ID |
Date |
Author |
Type |
Category |
Subject |
12909
|
Mon Mar 27 16:01:55 2017 |
Steve | Update | PEM | X arm AC set to 68F |
The X arm air conditioner was not regulating properly. The arm temp was warmer than usual. I requested thermistor calibration.
The mechanic reset the thermostate to 68F last week. It was 70-71F before.
The ETMX oplev laser now running 4 C lower at 30 C inside the enclousure.
The ETMX optical table top is 5 C cooler at 21 C
The ETMX concrete wall temp 20 C at 9am with flow bench on.
ETMY conrete wall temp 23 C at 9am |
8703
|
Thu Jun 13 22:31:12 2013 |
Manasa | Summary | Green Locking | X arm ALS | Stabilized ALS and beat frequency sweep realized.
Procedure:
1. Enable appropriate filter modules and set appropriate servo gains.
Servo module
|
Filters |
Gain |
C1:ALS-BEATX_FINE_I |
FM2 FM3 |
1.0 |
C1:ALS-BEATX_FINE_Q |
FM2 FM3 |
1.0 |
C1:ALS-BEATX_FINE_PHASE |
FM1 |
300.0 |
C1:ALS-BEATX_FINE_OFFSET |
- |
1.0 |
C1:ALS-XARM |
FM4 FM5 |
-0.25 |
C1:ALS-ETMX |
- |
1.0 |
2. Clear history of C1:ALS-BEATX_FINE_PHASE
3. Enable the servo loop. I had set limits on the servo loop and ramp time for gain switching so that I don't kick the ETMY hard.
Gains were decided such that the error signal C1:ALS-BEATX_FINE_PHASE_OUT was minimized.
4. Beat frequency sweep is realized by stepping up on C1:ALS-BEATX_FINE_OFFSET_OFFSET (from 0 to 2100 in this case).
Video1 shows the difference that can be seen at the RF spectrum analyzer when ALS is enabled.
Video2 shows the beat frequency sweep as seen on the spectrum analyzer.
I could not get 'getdata' to work as I wanted. So I have attached the error signal trend before and after the ALS servo loop is enabled.
Thank you Jenne for helping retrieve more sensible data!
More info:
The beat note is very strong and we can clearly see its harmonics as well. Attached is the picture showing the several harmonics.
________________________________
Peak frequency(MHz) Power(dBm)
________________________________
47 -2.77
93.5 -27.56
139 -32.75
185.4 -45.64
231.9 -57.10
278.4 -59.42
________________________________
To do:
1. Obtain IR resonance.
2. Check the digital anti-whitening filter after the beatbox.
3. The effect of the harmonics should be figured out.
4. Write scripts to enable ALS and findIRresonance. |
Attachment 1: IMG_0659.JPG
|
|
Attachment 2: ALS_stabilized_13June2013.png
|
|
8708
|
Fri Jun 14 07:06:19 2013 |
Koji | Summary | Green Locking | X arm ALS | It's nice that we are now able to scan the cavity again. We got close to PRMI+one arm one step further.
The calibration of the scan frequency and the evaluation of the in-loop/out-of-loop error signal in terms of (Hz/rtHz) would be necessary.
The beat amplitude looks actually huge aIthough I don't know where you are monitoring.
Talk to Jamie to figure out how much the signal should be at the monitoring point.
If it is more than we are supposed to have, put an attenuator somewhere. |
8824
|
Thu Jul 11 00:30:27 2013 |
manasa | Update | Green Locking | X arm ALS post-beatbox makeover | I ran a series of diagnostics on the X arm ALS to look at how the beatbox behaves after the makeover.
Diagnostic tests run:
1. X arm ALS in-loop spectrum
2. X arm ALS out-of loop spectrum
3. X ALS scan of the X arm cavity
The noise suppression looks better after the makeover at the lower frequencies. To suppress the noise at high frequencies, we would have to add more whitening filters. |
Attachment 1: XALS_inloop.pdf
|
|
Attachment 2: XALS_scan.pdf
|
|
Attachment 3: ALS_outloop.pdf
|
|
14132
|
Fri Aug 3 19:02:11 2018 |
gautam | Update | ASS | X arm ASS recovery | [koji, gautam]
After I effected the series resistance change for ETMX, the X arm ASS didn't work (i.e. IR transmission would degrade if the servo was run). Today, we succeeded in recovering a functional ASS servo .
So both arms have working dither alignment servos now. But remember that the Y arm ASS gains have been set for locking the Y arm with MC2 as the actuator, not ETMY.
Details:
- Koji pointed out that the demodulated signals from the ETM dither are only used to center the spot on the ETM, and that we should first run the servo with existing settings with the ETM pitch and yaw spot centering loops disabled.
- This improved TRX level from ~0.8 to 1.1
- Next, we tried increasing the LO amplitudes by x5 to account for the reduced actuation of the dither on ETMX
- We then re-enabled the two loops that were earlier disabled.
- This resulted in TRX degrading very quickly.
- So we decided to try going back to the nominal LO gains, and reducing the gain of the two ETM spot centering loops.
- This did the trick, TRX went from 1.1 --> ~1.23, which is the nominal maximum pre-vent value.
- The snap file used to recover the correct settings to run the dither alignment servos have been updated, the old one has been backed up with today's datestamp.
We then tried to maximize GTRX using the PZT mirrors, but were only successful in reaching a maximum of 0.41. The value I remember from before the vent was 0.5, and indeed, with the IR alignment not quite optimized before we began this work, I saw GTRX of 0.48. But the IR dither servo signals indicate that the cavity axis may have shifted (spot position on the ITM, which is uncontrolled, seems to have drifred significantly, the Pitch signal doesn't stay on the StripTool scale anymore). So we may have to double check that the transmitted beam isn't falling off the GTRX DC PD. |
14161
|
Tue Aug 14 00:50:32 2018 |
gautam | Update | ASS | X arm ASS still not quite right? | While working on the single arm alignment, I noticed that today, i was able to get the X arm transmission back to ~1.22, and the GTRX to 0.52. These are closer to the values I remember from prior to the vent. Running the dither alignment promptly degrades both the green and IR transmissions. Since the pianosa SL7 upgrade, I can't use the sensoray to capture images, but to me, the spot looks a little off-center in Yaw on ETMX in this configuration, I've tried to show this in the phone grab (Atm #2). Maybe indicative of clipping somewhere upstream of ITMX?
Anyways, I'm pushing onwards for now, something to check out in the daytime.
Quote: |
[koji, gautam]
After I effected the series resistance change for ETMX, the X arm ASS didn't work (i.e. IR transmission would degrade if the servo was run). Today, we succeeded in recovering a functional ASS servo .
We then tried to maximize GTRX using the PZT mirrors, but were only successful in reaching a maximum of 0.41. The value I remember from before the vent was 0.5, and indeed, with the IR alignment not quite optimized before we began this work, I saw GTRX of 0.48. But the IR dither servo signals indicate that the cavity axis may have shifted (spot position on the ITM, which is uncontrolled, seems to have drifred significantly, the Pitch signal doesn't stay on the StripTool scale anymore). So we may have to double check that the transmitted beam isn't falling off the GTRX DC PD.
|
|
Attachment 1: POXPOY.png
|
|
Attachment 2: IMG_7108.JPG
|
|
10865
|
Wed Jan 7 11:20:22 2015 |
Steve | Update | LSC | X arm T-QPD gets SM1 thread adapter | C1:SUS-ETMX_QPD is removed and internal SM1 thread adapter epoxied into position as it is at the Y end
This adapter will take FL1064-10 line filter holder
Line filter is attached and qpd needs alignment. |
9594
|
Tue Feb 4 00:42:18 2014 |
Koji | Update | General | X arm aligned for IR/GR | The X arm was also aligned for the IR by hand and ASS. Also the X end green PZT was aligned to make the TEM00 mode reasonably locked.
What I did:
- Looked at the ITMXF camera. It seemed that the green beam was hitting the mirror.
- Went to the end. Looked at the X end green REFL. Tuned coarse alignment of the ETMX so that the beam was (retro-)reflected to the Faraday and the REFL PD.
- Looked at the ETMX face from the view port. Tried to locate the spot from the ITMX by shaking the ITMX alignment with 0.1 and then 0.01 increments.
- After some struggle with the ETMX and ITMX alignment, resonant fringes were found on the ETMY face while I still looked at the ETMX.
- Once the ITMX/ETMX were aligned, the BS needed to be aligned. But of course there was no IR fringe.
- Returned to the original alignment of the ITMX to find the ITMX spot on the AS camera.
Then gradually moved the ITMX to the aligned value for the green while tracking the michelson alignment with the BS.
This made the AS spots at the upper left edge of the AS video image.
- This was enough to find the IR spikes at TRX. Then the ETMX was touched to maximize the transmission.
- Lock the cavity. Use the ASS to optimize the alignement.
- Once the arm mirrors were aligned, the Xend PZT was also adjusted to have TEM00 for the green beam.
Now I leave the IFO with ITMX/Y, ETMX/Y and BS aligned. As I wrote above, the AS spot is very high at the AS camera.
We need to revisit the AS steering (SR TTs?) to ensure the AS beam unclipped. |
6849
|
Thu Jun 21 15:36:51 2012 |
yuta | Update | Locking | X arm alignment | I aligned X arm so that the beam spot comes roughly on the center.
1. Use ITMX and ETMX (mainly ITMX) to make beam spot come on center of the optic using eyeball.
2. Use ETMX and BS to maximize TRX power (reached ~ 0.85)
3. Aligned green optics on X end. Transmission of X green measured at PSL table is now 255 uW and TEM00 has the most power.
It was not easy to increase X green transmission more because beam spot on green transmission PD is wiggly when X end table is opened. When closed, wiggliness is about the same for Y green and X green.
Turning off HEPA on the X end didin't helped, but there must be something bad in the X end table. If we couldn't figure out why, let's wait for PZTs to come for end tables.
Considering the laser power is different(X end 1 W, Y end 700 mW), X green transmission should reach ~400 uW. But I think we should go on to X beat search.
I placed green shutter for X end back for convenience. I put some spacers to adjust its height and avoid beam clipping.
[Steve, Yuta]
What causing wiggly X green transmission was the air flow from the air conditioner. When we turned it off, beam spot motion became quiet. Air flow from HEPA was not effecting much. |
4278
|
Sun Feb 13 15:02:23 2011 |
kiwamu | Update | Green Locking | X arm beam offcentering has been measured | The amounts of the X arm's beam off-centering have been measured by the A2L technique.
So now we are able to start aligning the IR beam axis in a quantitative way.
(motivation)
Since we saw big residual motions at 1 Hz, 16 Hz on both the green beat note signal and the IR PDH signal (see #4268 and #4211),
we are suspecting that these noise come from an angle to length coupling.
In order to minimize the angle to length coupling, one thing we can do is to bring the beam spots to the center of ITMX and ETMX more precisely.
To do it, we have to quantitatively know how well the beam spots are on the center of the optics. Therefore I started measuring the amount of the beam off-centering.
(method)
The A2L technique was used to measure the off-centering with the real-time lockin system, which has been recently embedded in the real-time code by Joe (see #4265).
The idea is the same as Yuta did before (see #3863).
But this time the excitation signal from the real-time oscillator was injected directly to the coil matrix on either ITMX or ETMX, at 18.13 Hz with the amplitude of about 400 cnt.
When the IR laser stays locked to the X arm, the LSC feedback signal is demodulated with the oscillator signal.
This demodulated signal gives the amount of the off-centering.
For this purpose I modified Yuta's A2L script such that we can use it also for the X arm.
(results)
I obtained the following values:
- ETMX
PIT = -1.61 mm
YAW = -0.918 mm
- ITMX
PIT = -3.76 mm
YAW = -2.24 mm
I used the same calibration factor as that of Koji calculated (see #3020) for MC, in order to convert the results from the coil gain to the off-centering.
These values are consistent with the spots appearing on the CCD monitors.
 |
8820
|
Wed Jul 10 11:27:02 2013 |
manasa | Update | Green Locking | X arm beatnote found | I found the beat note for X arm. I did not change anything this morning (to the best of my knowledge). Hooking up the spectrum analyzer, I could find the beatnote signal at the PD RF output, after the amplifier and also at the MON port of the beatbox. I still don't know what changed from the last night set of trials  |
8296
|
Thu Mar 14 17:31:57 2013 |
Manasa | Update | Green Locking | X arm green locked in TEM00 |
Quote: |
Manasa has done some work to get the Xgreen aligned, so I'll switch to trying to find that beatnote for now.
|
[Jenne, Manasa]
Aligned X-arm green in TEM00.
It was difficult to get the X-arm to lock in TEM00 earlier. Even when it locked, it was just a TEM00 flash. The green was mainly bad in pitch.
I started aligning with the arms flashing in IR and it was still not possible to lock. The second trial was with the arms locked in IR; I lost the green lock when the arms were aligned for IR. I aligned by overlapping the ingoing green with the reflected green visible on the steering mirror. This got the green to lock in TEM00; but still it would only stay that way for 30 sec. Jenne pointed me to Yuta's suggestion of increasing the green refl PD gain. Once I increased the PD gain, the mode stabilized.
I went ahead and centered ALS TRX on the PD and GTRX camera (on the PSL table). ALS_TRX reads 2000+ counts.
EDIT by Jenne: This corresponds to a power of 550uW on the PSL table, measured before the first out-of-vac steering mirror. |
8298
|
Fri Mar 15 01:57:02 2013 |
Jenne | Update | Green Locking | X arm green locked in TEM00 | This work earlier today had required moving the harmonic separator back closer to its original position, so that the green could get through without clipping. I locked the Xarm (overriding the trigger) and realigned TRX to the PD and camera. |
9233
|
Fri Oct 11 00:37:23 2013 |
manasa | Update | Green Locking | X arm green locking modes | [Masayuki, Manasa]
We have stabilized the ALS for Y arm and concluded that although the PDH servo could be stabilized, it drifts and loses stability over a span of few hours. (See masayuki's elog today)
We wanted to follow the same systematic procedure like in the previous elog to look at the condition of the X arm as well.
In order to stabilize the green PDH servo, we held the arm using the IR PDH and aligned the end-green to the X arm.
We see 2 TEM00-like modes and one oblong TEM00+TEM01 mode that can lock to the cavity. It is not clear to me as yet as to how to differentiate between these 2 TEM-00 like modes and how we should decide between them.
One of the TEM00-like mode is strongly matched to the arm cavity. Normalized GTRX measures 0.6 counts. The other TEM00-like mode is weakly matched to the cavity. Normalized GTRX measures 0.12 counts. This might be the reason why Jenne and Masayuki were seeing a lower beat amplitude. Camera images are shown below.
 
On another note, we found that an oblong mode (looks like a TEM00+TEM10 mode) also locks to the cavity. The mode looks weird in that, only one half of the mode is seen moving due to seismic noise and the other part does not. I am not sure how I can describe this...so here is a 10 second video of how this mode looks like.
|
12478
|
Thu Sep 8 22:12:36 2016 |
gautam | Update | SUS | X arm in place, locked to green, IR flashes visible | Detailed elog to follow but summary of todays activities:
- ITMX and ETMX are back in their respective positions
- F.C was peeled, OSEMs were inserted after releasing EQ stops
- X arm was aligned to green
- IMC was locked, BS was used to adjust IR input pointing till beam was cleanly passing through irides (slid on to the tower)
- After best efforts for today - we see flashes as judged from TRX signal and also POX11_I. Unfortunately these are really weak and we can't lock, let alone see anything on the screens. Tomorrow we can try some more fine alignment
|
11111
|
Fri Mar 6 14:51:59 2015 |
ericq | Update | General | X arm linewidth, loss | The fit FWHM is 10.444kHz +-55Hz.
If we take the FSR from ELOG 9804, this implies an Xarm fineese of 380 +- 2.
Assuming an ITMX transmission of 1.4%, this means an Xarm loss of 240 +- 90ppm.
This is substatially lower than the ~500ppm I had measured via the unlocked/locked ASDC power method, but still pretty high.
Since we were able to get continuous frequency counter values into the digital system, I decided to give it a quick spin with a calibrated single arm ALS scan. This should be repeated when amplifiers are in place, because the Y IR beatnote is wandering around in a way I don't trust and I'm not sure if the frequency counters have good absolute calibration...
Neverthess, I did a 5 minute scan through the Xarm, and fit it nicely to a lorentzian peak.

|
4191
|
Mon Jan 24 02:58:46 2011 |
kiwamu | Update | Green Locking | X arm locked ! | I succeeded in green-locking the X arm by feeding back the beat signal to ETMX.
Here are some quick reports. Some more details will be posted tomorrow.
The below shows a time series data of the PLL feedback signal when the servo was acquiring the lock.

At t = -2 sec. I started feeding back the signal to ETMX with the gain 50 times smaller than its nominal.
Then at t = 0 sec.I switched on a low frequency boost (pole 0.1Hz and zero 1Hz) to make it more robust.
At t = 3 sec. I increased the gain to the nominal.
Finally the UGF became ~ 60 Hz according to my open loop measurement by diaggui.
However I couldn't make the UGF higher than 60Hz because the more gain caused a instability for some reasons.
Here is a diagram for the green locking.
I used the same VCO box as we setup on the last Friday (see #4189).
 |
4192
|
Mon Jan 24 09:33:08 2011 |
rana | Update | Green Locking | X arm locked ! | Very cool.
But the PLL seems very fishy to me. The ZP-3MH needs 13 dBm to operate correctly. You should change the MODLEVEL input of the VCO so as to make the LO input of the mixer go up to 13 dBm. Then the input from the PD should be ~0 dBm.
Also, the PLL diagram seems to show that you have a 1/f^2 loop: 1/f from the SR560 and 1/f from the Hz->rad conversion ?? |
4193
|
Mon Jan 24 10:19:21 2011 |
Koji | Update | Green Locking | X arm locked ! | Well... The ALS loop is engaged and the error was suppressed.
So, how is the IR error signal stabilized when the IR is brought in to the resonance?
I can see the linear trend of 0.1V/s from 5s to 10s. This corresponds to 100kHz/s and 13nm
for the residual beat drift and the arm length motion, respectively. That sounds huge. The DC gain must be increased. |
4153
|
Fri Jan 14 01:55:26 2011 |
kiwamu | Update | LSC | X arm locked with C1LSC digital control | [Koji, Kiwamu]
We succeeded in locking the X arm with the C1LSC digital control.
As we did on the day before yesterday, the feedback signal goes to MCL (#4141), but this time the signal is transfered from C1LSC through the RFM.
(key points)
- checking the state of the analog whitening filters at C1LSC rack.
We took the transfer function of them and found that they were always on regardless of the clicking any buttons on medm.
To cancel the filter shape of the whitening, we put an unWhitening filter so that these transfer functions becomes flat in total.
The whitening filter approximately has : pole:150Hz, pole:150Hz, zero:15Hz, zero:15Hz (although these numbers came from by our eye ball fitting)
- demodulation phase adjustment
We performed the same measurement as that of Suresh and Koji did yesterday (#4143) to adjust the phase of the PDH demodulation.
By changing the cable length we roughly adjusted the I-phase to eventually ~10 deg, which is close enough to 0 deg.
(probably some more efforts should be made as a part of daytime tasks)
Note that we are currently using the REFL33 demodulation board for this purpose (#4144). The LO power we put is about 16dBm.
The angle between I and Q at 11MHz is actually almost 90 deg.
This fact has been confirmed by putting a sinusoidal signal with a slightly different frequency (~100Hz) from that of the LO onto the RF input.
- attenuation of RF signal
Since the PDH signal taken by C1LSC's ADC had been saturated somewhat, we introduced a ND filter of 10 on the photo diode to attenuate the RF signal.
As a result the amplitude of the PDH signal on dataviewer became more reasonable. No more saturations.
(some notes)
unWhitening filter pole:15Hz. pole:15Hz, zero:150Hz, zero:150Hz
C1LSC_MC_FM1 pole:1kHz, zero:10Hz
Gain in digital control G ~ -1
measured UGF ~ 200-300 Hz
measured RFM delay ~ 125 usec |
4156
|
Fri Jan 14 12:34:08 2011 |
Koji | Update | LSC | X arm locked with C1LSC digital control | My feeling was that the saturation was caused by the LSC whitening filter which was always on.
Once the LSC whitening filter is controlled from C1LSC, we would be able to remove the attenuator.
Quote: |
- attenuation of RF signal
Since the PDH signal taken by C1LSC's ADC had been saturated somewhat, we introduced a ND filter of 10 on the photo diode to attenuate the RF signal.
As a result the amplitude of the PDH signal on dataviewer became more reasonable. No more saturations.
|
|
5360
|
Wed Sep 7 17:28:41 2011 |
kiwamu | Update | LSC | X arm loss measurement : not good | The measurement itself wasn't good.
I looked at the full 2 kHz data which was taken during the time when I was running the arm loss script on the X arm.
The plot below shows the raw data. The X arm was locked and unlocked sequentially several times.
The ASDC power didn't show a significant difference between the state where it is locked and unlocked.
I am not sure why, but It could be because of a misalginment or some kind of mode-mismatching, which can decrease the coupling efficiency of light going into the cavity.

(some notes)
The raw data were analyzed.
I split the ASDC data into two data, (1) low power state, when the cavity is locked (2) high power state, when the cavity is unlocked.
Then each state was averaged to estimate the averaged ASDC power in each case.
The number I obtained are :
ASDC when X arm was locked = 54.77755 cnts
ASDC when X arm was unlocked = 55.45830 cnts
Those numbers correspond to a round trip loss of 78.780778 ppm, which sounds too small for me.
Quote from #5359 |
To see what was going on I will look at the trend data.
|
|
14552
|
Thu Apr 18 23:10:12 2019 |
gautam | Update | Loss Measurement | X arm misaligned | Yehonathan wanted to take some measurements for loss determination. I misaligned the X arm completely and we installed a PD on the AS table so there is no light reaching the AS55 and AS110 PDs. Yehonathan will post the detailed elog. |
6859
|
Sat Jun 23 02:29:18 2012 |
yuta | Update | Green Locking | X arm mode scan results | X arm finesse is 416 +/- 6, mode-matching ratio is 91.2 +/- 0.3%
I did mode scan for X arm just like we did for Y arm (see elog #6832)
Servo design:
Servo filters are as same as Y arm.
UGF and phase margin of X arm ALS are 100 Hz and 14 deg.
For phase tracking loop, they are 1.5 kHz and 56 deg.
Raw data from the mode scan:

Fitted peaks and finesse:

By taking the average,
F = 416 +/- 6 (error in 1 sigma)
(For Y arm, it was 421 +/- 6. See elog #6832)
Mode matching ratio:
From X arm 8FSR measurement using phase tracker, peak heights are
TEM00 0.834, 0.851, 0.854, 0.852, 0.876, 0.850, 0.855, 0.878
TEM01 0.031, 0.031, 0.017, 0.017, 0.009, 0.014, 0.009, 0.011
TEM02 0.053, 0.052, 0.057, 0.058, 0.061, 0.060, 0.061, 0.059
TEM03 0.011, 0.010, 0.010, 0.007, 0.006, 0.005, 0.006, 0.005
So, the mode-matching ratio is
MMR = 89.7%, 90.1%, 91.0%, 91.2%, 92.0%, 91.4%, 91.8%, 92.1%
By taking the average,
MMR = 91.2 +/- 0.3 (error in 1 sigma)
(for Y arm, it was 86.7 +/- 0.3 %. See elog #6828)
Discussion:
- Mode matching ratio for both X and Y arm is ~90%, which is not so great, but OK. It seems like there's no huge clipping or mode-mismatch from MC to ITMs. I think we should go next for PRMI investigation.
- Measured finesse seems too low compared with the design value 450. If we believe power transmission of ITM and ETM are 0.0138 and 1.37e-5, the measured finesse tells you that there's ~0.1% loss(F = 2*pi/(T_{ITM}+T_{ETM}+T_{loss})). We need some evaluation for the linearity of the sweep, before concluding that there's 0.1% loss for each arm. Using FINE_I/Q signal for calibration, or installing frequency divider for monitoring actual beat frequency would help.
Things to do for the beat setup:
- Amplifiers after beat PDs shouldn't be on the PSL table. Move them near the beatbox.
- Install DC PD (and camera?) at un-used port of the beat BS for monitoring green transmission power.
- Make nice MEDM screens for our new phase tracking ALS.
- Make a script to sweep arm length with ALS and find IR resonance.
- Look into X end table. Beam spot of the X green transmission is wiggly when X end table is opened and there's air flow. |
10680
|
Thu Nov 6 12:53:09 2014 |
diego | Update | ASC | X arm restored | [Diego, Koji]
X arm has been restored, after modifying the two parameters mentioned in http://nodus.ligo.caltech.edu:8080/40m/10676 (C1SUS_ITMX: LSC/DAMP and LSC/BIAS); after that, a manual re-alignment of ETMX was necessary due to heavy PIT misalignment. I will check the ALS lock once work on the Y arm is done.
|
12474
|
Tue Sep 6 20:45:14 2016 |
gautam | Update | SUS | X arm test masses back in chamber | [Teng, Johannes, Lydia, gautam]
- The goal was to peel F.C. off both the X arm test masses and start work on aligning the arm
- However the F.C. peeling wasn't successful - Johannes spotted spme residual junk close to the center of the optic on ITMX and I saw a whole bunch of specks in and around the center of the ETM (see Attachment #1)
- Moreover, the PRM LR OSEM issue meant that we decided to re-paint the X arm optics and only take it off after debugging this OSEM PD issue
- Attachment #2 and #3 show the AR and HR face of the ITM respectively after F.C painting
- Attachment #4 shows the ETM HR face after HR painting
- Both towers have been moved, so any pre-emptive levelling has probably gone out the window, just something to be aware of when we put the towers back in place....
- There looks to be some filaments of F.C towards the edge of both the ITM and the ETM. These have been successfully removed with isopropanol + optical tissue, we should take care to do so before peeling the F.C....
|
Attachment 1: IMG_3137.JPG
|
|
Attachment 2: IMG_3143.JPG
|
|
Attachment 3: IMG_3142.JPG
|
|
Attachment 4: IMG_3148.JPG
|
|
10702
|
Wed Nov 12 15:14:09 2014 |
manasa | Update | General | X end AUX laser fiber setup | I looked at the endtable for possible space to setup optics in order to couple the X end laser into a PM fiber.
Attached is the layout of where the setup will go and what are the existing stuff that will be moved.

|
10831
|
Mon Dec 22 17:06:14 2014 |
manasa | Update | General | X end AUX laser fiber setup |
Quote: |
I looked at the endtable for possible space to setup optics in order to couple the X end laser into a PM fiber.
Attached is the layout of where the setup will go and what are the existing stuff that will be moved.

|
Since we will not be doing any major locking, I am taking this chance to move things on the X end table and install the fiber coupler.
The first steering mirror shown in the earlier elog will be a Y1 (HR mirror) and the second one will be a beam sampler (similar to the one installed at the Y endtable for the fiber setup).
Configuration:
Doubler --> Y1 ---> Lens (f=12.5cm) ---> Beam sampler --->Fiber coupler
The fiber coupler mount will be installed in the green region to the right of the TRX camera.
This work will involve moving around the TRX camera and the optic that brings the trans image on it.
Let me know if this work should not be done tomorrow morning for any reason. |
10834
|
Tue Dec 23 13:18:37 2014 |
manasa | Update | General | X end AUX laser fiber setup |
Quote: |
Since we will not be doing any major locking, I am taking this chance to move things on the X end table and install the fiber coupler.
The first steering mirror shown in the earlier elog will be a Y1 (HR mirror) and the second one will be a beam sampler (similar to the one installed at the Y endtable for the fiber setup).
Configuration:
Doubler --> Y1 ---> Lens (f=12.5cm) ---> Beam sampler --->Fiber coupler
The fiber coupler mount will be installed in the green region to the right of the TRX camera.
This work will involve moving around the TRX camera and the optic that brings the trans image on it.
Let me know if this work should not be done tomorrow morning for any reason.
|
I was working around the X endtable and PSL table today.
1. Y1 mirror, beam sampler and the fiber coupler have been installed.
2. Removed TRX camera temporarily. The camera will be put back on the table once we have the filter for 532nm that can go with it.
3. Removed an old fiber mount that was not being used from the table.
4. Lowered the current for X end NPRO while working and put it back up at 2A before closing.
5. The fibers running from the X end to the PSL table are connected at an FC/APC connector on the PSL table.
6. Found the HEPA left on high (probably from yesterday's work around the PSL table). I have brought it back down and left it that way.
I have not installed the coupling lens as yet owing to the space restrictions - not enough space for footprint of the lens. I have to revisit the telescope design again. |
12526
|
Fri Sep 30 19:53:07 2016 |
gautam | Update | endtable upgrade | X end IR pickoff fiber coupled | [johannes, gautam]
Today we re-installed the fiber coupler on the X-endtable to couple some of the PSL light into a fiber that runs to the PSL table, where it is combined with a similar PSL pickoff to make an IR beat between the EX AUX laser and the PSL. The main motivation behind this was to make the process of finding the green beatnote easier. We used JAMMT (just another mode matching tool) to calculate a two lens solution to couple the light into the collimator - we use a +200mm and -200mm lens, I will upload a more detailed mode matching calculation + plot + picture soon. We wanted to have a beam waist of 350um at the collimator, a number calculated using the following formula from the Thorlabs website:

where d is the diameter of the output beam from the collimator, f is the collimating lens focal length and MFD is 6.6um for the fiber we use.
There is ~26mW of IR light coming through the BS after the EX AUX - after playing around with the 6 axis stage that the coupler is mounted on, Johannes got the IR transmission to the PSL table up to ~11.7mW. The mode matching efficiency of 45% is certainly not stellar, but we were more curious to find a beat and possibly measure the X arm loss so we decided to accept this for now - we could probably improve this by moving the lenses around. We then attenuated the input beam to the fiber by means of an ND filter such that the light incident on the coupler is now ~1.3mW, and the light arriving at the PSL table from the EX laser is ~550uW. Along with the PSL light, after the various couplers, we have ~500uW of light going to the IR beat PD - well below its 2mW threshold.
The IR beat was easily found with the frequency counter setup. However, there was no evidence of a green beat. So we went to the PSL table and did the near-field-far-field alignment onto the beat PD. After doing this, we were able to see a beat - but the amplitude was puny (~-60dBm, we are more used to seeing ~-20dBm on the network analyzer in the control room). Perhaps this can be improved by tweaking the alignment onto the PD while monitoring the RF output with an oscilloscope.
Moreover, the green PDH problems with the X end persist - even though the arm readily locks to a TEM00 mode, it frequently spontaneously drops lock. I twiddled around with the gain on the uPDH box while looking at the error signal while locked on a oscilloscope, but was unable to mitigate the situation. Perhaps the loop shape needs to be measured and that should tell us if the gain is too low or high. But ALS is getting closer to the nominal state...
Johannes is running his loss measurement script on the X arm - but this should be done by ~10pm tonight.
|
9455
|
Thu Dec 12 00:21:04 2013 |
Koji | Update | Green Locking | X end PDH box oscillation issue solved (Re: screwed up the end PDH box) | What a such long pain we suffered.
After more than three years from Kiwamu's discovery, the PDH box 50kHz oscillation issue was finally solved.
This "weird peak at 50kHz" was caused by the oscillation of the voltage regulator (ON's MC7912).
As it imposed common noise almost everywhere, it screwed up transfer function measurements
like EricQ saw recently.
The negative voltage regulator (79XX) tends to get unstable than the positive counter parts (78XX).
The oscillation was removed by adding 22uF electrolytic capacitor between the output pin (pin3) and the ground pin (pin1) of MC7912.
This is indeed more than 20 times of the specification you can find in the data sheet. |
4297
|
Tue Feb 15 06:59:57 2011 |
Suresh, Jenne | Omnistructure | General | X end enclosure left open | [Jenne, Suresh]
Jenne found the X-end table enclosure had been left open. She replaced the lid on it.
|
10730
|
Fri Nov 21 11:41:24 2014 |
manasa | Update | General | X end fiber insulated and on cable tray |
Quote: |
Quote: |
[Steve, Diego, Manasa]
Since the beatnotes have disappeared, I am taking this as a chance to put the FOL setup together hoping it might help us find them.
Two 70m long fibers now run along the length of the Y arm and reach the PSL table.
The fibers are running through armaflex insulating tubes on the cable racks. The excess length ~6m sits in its spool on the top of the PSL table enclosure.
Both the fibers were tested OK using the fiber fault locator. We had to remove the coupled end of the fiber from the mount and put it back in the process. So there is only 8mW of end laser power at the PSL table after this activity as opposed to ~13mW. This will be recovered with some alignment tweaking.
After the activity I found that the ETMY wouldn't damp. I traced the problem to the ETMY SUS model not running in c1iscey. Restarting the models in c1iscey solved the problem.
|
AP Armaflex tube 7/8" ID X 1" wall insulation for the long fiber in wall mounted cable trays installed yesterday.
The 6 ft long sections are not glued. Cable tied into the tray pressed against one an other, so they are air tight. This will allow us adding more fibers later.
Atm2: Fiber PSL ends protection added on Friday.
|
[Steve, Manasa]
Two 70m long fibers are now running through armaflex insulating tubes along the X arm on the cable racks. The excess length of the fiber sits in its spool on top of the PSL enclosure.
Fibers were checked after this with the fiber fault locator (red laser) and found OK. |
10739
|
Mon Dec 1 15:41:38 2014 |
Steve | Update | General | X end fiber insulated and on cable tray |
Quote: |
Quote: |
Quote: |
[Steve, Diego, Manasa]
Since the beatnotes have disappeared, I am taking this as a chance to put the FOL setup together hoping it might help us find them.
Two 70m long fibers now run along the length of the Y arm and reach the PSL table.
The fibers are running through armaflex insulating tubes on the cable racks. The excess length ~6m sits in its spool on the top of the PSL table enclosure.
Both the fibers were tested OK using the fiber fault locator. We had to remove the coupled end of the fiber from the mount and put it back in the process. So there is only 8mW of end laser power at the PSL table after this activity as opposed to ~13mW. This will be recovered with some alignment tweaking.
After the activity I found that the ETMY wouldn't damp. I traced the problem to the ETMY SUS model not running in c1iscey. Restarting the models in c1iscey solved the problem.
|
AP Armaflex tube 7/8" ID X 1" wall insulation for the long fiber in wall mounted cable trays installed yesterday.
The 6 ft long sections are not glued. Cable tied into the tray pressed against one an other, so they are air tight. This will allow us adding more fibers later.
Atm2: Fiber PSL ends protection added on Friday.
|
[Steve, Manasa]
Two 70m long fibers are now running through armaflex insulating tubes along the X arm on the cable racks. The excess length of the fiber sits in its spool on top of the PSL enclosure.
Fibers were checked after this with the fiber fault locator (red laser) and found OK.
|
X-arm AP Armaflex tube insulation is cable tightened into cable tray. Only turning 6 ft sections are taped together.
Remaining things to do: install ends protection tubing |
7274
|
Fri Aug 24 21:00:40 2012 |
Koji | Update | LSC | X end green investigation | I checked and fixed the X end green situation. Now the X green beam is locked with TEM00.
There are various reasons it did not lock nicely.
- The IR beam axis was changed by Yoichi and Rana (ELOG #7169). So the green axis also had to be changed.
- The end green optics is really "BS". Anytime I see it, I feel disgusted. Because of 3D steering mirrors, cross couplings
between yaw and pitch are big. This makes the alignment hard.
- Even with acceptable alignment, the lock was only momentarily. I found the slow control was on. This pushed the frequency
too much and made the lock unstable.
- The slow control screen was broken as Jamie changed the model names but did not fix the slow screens.
- Jamie saids (ELOG #7011): Fix the c1sc{x,y}/master/C1SC{X,Y}_GC{X,Y}_SLOW.adl screens.
I need to figure out a more consistent place for those screens.
Now some action items are left:
- IR TRX is not aligned.
- X end green needs precise alignment.
- PSL GR TRX is not aligned.
These will be checked on Sunday.
- End green setup is horrible. => Manasa and I should work on this together. |
17530
|
Wed Mar 29 19:19:41 2023 |
Koji | Update | ALS | X end green now indefinitely locking | Stable lock of the X End green laser was recovered.
- The biggest issue was that the laser PZT input had been terminated with a 50ohm at the laser head. (See Attachment 1: The terminator has already been removed in the photo.) Since the PZT output of the servo box (output impedance 10Ohm) goes through 680Ohm at the summing node for the modulation, the PZT output was attenuated by a factor of 15. This made the required servo gain for locking more than the box could deliver. More importantly, the PZT range (in terms of the laser frequency) was also limited. Momentary locks were still possible with the reduced range and gain. However, the actuation signal hit the rail within a few seconds because of the pendulum motion.
Once the terminator was removed from the head, the Xarm was locked with the green laser like a charm.
- On the way to the resolution, I had to go through the full scrutinization of the loop components one by one. Here is the record of the findings:
- Inspected the green Refl PD (Thorlabs PFA36A). The gain setting of the PD was 40 dB, and the unlocked output voltage was 10.8 V. This is not only very close to saturation, but also the bandwidth drops below the modulation frequency (150 kHz according to Thorlabs' manual). The gain was changed to 20dB. This made the unlocked PD output to be 1.08V and the BW was expected to be 1MHz.
- Checked the LO setting. The box has a label saying "LO 7dBm". The function generator setting of "0.66 Vrms" resulted in 7.0dBm at the mixer LO input. So this number is used. Exactly the same amount goes to the PZT summing node.
- Checked the mod freq. The PDH error signal amplitude was maximized at 278.5kHz (mixer output observed with 50Ohm: 46.0mV), however, the signal looked distorted from the text-book shape of the PDF error. This means that the demod phase was not optimized.
The mod freq of 287.5kHz made the PDH error signal look better while the response was weaker (mixer out: 31.2mV). It turned out that the cavity locking didn't like these mod freq between 280kHz~290kHz. The momentary lock stretches showed a lot of quasi-sinusoidal fluctuation ~600Hz in the error and transmission signals. Instead, the modulation of 210.5kHz was used. This made the error signal during lock stretches clean and tight.
- Box inspection: Checked the signal ratio between the error in and the error mon. The monitor gain seemed x20~x21. The PZT output and the PZT mon had identical gains. The transfer function of the box was measured with the gain knob changed from 0.00 to 7.00 where the transfer function started to get distorted with the given input. The gain was increased by 5dB/turn (i.e., 1 turn increases the gain by 5dB). ? It does not match with the info on the schematic and the datasheet? Anyways, the gain knob is working fine.
- To resurrect the SLOW THERMAL servo, the monitor channels were connected to the DAQ interface. The existing slow channel servo/setting worked fine, wh
- Usual caution: a slight touch to the satellite amp caused the UR OSEM PD completely black out. It means that just your presence at the X end can make some changes to the suspension.
|
Attachment 1: PXL_20230330_021330919.jpg
|
|
Attachment 2: PXL_20230330_021305080.jpg
|
|
Attachment 3: PXL_20230330_021024076.jpg
|
|
Attachment 4: PXL_20230330_021029670.MP.jpg
|
|
Attachment 5: PXL_20230330_021048654.jpg
|
|
Attachment 6: Screen_Shot_2023-03-29_at_23.19.16.png
|
|
17540
|
Tue Apr 11 19:46:45 2023 |
Radhika | Update | ALS | X end green now indefinitely locking | I measured the OLTF of the XAUX-PDH loop [Attachment 1] now that the green laser is stably locking. I injected an excitation (100mVpp) at the error point of the loop using a Moku:Go. The excitation was summed with the PDH error signal (alpha) using an SR560, and the summed signal (beta) was sent to the PDH servo. (The Moku excitation was buffered with another SR560.) The transfer function beta/alpha was measured on the Moku.
The loop has a UGF of 26.3 kHz, and a phase margin of ~25º (using 1/1-OLG convention).
Next steps:
- Replace PDH servo demod + controller with Moku:Go lock-in amplifier (ensure loop shape is maintained)
- Deploy digital filters to further increase loop bandwidth/phase margin
|
Attachment 1: XEND_AUX_PDH_OLTF.pdf
|
|
Attachment 2: 100mV_MokuFrequencyResponseAnalyzerData_20230411_184527_Traces.csv
|
% Moku:Go Frequency Response Analyzer
% Channel 1, DC coupling, 10 Vpp range, amplitude 100 mVpp, offset 0.000 0 V, phase 0.000 deg
% Channel 2, AC coupling, 10 Vpp range, amplitude 2 mVpp, offset 0.000 0 V, phase 0.000 deg
% Logarithmic sweep from 1.000000 MHz to 99.99997 Hz with 1,024 pts, dynamic amplitude mode off, measuring fundamental, normalization off
% Averaging time 2.00 ms, 1 cycles; Settling time 100 us, 1 cycles
% Acquired 2023-04-11 T 18:45:27 -0700
% Frequency (Hz), Channel 1 Magnitude (dB), Channel 1 Phase (deg), Channel 2 Magnitude (dB), Channel 2 Phase (deg)
1.00000000e+06, 0.0000e+00, 0.0000e+00, -6.8386e+01, 5.7147e+01
9.91037142e+05, 0.0000e+00, 0.0000e+00, -6.9750e+01, 7.4405e+01
9.82154618e+05, 0.0000e+00, 0.0000e+00, -6.6775e+01, 7.2207e+01
... 1022 more lines ...
|
10873
|
Wed Jan 7 19:49:09 2015 |
manasa | Update | General | X end space issues | I have attached a photo of the ETMX table. The path of the 1064nm light rejected after the doubler and the green light are indicated for reference.
The fiber mount can only be mounted in the green space shown in the picture.
Calculating lens solutions for coupling the 1064nm light rejected after the doubler into the fiber, a lens of f=12.5cm should be placed at z=15.31cm (measured from the waist in the doubler crystal) gives the best ~80% coupling. This falls in the blue region where there is not enough space to mount a lens.
The region marked in orange has enough room for a lens; but the lens solutions give a coupling <10% which means there will be light scattering everywhere.
I am open to any suggestions on how to go about this.

|
10955
|
Thu Jan 29 10:14:21 2015 |
manasa | Update | General | X end space issues | It is certain that we have space issues at the X end that has been preventing us from sticking in a lens to couple light into the fiber.
The only way out is to install a platform on the table where we can mount the lens. I have attached the a photo of how things look like at the X end (attachment 1) and also a drawing of the platform that which can hold the lens (attachment 2). Additional support to the raised platform will be added depending on how much space we can clear up on the table by moving the clamping forks of the doubler.
Steve and I have been able to gather parts that can be put together into something similar to what is shown in the drawing.
Proposed modifications to the X end table:
1. The side panels of the table enclosure will come out while putting in the new platform.
2. The clamping forks for the doubling crystal will be moved.
Let me know of any concerns about the proposed solution. |
Attachment 1: photo.png
|
|
Attachment 2: drawing.png
|
|
12056
|
Wed Mar 30 17:38:52 2016 |
gautam | Update | endtable upgrade | X end table proposed layout | Attachment 1: This is a photo of the current X end table optical layout with the beampaths of the various sub-systems overlaid. For the labels, see Attachment #2.
Attachment 2: This is a summary of all the optical components that are currently being used. I've noted some things we may want to change when we effect the swap. The important ones are:
- Switch out all 1" and 2" optic mounts which are not of the Polaris type to the Polaris type. I have checked that we have sufficient numbers of these in hand.
- Adjust the collimating lens of the fiber collimating telescope to get a better mode
- Many of the labels are probably outdated, now would be a good time to update them
- For the mode-matching of the AUX IR into the doubling crystal, a la mode suggests a better (i.e. less sensitive to lens position) solution is effected with L2 as a 100mm fl lens rather than 88.3mm. I did not change this during the laser swap in order to minimize the number of components changed. Since we are doing a wholesale change now, it may not be a bad idea to swap this out as well. I have checked that we have a suitable AR1064 coated lens.
- Some optics probably need to be cleaned...
- PZT mirror 2 has a new mount ready that is the "correct" height so we don't have to keep using makeshift stacked posts.
- The plan as it stands is to use the green coloured mount for the IR faraday (IO-5-1064-HP).
Have I missed anything important?
Attachment #3: I've made a CAD drawing of the proposed new layout and have overlaid the beampath in an amateur way because I couldn't figure OptoCad out - I figure this will suffice for now. I have adopted elements from the current Y-end layout, but have used Anders' mode-matching solution (same lenses, same positions of optics) to make sure we have good Guoy phase separation between the two PZT steering mirrors. Some notes:
- I've tried to palce the optics for the AUX IR into the doubler and subsequent steering of green into the arm cavity as per the mode matching solution. These should be pretty accurate, and the layout suggests we have some room to maneuver
- The Green REFL beampath is exaggerated but I think we have enough room to place Y16 appropriately and steer the reflected beam into the PDA36A
- We need two more 1" 1064nm coated mirrors for the initial steering into the doubling oven, I have checked we have these in hand.
- The IR pickoff into the fiber coupler may change somewhat once we change the mode and redo the mode-matching calculations. But again, I think we have sufficient room to implement a workable solution.
- After accounting for the fact that the new endtable will be a little closer to the vacuum chamber, Y12 in the proposed layout will be ~10cm further away from ETMX than it is currently. But as discussed at the meeting today, the Rayleigh range of the green beam should be large enough here such that this shouldn't be a significant change.
Steve says the table is ready - so if we are happy with this layout, we can move forward... |
Attachment 1: ETMX_3x2.JPG
|
|
Attachment 2: layout_details_20160328.pdf
|
|
Attachment 3: ETMX_proposed_layout.pdf
|
|
10457
|
Fri Sep 5 04:07:44 2014 |
ericq | Update | Green Locking | X end uPDH Box Replaced | Just a quick note, plots and data will come tomorrow:
I grabbed an unused uPDH board from the ATF (thanks Zach!), and re-stuffed almost the entire thing to match Jenne's latest schematic for the y end box. I also threw some 22uF caps on the regulators, as Koji did with the previous box, to eliminate some oscillations up in the high 10s of kHz. I replaced the tragedy of a box that I created on Wednesday with this new box. The arm locks pretty stably with the boost on, 30 degrees of phase margin with 10kHz UGF, and locks pretty darn reliably.
Now we should now have two nicely boosted PDH loops. I'll do a noise/loop breakdown again in the upcoming days. |
12073
|
Wed Apr 13 00:56:07 2016 |
gautam | Update | endtable upgrade | X endtable repopulation | Over the last couple of days, I've been working on restoring the optical layout on the X-endtable. Some notes about the status as of today:
Lightwave NPRO output power
The output power from the lightwave NPRO is about 210mW (as measured with the calorimeter). This is significantly lower than the value of ~300mW reported in this elog. It may be that the laser crystal temperature has changed compared to that measurement, but the "ADJ" parameter is at 0, both today and in that measurement. The laser has also been on for more than a day now, that should be sufficient time for the crystal to equilibriate to its final operating state? Is such a large change in output power possible just because of a change in laser crystal temperature? Or did the laser really lose ~1/3rd of its output power over the last two months?
Alignment into IR Faraday, and changes to the planned layout
I've set up the layout until steering the beam through the IR faraday. The input power into the IR Faraday is ~210mW. The output power is ~186mW, after optimizing the angle of the HWP. These numbers seem consistent with what I had reported in this elog (although this was for the Innolight NPRO). The alignment looks reasonably good to the eye as well.
I've made one change to the planned layout (latest version here). Y1 is now a 2" 99% reflective for S polarization beam splitter, instead of a 1" HR mirror. I made this change because we want some light from the NPRO to be transmitted through this optic to couple into the fiber eventually, for the IR beat. I measured the transmitted power to be ~1.5mW, which is around what we were coupling into the fiber before, and should suffice now. The Lightwave NPRO datasheet (page 4) suggests that the polarization of the output of the laser is S, and the measured power before and after this optic suggests that it is working as advertised. This means that HWP 1 also has to be moved downstream (to rotate the polarization so as to maximize transmission through the IR faraday). Space constraints meant that I could not mount HWP 1 on the baseplate+3/4" OD post assembly which is what we want where possible on the new table, so for this optic, I used a 1" OD post and a fork. There may be a couple of other optics in the final layout where space constraints dictate we compromise in this way.
I've also installed beam dumps for the rejected light from the Faraday. For now, these are the old beam dumps. They looked reasonably intact. I believe we have a bunch of new beam dumps on hand as well, so these can be swapped out if deemed necessary.
Cleaning of optics
All the optics are being cleaned using first contact before being installed on the table.
As I found out the hard way, it is not a good idea to clean small optics like half-wave plates while in their mounts. The first contact tends to bond to the frame while drying, and doesn't come off cleanly. Koji helped me clean the offending pieces (he used tweezers to manually remove the residual first contact, and then some acetone to clean up any remaining residue). Subsequently, he re-cleaned these optics, again using first contact, but this time being careful not to extend all the way out to the edge of the optic. The idea is to cover as much area as possible with first contact, while staying clear of the edge. This approach worked reasonably well.
The next major step is to achieve optimal alignment into the doubler. I've placed the doubler on the table in it's approximate final position, I wanted to make sure the enclosure support wasn't in the way (it isn't). The cable from the oven won't run all the way to the Thorlabs temperature controller in it's usual place, we need to either extend the cable, or figure out a new place where we can keep the temperature controller. |
12074
|
Wed Apr 13 11:00:28 2016 |
Steve | Update | endtable upgrade | X endtable repopulation | ETMX optical table is grounded to ETMX chamber through 1 Mohms
The doubling oven temp controller is installed to reach its cable.
|
10981
|
Thu Feb 5 15:21:25 2015 |
manasa | Update | General | X endtable work | [EricG, Manasa]
We were at the X end today trying to couple AUX X light into the fiber.
The proposed plan still did not give a good beampath. The last steering mirror before the fiber coupler was sticking out of the table enclosure. I tried a few other options and the maximum coupling that I could get was ~10%
I am working on plan C now; which would be to use fixed mount mirrors and steer the beam to the space created by Koji near the IR trans path and use a set of lenses instead of a single lens. I will elog more details after some modematching calculations.
We moved one of the clamps for the doubling crystal to make space. Also, the NPRO current was reduced during this work.
I reset things to how they were before I touched the table. I ensured that the green power was still the same (~3mW) after the doubler and that it could lock to the arm in TEM00. |
10978
|
Wed Feb 4 21:09:43 2015 |
manasa | Update | General | X endtable work scheduled tomorrow | The X end fiber setup will be put together tomorrow morning. Let me know if there are any concerns.
Quote: |
It is certain that we have space issues at the X end that has been preventing us from sticking in a lens to couple light into the fiber.
The only way out is to install a platform on the table where we can mount the lens. I have attached the a photo of how things look like at the X end (attachment 1) and also a drawing of the platform that which can hold the lens (attachment 2). Additional support to the raised platform will be added depending on how much space we can clear up on the table by moving the clamping forks of the doubler.
Steve and I have been able to gather parts that can be put together into something similar to what is shown in the drawing.
Proposed modifications to the X end table:
1. The side panels of the table enclosure will come out while putting in the new platform.
2. The clamping forks for the doubling crystal will be moved.
Let me know of any concerns about the proposed solution.
|
|
5133
|
Sun Aug 7 14:01:58 2011 |
kiwamu | Update | Green Locking | X green beam re-aligned | [Jenne / Kiwamu]
The X green beam has been realigned to compensate the effect of the ETMX repositioning.
After the alignment we became able to lock the 00 mode with the X green beam.
For the alignment:
spot position on the ETMX mirror = within ~ 1 cm. This number is strictly constrained by a homemade aluminum iris that Jamie put last Friday.
spot position on the ITMX mirror = unknown, but looks pretty good on the CCD camera.
spot position on the PSL table = ~ 1 mm downward from what it used to be. The horizontal alignment is perfect.
Conclusions :
The X green beam again became a reference of the beam axis.
The ETMX suspension tower is in a good place.
Quote from #5127 |
Kiwamu will work on the green alignment over the weekend. Assuming everything works out, we'll try the same procedure on ETMY on Monday.
|
|
5135
|
Sun Aug 7 18:41:12 2011 |
Jenne | Update | Green Locking | X green beam re-aligned |
Quote: |
The ETMY suspension tower is in a good place.
Quote from #5127 |
Kiwamu will work on the green alignment over the weekend. Assuming everything works out, we'll try the same procedure on ETMY on Monday.
|
|
You meant ETMX, right? ETMY still hasn't been touched.
KI : sorry, I meant ETMX. I fixed the entry. |
6342
|
Wed Feb 29 20:27:00 2012 |
Jenne | Update | Green Locking | X green beat - found it! | Found it!
The actual temperature of the Xend laser is 0.02 C higher than anticipated based on the formula in elog 3759. Both the PSL and the Xend laser are at their nominal diode currents (2.100 A for the PSL, 2.003 A for Xend), so the curves should be used as they are. The PSL temp (when the slow servo offset is ~0) is 31.71 C. Using curve 2 from elog3759, the Xend laser should be 37.78, which I found was +10 counts on the Xgreen slow servo offset.
Right now the Xend laser is at 37.80 C, and the beat is around 30 MHz. This is +80 counts on the Xgreen slow servo. +60 counts gave me ~80 MHz. When (a few minutes ago) the MC unlocked and relocked, it came back to a slightly different place, so the temp of the Xend laser had to go up a few 10's of counts to get the same beat freq. Right now the PSL slow servo offset is 0.076 V.
The HP8591E is set with ResBW=100kHz, Ref Level= -39dBm (so I'm not attenuating my input signal!). The largest peak I see for the beatnote is -66dBm. The nose floor around the peak is -83dBm. Trace (trace button!) A is set to MaxHoldA, and Trace B is set to ClearWriteB, so B is giving me the actual current spectrum, while A is remembering the peak value measured, so it's easier to see if I went past the peak, and just didn't see it on the analyzer.
Also, I went back and realigned the beams earlier, to ensure that there was good overlap both near the BS which combines the PSLgreen and Xgreen beams, and at the PD. The overlap I had been looking at was okay, but not stellar. Now it's way better, which made the peak easier to see. Also, also, the waveplate after the doubling oven on the PSL table is still rotated so that I get max power on the Xgreen side of things, and not much at all on the Ygreen side. I'll need to rebalance the powers, probably after we make sure we are seeing the beatnote with the BeatBox.
Next Steps:
Lay a cable from the BBPD to the BeatBox in 1X2, make the BeatBox do its thing.
Use the dichroic locking to do a sweep of the Xarm. |
8666
|
Mon Jun 3 21:09:55 2013 |
Manasa | Update | Green Locking | X green beatnote through beatbox | I connected the X green beat PD output back to the beatbox, did the usual PSL alignment for green and searched for the beat note from the RFmon of the beatbox.
Yuta had used a power splitter which took Xbeat-RFmon and Ybeat-RFmon and used the SUM port to monitor the beat signals. I have removed this splitter and just used the X beat RFmon.
I found the beat note with:
Beat@58.7MHz : Amplitude -30dBm
C1:ALS-TRX_OUT16 = 3000 counts
C1:PSL-FSS_SLOWDC = 0.2250
PSL temperature = 31.52 degC
X- green temperature = 39.34 degC (OFFSET = 5140)
Next
Beatbox calibration

|
10000
|
Wed May 28 17:51:48 2014 |
manasa | Update | LSC | X green broadband PD NOT working |
Quote: |
Grr. I am very frustrated. After lunch I redid alignment for both X and Y green systems (Yarm both at the end and on the PSL table, Xarm just on the PSL table). After that realignment work, I cannot find a beatnote for the Xarm!!!
At this point, I still hadn't touched anything on the X path (except the PZT input steering mirrors, remotely from the control room). The beatnote was about the same size as it was on Friday, around -27dBm. I went onto the PSL table and did the same alignment procedure that I had just done for the Yarm: Remove the green trans PD and the accompanying lens so that I get far-field spots on the wall, and then steer the PSL green and the X green spots until they are nicely overlapped at both the camera (near-field) and on the wall. I looked at the DC output of the beat PD, and centered the beam on the diode. I put back the thorlabs DC transmission PD and the lens, and centered the beam on that. However, after this work, I cannot find a beatnote for the X arm! I still see the nice big Ygreen beatnote, and I have the PSL and Xend temperatures where they usually are ( abs(FSS Slow) < 0.1, and X end Slow around 10,090. ) I scanned -10,000 counts, and +5,000 counts from there, and still don't find a beatnote!
I went back inside, and I don't see an RF signal coming into the beatbox from the Xarm. It's not the cable's fault though, since I then hooked the RF output of the beat PD to a 'scope, and still didn't see any beatnote. The DC path of the PD is definitely seeing things, because when I switch the 'scope over to the DC output of the Xbeat PD, and I block/unblock the beam, I see the voltage step up and down as expected.
I have not pulled out the Xgreen broadband PD, but unless someone else has a good idea of what to check, that might be one of the next things to do. 
Ideas of things I could try:
* Put the X broadband PD on the Y beatnote path to see if I see the same Y beatnote (use the port where the Y green trans PD is, since it has the coaligned beams, and a lens).
* Open the PD and see if anything on the RF path is fried.
* Move the Y PD over to the X path, to see if it sees the beatnote.
* ????
|
I made my attempts trying to figure out what was wrong.
Checking if we are at the right X end laser temperature:
I aligned the arms and found the Y beatnote.I blocked the light falling on the X beat PD so that the RF analyser was only looking at the output from the Y beat PD. AT the RF analyser, I found the strong Y-PSL beatnote, the X-Y beat note and a weak X-PSL beatnote. This confirmed that we have the X end laser at the right temperature to be able to detect the beatnote. Unblocking the light on the X beat PD did not bring in any additional peaks.
Checking the RF cabling from the X beat PD to the beat box:
I swapped the RF cables such that the signal from the Y beat PD output was going to the X input on the beatbox. I could still see the beatnote on the RF analyser. This confirmed that there aren't any broken RF cables along the X path.
Checking X green PSL alignment:
I replaced the X beat PD with the Y beat PD to check if the alignment of X&PSL green are alright. I could find the X beat note this way without any alignment tweaking.
I suspect we probably have some RF component burnt in the X beat PD. Do we have any spares lying around? There is a Koji's box with a PD having the same serial number.
IFO status report for anyone who is looking to do some locking tonight :
The Y beat PD is back along the Y path and I have confirmed the presence of Y-PSL beat note after replacing the PD.
The X beat PD has been removed and now rests on the electronics bench for checking.
While aligning the arms today, I noticed that enabling LSC would cause misalignment of the ETMY suspension. I haven't tried to find out what has been causing this. Could be something similar to what was noticed with the ETMX suspension a couple of weeks ago elog9969. |
|