ID |
Date |
Author |
Type |
Category |
Subject |
9748
|
Mon Mar 24 22:13:37 2014 |
steve | Update | General | Power Failure |
Quote: |
I'm checking the status from home.
P1 is 8e-4 torr
nodus did not feel the power outage (is it APS supported?)
linux1 booted automatically
c1ioo booted automatically.
c1sus, c1lsc, c1iscex, c1iscey need manual power button push.
|
9:11pm closed PSL shutter, turned Innolight 2W laser on,
turned 3 IFO air cond on,
CC1 5.1e-5 torr, V1 is closed, Maglev has failed, valve configuration is " Vacuum Normal " with V1 & VM1 closed, RGA not running, c1vac1 and c1vac2 were saved by UPS,
(Maglev is not connected to the UPS because it is running on 220V)
reset & started Maglev.........I can not open V1 without the 40mars running...........
Rossa is the only computer running in the control room,
Nodus and Linux1 was saved by UPS,
turned on IR lasers at the ends, green shutters are closed
It is safe to leave the lab as is.
|
Attachment 1: poweroutage.png
|
|
9747
|
Mon Mar 24 21:36:28 2014 |
Koji | Update | General | Power Failure |
I'm checking the status from home.
P1 is 8e-4 torr
nodus did not feel the power outage (is it APS supported?)
linux1 booted automatically
c1ioo booted automatically.
c1sus, c1lsc, c1iscex, c1iscey need manual power button push. |
9746
|
Mon Mar 24 19:42:12 2014 |
Charles | Frogs | VAC | Power Failure |
The 40m experienced a building-wide power failure for ~30 seconds at ~7:38 pm today.
Thought that might be important... |
9745
|
Mon Mar 24 10:41:46 2014 |
Steve | Update | VAC | RGA scan at day 50 |
|
Attachment 1: pd77m50dRGA.png
|
|
9744
|
Sun Mar 23 14:20:07 2014 |
rana | HowTo | LSC | BLRMS screens |
We should make screens like this for the LSC signals, errors, ALS, etc. |
Attachment 1: blrms.png
|
|
9743
|
Fri Mar 21 14:59:44 2014 |
steve | Update | PEM | controling dust |
Please take a look at the table top with the flashlight before removing it. If it is dusty, wipe it down with dry lint free cloth in the box.
There is one box with flash light and wiper at AP, ETMY & ETMX optical tables. |
Attachment 1: IMG_0044.JPG
|
|
9742
|
Fri Mar 21 01:54:32 2014 |
ericq | Update | LSC | Some early CARM modeling |
I've been getting a simulation going with the eventual goal of simulating CESAR-type signals for CARM. So for I've only been using MIST, though I'm still thinking about what to do for a fully time domain approach. (For example, maybe a mixture of simulink and analytical equations? We'll see how painful that gets...)
Anyways, with the parameters I have for the 40m, I've set up a simulation, where I can do things like a "static" CARM scan.
(i.e. PRMI perfectly locked. Ask what different PDs see if the arms were just statically sitting at some CARM offset)

PDH signals are there in the REFL diodes. The coupled line width here looks smaller than the ~40pm number I've heard before, so I should check my parameters. (Likely culprit, I'm using nominal R and T for the arm cavities)
I've also done the slightly more sophisticated thing of looking at the transfer function from CARM motion to different PDs, at different CARM offsets. For TRX and REFLDC, these seem to match up qualitatively to some plots that Kiwamu has done for aLIGO, with frequencies shifted by the relative arm length factor of 100. (Q's left, K's right, Y-axis on mine are W/m with 1W input the IFO)
 

We can also look at the PDH diodes (revised from my initial post. Had an error in my code):
 
That's where I've gotten so far!
|
9741
|
Thu Mar 20 11:16:16 2014 |
Steve | Update | General | new projector |
[ Manasa, Ericq and Steve ]
Vivitek D952HD with 186 hours installed. |
9740
|
Wed Mar 19 21:37:45 2014 |
manasa | Summary | LSC | Attempt to lock PRMIsb with REFL165I&Q |
I tried to repeat Koji's PRMI lock using REFL165I/Q. I was not able to lock PRMI stably. All I could get was momentary PRMI sb locks (few seconds) using AS55Q for MICH and REFL165Q for PRMI. I tried to transition MICH locks from AS55Q to REFL165I/Q and this did not work well; I lost even the momentary locks.
The demod phases for both AS55 and REFL165 were also very different.
Input ports:
AS55 WHTN: 21dB demod phase -78.7deg
REFL165 WHTN: 45dB demod phase -80.7deg
Input matrix:
AS55Q x1.00 MICH
REL165Q x+0.14
Triggers:
MICH POP110I 100up/10down / FM Trig FM2/3/6/7/9 35up 2down 5sec delay
PRCL POP110I 100up/10down / FM Trig FM2/3/6/9 35up 2down 0.5sec delay
Servo:
MICH OFS 0.0 / Gain -10 / Limiter ON
PRCL OFS 0 / Gain -0.023 / Limiter ON
Output matrix:
MICH ITMX -1.0 / ITMY +1.0
PRCL PRM 1.0
|
9739
|
Tue Mar 18 21:19:22 2014 |
Koji | Summary | IOO | MC spot positions checked |
MC spot sposition script was ran
/opt/rtcds/caltech/c1/scripts/ASS/MC/mcassMCdecenter
Found no notable beam position change before and after the earthquake
|
Attachment 1: MCASS.png
|
|
9738
|
Tue Mar 18 18:05:57 2014 |
Steve | Update | SUS | ETMX osem UR |
UR osem IR shield glass is pushed back. It came out of its clip holder. The magnet is free.
Atm2, UL & LL magnets centered poorly. Almost hinging on opposite sides.
UR & LR centered well. There have plenty of room to move in an earth quake. |
Attachment 1: ETMX_URcollapsing.jpg
|
|
Attachment 2: ETMX_outoffocus.JPG
|
|
Attachment 3: URisFREE.png
|
|
9737
|
Tue Mar 18 08:31:37 2014 |
Steve | Update | SUS | 4.4M local earthquake & its miracle |
Quote: |
It was little bit surprising to me but Rana's professorial rock'n roll excitation released its sticking on the unconfirmed thing by unconfirmed reason.
I aligned the Xarm manually and via ASS.
Now we are back in the normal state.
|
This recovery proceeder deserves a pattern
Note: IR shield glass position variations, Atm4 |
Attachment 1: miracel.png
|
|
Attachment 2: ETMX_LR.JPG
|
|
Attachment 3: ETMX_LR-UL.JPG
|
|
Attachment 4: ETMX_UL-LL-S-IRshield.jpg
|
|
9736
|
Tue Mar 18 00:51:02 2014 |
Jenne | Update | SUS | 4.4M local earthquake |
I am really, really happy to hear that it was just a sticking situation. Really happy. |
9735
|
Mon Mar 17 21:55:36 2014 |
Koji | Update | SUS | 4.4M local earthquake |
It was little bit surprising to me but Rana's professorial rock'n roll excitation released its sticking on the unconfirmed thing by unconfirmed reason.
I aligned the Xarm manually and via ASS.
Now we are back in the normal state. |
9734
|
Mon Mar 17 20:44:42 2014 |
rana | Update | SUS | 4.4M local earthquake |
|
9733
|
Mon Mar 17 20:14:34 2014 |
Koji | Summary | General | IFO recovery / confirmed ETMX in trouble |
I tried to take the photos of the magnets from outside. So far most suspicious was LL.
Otherwise, the magnets are OK.
(The SD magnet is the one with most reasonable response.)
Steve will try to take much more zoomed photo with Olympus. But the LL coil already showed some response in my observation in the morning.
   
|
9732
|
Mon Mar 17 12:31:58 2014 |
manasa | Update | CDS | fb timing was off |
Off again. Restarted ntp on fb. |
9731
|
Mon Mar 17 12:02:55 2014 |
Koji | Summary | General | IFO recovery / confirmed ETMX in trouble |
I confirmed that we need to vent the chambers.
All of the mirrors have been aligned except for ETMX.
ETMX does not respond to the excitation by the UR and LR coils. Likely that the magnets are knocked off, or stuck in the coil.
PRM/SRM oplevs are too much off and can't be turned on. Need realignment of the beams on the QPDs.
- FB was down. FB restarted ("telnet fb 8087", then type shutdown)
- Aligned the MC mirrors.
- Aligned PRM. Look at the REFL. It was slightly mislisligned.
- AS has no beam. The Y arm was resonating with the green. So I determined that the TTs were the misaligned guys.
- Touched TT2 pitch with an increment of 0.1. Immediately the AS beam spot for ITMY was found. And the arm was resonating.
- The RM was further aligned. The bias sliders were saved and then the PRM was misaligned.
- Yarm was locked on TEM01. The ASS maximized the transmission for TEM01, and then the arm was locked on TEM00.
The ASS aligned the arm and TTs. These values were saved.
- Yarm was aligned and I can see the AS spot. So I believe the BS is still well aligned.
- Aligned the PRM to reduce the ghost beams.
- Moved the ITMX to have Michelson fringes properly.
- Also aligned the SRM.
- Now ETMX was checked. Played with the alignment biases to see if the mirror was sticking on the coils. The mirror can rock a little, but it did not come back.
- Then, checked each magnets. 0.8Hz 1000cnt signals were injected to each coils (cf. C1:SUS-PRM_**COIL_EXC) to see how the mirror could react.
The OSEM output and green spot on the ETMX cage were observed.
- Saw some response by actuating the UL, LL, SD coils.
- Saw no response from the UR and LR coils. They show the OSEM output of zero. Does this mean the magnets fell in the coils?
//Manasa// MC spot positions measured and they look alright with not much change from before the earthquake (attach) |
Attachment 1: MCspots.png
|
|
9730
|
Mon Mar 17 10:50:58 2014 |
steve | Update | safety | 2014 annual crane certification |
Quote: |
Quote: |
We had our annual safety inspection today. Our SOPs are outdated. The full list of needed correction will be posted tomorrow.
The most useful found was that the ITMX-ISCT ac power is coming from 1Y1 rack. This should actually go to 1Y2 LSC rack ?
Please test this so we do not create more ground loops.
|
Annual crane inspection is scheduled for 8-11am Monday, March 17, 2014
The control room Smart UPS has two red extension cords that has to be removed: Nodus and Linux1
|
KroneCrane Fred inspected and certified the 3 40m cranes for 2014. The vertex crane crane was load tested at fully extended position. |
Attachment 1: 2014craneCert.jpg
|
|
9729
|
Mon Mar 17 09:27:05 2014 |
Steve | Update | SUS | 4.4M local earthquake |
It looks like that ETMX have 2 sticky magnets.
|
Attachment 1: 4.4M-Encino.png
|
|
Attachment 2: 3.9-4.4Meq.png
|
|
Attachment 3: EQdamage4.4.png
|
|
9728
|
Fri Mar 14 12:18:55 2014 |
Koji | Update | LSC | Composite Error Signal for ARms (9) |
Asymptotic cooling of the mirror motion with CESAR was tested.
With ALS and the full control bandwidth (UGF 80-100Hz), the actuator amplitude of 8000cnt_pp is required.
Varying control bandwidth depending on the noise level of the signal, the arm was brought to the final configuration with the actuator amplitude of 800cnt_pp. |
Attachment 1: asymptotic_cooling.pdf
|
|
9727
|
Fri Mar 14 10:31:10 2014 |
jamie | Update | Green Locking | ALS Slow servo settings |
Quote: |
Q and I have started to...
|
Ha! |
9726
|
Fri Mar 14 09:44:34 2014 |
Steve | Update | LSC | morning lock |
|
Attachment 1: 2hrsMorningLock.png
|
|
9725
|
Thu Mar 13 16:05:48 2014 |
steve | Update | safety | safety audit 2014 |
Quote: |
We had our annual safety inspection today. Our SOPs are outdated. The full list of needed correction will be posted tomorrow.
The most useful found was that the ITMX-ISCT ac power is coming from 1Y1 rack. This should actually go to 1Y2 LSC rack ?
Please test this so we do not create more ground loops.
|
Annual crane inspection is scheduled for 8-11am Monday, March 17, 2014
The control room Smart UPS has two red extension cords that has to be removed: Nodus and Linux1 |
Attachment 1: T1400198-v1-CIT-LIGO_Labs_Annual_Safety_Audit_2014.pdf
|
|
9724
|
Thu Mar 13 01:18:00 2014 |
Jenne | Update | LSC | Composite Error Signal for ARms (8) |
[Jenne, EricQ]
As Koji suggested in his email this afternoon, we looked at how much actuator range is required for various forms of arm locking: (1) "regular" PDH lock aquisition, (2) ALS lock acquisition, (3) CESAR cooling.
To start, I looked at the spectra and time series data of the control signal (XARM_OUT) for several locking situations. Happily, when the arm is at the half fringe, where we expect the 1/sqrt(TRX) signal to be the most sensitive (versus the same signal at other arm powers), we see that it is in fact more quiet than even the PDH signal. Of course, we can't ever use this signal once the arm is at resonance, so we haven't discovered anything new here.

EricQ then made some violin plots with the time series data from these situations, and we determined that a limit of ~400 counts encompasses most of the steady-state peak-to-peak output from locking on the PDH signal.
 
[ericq: What's being plotted here are "kernel density estimates" of the time series data of XARM_OUT when locked on these signals. The extent of the line goes to the furthest outlier, while the dashed and dotted lines indicate the median and quartiles, respectively]
I tried acquiring ALS and transitioning to final PDH signals with different limiters set in the Xarm servo. I discovered that it's not too hard to do with a limit of 400 counts, but that below ~350 counts, I can't keep the ALS locked for long enough to find the IR resonance. Here's a plot of acquiring ALS lock, scanning for the resonance, and then using CESAR to transition to PDH, with the limit of 400 counts in place, and then a lockloss at the end. Even though I'm hitting the rails pretty consistently, until I transition to the more quiet signals, I don't ever lose lock (until, at the end, I started pushing other buttons...).

After that, I tried acquiring lock using our "regular" PDH method, and found that it wasn't too hard to capture lock with a limit of 400, but with limits below that I can't hold the lock through the boosts turning on.
 
Finally, I took spectra of the XARM_OUT control signal while locked using ALS only, but with different limiter values. Interestingly, I see much higher noise between 30-300 Hz with the limiter engaged, but the high frequency noise goes down. Since the high frequency is dominating the RMS, we see that the RMS value is actually decreasing a bit (although not much).

We have not made any changes to the LSC model, so there is still no smoothing between the ALS and IR signals. That is still on the to-do list. I started modifying things to be compatible with CARM rather than a single arm, but that's more of a daytime-y task, so that version of the c1lsc model is saved under a different name, and the model that is currently compiled and running is reverted as the "c1lsc.mdl" file. |
9723
|
Wed Mar 12 08:40:42 2014 |
Steve | Update | VAC | RGA scan at day 35 |
|
Attachment 1: pd77m35d.png
|
|
9722
|
Tue Mar 11 21:38:43 2014 |
manasa | Update | Computer Scripts / Programs | IFO configure scripts in burt modified |
I have modified the IFOconfigure scripts and the corresponding .req files for the X arm and Y arm in burt. I have also added configure scripts to save and restore LSC settings for locking the arms using ALS error signals.
The settings are yet to be saved and the scripts should also be checked if they are working as required. |
9721
|
Tue Mar 11 19:38:26 2014 |
manasa | Update | Green Locking | ALS Slow servo settings |
Quote: |
Nic, Jenne, EricQ, and Koji should describe the demonstartion of CESAR achieved tonight.
|
Q and I have started to use the ALS slow servo for the end aux lasers while locking the arms using ALS. The servo prevents us from hitting the limits of the PZT range for the end lasers and a better PDH locking.
But keeping the servo ON causes the slow output to drift away making it hard to find the beat note everytime the arm loses lock. The extensive beat note search following the unlock can be avoided by clearing history of the slow servo. |
9720
|
Tue Mar 11 19:07:24 2014 |
ericq | Update | Electronics | High gain Trans PD electronics change |
Speaking of the whitening board, I had neglected to post details showing the the whitening was at least having a positive effect on the transmon QPD noise. So, here is a spectrum showing the effects that the whitening stages have on a QPD dark noise measurement like I did in ELOG 9660, at a simulated transmission level of 40 counts.
The first whitening stages gives us a full 20dB of noise reduction, while the second stage brings us down to either the dark noise of the QPD or the noise of the whitening board. We should figure out which it is, and fix up the board if necessary.

The DTT xml file is attached in a zip, if anyone wants it. |
Attachment 2: sqrtinvWhitening.zip
|
9719
|
Tue Mar 11 18:34:11 2014 |
Jenne | Update | LSC | Composite Error Signal for ARms (7) |
[Nic, Jenne, EricQ, and Koji]
We have used CESAR successfully to bring the Xarm into resonance. We start with the ALS signal, then as we approach resonance, the error signal is automatically transitioned to 1/sqrt(TRX), and ramped from there to POX, which we use as the PDH signal.
In the first plot, we have several spectra of the CESAR output signal (which is the error signal for the Xarm), at different arm resonance conditions. Dark blue is the signal when we are locked with the ALS beatnote, far from IR resonance. Gold is when we are starting to see IR resonance (arm buildup of about 0.03 or more), and we are using the 1/sqrt(TRX) signal for locking. Cyan is after we have achieved resonance, and are using only the POX PDH signal. Purple is the same condition as cyan, except that we have also engaged the low frequency boosts (FM 2, 3, 4) in the locking servo. FM4 is only usable once you are at IR resonance, and locked using the PDH signal. We see in the plot that our high frequency noise (and total RMS) decreases with each stage of CESAR (ALS, 1/sqrt(TR) and PDH).
To actually achieve the gold noise level of 1/sqrt(TR), we first had to increase the analog gain by swapping out a resistor on the whitening board.

The other plots attached are time series data. For the python plots (last 2), the error signals are calibrated to nanometers, but the dark blue, which is the transmitted power of the cavity, is left in normalized power units (where 1 is full IR resonance).
In the scan from off resonance to on resonance, around the 58 second mark, we see a glitch when we engage FM4, the strong low frequency boosts. Around the 75 second mark we turned off any contribution from 1/sqrt(TR), so the noise decreases once we are on pure PDH signal.
In the scan through the resonance, we see a little more clearly the glitch that happens when we switch from ALS to IR signals, around the 7 and 12 second marks.
We want to make some changes, so that the transition from ALS to IR signals is more smooth, and not a discrete switch.
|
Attachment 2: Screenshot-1.png
|
|
Attachment 3: ScanFromOffToOnResonance.pdf
|
|
Attachment 4: ScanThroughResonance.pdf
|
|
9718
|
Tue Mar 11 18:33:21 2014 |
Koji | Update | LSC | Composite Error Signal for ARms (6) |
Today we modified the CESAR block.
- Now the sign(X) function is in a block.
- We decided to use the linearization of the PDH.
- By adding the offset to the TR signal used for the switching between TR and PDH, we can force pure 1/sqrt(TR) or pure PDH to control the cavity. |
Attachment 1: 14.png
|
|
9717
|
Tue Mar 11 15:21:08 2014 |
Koji | Summary | LSC | Composite Error Signal for ARms (1) |
True. But we first want to realize this for a single arm, then move onto the two arms case.
At this point we'll need to incorporate frequency dependence. |
9716
|
Tue Mar 11 15:19:45 2014 |
Jenne | Update | Electronics | High gain Trans PD electronics change |
As part of our CESAR testing last night, we had a look at the noise of the 1/sqrt(TR) signal.
Looking at the time series data, while we were slowly sweeping through IR resonance (using the ALS), Rana noted that the linear range of the 1/sqrt(TR) signal was not as wide as it should be, and that this is likely because our SNR is really poor.
When a single arm is at a normalized transmission power of 1, we are getting about 300 ADC counts. We want this to be more like 3000 ADC counts, to be taking advantage of the full range of the ADC.
This means that we want to increase our analog gain by a factor of 10 for the low gain Thorlabs PDs.
Looking at the photos from November when I pulled out the Xend transmission whitening board (elog 9367), we want to change "Rgain" of the AD620 on the daughter board. While we're at it, we should also change the noisy black thick film resistors to the green thin film resistors in the signal path.
The daughter board is D04060, S/N 101. The main whitening board for the low gain trans QPD is D990399, RevB, S/N 104.
We should also check whether we're saturating somewhere in the whitening board by putting in a function generator signal via BNC cable into the input of the Thorlabs whitening path, and seeing where (in Dataviewer) we start to see saturation. Is it the full 32,000 counts, or somewhere lower, like 28,000?
Actually the gain was changed. From gain of 2 (Rgain = 49.4kOhm) to 20 (Rgain = 2.10kOhm), Corresponding calibration in CDS was also changed by locking the Xarm, running ASS, then setting the average arm power to be 1. Confirmed Xarm is locking. And now the signal is used for CESAR. We see emperically that the noise has improved by a factor of approximately 10ish. |
Attachment 1: IMG_1309.JPG
|
|
9715
|
Tue Mar 11 15:14:34 2014 |
den | Summary | LSC | Composite Error Signal for ARms (1) |
Quote: |
The composite error signal

|
Very nice error signal. Still, I think we need to take into account the frequency shape of the transfer function TR -> CARM. |
9714
|
Tue Mar 11 15:01:28 2014 |
ericq | Update | Computer Scripts / Programs | MC Signal Monitoring |
Two weeks ago (Feb 26) I took the "Q MON" output of the demodulator that sends its Q output to the MC servo board as the error signal, and connected it to an SR785, so we can occasionally monitor the error signal noise. (Also, I did not appropriately ELOG the fact I touched things...)
I'm working on an automated script to do the monitoring, but the wireless router that the SR785 is connected is wicked slow. I should run an ethernet cable to it...
I'm just figuring I'll look at the full span (~100kHz) spectrum every ten minutes, and compare it to some nominal spectrum from a known-good time, and the last few hours. |
9713
|
Tue Mar 11 14:49:01 2014 |
Koji | Summary | LSC | Important notice on the XARM servo |
The nominal gain of the XARM for the POX11 error signal is 0.03 (instead of previous 0.3)
This is due to my increase of the gain in FM4 by 20dB so that we can turn of FM4 without changing the UGF when it is at 150Hz.
The YARM servo was not yet touched. |
9712
|
Mon Mar 10 21:16:56 2014 |
Koji | Summary | LSC | Composite Error Signal for ARms (5) |
After making the CDS modification, CESAR was tested with ALS.
First of all, we run CESAR with threshold of 10. This means that the error signal always used ALS.
The ALS was scanned over the resonance. The plot of the scan can be found in EricQ's elog.
At each point of the scan, the arm stability is limited by the ALS.
Using this scan data, we could adjust the gains for the TR and PDH signals. Once the gains were adjusted
the threshold was lowered to 0.25. This activates dynamic signal blending.
ALS was stabilized with XARM FM1/2/3/5/6/7/9. The resonance was scanned. No glitch was observed.
This is some level of success already.
Next step was to fully hand off the control to PDH. But this was not successfull. Everytime the gain for the TR was
reduced to zero, the lock was lost. When the TR is removed from the control, the raw PDH signal is used fot the control
without normalization. Without turning on FM4, we lose 60dB of DC gain. Therefore the residual motion may have been
too big for the linear range of the PDH signal. This could be mitigated by the normalization of the PDH signal by the TR. |
9711
|
Mon Mar 10 21:16:13 2014 |
Koji | Summary | LSC | Composite Error Signal for ARms (4) |
The LSC model was modified for CESAR.
A block called ALSX_COMBINE was made in the LSC block. This block receives the signals for ALS (Phase Tracker output), TRX_SQRTINV, TRX, POX11 (Unnormalized POX11I).
It spits out the composite ALS signal.
Inside of the block we have several components:
1) a group of components for sign(x) function. We use the PDH signal to produce the sign for the transmission signal.
2) Hard triggering between ALS and TR/PDH signals. An epics channel "THRESH" is used to determine how much transmission
we should have to turn on the TR/PDH signals.
3) Blending of the TR and PDH. Currently we are using a confined TR between 0 and 1 using a saturation module. When the TR is 0, we use the 1/SQRT(TR) signal for the control,
When the TR is 1, we use the PDH signal for the control.
4) Finally the three processed signals are combined into a single signal by an adder.
It is important to make a consideration on the offsets. We want all of ALS, 1/SQRT(TR), and PDH to have zero crossing at the resonance.
ALS tends to have arbitorary offset. So we decided to use two offsets. One is before the CESAR block and in the ALS path.
The other is after the CESAR block. Right now we are using the XARM servo offset for the latter purpose.
We run the resonance search script to find the first offset. Once this is set, we never touch this offset until the lock is lost.
Then for the further scanning of the arm length, we uused the offset in the XARM servo filter module. |
Attachment 1: ss1.png
|
|
Attachment 2: ss2.png
|
|
Attachment 3: CESAR_OFFSETS.pdf
|
|
9710
|
Mon Mar 10 21:14:58 2014 |
ericq | Summary | LSC | Composite Error Signal for ARms (3) |
Using Koji's mathematica notebook, and Nic's python work, I set out to run a time domain simulation of the error signal, with band-limited white noise added in.

Basically, I sweep the displacement of the cavity (with no noise), and pass it to the analytical formulae with the coefficients Koji used, with some noise added in. I also included some 1/0 protection for the linearized PDH signal. I ran a sweep, and then compared it to an ALS sweep that Jenne ran on Monday; reconstructing what the CESAR signal would have looked like in the sweep.
The noise amounts were totally made up.
They matched up very well, qualitatively! [Since the real sweep was done by a (relatively) noisy ALS, the lower noise of the real pdh signal was obscured.]
 
Given this good match, we were motivated to start trying to implement it on Monday.
At this point, since we've gotten it working on the actual IFO, I don't plan on doing much more with this simulation right now, but it may come in handy in the future... |
9709
|
Mon Mar 10 21:13:43 2014 |
nicolas | Summary | LSC | Composite Error Signal for ARms (2) |
In order to better understand how the composite signal would behave in the presence of noise, I decided to do a simple analysis of the cavity signals while sweeping through resonance.
My noise model was to just assume that a given signal has some rms uncertainty (error bars) and use linear error propagation to propagate from simple signals to more complicated ones.
I used the python package uncertainties to do the error propagation.
I assumed that the ALS signal, the cavity transmission, and the cavity PDH error signal all have some constant noise that is independent of the cavity detuning. Below is a sweep through resonance (x axis is cavity detuning in units of radians).

The shaded region represents the error on each signal.
Next I calculated the 'first order' calculated error signals. These being a raw PDH, normalized PDH, an inverse square root trans, and the normal ALS again. I tuned the gains so they match appropriately.
Here, one can see how the error in the trans signal propagates to the normalized and trans signals and becomes large are the fractional error in the trans signal becomes large.

Next I did some optimization of linear combinations of these signals. I told the code to maximize the total signal to noise ratio, while ensuring that the overall signal had positive gain. I did this again as a function of the cavity detuning.
Each curve represents the optimized weight of the corresponding signal as a function of detuning.

So this is roughly doing what we expect, it prefers ALS far from the resonance, and PDH close to the resonance, while smoothly moving into square root trans in the middle.
It's a little fake, but it gives us an idea of what the 'best' we can do is.
Finally I used these weights to recombine the signals into a composite, to get an idea of the noise of the overall signal. At the same time, I plot the weighting proposed by Koji's mathematica notebook (using trans and 1-trans, and a hard switch to ALS).

So as one can see, at least for the noise levels I chose, the koji weighting is not much worse than the 'optimal' weighting. While it is much simpler.
The code for all this is in the svn at 40mSVN/nicolas/workspace/2014-03-06_compositeerror |
9708
|
Mon Mar 10 21:12:30 2014 |
Koji | Summary | LSC | Composite Error Signal for ARms (1) |
The ALS error (i.e. phase tracker output) is linear everywhere, but noisy.
The 1/sqrt(TR) is linear and less noisy but is not linear at around the resonance and has no sign.
The PDH signal is linear and further less noisy but the linear range is limited.
Why don't we combine all of these to produce a composite error signal that is linear everywhere and less-noisy at the redsonance?
This concept was confirmed by a simple mathematica calculation:
The following plot shows the raw signals with arbitorary normalizations
1) ALS: (Blue)
2) 1/SQRT(TR): (Purple)
3) PDH: (Yellow)
4) Transmission (Green)

The following plot shows the preprocessed signals for composition

1) ALS: no preprocess (Blue)
2) 1/SQRT(TR): multiply sign(PDH) (Purple)
3) PDH: linarization with the transmission (If TR<0.1, use 0.1 for the normalization). (Yellow)
4) Transmittion (Green)
The composite error signal

1) Use ALS at TR<0.03. Use 1/SQRT(TR)*sign(PDH)*(1-TR) + PDH*TR at TR>0.03
2) Transmittion (Purple)
|
Attachment 1: composite_linear_signal.nb.zip
|
9707
|
Mon Mar 10 12:49:27 2014 |
Jenne | Update | IOO | PMC input pointing misaligned |
I don't know why, but as you can see in Steve's plot from earlier this morning, the PMC transmission has been going down significantly all weekend. The PMC refl camera was very bright. I tweaked up the alignment (mostly pitch), and now we're back to normal.
The IMC was having trouble staying locked all morning, and I'm hoping that this PMC adjustment will help - the MC already looks better, although it's only been a few minutes. |
9706
|
Mon Mar 10 11:42:36 2014 |
Jenne | Update | CDS | fb timing was off |
fb timing was off again. |
9705
|
Mon Mar 10 09:28:48 2014 |
steve | Update | IOO | IOO pointing 2 days |
Morning seconds without adjustment.
|
Attachment 1: IOOpointing2d.png
|
|
Attachment 2: morningseconds.png
|
|
9704
|
Fri Mar 7 16:56:17 2014 |
steve | Update | IOO | IOO qpds centered |
Quote: |
Quote: |
The IMC has not been behaving well since this morning and totally not happy when Q was finishing his measurements. The WFS servo had large offsets in pitch. Looking back at the trend and using ezcaservo to restore the suspensions did not help.
I realigned the IMC and brought TRANS SUM to ~18000 and MCREFL to < 0.5. The spot positions are not very good; nearly 2 mm off in pitch on MC1 and MC3. But after the alignment of MC, the WFS servo offsets were below +/-20.
The MC has been locked stably with WFS servo ON for the last few hours.
P.S. I did not touch the WFS pointing or reset the WFS offsets.
|
MC remained locked with WFS enabled all through last night and this morning. Koji dropped by and looked at the MC. The MC WFS servo, though stable, was at the edge of becoming unstable. This was because I did not touch the WFS pointing on the QPDs yesterday after realigning. So I recentered the WFS, reset the WFS filterbank offsets and reenabled the servo.
I measured the spot positions on MC mirrors for reference.
Spot positions in mm (MC1,2,3 pit MC1,2,3 yaw): [1.405767579680834, 0.79369009503571208, 1.3220430681427462, -1.2937873599406551, -1.1704264340968924, -1.2518046122798692]
|
|
Attachment 1: IOOqpdCentered.png
|
|
9703
|
Fri Mar 7 16:13:03 2014 |
Steve | Update | VAC | pumping speed is recovered |
Quote: |
Valve configuration: Vacuum Normal is reached in really 4 days if we do not count overnight rest of roughing.
VA6 and VC2 are reconnected. I'm turning on the RGA next
All 4 ion pumps were vented with air and pumped down to ~ 1e-4 Torr
Ion pumps gate valve control cables are connected and their pumps are disconnected.
The black relay box was tested repeatedly and it stopped misbehaiving.
We were at atmosphere for 13 days. Chamber BS, ITMX, ITMY and ETMY were opened.
Al foil "cups" were placed on the back side OSEMs of PRM.
|
Pd 76 and 77 are compared at 30 days of pumping. We spend 13-14 days at atmosphere before each.
Pump down 76 was with leaky ion pump gate valve. The ion pumps are not in use for years so they accumulated some higher pressure PLUS the valve switching caos at computer reboot most likely
increased the ion pumps pressures to about 10-20 torr
I think one of the ion pump gate valve was not sealing well. This leak was holding back pump down speed at pd76 |
Attachment 1: pumpingSpeed.png
|
|
9702
|
Fri Mar 7 00:43:34 2014 |
manasa | Update | LSC | ALS C&D locked (on MC2 and ETMs) and IR resonance obtained |
[EricQ, Manasa]
ALS common locked by actuating on MC2 and ALS Differential locked by actuating on ETMX and ETMY (Stable lock acquired for over an hour). 
Common and Differential offsets were swept to obtain IR resonance in both the arms (arms stayed on resonance for over 15 minutes). 
Procedure:
1. Configured LSC settings to allow locking using ALS error signals.
2. Locked common and differential using ALS error signals
Aux matrix
ALSX ALSY
------------------------------
XARM 1 -1
YARM 1 1
-----------------------------
X arm servo settings:
FIlters: FM1, FM5, FM6, FM7, FM9
Gain = -8.0
Y arm servo settings:
Filters: FM1, FM5, FM6, FM7, FM9
Gain = +8.0
Out matrix
XARM YARM
------------------------
ETMX 1 0
ETMY 0 1
------------------------
3. Transitioned CARM control output to actuate on MC2 instead of ETMX
SUS-MC2_LSC servo gain = 1.0
The transition was done in very small steps : actuating on MC2 in -0.01 steps at the outmatrix upto -1.0 while reducing the ETMX actuation to 0 simultaneously.
DARM still stayed locked only with actuation on ETMY.
4. Transitioned DARM control to ETMX and ETMY.
Used ezcastep to step up DARM control (Y arm output) actuation on ETMX and step down the actuation on ETMY.
Final output matrix
Xarm Yarm
-----------------------
ETMX 0 -0.5
ETMY 0 0.5
MC2 -1.0 0
-----------------------
Noise plot in attachment.
5. Finding arm resonance
Used ezcastep to gradually build up offsets in CARM (LSC-XARM_OFS) to find IR resoance in one arm (Y arm).
Introducing a small (order of 0.5) DARM offset (LSC-YARM_OFS) shifted the Y arm off-resonance.
Used CARM offset to get back the Y arm to resonance.
Changing CARM and DARM offsets alternately while tracking the Y arm resonance got us to a point where we had both the arms resonating for IR.
6. At this point the MC decided to give up and we lost lock. 
Notes:
1. We found that the WFS2 YAW output filterbank had the output switched OFF (probably accidentally by one of us). This was reenabled. Please be careful while manually turning ON and OFF the MC WFS servos. |
Attachment 1: ALS_MC2CARM.pdf
|
|
9701
|
Thu Mar 6 19:17:05 2014 |
manasa | Update | IOO | MC calmed down |
Quote: |
The IMC has not been behaving well since this morning and totally not happy when Q was finishing his measurements. The WFS servo had large offsets in pitch. Looking back at the trend and using ezcaservo to restore the suspensions did not help.
I realigned the IMC and brought TRANS SUM to ~18000 and MCREFL to < 0.5. The spot positions are not very good; nearly 2 mm off in pitch on MC1 and MC3. But after the alignment of MC, the WFS servo offsets were below +/-20.
The MC has been locked stably with WFS servo ON for the last few hours.
P.S. I did not touch the WFS pointing or reset the WFS offsets.
|
MC remained locked with WFS enabled all through last night and this morning. Koji dropped by and looked at the MC. The MC WFS servo, though stable, was at the edge of becoming unstable. This was because I did not touch the WFS pointing on the QPDs yesterday after realigning. So I recentered the WFS, reset the WFS filterbank offsets and reenabled the servo.
I measured the spot positions on MC mirrors for reference.
Spot positions in mm (MC1,2,3 pit MC1,2,3 yaw): [1.405767579680834, 0.79369009503571208, 1.3220430681427462, -1.2937873599406551, -1.1704264340968924, -1.2518046122798692] |
Attachment 1: MC_spots_Mar6.png
|
|
9700
|
Thu Mar 6 17:34:03 2014 |
rana | Update | SUS | Oplev Tuning - Cartoon cost function |
Quote: |

|
In addition, we have to make sure to not let the suspension DACs saturate and make sure that the impulse response time of the OL servo is short; otherwise the lock acquisition kicks or bumps can make it wiggle for too long. |
9699
|
Thu Mar 6 11:43:17 2014 |
steve | Update | General | Guralp control unit |
CMG-40T handheld controller unit is missing its power supply. In order to zero the instrument one has to apply plus and minus DC voltage.
The wiring on this 10 pin Amphenol PT02E-12-10P is shown. |
Attachment 1: GuralpCU.jpg
|
|
Attachment 2: BAH.jpg
|
|