40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 66 of 355  Not logged in ELOG logo
ID Date Authorup Type Category Subject
  7551   Mon Oct 15 22:16:09 2012 JenneUpdateSUSAll oplev gains turned to 0

Steve has promised to fix up all of the oplevs, but it hasn't happened yet, so I've turned all of the oplev gains to zero, so that when the optics are restored we don't have to quickly click them off.

Oplev values that were changed to zero:

PRM P=0.15, Y=-0.3

SRM P=-2.0, Y=2.0

BS P=0.2, Y=-0.2

ITMY P=2.1, Y=-2.0

ITMX P=1.0, Y=-0.5

ETMX P=-0.2, Y=-0.2

ETMY P=0.5, Y=0.6

Also, PRCL was changed in the LSC input matrix from REFL33I to AS55I, since there is no REFL beam out of the IFO :(

  7552   Mon Oct 15 22:24:45 2012 JenneUpdateComputersLots of new White :(

Evan and I are starting to lock, and there is lots of new, unfortunate white stuff on several different screens.

C1:TIM-PACIFIC_STRING is gone, C1:IFO-STATE (MC state) is gone, C1:LSC-PZT..._requests are gone (all 4 of them), C1:PSL-FSS_FASTSWEEPTEST from the FSS screen is gone (although I'm not sure that that one is newly gone), lots of the WF AA lights on the LSC screen are gone.

Those are the things I find in a few minutes of not really looking around.

EDIT:  IPPOS is also gone, so I can't see how my current alignment relates to old alignments.

  7554   Tue Oct 16 00:33:29 2012 JenneUpdateLSCPOP lens placed on POX table, 2 PRMI movies

[Evan, Jenne]

We aligned the PRMI.  We definitely can lock MICH, but we're not really sure if PRCL is really being locked or not.  I don't think it is.

Anyhow, we found 2 different places on the AS camera that we can align the PRMI.  One (middle, right hand side of the camera), we see the same weird fringing that we've been seeing for a week or two.  The other (lower left side of the camera), we see different fringing, almost reminds me more of back in the day a few months ago when the beam looked like it was expanding on each pass.  As I type, Evan is uploading the movies to youtube.  I *still* don't know how to embed youtube videos on the elog!

Also, we found both forward-going and backward-going POP beams coming out onto the POX table.  We placed the 2" lens in the path of the backwards beam, so that we can find it again.  We can't see it on an IR card, but if we put some foil where we think the beam should be, we can use a viewer to see the spot on the foil.  Poking a hole in the foil made an impromptu iris.

Youtube videos:

Lower left on camera

Middle right on camera

  7556   Tue Oct 16 11:38:17 2012 JenneUpdateLSCMore PRMI notes from last night

Quote:

How can you lock the PRMI without the REFL beams? c.f. this entry by Kiwamu
Which signals are you using for the locking?

I think the first priority is to find the fringes of the arms and lock them with POX/POY.

As for the POP, make sure the beam is not clipped because the in-vac steering mirrors
have been supposed to be too narrow to accommodate these two beams.

I was using AS55I for PRCL, and AS55Q for MICH.  I snuck that into the last line of an unrelated elog, since I did both things at the same time: see elog 7551.  Kiwamu's measurements (elog 6283) of the PRMI sensing matrix show that the PRCL and MICH signals are almost orthogonal in AS55 (although the optickle simulation doesn't agree with that...)  He was able to lock PRMI with AS55 I&Q (elog 6293), so I thought we should be able to as well.  Locking the PRMI was supposed to help tune the alignment of the PRM, not be the end goal of the night.  Also, we only tried locking PRCL in the "middle right" configuration, not the "lower left" configuration, but we were seeing what looked like recycling flashes only in the "lower left" configuration.

I agree in principle that we should be working on the arms. However, since we can't use the old steer-the-beam-onto-the-cage trick to find the beam, I was hoping that we could steer the beam around and see some light leaking out of the ETM, onto the end table.  However, with the 1% transmission of the ITMs and ~10ppm transmission of the ETMs, there's not a lot of light back there.  I was hoping to align the PRMI so that I get flashes with a gain of 10 if I'm lucky, rather than just the 5% transmission of the PRM.  With the PRMI aligned, I was expecting:

(1W  through Faraday) * (10 PR gain) * (0.5 BS transmission) * (0.01 ITM transmission) * (10ppm ETM transmission) = 0.5uW on the ETM tables during PRCL flashes

I was hoping that things would be well enough aligned that I could just go to the end table, and see the light with a viewer, although as I type this, I realize that if the beam was not on the end table (or even if it was...) any time I move the PZTs, I'd have to completely realign the PRMI in order to see the flashes.  This seems untenable, unless there are no other options.

We then got sidetracked by trying to see the POP beam, and once we saw the POP beam we wanted to put something down so we could find it again.  POP is also small, but not as small as expected at the end:

(1W  through Faraday) * (10 PR gain) * (20ppm PR2 transmission) = 0.2mW on POP during PRCL flashes.

POP was very difficult to see, and we were only able to see it by putting the foil in the beam path, and using a viewer.  I think that we once were able to see it by looking at a card with the viewer, but it's much easier with the foil.  I'd like to find an iris that is shiny (the regular black iris wasn't helpful), to facilitate this alignment.  Since we were just looking at the reflection off of the foil, I have no comment yet about clipping vs. not clipping.  I do think however that the forward-going beam may have been easier to find....when the PRMI alignment drifted, we lost the beam, but I could still see the forward-going beam.  Probably I should switch to that one, since that's the one that was lined up with the in-vac optics. 

Summary:

Ideas are welcome, for how to align the beam to the Yarm (and later to the Xarm), since our old techniques won't work.  Aligning the PRMI was a distraction, although in hopes of getting flashes so we could see some light at the end tables.  I'm going to go see if I can look through a viewport and see the edges of the black glass aperture, which will potentially be a replacement for the steering-on-the-cage technique, but if that doesn't work, I'm running out of ideas.

  7557   Tue Oct 16 11:54:05 2012 JenneUpdateIOOMC alignment??

The MC won't survive the boosts right now.  Pizza meeting is in a minute, and I won't be back to the lab before ~3:30 because of the seminar / a meeting, so someone else is welcome to try to fix it. Otherwise I'll have a look later on.

I'm leaving the autolocker disabled, so that it won't try any funny business.  WFS are off, so that they don't need to be turned off by the down script.

  7569   Wed Oct 17 18:41:27 2012 JenneUpdateCamerasCamera looking at ETMY baffle

The camera titled "watec_mobile" is looking at the front of the black glass baffle (i.e. the side facing the ITM) on the ETMY table.  This required (for my quick hacky solution) removing the regular ETMYF camera.  Steve has a genius plan (I think) so that we can have both at the same time.  Anyhow, eventually we'll move the black glass back, so we'll be back to needing just one camera.

After dinner, I'll try aligning the Yarm.

  7572   Thu Oct 18 03:45:56 2012 JenneUpdateIOOMCL, WFS triggers

Quote:

I've added MCL and WFS stop triggers into C1MCS/SUS model. Threshold value of MC_TRANS can be changed in the text entry located in MC2_POSITION medm screen. I tried 2 cases: trigger either blocks signal before MCL filter bank input or after output. Due to filter history in the 1 case MC2 was still slightly disturbed (C1:SUS-MC2_ULPD_VAR ~= 15) right after unlock. In the second case there was no disturbance as we zero output signal, but then I had to add "clear history" command to the mcup script.

WFS triggers block the signal before ASCPIT/YAW filter bank.

 I've redone the WFS triggers.  I left the MCL trigger alone (for now....I'll come back to it). 

The trigger was setup such that (a) it was totally unclear what was going on, by looking at the WFS screen.  Koji and I spent some time confused before I remembered that Den did this work recently.  Also, for some reason, the triggers were just plain thresholding, not a schmidt trigger, so any time the cavity flashed, the WFS came on.  Since the cavity can flash before the mcdown script has a chance to turn off the WFS servos, the outputs of the WFS filters are trying to output thousands of counts, and the signal goes through any time the cavity flashes.  Not so good.

I have removed the triggering for the angular DoFs from the mcs model (leaving the MCL triggering for now).  I have put new triggering into the ioo model, at the error point of the WFS loops.  The idea is that if the cavity unlocks, we don't want to lose the current pointing of the mirrors.  If the WFS servos were doing a lot of DC work, the bias sliders won't have the full information about where we want the mirrors to point.  Since we have the integrators in FM1, removing the input signal should freeze the output signal.  I need to modify the WFS on / off script so that this doesn't get turned off every lockloss.

Also, I have implemented (for the first time in a useful model, although I've done some testing in the tst model) the "wait" delay between a cavity locking and the trigger going through.  The idea is that we don't necessarily want the WFS to come on simultaneously with the cavity lock.  Since the wait delay resets any time it is un-triggered, this also prevents any signals from going through during cavity flashes.  The wait block has 3 inputs:  (1) a trigger, the output of some kind of trigger block, (2) a number of seconds to wait and (3) the model rate in Hz.  The model rate should be set with a constant in the model, the trigger passed from the trigger block, and the wait time in seconds should be available as an epics input. 

So far it looks like it's working as I expect, although I'm honestly too tired to do enough testing that I'm satisfied with, so I'm leaving the WFS off for the night.

  7573   Thu Oct 18 03:57:20 2012 JenneUpdateLockingAlignment is really bad??

The goal of the night was to lock the Y arm.  (Since that didn't happen, I moved on to fixing the WFS since they were hurting the MC)

I used the power supplies at 1Y4 to steer PZT2, and watched the face of the black glass baffle at ETMY.  (elog 7569 has notes re: camera work earlier)  When I am nearly at the end of the PZT range (+140V on the analog power supply, which I think is yaw), I can see the beam spot near the edge of the baffle's aperture.  Unfortunately, lower voltages move the spot away from the aperture, so I can't find the spot on the other side of the aperture and center it.  Since the max voltage for the PZTs is +150, I don't want to go too much farther.  I can't take a capture since the only working CCD I found is the one which won't talk to the Sensoray.  We need some more cameras....they're already on Steve's list.

When the spot is a little closer to the center of the aperture than the edge of the aperture (so the full +150V!!), I don't see any beam coming out of AS....no beam out of the chamber at all, not just no beam on the camera.  Crapstick.  This is not good.  I'm not really sure how we (I?) screwed up this thoroughly.  Sigh.  Whatever ghost REFL beam that Kiwamu and Koji found last week is still coming out of REFL.

Previous PZT voltages, before tonight's steering:  +32V on analog power supply, +14.7 on digital.  This is the place that the PRMI has been aligned to the past week or so.

Next, just to see what happens, I think I might install a camera looking at the back (output) side of the Faraday so that I can steer PRM until the reflected beam is going back through the Faraday.  Team K&K did this with viewers and mirrors, so it'll be more convenient to just have a camera.

Advice welcome.

  7577   Fri Oct 19 00:55:35 2012 JenneUpdateComputersc1lsc is down (at least all of the models)

When Evan and I were dithering the BS and ITMY (see his elog), I noticed that c1lsc was acting weird.  the IOP was the only one with the blinky heartbeat.  The IOP was all green lights, but all the other models had red for the fb connection, as well as the rightmost indicator (I don't know what that one is for).  I logged on to c1lsc and ran 'rtcds restart all'.  The script didn't get anywhere beyond saying it was beginning to stop the 1st model (sup, the bottom one on the lsc list).  Then all of the cpus went white.  I can still ping c1lsc, but I can't ssh to it.

I'm not sure what to do here Jamie.  Heelp. 

  7578   Fri Oct 19 01:11:18 2012 JenneUpdateIOOMCL, WFS triggers

Quote:

Quote:

 

 I've redone the WFS triggers.  I left the MCL trigger alone (for now....I'll come back to it). 

The trigger was setup such that (a) it was totally unclear what was going on, by looking at the WFS screen.  Koji and I spent some time confused before I remembered that Den did this work recently.  Also, for some reason, the triggers were just plain thresholding, not a schmidt trigger, so any time the cavity flashed, the WFS came on.  Since the cavity can flash before the mcdown script has a chance to turn off the WFS servos, the outputs of the WFS filters are trying to output thousands of counts, and the signal goes through any time the cavity flashes.  Not so good.

 Your schmitt trigger has 2 threshold values - min and max. Set thresholding value in my trigger to the max of your schmitt trigger and you get the same behavior for MC,  triggers are not supposed to turn anything on in this realization as they do for locking with flashing.

 The problem is that the WFS were being engaged with your triggers every time the MC flashed.  That wasn't a schmidt trigger thing, but I like the schmidt trigger better anyway.

 

Anyhow, it's turned on, and it works really well.  It's kind of awesome.  I'm really excited to start using the wait block to start pushing even more of the locking out of scripts and into the real time system.

  7584   Fri Oct 19 19:26:42 2012 JenneUpdateVACvent has started

Quote:

Quote:

 

 PSL shutter closed, manual block in place, HV turned off. P1 is at 200 Torr now.  Jenne is taking over here.

 Valves closed, 500 torr.  Steve will finish off Monday morning, then we'll take off doors and get to work.

  7590   Mon Oct 22 21:20:36 2012 JenneUpdatePSLBeam attenuation optics in place

[Jenne, Raji (before dinner)]

We put the beam attenuation optics in place.  Before putting any optics down, I centered the IOO QPDs, then adjusted the HWPs and PBS such that we remained centered on those QPDs.

Now, I'm about to unblock the beam and let ~100mW into the vacuum so I can lock the MC.  Steve and Manasa were putting on the light access connector when I left earlier, so I'm excited to use it!

  7591   Tue Oct 23 00:41:55 2012 JenneUpdateAlignmentMC locked, spots centered

[Jenne, Raji]

We replaced the MC Refl path BS with the Y1, as usual, so that the full ~100mW goes to the REFL PD, so we don't have WFS or MC refl camera. 

The MC spots were all outside of 1mm, and some were beyond 2mm (for MC1,3, P,Y....MC2 is of course free since we have more DoFs than we need), so we touched (very, very slightly) the zigzag mirrors on the PSL table.  We realigned the MC, and now the spots are centered to my satisfaction.

MC1,2,3 Pit,    MC1,2,3 Yaw (in mm):

[0.46444020918749457, 8.2634316545130009, -0.41417975237831089, -0.89401481457980592, -0.9323196976382162, -1.543145765853893]

MC2 is way off in pitch, according to this measurement, and it's been consistently going down as we move the MC2 spot in the same direction (up on the monitor), but since we started at +15mm and are now at +8, and we've gone quite a ways, I'm not sure that we really want to go all the way to 0.  Anyhow, MC1 and MC3 are the ones which define our input pointing, so we're quitting for tonight.

We will turn on the PZTs and begin with the official vent list for dummies tomorrow.

Attachment 1: Screenshot-Figure_1.png
Screenshot-Figure_1.png
  7601   Tue Oct 23 18:12:18 2012 JenneUpdateAlignmentTip tilt wires - the truth

Quote:

At this point I cried foul.  This is not an acceptable situation.  Very little stimulation to the tip-tilts can repoint the beam inside the PR cavity.

Steve says that the TT weights, which will attach to the base of the TT mirror mounts and should help keep the mirrors vertical and not hysteretic, are being baked now and should be available tomorrow. 

 We also wrote down the serial numbers (top center of each TT, inscribed by hand) for what tip tilt is installed where.  I then went through the elog to determine which TT was suspended with what kind of wire (thick or thin).  Summary: all installed tip tilts have thick wire, 0.0036" diameter.

As noted in elog 3295, we had found that there was similar hysteresis whether we used the thick or the thin wire, so we had decided not to go back and re-suspend every optic.

Also, since we will redo the pitch balance tomorrow with the new hardware tomorrow, I think we should put in the new LaserOptik mirrors at the same time.  We have not yet gotten phase maps of them, but we might as well do this rebalancing once, rather than twice.

 

As-installed tip tilt list
Serial number Installed as Wire thickness Notes, elog reference
001 SR 3 0.0036" See elog 3437
002 SR 2 0.0036" See elog 3295
003 PR 2 0.0036" No elog, but inferred since there were 4 with thick wire, and #004 is the thin wire one.  Elog 3437 has notes on the 4 thick, 1 thin situation.
004 spare, dirty originally 0.0017", but looks redone with thicker wire See elog 3295
005 PR 3 0.0036" Was supposed to be spare according to elog 3437, but was installed.  See elog 3437

 

  7602   Tue Oct 23 18:18:29 2012 JenneUpdateSUS PITCH damping needed

Quote:

Quote:

 

Too bad - I thought it would at least give a little damping. Since we want the viscous-like energy loss to be ~49x larger, we need to have the field modulation in the damper (not dumper) increase by ~7.

 I've made SolidWorks models of damping bracket and eddy current disk. They will me manufactured and used instead of old ones. New bracket will be mounted in exactly the same place where the old one was. Drawings might not be complete but all dimensions are in the models so we can fix drawing tomorrow before going to machine shop.

I think we can use ring magnets for passive damping. Then we won't have the vent problem. I've found some at K&J Magnetics, we can get them any time. Magnets are Ni-Cu-Ni (fine for vacuum?) Diameter is 3/8'' with advertised tolerence 0.004'', so they should fit the holes.

 Den mentioned that the disks will have threaded holes, and that he has made a note to that effect on the paper copy of the drawing that he will bring to Mike at the shop.  Also, all threaded holes in the new plate are marked on the paper copy.

  7603   Tue Oct 23 18:21:21 2012 JenneUpdateAlignmentPower supply at OMC removed

Quote:

Quote:

Manasa and Raji hooked up HV power supplies to the PZTs and set them to the middle of their ranges (75 V).

 [Raji, Manasa]

The high-voltage power supply from the OMC was removed to replace one of the PZT power supplies. The power supply terminals were connected to the rear connection ports as per instructions from the manual (TB1 panel: port 3 - (-)OUT and port7 - (+)OUT). They were both switched  on and set to deliver (75V) to the PZTs.

 

 This means that the low voltage dual supply which was wired in series (so could supply a max of 63V = 2*31.5V) has been replaced with the OMC power supply.  This is okay since we haven't turned on the OMC PZTs in a long, long time.  This is *not* the power supply for the output pointing PZTs.  When she says "both", she means the new HV supply, as well as the HV supply that was already there, so both pitch and yaw for PZT2 are being supplied with 75V now.

  7606   Wed Oct 24 11:49:07 2012 JenneUpdateAlignmentGame plan for the day

Jamie has arranged for phase map measurements this afternoon, so I will take the 6 dichroic LaserOptik optics over to Downs at 1:15 this afternoon.

Team Jamie+Nic will lead the effort to clamp down dog clamps as placement markers for all 4 in-vac passive TTs, and then pull all 4 TTs out of the chambers.  They plus Den will move the TTs to the Cleanroom, and will start to install the new pitch alignment hardware. 

When I return with the optics, we will install them in the TTs and re-balance them.  Then we can put them back in the chambers and get back to work on alignment.  

After we re-install the TTs, we will need to check the leveling of all 3 corner tables, just to be sure.

  7608   Wed Oct 24 14:19:01 2012 JenneUpdateAlignmentPhase map summary of LaserOptik mirrors

Quote:

Jamie has arranged for phase map measurements this afternoon, so I will take the 6 dichroic LaserOptik optics over to Downs at 1:15 this afternoon.

Team Jamie+Nic will lead the effort to clamp down dog clamps as placement markers for all 4 in-vac passive TTs, and then pull all 4 TTs out of the chambers.  They plus Den will move the TTs to the Cleanroom, and will start to install the new pitch alignment hardware. 

When I return with the optics, we will install them in the TTs and re-balance them.  Then we can put them back in the chambers and get back to work on alignment.  

After we re-install the TTs, we will need to check the leveling of all 3 corner tables, just to be sure.

 Raji took the optics over. They were all measured at 0 deg incidence angle, although we will use them at the angles required for the recycling folding mirrors.  Here's the summary from GariLynn:

In general all six pieces have a radius of curvature of around -700 meters.

They all fall off rapidly past 40 mm diameter.  Within the 40 mm diameter the rms is ~10 nm for most.  I can get finer analysis if you have something specific that you want to know. 
 
All data are saved in Wyko format at the following location:
Gari
  7610   Wed Oct 24 17:02:01 2012 JenneUpdateAdaptive Filteringmicrophone noise

Quote:

  We have to change the sample rate and AA filter for the mic channels before going too far with the circuit design.

 To save the mic channels at higher than 2k (which we should do), we either have to move them to a different model, change the rate of the PEM model, or see if you can save data faster than the model runs (which I can't imagine is possible).

  7625   Thu Oct 25 20:44:11 2012 JenneUpdateSUSTip tilts in progress

Jamie and I spent some time with tip tilt SN001 this afternoon.  This was installed as SR3, so I was going to put a new LaserOptik mirror in there.  I accidentally snapped one of the wires (I forgot how strong the magnets are - one zipped from the mirror holder and captured the wire).  Jamie and I put the new LaserOptik mirror in, with the wedge correct, but we need to re-resuspend it with the 0.0036" wire tomorrow.  We'll also keep working on re-pitch aligning the other optics.

PR2 needs to be put back as a G&H, and we need to put a LaserOptik mirror into PR3.

  7630   Fri Oct 26 10:44:25 2012 JenneUpdateSUSTip tilts in progress

Quote:

Jamie and I spent some time with tip tilt SN001 this afternoon.  This was installed as SR3, so I was going to put a new LaserOptik mirror in there.  I accidentally snapped one of the wires (I forgot how strong the magnets are - one zipped from the mirror holder and captured the wire).  Jamie and I put the new LaserOptik mirror in, with the wedge correct, but we need to re-resuspend it with the 0.0036" wire tomorrow.  We'll also keep working on re-pitch aligning the other optics.

PR2 needs to be put back as a G&H, and we need to put a LaserOptik mirror into PR3.

 We resuspended SN001 this morning with 0.0036" wire.  We did as Koji suggested, and flipped the wire clamp so the suspension point is a little higher, so we'll see if that helps.  We put LaserOptik mirror SN1 into this TT001.

We put the G&H mirror back into TT004, which is PR2.  We also put a LaserOptik mirror (SN5) into TT005, which is SR3.

Jamie is working on re-pitch aligning TT004 and TT005 (we already did 001), then we can re-install them in the vacuum system later this afternoon.

  7631   Fri Oct 26 13:08:14 2012 JenneUpdateSUSTip tilts in progress

Quote:

Quote:
 

Jamie is working on re-pitch aligning TT004 and TT005 (we already did 001), then we can re-install them in the vacuum system later this afternoon.

 The tip tilts have all been pitch-adjusted now, and they have all been put back onto the tables, with the same serial numbers in the same places as we took them out.  Jamie also re-leveled the BS table.

Raji and I will align things after I finish measuring the MC spot positions.

  7632   Fri Oct 26 16:57:30 2012 JenneUpdateAlignmentPR2 aligned, PR3 mostly aligned

[Raji, Jenne]

After lunch we began where Raji and Jamie had left things.  PR2 was unfortunately pitched down so far that it was almost hitting the table just in front of PR3.  I loosened the 4 clamp screws that hold the wire clamp assembly to the mirror holder, and tapped it back and forth until I was within hysteresis range, re-tightened, then tapped the top and bottom until we were at the correct beam height just in front of PR3.  I also had to unclamp it from the table and twist the base a tiny bit, since the beam was closer to hitting the beam tube than the optic.  Finally, however, PR2 is adjusted such that the beam hits the center of PR3.

Moving on to PR3, the pitch looked good while we were looking at the aperture placed near the face of ITMY, so we left that alone.  The beam is off in yaw though.  Several times I unclamped the tip tilt from the table, and twisted it one way or another, but every time when I tighten the dog clamps, I'm too far off in yaw.  The beam points a little too far south of the center of ITMY, so we were putting the beam a little north of the center before I clamped it, but even tightening the screws in the same order, by the same amount each time, causes a different amount of slipping/twisting/something of the TT mount, so we never end up directly in the center of the ITM.  It seems a little like a stochastic process, and we just need to do it a few more times until we get it right. 

We left it clamped to the table, but not in it's final place, and left for JClub.  On Monday morning we need to go back to it.  As long as we're pretty close to centered, we should probably also have someone at ETMY checking the centering, because we need to be centered in both ITMY and ETMY.

We have not touched the SR tip tilts, so those will obviously need some attention when we get to that point.

  7641   Mon Oct 29 18:50:02 2012 JenneUpdateAlignmentPRC aligned, Yarm almost aligned

[Jamie, Jenne, Raji, with consultation from Nic, Ayaka and Manasa]

We went back and re-looked at the input alignment, and now we're "satisfied for the moment" (quote from Jamie) with the PRC alignment.  Also, by adjusting the PR folding mirrors, we are almost perfectly aligned to the Yarm.

What we did:

Set PRM DC biases to 0 for both pitch and yaw.

Aperture was attached to PRM cage, double aperture was attached to BS cage, free-standing aperture was placed in front of PR2. 

Adjusted PZT1, PZT2 such that we were centered on PZT2, and through apertures at PRM and PR2.   This was mainly for setting beam height in PRC.

Checked centering on PZT1, MMT1, MMT2, PZT2.

Adjusted PRM pitch bias and PZT2 yaw such that REFL beam was retro-reflected from PRM.

Checked that REFL beam came nicely out of Faraday.

Checked that beam was still going through center of PRM aperture, and pitch height at PR2 was good.

Moved PR2 sideways until beam hit center in yaw of PR2.

Twisted PR2 such that beam was hitting center of PR3.

Moved and twisted PR3 (many times) so that beam went through BS input and output apertures, and through center of ITMY aperture.

Found that beam was just getting through black glass aperture at ETMY, top left corner, if looking at the face of ETM from ITM.

Locked down dog clamps on PR2.

This required some re-adjustment of PR3.  Re-did making sure going through BS apertures and ITMY aperture, locked down PR3 dog clamps.

Found that we are centered in yaw at ETMY, a little high in pitch on ETMY.

Replaced all of the light doors, to take a break.  4 hours in bunny suits seemed like enough that we earned a break.

This all sounds more straighforward than it was.  There was a lot of iteration, but we finally got to a state that we were relatively happy with.

 

What we will do:

Tweak PZT2 a *tiny* bit in pitch, ~0.5 mrad, so that the beam goes through the ETMY aperture.

See if we can align EMTY and ITMY to get multiple bounces through the Yarm.

Remove ETMX heavy door, steer BS such that we're getting through the center of an aperture at ETMX.

Align ETMX and ITMX such that we get multiple bounces through the Xarm.

Check SRM, AS path alignment.

Check REFL out of vac alignment.

Check other pickoffs.

Check all oplevs.

Check IPPOS/IPANG

 

We have a open-sided 2" mirror mount that we are considering using for the POY pick-off mirror.  This might help us get a little more clearance in the Y-arm of the Michelson.  Problem is the mount is not steerable, so we need to determine if that's doable or not.

 

  7642   Tue Oct 30 11:51:45 2012 JenneUpdateAlignmentPRC aligned, Yarm almost aligned

[Raji, Jenne]

We tweaked PZT2, PZT1 (yaw only), and PR3 (pitch only) to get the beam ~centered on the BS aperture, the ITMY aperture, and the ETMY aperture.

After lunch I'll tweak up the MC alignment, since, although the spots are in the right places, the transmitted beam could be higher power.  This will make it easier to check our pointing, especially since the ETMY spot is larger than our aperture, but the beam is dim.

We're getting there!

  7647   Wed Oct 31 17:18:34 2012 JenneUpdateAlignmentprogress, then setback - trend of BS table shift

Here is a two hour set of second trends of 2 sensors per mirror, for BS, PRM, ITMY and MC1.  You can see about an hour ago there was a big change in the BS and PRM suspensions, but not in the ITMY and MC1 suspensions.  This corresponds as best we can tell with the time that Jamie was figuring out and then fixing PZT2's mount.  You can see that the table takes some time to relax back to it's original position.  Also, interestingly, after we put the doors on ~10 or 20 minutes ago, things change a little bit on all tables. This is a little disconcerting, although it's not a huge change.

Attachment 1: PRM_BS_table_bumped_ITMY_MC1_no_big_change_2hoursBack.png
PRM_BS_table_bumped_ITMY_MC1_no_big_change_2hoursBack.png
  7658   Thu Nov 1 19:28:48 2012 JenneUpdateAlignmentaligned, AS beam on camera

After everyone else did the hard work, I moved the AS first-on-the-table steering mirror sideways a bit so the AS beam is on the center of the mirror, then steered the beam through the center of the lens, onto the 2" 99% BS.  I also moved the camera from it's normal place (the 1% transmitted through that BS) to the AS110 PD path, as we did last vent.  We'll need to put it back before we go back to high power.

  7662   Fri Nov 2 14:37:36 2012 JenneUpdateGeneralOpen-sided mount - why

Quote:

Quote:

Suprema- SS clear edge mirror mount 2" diameter is modified for 40m vacuum use. One left and one right handed one. It's adjustment screw housing is bronze! It is not ideal for out gassing.

It will be baked and scanned. If it passes we should use it.

We may need these to bring out some pick-off beams.

 I vote against it. We don't know about the grease inside the screw bushings - scans are not everything if adjusting the screw loosens up the grease. If we need more pick off mirrors lets just make some of the kind that we already use inside for the 2" optics.

 I think Steve had these prepared in response to my question a few days ago of how badly do we need adjustability for the POX/POY mirrors?  We already have cleaned open-sided mounts that have no adjustment screws.  So as long as the beam reflects off the ITMs horizontally (which it should), we can do yaw adjustment by twisting the whole mount. We don't need super fine yaw adjustment, we just need to get the beam out, so this is probably good enough.

We should put the POY mirror on this open-sided mount (the one without screws) some time.  Perhaps even today.

  7664   Fri Nov 2 19:59:54 2012 JenneUpdateAlignmentPOP, POX, POY, IPPOS, IPANG, REFL all coming out of vac

[Evan, Jenne, Jamie]

We used the green laser pointer technique to adjust the POP steering mirrors behind PR2 to get the POP backward beam out onto the table (rather, the mirrors were adjusted so that the green laser pointer, mounted on the POX table, was co-aligned with the beam between PR2 and PR3).

We were unable (why? I feel like it wasn't so hard last time) to see the POX beam, with a camera pointed at an IR card.  We ended up just waving a lens-free CCD camera around on the POX table where we expected POX to be, found the beam, and decided that if the beam was getting to the table, that was good enough.

We then waved the camera around on the POY table, and found the POY beam on the table.  We also moved ITMY up and down in pitch, and saw that the POY beam was moving, so we were satisfied that we had the correct beam.  We should go back and do this same check with POX, although I'm pretty sure that we already have the correct beam.  But checking is good.

We confirmed that IPPOS was coming out of the chambers.  I didn't end up touching any in-vac mirrors for IPPOS, since they all looked centered, and the beam on the table was already centered on the steering mirror on the out-of-vac table.

We got IPANG out of the chamber to the ETMY table.  IPANG has, after the pickoff window, an adjustable mirror, and then a fixed mirror on the BS table.  The beam was very close to the edge, in yaw, on that fixed mirror.  Jamie unclamped it and moved it so the beam was centered, then twisted it until I got beam back down at the end, centered on the first steering mirror down there.  Then Evan and I got the beam centered on the other steering mirror on the in-vac ETMY table, and got the mirror to ~the center of the first out-of-vac steering mirror.  Then Evan adjusted the other steering optics so the beam was hitting the QPD.

We then got the real REFL beam out of the chambers.  I still don't know what that ghost/fake beam is.  Anyhow, we moved PRM around, and saw that the real REFL beam moves, while the fake one doesn't.  We adjusted the adjustable REFL steering mirror in-vac such that the real REFL beam came out to the table.  Once on the AP table, we moved the PRM around again, just to be doubly/triply sure that we had the correct beam.  We put a beam splitter (found on the SP table) after the lens in the REFL path on the AP table, and put the camera on the reflected side of that BS.  This is because, like the AS port, the beam is too dim at the normal camera spot (which for REFL is the transmission through a Y1 mirror).

Jamie has centered IPPOS and IPPANG QPDs, so we should look at the weekend trend come Monday, to see what things look like, and how they drift, if at all.

 

On Monday, we should:

* Check the alignment, and the centering of beams on all mirrors one last time

* Remove all apertures from suspended optics (I think BS and PRM may be the only two that have them at this time)

* Check oplev paths for all mirrors

* Check all pickoffs / beams that need to come out of the vacuum

* Start putting on doors

  7665   Fri Nov 2 20:41:53 2012 JenneUpdateAlignmentAS, REFL camera shots

These don't show anything too interesting, but we're including them to show where the beams are right now on the cameras, so we can compare on Monday.

 

AS:

AS_2Nov2012.png

 

REFL:

REFL_2Nov2012.png

 

 

  7668   Mon Nov 5 09:53:35 2012 JenneUpdateAlignmentAS, REFL camera shots

Today's photos:

AS:

AS_5Nov2012.png

 REFL:

REFL_5Nov2012.png

IPANG / IPPOS trends:

IPPOS_IPANG_weekend_trend.png

 

 c.f. screen caps from Friday:

Quote:

These don't show anything too interesting, but we're including them to show where the beams are right now on the cameras, so we can compare on Monday.

 

AS:

AS_2Nov2012.png

 

REFL:

REFL_2Nov2012.png

 

 

 

  7683   Wed Nov 7 15:51:44 2012 JenneUpdateAlignmentJamie's tip tilt proposal

Steve's elog 7682 is in response to the conversation we had at group meeting re: Jamie's proposed idea of re-purposing the active tip tilts.

What if we use the active TTs for the PR and SR folding mirrors, and use something else (like the picomotors that Steve found from the old days) for our input steering?

  7685   Wed Nov 7 19:07:45 2012 JenneUpdateCamerasNo beam seen on external camera views

I have written some scripts which collect photos, then average them together, and subtract out an averaged background (as Rana described in elog 7678). 

I am not seeing any beam spots on any of the resulting pictures. 

 

The script to get 500 pictures is

.../scripts/general/videoscripts/videocapture50

and it's inputs are {name of camera} {folder to save in} {noBeam or withBeam}, where noBeam and withBeam indicate whether or not the PSL shutter is closed.  For the saved photos to work nicely with the Matlab script, the folder to save in should be in the format (Month_day_year/CAMERA).  So today's ITMYF pics, for example, are in Nov_7_2012/ITMYF/ .

So, you run it once with the shutter open, and once again with the shutter closed.

 

To create the new picture, open ImageBkgndSubtractor.m in the same .../scripts/general/videoscripts folder, edit the top few lines (month, day, year, camera name).  Run it, and it will read through all the pictures and supply a background-subtracted output, and save the output (as well as a version where every pixel value is multiplied by 3) in the same folder as the 500 pictures.

The pictures are all saved in

/opt/rtcds/caltech/c1/scripts/general/videoscripts/photos

so really, for my example above, it would be /opt/rtcds/caltech/c1/scripts/general/videoscripts/photos/Nov_7_2012/ITMYF/, with 2 subfolders, noBeam and withBeam, and the final pictures are saved in /opt/rtcds/caltech/c1/scripts/general/videoscripts/photos/Nov_7_2012/ITMYF/ .

 

In other, semi-unrelated news, the ITMXF camera has been not working for a while.  The bottom right quad on the test mass tv has been dark for at least a week or two.  Steve, when you have a chance (after the oplevs are all taken care of), can you see if there's something obvious that's wrong?

Here are the background subtracted photos that I've taken today:

BS_PRM_7Nov2012_SpotImage_pixelsTimes3.png

ETMXF_7Nov2012_SpotImage_pixelsTimes3.png

ETMYF_7Nov2012_SpotImage_pixelsTimes3.png

ITMYF_7Nov2012_SpotImage_pixelsTimes3.png

MC2F_7Nov2012_SpotImage_pixelsTimes3.png

MC2F is included, even though you can see the spot usually, just to prove that I'm not trying to subtract away the spot!  You just can't see it in any other picture.

  7689   Thu Nov 8 20:11:54 2012 JenneUpdateAlignmentAS steering moved out of POY's way, 2 green beams onto PSL table

[Jenne, Jamie, Manasa, Ayaka]

 

Flipped mount of OM2, moved OM2 behind POY pickoff so we're out of the way of POY.  Adjusted and recovered rest of AS path.

We found that IPANG was not on its photodiode, but determined that it was centered on all of the in-vac mirrors, and that it was just a little bit of steering on the ETMY end out-of-vac table that needed to be done.

Got green flashes in Yarm, moved down periscope to the north by ~1 inch in order to get y green out to PSL table.  This also involved moving the steering mirror on the IOO table immediately after the down periscope to match.  We measured the MC spot positions before and after touching the periscope, and there was no significant change. 

Aligned X green to X arm (centered on ITMX, ETMX, although no flashes since we didn't move ETMX's biases around), then made sure it was centered on all of its steering mirrors, and came out of the vacuum.

Manasa took photos of all test mass chambers and the BS chamber, so we can keep up-to-date CAD drawings. 

Oplevs and IPPOS/IPANG are being centered as I type.  Manasa and Ayaka are moving the lens in front of IPANG such that we have a slightly larger beam on the QPD.

 

In the morning, Jamie is going to put apertures back on 2 of the suspended mirrors for one last check that moving things on the IOO table didn't do anything bad, but since the AS and REFL beams on those cameras didn't move significantly, we think things are fine. 

Heavy doors go on in the morning, and access connector at ~1pm, if not before lunch.  Then Steve will start pumping early Monday morning!  Hooray!

 

PS, for reference,

AS: AS_8Nov2012.png

REFL: REFL_8Nov2012.png

  7699   Tue Nov 13 00:15:08 2012 JenneUpdateIOOInput and Output PZTs turned on

I turned on the power supplies for the output PZTs and pitch and yaw for PZT2.  This is back to the condition that we had during atmosphere alignment, so after Ayaka has finished tweaking the MC, we can have a look at alignment of the interferometer.

  7700   Tue Nov 13 00:19:51 2012 JenneUpdateCamerasITMX is just fine

Quote:

In other, semi-unrelated news, the ITMXF camera has been not working for a while.  The bottom right quad on the test mass tv has been dark for at least a week or two.  Steve, when you have a chance (after the oplevs are all taken care of), can you see if there's something obvious that's wrong?

 While helping Charles string the ethernet cable (elog 7698) for the power strip in the vertex (from 1Y1 to 1X1), I looked at the ITMX power cord.  It was connected to the same power strip as the illuminator, which, since the illuminator was turned off by turning off the power strip, meant no power was going to the camera.  Since Charles is very close to having the new power strips set up, I unplugged the ITMX illuminator and turned the power strip back on.  ITMX camera is back to normal.

  7701   Tue Nov 13 00:27:50 2012 JenneUpdateElectronicsEthernet Illuminator Control

Quote:

 

 The ethernet power strips have been installed. 192.168.113.110 is on ETMX, 192.168.113.111 is on ETMY and 192.168.113.112 is on the vertex. I have also written an EPICS file "illuminator_control.adl" (currently stored in my named directory) that allows a user to turn individual sockets on and off at each of the three locations. Some short tests have indicated that everything is in working order.

Currently, no illuminators are hooked up to the power strips. However, the power control will most likely be ready for use tomorrow, granted I can find and use extension cords so that the illuminators might reach their respective power strips.

 I'm sure Charles meant to also say that he connected the ETM power strips to the ethernet switches in those racks.  For the vertex, the ethernet switch is in 1X2, but there isn't space in there, so the power switch was installed in 1Y2.  The vertex ethernet cable is along the overhead inside cable tray.

I'm not sure what we want to do about connecting the new power strips to the illuminators.  No illuminator is close enough that its built-in cable can reach the power strip, so we'll need extension cables or some such.  Charles is going to ask Steve about the plan tomorrow.

  7709   Tue Nov 13 22:12:15 2012 JenneUpdateIOOTurn off MCL path before doing MC spot measurement

[Jenne, Den]

We have discovered that the MCL loop squishes the length fluctuations that result from the MC spot measurement angular dither.  This is good, in that the MCL is doing what it ought to do.  However, we need to turn it off before doing a spot measurement.

  7710   Wed Nov 14 00:56:19 2012 JenneUpdateIOOWFS on, PZT2 ~set, AS camera back

WFS are back on, and working nicely.  Den and I had seen a problem (which I had seen before) that when you turn on the integrators in the WFS loops, the MC Refl value gets worse (goes up).  Koji reminded me that he had a nice elog (7452) on how to get the MC awesome.  Ayaka and I last night stopped after Step 7.  Step 8, zeroing the WFS offsets, is the secret important thing that I keep forgetting.  I ran the script /opt/rtcds/caltech/c1/scripts/MC/WFS/WFS_FilterBank_offsets, then turned on the WFS loop, and the WFS are working just fine now.   

I'm back to wishing that I had control over PZT1.  I went back and forth for a while between 1Y4 and the ETMY table to get the IPANG beam centered on the QPD.  Initially, the beam was coming out of the vacuum a little high.  The digital HV power supply is pitch, and the analog HV power supply is yaw.  When I get the beam reasonably well centered on IPANG, with a beautiful, non-clipped, beam, the beam is much too high on IPPOS.  The beam barely hits the top of the first out of vac steering mirror, and then is too high on the QPD.  It looks like (based on the sum value) that the beam is on the diode, just entirely in the upper 2 quadrants.  But if I try to fix this, since IPANG has a longer lever arm, the IPANG beam doesn't come out of the vacuum.  I have compromised by getting the beam on IPANG, centered in pitch and ignoring IPPOS pitch.  For yaw, since moving PZT2 only makes one of POS or ANG good at a time, I split the difference, so the beam is not really centered in yaw on either, but it's close on both.

AS beam is back on the camera.  I forgot that, especially since the beam at REFL is pretty bright, I had moved the AS and REFL cameras out of the beam so we didn't crispy-fry the CCDs.  Therefore, the camera spots are no longer a reference of spot position. We can still eyeball the position on, say, the 2" lens, but that's not any kind of accurate.  I put the AS camera back to it's normal place, so the AS beam is going toward AS55 and AS110, and a little bit is going to the camera.  I have not yet aligned the beams actually on to the PDs.

REFL beam is dumped by a razor dump after the 2" lens.  Manasa did some work (elog 7666) to the REFL path, and I'm not 100% sure how it was before, so I leave it to her to please work on during the day tomorrow.  I think we need to put back the Y1 that used to be there, but I don't know where the optic is.  I put a yaw dither on the PRM with awggui, and saw the REFL beam moving at my 2Hz dither frequency, so this time we actually have a useful beam coming out.

  7712   Wed Nov 14 13:58:18 2012 JenneUpdateIOOTurn off MCL path before doing MC spot measurement

Quote:

Quote:

We have discovered that the MCL loop squishes the length fluctuations that result from the MC spot measurement angular dither.  This is good, in that the MCL is doing what it ought to do.  However, we need to turn it off before doing a spot measurement.

 This is totally non-sensical statement, of course.

We might also say that the DARM loop totally squishes the GW signal and so its better not to have any feedback in the interferometer.

 Hmmm, indeed.  To keep MC_L on, we should be looking at the control signal rather than the error signal.  I think Den has MC_L running nicely such that it always comes on when the MC locks, so I can switch it.

  7715   Thu Nov 15 03:09:08 2012 JenneUpdateAlignmentNo good progress, many options
I didn't make any concrete progress today. AS and REFL cameras are in place, and Manasa has put ND 0.5 filters on both. I used a 
camera to look at the back of the Faraday, and aligned PRM such that it was retroreflecting, and then tried to align ITMY to have once 
fringes with the PRM at that place. I failed in this, since the AS beam on the AS table was starting to dall off the first mirror on the table. 
I then restored all the suspensions to where they were before I started touching them today. 

I moved ETMY face camera so that it is looking at the front of the black glass, but it's hard to tell where the beam is.  I was thinking 
about setting up a temporary camera to look at the back of ITMY to help guide PZT steering, but I haven't done this yet. 

Koji and I then talked about the several different options I have for references, and how many different knobs I  can turn. I'm sleeping 
on it for now, and hopefully I'll have more insight on what to do tomorrow. 
  7719   Fri Nov 16 09:57:57 2012 JenneUpdateGreen Lockingxarm locked

Quote:

We aligned and locked xarm for green.

 

 That's really, really awesome!

 

  7723   Mon Nov 19 15:12:52 2012 JenneUpdateAlignmentYarm locked IR

Quote:

Quote:

POY11 does not go out of the vacuum

 It does but slighty low and does not get on mirrors. We need to change optic mounts to adjust the height. Red is flashing in yarm at 00 and 10 modes. TRY is ~0.4-0.5.

I've adjusted BS angle, camera and TRX PD at ETMX table so I can see red flashing at 03 mode while green is locked to 00 and its transmission is maximized. I thought that by adjusting BS angle, I will be able to align red to 00 not disturbing green, but this was not the case. Maximum TRX I could get was 0.1. I've adjusted POX to get into PD and I can see PDH signal though I can't lock as cavity is still misaligned for red.

 [Ayaka, Jenne]

We put the POY beam onto the POY PD.  The Yarm is currently locked on IR with ~0.65 transmission.

 

  7724   Mon Nov 19 15:15:22 2012 JenneUpdateSUSAll oplev gains turned on

Quote:

Oplev values that were changed to zero:

PRM P=0.15, Y=-0.3

SRM P=-2.0, Y=2.0

BS P=0.2, Y=-0.2

ITMY P=2.1, Y=-2.0

ITMX P=1.0, Y=-0.5

ETMX P=-0.2, Y=-0.2

ETMY P=0.5, Y=0.6

Also, PRCL was changed in the LSC input matrix from REFL33I to AS55I, since there is no REFL beam out of the IFO :(

 Ayaka and I restored all of the oplev gains to these values.  The exception is ETMY, which has both gains negative.  I am unsure if this is a transcription error on my part, or if something physical has changed.  The layout of the ETMY oplev was modified (since Rana took out the offending lens) but that shouldn't affect the sign of the gains.

  7727   Mon Nov 19 20:17:53 2012 JenneConfiguration40m UpgradingEndtable upgrade for auxiliary green laser : ETMX layout on new table

 

For convenience, I would include a steering mirror in front of the TRX PD.  Also, don't we usually have lenses in the oplev paths?  Also, also, do we need lenses in front of the green refl and TRX PDs?  Do you have a place in mind for the shutter?  Is there a way to compactify the layout a little bit, so that even if the lenses are different for each table, the general layout for both ETMX and ETMY is the same (with an empty space on ETMX where IPANG belongs on ETMY)?  I'm sure it is, since you've talked to Steve about this, but just to check: is the green refl PD far enough away from the edge of the table to accommodate the fancy new box?

  7745   Mon Nov 26 18:36:17 2012 JenneUpdatePEMBLRMS back

Quote:

 I got two seismometers and one microphone back from Tara.

They are now near the Gurlap under the MC.

 I have finally plugged GUR1 back in....it is down at ETMY for now, since that's where the cable was.  BLRMS are back up on the projector.

  7746   Mon Nov 26 18:56:34 2012 JenneHowToComputersData logging suggestions

We've been talking for a while about how we want to store data.  I'm not in love with keeping it on the elog, although I think we should always be able to reference and go back and forth between the elogs and the data.

I have made a new folder: /data    EDIT: nevermind.  I want it to be on the file system just like /users, but I don't know how to do that.  Right now the folder is just on Ottavia. Jamie will help me tomorrow.

In this folder, we will save all of the data which goes into the elog. 

I propose that we should have a common format for the names of the data files, so that we can easily find things.

My proposal is that one begins ones elog regarding the data to be saved, and submit it immediately after putting in the first ~sentence or so. One should then make a new folder inside the data folder with a title "elog#####_Anything_Else_You_Want" Then, data (which was originally saved in ones own users folder) should be copied into the /data/elog#####_AnythingElse/ folder. Also in that folder should be any Matlab scripts used to create the plots that you post in the elog.  One should then edit the elog to continue making a regular, very thorough elog, including the path to the data.  Elog should include all of the information about the measurement, state of the IFO (or whatever you were measuring), etc. 

Riju will be alpha-testing this procedure tonight.  EDIT: nevermind...see previous edit.

  7752   Tue Nov 27 03:26:00 2012 JenneUpdateGeneralPOP beam on POP55, POP camera

Quote:

Some obvious things to be fixed

-  We need POP55 and POP CCD for diagnoses.

 Done.  The beam is also going vaguely in the direction of POP110, but I can't see the beam, so it's tricky. 

Order of operations:

1. Find POP on the table, place iris so I wouldn't forget.  Find beam by putting big IR card where I think beam should be, look at IR card with IR viewer.

2. Move and re-clamp 2" lens so beam is on center of lens.

3. Move and re-clamp 1st (2") mirror so that beam is on center of mirror. 

4. Remove BS-33% so that all the beam goes to POP55, steer that 1st mirror so beam is on POP55's little mirror.  Align little mirror so beam is centered on POP55 (as seen by looking at PD with viewer, finding "edges" of PD, going back to center).

5. Put BS-33% back in place.  The reflected portion of this beam is not possible to see using card+viewer technique.

6. Remove BS-50% that reflects half of this beam to POP110.  Find beam reflected from BS-33% by waving POP camera around.  Steer BS-33 until beam goes back in the direction that the camera used to be mounted.  Adjust camera mount and BS-33 so that beam is on camera.

7. Put BS-50% back in place.  Steer it around with voltmeter on PD to see if beam ever hits PD.  Unsuccessful. Give up, since we have POP55, and POP camera.

8. Make a youtube video: POP, AS, REFL, ITMXF (all on Quad3) - PRMI coarsely aligned, no IFO parts locked.  MICH was locked earlier, but not during video time.

  7816   Wed Dec 12 16:52:12 2012 JenneUpdateAlignmentPR2_face, PR3_back cameras in place

I have setup cameras looking at the back of PR3 (through the north viewport on the MC chamber) and the face of PR2 (through the north viewport on the ITMX chamber). We would like a view of the face of PR3, but that isn't possible without placing another in-vac mirror.  The best we can do is the current PRM_BS camera setup, which sees a small portion of the PR3 face.  Most of the face is obscured by the PRM itself. 

I have taken images with the PRM misaligned.  The spot near the top of PR2 is the first reflection from the pitch-misaligned PRM, so it should be ignored for the purposes of trying to see the straight-shot, no PRM beam.

Images are taken with my videocapture50 script, in ..../scripts/general/videoscripts.  This takes 10 sets of 50 images and saves them.  Then ImageBkgndSubtractor.m located in the same folder takes the images (you must edit the beginning of the script to tell it where the images are), averages the noBeam images (PSL shutter closed), and averages the withBeam images, and subtracts them.  Results below:

PR2_face_12Dec2012_SpotImage.pngPR2_face_12Dec2012_SpotImage_pixelsTimes3.png

PR3_back_12Dec2012_SpotImage.pngPR3_back_12Dec2012_SpotImage_pixelsTimes3.png

  7828   Fri Dec 14 03:15:49 2012 JenneUpdateSUSTT angle of incidence

But have you looked yet at what angle we need? The first input string mirror has a quite small incidence angle. The other input steering mirror maybe borderline, based on your estimates. Also, have we considered just having new brackets made and cleaned? Is there a reason we would prefer to modify the ones we have?

ELOG V3.1.3-