Sat Sep 13 23:13:01 2008 |
Alberto | Update | General | abs cavity length experiment |
The Y arm was locked all time today but, suddenly, this afternoon it lost lock and since then I've been unable to restore it. I tried unsuccessfully the Restore and the Align scripts several times, although the position of PZT steering mirrors were good (as in the snapshot). I tried things like unlocking/locking the MC, the FSS reference cavity, the PMC but it didn't work. Then I decided to switch to the X arm. Locking it was a piece of cake compare to Y. I'm going to start measuring the FSR of the X arm. |
944
|
Fri Sep 12 11:09:20 2008 |
Alberto | Update | General | abs cavity length experiment |
I'm leaving the lab for a couple of hours. I shut the NPRO. The interferometer is available to anyone. |
943
|
Thu Sep 11 23:28:35 2008 |
alberto | Update | General | abs cavity length experiment |
The MC lost lock for some reason not related to either the FSS or the PMC I'm done with my measurement for tonight. I've shut the NPRO beam before leaving. |
942
|
Thu Sep 11 14:27:53 2008 |
steve | Bureaucracy | SAFETY | safety training |
Peter Kalmus received 40m specific safety training last week. |
941
|
Thu Sep 11 11:29:14 2008 |
josephb | Configuration | Computers | Final netgear switch in place in 1Y2 |
I've placed the final (of 4) Netgear prosafe 24 port switch at the very top of 1Y2. At that location, there are no holes left to screw into, so it has 4 rubber feet and is sitting on the top most signal generator. It has been plugged in and connected to the control room hub with a labeled cat6 ethernet cable.
Its IP address has been set to 131.215.113.253, and has the usual controls password if using the "Smart Wizard Discovery Tool" which comes on the Netgear CD. The CD can be found in the Equipment manuals filing cabinet under Netgear. This program unfortunately only runs on a window PC.
To Do: Fix the C1:ASC ethernet connection which is currently coming straight out the front door and connected to the 1X4 switch (again through the front door). |
940
|
Wed Sep 10 19:53:53 2008 |
Alberto | Update | General | abs length experiment |
Update of the last days work on the experiment to measure the absolute length of the cavities.
I'm trying to repeat the same measurement that Koji did on the Y arm, before switching to the X arm.
I switched to the PHD universal box for the PLL control between the main laser and the secondary laser. I found a good gain value for the servo and now I can set the frequency of the beat to any value as long as I do it slowly turning the LO frequency from the knob on the Marconi.
I laid down a 50m BNC cable from the Y end to near the BS chamber, where all the abs length equipment is. I matched the two laser beams changing the alignment of the injection steering mirror at the the dark port on the AP table. I then locked the Y arm cavity. When I first tried to do that, the locking script didn't work because the beam was off of the 'sweet spot' where Rob had set it on Monday. It turned out that aborting the script during one of its previous run, had changed the alignment of the PZT steering mirrors. So with Rob I brought them back near the positions as in the snapshot and then saved a new one with the latest values.
Eventually I could set the beat frequency to the FSR of the arm cavity and saw it in transmission at the ETMY.
Now I'm working on the LabView interface for the GPIB data acquisition board. |
939
|
Wed Sep 10 13:28:25 2008 |
Yoichi | Summary | Electronics | IOO rack lost -24V (recovered) |
Alberto, Yoichi
This morning, the MC suddenly started to be unwilling to lock.
I found a large offset in the MC servo board.
It turned out that -24V was not supplied to the Eurocard crate of the IOO rack.
We found two loose cables (violet, that means -24V) around the cross connects with fuses.
We connected them back, and the -24V was back.
The MC locks fine now, and Alberto can continue his arm length experiment. |
938
|
Wed Sep 10 08:57:03 2008 |
steve | Update | General | etmy illuminator turned off |
The ETMY illuminator was left on yesterday.
I just turned it off. |
937
|
Mon Sep 8 15:38:57 2008 |
Yoichi | Configuration | PSL | POY RF amp is back to its original task |
I temporarily fixed the busted ZHL-32A RF amplifier's power connector by simply soldering a cable to the internal circuit and pulling the cable out of the box through a hole for the power connector.
So I released the POY RF amplifier from the temporary duty of serving the FSS RF distribution and put it back to the original task,
so that Rob can finally re-start working on the lock acquisition.
Now the temporarily fixed ZHL-32A is sitting next to the IOO rack along with the power supply and a Stanford signal generator.
Please be careful not to topple over the setup when you work around there. They will be there until Peter's Wentzel RF box arrives. |
936
|
Mon Sep 8 13:47:35 2008 |
steve | Update | PEM | thermostate setting changed |
Quote: |
Quote: | Some one changed the thermostat (old control room ) setting behind 1Y6 from 73 to 79F
It should be in the elog. |
In fact, it is. I demand satisfaction for the injury to my elogging reputation! |
Thermostate setting was changed from 79F to 77F behind 1Y6 |
935
|
Mon Sep 8 10:57:49 2008 |
steve | Update | IOO | the psl and mc are back to normal |
The alarm handler is silent this morning.
This is almost unbelievably pleasant after two mount of harassment.
The MC did not lose lock for three days.
Atm1: the new fss layout
Atm2: PMC with lead brick
Atm3: 3 days plot |
Attachment 1: fss.png
|
|
Attachment 2: pmcbrick.png
|
|
Attachment 3: brick.jpg
|
|
934
|
Fri Sep 5 15:09:50 2008 |
rana | Update | PEM | thermostate setting changed |
Quote: | Some one changed the thermostat (old control room ) setting behind 1Y6 from 73 to 79F
It should be in the elog. |
In fact, it is. I demand satisfaction for the injury to my elogging reputation! |
933
|
Fri Sep 5 10:36:34 2008 |
steve | Update | PEM | thermostate setting changed |
Some one changed the thermostat (old control room ) setting behind 1Y6 from 73 to 79F
It should be in the elog.
The temp changed from freezing 20 to sunny 25 C |
932
|
Fri Sep 5 09:56:14 2008 |
josephb, Eric | Configuration | Computers | Funny channels, reboots, and ethernet connections |
1) Apparently the I00-ICS type channels had gotten into a funny state last night, where they were showing just noise, exactly when Rana changed the accelerometer gains and did major reboots. A power cycle of the c1ioo crate and appropriate restarts fixed this.
2) c1asc looks like it was down all night. When I walked out to look at the terminal, it claimed to be unable to read the input file from the command line I had entered the previous night ( < /cvs/cds/caltech/target/c1asc/startup.cmd). In addition we were unable to telnet in, suggesting an ethernet breakdown and inability to mount the appropriate files. So we have temporarily run a new cat6 cable from the c1asc board to the ITMX prosafe switch (since there's a nice knee high cable tray right there). One last power cycle and we were able to telnet in and get it running. |
931
|
Fri Sep 5 08:34:03 2008 |
steve | Update | PSL | MZ locked |
The MC is happy.
The MZ can be locked if you move the slider by hand. |
Attachment 1: mzhv.jpg
|
|
930
|
Thu Sep 4 18:02:34 2008 |
rana, josephb | Configuration | PEM | Accelerometer gains increased by 10 |
We increased the Accelerometer gains by 10 by modifying the C1ADCU_PEM.ini file.
[C1:PEM-ACC_MC1_X]
chnnum=15014
gain = 10
etc.
The plot shows the before and after for one channel. The ADC noise floor is ~10^-2 counts/rHz in this plot so now
we can do much better noise subtraction. |
Attachment 1: acc.png
|
|
929
|
Thu Sep 4 17:44:27 2008 |
Yoichi | Update | PSL | FSS error signal spectrum |
Attached is a spectrum of the FSS error signal.
There are a lot of sharp peaks above 100kHz (the UGF of the servo is about 200kHz).
These are mostly harmonics of 77kHz. They are the current suspects of the FSS slew rate saturation.
I remember when I blocked the light to the PD, these peak went away. So these noises must be
in the light. But I checked it a few weeks ago. So I will re-check it later.
One possible source of the lines is a DC-DC converter in the NPRO near the crystal.
We will try to move the converter outside of the box. |
Attachment 1: FSS-Error-Spe.png
|
|
928
|
Thu Sep 4 17:17:03 2008 |
Yoichi | Update | IOO | MC open loop TF |
I measured open loop transfer functions of the MC servo.
The UGF was about 30kHz. Since there was some gain margin at higher frequencies, I increased
the input gain of the MC servo board from 19dB to 22dB. Now the UGF is 40kHz and we have more
phase margin (~30deg). |
Attachment 1: MC-OPLTF.png
|
|
927
|
Thu Sep 4 17:12:57 2008 |
Yoichi | Update | PSL | FSS open loop TF |
I changed the gain settings of the FSS servo.
Now the Common Gain is 5dB (the last night it was 2dB) and the Fast Gain is 12dB (formerly 16dB).
I measured the open loop TF with this setting (the attachment).
I also plotted the OPLTF when CG=2dB, FG=20.5dB. With this setting, the MC looses lock every 30min.
You can see that the OPLTF is smoother with FG=12dB.
When the FG is high, you can see some structure around 250kHz. This structure is reproducible.
This may be some interruption from the fast path to the PC path through a spurious coupling. |
Attachment 1: FSS-OPLTFs.png
|
|
926
|
Thu Sep 4 17:03:25 2008 |
Yoichi | Update | PSL | RF oscillator noise comparison |
I measured current spectra of the RF signal going to the FSS EOM.
The attachment compares the spectra between a Stanford signal generator and a Marconi.
I borrowed the Marconi from the abs. length measurement experiment temporarily.
The measurement was done using the signal going to the EOM. That means the spectra include
noise contributions from the RF amp., splitter and cables.
21.5MHz peak was not included because that would overload the ADC and I would have to use a large attenuation.
This means the measurement would be totally limited by ADC noise everywhere except for 21.5MHz.
I noticed that with the Marconi, the FSS is a little bit happier, i.e. the PC path is less loaded
(0.9Vrms with Stanford vs. 0.7Vrms with Marconi). But the difference is small.
Probably the contribution from the 77kHz harmonics in the laser light is more significant (see entry #929).
Also the peaks in the Stanford spectrum are not harmonics of 77kHz, which we see in the FSS error signal.
I returned the Marconi after the measurement to let Alberto work on the abs. length measurement. |
Attachment 1: RFSpectra.png
|
|
925
|
Thu Sep 4 16:24:56 2008 |
rana | Configuration | Computers | Attempt to increase gain for C1:PSL-ISS_INMONPD_F via 110B |
Quote: | We were attempting to increase the gain on the channel C1:PSL-ISS_INMONPD_F in preparation to do a scan of the PMC at very low input power.
|
According to the Wikipedia, certain esoteric mathematical
operations lead to the result that 4000 x 10 > 32768. |
924
|
Thu Sep 4 14:43:58 2008 |
Jenne | Update | PSL | PMC Open Loop Gain |
I have measured the PMC's open loop gain. UGF is 629.7Hz, with a phase margin of 53 degrees.
I injected into FP2 on the front panel, and measured MixOut/Source from 100Hz to 100kHz using the SR785. I did this both when the loop was open, and when the loop was closed (open the loop by enabling FP1, which breaks the loop).
We have 2 transfer functions involved: The actual open loop gain of the PMC servo loop (G1), and the gain between FP2 and the MixerOut monitor point (G2). This gives us:
TF(closed loop) = G2*(1+G1)
TF(broken loop) = G2
G1 = TF(closed)/TF(broken) - 1
This G1 is the final open loop gain, and it is plotted below. |
Attachment 1: OpenLoopTF04Sept2008.png
|
|
923
|
Thu Sep 4 13:48:50 2008 |
Yoichi | Update | PSL | FSS modulation depth |
I scanned the reference cavity with the NPRO temperature (see the attached plot).
The power ratio between the carrier and the sideband resonances is about 26.8.
It corresponds to gamma=0.38.
The RF power fed into the EOM is now 14.75dBm (i.e. 1.7V amplitude). The NewFocus catalog says 0.1-0.3rad/V. So
gamma=0.38 is a reasonable number.
|
Attachment 1: RCScan.png
|
|
922
|
Thu Sep 4 11:33:25 2008 |
josephb, Eric, Jenne | Configuration | Computers | Attempt to increase gain for C1:PSL-ISS_INMONPD_F via 110B |
We were attempting to increase the gain on the channel C1:PSL-ISS_INMONPD_F in preparation to do a scan of the PMC at very low input power.
We started by adding a line to the C1:IOOF.ini file in /cvs/cds/caltech/chans/daq/ under that channel that said "gain=10.0". Before touching anything, the channel was outputting around 4000 counts.
We hit the reconfig button for c1iovme16k, then rebooted c1iovme (which turned out to do nothing) and then the framebuilder, in a method consistent with the wiki. This turned out to put the channel in an odd state, where it was showing very rapid, random spikes, virtually but still around 4000ish counts. We returned the file back to its original format, hit reconfig, and then rebooted the framebuilder. The channel however, was still behaving in the same broken way.
After poking around the PSL table, looking at some direct outputs, we came back and rebooted c1iovme and the framebuilder again, which fixed the channel, such that it was reading out correctly. Taking this as a sign that maybe we should reboot the framebuilder, then c1iovme to get the channel to load changes, we changed the file again to have "gain=10.0". Upon reboot of the framebuilder, the channel was still reading out fine, but at the same level. So we continued with the reboot of c1iovme. This still had no effect on the channel output.
The ini file has been set back at this point, however since Yoichi is working, I'm holding off doing a reconfig and reboot on the framebuilder until later. |
921
|
Thu Sep 4 10:13:48 2008 |
Jenne | Update | IOO | We unlocked the MC temporarily |
[Joe, Eric, Jenne]
While trying to diagnose some DAQ/PD problems (look for Joe and Eric's entry later), we unlocked the PMC, which caused (of course) the MC to unlock. So if you're looking back in the data, the unlock at ~10:08am is caused by us, not whatever problems may have been going on with the FSS. It is now locked again, and looking good. |
920
|
Thu Sep 4 07:46:10 2008 |
Yoichi | Update | IOO | MC is now happy |
The MC has been locked for more than 12 hours continuously now !
Changes I made yesterday were:
(1) Removed the 20dB attenuator before the EOM.
(2) Reduced the Fast Gain from 21dB to 16dB, which made the PC to be a little bit more loaded (~0.6Vrms).
As Rana pointed out in the meeting, setting the Fast Gain a bit lower may have put the FSS in a stabler state. |
Attachment 1: MC-lock.png
|
|
919
|
Thu Sep 4 07:29:52 2008 |
Yoichi | Update | PSL | c1iovme power cycled |
Quote: | Entry 663 has a plot of this using the PSL/FSS/SLOWscan script. It shows that the SB's were ~8x smaller than the carrier.
P_carrier J_0(Gamma)^2
--------- = ------------
P_SB J_1(Gamma)^2
Which I guess we have to solve numerically for large Gamma? |
P_carrier/P_SB = 8 yields gamma=0.67. |
918
|
Thu Sep 4 00:38:14 2008 |
rana | Update | PSL | c1iovme power cycled |
Entry 663 has a plot of this using the PSL/FSS/SLOWscan script. It shows that the SB's were ~8x smaller than the carrier.
P_carrier J_0(Gamma)^2
--------- = ------------
P_SB J_1(Gamma)^2
Which I guess we have to solve numerically for large Gamma? |
917
|
Wed Sep 3 19:09:56 2008 |
Yoichi | DAQ | Computers | c1iovme power cycled |
When I tried to measure the sideband power of the FSS using the scan of the reference cavity, I noticed that the RC trans. PD signal was not
properly recorded by the frame builder.
Joe restarted c1iovme software wise. The medm screen said c1iovme is running fine, and actually some values were recorded by the FB.
Nonetheless, I couldn't see flashes of the RC when I scanned the laser frequency.
I ended up power cycling the c1iovme and run the restart script again. Now the signals recorded by c1iovme look fine.
Probably, the DAQ boards were not properly initialized only by the software reset.
I will re-try the sideband measurement tomorrow morning. |
916
|
Wed Sep 3 18:45:01 2008 |
Alberto | Configuration | | PD3 gain |
Alberto, Yoichi,
We found that the PD3 servo was unstable with a gain of 1, so we switched it to 0.5 |
915
|
Wed Sep 3 18:43:19 2008 |
Yoichi | Configuration | Electronics | Two more active probes |
I found two active probes, an HP41800A and a Tektronix P6201.
Thanks John for telling me you saw them before.
Now we have three active probes, wow !
We have to find or make a power supply for P6201.
The manual of the probe is attached. |
Attachment 1: Tektronix-P6201-active-probe.pdf
|
|
914
|
Wed Sep 3 12:26:49 2008 |
Eric | Summary | Cameras | Weekly Summary |
Finished up simulating the end mirror error in order to test the whether the fitting code still provides reasonable answers despite the noise caused by the defects on the end mirror. The model I used to simulate the defects is far from perfect, but its good enough given the time I have remaining, and I have no reason to believe the differences between it and the real noise would cause any radical changes in how the fit operates. A comparison between a modeled image and a real image is attached. Average error (difference between the estimated value and the real value) for each of the parameters is
For the fit:
Max Intensity: 2767.4 (Max intensities ranged from 8000 to 11000)
X-Position: 0.9401 pixels
X Beam Waist: 1.3406 pixels (beam waists ranged from 35 to 45)
Y-Position: 0.9997 pixels
Y Beam Waist: 1.3059 pixels (beam waists ranged from 35 to 45)
Intensity Offset: 12.7705 (Offsets ranged from 1000 to 4000)
For the center of mass calculation (with a threshold that cut off everything above 13000)
X-Position: 0.0087 pixels
Y-Position: 0.0286 pixels
Thus, the fit is generally trustworthy for all parameters except for maximum intensity, for which it is very inaccurate. Additionally, this shows that the center of mass calculation actually does a much better job than the fit when this much noise is in the image. For the end mirrors, the fit is really only useful for finding beam waist, and even this is not extremely accurate (~3% error). All the parameters for the modeling is on the svn in /trunk/docs/emintun/MatLabFiles/EndMirrorErrorSimulation.txt.
Finished working on the calculations that convert a beam misalignment as measured as a change in the beam position on the two mirrors to a power loss in the cavity. Joe calculated the minimum measurable change in beam position to be around a tenth of a pixel, which corresponds to half a micron when the beam is directly incident on the camera. This gives the ability to measure fractional power losses as low as 2*10^-10 for the 40m main arm cavities. To me, this seems unusually low, though it scales with beam position squared, so if anything else limited the ability to measure changes in the beam position, it would have a large effect on the sensitivity to power losses. Additionally, it scales inversely with length, so shorter cavities provide less sensitivity.
This morning Joe and I tested the ability for the camera code to servo the ITMX in order to change the beam's position on the ETMX. Two major things have been changed since the last time we tried this. First, the calculated beam center that gets output to the EPICS channels now first goes through a transform that converts it from pixels into physical units, and should account for the oblique angle of the camera. The output to the EPICS channels should now be in the form of 'mm from the center of the optic', although this is not very precise at the moment. The second thing that was changed was that the servo was run with a modified servo script that included options to set a minimum, maximum, and slew rate in order to protect the mirrors from being swung around too much. The servo was generally successful: for a given x-position, it was capable of changing the yaw of ITMX so that the position seen on the camera moved to this new location. The biggest problem is that the x and y dimensions do not appear to be decoupled (the transform converting it to physical units should have done this), so that modifying the yaw of the mirror changed both the x and y positions (the y about half as much) as output by the camera. This could cause a problem when trying to servo in both dimensions at once, since one servo could end up opposing the other. I don't know the cause of this problem yet, since the transform that is currently in use appears to be correctly orienting the image. |
Attachment 1: SimulatedErrorComparison.png
|
|
913
|
Tue Sep 2 22:43:16 2008 |
Yoichi | Configuration | PSL | Updated FSS open loop TF |
Since the LO level of the FSS servo was too low, I replaced the RF oscillator board with a combination of
a Stanford signal generator and an RF amplifier.
Right now, the POY RF amplifier is used for this purpose temporarily.
Now the LO level is about 16dBm. The RF power going into the EOM is attenuated by 20dB from the LO level.
I played with the cable length to get the phase right.
Then I was able to lock the FSS with the new RF signal source.
Attached is the open loop transfer function of the current FSS. Now the UGF is a bit above 200kHz, a factor of 2 improvement.
This gain was achieved with the common gain slider at 13.5dB and the fast gain = 30dB.
With the old RF oscillator board, UGF=100kHz was achieved with the common gain =30dB. Therefore, the increase of the LO gave
us a large signal gain.
Increasing the gain further, again ,makes the PC path crazy.
Rich suggested that this craziness was caused either by the slew rate limit of the PA85 or the output voltage limit of the bypass Op-amp(A829)
is hit.
TO DO:
* Look at the error signal spectrum to see if there is any signal causing the slew rate saturation at high frequencies.
* Find out what the RF signal level for the EOM should be. 20dB attenuation is an arbitrary choice.
* Find out the cross over frequency. Determine where the fast gain slider should be.
etc ... |
Attachment 1: OPLTF.png
|
|
912
|
Tue Sep 2 14:28:41 2008 |
Yoichi | Update | PSL | FSS EOM driving signal spectra |
Rich advised me to change the +10V input of the FSS crystal frequency reference board from whatever voltage supply we use now to a nice one.
This voltage is directory connected to the signal lines of both LO and RF output amps. Therefore, fluctuations in the voltage directly appear
in the outputs, though DC components are cut off by the AC coupling capacitors.
I changed the source of this voltage from the existing Sorensen one to a power supply sitting next to the rack.
The attached plots shows the difference of the RF output spectra between the two 10V sources.
The low frequency crap is almost gone in the new 10V spectrum.
I tried to increase the FSS gain with the new 10V, but still it goes crazy. I suspect it is because the LO power is too low. |
Attachment 1: RFDrive1.png
|
|
Attachment 2: RFDrive2.png
|
|
911
|
Tue Sep 2 10:09:03 2008 |
steve | Update | PSL | head temp is cooling down |
The chiller was over flowing this morning.
800 cc of water was removed.
PSL-126MOPA_HTEMP peaked at 20.7 C (normal is 18.7 C) |
910
|
Tue Sep 2 09:58:42 2008 |
Yoichi | Configuration | PSL | FSS on an auxiliary loop |
Quote: | Summary: The FSS is now temporarily disabled. Naturally, the MC won't lock. I will fix it tomorrow morning.
|
Now I removed the power splitters for the aux. reference cavity servo. The FSS is back and the MC locks.
I'm now returning one of the active high-impedance probes to the Wilson house. They need it today.
We are left with only one active probe. If anyone finds another active probe in the 40m lab.,
please let me know (according to Rana we should have one more). |
909
|
Tue Sep 2 07:58:34 2008 |
rana | Summary | PSL | FSS & PMC LO trends for 2 years |
The attached plot is a 2 year minute trend of the EPICS readback of the PMC & FSS LO Monitors (FSS_LODET & PMC_LODET).
Clearly the FSS LO has been dying for at least 2 years. The step up from 10 months
ago is probably when Rob removed a 3dB attenuator from in front of the box. |
Attachment 1: psl-lo-trend.png
|
|
908
|
Mon Sep 1 19:23:17 2008 |
Yoichi | Configuration | PSL | FSS on an auxiliary loop |
Summary: The FSS is now temporarily disabled. Naturally, the MC won't lock. I will fix it tomorrow morning.
Today, I did the 4th item of my TO DO list.
Using a mini-circuit mixer and two SR560s, I constructed an auxiliary servo loop for the reference cavity.
With this loop, I was able to lock the reference cavity without using the FSS box.
By locking the reference cavity with this auxiliary servo, I was able to measure the PC path transfer function.
I will post the analyzed results later.
I borrowed the PD RF and the LO signals from the main FSS loop by power splitters. Therefore, the gain of the main FSS loop
is now about 3dB low. I tried to compensate it by increasing the EOM modulation depth, but the PC path is still a bit noisy.
Probably the already too low LO power is now seriously low (the LO power cannot be changed from EPICS).
Because I did not want to leave the PC path with large output overnight (it will heat up the PA85, and might cause damage, though unlikely),
I disabled the FSS for now.
|
907
|
Mon Sep 1 04:34:00 2008 |
rana | Update | PSL | FSS loop transfer functions |
I started from 6th item in Yoichi's todo list.
1) Increased the setpoint of the thermostat next to the framebuilder from 73F to 79F. Its freezing over there
in the room with the drill press. Steve's illegal mercury thermometer is reading 19 C.
2) Looked the RFPD's output spectrum using the 20 dB coupled output from the coupler that's in-line.
The first attached PDF file (n.pdf) has several plots:
page 1: 0-500 MHz anomolous peaks at 138 & 181 MHz but nothing too crazy
page 2: 0-100 MHz 80 MHz peak is RF pickup from the VCO Driver - not on the light
page 3: 10-30 MHz totally nuts
page 4: 18-25 MHz that's just wrong
The RF spectrum should only have some action around 21.5 MHz and a little peak at 2x 21.5 MHz. All that extra
junk means that something is broken!
3) To see if I could rid of any of the 80 MHz signal or any of that other trash from 18-25 MHz, I wound the RF cable
around a large toroidal ferrite core. This should have given us many uH of inductance for any signals common to
both the center and shield of the cable with no effect on the differential RF signals. There was no effect.
4) Next went to look at the 21.5 MHz Crystal Oscillator Reference card (D980353...I bet you can't figure out how
this one works). These have the Mini-Circuits SMA 30 MHz low pass (SLP-30) filters on both the LO and EOM outputs.
FSSLO.PNG shows the waveform after 20 dB attenuation going into a scope terminated with 50 Ohms.
FSSLO-Spec.png shows the spectrum of this signal - its pretty distorted. Here's the levels
f (MHz) | before filter (dBm) | after filter (dBm)
---------------------------------------------------
21.5 | -12.8 -13.1
43 -24 -46
64.5 -50 < -80
86 -64 < -80
This would be OK after the filter, but the level is very low. Only 7dBm (accounting for my 20 dB att) !!
The FSS uses a JMS-1H mixer which needs, as everyone knows, a +17 dBm LO signal. Que lastima.
There seems to be something wrong already, but wait...
5) PC25.PNG shows the output signal going to the EOM from 0 - 25 MHz. The step that's visible there at
around 10 MHz is just something inherent to the analyzer (??). But see all that crap there down below
5 MHz ? That is NOT supposed to be there.
pc.pdf shows on the first page the comparison in EOM drive with 2 different slider values on the
RF AM adjust screen for the FSS. But page 2 is the punchline of this long entry: There is a bunch of
excess junk on the drive signal going to the FSS's phase modulator. The FSS is then trying to handle
this extra frequency noise and getting into trouble.
We have to fix this board. I have also ordered a few SBP-21.4 from mini-circuits (SMA bandpass around 21.4 MHz)
just in case. Another option is to just replace this thing with a Marconi and an RF amp.
|
Attachment 1: n.pdf
|
|
Attachment 2: FSSLO.PNG
|
|
Attachment 3: FSSLO-Spec.png
|
|
Attachment 4: PC25.png
|
|
Attachment 5: pc.pdf
|
|
906
|
Sat Aug 30 13:28:01 2008 |
rana | Configuration | PSL | PMC: List of changes |
This is a list of changes made to the PMC board while we had it out for modifying the notch:
- LC-LC 4th order low pass filter
- Replace the AD797 (U2) with an OP27. AD797's are bad - do not use them anywhere for any reason. The OP27 is slower and has a 3x worse input noise but doesn't compromise the bandwidth or noise performance of the PMC by any significant amount. The rule is: use OP27 everywhere unless you have a very good reason why not.
- There is no 'H1' jumper on board. R9 is 90.9 Ohms and R2=900 Ohms so that the U2 stage has a gain of 10.
- Cut a trace and inserted a 500 Ohm resistor between U2-pin6 and U5A-pin2 (the AD602). The AD602 has a 100 Ohm input impedance which cannot be driven without limiting by the AD797 or the OP27. The 500 Ohm resistor makes it a driveable load for low level signals which is all that should be there since its the error point of the servo. it also becomes a 6:1 voltage divider. Since the AD602 has a fixed output voltage noise of 100 nV/rHz, this will limit the noise performance if the VGA gain is less than 20 dB, but whatever.
- R11 7.87k -> 1.74k, R12 = 78.7k -> 700k. This increases the high frequency gain of that stage by 7.87/1.74 = 4.5 and lowers the low frequency pole from 2 to 0.2 Hz to give the PMC some more staying power at DC. The loop shape is now 1/f^2 in the 9-480 Hz band and so the phase dips enough to make it almost conditionally stable, but not quite.
- C26 changed from ??? + a 30 pF trim cap into a fixed NNN pF cap to set the notch frequency for the 14.5 kHz body mode that we measured. Once our brick configuration is more settled we can increase the Q of this notch from small to big.
- Grounded pin 5 of U14 & U15 (AD620). These have sometimes been used as "differential" drivers in LIGO by connecting this reference voltage pin to the remote ground of the next board. This has always lead to insidious oscillation and noise. This beauty also has an output noise of 100 nV/rHz. Just never use this chip if you can help it; we can make true differential drivers - we have the technology.
Of course, we didn't have a current version of a schematic sitting around so I printed out a Rev E schematic and marked it up with red pen. I'll post pictures later and put the schematic into the PSL schematics notebook. Would be useful to take the old schematic and update it in Acrobat so that we have something electronic. |
905
|
Fri Aug 29 22:57:48 2008 |
Yoichi | Update | PSL | FSS loop transfer functions |
I've been measuring a bunch of transfer functions of the FSS related stuffs.
There are a lot to be analyzed yet, but here I put one mystery I'm having now.
Maybe I'm missing something stupid, so your suggestions are welcome.
Here is a conceptual diagram of the FSS control board
TP3 TP4
^ ^
| |
RF PD -->--[Mixer]-----[Sum Amp]------>--[Common Gain]--->----[Fast Gain]----[Filter]--> NPRO PZT
^ | ^ | |
| V | V |
LO ---->------- TP1 IN TP2 -->---[Filter]--[High Volt. Amp.] --> Phase Corrector
What I did was first to measure a "normal" openloop transfer function of the FSS servo.
The FSS was operated in the normal gain settings, and a signal was injected from "IN" port.
The open loop gain was measured by TP1/TP2.
Now, I disconnected the BNC cable going to the phase corrector to disable the PC path and locked the ref. cav.
only using the PZT. This was done by reducing the "Common Gain" and "Fast Gain" by some 80dB.
Then I measured the open loop gain of this configuration. The UGF in this case was about 10kHz.
I also measured the gain difference between the "normal" and "PZT only" configurations by injecting
a signal from "IN" and measuring TP3/TP2 and TP4/TP3 with both configurations (The signal from the Mixer was
disconnected in this measurement).
The first attachment shows the normal open loop gain (purple) and the PZT only open loop gain scaled by the
gain difference (about 80dB). The scaled PZT open loop gain should represent the open loop gain of the PZT
path in the normal configuration. So I expected that, at low frequencies, the scaled PZT loop TF overlaps the normal
open loop TF.
However, it is actually much larger than the normal open loop gain.
When I scale the PZT only TF by -30dB, it looks like the attachment #2.
The PZT loop gain and the total open loop gain match nicely between 20kHz and 70kHz.
Closer look will show you that small structures (e.g. around 30kHz and 200kHz) of the two
TFs also overlap very well. I repeated measurements many times and those small structures are always there (the phase is
also consistently the same). So these are not random noise.
I don't know where this 30dB discrepancy comes from. Is it the PC path eating the PZT gain ?
I have measured many other TFs. I'm analyzing these.
Here is the TO DO list:
* Cavity response plot from AOM excitation measurements.
* Cavity optical gain plot.
* Reconstruct the open loop gain from the electric gain measurements and the optical gain above.
* Using a mixer and SR560(s), make a separate feedback circuit for the PZT lock. Then use the PC path
to measure the PC path response.
* See the response of the FSS board to large impulse/step inputs to find the cause of the PC path craziness.
etc ... |
Attachment 1: OPLTFs.pdf
|
|
Attachment 2: OPLTFsScaled.pdf
|
|
904
|
Fri Aug 29 18:24:48 2008 |
rana | HowTo | PSL | PMC: PZT Calibration |
I calibrated the PMC PZT at DC by using 'trianglewave' to drive the DC offset slider
and reading back PMC_PZT and PMC_TRANSPD_F (both are DC coupled DAQ channels).
The attached PDF illustrates the method: look at the voltage required to span 1 FSR and then divide.
PMC_cal (m/V) = (1064 nm)/2 / V_FSR The calibration for our PZT is therefore 10.4 nm/V.
The full scale (0-300 V) range is 3.1 microns.
From Jenne's elog entry we know that the series resistor to the PZT is 63.6 kOhms. The PZT is labeled as
having a capacitance of 279 nF. So the PMC drive's pole frequency is 1/2/pi/63.6e3/279e-9 = 9 Hz +/- 0.5 Hz.
The cable capacitance is ~20 pF/foot so its not significant for this.
The template file is Templates/PMC-PZTcal.xml.
Using the above calibrations, also plot the calibrated PMC ERR and PZT spectra. |
Attachment 1: pmc-pzt-cal.pdf
|
|
Attachment 2: mcf.png
|
|
903
|
Fri Aug 29 17:39:25 2008 |
rana | Configuration | PSL | PMC: ADC Channels |
The attached PNG shows the PMC error and controls signals with no calibration.
There are 3 states:
DARK - RF input disabled & output blanked. This should be a measure of the ADC noise
(-10 dB) - This is with the gain slider down at 5 dB instead of the nominal 15 dB.
Looks like the Generic DAQ board whitening is good enough for these signal levels above ~1 Hz.
From the low and high gain spectra it also looks like the UGF is ~500 Hz with the gain at 15 dB. |
Attachment 1: mcf.png
|
|
902
|
Fri Aug 29 16:35:18 2008 |
Yoichi | Configuration | PSL | beam block distorted |
Quote: | There was a beam block after the Mach Zender. Who or what put this there?
The going to the MC now looks distorted as if someone has left something funny in the beam or maybe the new PMC has started to degrade??
Use the ELOG people...its good for you. |
The apparent distortion of the MC refl. was caused by mis-alignment of the MC mirrors.
Because the MC1 was mis-aligned, the reflected light was clipped by a steering mirror.
I restored the MC angle bias values from the conlog history and now the MC locks.
According to conlog, the MC alignment was changed at around 18:30 on Thursday PDT.
It could have been caused by the computer reboots. |
901
|
Fri Aug 29 15:01:45 2008 |
steve | Update | PSL | MOPA_HTEMP in increasing |
The laser chiller temp is 21.9C ( it should be 20.0C )
Control room temp 73F ok, no obvious block
Ops, there is a piece of paper blocking the intake of the chiller
This is a four day plot. The paper was blocking the air flow all day. |
Attachment 1: htcl.jpg
|
|
900
|
Fri Aug 29 12:43:44 2008 |
josephb | Summary | Computers | c1susvme1 down |
Around noon today, c1susvme was having problems. The C0DAQ_RFMNETWORK light was red. The status light was off, the sig det light was amber and the own data light was green. I could also ssh in, but could not not run startup. I switched off the watchdogs for c1susvme2 (the watchdogs for c1susvme1 had already been tripped), and manually power cycled the crate.
However, when c1susvme1 when it came back up it had not mounted the usual cvs/cds/ directories. c1susvme2 did however. c1susvme1 has been on the new network for awhile, while c1susvme2 was switch over today. So apparently switching networks doesn't help this particular problem.
I did a remote reboot of c1susvme1, and it came up with the correct files mounted. Both machines ran their approriate startup.cmd files and are currently green. |
899
|
Fri Aug 29 12:41:26 2008 |
josephb, Eric | Configuration | Computers | More front ends moved to new network |
Used Cat6 cables to finish moving all the front ends in 1Y4 and 1Y5 over to the new GigE network switches, specifically to the switch in 1Y6. This included the ones labeled c1susvme2, c1sosvme, and c1dscl1epics0. |
898
|
Fri Aug 29 11:05:11 2008 |
josephb | Summary | Computers | c1asc was down this morning |
I had to manually reboot c1asc this morning, as for some unknown reason its status was red, and the fiber lights on the board were status:red, sig det:amber, own data: nothing. Shut the crate down, turned it back on, heard a beep, then followed wiki reboot instructions. Seems to be working now. |
897
|
Fri Aug 29 11:01:49 2008 |
josephb | Configuration | Computers | Attempt to change a channel gain in ICS-110B |
As noted earlier by Rana, I was playing around with the /cvs/cds/caltech/chans/daq/C1IOOF.ini file with help from Rob. I had made a backup before hand and saved it as C1IOOF.ini.Aug-28-2008. (I have since been informed that C1IOOF.ini.082808 would have been prefered as a name).
We had been trying to up the gain in the C1: PSL-ISS_INMONPD_F in order to do a very low power PMC sweep, in an attempt to get clean modes for fitting. Initially we pressed the reconfig button on the C0DAQ_DETAIL screen, but all that seemed to do was change the Config File CRC. We proceeded to reboot fb40m remotely. However, any change to the ini file (even an extra space at the end of the file) caused a 0x2000 status for C1IOVME16k on the C0DAQ_DETAIL screen. At the time I presumed it was comparing the CRC of the ini-file to something else.
Digging around on in Alex's webspace at http://www.ligo.caltech.edu/~aivanov/ , I found the NDS Access page, which indicated that 0x2000 was a conflict between the front-end and frame builder .ini files.
"There is also status bit 0x2000 which gets added when the DCU configuration is different in front-end and frame builder. That is you can change and .ini file an then reload DAQ configuration with Epics button, which reconfigures the front-end, but leaves frame builders with invalid old configuration. They will detect this change and set the status to 0x2000 to indicate this condition. You will have to restart frame builders to pick up new .ini file and set status back to zero for the affected DCU."
It was when I was going to try reseting the c1iovme via the C0DAQ_RFMNETOWRK medm screen that we realized the EPICS controls were not responding properly. The .ini file was returned to its original form, and mass reboots commenced. |
896
|
Fri Aug 29 10:20:32 2008 |
Yoichi | Configuration | PSL | beam block distorted |
Quote: | There was a beam block after the Mach Zender. Who or what put this there?
The going to the MC now looks distorted as if someone has left something funny in the beam or maybe the new PMC has started to degrade??
Use the ELOG people...its good for you. |
I put the block. I was frequently reaching to the FSS box to change the test point probes. I put the block to protect my hands/clothes from being burnt accidentally. |