the AP table. I also opened the auxiliary NPRO shutter. The auxiliary beam is on its path on the AP table and PSL table.

 Closing the AP table and the NPRO shutter now.

  2325   Wed Nov 25 03:05:15 2009 robUpdateLockingMeasured MC length

Quote:

What I meant was the VCO driver, not the FSS box.

As for the frequency, all written numbers were the Marconi displays.
The number on the frequency counter was also recorded, and so will be added to the previous entry shortly... 

Quote:

I propose that from now on, we indicate in the elog what frequencies we're referring to. In this case, I guess its the front panel readback and not the frequency counter -- what is the frequency counter readback? And is everything still locked to the 10 MHz from the GPS locked Rubidium clock?

Plus, what FSS Box? The TTFSS servo box? Or the VCO driver? As far as I know, the RC trans PD doesn't go through the FSS boxes, and so its a real change. I guess that a bad contact in the FSS could have made a huge locking offset.

 

 

Locking has gone sour.  The CARM to MCL handoff, which is fairly early in the full procedure and usally robust, is failing reliably. 

As soon as the SUS-MC2_MCL gain is reduced, lock is broken.  There appears to be an instability around 10Hz.  Not sure if it's related.

ELOG V3.1.3-