ID |
Date |
Author |
Type |
Category |
Subject |
8983
|
Wed Aug 7 23:40:49 2013 |
gautam | Update | CDS | X-End Green ASS - A first update | I have done some preliminary testing of the X-End Green ASS Servo. I will write a more detailed elog about this soon, but I thought I'd note down the important stuff here.
Yesterday, while we were venting, I aligned the X-arm to the green using the sliders on IFOalign, maximizing the transmission. Then I retook a power spectrum so as to determine the LO frequencies. Jenne pointed out that LO frequencies should not be integers (it usually suffices to append a .098725 or something to the frequency) so I made the necessary changes.
I did a first run of the servo yesterday, and more runs today. Notable points:
- I was able to lock to 00 from a 08 or 09 mode using the PZT sliders
- The green transmission having locked to 00 was ~0.2. I then ran the servo and got it up to ~0.4 and then 0.6 (see time series plot attached). The servo was able to recover this level of transmission after misaligning the steering mirrors using the PZT sliders.
- This was not the optimal transmission level as when Koji moved ETMX a little, the transmission improved.
- The actuators are degenerate. Most of the time, only two of the four servos are doing anything significant. This is probably because of the fact that the two steering mirrors are so close to each other, that moving one or the other produces virtually the same effect. I do however have some cool videos of mode-hopping :)
- The range of actuation of the PZTs is probably not enough to maximize the green transmission from an arbitrary state because of point 4 (i.e. we need to move one mirror in some direction a lot, and move the other a lot to compensate for the change, and the overall gain in input pointing/alignment is marginal). It may be that things will be slightly better at the Y-end. It would also be interesting to see if there is any improvement in the servo performance by dithering the cavity mirrors as opposed to the PZT mirrors.
- To this end, I tried modifying the c1asx model to incorporate an option to dither the cavity mirrors. The plan was to make a second set of LOs in the model that output to ITMX and ETMX suspensions. However, for some reason, when I recompiled the model and restarted it, c1iscex crashed. Parity has now been restored. Note that in order to accommodate the new LOs, I had to make changes to C1SUS, C1RFM and C1SCX as well. I have since removed all my additions, saved, built and installed these models, but have not restarted them (with the exception of C1SCX which restarted when I manually restarted c1iscex).
- The plan tomorrow is to try incorporating cavity dither into the model again. This time, I'll try grabbing the LO-related signals from c1ass directly, as I am not clear why my approach did not work.
More details to follow.

|
8982
|
Wed Aug 7 22:18:43 2013 |
Koji | Update | ASC | ASS update | While Gautam is working on the Xarm green ASS...
The EPICS monitor points for the ASS actuators were added to the ASS model.
This will be used for the offloading the ASS actuations to the alignment biases.
As this modification allowed us to monitor the actuation apart from the dithering,
now we can migrate the ASS actuation to the fast alignment offset on the suspension.
This modification to the offset moving scripts were also done.
|
8981
|
Wed Aug 7 21:52:11 2013 |
Jenne | Update | SUS | SRM coils fine - problem with slow bias slider | [Koji, Jenne]
We have looked a little more at the SRM situation. We aligned the SRM, and then aligned the oplev, so that we had a convenient monitor of the optic's motion.
When we use the _COMM channels, which are the usual ones on the IFO_ALIGN screen, the pitch slider makes pitch motion, but the yaw slider makes the oplev spot move ~45degrees from horizontal.
However, when we use the bias channels that are in the front end model, parallel to the ASC path, pitch moves pitch, and yaw moves pure yaw.
So, we conclude that the SRM coils are fine, and there is something funny going on with the slow part of the actuation.
Koji restarted the slow computer susaux, and burt restored it, but that did not fix the situation. We went inside and looked at all of the ribbon cable connections, and pushed them all in, but that also has not fixed things.
We have been looking at D010001-b, the coil driver board, and we think that's where the summing resistor network between the slow bias slider, and the coil outputs from the fast model exists. (It's not 100% clear, but we're confident that that's what is going on).
Tomorrow, we will pull the SRM's coil driver board, and see if any of the components in the slow slider path, before the summing point, look burned / broken / bad. |
8980
|
Wed Aug 7 19:16:20 2013 |
Jenne | Update | ASC | ASS setting up accelerated more | I have furthered Koji's work, and moved the filter on/off state for all the filter banks also to the burt snapshot.
Turning on the ASS is now much faster than it was originally, with the ezcawrites in series. |
8979
|
Wed Aug 7 15:51:53 2013 |
Alex Cole | Configuration | Electronics | RF Switch Change | For the photodetector frequency response project, our new RF Switch Chassis (NI pxie-1071) arrived today. I took the switches out of the old chassis (Note for future generations: you have to yank pretty darn hard) and put them in the new chassis, which I mounted in rack 1Y1 as pictured.
The point of this new chassis is that its controller is compatible with our control room computer setup. We will be able to switch the chassis using TCP/IP or telnet, aiding in our automation of the measurement of photodetector frequency response. |
Attachment 1: photo_(2).JPG
|
|
8978
|
Wed Aug 7 15:36:29 2013 |
Steve | Update | VAC | TP2 drypump replaced |
TP2's fore line - dry pump replaced at performance level 600 mTorr after 10,377 hrs of continuous operation.
Where are the foreline pressure gauges? These values are not on the vac.medm screen.
The new tip seal dry pump lowered the small turbo foreline pressure 10x
TP2fl after 2 day of pumping 65mTorr |
Attachment 1: forelinesPs.jpg
|
|
8977
|
Wed Aug 7 15:32:37 2013 |
Koji | Update | ASC | ASS setting up accelerated (slightly) | I moved bunch of ezcawrite from the ASS Dither On script to a snapshot file.
This accelerated a half of the "up" time but still switching part is not in the snapshot.
If you find anything wrong with ASS, please notify me. |
8976
|
Wed Aug 7 10:51:49 2013 |
Steve | Update | VAC | Pumpdown at 26hrs | Cold cathode gauge CC1 -h (horizontal) just coming on 9.2e-5 Torr
P2 is the fore line pressure of the maglev. One can see the 4 Torr load during switching over to turbo pumping.
CC4 5e-9 Torr is the performance of the maglev pumping on the RGA only.
The annuloses are not pumped now. They are closed off to see how much outgassing plus leak they have.
Configuration: vacuum normal, annuloses not pumped
Condition: normal
Precondition: 14 days at atm, IOO chamber north door was taken off as a new entrance, the ETMX chamber was not opened.
What is new in the vacuum system: new P1 pirani gauge, gold plated clean allen wrench and ..........what else was dropped?
Note: the wireless laptop did not fail once all day yesterday. I want to give credit to the person who is responsible for this.
|
Attachment 1: 26hr.png
|
|
Attachment 2: 14datm.png
|
|
8975
|
Wed Aug 7 10:09:30 2013 |
Koji | Update | LSC | Arms locked in IR, aligned. IFO at nominal power | I have a concern about the SRM suspension. The yaw alignment bias produces huge pitch coupling.
This could be a connector issue or the rubbing of the mirror on the EQ stops.
We have the photos of the magnets and they were not touching the OSEMs. |
8974
|
Tue Aug 6 19:53:15 2013 |
Jenne | Update | LSC | Arms locked in IR, aligned. IFO at nominal power | [Koji, Manasa, Jenne]
The Y arm was locked in IR, and we saw flashing in the Xarm (Gautam had the Xarm for green work when we began). I checked IPANG, and the beam was beautifully unclipped, almost perfectly centered on the first out of vacuum mirror. I aligned the beam onto the QPD.
We then swapped out the MC Y1 that we use at low power, and replace the usual 10% BS, so that we wouldn't crispy-fry MC REFL. Manasa adjusted the half wave plate after the laser, to maximize the power going toward the PMC. We relocked the PMC, and see transmission of ~0.84, which is at the high side of what we usually get. The beam was aligned onto MC REFL and centered on the WFS, and the MC was locked at nominal power. Koji tweaked up the alignment of the MC, and ran the WFS offset script. I aligned beam onto POP QPD and POP110 coarsely (using a flashing PRC, not a locked PRM-ITMY cavity, so the alignment should be rechecked). The arms have both been locked and aligned in IR....the green beams need to be steered to match the current cavity axis.
The AS beam, as well as REFL and POP, are all coming out of the vacuum nicely unclipped.
Notes: When Koji was aligning the SRM to get the SRC cavity roughly aligned (the AS flashes all overlapping), we noticed that there is some major pitch-yaw coupling. Serious enough that we should be concerned that perhaps some connector is loose, or an actuator isn't working properly. This should be checked.
Moral of the story: Coarse alignment of all mirrors is complete after pump-down and we have IR locked and aligned to both arms at nominal power.
Still to do:
* Restore PRM, align beam onto the REFL PDs.
* Lock PRM-ITMY cavity, align beam onto POP PDs.
* Align AS beam onto AS55.
* Recenter all oplevs.
* Recenter IPPOS and IPPANG at nominal power.
* Start locking!! |
8973
|
Tue Aug 6 16:45:31 2013 |
Steve | Update | VAC | Pumpdown is completed | IFO P1=1mT PSL shutter is opened
|
Attachment 1: 8hrPd76.png
|
|
Attachment 2: 1mT.png
|
|
8972
|
Tue Aug 6 16:36:51 2013 |
gautam | Update | CDS | Choosing LO Amplitudes and Frequencies-revised | I redid the power spectrum measurement for the X-arm green transmission after aligning the arm to green using the ITMX/ETMX Pitch and Yaw sliders on IFOalign.
The Y-axis now reflects the relative intensity noise (RIN), which I obtained by taking the average value of the X-arm green transmission using tdsavg. Based on this measurement, I have now picked four new frequencies at which to try and modulate the PZT mirrors: 10, 19, 34 and 39 Hz. Bandpass filters in the LIA stage have been appropriately modified.
Power Spectrum:

|
8971
|
Tue Aug 6 12:43:23 2013 |
Alex Cole | Configuration | Electronics | AS Table and Rack 1Y1 Additions | For the photodetector frequency response project, I finished the construction of our baluns chassis and mounted it in rack 1Y1 (1st picture).
After consulting with Jenne, I mounted the fiber launcher for REFL165 on the AS table such that it would not cause an obstruction. I aligned the launcher using a multimeter to monitor the DC output of REFL165, but looking at the data I got, it seems I need to do a better alignment/focusing job to get rid of a bunch of noise. |
Attachment 1: photo_1_(5).JPG
|
|
Attachment 2: photo_2_(5).JPG
|
|
8970
|
Tue Aug 6 10:06:29 2013 |
Steve | Update | VAC | Pumpdown is at 500Torr | The pumpdown has started at 8:38am
Manasa was here to confirm good alignment of the IFO
Inner jam nuts of AC bellow were torqued to 45 ft/lbs and door nuts were check on opened chambers.
Annulos were roughed down to 500 mTorr
Oplev servos turned off, PSL and green shutters closed before pumpdown started
|
Attachment 1: befPd.png
|
|
8969
|
Tue Aug 6 07:52:58 2013 |
manasa | Update | General | Ready to pump down | I did an alignment check of the IFO before we start pumping down.
Arms were locked. PRM and SRM were aligned. Green was aligned to the arms for reference during the pump down.
Steve! It's a GO!
MC spot positions:

OSEM:

Oplevs were all centered yesterday and haven't drifted much. So I left them as is.

QPDs (IPPOS aligned from yesterday.

|
8968
|
Mon Aug 5 19:10:01 2013 |
Koji | Update | CDS | Choosing LO Amplitudes and Frequencies | - I suppose the green transmission paths were thoroughly inspected and aligned in prior to the measurement
- Of course it is a BAD idea to use 60Hz as the LO frequency.
- Power spectra should be plotted as "RIN (relative intensity noise)" as the DC of 1 and 100 gives you 100 times different power spectra for the same beam.
Don't forget to subtract the offset from your DC values. |
8967
|
Mon Aug 5 18:48:44 2013 |
gautam | Configuration | endtable upgrade | Full range calibration of PZT mounted mirrors for Y-endtable | I had prepared two more PZT mounted mirrors for the Y-end some time back. These are:
- A 2-inch CVI mirror (45 degree, HR and AR for 532nm, was originally one of the steering mirrors at the X-endtable, and was removed while switching those out for the PZT mounted mirrrors).
- A 1-inch Laseroptik mirror (45 degree, HR and AR for 532nm).
I used the same QPD set-up and the methodology described here to do a full-range calibration of these PZTs. Plots attached. The calibration constants have been determined to be:
CVI-pitch: 0.316 mrad/V
CVI-yaw: 0.4018 mrad/V
Laseroptik pitch: 0.2447 mrad/V
Laseroptik yaw: 0.2822 mrad/V
Remarks:
- These PZTs, like their X-end counterparts, showed evidence of drift and hysteresis. We just have to deal with this.
- One of the PZTs (the one on which the CVI mirror is mounted) is a used one. While testing it, I thought that its behaviour was a little anomalous, but the plots do not seem to suggest that anything is amiss.
Plots:
CVI YAW CVI PITCH

Laseroptik YAW Laseroptik PITCH
|
8966
|
Mon Aug 5 18:18:32 2013 |
gautam | Update | CDS | Choosing LO Amplitudes and Frequencies | In order to decide what frequencies to dither the 4 degrees of freedom (M1-pitch&yaw, M2-pitch&yaw) at, I took the power spectrum of the X and Y-arm green transmission (C1:ALS-TRX_OUT, C1:ALS-TRY_OUT). Plots showing the power spectra are attached. Looking at the power spectra, I would think that for the X-arm, it would be okay to dither at 40, 50, 60 and 70 Hz. In order to check if the piezos could respond to these frequencies, I used my QPD setup and shook the PZTs with a 100Hz, 1Vpp sinusoid, and saw that the spot moved smoothly on the QPD.
As for choosing the modulation amplitude, I did a simplistic approximation assuming that the misalignment only rotates the beam axis relative to the cavity axis, and determined what angle coupled 10% of the power into the next eigenmode. Assuming that this is small enough such that if we are already locked to TEM00, the dither won't kick it up to some higher-order mode, the LO amplitude should be in the range of 30-60 digital counts (determined using the PZT calibration constants determined here. This corresponds to a sine-wave of ~50mV amplitude reaching the PZTs (after HV amplification). I am not sure if this is too small, but according to the PZT datasheet, these platforms are supposed to have a resolution of 0.02 urad, which would correspond to the input signal changing by ~0.1 mV, so this signal should be capable of dithering the tip-tilt.
I have already added band-pass filters centered at these frequencies to the model (with a passband of 5Hz, 2Hz on either side), and low-pass filters to pull out the DC component of the output of the lock-in amplifiers. It remains to tune the gains of the filter stages. These parameters (frequency, amplitude of the LOs) may also have to be changed after tests). Hopefully the PZTs can be plugged in tomorrow, and I can try and make a measurement of the output matrix.
Koji also suggested that it may be good to have a path in the model that feeds back to the PZTs by dithering the cavity mirrors as opposed to the PZT mounted mirrors. I will work on incorporating this into the SIMULINK model (c1asx.mdl) and also into the master medm screen.
Notes:
- The spot size of the X-arm green transmission on the PD was larger than the active surface. I moved the GTRX PD a little back and put in a lens (KPX085, 62.9mm FL, AR.14) in front of the PD, such that the spot is now occupying about 1/4th of the active surface area. The lens was mounted in a Thorlabs LMR1mount, and has been labelled.
- I made a slight change to the SIMULINK model, so as to calibrate the PZT sliders to (approximately) volts (I added a multiplier block that multiplies the slider value by constant value 3267.8). The idea is that we can approximately relate the slider value to tilt, knowing the calibration constant in mrad/V for the PZTs.
Power Spectra of Arm Green Transmission:

|
8965
|
Mon Aug 5 18:02:34 2013 |
manasa | Update | General | Close up list checked |
-
Center beam on all AS optics
-
GET CAMERA IMAGES OF EVERYTHING
- We must get images right before closing, right after closing, etc.
Make sure REFL is clear
- dither PRM, see motion on AP tables
Make sure AS is clear
- dither BS/ITM, see motion on AP tables
Using IPANG/POS pick-off mirrors, center beams on:
IPPOS
IPANG (aligned low in pitch)
Check green alignment in the arms and make sure the transmitted green reaches the PSL table.
-
Check all OpLevs centered, in and out of vacuum
[Jenne, Manasa]
IPANG needed to be re-aligned today. Heavy doors are in place and bolts tight (torque 25 & 45).
Steve! We are ready for pump down!
I will check the IFO alignment once again early tomorrow morning before Steve starts pumping down.
|
8964
|
Mon Aug 5 11:53:45 2013 |
Evan | Update | ISS | CTN Servo - Explicit Requirement and Proposed Servo | I goofed on the transfer function requirement by not giving you the plant transfer function, which looks to be about 0.014 V/V, independent of frequency (PSL:1278). This needs to be compensated for in the electronic transfer function. |
8963
|
Mon Aug 5 10:50:48 2013 |
Steve | Update | VAC | RGA background | RGA background at day 12 of this vent . The maglev is pumping on the rga through VM2
|
Attachment 1: v75bgMd12.png
|
|
8962
|
Fri Aug 2 22:51:10 2013 |
Jenne | Update | General | All vent tasks complete, just need oplev check | [Manasa, Koji, Jenne]
We went into the BS and IOO chambers, and aligned the green beams such that they came out of the vacuum chamber. The idea here was to get the beams at the same height, but slightly offset in yaw. This required moving the Periscope on BS table, PBS in front of that periscope, the Periscope on the IOO table, and 2 steering mirrors on the IOO table after the 2nd periscope. The tables were not releveled, although we have aligned the full interferometer to this situation, so we do not want to touch the tables. The MC spot positions are still consistent with those measured earlier this afternoon, before this work, so I'm not concerned.
We confirmed that both green beams are hitting a good place (centered in pitch, and just left and right of center in yaw) on the mirror in the OMC chamber, and are getting to the center of the first mirror on the PSL table. We then coarsely aligned the beams on the PSL table.
We then relocked and aligned the arms for IR, and checked that the AS beam is centered on the mirrors in the BS chamber, and that the beam is coming out, and to the AS table. I touched the last mirror before PZT3 a small amount in yaw, and then PZT3 in pitch and yaw, until we saw the beam recentered on the first mirror on the AS table. At that point, we were also back to the center of the AS camera (which is good, since Koji had aligned all of that the other day). So, the AS beam is good.
We checked IPPOS, and have centered the beam on all the mirrors, and aligned the beam onto the QPD.
We checked IPANG, by looking through the viewports at the mirrors in the ETMY chamber. We are now centered in yaw, but clipping a bit low. This is what we want, since we always end up drifting high during the pump-down.
We see a nice, unclipped REFL beam on the camera.
We see a beam a little high on the POP camera, but Koji looked on the table with a card, and saw the beam....we just need to do minor alignment on the out of vac mirrors.
We checked again that the green TEM00 beams from both arms come to the PSL table.
We are getting POX and POY out, since we are using them to lock and align the arms.
Manasa and Koji recovered one clean allen key from the bottom of the chambers, but one remains, as a sacrifice to the vacuum gods.
I believe that, with the exception of checking the oplevs and taking photos of PR3, and the green steering optics, we have finished all of our vent tasks. We should do a quickie alignment on Monday, check the oplevs, take some photos, and put on the heavy doors. Pumping can start either Monday afternoon or Tuesday morning.
|
8961
|
Fri Aug 2 21:59:36 2013 |
Charles | Update | ISS | Finalized ISS Schematic (hopefully) | Attached is the finalized schematic. The general circuit topology should remain the same from this point forward, although individual component values are subject to change. I will also be adding some more annotations to ensure everything on the board is clear.
In general, I have finally included all of the correct components (i.e. front panel switches are now actually switches and front panel LEDs are now included). I also added an external 'Boost' switch, which can be used to enable or disable the boosts. The motivation for including this switch is that one might want to test functionality of the ISS without using the 'fancy' RMS detection and triggering circuitry. Additionally, one can disable the boosts when all the circuitry is stuffed in order to troubleshoot, so it essentially grants the board some flexibility in its operation.
I am now working on the PCB layout and I should hopefully have that done next week. |
Attachment 1: ISS_v3.pdf
|
|
Attachment 2: ISS_v3-Power_Reg.pdf
|
|
8960
|
Fri Aug 2 17:50:10 2013 |
Jenne | Update | General | PR3 wedge angle adjusted | [Jenne, Manasa, Koji]
Earlier today, we locked and aligned both the X and Y arms.
I then went into the BS chamber, put on the BS' aperture, and put an aperture along the AS path. (We had Michelson fringes, so I centered the aperture around the fringes. I used one of the brass ruler things that we use to center the beam on ITMs and ETMs, on a riser. I put this aperture at the edge of the BS table, after the AS beam is launched toward the OMC chamber. The idea was to replace PR3 such that I could get the beam back through the BS aperture, and the brass ruler aperture, in hopes that we would see arm flashes, and not have to open the ITMY and ETMY heavy doors.)
I set references on the table so that I could put PR3 back in its original position, then removed PR3 from the chamber.
Steve set up a HeNe for me, that we pointed through the optic. The ghost beam was very high, indicating (as expected) that the wedge was not perfectly horizontal.
I took the suspension off of the cage and laid it down, as I have in the past.
I removed the optic from the suspension, to try to figure out which was the fat vs. skinny side. I noticed that there are very faint marks on the actual fat and skinny sides of the optic. (Mpral - for the LaserOptik mirrors, look for the faint lines that are the full width of the barrel, not the placement of the arrow which marks the HR side). I put the optic back in (HR side toward the back, fat side on the left (as you look at the face of the optic), which is consistent with the picture in the Optical Layout page of the Wiki, near the bottom.) the optic holder ring.
I put the suspension back on the cage, and saw that the HeNe's ghost beam was now nearly horizontal relative to the straight-through beam. Excellent. Also, the pitch balancing didn't seem to change noticably, which I determined was within "poking" distance of where we need it to be.
I put PR3 back onto the BS table, and adjusted it around until I got the beam through both the BS aperture, and the one on the AS path. As usual, this took quite a while, but as soon as I got through both of those apertures (really at the same place, not close to being through them, but as close as I could tell by eye - this is what took forever), Koji and Manasa saw flashes in the Yarm! Yay!
Since I had to move PR3 in angle a tiny bit, I reset the references, then dogged down PR3. We still had flashes, this time in both arms, so we closed up the light doors.
We have now locked and aligned both arms in IR after the adjustment of PR3, and see both arms' green at 01 or 02. We are about to start checking the green positioning on the periscopes. We will also need to check the AS path, as well as IPPOS and IPANG before we close up. We see REFL on the camera.
Separately - Manasa remembered that 2 clean things were dropped yesterday - a screw, and an allen key. Since they're both Clean, we're not too worried, although she thinks a long-armed person may be able to reach the allen key. |
8959
|
Thu Aug 1 22:58:45 2013 |
Charles | Update | ISS | CTN Servo - Explicit Requirement and Proposed Servo | In PSL elog 1270, Evan elucidated the explicit requirements for the CTN ISS board. Essentially, the transfer function of the ISS should be something like:
TF_mag = (Unstabilized RIN) / (Calculated RIN Requirement)
I took Evan's data and did exactly this. I then designed a servo (using the general design I proposed here) to meet this requirement with a safety factor of ~10. By safety factor, I mean that if the ISS operates exactly according to theory, it should suppress the noise by a factor of 10 more than what is necessary/set out by the requirement. Below is a plot of the loop gain obtained directly from the requirement (the above expression for TF_mag) and the transfer function of the servo I am proposing.

I don't have the actual schematics attached as I was working with a LISO file and have yet to update the corresponding Altium schematic. The LISO file is attached and I will add the schematics later, although one can reference the second link to find a simple drawing. |
Attachment 2: CTNServo_v3.fil
|
# Stage 1
r R31 1.58k in n_inU3
op U3 ad829 p_inU3 n_inU3 outU3
r R35 1k p_inU3 gnd
c C33 1u p_inU3 gnd
c C32 10n n_inU3 outU3
r R34 158k n_inU3 outU3
# Stage 2
#r R41 15.8 outU3 n_inU4U5
... 24 more lines ...
|
8958
|
Thu Aug 1 22:49:31 2013 |
manasa | Update | General | Green status after PR3 flipping investigated | [EricQ, Koji, Manasa]
We opened the BS chamber to check the status of the green beams. The X green has 3 steering mirrors before they hit periscope1 and the Y green transmits through all the optics giving no way to steer it.
We agreed to start fixing the Y green. The wedge angle of PR3 is steering the transmitted beam away in both pitch and yaw. Since we are restricted only to yaw movement (done by moving the periscope), we want he wedge angle to be oriented in the yaw as well.
Right now, the wedge is oriented at about 20-30 deg off (The mark on the side of the mirror does not indicate the wedge). So we see a pitch as well as yaw misalignment in the transmitted beam. The pitch misalignment is making the beam fall off the mirrors in periscope2.
We have decided to get the wedge angle set right for PR3 and redo the alignment for IR. Once we are aligned for the IR, we will modify the green layout. |
8957
|
Thu Aug 1 21:28:09 2013 |
gautam | Update | CDS | Slow channels set-up in ALS | The following slow channels have been added and are now being recorded by FB.
C1:ALS-X_OVEN_TEMP
C1:ALS-Y_OVEN_TEMP
C1:ALS-BEATX_FREQ
C1:ALS-BEATY_FREQ
Details:
In order to integrate the data collected by the Raspberry-Pi from the Y-end doubling oven temperature controller and also the data from the frequency counter which will be hooked up to monitor the beat frequency, Koji helped me set up some slow EPICS record channels (in ALS as we felt this was most appropriate). The procedure for setting up slow channels was as follows (virtually identical to what is detailed in this elog:
- Add the channel names to the file C0EDCU.ini (path = /opt/rtcds/caltech/c1/chans/daq/C0EDCU.ini).
- Make a database (.db) file so that these channels are actually recorded (path = /cvs/cds/caltech/target/c1aux/als.db).
- Restart framebuilder.
- Verify that the channels indeed exist and can be read and written to using ezcaread and ezcawrite.
I will now integrate these channels into my scripts, and make some simple MEDM screens.
|
8956
|
Thu Aug 1 20:58:56 2013 |
gautam | Update | CDS | New model for endtable PZTs-MEDM Screens setup |
I have made some minor changes to the model, made all the MEDM screens, and linked monitors on these to the appropriate channels. I have borrowed heavily from the C1ASS MEDM screens (particularly for the small filter modules-it was convenient to just copy and paste an existing module, and edit the channel names using EMACS/GEDIT), and have edited these to suit the needs of this servo. Some features:
- The feedback signal (only the output of the servo to the PZTs, plus any contribution from the on-screen sliders, and not including the LO output) is monitored with both a slow (using CDS_EPICS_OUTPUT block from the CDS_PARTS library) and fast channel (using Test Point from the same library). The idea is that it would be useful to know the output to the PZTs such that if coarse adjustment ever needs to be done at the endtable, the PZTs can be restored to the middle of its operating range by means of the sliders.
- Sliders are incorporated into the master screen for adjusting the output to the PZTs. There are text-input fields below the sliders as well, which control the same channel.
- I have removed the 4 remaining excitation points to the DAC set up in C1SCX, and have relocated them to channels 12-15 of the DAC in C1ASX.
I think I am now ready to take some measurements and try and optimize this servo. There is no green transmission at the PSL table at the moment, so not much can be done, though the first step would be to take the power spectrum of the error signal, which would help me decide the appropriate frequencies for the LOs. I would then have to add the appropriate filters to the model. The last, and most difficult step, would be the measurement of the output matrix, though Koji has given me some ideas about how this measurement can be done. I also have a template script ready, though I will only finalise this after optimising the servo and running it a couple of times manually.
Attached are screenshots of the MEDM screens.

|
8955
|
Thu Aug 1 18:55:20 2013 |
Alex Cole | Update | Electronics | Preliminary Photodetector Frequency Reponse Measurements |
Quote: |
[Eric, Alex]
We used our setup from yesterday (elog #8940) to measure transimpedance measurements for AS55, REFL11, REFL33, and REFL55, using our Newport 1611 FC-AC as reference. We connected the fibers to their respective telescopes such that the beams focused on their photodetectors, using a multimeter to maximize photodetector DC output. Plots are attached. At first glance, the poles seem to be where they're supposed to be.
Note that the procedure used today is similar to what the eventual automated procedure will be. The main differences are (1) The RF Switch will be used rather than manual switching (2) NWAG4395A will be used to collect data rather than netgpibdata (3) Data will be fit using vectfit4.m and compared to some canonical set.
|
[Alex, Eric]
Today I spent some time mounting the launcher and performing the same data collection for POX11. I think I still need to focus the launcher so the photodetector gets a good signal, but the data from today wasn't too bad. Additionally, I worked on matlab scripts to improve PDFR data analysis.
This time I collected data from the network analyzer using NWAG4395A in the netgpibdata directory. The advantage of this is that the computer tells the network analyzer to perform the sweep as well as retrieving the data.
For analysis, I improved my implementation of vectfit4.m so that it focuses in on the particular photodetector's predicted peaks and thus ignores much of the noise, giving a better fit. The raw data is the red circles in the 2nd attachment, while the fit is the blue line. I also had the program return the frequency value of the peak. For POX11, this was 1.106e+07 Hz.
I also finagled copies of existing programs to enable one to plot multiple transfer functions on the same axes. This function is /users/alex.cole/plottwo.m. I will eventually use this to compare new data to some canonical data so that we may monitor photodetector performance over time.
The eventual plan is to generate two plots per photodetector, one of which will compare new data to the canonical set, the other of which will show the fit of the data. Both will have subplots that zoom in around regions of interest (known peaks and notches), and the plot which displays the canonical set will also have Q's of peaks and their locations. |
Attachment 1: POX11.jpg
|
|
Attachment 2: POX11fit.jpg
|
|
8954
|
Thu Aug 1 16:33:39 2013 |
Sujan | Update | PEM | Guralp2 seismometer installed at North side of POX table | 1)Power to the seismometers were turned down,
2)Guralp2 was moved to North side of POX table
3)Guralp2 was aligned in N-s Direction and leveled before connecting
4)Power to seismometers was turned on once Guralp2 was connected

|
8953
|
Thu Aug 1 16:03:52 2013 |
Steve | Update | General | reopen BS & IOO |
Quote: |
[Koji, Manasa]
We missed to check that we had the green transmitted to the PSL after flipping the SRC and PRC folding mirrors.
There is no green transmission reaching the PSL even after locking the arms to green.
We should fix this tomorrow. The BS heavy door should come off.
Steve! Do not start pump down tomorrow !
|
We closed the chambers last night with heavy doors and reopened it today. |
Attachment 1: IOOnorth.jpg
|
|
8952
|
Thu Aug 1 15:28:44 2013 |
gautam | Update | CDS | New model for endtable PZTs-problem solved |
Quote: |
I don't know what's going on here (why the channels are white), and I don't yet have a suggestion of where to look to fix it but...
Is there a reason that you're making a new model for this? You could just use and existing model at c1iscex, like the c1scx, and put your stuff in a top-names block. Then you wouldn't have to worry about all of the issues with adding and integrating a new model.
|
Koji just fixed this.
It seems that the new model's channels were not automatically added to the master file in the framebuilder (/opt/rtcds/caltech/c1/target/master). Adding the following two lines to the master file fixed the problem;
/opt/rtcds/caltech/c1/chans/daq/C1ASX.ini
/opt/rtcds/caltech/c1/target/gds/param/tpchn_c1asx.par
The box is now green. It looks like C1ASX.ini is created automatically in /opt/rtcds/caltech/c1/chans but the master file needs to be manually edited. The channels are now showing up on dataviewer etc. I have updated the information on the wiki page.
|
8951
|
Thu Aug 1 15:06:59 2013 |
jamie | Update | CDS | New model for endtable PZTs |
Quote: |
I have made a new model for the endtable PZT servo, and have put it in c1iscex. Model name is c1asx. Yesterday, Koji helped me start the model up. The model seems to be running fine now (there were some problems initially, I will post a more detailed elog about this in a bit) but some channels, which are computer generated, don't seem to exist (they show up as white blocks on the MEDM GDS_TP screen). I am attaching a screenshot of the said screen and the names of the channels. More detailed elog about what was done in making the model to follow.

Channel Names:
C1:DAQ-DC0_C1ASX_STATUS (this is the channel name for the two leftmost white blocks)
C1:DAQ_DC0_C1ASX_CRC_CPS
C1:DAQ-DC0_C1ASX_CRC_SUM
|
I don't know what's going on here (why the channels are white), and I don't yet have a suggestion of where to look to fix it but...
Is there a reason that you're making a new model for this? You could just use and existing model at c1iscex, like the c1scx, and put your stuff in a top-names block. Then you wouldn't have to worry about all of the issues with adding and integrating a new model. |
8950
|
Thu Aug 1 13:09:17 2013 |
gautam | Update | CDS | New model for endtable PZTs-procedure |
These are roughly the steps I followed in setting up the new model for the endtable PZT servo - C1ASX.
Simulink model:
I made a SIMULINK model of the servo, using MATLAB R2013a. The path to the model is /opt/rtcds/caltech/c1/userapps/release/isc/c1/models/c1asx.mdl. I am listing the parameters set on the CDS_PARAMETERS block:
- host = c1iscex
- site = c1
- rate = 16k
- dcuid = 44 (which I chose after making sure that this dcuid was not used on this list which was last updated end Feb 2013)
- specific_cpu = 5 (again chosen after checking the available CPUs in the above list).
- adc_Slave = 1
- shmem_daq = 1
- no_rfm_dma = 1
- biquad = 1
Making, Compiling and Installing the Model:
After saving the model, I ssh-ed into c1iscex and ran the following commands:
rtcds make c1asx - this gave me a whole bunch of errors initially, which I tracked down to a naming problem in some of the from and goto flags: there should not be any spaces.
rtcds install c1asx
rtcds start c1asx - this gave me an error which said something like 'can't start/stop model.' Koji pointed out that given that a new model is being started, there is an additional step involved, which is to add the model name to the rtsystab file (this is located at /diskless/root/etc/rtsystab on framebuilder, and is mirrored in the various computers. It would be advisable to make sure that the changes are mirrored in the corresponding file on the computer in which the new model is being installed).
After adding the model name to the rtsystab file, I tried running rtcds start c1asx again. This time, no errors were output, but the model was not up and running as verified by looking at the C1:ASX_GDS_TP medm screen.
Debugging
Koji suggested making a simple model (1 CDS parameters block, 1 ADC block and 2 filter modules, appropriately terminated) and see if that starts up, which it did. I then tried adding my servo minus the DAC block and recompiled and restarted the model. This too worked fine. I figured that the next logical step would be to add the DAC block to the model, and restart the model. But when I tried this, c1iscex crashed .
Jenne helped in restoring things to a working state (we reverted the c1asx model to just 2 filter modules, and went to the X-end and restarted the computer. This did not work the first time so I went back in and restarted it again, at which point we were able to ssh into c1iscex again and restart the four models running on it).
Since Manasa and Koji were working on getting things set up for the pumpdown,I did not try anything again till later in the evening, when Koji helped in debugging the problem further. In the meantime, at Jenne' suggestion, I made the model once again in MATLAB R2010b. In the evening, when I tried restarting the model, Koji suggested that the DAC channels in c1asx may be used by other models, at which point I realised I had set up excitation points on channels 8 through 15 of the DAC in c1scx (detailed here) in order to test the hardware at 1X9. I removed the excitation points from channels 8-11 of the DAC block in c1scx (these are the ones used in c1asx), and recompiled and restarted c1asx (using the above sequence of commands). I then tried recompiling and starting c1asx once more, and this time, it worked . At least, the GDS_TP screen suggests that the model is running alright, except for the fact that some computer generated channels seem to be missing. This problem is unresolved for now, and probably has something to do with the fact that C1ASX channels do not appear in Dataviewer.
I do not think this has to do with restarting framebuilder (I did the usual telnel fb 8088 followed by shutdown). In any case, I have added the new model to the CDS_FE_STATUS screen, and will continue to debug the same. I have also got a template medm screen (work in progress) which I will elog about soon as I get it done.
Note to self: There are 4 more excitation channels still hooked up to the DAC (channels 12-15) in the c1scx model. I plan to remove these and put them in c1asx.
|
8949
|
Thu Aug 1 12:12:35 2013 |
gautam | Update | CDS | New model for endtable PZTs | I have made a new model for the endtable PZT servo, and have put it in c1iscex. Model name is c1asx. Yesterday, Koji helped me start the model up. The model seems to be running fine now (there were some problems initially, I will post a more detailed elog about this in a bit) but some channels, which are computer generated, don't seem to exist (they show up as white blocks on the MEDM GDS_TP screen). I am attaching a screenshot of the said screen and the names of the channels. More detailed elog about what was done in making the model to follow.

Channel Names:
C1:DAQ-DC0_C1ASX_STATUS (this is the channel name for the two leftmost white blocks)
C1:DAQ_DC0_C1ASX_CRC_CPS
C1:DAQ-DC0_C1ASX_CRC_SUM |
8948
|
Wed Jul 31 21:12:05 2013 |
manasa | Update | General | Pump down called off | [Koji, Manasa]
We missed to check that we had the green transmitted to the PSL after flipping the SRC and PRC folding mirrors.
There is no green transmission reaching the PSL even after locking the arms to green.
We should fix this tomorrow. The BS heavy door should come off.
Steve! Do not start pump down tomorrow ! |
8947
|
Wed Jul 31 17:02:17 2013 |
Alex Cole | Update | Electronics | Preliminary Photodetector Frequency Reponse Measurements | [Eric, Alex]
We used our setup from yesterday (elog #8940) to measure transimpedance measurements for AS55, REFL11, REFL33, and REFL55, using our Newport 1611 FC-AC as reference. We connected the fibers to their respective telescopes such that the beams focused on their photodetectors, using a multimeter to maximize photodetector DC output. Plots are attached. At first glance, the poles seem to be where they're supposed to be.
Note that the procedure used today is similar to what the eventual automated procedure will be. The main differences are (1) The RF Switch will be used rather than manual switching (2) NWAG4395A will be used to collect data rather than netgpibdata (3) Data will be fit using vectfit4.m and compared to some canonical set. |
Attachment 1: REFL11.jpg
|
|
Attachment 2: REFL33.jpg
|
|
Attachment 3: REFL55.jpg
|
|
Attachment 4: AS55.jpg
|
|
8946
|
Wed Jul 31 10:27:00 2013 |
Steve | Update | VAC | instrument grade air is discontinoued | We have been using Air Liquide Instrument Grade Air for venting the 40m vacuum envelope. It is no longer available.
The replacement is Alphagas 1, total hydrocarbon <0.1% ppm
LINK TO Alphagaz

Air Liquide > ALPHAGAZ™ Specifications

ALPHAGAZ™ Specifications
ALPHAGAZ™ range of Pure Gases for Analysis: two purity levels tailored to cover all your analytical needs.
Because each analytical application is sensitive to critical impurities that can affect your results, tests in our research center and analyzer manufacturers' studies confirm that the main impurities are moisture, oxygen and hydrocarbons.
Based on these critical impurities, Air Liquide has developed ALPHAGAZ™ brand with a Maximum Impurity Level (H2O, O2, CnHm).
- The Maximum Impurity Level is the same across the whole product range. This constitutes a guarantee of quality with regard to major critical impurities and contamination tracers. ALPHAGAZ™ 2 offers an even greater guarantee in terms of purity (CO, CO2, H2).
- The Minimum Total Purity is a prerequisite for product quality, but we do not stop there: Air Liquide selects and controls its gas sources by defining specifications regarding impurities that are critical for your analysis or application.
ALPHAGAZ™ 1 |
ALPHAGAZ™ 1 |
ALPHAGAZ™ 2 |
ALPHAGAZ™ 2 |
ALPHAGAZ™ AUTO IV |
ALPHAGAZ™ CO2 SFC |
He, N2, H2,
Ar, O2 |
Air |
He, N2, H2,
Ar, O2 |
Air |
Air*, N2, O2 |
CO2 |
Specifications (ppm vol): |
H2O < 3 ppm
O2 < 2 ppm
CnHm < 0.5 ppm |
H2O< 3 ppm
CnHm< 0.1 ppm
CO2 < 1 ppm
CO < 1 ppm |
H2O < 0.5 ppm
O2 < 0.1 ppm
CO2 < 0.1 ppm
CO < 0.1 ppm
CnHm < 0.1 ppm
H2 < 0.1 ppm |
H2O < 0,5 ppm
CnHm < 50 ppb
CO2 < 0,1 ppm
CO < 0,1 ppm
NOx < 10 ppb *
SO2 < 10 ppb* |
CnHm < 0.2 ppm
CO2 < 0.5 ppm
CO < 0.5 ppm
NO/NOx < 0.1 ppm |
H2O < 5 ppm
O2 < 2 ppm
CnHm < 2 ppm
NVOC* < 2 ppm
CO < 5 ppm
NO + NO2 < 2 ppm
S Total < 1 ppm
H2 < 0.5 ppm |
|
20 ± 1 % O2 (Europe)
21 ± 1% O2 (North America) |
|
20,9 ± 1 % O2
*: Detection limit |
*20 ± 1 % O2 |
*NVOC: Non Volatile Organic Compounds |
Minimum Total Purity |
|
8945
|
Wed Jul 31 09:51:43 2013 |
Steve | Update | safety | Sujan receives safety training | Sujan got 40m specific basic safety training this morning. |
Attachment 1: IMG_0031.JPG
|
|
8944
|
Wed Jul 31 00:59:58 2013 |
manasa | Update | General | ITMX chamber vented/Closing plan | My Plan for closing:
Today: I will work on the ITMX oplev situation today and go through the vent close-up checklist as far as I can get.
[Alex, Sujan, Manasa]
The ITMX oplev steering mirrors were laid out such that they were out of the way of the BS oplev leakage. But the halo associated with the He-Ne laser does exist even now. I conclude that this is something that can be dealt with after we pump down as well. So I did not change the ITMX oplev optics on the POX table.
BS, PRM, ITMY and SRM oplevs were aligned and centered.
We want to do IPANG and IPPOS alignment when the IFO is aligned satisfactorily and right before we put the heavy doors.
The arms were aligned and ASS'd before I went in to fix the oplevs. I haven't done anything but deal with the oplevs tonight. So I am being lazy by assuming the alignment is still good and calling it a night.
Tomorrow: We will do the final alignment check for the arms, PRC and SRC with the light doors on. Check IPANG and IPPOS. The access connector and heavy doors should go in place late afternoon.
Thursday: We will start pumping down early in the morning on Thursday. |
8943
|
Tue Jul 30 19:44:05 2013 |
gautam | Configuration | endtable upgrade | Second mirror glued to PZT and mounted |
I have glued a fourth mirror to a PZT (using superglue) and inserted it into a modified mount. This is to be used together with the 1-inch Laseroptik mirror I had glued a couple of weeks back at the Y-endtable. I will be calibrating both these mirrors tonight such that these are ready to put in as soon as we are pumped down.
The mirror was one of those removed from the X-endtable during the switch of the steering mirrors. It is a CVI 2-inch mirror, with HR and AR coatings for 532 nm. |
8942
|
Tue Jul 30 19:40:47 2013 |
gautam | Configuration | endtable upgrade | DAC-PZT Driver Board Output Signal Chain Tested |
[Alex, Gautam]
The signal chain from the DAC output to the output of the PZT driver board (including the HV supply) has been verified.
I had installed the two boards in the eurocrate yesterday and laid out the cables from 1X9 to the endtable. The output of the AI board had been verified using the monitor port on the front panel, but the output from the PZT driver board was yet to be checked because I had not connected the HV supply yesterday.
When I tried this initially today, I was not getting the expected output from the monitor channels on the front panel of the PZT driver board, even though the board was verified to be working. Alex helped debug the problem, which was identified as the -15V supply voltage not making it onto the board.
I changed the slot the board was sitting in, and used a long screw to bolt the board to the crate. Both the AI board and the PZT driver board seem to be slightly odd-sized, and hence, will not work unless firmly pushed into the eurocrate and bolted down. This would be the first thing to check if a problem is detected with this system.
In any case, I have bolted both boards to the eurocrate, and the output from the PZT driver board is as expected when I sent a 10Vp sine wave out from the DAC. I think the cables can now be hooked up to the PZTs once we are pumped down. |
8941
|
Tue Jul 30 18:56:31 2013 |
manasa | Update | General | ITMX chamber vented/Closing plan | [Koji,Manasa]
We removed the ITMX heavy door to fix the oplev situation.
My Plan for closing:
Today: I will work on the ITMX oplev situation today and go through the vent close-up checklist as far as I can get.
Tomorrow: We will do the final alignment check tomorrow with the light doors. The access connector and heavy doors should go in place late afternoon.
Thursday: We will start pumping down early in the morning on Thursday. |
8940
|
Tue Jul 30 16:21:46 2013 |
Alex Cole | Update | Electronics | Photodetector Input Modulation | [Eric, Alex]
We successfully used our system to modulate the input to a single photodetector. The RF Out of the network analyzer went to the Mod In of our laser, which was operating at 98 mA. The laser's output was sent to our 1x16 optical splitter. This provided input signals for both our reference detector and AS55. Our reference detector's output was sent to the network analyzer's R input, while the AS55's output was sent to the network analyzer's A input.
We still need to work out the specifics of how the modulation works. Specifically, we want to look at the amplitude of the network analyzer's output. Additionally, we may have been saturating our reference detector, causing noise problems. |
8939
|
Tue Jul 30 15:38:24 2013 |
manasa | Update | General | ITMX oplev |
Quote: |
I wonder what optics is causing the halo on the oplev beam.
It this comes from any uncoated lens (or similar) it should be identified.
|
We identified this to be coming from the uncoated concave lens that we have right after the He-Ne laser which should also be replaced in addition to the other problems with the oplev. |
8938
|
Tue Jul 30 13:41:35 2013 |
Koji | Update | General | ITMX oplev | I wonder what optics is causing the halo on the oplev beam.
It this comes from any uncoated lens (or similar) it should be identified. |
8937
|
Tue Jul 30 11:30:55 2013 |
manasa | Update | General | ITMX oplev |
Quote: |
[Koji, Manasa, Sujan]
The ITMX oplev was also checked from the windows without removing the door.
The beam is actually hitting the right positions of the steering mirror and the test mass
although there are lots of stray beams because of the BS oplev and some halo assciated with the ITMX He-Ne laser(!?).
|
[Steve, Gautam, Manasa]
While we checked the ITMX oplev situation yesterday, we found that the beam hitting the ITM and the in-vacuum steering mirrors had a halo around them. We used the set of irides in the path of the ingoing beam and cut the stray light around the beam. This reduced the intensity of the halo around the mirrors. We noticed that the halo accounted for 2000 counts of 6500 at the oplev QPD. We tried changing the laser and this did not make the situation any better.
Also there are a couple or more strong stray beams from the BS oplev.
Thoughts:
I suspect that the BS oplev leakage is messing with the ITMX oplev. Why?? We have been seeing the breathing of transmitted beam from the X arm cavity and the shadow sensor readings have been bigger than usual since the last vent (when we changed the BS oplev path). Also, the ingoing oplev beam is close to clipping at the PR2 stack.
I think it would be best to open the ITMX chamber and modify the in-vac steering mirror layout of the oplev. |
8936
|
Tue Jul 30 00:21:15 2013 |
Koji | Update | General | Vent progress | [Koji, Manasa, Sujan]
Tomorrow we'll make final checks of the optics inside the chamber.
Then we will pump down the chamber.
- AS beam steering
The AS beam was aligned on the steering mirror (Piezo Jenna i.e. PZT4).
At this point we already found the beam on the AS CCD.
After checking the spots in the OMC chamber, we determined the beam is not clipped inside.
On the AS table, there is a weird thick lens with short focal length (like f=37.5mm), this actually clipped the beam significantly
[I take the blame for it from the last vent. The lens was installed thinking it would be temporary which it was not. -Manasa].
We removed this lens and put the CCD camera back so that the spot on the camera became smaller.
- SRM debiasing
First of all, the bias of the SRM was moved to zero.
Then the tower was rotated so that the beam is aligned in yaw.
After fine alignment of the SRM, the yaw slider was changed from -5.8 to 1.2.
The OSEMs on the SRM suspension was also checked. The OSEM insertions were
adjusted so that they show the half values of the open calibration
(the values can be found on the 40m wiki)
SRM / ITMY / BS / PRM oplev paths in the chamber were checked.
The beams were aligned on the steering mirrors and the QPDs on the optical table.
The ITMX oplev was also checked from the windows without removing the door.
The beam is actually hitting the right positions of the steering mirror and the test mass
although there are lots of stray beams because of the BS oplev and some halo assciated with the ITMX He-Ne laser(!?).
- Checking the alignment
After we closed the light doors, we aligned the MC, the arms and the recycling mirrors.
Now the everything is aligned. |
8935
|
Mon Jul 29 21:57:45 2013 |
gautam | Configuration | endtable upgrade | Hardware installed at 1X9 | The following hardware has been installed on rack 1X9;
- KEPCO high voltage power supply (kept in a plastic box at the bottom of the rack, with the 3m SMA cable carrying 100V running along the inside side wall of the rack). The HV supply has not been connected to the driver board yet.
- AI board D000186 installed in top eurocrate. The board does not seem to fit snugly into the slot, so I used a longish screw to bolt the front panel to the eurocrate.
- PZT driver board D980323 installed in top eurocrate adjacent to the AI board.
- Six 11m SMB-LEMO cables have been laid out from 1X9 to the endtable. I have connected these to the PZT driver board, but the other end (to the PZTs) is left unconnected for now. They have been routed through the top of the rack, and along the cable tray to the endtable. All the cables have been labelled at both ends.
I have also verified that the AI board is functional in the eurocrate by using the LEMO monitoring points on the front panel.
The driver boards remain to be verified, but this cannot be done until we connect the HV supply to the board.
|
8934
|
Mon Jul 29 16:15:13 2013 |
Koji | Update | VAC | AC removed | Prior to the access connector removal, Manasa and I aligned the IFO mirrors.
The arms were locked and aligned by ASS. |
|