ID |
Date |
Author |
Type |
Category |
Subject |
8955
|
Thu Aug 1 18:55:20 2013 |
Alex Cole | Update | Electronics | Preliminary Photodetector Frequency Reponse Measurements |
Quote: |
[Eric, Alex]
We used our setup from yesterday (elog #8940) to measure transimpedance measurements for AS55, REFL11, REFL33, and REFL55, using our Newport 1611 FC-AC as reference. We connected the fibers to their respective telescopes such that the beams focused on their photodetectors, using a multimeter to maximize photodetector DC output. Plots are attached. At first glance, the poles seem to be where they're supposed to be.
Note that the procedure used today is similar to what the eventual automated procedure will be. The main differences are (1) The RF Switch will be used rather than manual switching (2) NWAG4395A will be used to collect data rather than netgpibdata (3) Data will be fit using vectfit4.m and compared to some canonical set.
|
[Alex, Eric]
Today I spent some time mounting the launcher and performing the same data collection for POX11. I think I still need to focus the launcher so the photodetector gets a good signal, but the data from today wasn't too bad. Additionally, I worked on matlab scripts to improve PDFR data analysis.
This time I collected data from the network analyzer using NWAG4395A in the netgpibdata directory. The advantage of this is that the computer tells the network analyzer to perform the sweep as well as retrieving the data.
For analysis, I improved my implementation of vectfit4.m so that it focuses in on the particular photodetector's predicted peaks and thus ignores much of the noise, giving a better fit. The raw data is the red circles in the 2nd attachment, while the fit is the blue line. I also had the program return the frequency value of the peak. For POX11, this was 1.106e+07 Hz.
I also finagled copies of existing programs to enable one to plot multiple transfer functions on the same axes. This function is /users/alex.cole/plottwo.m. I will eventually use this to compare new data to some canonical data so that we may monitor photodetector performance over time.
The eventual plan is to generate two plots per photodetector, one of which will compare new data to the canonical set, the other of which will show the fit of the data. Both will have subplots that zoom in around regions of interest (known peaks and notches), and the plot which displays the canonical set will also have Q's of peaks and their locations. |
Attachment 1: POX11.jpg
|
|
Attachment 2: POX11fit.jpg
|
|
8971
|
Tue Aug 6 12:43:23 2013 |
Alex Cole | Configuration | Electronics | AS Table and Rack 1Y1 Additions |
For the photodetector frequency response project, I finished the construction of our baluns chassis and mounted it in rack 1Y1 (1st picture).
After consulting with Jenne, I mounted the fiber launcher for REFL165 on the AS table such that it would not cause an obstruction. I aligned the launcher using a multimeter to monitor the DC output of REFL165, but looking at the data I got, it seems I need to do a better alignment/focusing job to get rid of a bunch of noise. |
Attachment 1: photo_1_(5).JPG
|
|
Attachment 2: photo_2_(5).JPG
|
|
8979
|
Wed Aug 7 15:51:53 2013 |
Alex Cole | Configuration | Electronics | RF Switch Change |
For the photodetector frequency response project, our new RF Switch Chassis (NI pxie-1071) arrived today. I took the switches out of the old chassis (Note for future generations: you have to yank pretty darn hard) and put them in the new chassis, which I mounted in rack 1Y1 as pictured.
The point of this new chassis is that its controller is compatible with our control room computer setup. We will be able to switch the chassis using TCP/IP or telnet, aiding in our automation of the measurement of photodetector frequency response. |
Attachment 1: photo_(2).JPG
|
|
8990
|
Fri Aug 9 16:49:35 2013 |
Jenne, manasa | Update | Electronics | Post-vent alignment cont'd - RFPDs |
Notes to the fiber team:
I am aligning beam onto the RFPDs (I have finished all 4 REFL diodes, and AS55), in preparation for locking.
In doing so, I have noticed that the fiber lasers for the RFPD testing are always illuminating the photodiodes! This seems bad! Ack!
For now, I blocked the laser light coming from the fiber, did my alignment, then removed my blocks. The exception is REFL55, which I have left an aluminum beam dump, so that we can use REFL55 for PRM-ITMY locking, so I can align the POP diodes.
EDIT: I have also aligned POP QPD, and POP110/22. The fiber launcher for POP110 was not tight in its mount, so when I went to put a beam block in front of it and touched the mount, the whole thing spun a little bit. Now the fiber to POP110 is totally misaligned, and should be realigned.
What was done for the alignment:
1. Aligned the arms (ran ASS).
2. Aligned the beam to all the REFL and AS PDs.
3. Misaligned the ETMs and ITMX.
4. Locked PRM+ITMY using REFL11.
The following were modified to enable locking
(1) PRCL gain changed from +2.0 to -12.
(2) Power normalization matrix for PRCL changed from +10.0 to 0.
(3) FM3 in PRCL servo filter module was turned OFF.
5. POP PDs were aligned. |
9004
|
Tue Aug 13 11:40:19 2013 |
Alex Cole | Summary | Electronics | RFPD Demod Filter Frequency Response Measurement |
For the RF PD Frequency Response Measurement project, we get each PD signal from the "PD RF Mon" output of each demodulator board corresponding to our PD under test. Therefore we can't neglect the frequency response of various filters inside the demodulator board. I used our Agilent 4395 Network Analyzer to gather frequency response data for each demodulator board being considered for the RFPD frequency response project (AS55, REFL11, REFL33, REFL55, REFL165, POX11, POP22, POP110).
The NA swept over a frequency range of 1-500 MHz. Data was collected using NWAG4395A (from the netgpibdata directory). It should be noted that the command line options -a 16 -x 15 (averaging=16 and excitation amplitude=15 dBm[the max]), in addition to the usual command line options described in the help file, were used to minimize noise.
The data is located in /users/alex.cole. The file names are in the format [PDNAME]DemodFilt_1000000.dat (e.g. REFL11DemodFilt_1000000.dat). Results for POP110 are shown below. |
Attachment 1: photo_(3).JPG
|
|
Attachment 2: test.jpg
|
|
9005
|
Tue Aug 13 11:54:40 2013 |
Alex Cole | HowTo | Electronics | RF PD Fiber-Coupled Laser Operation |
This post pertains to the fiber-coupled diode laser mounted in rack 1Y1.
To turn the laser on, first turn the power supply's key (red) to the clockwise. Then make sure that the laser is in "current" mode by checking that the LED next to "I" in the "Laser Mode" box in lit up. If the light is not on, press the button to the right of the "I" light until it is. Now press the output button (green). This is like removing the safety for the laser. Then turn the dial (blue) until you have your desired current. Presently, the current limit is set to around 92 mA.
To turn the laser off, dial the current back down to 0mA and turn the key (red) counterclockwise. |
Attachment 1: photo_(4).pdf
|
|
9006
|
Tue Aug 13 13:30:41 2013 |
Alex Cole | Configuration | Electronics | Cable Routing |
I routed cables (RG405 SMA-SMA) from several demodulator boards in rack 1Y2 to the RF Switch in rack 1Y1 using the overhead track. Our switch chassis contains two 8x1 switches. The COM of the "right" switch goes to channel 7 of the "left" switch to effectively form a 16x1 switch. The following is a table of correspondences between PD and RF Switch input.
PD |
Left/Right Switch |
Channel Number |
REFL11 |
R
|
0 |
POX11 |
L |
0 |
AS55 |
R |
1 |
REFL55 |
R |
7 |
POP22 |
R |
6 |
REFL165 |
R |
5 |
REFL33 |
L |
7 |
ThePOP110 demod board has not yet had a cable routed from it to the switch because I ran out of RG405.
We should also consider how important it is to include MCREFL in our setup. Doing so would require fabrication of a ~70 ft RG405 cable. |
Attachment 1: photo_(6).JPG
|
|
9059
|
Fri Aug 23 21:01:38 2013 |
Alex Cole | HowTo | Electronics | Automated Photodetector Frequency Response System |
This post describes how to use the Automated Photodetector Frequency Response System.
On the mechanical side, turn on:
-the diode laser (in rack 1Y1)
-the RF Switch (in rack 1Y1)
-the reference PD (under the POY table)
-the AG4395A Network Analyzer
The NA’s RF output should go to the laser’s modulation input, the reference PD’s output should go to the NA’s R input, and the RF Switch Chassis’s output (which is the combination of the two switches’ COM channels using a splitter) should go to the NA’s A input.
Once this is done, navigate into /users/alex.cole and run PDFR.sh. This script collects data for each photodetector under consideration by switching using a python script and communicating with the NA via GPIB. It then sends all the data to RF.m, which fits the functions, plots the latest data against canonical data, and saves the plots to file.
The fitting function, fit.m, also outputs peak frequency to the command line. This function uses PD name data (e.g. ‘REFL33’) to choose an interval with minimal noise to fit.
The main script prompts the user to press enter after each NA sweep to make sure that measurements don’t get interrupted/put out of order by RF switching.
Once you're done, you should turn off the laser, NA, RF Switch, and reference PD.
Troubleshooting
Sometimes, the NA throws up and doesn’t feel like running a particular sweep. If this happens, it’s a good idea to keep the matlab script from trying to analyze this PD’s data. Do this by opening up RF.m and commenting out the calls to ‘fit’ and ‘canonical’ for that PD.
If fit.m complains about a particular set of data, it is often the case that the N/P ratio (where N is order of approximation and P is number of points in the interval) is too high. You can fix this by reducing N or making the PD’s frequency range (chosen in the fnew_idx line) larger.
Choosing a single PD
If you only want to grab the transfer function for one PD, first look up which switch input it belongs to. This information is contained in /users/alex.cole/switchList. To turn the switch to a particular input, type something like:
python rf.py “ch7”
This command uses TCP/IP to tell the switch to look at channel 7. Switch input numbers range from 1 to 16, though not all of them are in use.
Once the switch is looking at the correct input, you can run a sweep and download the data by typing /opt/rtcds/caltech/c1/scripts/general/netgpibdata/NWAG4395A -s 1000000 -e 500000000 -c 499000000 -f [filestem for output] -d [path of directory for output] -i 192.168.113.108 -g 10 -x 15. |
9062
|
Mon Aug 26 18:55:18 2013 |
Jenne | Configuration | Electronics | putting together a 110 MHz LSC demod board for AS |
I have modified one of the spare demod boards that was sitting above the electronics bench (the one which was unlabeled - the others say 33MHz, 55MHz and 165MHz) to be the new AS110 demod board. In place of the T1 coil, and the C3 and C6 resistors, I have put the commercial splitter PSCQ-2-120+. In place of U5 (the low pass for the PD input) I have put an SCLF-135+.
In order to figure out how to make the pinout of the PSCQ match up with the available pads from T1, I first pulled the "AS11" board (it's not something that we use, so it would be less of a tragedy if something happened while I had the board pulled). However, while the PCB layout is the same, the splitter for the low frequencies (PSCQ-2-51W) has a different pinout than the one I need for the 110MHz. So, I put AS11 back, and pulled the POP110 board. (After I noted the pinout on POP110, I reinstalled that board. To get it out, I had to unplug the I and Q outs of POP22, but I have also replugged those in).
For my new AS110 demod board, I copied the pin connections on POP110. I have made a little diagram, so you can see what pins went where. The top 2 rectangles are the "before" installation cartoon, and the bottom is the "as installed" cartoon.

The one thing that must be noted is that, because of the pinout of the splitter and the constraints of the board layout, the +0 degrees (I-phase) output of the splitter is connected to the Q channel for the rest of the demod board. This means that the +90 degrees (Q-phase) output of the splitter is connected to the I channel for the rest of the demod board. This is not noted for POP110, but is true for both: The I and Q channels of the 110 MHz demod boards are switched. In practice, we can handle this with our digital phase rotation.
Daytime tomorrow, I will test my new board as Suresh did in elog 4736. Before we get to use AS110, we need (a) some LO juice from the RF distribution box, and (b) a spot to plug the board in, in the LSC rack. Meditating on how those are going to happen are also tasks for daytime tomorrow. |
9067
|
Mon Aug 26 20:13:17 2013 |
rana | Configuration | Electronics | putting together a 110 MHz LSC demod board for AS |
Quote: |
I have modified one of the spare demod boards that was sitting above the electronics bench (the one which was unlabeled - the others say 33MHz, 55MHz and 165MHz) to be the new AS110 demod board. In place of the T1 coil, and the C3 and C6 resistors, I have put the commercial splitter PSCQ-2-120+. In place of U5 (the low pass for the PD input) I have put an SCLF-135+.
|
OK, but what kind of filter should we be actually using? i.e. what purpose the 135 MHz low pass serve in contrast to a PHP-100+ ? |
9069
|
Tue Aug 27 15:31:48 2013 |
Jenne | Configuration | Electronics | putting together a 110 MHz LSC demod board for AS |
Quote: |
Quote: |
I have modified one of the spare demod boards that was sitting above the electronics bench (the one which was unlabeled - the others say 33MHz, 55MHz and 165MHz) to be the new AS110 demod board. In place of the T1 coil, and the C3 and C6 resistors, I have put the commercial splitter PSCQ-2-120+. In place of U5 (the low pass for the PD input) I have put an SCLF-135+.
|
OK, but what kind of filter should we be actually using? i.e. what purpose the 135 MHz low pass serve in contrast to a PHP-100+ ?
|
Hmmm. Indeed. This is just cutting off higher frequency stuff, but anything from other lower sidebands still gets through. I should actually stick in the SXBP-100's, which will band pass from 87-117 MHz. These have an insertion loss at 100 MHz of 1.64 dB.
Jamie ordered 2 of these, so I can put one in each of AS110 and POP110. |
9071
|
Tue Aug 27 17:32:52 2013 |
Jenne | Configuration | Electronics | putting together a 110 MHz LSC demod board for AS |
I measured the phase split between the I and Q signals of my AS110 board. To do so, I plugged the board into an empty slot next to the PD DC readout / whitening board in the LSC rack. I borrowed the POP110 local oscillator, and used a Marconi to generate a "PD input". (I'm roughly following what Suresh did in elog 4736). Our 11MHz is currently 11.066134MHz, so I had the Marconi going at 110.662340 MHz (1kHz from 10*11MHz), and I had the Marconi source at -13dBm.
I took a transfer function using the SR785 between the I and Q outs of the AS110 demod board, and got a magnitude misbalance of 0.809 dB, and a phase split of 110.5 degrees. This isn't so close to 90 degrees, but this may be a problem with the splitter that we're using, as Suresh detailed in elog 4755. In that elog, he measured a phase split of POP110 of 105 degrees, unless the power going into the splitter was pretty high. As with POP110, since I expect that we'll usually only look at one channel (I, for instance), this isn't such a big deal for AS110.
I have left, for now, the board in the empty slot. It looks like (I'm going to go check) there are 3 open channels on the whitening board that has the PD DC signals. So, the only thing left to figure out is how we want to get some local oscillator action for this new board.
EDIT: Yes, those channels are available. Right now (as a remnant from testing the whitening filters waaaay back in the day) they are called C1:LSC-PDXXX I, Q, DC. I'll use 2 of those for the AS110 I and Q. |
9072
|
Tue Aug 27 18:21:35 2013 |
Jenne | Configuration | Electronics | 110 MHz LO options |
As I see it, we have a few options for getting the 110 MHz LO to both the POP110 and AS110 demod boards.
The current situation is described by Kiwamu in elog 5746. The 55 MHz signal comes into the box, and is split 4 ways, with each path having 19.7 dBm. One of these 4 is for 110. It has a 2dB attenuator (giving us ~17.7 dBm), and then it goes to an MK-2 frequency multiplier. I'm a little lost on why we're giving the MK-2 17 dBm, since it says that it can handle an input power between 1 - 15 dBm. It has ~16 dB conversion loss, so the 110 output of the distribution board has (according to the drawing) 1.9 dBm. The demod boards have a 10 dB attenuator as the first element on the LO path, so we're giving the ERA-5 -8 dBm.
We can either amplify the 110 leaving the distribution box, split it, and then attenuate it to the appropriate level for the demod boards, or we can change the attenuators on the POP110 and AS110 demod boards.
Since we seem to be over driving the 2x frequency multiplier, I think I should change the 2dB attenuator to a 5dB attenuator, so we're giving the 2x multiplier ~15 dBm. The conversion loss of ~16 dB means we'll have -1 dBm of 110 MHz. I want to amplify that by ~10 dB, to give 9 dBm. Attenuate by 5 dB to get to 4 dBm, then split into 2, giving me 2 110 MHz spigots, each of ~1 dBm. Since the demod boards expect between 0-2 dBm for the LO's, this should be just fine.
Thoughts, before I start scrounging parts, and pulling the RF distribution box? |
9073
|
Tue Aug 27 18:58:52 2013 |
Koji | Configuration | Electronics | 110 MHz LO options |
- Do we have an appropriate amplifier?
- True challenge could be to find a feedthrough for the new port. (or to find a space for the amplifier in the box)
- PDXXX channels is on the DC whitening filter module. There could be some modification on this module (like diabling the whitening gain selector).
- We don't have AS11 and AS165, and so far it is unlikely to use AS11. i.e. The feedthrough, the slot on the crate, the whitening, and the channels can be trasnsition from 11 to 110.
Quote: |
I want to amplify that by ~10 dB, to give 9 dBm. Attenuate by 5 dB to get to 4 dBm, then split into 2, giving me 2 110 MHz spigots, each of ~1 dBm.
Thoughts, before I start scrounging parts, and pulling the RF distribution box?
|
|
9095
|
Mon Sep 2 16:46:32 2013 |
Jenne | Update | Electronics | RF distribution box is on the bench |
I have pulled the RF distribution box out of the rack, so I can look at it, and modify it to have 2 110 MHz spigots. I'm going to make the mods as in elog 9072.
Before I pulled the distribution box, I turned off the RF Generation Box, so don't be surprised that the MC will not lock. I have terminated the cables that bring the 11 and 55 MHz signals from the generation box to the distribution box, so if someone does turn on the generation box, there won't be bad reflections.
To get the box out, in addition to unplugging all of the cables that go to the distribution box, I had to disconnect 2 of the ADC ribbon cables from the top row of RFPD demod / whitening / ADC boards, since they were in the way. Everything is labeled, so it should be easy to put back together.
Note to Future Jenne: Past Jenne put the screws needed for those ADC cables and to hold the box in the rack, in the plastic box that is on the floor in front of the LSC rack.
Also, I measured the 110 MHz port before I pulled the board, so I would know what my "before" looked like. I was using the 300MHz 'scope's measurement functions, so these are in volts, not dBm. Amplitude = 1.33V, RMS = 456 mV, freq = 109.4-111.9 MHz |
9096
|
Mon Sep 2 18:06:21 2013 |
Jenne | Update | Electronics | RF distribution box: 110 MHz LO options |
After scrounging for parts, and opening up the box, I have modified my proposal to the following:

Note that the freq multiplier is supposed to take, at maximum, 15 dBm. The reason I put the 5 dB attenuator, then an amplifier, then another attenuator is that I don't know of / can't find easily a 10 dB amplifier with the usual case type on the MiniCircuits site. (If anyone knows of one off the top of their head, that would be handy. Then I'd remove the attenuator between the multiplier and the amplifier, and make the 10 dB attenuator a 5 dB.)
Anyhow, the ZFL-500HLN can only output 16 dBm of power, and I don't think I have space for another ZHL-2 (which can output up to 26 dBm) inside the box, so I put an attenuator before, as well as after, the amplifier.
I think I have space inside the box for all the bits and pieces I'll need, although to do things correctly, I need to drill holes in the teflon mounting plate to mount the amplifier and splitter.

I also think that I have space on the front panel to put another isolated SMA feedthrough.

I have, on my desk, all the parts (except for mounting screws, and cables between things) to make these modifications to the distribution box. |
9099
|
Tue Sep 3 21:08:13 2013 |
Jenne | Update | Electronics | RF distribution box: 110 MHz LO options |
The RF distribution box is still on the bench, so again, don't be surprised that the MC doesn't lock.
I have completed my modifications as proposed in elog 9096, but I want to do a couple of quickie tests in the morning before I declare it ready for service. |
9100
|
Tue Sep 3 21:10:36 2013 |
Jenne | Configuration | Electronics | putting together a 110 MHz LSC demod board for AS |
Quote: |
I should actually stick in the SXBP-100's, which will band pass from 87-117 MHz.
|
I have removed the 135 MHz low pass from my new AS110 demod board, but these SXBPs have different feet than the SCLFs, so I want to confirm with Koji or someone that I can solder them in the same way, before I get carried away and destroy anything. I should be able to finish this up tomorrow, plug in the demod board and the distribution box, and try out AS110 triggering, etc, tomorrow night. |
9101
|
Wed Sep 4 16:06:40 2013 |
Jenne | Update | Electronics | RF distribution box: Reinstalled |
I have reinstalled the RF distribution box, as well put in the AS110 demod board. I plugged everything back in, and turned it all on.
The switch on the distribution box may be starting to fail. When I was turning the box on, I could depress the button, and see the blue glow, but it wouldn't catch, so when I removed my finger, the glow went away. I was afraid that I'd have to pull the box, but after a few more button toggles, I got it to stay on. I'm leaving it for now, but we should remember that this may be a problem.
I will look at the phases of all the PDs, but none should need changing except POP 110. Every other PD has the exact same cables as before. |
9106
|
Wed Sep 4 21:44:54 2013 |
Not Jenne | Update | Electronics | RF distribution box: Reinstalled |
Temporary fix for the switch: give a bit of oil to the button
Permanent fix: buy better switches. |
9208
|
Sun Oct 6 22:27:35 2013 |
rana | Frogs | Electronics | MC3 LL sensor cable was loose |
I noticed that the MC3 LL sensor was apparently dead according to its suspension screen. Since it was only the fast ADC channel and not the SLOW PDmon, I could tell that it was just in the ADC cabling. I pushed in a few of the MC3 sensor cables on the front and back of the PD whitening board and it came back OK. According to this trend of the past 40 days and 40 nights, it started slipping on this past Wednesday morning.
Was anyone walking near MC2 or the suspension electronics racks before noon on Wednesday (Oct. 2nd)? |
Attachment 1: MC3_LL.png
|
|
9350
|
Tue Nov 5 19:50:07 2013 |
manasa | Update | Electronics | IOO rack +/-5V power supplies |
The power supply to the ADC box on the IOO rack (that reads the beat I & Q signals) was pulled out because it did not run through any fuse and was connected directly to the power supply.
There were already connections running from the +/-5 V power supply. They were powering the mode cleaner demod board rack. In order to remove the ADC power connectors from the power supply, I notified Jenne in the control room because turning off the power supply would affect the MC. I switched off the +/-5V power supplies at the same time. The ADC power connectors were removed. The +/-5V power supplies were then turned ON again at the same time. Jenne relocked the MC after this.
I have still not connected the ADC to the fuse rack power supply because this requires the +/-5V power supplies to be turned OFF again in order to pull out new connections from the fuse rack and I need to make a new ADC power connector with thicker wires. |
9356
|
Wed Nov 6 15:59:41 2013 |
manasa | Update | Electronics | IOO rack +/-5V power supplies |
Quote: |
The power supply to the ADC box on the IOO rack (that reads the beat I & Q signals) was pulled out because it did not run through any fuse and was connected directly to the power supply.
There were already connections running from the +/-5 V power supply. They were powering the mode cleaner demod board rack. In order to remove the ADC power connectors from the power supply, I notified Jenne in the control room because turning off the power supply would affect the MC. I switched off the +/-5V power supplies at the same time. The ADC power connectors were removed. The +/-5V power supplies were then turned ON again at the same time. Jenne relocked the MC after this.
I have still not connected the ADC to the fuse rack power supply because this requires the +/-5V power supplies to be turned OFF again in order to pull out new connections from the fuse rack and I need to make a new ADC power connector with thicker wires.
|
I switched OFF the +/-5V power supplies on the IOO rack to hook up the ADC power connectors through 250mA fuses to +/-5V. Since these power supplies were powering the MC demod boards, MC remained unlocked during the process. I turned the power supplies back ON and MC relocked itself after this. |
9377
|
Wed Nov 13 18:37:19 2013 |
rana | Configuration | Electronics | DAC available in c1lsc IO chassis for DAFI |
The first picture shows that there is indeed a DAC next to the ADC in the LSC IO chassis. The second picture shows how there are two cables, each one carrying 8 channels of DAC. The third one shows how these come out of the coil drivers to handle the Tip/Tilt mirrors which point the beam from the IMC into the PRC. It should be the case that the second Dewhitening filter board can give us access to the next 8 channels for use in driving an audio signal into the control room or an ISS excitation. |
9563
|
Tue Jan 21 19:41:59 2014 |
Jenne | Update | Electronics | RF distribution box power button fail |
Rana, Gabriele and I are trying to measure the FSR of the PRC (elog about that later), and we turned off the power to the RF generation box so that we could switch cables at the EOM combiner. However, as in elog 9101, the power button won't latch when we try to turn the power back on. All 3 of us tried, to no avail. For our measurement, poor Gabriele is standing holding the button pushed in, so that we can have some RF sidebands.
Tomorrow, we'll have to pull the RF generation box, and put in a better switch. |
9566
|
Wed Jan 22 16:36:45 2014 |
ericq | Update | Electronics | RF distribution box power button fail |
Quote: |
Rana, Gabriele and I are trying to measure the FSR of the PRC (elog about that later), and we turned off the power to the RF generation box so that we could switch cables at the EOM combiner. However, as in elog 9101, the power button won't latch when we try to turn the power back on. All 3 of us tried, to no avail. For our measurement, poor Gabriele is standing holding the button pushed in, so that we can have some RF sidebands.
Tomorrow, we'll have to pull the RF generation box, and put in a better switch.
|
I replaced the stupid broken fancy button with a simple sturdy switch. I had to file out the hole in the chassis a bit, but the switch is pressed in tightly and securely. I put the box back in the rack, but the power cable was coming directly from the power supplies with no fuses. The box was drawing ~.9 and 1.5 Amps from two supplies, so I put 2A fuses on both. Plugged everything back in, and the mode cleaner locks, so it looks like all is well.
RXA: When its so close, I prefer to size it up by 1 step. Please change to 5A fuses. Otherwise, we may blow them from power glitches.
Q: 5A fuses have been swapped in |
9637
|
Fri Feb 14 02:09:55 2014 |
ericq | Update | Electronics | Transmon QPD whitening |
[Quick post, will follow up with further detail later. Excuse my sleepy ELOG writing]
Goal: Check out the transmon QPD signal chain; see if whitening works. Assess noise for 1/sqrt(TRX/Y) use.
First impression: Whitening would not switch on when toggling the de-whitening. The front monitors on the whitening boards are misleading; they are taken a few stages before the real output. ADC noise was by far the limiting noise source.
I updated the binary logic in the c1scx and c1scy to actually make the binary IO module output some bits.
After consulting a secret wiring diagram on the wiki, not linked on the rack information page (here), I worked out which bits correspond to the bypass switches in the whitening board ( a fairly modified D990399, with some notes here)
Now, FM1 and FM2 (dewhitening filters on the ETM QPD quadrants) trigger the corresponding whitening in the boards. Here's a quick TF I took of the quadrant 1 board at ETMY. (I should take a whitening+dewhitening TF too, and post it here...)

Seems to roughly work. Some features may be due to non-accounted for elements in the anti-imaging of the DAC channels I used for the excitation, or such things. The board likely needs some attention, and at least a survey of what is there.
I also need to take dark noise data, and convert into the equivalent displacement noise in the 1/sqrt(TRX/Y) error signals. For the no-whitening ADC noise, I estimated ~1pm RMS noise on a 38pm linewidth of PRFPMI arms. |
9642
|
Mon Feb 17 20:35:19 2014 |
ericq | Update | Electronics | Transmon QPD whitening |
My apologies for all of that crap I left at the Y-end... I cleaned the rest of it up today.
I took transfer functions of the four ETMY QPD whitening channels today. (Attempted the ETMX ones too, but had troubles driving the board; detailed below). I've attached a zip with the DTT xml files for the cases of no whitening / 1 whitening stage / both whitening stages engaged. Here's a plot of both whitening stages engaged.

Given the way I measured, the DAC output anti-imaging is in the TFs as well. ( This is a D000186 board; with something like a 4th order elliptic LP, but I need to look at the board / fit the TF to see the parameters, there are different revisions with different filter shapes.)
The c1scy model had excitation blocks on some of the unused DAC channels (C1:SCY-XXX_CHAN9 etc.), but these were in the second DAC output connection, and not cabled up. However, the 8th channel on the DAC had no connection in the simulink model, so I added another excitation block there (C1:SCY-XXX_CHAN8), and used the anti-imaging front panel lemo connector to drive the input of the whitening board.
I also added a similar channel to the SCX model, but no data would show up in the channel as viewed by data viewer (though the channel name was black), or in analog world. There's the additional weirdness that the SCY excitation channels show up under SCX in DTT and awggui... I'm not entirely sure what's going on here.
I still need to look at the noise, and peek inside the boards, to check for homemade modifications and see if there are bad things like thick film resistors that may be spoiling the noise performance... |
Attachment 2: ETMY_QPD_whitening.zip
|
9661
|
Mon Feb 24 13:21:00 2014 |
Jenne | Update | Electronics | Measured REFL165 demod board |
I measured the REFL 165 demod board's I/Q separation.
Our 11MHz signal is currently 11.066092 MHz, so I put a signal to the RF input of the REFL165 demod board at 165.992380 MHz (15*11 MHz + 1kHz), with a signal of -13 dBm.
I then used the SR785 to measure the transfer function between the I and Q output channels.
I got 82.7 degrees, at -0.64 dB. (I don't remember now if I had I/Q, or Q/I, not that it really matters). So, it seems that the REFL165 demod board has good separation, and at least isn't totally broken. |
9666
|
Mon Feb 24 17:59:31 2014 |
RANA | Update | Electronics | Measured REFL165 demod board |
Demod boards should be at 90 deg, not 82.7 or 12 or yellow or ****. We should re-inject the RF and then set the D Phase in the filter module to make the signals orthogonal. 165 is a challenging one to get right, but its worth it since the signals are close to degenerate already. |
9716
|
Tue Mar 11 15:19:45 2014 |
Jenne | Update | Electronics | High gain Trans PD electronics change |
As part of our CESAR testing last night, we had a look at the noise of the 1/sqrt(TR) signal.
Looking at the time series data, while we were slowly sweeping through IR resonance (using the ALS), Rana noted that the linear range of the 1/sqrt(TR) signal was not as wide as it should be, and that this is likely because our SNR is really poor.
When a single arm is at a normalized transmission power of 1, we are getting about 300 ADC counts. We want this to be more like 3000 ADC counts, to be taking advantage of the full range of the ADC.
This means that we want to increase our analog gain by a factor of 10 for the low gain Thorlabs PDs.
Looking at the photos from November when I pulled out the Xend transmission whitening board (elog 9367), we want to change "Rgain" of the AD620 on the daughter board. While we're at it, we should also change the noisy black thick film resistors to the green thin film resistors in the signal path.
The daughter board is D04060, S/N 101. The main whitening board for the low gain trans QPD is D990399, RevB, S/N 104.
We should also check whether we're saturating somewhere in the whitening board by putting in a function generator signal via BNC cable into the input of the Thorlabs whitening path, and seeing where (in Dataviewer) we start to see saturation. Is it the full 32,000 counts, or somewhere lower, like 28,000?
Actually the gain was changed. From gain of 2 (Rgain = 49.4kOhm) to 20 (Rgain = 2.10kOhm), Corresponding calibration in CDS was also changed by locking the Xarm, running ASS, then setting the average arm power to be 1. Confirmed Xarm is locking. And now the signal is used for CESAR. We see emperically that the noise has improved by a factor of approximately 10ish. |
Attachment 1: IMG_1309.JPG
|
|
9720
|
Tue Mar 11 19:07:24 2014 |
ericq | Update | Electronics | High gain Trans PD electronics change |
Speaking of the whitening board, I had neglected to post details showing the the whitening was at least having a positive effect on the transmon QPD noise. So, here is a spectrum showing the effects that the whitening stages have on a QPD dark noise measurement like I did in ELOG 9660, at a simulated transmission level of 40 counts.
The first whitening stages gives us a full 20dB of noise reduction, while the second stage brings us down to either the dark noise of the QPD or the noise of the whitening board. We should figure out which it is, and fix up the board if necessary.

The DTT xml file is attached in a zip, if anyone wants it. |
Attachment 2: sqrtinvWhitening.zip
|
9823
|
Thu Apr 17 16:04:40 2014 |
Jenne | Update | Electronics | High gain Trans PD electronics change |
I have made the same modification to the Yarm trans PD whitening board as was done for the xend, to increase our SNR. I put in a 2.1kOhm thin film resistor in the Rgain place.
When I was pulling the board, the ribbon cable that goes to the ADC had its connector break. I redid the ribbon connector before putting the board back.
I see signals coming into the digital system for both the high gain and low gain Y transmission PDs, so I think we're back. I will re-do the normalization after Jamie is finished working on the computers for the day. |
9828
|
Fri Apr 18 09:40:28 2014 |
Steve | Update | Electronics | it is a kick |
10 days of on-off glitching? It is a kick. The LSC is off, so it must be the ALS |
Attachment 1: theLifeofGlithes.png
|
|
Attachment 2: glitchingEndsMC2.png
|
|
Attachment 3: itisaKick.png
|
|
9991
|
Sat May 24 22:56:57 2014 |
Jenne | Update | Electronics | Amplifier removed from BeatX path |
I just realized that I forgot to elog this, but yesterday afternoon I bypassed the amplifier in the BeatX path, and now the X beatnote is about -27dBm. Arms lock nicely with ALS. |
9992
|
Mon May 26 07:59:23 2014 |
Koji | Update | Electronics | Amplifier removed from BeatX path |
And the out-of-loop level of the ALSX compared with the previous measurement is ...?
Quote: |
I just realized that I forgot to elog this, but yesterday afternoon I bypassed the amplifier in the BeatX path, and now the X beatnote is about -27dBm. Arms lock nicely with ALS.
|
|
9995
|
Tue May 27 11:58:45 2014 |
Jenne | Update | Electronics | Amplifier removed from BeatX path |
Sorry, I had been in a hurry when I worked on this last week, and again when I wrote the elog, but I wanted to at least put in a note for any weekend workers.
The ALS beatnote setups need alignment on the PSL table. However, even at very low RF beat frequency, the X beatnote now at low frequencies matches our best measurement from last week. The "HEPA off" (teal and purple) measurements are from last week, and the red and blue are from this week. The X beatnote was 10MHz and the Y beatnote today was 31MHz.

|
10008
|
Mon Jun 9 09:51:11 2014 |
Sai Akhil | Update | Electronics | Frequency Error Measurement of UFC-6000 RF Frequency Counter |
Motivation:
To test the precision of Mini-Circuits Model UFC-6000 RF Frequency Counter which will be used for recording the beat note for the Frequency Offset Locking Loop(FOLL).
Setup:
Mini Circuits RF Frequency Counter Model UFC-6000 has three I/O ports:
1)REF IN,2)USB INTERFACE,3)RF IN.
The USB INTERFACE is connected to the PC(Windows/Linux) through a USB cable.
The test RF input from an SRS Function Generator(Model DS 345 with tested precision up to 1mHz)is fed in through RF IN using an SMA cable with an SMA-BNC adaptor to connect to the Function Generator.
The REF IN is open since we want to test the counter.
What I did:
1. First interfaced the counter with the PC with windows OS.
2. Installed the user friendly GUI on my Laptop so as to record the data from the counter into a .txt file.
3. Gave an RF input through the function generator and recorded the response of the counter for different frequencies ranging from 1MHz to 30MHz.
4.Analyzed the collected data by plotting the histograms(attached) in Matlab(script attached in .zip file)
What was Expected:
The measurement statistics of the instrument would give knowledge about the error and tolerance in the measurement which will be helpful to negotiate the error when the counter is being used in the setup.
Results:
The obtained plots(for sampling time of 1s) are attached in a figure.
The measurement error of the frequency counter for 1s sampling time is:
data file Frequency Mean in MHz Standard Error(+/-)in Hz
1MH.txt 1MHz 0.999999846 0.0109
5MHz.txt 5MHz 5.000000293 0.0134
10MHz.txt 10MHz 10.00000148 0.0108
15MHz.txt 15MHz 15.0000018 0.0072
20MHz.txt 20MHz 20.00000053 0.0259
30MHz.txt 30MHz 30.00000146 0.0230
The measurement error of the UFC-6000 RF Frequency Counter is in the order of 0.01-0.02 Hz. This error varies at different frequencies as inferred from the table.
The error for different sampling times of the FC are also plotted.
Plan:
To complete interfacing the counter with the Raspberry-pi.
Make this Frequency Counter talk to EPICS through slow channels.
|
Attachment 1: Data_and_Script_Files.zip
|
Attachment 2: Error_Measurement_FC.pdf
|
|
Attachment 3: Error_freq.jpg
|
|
10009
|
Mon Jun 9 10:55:48 2014 |
Nichin | Summary | Electronics | BBPD D1002969-v8 transimpedence measurement |
My SURF week-1 work...
Motivation:
To measure the transimpedence of the Broadband photodiode (D1002969-v8), using a New focus photodiode (1611) as reference. The amplitude modulated Jenne Laser (1.2mW) was used.
The steps involved in getting the transimpedence are as follows:
Acquiring data
- Get 2 sets of data from Network Analyzer Agilent 4395: One set of data will be for the transfer function of Ref PD over RF out. The other set for Test PD over Ref PD.
- The following conditions were set:
1) Frequency sweep range: 1MHz to 200 MHz.
2) Number of Points sampled in the range: 201
3) Type of sweep: Logarithmic
- Set the NA to give the corresponding transfer function values in dB and also Phase response in degrees.
- Save the data into floppy disk for processing on the computer (The wireless way of acquiring data was not working when the experiment was conducted )
Plotting
- The matlab code attached (TransimpedencePlot.m) will then give plots for the absolute values of transimpedence in V/A.
- Logic involved in the code:
- Transimpedence = Voltage response / (Responsivity of the photodiode * Power incident)
- Responsivity for BBPD is taken as 0.1 A/W and for NF1611 as 0.68 A/W as given in their datasheets.
- Voltage response of Test PD w.r.t RF output of NA (in dB) = Voltage response of Test PD w.r.t Ref PD (in dB) + Voltage response of Ref PD w.r.t RF output of NA (in dB)
Results
The Plots of transimpedence obtained are attached (results.pdf) . The results obtained for BBPD is consistent with the ones obtained before, but the same method and code gives a different transimpedence for 1611.
The transimpedence of NF 1611 was obtained to be around 4-5 V/A which is very much off-track compared to the one given in the datasheet (elog: 2906).
The transimpedence of Broadband photodiode (D1002969-v8) was around 1200 - 1300 V/A for most of the range, but the value started falling as the frequency approached 100 MHz. This result is consistent with DCC document: T1100467-v2.
|
Attachment 1: PD_transimpedence_measurement.png
|
|
Attachment 2: results.pdf
|
|
Attachment 3: code_and_data.zip
|
10012
|
Mon Jun 9 16:55:31 2014 |
Koji | Summary | Electronics | BBPD D1002969-v8 transimpedence measurement |
How is the modulation depth assumed in the calculation?
If you don't know the modulation depth, you can't calibrate the transimpedance of each PD individually. |
10034
|
Thu Jun 12 16:56:31 2014 |
Nichin | Update | Electronics | PD Inspection |
I and Eric Gustafson inspected the automated PD frequency response measurement system which Alex Cole built last summer. We just lifted the tops off the tables [AS table, POY table and ITMX table] and looked at the alignment checking to see if the correct optical fibers from the fiber splitter were illuminating the correct photodiodes. We did not change anything at all and put the covers back on the tables.
The PDF attached shows the state of each PD fiber pair. The fibers labeled REFL11 and REFL55 were reversed and illuminating the wrong photodiodes.
We will do a manual measurement of REFL33 tomorrow using the network analyzer and the modulatable laser but not the RF switch. Afterward we will check to make sure the RF cables are connected to the correct channels of the RF switch according to the switch list (/users/alex.cole/switchList). |
Attachment 1: Inspection_PD_Freq_Resonse_system_12th_June_2014.pdf
|
|
10036
|
Fri Jun 13 11:33:55 2014 |
Akhil | Update | Electronics | Characterization of UFC-6000 RF Frequency Counter |
Goal:
To characterize the Mini-Circuits RF FC (Model UFC-6000) and plot Bode Plots at varying Modulation frequencies.
Work Done:
Here are the list of measurements(files attached) taken from FC using SRS(Model DS345) Synthesized Function Generator for a Sinusoidal signal at different Modulation Frequencies ranging from 0.01 Hz to 1 KHz:
Carrier Frequency Modulation Depth Attached measurement Folder
5 MHz Δ f = 5 MHz Bode_5
10 MHz Δ f = 10 MHz Bode_10
20 MHz Δ f = 20 MHz Bode_20
The measured data will be used to estimate:
1) Transfer Function of FC
2) Quantization noise from Power Spectral Density(PSD) vs Hz
To Do Next:
1)Complete interfacing the Pi with EPICS.
2)Make bode plots (Matlab script attached) and PSD plots and estimate the control parameters for optimal design of FOLL PID loop.
|
Attachment 1: Bode_Plots.zip
|
10037
|
Fri Jun 13 18:16:00 2014 |
Nichin | Update | Electronics | Changes to the PD frequency response measurement system |
As we had planned yesterday (ELOG 10034) I and Eric Gustafson wanted to manually measure the transimpedence for REFL33. But on closer inspection I found the RF signal cable coming from the Photodiode REF DET (mounted on the POY table), that we were supposed to plug into the network analyzer, did not have an SMA connector at the end. There was just the Teflon and metal part sticking out of the insulation. So we disconnected the cable labeled REF DET from the PD and pulled it out to fix it. (POY table and from near the 1Y1 rack)
Being unable to locate any SMA male connectors in the 40m lab [pasternack PE4025], we headed over to Downs where Rich Abbott did a quick and awesome job of soldering the SMA connector and also teaching me in the process. I will write an ELOG on how to do a clean solder of the SMA connectors to the RF cable shortly for future reference.
Coming back to the 40m we rerouted the REF DET cable from near the 1Y1 rack and into the POY table. This job was done mostly by Eric. We were also unable to locate a torque wrench to tighten the cable at the PD’s end and had to leave it finger tight. Eric is planning to buy a new torque wrench as we will need it often.
Also, I cross checked the SwithList located at /users/alex.cole/switchList with the RF switch connections at 1Y1 rack and turns out it is consistent, except that at CH2 of the first switch where MC REFL was to be connected, there is a unlabeled cable. It might belong to the correct PD, but must be made sure of. The rest of the channels that are not mentioned in the list were unconnected on the RF switch.
Now instead of disconnecting REFL 33 to make measurements with the NA, we had to take out AS55 from the RF switch, as the former was very hard to remove without the torque wrench. Then Eric removed the optical fiber which was illuminating the AS55 (AS table) from its mount to hook it up to the power meter. But then we were not sure of how to operate the Laser diode controller (LDC 3744C) and decided to leave stuff as it is and continue either tomorrow or on Monday. Right now we closed the optical fiber of AS55 with a cap and it remains unmounted. The RF cables of REF DET and AS55 were left hanging near the 1Y1 rack. |
10058
|
Wed Jun 18 15:25:06 2014 |
Nichin | Update | Electronics | BBPD Transimepedence plot |
Motivation:
To measure the transimpedence of the Broadband photodiode (D1002969-v8), using a New focus photodiode (1611) as reference. The amplitude modulated Jenne Laser (1.2mW) was used @20mA
The steps involved in getting the transimpedence:
Acquiring data
- The following conditions were set on Network Analyzer Agilent 4395:
1) Frequency sweep range: 500KHz to 300 MHz.
2) Number of Points sampled in the range: 301
3) Type of sweep: Logarithmic
- Set the NA to give the corresponding transfer function value (output of BBPD over output of 1611) and also Phase response in degrees.
- Save the data into floppy disk for processing on the computer.
Plotting
- The matlab code attached (Trans_plot.m) will then give plots for the absolute values of transimpedence in V/A.
- Logic involved in the code will be presented clearly in a separate Elog.
Results
The Plots of transimpedence obtained are attached. The data and matlab code used is in the zip file.
The transimpedence of Broadband photodiode (D1002969-v8) was around 1200 - 1300 V/A for most of the range (2), but the value started falling as the frequency approached 200 MHz. |
Attachment 1: BBPD.png
|
|
Attachment 2: BBPD_readings_06-18-2014.zip
|
10059
|
Wed Jun 18 16:44:55 2014 |
Manasa | Update | Electronics | BBPD installed for BEATX detection |
This BBPD is the spare that we pulled out and is installed for ALSX-PSL beat note detection. |
10062
|
Wed Jun 18 18:16:26 2014 |
Nichin | Update | Electronics | Changes to the PD frequency response measurement system |
[Nichin, Eric G, Koji]
Continuing out work from elog:10037, we wanted to check if the frequency response of AS55. Having figured out exactly how to use the Laser diode controller (LDC 3744C), we hooked up a fiber power meter to the optical fiber illuminating AS55 (that we disconnected from its mount last Friday ) and raised up the current to 150mA to get almost 0.8mW power reading.
When aligning the fiber to illuminate the PD, we found that the beam was pretty wide. So we pulled out the collimator and tweaked it to get a focused beam. The fiber was mounted back and was aligned to get a maximum DC reading. The multimeter readout 30mV finally. Taking the transimpedence as 200ohm approx., the hot current is about 1.5mA.
Network analyzer was now connected to the modulation input of the laser and the RF output from REF DET and AS55 (inputs to RF switch at rack 1Y1) were connected as input to measure the transfer function. We got just noise on the scope of NA. So, then we tried REFL33 as the Input and still got nothing (We were also not sure if this PD was properly illuminated, we did not check). However the REF DET was giving a nice response on the scope. Turns out all the PDs were disconnected form the Demodulator (D990511) on rack 1Y2.
On closer inspection the RF cable between domodulator and RF switch that was labelled AS55 had a loose SMA connector at the switch end. I will have to fix that tomorrow . For the time being Koji connected the cable labelled REFL33 to the AS55 demodulator and we finally got a response form the AS55 PD on the NA. However no readings were recorded. The power supply to REF DET was turned off in the end as Eric G claimed that it has been ON for almost a year now, which is not a good thing. Also, we removed the modulation input from NA to the diode laser and terminated the input with a 50ohm terminator.
We planned to pull out and check each and every RF cable (especially the SMA ends for faulty soldering and loose connections) and fix/ replace them as needed. |
10065
|
Wed Jun 18 21:53:48 2014 |
Koji | Update | Electronics | Changes to the PD frequency response measurement system |
Not "hot" current but "photo" current. Is this my bad!?
It was me who told Nichin that the DC transimpedance was 200Ohm. But according to this entry I checked the RF transimpedance of AS55 before.
In my code, the DC transimpedance was defined to be 50Ohm. If we believe it, 30mV corresponds to 0.6mA.
Quote: |
The multimeter readout 30mV finally. Taking the transimpedence as 200ohm approx., the hot current is about 1.5mA.
|
|
10070
|
Thu Jun 19 12:10:18 2014 |
Akhil | Update | Electronics | Gain plots to Characterize UFC-6000 RF Frequency Counter |
The goal is to characterize the Mini-Circuits RF FC (Model UFC-6000) by plotting Gain Plots.
Work Done:
The sampling rate of the UFC-6000 RF FC is 1s (should look into making the sampling time smaller). So to satisfy Nyquist criterion, the maximum modulation frequency is 0.5 Hz beyond which aliasing effects are seen.
The measurements taken (mentioned in my previous elog) are used to plot Gain vs Modulation frequency for carrier frequencies of 5 MHz and 25 MHz.
Calculations:
A modulated signal can be represented as X(t)= A*sin (Fc*t+D*sin(Fm*t+phase1)) where Fc and Fm are carrier and modulation frequencies respectively and D is the modulation depth.
This signal Y(t) is input to the FC and the output frequencies of the FC are recorded.
Let the output of the FC is Y(t)= A'*sin(Fc*t+D'*sin(Fm'*t+phase2));
Gain = D'/D;
phase = phase2 - phase1;
D' is calculated by subtracting the carrier frequency from the output frequency and calculating the amplitude of the resulting fitted sine wave.
The phase can be calculated if the phase of the input is known(which will be done next).
Plotting the Bode plots would give response of the FC to modulation.
The plots generated will be used to estimate:
1) Transfer Function of FC to be known to build an FOL-PID loop.
2) Quantization noise from Power Spectral Density(PSD) vs Hz.
To Do Next:
1)Calculate the phase difference to complete the Bode plot. This would require interfacing of the ADC on raspberry pi.
2)Estimate the quantization noise from the digital output.
|
Attachment 1: GainVsFreq.png
|
|
10079
|
Fri Jun 20 11:41:18 2014 |
Nichin | Update | Electronics | Transimpedence measurement-BBPD |
EDIT: Please ignore the following data. The revised data and plot are in Elog 10089
Yesterday evening, I conducted the same measurements done in Elog-10059 using the same REF PD (NF 1611) and the same model of BBPD, but on different piece that needed to be checked.
I moved the NA from near rack 1Y1 to the Jenne laser table and back again after the readings were done.
Acquiring data
- The following conditions were set on Network Analyzer Agilent 4395:
1) Frequency sweep range: 1MHz to 300 MHz.
2) Number of Points sampled in the range: 201
3) Type of sweep: Logarithmic
- Set the NA to give the corresponding transfer function value (output of BBPD over output of 1611) and also Phase response in degrees.
- Save the data into floppy disk for processing on the computer.
Results
The Plots of transimpedence obtained are attached. The data and matlab code used is in the zip file.
The transimpedance of Broadband photodiode (D1002969-v8) was around 50kV/A-70kV/A (Unusually high) for most of the range (2), but the value started falling as the frequency approached 200 MHz.
The high impedance might be because the PD is faulty.
|
Attachment 1: BBPD_readings_06-19-2014.zip
|
Attachment 2: BBPD_transimpedence_19thJune2014.pdf
|
|
10082
|
Fri Jun 20 16:36:44 2014 |
Nichin | Update | Electronics | RF cables removed |
[Nichin, Eric G]
As mentioned in Elog 10062, we found RF cables running between demodulators in rack 1Y2 and RF switch in 1Y1 to have bad SMA connectors (No shield / bad soldering / no caps).
we pulled out all the cables belonging to PD frequency response measurement system , 8 in total, and all of them about 5.5m in length.
Their labels read :
REFL33, REFL11, REFL55, AS55, POX11, REFL165, POP22 and POP110.
All of them are now sitting inside a plastic box in the contorl room.
On another note, instead of fixing all the cables ourselves, Steve and Eric G decided to order custom made RF cables from Pasternack as professionally soldered cables are worth it. We have placed an order for 2 cables (RG405-550CM) to check out and test them before we order all of the cables. |