40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 126 of 344  Not logged in ELOG logo
ID Date Author Type Categoryup Subject
  7436   Mon Sep 24 23:45:39 2012 ranaUpdateGeneralPOP in-vac optics aligned, POY beam coming out of vac

  What was the reasoning / resolution of the POP forward/backward beam? Are we going to have the right beam for DRMI locking?

  7437   Tue Sep 25 14:29:07 2012 JenneUpdateGeneralPOP in-vac optics aligned, POY beam coming out of vac

Quote:

  What was the reasoning / resolution of the POP forward/backward beam? Are we going to have the right beam for DRMI locking?

From Koji's email to me:

"With the backward beam you can see the returning beam even when the PRM is misaligned. That's the only difference. Once the PRM is aligned both beams have the same information."

So, we should be fine.

  7476   Thu Oct 4 08:39:58 2012 SteveUpdateGeneralpower outage

There had to be a power outage. Laser and air condition turned back on. The vacuum is OK

Sorensen DC power supplies were tripped, so they were reset: at AUX OMC South 18V and 28V for RF PS and at 1X1 24V

 

Power Outage confirmed:

** Notification **

 

CALIFORNIA INSTITUTE OF TECHNOLOGY

                 FACILITIES MANAGEMENT

 

**PLEASE POST**

 

 

Building:         Campus

 

Date:             Thursday October 04,2012

 

This morning at 2:17 a.m. much of the City of Pasadena including our Campus experienced a electric power sag of short duration, approximately 1/10 of a second. The cause was a fault on one of Pasadena’s 17KV circuits. Some sensitive equipment have been impacted.

                 

Contact:          Mike Anchondo x-4999

 

Attachment 1: Oct4R2012.png
Oct4R2012.png
  7493   Fri Oct 5 16:21:48 2012 SteveUpdateGeneralfreshmen visiters

40 plus freshmen visited the 40m today

Attachment 1: IMG_1693.JPG
IMG_1693.JPG
Attachment 2: IMG_1694.JPG
IMG_1694.JPG
  7527   Thu Oct 11 11:20:05 2012 janoschUpdateGeneralbeam shape simulation, PRC

I started to create a Finesse model of the PRC cavity. We have the phase maps for the PRC and the two ITMs. I could not find anything for PR2,3 and BS. All files can be found in my SVN folder /janosch/PRC40m. I used the AutoCAD model to determine angles of incidence and distances. These numbers are largely inconsistent with numbers that you can find elsewhere on the 40m wiki, but this certainly depends on what accuracy is required for interferometer alignment and I don't understand anything about alignment.

The phase maps come in a format that needs to be modified before they can be used in Finesse. I have started with this work, but maybe someone else can take over. The phase maps show tilts and the PRC also has the curvature. These have to be subtracted out before the maps can be loaded into Finesse. I asked GariLynn for the code that they use. The Finesse model (MichPRC_40m.kat) does not load the phasemaps yet, and I just wrote some random parameter values for the TEM00 input beam to the PRC. So these Gauss parameters need to be corrected.

I will only go on with this work if Rana tells me that I should do so, otherwise it is on hold until we have a volunteer.

  7533   Thu Oct 11 21:26:40 2012 janoschUpdateGeneralPRC phase maps

Just some plots. There is nothing new here except for the fact that I learned how to analyze phase maps myself and how to prepare them for Finesse. In other words, everything is ready for a Finesse simulation.

These phase maps show the raw measurement of ITMY, ITMX and PRC:

ITM03_HR_raw.pngITM04_HR_raw.pngPRM02_HR_raw.png

Subtracting out the tilt from all phase maps, and the curvature from the PRC (I found the fit 121m consistent with previous fits), the one obtains the following residuals that can be used in Finesse (order is again ITMY, ITMX and PRC):

ITM03_HR_untilted.pngITM04_HR_untilted.pngPRM02_HR_untilted_uncurved.png

Attachment 3: PRC_40m.png
PRC_40m.png
  7534   Fri Oct 12 01:56:26 2012 kiwamuUpdateGeneralAlignment situation of interferometer

[Koji / Kiwamu]

 We have realigned the interferometer except the incident beam.

 The REFL beam is not coming out from the chamber and is likely hitting the holder of a mirror in the OMC chamber.

So we need to open the chamber again before trying to lock the recycled interferometers at some point.

 

--- What we did

  •  Ran the MC decenter script to check the spot positions.
    • MC3 YAW gave a - 5mm offset with an error of about the same level.
    • We didn't believe in this dither measurement.
  •  Checked the IP-POS and IP-ANG trends.
    • The trends looked stable over 10 days (with a 24 hours drift).
    • So we decided not to touch the MC suspensions.
  • Tried aligning PRM
  • Found that the beam on the REFL path was a fake beam
    • The position of this beam was not sensitive to the alignment of PRM or ITMs.
    • So certainly this is not the REFL beam.
    • The power of this unknown beam is about 7.8 mW
  • Let the PRM reflection beam go through the Faraday
    • This was done by looking at the hole of the Faraday though a view port of the IOO chamber with an IR viewer.
  • Aligned the rest of the interferometer (not including ETMs)
    • We used the aligned PRM as the alignment reference
    • Aligned ITMY such that the ITMY reflection overlaps with the PRM beam at the AS port.
    • Aligned the BS and SRM such that their associated beam overlap at the AS port
    • Aligned ITMX in the same way.
    • Note that the beam axis, defined by the BS, ITMX  and SRM, was not determined by this process. So we need to align it using the y-arm as a reference at some point.
    • After the alignment, the beam at the AS port still doesn't look clipped. Which is good.

 

---- things to be fixed

   - Align the steering mirrors in the faraday rejected beam path (requires vent)

   - SRM oplev (this is out of the QPD range)

   - ITMX oplev (out of the range too)

  7558   Tue Oct 16 16:46:20 2012 Vent Czar and CzarinaSummaryGeneralvent stuff

We're discussing the plan for the next vent.  Now that we have all the tip-tilt stuff in hand, we should get ready for the big TT installation vent ASAP.  The question is what remains to be done, and when can we be ready

Stuff that needs doing:

* characterize TTs, check electrical connection of quadrupus cables, bake cables and base plates - a week or so (not including baking)

* phase map of LaserOptik mirrors (decide on needed resolution) - a week or so?

* get ready to swap G&H mirrors with LaserOptik mirrors (PR2, PR3, SR2, SR3)

* assembly of black glass beam dumps - one afternoon

* green periscope moving - day or so to make sure greens are resonant in arms, so we have something to align to

* redraw of REFL path?? (rotate 2nd refl mirror to reflect beam to the east, then add mirror to get beam out of IOO chamber viewport), cut hole in BS oplev table's box, install black beam tube

* FI camera mount - make sure it's here, baked

* active TT pitch damping plan

* passive TT weight addition and pitch alignment plan

* camera plan for taking in-chamber photos without touching the tables

* look at layout of ITMY table.  POY pick-off too close to main beam.  Can we move POY pick-off to BS table?

* remount black glass baffles on SOS cages.

 

Stuff to bake:

* Quadrupus cables

* TT base plates

* FI mirror mount for FI camera

* stuff to add to active TTs

* stuff to add to passive TTs

 

  7560   Tue Oct 16 17:13:23 2012 CzarinaSummaryGeneralvent stuff - 4 paths

I see 4+ possible paths for us to take, in terms of a possible vent in the next few weeks:

No Vent - Just do FPPRMI, using AS55

Mini Vent - Fix REFL path, nothing else.  ~1 day at atmosphere

Medium Vent - Fix REFL path, swap G&H mirrors for LaserOptik mirrors (so also resuspend passive TTs, maybe add pitch adjustment option). ~1 week or so at atmosphere - do this rather than Mini if Jan's Finesse calc says the G&H mirrors are too rough

Mega Vent - Fix all the things, do all the things.  Long time at atmosphere

The "+" is to take into account all the possible variations on "medium vent".  The No, Mini and Medium options assume we'll do the Mega option later, just not immediately.

  7563   Wed Oct 17 08:01:54 2012 SteveUpdateGeneraldo not leave op-table open

ETMX optical table was left open. Burned toast award goes to ......?

  7565   Wed Oct 17 08:05:10 2012 DenUpdateGeneraldo not leave op-table open

Quote:

ETMX optical table was left open. Burned toast award goes to ......?

 me

  7627   Thu Oct 25 22:52:07 2012 janoschUpdateGeneraltip-tilt phase maps

Now that I read Koji's last elog about phase maps, I am not sure if these are still required, but here they are (the tilt-removed phase maps of the Laser Optik mirrors), first 1, 2, 3:

sn1Laseroptik_untilted.pngsn2Laseroptik_untilted.pngsn3Laseroptik_untilted.png

Then 4,5,6:

sn4Laseroptik_untilted.pngsn5Laseroptik_untilted.pngsn6Laseroptik_untilted.png

So they all have an elevated center. I am not sure why the phase maps of mirrors 5 and 6 are slightly smaller in dimension. Anyhow, all mirrors have quite strong aberrations. Also, there is no big difference between the mirrors. Check for yourself, but be careful with the colors since the scales are all different.

  7628   Thu Oct 25 23:00:44 2012 ManasaUpdateGeneraltip-tilt phase maps

Are these maps drawn from the data we extracted using Image SXM??

  7629   Thu Oct 25 23:14:42 2012 janoschUpdateGeneraltip-tilt phase maps

Quote:

Are these maps drawn from the data we extracted using Image SXM??

 Indeed. So the only manipulation that I did was to remove the tilt (since this should usually be seen as an artifact of the measurement, or better, we can assume that tilt is compensated by alignment). I did not remove the curvature.

  7638   Mon Oct 29 11:27:42 2012 ManasaUpdateGeneraltip-tilt phase maps

 [Jan, Manasa]

Below are phasemaps for the tip-tilts with both tilt and RoC removed. We have not used Koji's code; but tweaked the earlier code to remove curvature.

The RoC values matched approximately to that quoted by Gari Lynn ~700m.

RoC of tip-tilts
Mirror
RoC (m)
SN1 748.7176
SN2 692.7408
SN3 707.0336
SN4 625.5152
SN5 672.5340
SN6 663.7791

 

Phasemaps

The color scale for height are not the same for all mirrors.

 

SN1, SN2 and SN3

sn1_UC_UT.pngsn2_UC_UT.pngsn3_UC_UT.png

SN4, SN5 and SN6

sn4_UC_UT.pngsn5_UC_UT.pngsn6_UC_UT.png

  7639   Mon Oct 29 14:57:41 2012 janoschUpdateGeneraltip-tilt phase maps

Quote:

 [Jan, Manasa]

Below are phasemaps for the tip-tilts with both tilt and RoC removed. We have not used Koji's code; but tweaked the earlier code to remove curvature. 

 The posted residual phase maps show circular contours since the data came with relatively low resolution in height. This is ok for what we want to do with these phase maps (i.e. simulating higher-order mode content in the PRC using Finesse). Better resolution is only required if you want to understand in detail optical scattering out of the cavity. Anyhow, the circular artifacts can be removed by first interpolating the phase maps to a higher lateral resolution, and then performing tilt and curvature subtraction. So we will soon have better looking phase maps posted. Then we should think about what type of Finesse simulation we could run. Certainly one simulation is to look at the beam shape in the PRC, but more interesting could be how sensitive the shape is to mirror alignments. The current simulation shows a mode that resembles the TEM01, but I have not yet tried to find optimal alignment of the mirrors (in simulation) to search for the TEM00 mode.

  7656   Thu Nov 1 17:37:00 2012 SteveUpdateGeneralbronze bushing for 40m vac

Suprema- SS clear edge mirror mount 2" diameter is modified for 40m vacuum use. One left and one right handed one. It's adjustment screw housing is bronze! It is not ideal for out gassing.

It will be baked and scanned. If it passes we should use it.

We may need these to bring out some pick-off beams.

Attachment 1: IMG_1764.JPG
IMG_1764.JPG
  7660   Fri Nov 2 03:28:54 2012 ranaUpdateGeneralbronze bushing for 40m vac

Quote:

Suprema- SS clear edge mirror mount 2" diameter is modified for 40m vacuum use. One left and one right handed one. It's adjustment screw housing is bronze! It is not ideal for out gassing.

It will be baked and scanned. If it passes we should use it.

We may need these to bring out some pick-off beams.

 I vote against it. We don't know about the grease inside the screw bushings - scans are not everything if adjusting the screw loosens up the grease. If we need more pick off mirrors lets just make some of the kind that we already use inside for the 2" optics.

  7662   Fri Nov 2 14:37:36 2012 JenneUpdateGeneralOpen-sided mount - why

Quote:

Quote:

Suprema- SS clear edge mirror mount 2" diameter is modified for 40m vacuum use. One left and one right handed one. It's adjustment screw housing is bronze! It is not ideal for out gassing.

It will be baked and scanned. If it passes we should use it.

We may need these to bring out some pick-off beams.

 I vote against it. We don't know about the grease inside the screw bushings - scans are not everything if adjusting the screw loosens up the grease. If we need more pick off mirrors lets just make some of the kind that we already use inside for the 2" optics.

 I think Steve had these prepared in response to my question a few days ago of how badly do we need adjustability for the POX/POY mirrors?  We already have cleaned open-sided mounts that have no adjustment screws.  So as long as the beam reflects off the ITMs horizontally (which it should), we can do yaw adjustment by twisting the whole mount. We don't need super fine yaw adjustment, we just need to get the beam out, so this is probably good enough.

We should put the POY mirror on this open-sided mount (the one without screws) some time.  Perhaps even today.

  7663   Fri Nov 2 16:09:14 2012 SteveUpdateGeneralOpen-sided mount - why

Quote:

Quote:

Quote:

Suprema- SS clear edge mirror mount 2" diameter is modified for 40m vacuum use. One left and one right handed one. It's adjustment screw housing is bronze! It is not ideal for out gassing.

It will be baked and scanned. If it passes we should use it.

We may need these to bring out some pick-off beams.

 I vote against it. We don't know about the grease inside the screw bushings - scans are not everything if adjusting the screw loosens up the grease. If we need more pick off mirrors lets just make some of the kind that we already use inside for the 2" optics.

 I think Steve had these prepared in response to my question a few days ago of how badly do we need adjustability for the POX/POY mirrors?  We already have cleaned open-sided mounts that have no adjustment screws.  So as long as the beam reflects off the ITMs horizontally (which it should), we can do yaw adjustment by twisting the whole mount. We don't need super fine yaw adjustment, we just need to get the beam out, so this is probably good enough.

We should put the POY mirror on this open-sided mount (the one without screws) some time.  Perhaps even today.

 The hex adjustment screw  is removed from the brass bushing for degreasing, cleaning and baking. Newport says the  bushing is is brass not bronze.

Brass outgassing rate is 10x higher than copper that is truth. The surface area is small so  Krytox grease is ok if it is really needed in the 40m. ( It was ok in the past and it is still there )

  7670   Mon Nov 5 13:28:15 2012 jamieUpdateGeneral40m DCC document tree

Link to the new 40m DCC Document Tree: E1200979

  7693   Fri Nov 9 11:38:38 2012 jamieUpdateGeneralwe're closing up

After a brief look this morning, I called it and declared that we were ok to close up.  The access connector is almost all buttoned up, and both ETM doors are on.

Basically nothing moved since last night, which is good.  Jenne and I were a little bit worried about how the input pointing might have been effected by our moving of the green periscope in the MC chamber.

First thing this morning I went into the BS chamber to check out the alignment situation there.  I put the targets on the PRM and BS cages.  We were basically clear through the PRM aperture, and in retro-reflection.

The BS was not quite so clear.  There is a little bit of clipping through the exit aperture on the X arm side.  However, it didn't seem to me like it was enough to warrant retouching all the input alignment again, as that would have set us back another couple of days at least.

Both arm green beams are cleaning coming out, and are nicely overlapping with the IR beams at the BS (we even have a clean ~04 mode from the Y arm).  The AS and REFL spots look good.  IPANG and IPPOS are centered and haven't moved much since last night.  We're ready to go.

The rest of the vertex doors will go on after lunch.

  7694   Fri Nov 9 17:15:05 2012 Manasa, Steve, AyakaUpdateGeneralWe're closed! Pumping down monday morning

Quote:

After a brief look this morning, I called it and declared that we were ok to close up.  The access connector is almost all buttoned up, and both ETM doors are on.

Basically nothing moved since last night, which is good.  Jenne and I were a little bit worried about how the input pointing might have been effected by our moving of the green periscope in the MC chamber.

First thing this morning I went into the BS chamber to check out the alignment situation there.  I put the targets on the PRM and BS cages.  We were basically clear through the PRM aperture, and in retro-reflection.

The BS was not quite so clear.  There is a little bit of clipping through the exit aperture on the X arm side.  However, it didn't seem to me like it was enough to warrant retouching all the input alignment again, as that would have set us back another couple of days at least.

Both arm green beams are cleaning coming out, and are nicely overlapping with the IR beams at the BS (we even have a clean ~04 mode from the Y arm).  The AS and REFL spots look good.  IPANG and IPPOS are centered and haven't moved much since last night.  We're ready to go.

The rest of the vertex doors will go on after lunch.

Jamie and Steve got the ETM doors on this morning.

We got the other heavy doors including the ITMs, BS and the access connector in place.

If nobody raises any concerns in reply to this elog, Steve will assume it as a green signal and will start pumping down first thing Monday morning after the final check on the access connector bellow screws.

 

Steve! 

Ayaka and I got the ITMY and BS door closed at 45foot pounds just now. 

  7732   Tue Nov 20 15:11:22 2012 SteveUpdateGeneralfew more sensing cards

New  Lumitek IR Sensor Cards are here. We got 2 pieces of Q-11-T (2" x 2"), 2 pieces of Q-11-T (0.75" x 0.75")  and one Q-11 (4" x 5")

  7740   Sat Nov 24 22:14:08 2012 KojiUpdateGeneralAP table cleaning up

On Wednesday (21st) night, I checked the AP table as I wanted to try PRMI locking. 
It was difficult to work with the table as there were so many unnecessary components on it.
Also the beams went through complicated paths as they have funny angles. 

So I decided to clean up of IMC REFl WFS, IFO REFL, and IFO AS paths.
I found that the AS beam was highly astigmatic as the beam went through a (too-much-) tilted lens.
I made several blocked optical paths for REFL and AS for future extension of the detection system.

The current status of the table was uploaded below.

The optical spectrum analyzers and the aux NPRO were left untouched but they should be moved
somewhere (either on the table or outside) which does not disturb the other optical paths.
 


After the cleaning, I started locking PRMI. I could lock PRMI stably. But I could not figure out how
the intra-cavity mode looked like as I did not have the POP camera. The power recycling
gain was not quantitatively evaluated as I did not have POP and I wasn't sure how the beam was aligned at POX/POY.


We need to know:

- Quantitative evaluation of the beam shape in the PRC

- Quantitative evaluation of the power recycling gain

Some obvious things to be fixed

- The POX whitening filters seem not switching. This issue should be checked at the circuit module itself and at the BIO.

- The POX beam is not well focused on the PD. This was particularly clear when PRMI was locked with carrier.

- The POP beam is going nowhere. We need POP55 and POP CCD for diagnoses.

I haven't checked ITMY table.

Attachment 1: APTable121124.pdf
APTable121124.pdf
Attachment 2: APTable121124.graffle.zip
  7752   Tue Nov 27 03:26:00 2012 JenneUpdateGeneralPOP beam on POP55, POP camera

Quote:

Some obvious things to be fixed

-  We need POP55 and POP CCD for diagnoses.

 Done.  The beam is also going vaguely in the direction of POP110, but I can't see the beam, so it's tricky. 

Order of operations:

1. Find POP on the table, place iris so I wouldn't forget.  Find beam by putting big IR card where I think beam should be, look at IR card with IR viewer.

2. Move and re-clamp 2" lens so beam is on center of lens.

3. Move and re-clamp 1st (2") mirror so that beam is on center of mirror. 

4. Remove BS-33% so that all the beam goes to POP55, steer that 1st mirror so beam is on POP55's little mirror.  Align little mirror so beam is centered on POP55 (as seen by looking at PD with viewer, finding "edges" of PD, going back to center).

5. Put BS-33% back in place.  The reflected portion of this beam is not possible to see using card+viewer technique.

6. Remove BS-50% that reflects half of this beam to POP110.  Find beam reflected from BS-33% by waving POP camera around.  Steer BS-33 until beam goes back in the direction that the camera used to be mounted.  Adjust camera mount and BS-33 so that beam is on camera.

7. Put BS-50% back in place.  Steer it around with voltmeter on PD to see if beam ever hits PD.  Unsuccessful. Give up, since we have POP55, and POP camera.

8. Make a youtube video: POP, AS, REFL, ITMXF (all on Quad3) - PRMI coarsely aligned, no IFO parts locked.  MICH was locked earlier, but not during video time.

  7755   Tue Nov 27 17:45:42 2012 SteveUpdateGeneralspare optics of AP table moved to cabinet

Spare optics from the AP table were moved to glass cabinet in the east arm. I'm not sure this is the right place. We'll see what everybody thinks.

There were two UNMARKED optics. Shame on you! No pencil marks on the optics either. These optics were shipped to the FBI for finger tip analysis.

 

 

Attachment 1: IMG_1819.JPG
IMG_1819.JPG
  7765   Fri Nov 30 09:59:53 2012 SteveHowToGeneralHow not to

Clean cabinet S15 doors were left open. You have to lock it up!

  7830   Fri Dec 14 03:49:02 2012 AyakaUpdateGeneralEarthquake

There was an earthquake around 2:30 am. Now all the mirrors except SRM are damped.

earthquake.pngseis.pdf

  7842   Mon Dec 17 20:22:07 2012 JamieUpdateGeneralVent plan: WE VENT TOMORROW

We will vent tomorrow (12/18)

After lengthy discussion, I have determined that we should vent now, with the primary goal of this vent to replace the input steering PZTs with the new active tip-tilts.  Since we still don't understand exactly what's going on with the PRC, it's unclear what we would do to attack the problem.  We need to do more modeling and measurements first.  We should limit the goal of this vent to replacing the PZTs, and then close up and do more measurements with better modeling and improved input point in hand.

There is limited time this week before everyone leaves for two-week holidays on Thursday or Friday.  The reason to not vent would be that we don't want to leave the IFO at air during the two week holiday.  People seem to think that this is not a problem, so we don't gain anything by waiting.  Therefore we vent now and do what we can before people take off.

The goal for this week is to replace PZT1 only:

Now: Jenne and Manasa are doing vent prep.  Manasa is lowering the input power and preparing the mode cleaner.  Jenne centered IPPOS and IPANG.  This will allow us to check how the input alignment changed during the vent.

12/18: Steve is going to start the vent as soon as he gets back from the dentist, at around 10am.  He will regulate the vent such that we are ready to lock the mode cleaner by 4pm.  At that time we will lock the MC and recheck the input alignment with IPPOS/ANG, with the idea being to see if anything moves during the vent.

12/19: First thing in the morning we take off the access connector ONLY.  The access connector is all we'll need to replace PZT1.    Put light door on the OMC chamber side immediately, since we won't need to get in there at all.  We won't need the light access connector.

For the rest of Wednesday we'll remove PZT1 and install the new active TT.  We'll be using the current out-of-vac cable for PZT1 for the new TT.  We should only have to modify the rack end of the cable to accommodate the coil driver.  This should be a small modification.  Given that we have no wiring diagrams we'll have to pin it out in situ.

12/20: Hopefully finish up TT installation.

Jenne leaves 12/20, Manasa and Jamie leave 12/21.  We will either leave light doors on access connector holes, or possibly Rana, Koji, and Steve will replace access connector on Friday so that we can pump down to 1 Torr or something so that we leave it there over the holiday.

After we return from vacation:

PZT2/TT2 installation.  This will be less straight forward since the new TT has a bigger foot print than PZT2 and will block the PRM optical levers.  We'll need one additional steering mirror to redirect the oplev around the TT.  See elog 7815.

Once the new TTs are installed, we'll reevaluate where we're at.  If PRC modeling has progressed and we have an idea of something to work on with the PRC, we can.  Otherwise, we'll just button up, pump down, and get on with some better PRC measurements.

  7843   Mon Dec 17 21:33:45 2012 ManasaUpdateGeneralVent plan: WE VENT TOMORROW

Quote:

Manasa is lowering the input power and preparing the mode cleaner.

I aligned the PSL angle and position QPDs...then attenuated the input laser power using HWP-PBS-HWP combo from 1.28W to 100mW following instructions from elog 6892 and elog 7299.

To enable MC locking, I replaced the 10% BS before the MC_REFL PD with Y1 mirror. I tweaked the steering mirrors at the PSL table a teeny tiny bit to enable MC locking. MC is now locked at low power with 1.0 transmission.

  7845   Mon Dec 17 22:42:27 2012 KojiSummaryGeneralThe projector lamp ended its life?

i just heard a rather large exploding sound in the control room.
I tried to locate the source and found the projector is not illuminating the wall anymore.
There is a slight smell of burning, but nothing is smoking.

Probably the lamp ended its life.

Rana and I just talked about the projector life time an hour ago! It must have been hearing!

  7846   Mon Dec 17 22:49:33 2012 ManasaSummaryGeneralThe projector lamp ended its life?

Quote:

i just heard a rather large exploding sound in the control room.
I tried to locate the source and found the projector is not illuminating the wall anymore.
There is a slight smell of burning, but nothing is smoking.

Probably the lamp ended its life.

Rana and I just talked about the projector life time an hour ago! It must have been hearing!

 LOL

We should try purchase a projector with LED this time...longer lifetime! I guess the price of replacing the lamp in the one we have will be more or less same as a new one!

  7847   Tue Dec 18 00:45:19 2012 KojiSummaryGeneralThe projector lamp ended its life?

...Nah. The projector is pretty new (t<1yr) and this is the first time to have the lamp busted after the installation last year in Jan.

We just should purchase two bulbs.

  7848   Tue Dec 18 07:03:40 2012 ManasaUpdateGeneralVent plan: WE VENT TOMORROW

Quote:

 

12/19: First thing in the morning we take off the access connector ONLY.  The access connector is all we'll need to replace PZT1.    Put light door on the OMC chamber side immediately, since we won't need to get in there at all.  We won't need the light access connector.

 The access connector is all we'll need to replace PZT1.

Really? Can you be sure of the input pointing this way?

 

 

 

 

 

 

  7860   Wed Dec 19 21:35:33 2012 KojiSummaryGeneralFirst Contact Training with Margot

[Koji, Steve]

First Contact Training with Margot

  7863   Thu Dec 20 12:14:19 2012 JamieUpdateGeneralproblem with in-vac wiring for TTs

Nic and I discovered a problem with the in-vac wiring from the feed-thru to the top of the table.  Pin 13 at the top of the stack, which is one of the coil pins on the tip-tilt quadrapus cables, is *the* shield braid on the cable that goes to the feed-thru.  This effectively shorts one of our coil signals.

There are three solutions as we see it:

* swap pin 13 for something else at the top of the stack, and then swap it back somewhere else outside of the vacuum.

* swap *all* the pins at the top of the table to be the mirror.  We would then need to mirror our cables on the outside, but that's less of an issue.

* make a mirror adapter that sits at the top.  This would obviously need to be cleaned/baked.

None of these solutions is particularly good or fast.

  7864   Thu Dec 20 17:13:56 2012 JamieUpdateGeneralhow to deal with problem with in-vac wiring for TTs

So this is obviously a general problem for all the TTs.  Our in-vacuum wiring is unfortunately mirrored relative to that of aLIGO, or at least:

  • aLIGO: in-vacuum pin 1 tied to shield (T1200131)
  • 40m: in vacuum pin 13 tied to shield

And again, the problem is that pin 13 on the TT quadrapus cables is the one of the coil pins for one of the OSEMs.

I think the right solution is to make mirroring adapter cables for the TTs.  Modifying the pins on the stack-top brackets for just the TTs would leave us with a bunch of brackets that are different than all the rest, which I think is a bad idea.  Therefore we leave all the feed-thru-->bracket wiring the same, and make adapters.  I'll describe the adapters in a follow-up post.

The silver lining to this whole thing, if there is one, is that I wired the polarity on the out-of-vac adapter cable at the coil driver in such a way that the drive/send signal went to the grounded in-vac pin.  If I had by chance wired the polarity oppositely everything would probably have worked, except that the return for one of the coils would have gone through the cable shield and the chamber, rather than the return pin to the coil driver.  I'll let you image the problems that would have caused.

Making new adapters will take a little while, but I think we can proceed with the installation and alignment with a temporary setup in the mean time by taking advantage of the polarity I mention above.  We can temporarily swap the polarity so that we can drive current to the coil using pin 13.  This will allow us to complete the installation and do all the alignment.  Once the in-vac adapter cable arrives, we just put it in, fix the out-of-vac polarity, check that everything works as expected, and button up.

We'll pick up all this when we're back on Jan 7.  Steve will put in an order for the in-vac adapter cables ASAP.

I take full responsibility for this fuck up.  We've been unable to find any in-vac wiring diagrams, but I should have checked all of the wiring during the last vent so that we could have prepared for this ahead of time.  Sorry.

  7865   Thu Dec 20 18:33:52 2012 JamieUpdateGeneralin-vac adapter cables for TTs

We need short cables that mirror the pins:

invac-adapter-cable.pdf

The male side will plug into the 25-pin female on the stack-top bracket.  The tip-tilt quadrapus cable will plug into the female side. This will match up pin 1 on the tip-tilt cable, which is connected to it's shield, to pin 13 on the bracket, which is the shield of the cable that runs to the stack.

They need to be vacuum compatible.  Shorter length is preferred, and there is no minimum length (something like an all-in-one gender changer would be ideal, but probably expensive to have made).


 

  7868   Fri Dec 21 09:55:23 2012 SteveUpdateGeneralin-vac adapter cables for TTs

 

 Accu-Glass is closed till Jan 2, 2013

Atm2 would be the ideal solution Gender Adaptor Special with male - female sides and vented D-sub. How many are we getting 2 or 6 ?

Attachment 1: extensioncable.pdf
extensioncable.pdf
Attachment 2: genderchanger.pdf
genderchanger.pdf
  7885   Wed Jan 9 13:34:34 2013 KojiSummaryGeneralThe projector lamp ended its life?

[Koji, Manasa]

- A new projector lamp installed.

- The old lamp lasted 8751 "equivalent lamp hours".

- The old lamp was found being shattered inside. It contains mercury.
So next time you hear the explosion sound of the lamp, establish the ventilation of the room and escape for an hour.

  7891   Fri Jan 11 11:07:04 2013 ManasaUpdateGeneralIFO status update - PMC problems

I came in this morning to see that the PMC was down. The PZT voltage had drifted to below 50V. I adjusted the FSS slow controls to 0V and PZT was back at 126V.

PMC and IMC could eventually be locked.

History of PZT voltage behaviour in dataviewer over the last 24 hours shows it has been drifting everytime after it has been fixed.

  7892   Fri Jan 11 16:13:47 2013 JenneUpdateGeneralIFO status update - PMC fixed

Quote:

I came in this morning to see that the PMC was down. The PZT voltage had drifted to below 50V. I adjusted the FSS slow controls to 0V and PZT was back at 126V.

PMC and IMC could eventually be locked.

History of PZT voltage behaviour in dataviewer over the last 24 hours shows it has been drifting everytime after it has been fixed.

 FSS was saturating.  Fixed.

  7930   Wed Jan 23 18:16:11 2013 ManasaUpdateGeneralLaseroptik mirror - SN6

 I repeated the transmittance measurements of LaserOptik SN6 @1064nm.

Transmittance for s-polarization 

0 deg - 0.524
45 deg - 0.055

Transmittance for p-polarization

0 deg - 0.515
45 deg - 0.1047 0.01047

Raji's measurements are here.

Attachment 1: sn6_trans.png
sn6_trans.png
  7932   Wed Jan 23 20:24:05 2013 KojiUpdateGeneralLaseroptik mirror - SN6

Got confused (even after I talked with Manasa).

The plot shows the number ~0.01 or less at 45deg. But the number is the text does not match with the plot.

Please use the logarithmic scale for the vertical axis.
And more points between 35 to 50 deg please (like ~1deg spacing)

Don't we have the data sheet from the coater? Can we request it?

  7935   Wed Jan 23 22:02:25 2013 ManasaUpdateGeneralLaseroptik mirror - SN6

Quote:

Got confused (even after I talked with Manasa).

The plot shows the number ~0.01 or less at 45deg. But the number is the text does not match with the plot.

Please use the logarithmic scale for the vertical axis.
And more points between 35 to 50 deg please (like ~1deg spacing)

Don't we have the data sheet from the coater? Can we request it?

I corrected the typo in the text...however, I agree the plot was lame...Will get the data sheet made tomorrow! 

  7940   Thu Jan 24 15:16:50 2013 ManasaUpdateGeneralLaseroptik mirror - SN6

I repeated the transmittance measurements of Laseroptik SN6 at 1064nm. The rotation stage could only resolve 2 deg rotation (We should consider buying a better rotation stage).

s-polarization

Percentage transmittance

       0.177% 42 deg
       0.806% 44 deg
       0.57%   46 deg
       54.8%    0 deg

p-polarization

Percentage transmittance

       1.039% 42deg
       1.155% 44 deg
       1.159% 46 deg
        65.6%   0 deg

sn6_trans1.png 

  7941   Thu Jan 24 16:23:24 2013 KojiUpdateGeneralLaseroptik mirror - SN6

The mirror T is completely out of spec. We should find or request the data sheet of the mirror.

> We should consider buying a better rotation stage

I'm already on it

  7943   Thu Jan 24 16:34:56 2013 ranaUpdateGeneralLaseroptik mirror - SN6

  I have two questions:

1) Are we sure that the T measurement is not being compromised by some systematic? i.e. some leakage is making the apparent T appear too high.

2) IF the T is really so high, how should we decide whether or not to use this one rather than the G&H? Is the 532 nm property more important than the high recycling gain?

  7944   Fri Jan 25 08:10:42 2013 SteveUpdateGeneralLaseroptik mirrors

 Here are the German plots. Unfortunately they are not logarithmic.

 Ed: Proprietary data removed. Use wiki (Koji)

Steve uploaded data to the 40m wiki  / Aux_Optics on 02-07-2013

 

 

ELOG V3.1.3-