ID |
Date |
Author |
Type |
Category |
Subject |
6531
|
Thu Apr 12 23:12:16 2012 |
Suresh | Update | IOO | Beam Profile measurement: IPPOS beam: Possible Clipping |
WiQuote: |
[Suresh, Jenne]
The input beam is most probably being clipped at the Faraday Isolator.
Evidence:
.....
We plan to investigate this further to be sure..
..... |
I tried to determine an optimal WFS2YAW offset to be used so that we may avoid clipping.
Initially, I just measured the beam diameter as a function of offset. If the beam diameter would become independent of offset if it is not clipped. However a systematic effect became apparent when I shifted the beam on the detector to a slightly different location. So I repeated the measurements while recentering the beam to the same location everytime (centered at -1650+/- 50 for both H and V directions).
I have attached plots of the scans for both cases, with recentering and without. I have not been able to figure out what is going on since the beam diameter does not become independent of the offset. While the beam profile becomes more gaussian beyond offsets of about 7 or so, the beam diameter does not seem to follow a clear pattern. The measurements are repeatable (within one sigma) so the experimental errors are smaller than 1 sigma.
The photographs below show the improvement of Horizontal beam profile with WFS2Yaw offset. These seem to indicate a good gaussian beam for offsets beyond 7 or so. At offsets more than 12 the MC unlocks.
 |
 |
 |
 |
Offset = -2 |
Offset = 0 |
Offset = 2 |
Offset = 8 |
 |
 |
This seems to indicate that the beam diameter does not vary for WFS2Yaw offset > 8 |
But if we recenter the beam for each measurement this effect seems to vanish |
Will continue tomorrow. Jenne wants to do some IFO locking now.
|
6530
|
Thu Apr 12 22:04:17 2012 |
Mike J. | Update | Computers | New Hysteresis Model & Plots |
The new hysteresis model uses a triangle wave with offset zero points as the position function and a sinusoidal force function, creating a loop similar to this. Model is at /users/mjenson/matlab/ferro_hyst.mdl.
 
|
6529
|
Thu Apr 12 20:56:07 2012 |
Den | Update | PEM | daq |
GUR1 XYZ, GUR2 XYZ, MC_F channels are now recorded at 256 Hz.
EDIT by JCD: What Den means to say here is that (a) he modified some .ini files, and (b) he restarted the fb. |
6528
|
Thu Apr 12 14:48:44 2012 |
Suresh | Update | SUS | local damping and WFS |
WFS servo is moving the MC mirror angles to minimise TEM01 and TEM10 modes within the MC cavity. This means it will compensate not only for angular noise in the mirrors but also for the PSL beam pointing fluctuations. So the extra "noise" we see when WFS loops are on is because they are active below the WFS UGF of about 2 Hz. Also if the HEPA airflow is above 20% (of its max), the PSL beam jitter (caused by the airflow) will add broadband noise into the WFS servo loops and this will show up in the OSEM signals. See elog 5943 for details.
Quote |
......
Then I compared MC1, MC2, MC3 suspos, suspit, susyaw and susside positions with WFS on (black curve) and off (red curve). WFS add noise to MC1 and MC3 measured by osems (MC2 is fine though). WFS should change osem readings but is it a correct way to do this below 0.5 Hz (?) It looks like just a flat noise. Need to think about the conclusion.
|
|
6527
|
Thu Apr 12 08:49:14 2012 |
Den | Update | SUS | local damping and WFS |
I tried to figure out what can add noise below 0.5 Hz to the MC_F. I compared MC1, MC2, MC3 suspos, suspit, susyaw and susside positions with damping (black curves) and without (red curves). Local damping is fine.

Then I compared MC1, MC2, MC3 suspos, suspit, susyaw and susside positions with WFS on (black curve) and off (red curve). WFS add noise to MC1 and MC3 measured by osems (MC2 is fine though). WFS should change osem readings but is it a correct way to do this below 0.5 Hz (?) It looks like just a flat noise. Need to think about the conclusion.
 |
6526
|
Thu Apr 12 01:17:56 2012 |
Suresh | Update | IOO | Beam Profile measurement: IPPOS beam: Possible Clipping |
[Suresh, Jenne]
The input beam is most probably being clipped at the Faraday Isolator.
Evidence:
a) The beam scan of the IPPOS beam showed a nongaussian beam in the horizontal direction. This was visible in the beam scan since it overlays a gaussian-fit over the data.
b) I was able to remove this departure from gaussian profile by introducing an offset of 5 into the C1:IOO-WFS2_YAW_OFFSET.
c) We made a few measurements of the beam diameter as a function of distance at an offset of 7. At a distance of beyond 3 m the deviation from gaussian profile was once again apparent.
d) We increased the offset to 14 to remove this deviation.
e) When we measured the beam diameter again with this new offset the horizontal diameter and vertical diameters dropped by 2.sigma. Indicating there the beam was clipped till then.
f) We increased the offset to 16 and the beam diameter did not change further (within 1.sigma). Implying no more clipping, hopefully.
And then the earthquake stopped us from proceeding further.
We plan to investigate this further to be sure.. Data attached.
Subsidiary effects to keep track of:
1) Introducing an offset into the WFS loops decreases the coupling from PSL into MC.
2) If the beam is being clipped at the Faraday Isolator then the REFL beam would also show lesser clipping with WFS offsets. |
Attachment 1: BeamProfileData_IPPOS_2.xlsx
|
6525
|
Thu Apr 12 00:40:45 2012 |
Jenne | Update | SUS | PRM, BS oplevs off |
There's a beam dump after the HeNe on the BS oplev table, since the IPPOS measurement optics (steering mirrors) are in the way of the oplev beams.
Don't enable the BS or PRM oplevs!!!! We'll post a notice in the elog when the oplevs are back to normal.
Self, remember to disable the oplevs manually if they come on with any restore scripts. |
6524
|
Thu Apr 12 00:16:38 2012 |
Jenne | Update | Environment | Earthquake - moderate |
M4.7 - Santa Isabel, Mexico 2012-04-12 06:48:38 UTC
Mode Cleaner doesn't want to stay locked. Seismic is coming down from an earthquake ~20min ago.
We're in the process of measuring IPPOS, so this is obnoxious.
EDIT: Followed by a 6.2 and a 7.1 at 07:06UTC and 07:15UTC in the same area.
We're following the tried and true tradition of going home when there's an earthquake big enough that the MC won't stay locked.
Several optics have rung up, PRM is the only one which has tripped so far, because the side sensor has the extra gain, but the watchdog threshold is set for the face OSEMs. |
6523
|
Wed Apr 11 22:48:39 2012 |
rana | Update | Environment | BLRMS |

|
6522
|
Wed Apr 11 18:19:51 2012 |
Jenne | Update | IOO | Mode matching recollections and conclusions |
Another histogram. This one allows the MMT mirror positions to move, the MMT incident angle for both curved optics to change, and the MC waist size and position to change. The error quoted for the MC waist size measurement from 2010
was +\- 0.01mm, and the MC waist position was +\- 28mm.
This histogram is showing that we're pretty sensitive to the MC waist measurement, which is used to define the beam. We can be up to ~2% off in our ideal mode matching to the arms if we're using the incorrect initial beam for the telescope design.

|
6521
|
Wed Apr 11 17:56:26 2012 |
Jenne | Update | General | Summary of things to figure out with the IFO |
Questions:
Power recycling gain
* It should be ~40, but we observe/measure it to be ~7. Even if mode matching of ~50% is assumed, gain is calculated to be ~15
* Would like to measure PR gain independent of mode matching, if possible
Power recycling cavity mode matching
* Reflectivity of PRMI was measured to be ~50%. That's pretty high. What's going on?
* Even if we're mode matched to the arm, are we appropriately mode matched to the PRC?
Is beam from MC clipped in the Faraday?
* We had to use MC axis for input pointing since PZTs aren't totally working.
* Need to measure IPPOS beam for different MC alignments to see if horizontal waist measurement stays constant.
PRM flipped?
* Not likely, but it can't hurt to confirm for sure.
* Want to know, since it could give us a different plan for MMT moving than if the PRM is correct.
Thick optic non-normal incidence in IPPO - does this exaggerate astigmatism, which would help explain IPPOS measurement?
Is PRC waist same size / position as arm cavity waist, given the current "known" positions of all the optics?
* How is this effected by moving the PRM?
------------------------------------------------------------------------------------
Measurements to take and what information they give us:
IPPOS beam scan, with MC as-is
* Confirm (or not) IPPOS measurements from last week
IPPOS beam scan with different MC alignments
* Will tell us about Faraday clipping, if any
AS beam scan, misaligned PRM, misaligned SRM, misaligned ITMX, single bounce from ITMY
* Can only take this measurement if beam is bright enough, so we'll just have to try
* Will confirm IPPOS measurement, but includes going through the thick PRM, so can compare to calculated intra-PRC mode
REFL beam scan (already done....is the data satisfactory? If so, no need to redo), single bounce off of PRM
* Will tell us about the potential PRM flipping
* Need to compare with calculated mode at REFL port for flipped or non-flipped PRM
Look at POP camera, see 2nd pass through cavity
* Try to match 1st and 2nd pass. If they don't match, we're not well matched to PRC mode
Look at beam directly on ETMY cage, then beam from ETM, bounce off ITM, back to ETM cage
* If the beams are the same size, we're well matched to arm cavity mode
* Use fancy new frame-grabber.
----------------------------------------------------------------------------
MMT code things to calculate, and what information it gives us:
REFL beam path, for PRM flipping comparison
Thick IPPO non-normal incidence - I'm not sure how to do this yet, since I only know how non-normal incidence changes effective radii of curvature, and this is a flat optic, so *cos(theta) or /cos(theta) won't do anything to an infinite RoC
Compare PRC waist to arm cavity waist, using "known" optic positions
Mode matching sensitivity to MC waist measurements
Mode matching sensitivity to PRM position |
6520
|
Wed Apr 11 16:33:16 2012 |
Jenne | Bureaucracy | General | 40m Meeting Action Items |
Action Items from Last Week:
(non-existant)
-----------------------------------------------------------------------
Action Items this Week and LEAD PERSON:
Assemble and ship 4 TTs from LHO - SURESH
Prepare electronics for TTs (coil drivers) - JAMIE
In-air TT testing to confirm we can control / move TTs before we vent (starting in ~2 weeks) - SURESH
Connect TTs to digital system and controls, lay cables if needed - JAMIE with SURESH
OAF comparison plot, both online and offline, comparing static, adaptive and static+adaptive - DEN
Static-only OAF noise budget (Adaptive noise budget as next step) - DEN
Black glass: big baffle pieces to clean&bake, get small pieces from Bob, put into baskets, make new basket for 1" pieces, get to clean&bake - KOJI
IPPOS beam measurement - SURESH with JENNE
AS beam measurement (if beam is bright enough) - SURESH and JENNE
Mode matching calculations, sensitivity to MC waist measurement errors, PRM position - JENNE
Summary of IFO questions, measurements to take, and game plan - JENNE
Think up diagnostic measurement to determine mode matching to PRC while chambers are open, while we tweak MMT - JAMIE, JENNE, KOJI, SURESH |
6519
|
Wed Apr 11 14:34:48 2012 |
Jenne | Update | Environment | Plumbing guys are here |
Quote: |
I don't know why, but they're looking around on the roof, and inside our ceiling above the bathrooms.
|
Bob tells me that the carpenter is going to move the nitrogen bottles to the other side of the outside door, so that the plumbers can install a safety shower / eyewash right outside our door.
Also, the carpenter just mounted a new glass door cabinet from Bob's lab in the IFO room, so we have some new storage space. |
6518
|
Wed Apr 11 12:25:11 2012 |
Ryan | Update | Computers | Updating aLIGO Conlog |
Over the next few days, I will be working on upgrading the aLIGO Conlog install to include new bugfixes distributed by Patrick T. The currently running conlog *should* not be affected, but please let me know if it is (ryan.fisher@ligo.org). |
6517
|
Tue Apr 10 23:56:44 2012 |
rana | Update | Computers | Sensoray |
Now that Mike has got the Sensoray working, Jenne/Suresh should grab some new images of the ETM cage as Keiko did so that we can analyze them for another mode matching diagnostic. |
6516
|
Tue Apr 10 17:02:29 2012 |
Jenne | Update | IOO | Mode matching recollections and conclusions |
...Mostly just recollections at this point.
I re-looked at the mode matching's sensitivity to misplaced optics. Here is the plot that the original MMT code from 2010 spits out:

What this plot is telling us is that we should lose no more than 0.1% of mode matching "goodness" if we messed up the curved optic's positions by up to 2 cm. If we can't place optics to within 2 cm, we might as well go back to optics kindergarten, because that's pretty lame.
UPDATE: Here is a histogram using the new code, which definitely includes the non-unity index of refraction for the transmissive optics and the Faraday. The only optics which are permitted to move are the 2 curved optics, and they are allowed a stdev of 20mm. Again, we shouldn't be doing worse than ~99% mode matching, even if we're 2cm off from the MMT positions that we measured with a ruler. This histogram only has 300 iterations, since it takes quite a while (~0.5sec) to calculate each iteration. Note this is mode overlap using the measured MC waist, propagated through optics, compared to the ideal arm mode. This is completely ignoring the IPPOS measurements so far.

UPDATE 2: Allowed 5 degrees of incident angle motion for both curved optics, which changes the astigmatism of the beam downstream. Still, no big change from ~99% mode matching efficiency. Again, this doesn't include any information from the IPPOS measurements. 3000 iterations this time around, since I didn't need my computer. Curved optics still allowed to move back and forth by 2cm.
More meditations and conclusions to follow... currently running hist code to allow tilt of optics, to account for astigmatism changes also.
Suresh and I are going to do some beam measurements tomorrow with the beamscanner, and then we will do a few measurements with the razor blade technique, to confirm that we're doing things okey dokey. |
6515
|
Tue Apr 10 13:42:54 2012 |
Jenne | Update | Environment | Plumbing guys are here |
I don't know why, but they're looking around on the roof, and inside our ceiling above the bathrooms. |
6514
|
Tue Apr 10 11:08:29 2012 |
tara | Update | PSL | curved mirror behind AOM removed |
We removed the curved mirror behind the AOM (ROC=0.3m) on PSL table. The mirror is now in PSL lab. See PSL:905 for more detail. |
6513
|
Mon Apr 9 20:02:19 2012 |
Mike J. | Update | Computers | Sensoray |
The highest resolution available is 720x480 pixels. Bit depth of captured images and video is most likely 16 bits per pixel. Video may be captured raw as well, which will be necessary for image subtraction/enhancement, however it cannot currently be played raw. A captured image is shown below, along with MP4 video.
|
6512
|
Mon Apr 9 18:18:14 2012 |
kiwamu | Update | LSC | DRMI time series |
Here is a time series when the DRMI is being locked.
You can see that the AS110 goes up because the SRCL is engaged and amplifies the 55 MHz sidebands.
 |
6511
|
Mon Apr 9 17:03:38 2012 |
Den | Update | Environment | Lms vs Wiener |
I tried to figure out why offline LMS filter subtract seismic noise much better from MC_F then the Wiener filter. I did the calculations twice - with my codes and with Matlab in-build functions, the results are the same. So this is not a code error.
The coherence between GUR 1, 2 and MC_F is still poor. Wiener filter is linear and its performance is confined to the frequency ranges where we see coherence. Lms filter is non-linear and it may be possible to subtract the noise even if non-linear effects are present in the system.

I've checked seismometer readout box again. I've soldered 50 Ohms to plus and minus inputs to VERT 1,2 N/S 1,2, E/W 1,2 - GUR 1 and 2 use these channels. Then I put the box back and connected it to the ADC.

The plot shows that the readout box noise is below the ADC noise. It is possible that amplifiers introduce non-linear effects. To check this I plotted the coherence between OSEM sensors and GUR1X signal:

The coherence between OSEM sensors and GUR1X is pretty good, so may be witness path is not responsible for low coherence at 0.1 - 0.5 Hz between MC_F and GUR 1,2. IT seems that MC_F is bad at low frequencies. I terminated the input to the Channel 1 of the Pentek Generic board, where MC_F is plugged in.

ADC is also good. Something else is wrong. |
6510
|
Mon Apr 9 15:09:34 2012 |
Jenne | Update | LSC | Locked MICH |
Quote: |
I was going to try some locking, but things are a little too noisy.
Just so Kiwamu knows what I did today, in case he comes back....
I ran LSCoffsets, and aligned both X and Y arms and saved their positions, and aligned MICH, and saved the BS position.
I'll play with it more later, when there aren't trucks driving around outside that I can hear / feel in the control room.
|
After giving up on locking, the MC is getting unlocked every now and again (2 times so far in the last few minutes) from transient seismic stuff. |
6509
|
Mon Apr 9 15:02:30 2012 |
Jenne | Update | LSC | Locked MICH |
I was going to try some locking, but things are a little too noisy.
Just so Kiwamu knows what I did today, in case he comes back....
I ran LSCoffsets, and aligned both X and Y arms and saved their positions, and aligned MICH, and saved the BS position.
I'll play with it more later, when there aren't trucks driving around outside that I can hear / feel in the control room. |
6508
|
Sat Apr 7 06:58:34 2012 |
kiwamu | Update | LSC | DRMI lock : lost good alignment |
Somehow I lost the good alignment, where the lock can be frequently acquired and hence I didn't go further ahead.
I will try locking the DRMI during the weekend again. My goal is to take time series when the DRMI is being locked and sensing matrix.
Quote from #6506 |
Currently SRM is slightly misaligned such that the MICH optical gain at AS55Q doesn't increase so much with the presence of SRM.
With this condition I was able to acquire the lock more frequently than how it used to be on the Wednesday.
The next step is to gradually align SRM, to optimize the controls and to repeat this process several times until SRM is fully aligned.
|
|
6507
|
Sat Apr 7 02:01:29 2012 |
Mike J. | Update | Computers | Projector Cable Management |
I replaced the projector video and power cables with longer ones, and zip-tied them to the ceiling and wall so they don't block the image.

|
6506
|
Sat Apr 7 01:56:05 2012 |
kiwamu | Update | LSC | OSA signal in DRMI condition |
It wasn't a dream or illusion -- I was locking the DRMI to the right condition last Wednesday (#6489).
Here is a snap shot of the AS-OSA signal taken today when the DRMI was locked with the same control settings (#6489).
The blue curve is data taken when the PRMI was locked for comparison.
You can see that both the upper and lower 55 MHz sideband are amplified by the SRC.

(Some notes)
Currently SRM is slightly misaligned such that the MICH optical gain at AS55Q doesn't increase so much with the presence of SRM.
With this condition I was able to acquire the lock more frequently than how it used to be on the Wednesday.
The next step is to gradually align SRM, to optimize the controls and to repeat this process several times until SRM is fully aligned.
Quote from #6489 |
A first thing I tried was : flipping the control sign of the SRCL while keeping the same control setups for the PRCL and MICH.
Occasionally the DRMI was "sort of" locked and hence I believe this setup must be a good starting point.
|
|
6505
|
Sat Apr 7 01:45:02 2012 |
Mike J. | Update | Computers | Even Better Hysteresis Model and Plots |
The new hysteresis model is slightly based on the SHO equation, but with the force being out of phase with the position by an amount of hysteresis {x(t)=Amp*sin(freq*t), F(t)=Amp*sin(freq*t+Hyst)}. The new model can be found at /users/mjenson/matlab/hyst_v_3.mdl. Pictures are: new hysteresis model, x(t) subsystem in new model[xh''(t) only lacks -1 multiplier and includes hysteresis variable], new plots.
 .png?lb=40m&thumb=1)  |
6504
|
Sat Apr 7 00:31:12 2012 |
keiko | Update | LSC | RAM simulation for Full ifo |
I didn't understand how CARM can be decreased 2 orderes of magnitude and PRCL can be INCREASED by such small offsets (see the matrix quoted).
Apparently it was because of an optical-spring ish effect from the "detuning" (which is actually RAM position offsets). I put two plots which are CARM and PRCL tranfer functions to REFL f1 or POP f1, when there is a slight PRCL offset (0, 1e-14m, and 1e-15m cases are plotted). Looking at these plots, it was not a good idea to calculate the LSC matrix in DC because they are affected by this detuning a lot. I'll try f = 150 Hz for the matrix.
 
Quote: |
*** Iteration ***
%*.*.*. Resulting matrix w/ RAM .*.*.*
REFL f1 : 0.039125 -0.000000 0.003665 0.000005 -0.000007
AS f2 : 0.000010 1.000431 0.000009 -0.003500 -0.000002
POP f1 : 156.420221 -0.000246 15.586838 0.019406 -0.000154
POP f2 : 1.255806 -0.154275 0.047313 1.000008 0.024285
POP f2 : -34.814720 -0.006600 -1.884850 0.042950 1.000000
Offsets converged to:
PRCL = 2.1e-15, MICH = 1.1e-17, SRCL = -3.8e-15, CARM = 2.2e-16, DARM = 0
|
|
6503
|
Fri Apr 6 20:38:41 2012 |
Mike J. | Update | Computers | Sensoray |
Turns out that the "MPEG-4 VES" video format is just bad for captured video. Everything except "MP4" and "MPEG-TS" works for streaming, and "MP4" and "MPEG-TS" seem to be the only captured formats that can be viewed properly. |
6502
|
Fri Apr 6 20:24:31 2012 |
Mike J. | Update | Computers | Sensoray |
The Sensoray device is currently viewing Monitor 4 and plugged into Pianosa. The user interface is run at /home/controls/Downloads/sdk_2253_1.2.2_linux/python demo.py. It can preview and capture the video stream, however the captured files are terrible. I believe it has something to do with the bitrate, since the captured video with lower bitrates are not as bad as the ones with higher bitrates, but I am not certain. |
6501
|
Fri Apr 6 20:05:12 2012 |
Jamie | Summary | General | Laser Emergency Shutoff |
We reset the interlock and restarted the PSL. The end AUX lasers seem to have come back online fine. PMC and mode cleaner locked back up quickly. |
6500
|
Fri Apr 6 19:40:57 2012 |
Mike J. | Summary | General | Laser Emergency Shutoff |
I accidently shut off the laser at 19:34 with the emergency shutoff button while trying to tap into a video line for the Sensoray device. |
6499
|
Fri Apr 6 19:04:35 2012 |
Jenne | Update | PEM | STS releveled, GUR2 plugged in |
[Den, Jenne]
We were wondering why the STS-2 signal was funny. When I went to look at it, the X-axis indicator was pointing ~45deg from the x-axis, so that it was pointing between the arms of the IFO. Also, the bubble in the level was totally stuck on one side. We locked the masses, and I put the seismometer back to the correct orientation, and then leveled it. We unlocked the masses and turned the power back on, and hit the auto-zero button a few times. Right now the X-axis signal is fine, but Y and Z are still railed, but it's been like 24 seconds, not 24 hours since we last hit auto zero, so there's still some time to wait.
Also, GUR2 was unplugged on both ends of the cable. We plugged it back in. However, it looks like the *seismometer* labeled #1 is now plugged into *channels* GUR2, and the seismometer labeled #2 is plugged into channels GUR1. Recall that Den has only modified X, Y, Z for GUR1 channels, not any other channels in the breakout box. |
6498
|
Fri Apr 6 16:35:37 2012 |
Den | Update | Computers | c1ioo |
c1ioo computer can not connect to the framebuilder and everything is red in the status for this machine, C1:FEC-33_CPU_METER is not moving.
EDIT by KI:
We rebooted the c1ioo machine, but none of the ftont end model came back. It looked like they failed the burt process for some reasons according to dmesg.
Then we restarted each front end model one by one, and every time after immediately we restarted it we hit the 'BURT' button in the GDS screen.
Everyone came back to the normal operation. |
6497
|
Fri Apr 6 16:22:15 2012 |
Den | Update | Environment | seism box |
I've changed R2 resistor in the seism box for the VERT 1 channel from 464 Ohm to 1051 Ohm to reduce the gain of this channel by a factor of 2. This should help the GUR1Z signal not to be corrupted inside the AA box, so we can use it in the adaptive filtering. |
6496
|
Fri Apr 6 15:06:05 2012 |
Den | Update | IOO | 1 Hz resonance |
I think we can try to damp 1 Hz resonance more. In September it was not seen because of the digital noise. After we've figured it out, 1 Hz resonance began to be more clear (blue line).

Now applying oaf we reduce the effect of the stack and the 1 Hz resonance is even more clear:

|
6495
|
Fri Apr 6 14:39:21 2012 |
Jamie | Update | Computers | RAID array is rebuilding.... |
The RAID (JetStor SATA 416S) is indeed resyncing itself after a disk failure. There is a hot spare, so it's stable for the moment. But we need a replacement disk:
RAID disks: 1000.2GB Hitachi HDT721010SLA360
Do we have spares? If not we should probably buy some, if we can. We want to try to keep a stock of the same model number.
Other notes:
The RAID has a web interface, but it was for some reason not connected. I connected it to the martian network at 192.168.113.119.
Viewing the RAID event log on the web interface silences the alarm.
I retrieved the manual from Alex, and placed it in the COMPUTER MANUALS drawer in the filing cabinet. |
6494
|
Fri Apr 6 11:32:09 2012 |
Jenne | Update | Computers | RAID array is rebuilding.... |
Suresh reported to Den, who reported to me (although no elogs were made.....) that something was funny with the FB. I went to look at it, and it's actually the RAID array rebuilding itself. I have called in our guru, Jamie, to have a look-see. |
6493
|
Fri Apr 6 11:14:34 2012 |
Jenne | Update | Adaptive Filtering | static and adaptive |
Quote: |
I've run static and adaptive filters simultaneously. AA32 filters rotate the phase of the witness signals GUR1X and GUR1Y and now the performance of the static filter is worse. Next time I'll recalculate Wiener filter coefficients taking this into account. But still 2 filters together can deal with a stack better.

|
This is super awesome! I'm totally excited!! |
6492
|
Fri Apr 6 10:31:07 2012 |
Den | Update | Adaptive Filtering | static and adaptive |
I've run static and adaptive filters simultaneously. AA32 filters rotate the phase of the witness signals GUR1X and GUR1Y and now the performance of the static filter is worse. Next time I'll recalculate Wiener filter coefficients taking this into account. But still 2 filters together can deal with a stack better.

|
6491
|
Fri Apr 6 09:57:24 2012 |
Den | Update | Adaptive Filtering | static starts to work |
I made static filter to work to evaluate the actuator TF.. Here is the result of static filtering:

What I did:
I did offline simulation of the MC_F Wiener filtering using 2 witness signals - GUR1X and GUR1Y. I've downsampled the data from 2048 to 128 Hz and applied the Wiener filter with 10000 for each witness channel:
 
Result of the filtering Filter coefficients for gur1x and then gur1y
 
Gur1x -> MC_F transfer function Gur1y -> MC_F transfer function
Then using vectfit I approximated obtained transfer functions in the region 0.5 - 20 Hz. I used a window function and then weights to get a more precise result in this range using only 8 poles and zeros.
 
I obtained the zpk-model for each witness channel and entered it into the FOTON splitting it into 2 parts before that because FOTON does not like too long filters. These zpk-models are at the C1:OAF-STATIC_STATMTX_8_8 and C1:OAF-STATIC_STATMTX_8_9 filter banks.
GUR1X:
z =
7.527339430029315 +31.603999069997801i
7.527339430029230 -31.603999069997823i
27.897703898191267 + 0.000000000000071i
-6.437806394766186 + 9.893955654289517i
-6.437806394766159 - 9.893955654289510i
1.114401249545640 + 5.479278396987240i
0.176877296954015 + 0.000000000000006i
1.114401249545616 - 5.479278396987245i
p =
-0.407251778925379 + 6.263247012022007i
-0.407251778925379 - 6.263247012022007i
-0.230672968859081 + 6.846868757063707i
-0.230672968859081 - 6.846868757063707i
-2.871419857491615 +13.707864827517826i
-2.871419857491615 -13.707864827517826i
-2.134260618362721 +18.319129999777648i
-2.134260618362721 -18.319129999777648i
k =
4.113285626223658e-04
GUR1Y
z =
17.961416874092624 +13.631821252434328i
17.961416874092642 -13.631821252434353i
-8.788634771726304 + 7.653357335975781i
-8.788634771726285 - 7.653357335975777i
-0.037906973323273 + 5.133348020858679i
-0.164348392996182 + 3.588803405511463i
-0.164348392996187 - 3.588803405511474i
-0.037906973323277 - 5.133348020858679i
p =
-0.027577318242359 + 5.174655410828068i
-0.027577318242359 - 5.174655410828068i
-0.500384298611703 + 6.310552036591990i
-0.500384298611703 - 6.310552036591990i
-0.237055716999485 + 6.881204941979009i
-0.237055716999485 - 6.881204941979009i
-1.408223271160550 +14.874570175309771i
-1.408223271160550 -14.874570175309771i
k =
-2.723835471763049e-04
Then I approximated the reversed actuator TF and placed it to the C1:OAF-SUS_MC2_OUT filter bank. The gain to the static filter output is -1.
P.S. Also the static matrix was filled with 1 for some reason. Here is the script to fix it if if will be bad again
for i in {1..8}
do
for j in {1..28}
do
element="C1:OAF-STATIC_STATMTX_"$i"_"$j"_GAIN"
ezcawrite $element 0
done
done
|
6490
|
Thu Apr 5 18:24:55 2012 |
Den | Configuration | Adaptive Filtering | oaf starts to work |
Today I tried to make the lms filter to work online. I played around with the signals (GUR1_X and MC_F) to pre-whiten them and in the end the following configuration worked out:
1. mu = 0.03, tau = 1e-5, downsample=8, nCoeff = 4000, delay = 5 (sample-and-hold delay is not included in the new code, it should be added here!)
2. witness pass: AA32 = cheby1("LowPass", 4, 1, 32) AND 0.1:0
3. witness adaptation path: AA32 AND AI32 = cheby1("LowPass", 4, 1, 32) AND 0.1:0
4. error path: AA32 AND 0.1:0 AND anti_1Hz. Before I added anti_1Hz filter oaf did nothing. This filter tries to approximate the actuator transfer function. Note, it is not in the witness adaptation path. This is some sort of whitening.
5. correction path: AI32, gain = -1
Convergence time ~ 5 mins. The performance of the filter is far not perfect compared to the offline implementation. But it deals with a stack though.

|
6489
|
Thu Apr 5 07:19:16 2012 |
kiwamu | Update | LSC | DRMI locking |
I tried locking the DRMI to the signal-extraction condition with the new trigger by AS110.
A first thing I tried was : flipping the control sign of the SRCL while keeping the same control setups for the PRCL and MICH.
Occasionally the DRMI was "sort of" locked and hence I believe this setup must be a good starting point.
As a next step I will try some different gains and demodulation phase to make it more lockable.
(Time series)

The picture above is time series of some signals when the DRMI was barely locked.
The red arrows indicate the durations when the DRMI was sort of locked.
(Green curve) REFLDC becoming a high value state, which indicates that the carrier is anti-resonant.
(Red curve) ASDC becoming dark, which indicates the MICH is in the vicinity of the dark condition.
(Brown curve) AS110 becoming a high value state, which means the 55 MHz sidebands got amplified by the SRCL.
(Blue curve) POP22 becoming a high value state, which indicates that the 11 MHz sidebands are resonating in the PRC.
According to the measurement of AS110 when PRMI was locked ( #6488), the AS110 signal went up to ~ 1 counts or so.
On the other hand when the DRMI was locked the AS110 went to up more than 10 counts as shown in the plot above.
Therefore at least some kind of signal amplification is happening for the 55 MHz sidebands in the SRC.
Looking at the AS CCD, I found that the beam looked like a TEM01 mode (two beam spots at top and bottom) every time when the DRMI was locked.
(settings)
- REFL33I => PRCL G = -0.2
- AS55Q => MICH G = -6
- AS55I => SRCL G = 1 (G = -50 for the signal recycling condition)
- AS55 demod phase = 17 deg
|
6488
|
Thu Apr 5 06:27:51 2012 |
kiwamu | Update | LSC | AS110 sideband monitor installed |
[Jenne / Kiwamu]
We have installed a broad band PD in the AS path in order to monitor the 110 MHz signal associated with the SRC.
The PD is currently connected to the POP110 demodulation board and it seems working fine.
I know this is confusing but right now the signal appears as "POP110" in the LSC front end model.
- Installed a 50% BS at the AS path
- The AS beam is split to two path - one goes to AS55 and the other goes to the OSA.
- The new BS is installed on the way of the OSA branch therefore AS55 isn't affected by the new BS.
- Installed a PDA10A
- This is a silicon diode with a bandwidth of 150 MHz, and is fast enough to detect the 110 MHz signals.
- The 110 MHz signal seems going up to approximately -40 dBm according to a coarse measurement with an RF spectrum analyzer.
- Also a SMA-style high pass filter, HPF-100, was attached to the output to cut off unnecessary sidebands (e.g. 11, 22 MHz and etc.)
- Put a long BNC cable, which goes from the PD to LSC rack.
- The end of the cable at the LSC rack was directly connected to the POP110 demod board.
- The actual POP110 signal path is currently terminated by a 50 Ohm load and therefore this signal is unavailable.
- Adjustment of the demodulation phase
- The demod phase was adjusted to be 7 deg in the EPICS screen. This phase minimize the Q-signal.
- Locking PRMI with sidebands resonating makes the AS110 signal ~ a few counts and this level is still noticeable.
- Perhaps we may need to put an RF amplifier to get the signal bigger.
|
6487
|
Thu Apr 5 01:07:08 2012 |
Mike J. | Update | Computers | Hysteresis Plots |
Here are the hysteresis plots from the most recent model, which uses a modified harmonic oscillator equation y''=-(Frequency)2*y-Hysteresis. The hysteresis constant seems to change both the amplitude and equilibrium point of the pendulums, which is akin to changing the length of a pendulum without changing the frequency. This does not make sense. Perhaps the hysteresis value should be moved to the "spring" constant for the pendula and not restricted to a position-biasing value.

|
6486
|
Wed Apr 4 23:57:35 2012 |
keiko | Update | LSC | RAM simulation for Full ifo |
I'm still wondering whether iteration version or simple version is closer approximation to the real situation. Sorry for few explanations here. I will try to present those on Friday.
Anyway, here is the results for both:
%*.*.*. Original matrix w/o RAM .*.*.*
REFL f1 : 1.000000 0.000000 -0.000003 -0.000005 0.000007
AS f2 : 0.000002 1.000000 0.000009 -0.003522 -0.000002
POP f1 : -3954.521443 -0.000965 1.000000 0.019081 -0.000152
POP f2 : -32.770726 -0.154433 -0.072594 1.000000 0.024284
POP f2 : 922.393978 -0.006608 1.488319 0.042948 1.000000
*** Iteration ***
%*.*.*. Resulting matrix w/ RAM .*.*.*
REFL f1 : 0.039125 -0.000000 0.003665 0.000005 -0.000007
AS f2 : 0.000010 1.000431 0.000009 -0.003500 -0.000002
POP f1 : 156.420221 -0.000246 15.586838 0.019406 -0.000154
POP f2 : 1.255806 -0.154275 0.047313 1.000008 0.024285
POP f2 : -34.814720 -0.006600 -1.884850 0.042950 1.000000
Offsets converged to:
PRCL = 2.1e-15, MICH = 1.1e-17, SRCL = -3.8e-15, CARM = 2.2e-16, DARM = 0
(POP CARMs became so much smaller compared with the other matrix below, because the offsets are added al of 5 DoFsl at once here.)
*** no iteration, offsets added for each DoF separately ***
REFL f1 : 0.020611 -0.000000 0.003600 0.000005 -0.000007
AS f2 : 0.000002 1.000000 0.000009 -0.003522 -0.000002
POP f1 : 1842.776419 -0.000198 21.533358 0.019404 -0.000132
POP f2 : -32.700639 -0.153095 -0.072481 0.999995 0.024360
POP f2 : 922.393862 -0.006435 1.488298 0.042949 0.999982
Added offsets:
PRCL = 7.5e-15, MICH = 6.25e-16, SRCL = -1.4e-14, CARM = 4.5e-16, DARM = 0
* So far, I used to add all the offsets at once. This time I add CARM and get the CARM row, add PRCL and get the PRCL row... and so on.
I
Quote: |
Koji and Jamie suggested me to include the coupling between DoFs when I calculate the last matrix. So far, I just add all the pos-offsets of 5 DoFs and re-calculate the matrix again. However, once I add one DoF pos-offset, it could already change the LSC matrix therefore different pos-offset to the other four DoF, we must iterate this process until we get the equilibrium pos-offsets for 5 DoFs.
I also noticed an error in the optical configuration file. AM mod levels were smaller than that supposed to be because of the hald power going through the AM-EOMs in the MZI paths. Also I have put PM-Mods in the MZT path which gives the smaller mod indexes. So, smaller mod levels were applied both for PM and AM. As PM-AM ratio is still kept in this, so the matrices were not very wrong, I assume. I'll modify that and post the results again.
|
|
6485
|
Wed Apr 4 21:43:16 2012 |
Mike J. | Update | Computers | Better Hysteresis Model |
A better hysteresis model based on the simple harmonic oscillator equation. Useless variables have been removed and output can now be saved to workspace for plotting. The model is at "/users/mjenson/matlab/SHO_hyst.mdl". |
Attachment 1: SHO_hyst.png
|
|
6484
|
Wed Apr 4 13:25:29 2012 |
jamie | Configuration | PEM | PEM_SLOW (i.e. seismic RMS) channels aquiring |
Quote: |
I've added the PEM_SLOW.ini file to the fb master file, which should give us the slow seismic RMS channels when the framebuilder is restarted. Example channels:
[C1:PEM-RMS_ACC6_1_3]
[C1:PEM-RMS_GUR2Y_0p3_1]
[C1:PEM-RMS_STS1X_3_10]
etc.
|
The framebuilder seems to have been restarted, or restarted on it's own, so these channels are now being acquired.
Below is a minute trend of a smattering of the available RMS channels over the last five days.

|
6483
|
Tue Apr 3 22:50:37 2012 |
keiko | Update | LSC | RAM simulation for Full ifo |
Koji and Jamie suggested me to include the coupling between DoFs when I calculate the last matrix. So far, I just add all the pos-offsets of 5 DoFs and re-calculate the matrix again. However, once I add one DoF pos-offset, it could already change the LSC matrix therefore different pos-offset to the other four DoF, we must iterate this process until we get the equilibrium pos-offsets for 5 DoFs.
I also noticed an error in the optical configuration file. AM mod levels were smaller than that supposed to be because of the hald power going through the AM-EOMs in the MZI paths. Also I have put PM-Mods in the MZT path which gives the smaller mod indexes. So, smaller mod levels were applied both for PM and AM. As PM-AM ratio is still kept in this, so the matrices were not very wrong, I assume. I'll modify that and post the results again. |
6482
|
Tue Apr 3 15:50:58 2012 |
keiko | Update | LSC | RAM simulation for Full ifo |
Oops, Yesterday's results for DARM was wrong!
I got more convincing results now.
> (B) Are pos-offsets degrade the CARM and DARM so much (See, the quoted result below), is that true?
Here is the new results. It does change CARM a lot, but not DARM:
Matrix1 (normalised so that the diagonals are 1):
REFL f1 : 1.000000 0.000000 0.000008 -0.000005 0.000003
AS f2 : 0.000001 1.000000 0.000005 -0.003523 -0.000001
POP f1 : -3956.958708 -0.000183 1.000000 0.019064 0.000055
POP f2 : -32.766392 -0.154433 -0.072624 1.000000 0.024289
POP f2 : 922.415913 -0.006625 1.488912 0.042962 1.000000
(=Matrix 2)
Position offsets:
only CARM, 4.6e-16 (this number changed because I increased the resolution of the calculation)
Matrix3 (normalised by matrix 1):
REFL f1 : 0.039780 -0.000000 0.003656 0.000005 -0.000003
AS f2 : 0.000008 1.000017 0.000005 -0.003499 -0.000001
POP f1 : 159.146819 -0.000138 15.605155 0.019393 0.000055
POP f2 : 1.277223 -0.154415 0.047344 1.000008 0.024289
POP f2 : -35.422498 -0.006633 -1.886454 0.042963 1.000000
- CARM got a small position offset which degrades CARM signal 2 orders of mag (still the biggest signal in the sensor, though).
- DARM was not so bad, and probably the change of the DoF mixture is mostly not changed.
- Matrices don't change only with 1e-4 RAM. It changes with position offsets.
- I'll see how the matrix changes without position offsets but only with RAM effects, changing RAM levels.
- Again, above is C1 configuration, 1e-4 RAM level of PM level.
Quote: |
I add a flow-chart drawing what the scripts do and how the scripts calculate the LSC matrix.

(1) First, you calculate the LSC matrix WITHOUT RAM or anything, just for a reference. This is the first matrix shown in the quoted post.
(2) The script calculates the LSC matrix with RAM. Also, the PDH signals for all 5 DoF are calculated. The PDH signals have offsets due to the RAM effect. The operating position offsets are saved for the next round.
(3) The script calculates the LSC matrix again, with RAM plus the offset of the operation points. The matrix is shown in the last part of the quoted post.
Now I am going to check (A) LSC matrices (matrix 2, the second matrix of above chart) with different RAM levels (B) Are pos-offsets degrade the CARM and DARM so much (See, the quated result below), is that true?
|
|