40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 128 of 355  Not logged in ELOG logo
ID Date Author Type Categoryup Subject
  4950   Wed Jul 6 23:53:38 2011 kiwamuSummaryGeneralminutes of 40m meeting
Attachment 1: meeting.pdf
meeting.pdf meeting.pdf
  4952   Thu Jul 7 10:25:34 2011 steveHowToGeneralcleaning out refregerator

Please ask the owner unless  it is rotten. Do not put food into garbage can inside. Take them outside so you are not inviting ants !

  4981   Mon Jul 18 10:24:48 2011 steveHowToGeneralsolvent bottles for cleaning optics

Each bottle has matched seals. They are not interchangeable.

It is critical that the solvent do not reach the rubber bulb. Practice with the pipet.

In case of solvent touching the suction bulb: do not let the solvent go back into the bottle! Remove bulb, let it dry out and rinse pipet.

It is essential that the solvent bottle must be rinsed and refilled if it's content met with the rubber bulb.

Use glass syringe with SS needle in critical application: Hamilton ~0.1 ml

Attachment 1: P1080090.JPG
P1080090.JPG
  4989   Tue Jul 19 10:54:14 2011 steveUpdateGenerallarge sensor card can not be found

Please return sensor card to  laser log box so others can use it. We have only one larger fluorescent sensor card.

  5020   Fri Jul 22 17:01:41 2011 Iron ManFrogsGeneralProof that Alberto lived through his Iron Man!

Alberto_IronMan_small.jpgIronman Vineman 70.3 logo

print 208 Alberto Stochino 67/129 585/1376 36:02 1:52 830 6:41 2:38:58 21.1 296 4:58 56:33 2:13:40 - 5:40:19
 
  5026   Mon Jul 25 11:02:19 2011 kiwamuSummaryGeneralWeekly report

 Summary of the week ending July 10th.  Number of elog entries = 21

- SUS

 + The cutoff frequency of the high pass filters for the damping were set to 30Hz.
 + Turned off all the BounceRoll filters.
 + The BS oplev was checked and seemed healthy.
 

- LSC


 + All the measred data of the LSC whitening filters were fit.
 + All the zpk parameters are recorded on the wiki.

- ABSL

 

+ The setup completed.  
 + The freqeucy-lock of the ABSL laser was achieved with UGF of ~ 40kHz.
 + The temperature of the ABSL laser was adjusted to be 47.25 deg
 

- ALS

 (Fiber experiment)
 + The I-P curve of the ETMY laser was measred.
 + The current set point is 1.8 [A], which used to be 1.5 [A], corresponding to the output of power of 197 [mW] and 390 [mW] respectively.
 

  5027   Mon Jul 25 11:04:22 2011 kiwamuSummaryGeneralWeekly report

Summary of the week ending July 17th.  Number of elog entries = 20

- LSC
 * BO switching logic for the WF was installed on c1lsc
 * Channel mapping updated

- SUS
 * Oplev health check. Spectrum of each quadrant on every suspension was inspected and looked healthy.

- OAF
 * BNS interface board was attached to an AA board
 * The AA board was installed on 1X7. The Electro-optic fanout chassis on 1X7 is now sitting on a jack, this should be fixed.

- Fiber experiment (ALS)
 * Fibre from the ETMX and EMTY tables were routed to the PSL table

- Misc.
 * Alberto came over to the 40m with Wagonga
 
  5028   Mon Jul 25 11:06:38 2011 kiwamuSummaryGeneralWeekly report

Summary of the week ending July 24th.  Number of elog entries = 45


- LSC
 * Check of LSC WF switching
  -> some were switching, but the majority were not.

- SUS
 * Ran activateDQ.py for seeting some DQ channels of Oplevs
 * Turned ON all the offset buttons on the OL1, etc.
 * Rebuilt and restarted c1msc, c1sus, c1scx and c1scy as an update.
 * ETMY's shadow sensors look bad. Unknown noise below 3 Hz, which is higher than the usual floor by factor of 10.

- ABSL
 * The frequency lock was down.
 * The laser power into the RFPD had been too big, so it was reduced

- OAF
 * Seismometers were connected to the AA-board on 1X6
 * Most of the channels were acquired to the ADC, but some were not.

- Mode Cleaner
 * Gain of quadrants were checked.
 * Due to the SUS model update, the MC locking trigger hasn't worked correctly. This was fixed by changing ioo.db file

- Misc.
 * Virtual box was installed on Rossa. Altium is now available on Rossa.

  5046   Wed Jul 27 15:18:50 2011 kiwamuSummaryGeneralschedule

The vent will start from 1 st of August ! !

  5048   Wed Jul 27 15:42:41 2011 kiwamuSummaryGeneralmitues of 40m meeting : Task list

Quote:

The vent will start from 1 st of August ! 

 ++++ Task List for the vent preparation ++++

  + Preparation of beam dumps (Jamie / Steve)

  + Health check of shadow sensor and measurement of the cross-coupling  (Steve)

  + Measurement of the arm Lengths and estimation of the required precision (Kiwamu)

  + Alignment of the Y green beam (Suresh)

  + Alignment of the incident beam axis (Jenne)

  + Measurement of the MC spot positions  (Suresh)

  + Loss measurement of the arm cavities (Kiwamu / volunteers)

 

 ++++ Task List for the post-vent activity ++++

  + 3f RFPDs (Koji / Rana)

  + EOM resonant circuit (Kiwamu)  

  + Sophistication of the LSC model (Yoichi)

  + DRMI commissioning (Keiko / Anamaria)

 

  5078   Sun Jul 31 22:48:35 2011 kiwamuSummaryGeneralpreparation of the vent : status update

Status update for the vent preparation:

The punchline is : We can not open the chamber on Monday !

 

##### Task List for the vent preparation #####

  (not yet) Low power MC

  (not yet) Measurement of the arm lengths

  (not yet) Alignment of the Y green beam (#5066)

  (not yet) Measurement of the MC spot positions

  (80% done) Estimation of the tolerance of the arm length (#5076)

  (done) Preparation of beam dumps (#5047)

  (done) Health check of shadow sensors and the OSEM damping gain adjustment (#5061)

  (done) Alignment of the incident beam axis (#5073)

  (done) Loss measurement of the arm cavities (#5077)

Quote from #5048

Quote:

The vent will start from 1 st of August ! 

 

  5085   Mon Aug 1 22:30:59 2011 kiwamuSummaryGeneralWeekly report

Summary of the week ending July 31st.  Number of elog entries = 53


- SUS
 + ETMY-LR sensor looked strange. Something wrong.
 + Responses from the DC alignment bias to the shadow sensors and the oplevs were checked.
   --> ETMY shows the response with the opposite sign. Wired.
 + ETMY shadow sensors were examined in terms of the spectra.
   --> WF, AA and ADC noise looks reasonably low and not high enough to explain the low frequency noise.
 + Adjusted all the OSEM gains

- LSC
 + MICH noise budget is ongoing. WF filter needs to be greater than 21 dB to have dark noise of the PD greater than ADC noise
 + The arms became lockable

- CDS
 + modified and re-ran activateDQ.py.
 + c1iscex crashed for unknown reasons and we physically rebooted it.

- ALS (Fiber experiments)
 + PMC trans is sampled for the fiber beat-note measurement on the PSL table.
 + The beat-note signal between PSL and X end laser were obtained.
 + Some optics in the ETMY table were rearragend to have the Y green light aligned.

- ACS / ASS
 + incident beam axis has changed a lot.
 + X arm and Y arm ASS were reactivated.
  ---> The sign of some of the control gains had been wrong.
 + The incident beam axis and X/Y arm were re-aligned

- IOO/MCWFS
 + Some medm screens fixed.
 + Adjustment of the demodulation phase on each quadrant on WFS1 and WFS2 are done.
 + The sensing matrix (from optics to WFS sensors) were measured.

- OAF
 + c1pem was modified
 + plugged a seismometers to ADC through an AA board.
   --> channels are coming to the digital land
 
- Preparation for the invac work
 + 7 pieces of beam traps are available
 + Tolerance of the arm length is estimated to be +/- 2 cm.


- PSL/RefCav
 + ABSL is injected into the reference cavity. some flashing happened but no locking.
 + eddited the psl.db file to set EGUF and EGUL
 + turned RefCav heater and servo back on

  5089   Tue Aug 2 02:35:23 2011 kiwamuUpdateGeneralpreparation of the vent : status and plan

The vent will take place on Wednesday.

Plan for Tuesday :

  (Morning) Preparation of necessary items for the low power MC (Steve / Jamie)

  (Daytime) Measurement of the MC spot positions (Suresh)

  (Daytime) Arm length measurement (Jenne)

  (Nighttime) Locking of the low power MC (Kiwamu / Volunteers)

 

Plan for Wednesday :

  (Early morning) Final checks on the beam axis, all alignments and green light (Steve / Kiwamu / Volunteers )

  (Morning) Start the vent (Steve)

  (daytime-nighttime) Taking care of the Air/Nitrogen cylinders (Everybody !!)

 

Status of the vent preparation :

 

  (not yet) Low power MC

  (ongoing) Measurement of the arm lengths

  (ongoing) Measurement of the MC spot positions

  (80% done) Estimation of the tolerance of the arm length (#5076)

  (done) Alignment of the Y green beam (#5084)

  (done) Preparation of beam dumps (#5047)

  (done) Health check of shadow sensors and the OSEM damping gain adjustment (#5061)

  (done) Alignment of the incident beam axis (#5073)

  (done) Loss measurement of the arm cavities (#5077)

  5092   Tue Aug 2 11:52:44 2011 kiwamuSummaryGeneralSchedule

I have updated the 40m public calender.

Main change :

  + The vent starts from 3rd of August

  + Keiko and Anamaria (LSU) come from 13th of August

  5110   Wed Aug 3 14:40:32 2011 kiwamuSummaryGeneralminutes of 40m meeting : plan tomorrow

Tomorrow's main goal is : let the both X and Y green light come out from the chambers.

 

Plan of the in-vac work for tomorrow :

    - Removal of the access connector and the BS north door, starting from 9:00 AM. (requires 6 people)

   - If necessary, align ITMs and ETMs to get the green light nicely flashing / locked.

   - Take pictures of the BS and IOO table before installing / repositioning some optics.

   - Repositioning of the green periscope in the BS chamber to let the Y green light go through it.

   - Steer some green mirrors on the IOO table to let the Y green light come out from the chamber.

   - Steer some green mirrors on the BS table to let the X green light come out from the chamber.

   - Put some beam traps on the BS table

   - Leveling of the BS table. (Do we need to level the IOO table ? it will change the spot positions on the MC mirrors somewhat)

   - Take pictures again.

   - Extra jobs : if we still have some more times, lock MC and check the beam clearance at the Faraday. Also check some possible beam clippings for the IR beam.

   - Close the chamber with the light doors.

   - Softball game at 6:30 PM.

  5120   Fri Aug 5 02:00:09 2011 kiwamuSummaryGeneralsummary of today's invac work

[Jamie, Jenne, Suresh, Steve, Koji, Kiwamu]

We got two green beams coming out from the chambers !

 

Summary of today's invac work :

    - removed the access connector and the BS north door

    - realigned the X and Y arm to the green beams.

    - installed a HWP on the ETMY table to rotate the polarization of the green beam to P.

    - repositioned the first periscope on the BS table.

    - repositioned the second periscope on the IOO table.

    - steered some green mirrors on the IOO and OMC chamber to let the Y green beam come out to the PSL table.

    - installed a PBS in front of the first periscope to spatially overwrap two green beams.

    - adjusted the incident angle of the PBS to maximize the power of the Y green beam, which is transmitted through it.

   - steer two mirrors on the BS table to align the X green beam

   - installed two beam dumps, one is near the PBS to eliminate a ghost in the X green beam, and the other is on the back side of IPPOS/ANG pick off window.

   - closed the doors.

 

Some Notes:

 When steering the final green mirror on the OMC table, accidentally we changed the alignment of the MC incident mirror.

So the alignment of the incident beam going into MC has changed, and we haven't re-aligned it yet.

 During we were installing the PBS on the BS table, we found that the allowable incident angle for the Y beam is about ~ 55 deg, which maximizes the amount of the transmitted Y green.

Since the PBS had been considered to be 45 deg incident in our optical layout, this required several modifications in the green mirrors.

To have a clear X green beam path going into the PBS, we had to slide the PBS and periscope to the West.

The periscope is now sitting on the very edge of the BS table, and in fact ~ 20% of the bottom plate of the periscope is already sticking out.

Also since 30% of the area of the PBS's post is on a hole, which is somehow for the stack, we had to use three dog clamps instead of a folk clamps to make the contact tight.

  5122   Fri Aug 5 08:08:42 2011 kiwamuSummaryGeneralplan today

Today's main mission is : adjustment of the arm length

 

   + Open the ETMX(Y) door, starting from 9:00 AM

   + Secure the ETMX(Y) test mass by tightening the earthquake stops.

   + Move the ETMX(Y) suspension closer to the door side

   + Inspect the OSEMs and take pictures before and after touching the OSEMs.

   + Level the table

   + Adjust the OSEM positions

   + Move the ETMX(Y) suspension to have designed X(Y)arm length

   + Level the table again

   + Align the ETMX(Y) such that the green beam resonate

  5124   Fri Aug 5 14:03:12 2011 steveUpdateGeneralmalfunctioning crane at ETMY

The horizontal trolley drive stopped working  at the east end this morning. It is working intermittently. In the worst case we can take the door off with the manual -Genie- lift.

I'm working with Konecrane to solve  the wormgear drive problem.

  5127   Fri Aug 5 20:37:34 2011 jamieSummaryGeneralSummary of today's in-vacuum work

[Jamie, Suresh, Jenne, Koji, Kiwamu]

After this morning's hiccup with the east end crane, we decided to go ahead with work on ETMX.

Took pictures of the OSEM assemblies, we laid down rails to mark expected new position of the suspension base.

Removed two steering mirrors and a windmill that were on the table but where not being used at all.

Clamped the test mass and moved the suspension to the edge of the table so that we could more easily work on repositioning the OSEMs.  Then leveled the table and released the TM.

Rotated each OSEM so that the parallel LED/PD holder plates were oriented in the vertical direction.  We did this in the hopes that this orientation would minimize SD -> POS coupling.

For each OSEM, we moved it through it's full range, as read out by the C1:SUS-ETMX_{UL,UR,LL,LR,SD}PDMon channels, and attempted to adjust the positions so that the read out was in the center of the range (the measured ranges, mid values, and ultimate positions will be noted in a follow-up post).  Once we were satisfied that all the OSEMs were in good positions, we photographed them all (pictures also to follow).

Re-clamped the TM and moved it into it's final position, using the rails as reference and a ruler to measure as precisely as possible :

ETMX position change: -0.2056 m = -20.56 cm = -8.09 in (away from vertex)

Rebalanced the table.

Repositioned the mirror for the ETMX face camera.

Released TM clamps.

Rechecked OSEM centering.

Unblocked the green beam, only to find that it was displaced horizontally on the test mass about half an inch to the west (-y).  Koji determined that this was because the green beam is incident on the TM at an angle due to the TM wedge.  This presents a problem, since the green beam can no longer be used as a reference for the arm cavity.  After some discussion we decided to go with the TM position as is, and to realign the green beam to the new position and relock the green beam to the new cavity.  We should be able to use the spot position of the green beam exiting the vacuum at the PSL table as the new reference.  If the green X beam exiting at the PSL table is severely displaced, we may decide to go back in and move ETMX to tweak the cavity alignment.

At this point we decided that we were done for the day.  Before closing up, we put a piece of foil with a hole in it in front of the the TM face, to use as an alignment aperture when Kiwamu does the green alignment.

Kiwamu will work on the green alignment over the weekend.  Assuming everything works out, we'll try the same procedure on ETMY on Monday.

  5130   Sat Aug 6 03:10:05 2011 SureshSummaryGeneralSummary of today's in-vacuum work

The table below gives the OSEM positions as seen on the slow chanels C1:SUS-ETMX_{UL,UR,LL,LR,SD}PDMon

ETMX_OSEMs.png

Note that the side OSEM has the fast channel (OUTPUT) available and we used that to locate it.

When we began work the OSEMs were photographed so that we have a record of their locations till now.  It was difficult to get accurate estimate of the magnet offset inside the OSEM we could not see the screen on the camera while clicking.  We then took some pictures after finishing the work. These are given below

 

          Before_osem_adj.JPG         After_osem_adjustment.JPG

 

The picture of the left is from before OSEMs were moved. It can be seen that OSEMs are rotated to make sure that the magnets avoid touching the teflon sheets which hold the shadow sensors.  The picture on the right shows the positions of the OSEMs after we adjusted their positions.  This time we kept the teflon sheets vertical as shown to minimise the coupling between the Side and Axial directions.

We needed to reposition them once again after we moved the tower to the center of the table.

Pictures with more detail will be posted to the wiki later.

 

 

 

 

  5131   Sat Aug 6 13:38:02 2011 ranaSummaryGeneralSummary of today's in-vacuum work

This OSEM placement is just the OPPOSITE of what the proper placement is.

Usually, we want to put them in so that the LED beam is vertical. This makes the OSEM immune to the optic's vertical mode.

The orientation with the horizontal LED beam makes the immunity to the side mode better, but may spoil the vertical.

In reality, neither of these assumptions is quite right. The LED beam doesn't come out straight. That's why Osamu and I found that we have to put in some custom orientations.

Also, the magnet gluings relative to the OSEM bracket centers are not perfectly aligned. So...I am saying that the OSEMs have to be oriented empirically to reduce the couplings which we want to reduce.

 

  5132   Sun Aug 7 11:10:54 2011 SureshSummaryGeneralOSEM positioning

Quote:

This OSEM placement is just the OPPOSITE of what the proper placement is.

Usually, we want to put them in so that the LED beam is vertical. This makes the OSEM immune to the optic's vertical mode.

The orientation with the horizontal LED beam makes the immunity to the side mode better, but may spoil the vertical.

In reality, neither of these assumptions is quite right. The LED beam doesn't come out straight. That's why Osamu and I found that we have to put in some custom orientations.

Also, the magnet gluings relative to the OSEM bracket centers are not perfectly aligned. So...I am saying that the OSEMs have to be oriented empirically to reduce the couplings which we want to reduce.

 

 Kiwamu suggested that since the side resonance is at a lower frequency than the bounce (~17Hz)  we ought to worry about the side more than the bounce.  If this is okay we can reposition the OSEMs to minimise this coupling. 

More over, in the current position, the OSEM s will not sense the side motion!!  So we definitely need to reposition them.  Sorry! I was being a spatz. 

  5151   Tue Aug 9 03:05:05 2011 kiwamuSummaryGeneralweekly report

Summary of the week ending Aug 8th.  Number of elog entries = 56


- VAC
 + The vent started Wednesday morning
 + Repositioning of the green periscopes and associated mirrors are done.
 + Got both of the green beams coming out from  the chambers
 + Moved the ETMX suspension tower by -8.09 inch (away from vertex)
 + Fixed the alignment of the ETMX CCD mirrors
 + Recovered the X green beam axis for the latest ETMX position

- SUS
 + oplev centered prior to the vent

- LSC
 + ETMY_TRANS_QPD didn't respond at all, needs to be fixed
 + Old MZ PD (InGaAs 2mm, @29.5MHz) has been modified for REFL33.
  The 11MHz notch circuit is at the amp side instead of the diode side. This is ready for the installation
 + REFL165 PD has been made from the old 166MHz PD.

- IOO
 + IPPOS has been sick since 19th of July, 2011
 + IPANG is clipped on a pick-up mirror on the ETMY table. QPD itself is healthy.
 + The spot positions on the MC mirrors were measured prior to the vent.
   The results are almost the same as before within a few percent difference expect for the MC2 yaw.
 + An attenuator, consisting of two HWPs and a PBS, has been installed on the PSL table for the MC low power state.
 + a 10% BS in front of the MCREFL_RFPD was replaced by a perfect reflector for the low power mode.
 + The incident power for MC was decreased to 20 mW
 + The beam axis going to MC was misalgned due to the attenutor.
    Then the beam was aligned by touching two steering mirrors on the PSL table
 + MC is able to be locked in air. The reflection DC goes from 1.4 to 0.13 V when the MC is locked.

- ABSL
 + With the mass-kicking technique, the arm lengths were measured.
   Xarm =  37.5918 m, Yarm = 37.5425 m.

- Green locking
 + Y green beam is aligned to the Y arm
 + Locking of the Y green is not robust, it needs to be revisited

- OAF
 + Wiener Filtering was applied on the data collected from the X-arm for a duration of 1500 seconds.

- Misc.
 + The hazardous waste people are moving chemicals around outside our door, and have roped off our regular front door.
 + The horizontal trolley drive of the east end crane stopped working. It will be fixed.

  5152   Tue Aug 9 10:15:25 2011 kiwamuUpdateGeneralin-vac work plan today

We will move on to the vertex region today.

The goal of the vertex region work is to get the pick-off beams out the chambers, including POX/Y and POP.

The work will be in parallel to the ETM woks.

The first step will be : lock and align MC with the IR beam.

  5155   Tue Aug 9 15:14:46 2011 KojiUpdateGeneralin-vac work plan today

This morning Kiwamu and I have aligned the MC. Kiwamu aligned the last steering (on the OMC table) to recover the touch last week.
Then I have aligned the MC with MC1 and MC3 as the last steering did not help to get TEM00.

Before

C1:SUS-MC1_PIT_COMM = 2.6587
C1:SUS-MC1_YAW_COMM = 2.7471
C1:SUS-MC2_PIT_COMM = 3.486
C1:SUS-MC2_YAW_COMM = -1.1592
C1:SUS-MC3_PIT_COMM = -1.876
C1:SUS-MC3_YAW_COMM = 1.2829

After

C1:SUS-MC1_PIT_COMM = 2.7596
C1:SUS-MC1_YAW_COMM = 2.6627

C1:SUS-MC2_PIT_COMM = 3.486
C1:SUS-MC2_YAW_COMM = -1.1592
C1:SUS-MC3_PIT_COMM = -1.697
C1:SUS-MC3_YAW_COMM = 1.2901

Quote:

We will move on to the vertex region today.

The goal of the vertex region work is to get the pick-off beams out the chambers, including POX/Y and POP.

The work will be in parallel to the ETM woks.

The first step will be : lock and align MC with the IR beam.

 

  5158   Tue Aug 9 16:40:12 2011 steveUpdateGeneralclean room coat counts

I measured the particles coming off of new- unused clean room coat. Tyvek, Convertors, Allegiance #9393 measured  10 counts of 0.3 micron at 1 minute and 0 count at 7 minute.

The 0.5 micron size  measured 0 at both times. Jenne's used coat measured 20 counts of 0.3 micron  and 0 counts for 0.5 micron at 1 minute. ( counts / cf- min)

The HEPA tent background is consistently 0 when it's CP STAT 100 curtains are closed.

Attachment 1: P1080155.JPG
P1080155.JPG
  5193   Thu Aug 11 15:31:43 2011 steveUpdateGeneralmalfunctioning crane at ETMY

Quote:

The horizontal trolley drive stopped working  at the east end this morning. It is working intermittently. In the worst case we can take the door off with the manual -Genie- lift.

I'm working with Konecrane to solve  the wormgear drive problem.

 Gear box found, it's  lead time one week at $825    The crane may be functional round August 25

  5212   Fri Aug 12 16:52:49 2011 steveUpdateGeneralChamber illuminator switch

I'm looking for an ether net based  power switch for turning lights on and off for the vacuum system from MEDM screen.  This is what I found

Jamie please take a look at it.

 

 

  5242   Mon Aug 15 17:38:07 2011 jamieUpdateGeneralFoil aperture placed in front of ETMY

We have placed a foil aperture in front of ETMY, to aid in aligning the Y-arm, and then the PRC.  It obviously needs to be removed before we close up.

  5245   Tue Aug 16 04:28:09 2011 kiwamuUpdateGeneraltoday's work to do

The main goal of today is to extract the pick-off beams

Today's menu :

  + If necessary steer ETMs and ITMs to make the X and Y green beam flashing.

  + Open the IOO and OMC chamber and lock MC.

       => cover the place of the access connector by a large piece of aluminum foil. It will give a robust lock of MC.

  + Check the beam pointing down to Y arm by looking at the ETMY face camera.

        => If it's necessary align PZT1 and PZT2 from EPICS to make the IR beam flashing in the X arm.

  + Align BS and let the beam hit the center of ETMX to make the X arm flashing.

        => These alignment procedure will automatically gives us the MICH fringes on the AS CCD camera.

  + Rotate the SRM tower to get the SRMI fringes on the AS CCD camera.

        => This is because the required amount of the YAW correction on SRM is currently beyond the range of the DC bias.

  + Align PRM to get the PRMI fringes on AS CCD camera. Also make sure the beam comes back to the REFL CCD camera.

  + Lock the PRC to make POP/POX/POY bright enough.

     => Probably the REFL11 RFPD needs more power. To increase the power, just rotate the HWP, which is sitting before the RFPDs on the AS table.

     => If the signal on AS55 is too small, we can use REFL11_Q or REFL55 for the MICH lock.

  + OR inject and align the ABSL laser to make those pick-off beams bright enough.

     => This case we don't have to make the DRMI exactly on the resonance, what we need is just the DRMI flashing.

  + Align necessary optics for those pick-off beams.

     => In our definition (aLIGO definition) POP is the beam propagating from BS to PRM. Don't be confused by another one, which propagates from PRM to BS.

  + Steer two mirrors on the ETMY table for aligning IPANG. Also, steer some mirrors on the BS table for IPPOS.

    => IPANG has already reached the ETMY table, so ideally we don't have to steer a mirror on the BS table.

    => IPPOS/ANG are very visible with a sensor card.

  + Align some oplevs.

     => This work is relatively independent from the other tasks. Steve can take care of it.

  + Adjust the OSEM to their mid-range.

     => This work can be done anytime, but everytime we have to recover the alignment of the interferometer.

  5248   Tue Aug 16 11:49:17 2011 jamie, jenneUpdateGeneraltoday's work to do

>If necessary steer ETMs and ITMs to make the X and Y green beam flashing.

Green is now flashing in both X and Y arms

>Open the IOO and OMC chamber and lock MC.

Open, and cover in place. MC is flashing and locking.

 

  5250   Tue Aug 16 17:09:55 2011 JenneUpdateGeneraltoday's work to do

Quote:

  + Rotate the SRM tower to get the SRMI fringes on the AS CCD camera.

        => This is because the required amount of the YAW correction on SRM is currently beyond the range of the DC bias.

 Kiwamu aligned things for me, and I rotated the SRM tower so that the reflected beam was pretty much totally overlapping the incident beam.  The SRC was aligned to make sure things were good.  Now the DC bias for SRM Yaw is ~1.4, so we're totally good. 

To rotate SRM, Jamie had the idea of using 2 screws so I could push the tower on one side, and back off the screw an equal amount on the other side and push the tower to be touching both screws again, to ensure that I was rotating about the center of the tower and wasn't introducing any Pos action. 

While I was at it, I also moved the OSEM connector tower back to its normal place on the table, so it's not in the way of oplev beams.  It had been moved previously to accommodate ITMY near the door.

  5252   Wed Aug 17 03:10:06 2011 kiwamuUpdateGeneralcurrent status of in-vac work

[Jamie / Suresh / Kiwamu] 

The in-vac work is ongoing.

Before we run out our energy we are posting this entry to briefly report the current status.

- (done) BS earthquake stop adjustment.

- (done) PRM earthquake stop adjustment

- (done) MC spot position check => They are almost the same within 10 %.

- (done) Injection and alignment of the ABSL laser to make the beam brighter in the vertex region.

- (done) POY => We repositioned an in-vac steering mirror to get the POY beam hitting the center of the steering mirror.

                           It's now coming out from the chamber.

(done)  IPANG => realigned two mirrors on the ETMY table to get the IPANG out from the chamber. Now it's reaching the ETMY optical table.

                             It needs a final touch before we pump down.  We revisited it later in the night after realigning the IFO and it is well aligned now.

- (done)  POP => We have aligned the ABSL laser injected from the AS port to reach the REFL camera.  We turned it up to max power of 300mW and used it as a substitute for the PRC beam.

                         Even this was not enough to see anything in the POP beam path after the PR2 (tip-tilt).  So we used a green beam from the Y-arm as a guide of the POP beam path because the ABSL (POP) beam was too dim to work with.

                         We placed a lens and a CCD camera to detect the green and then blocked the Y-green.  It was then possible to see the ABSL-POP beam in the CCD camera.  The lens and the CCD are markers for this beam. 

                         Do not remove these markers unless absolutely necessary.

-(done) POX => We located the ABSL (mimicking POX) beam on the POXM1 mirror and adjusted the mirror to ensure that the beam exits at the right height and a convenient location on the POX table. 

- (0%) OSEM mid-range adjustment

- (0%) IPPOS

- (0%) oplev re-alignment

  5253   Wed Aug 17 06:42:38 2011 kiwamuUpdateGeneralin-vac work : the end is near

We will pump down the chambers on Thursday morning.

Today will be a day of the OSEM and oplev party.

 

 -- to do list for today --

 + OSEM mid-range adjustment

 + oplev realignment

 + placement of beam traps

 + extraction of IPPOS

 + table leveling

 + interferometer alignment

 + AM-PM mystery

 + preparation for drag and wipe

  5256   Wed Aug 17 15:55:01 2011 JenneUpdateGeneralin-vac work : the end is near

Quote:

We will pump down the chambers on Thursday Friday morning.

 All hands on deck at 9am Thursday for drag wiping and doors.  We'll do the 5 doors first (including drag wiping), then put on the access connector last.  Steve will then begin pumping early Friday morning.

  5258   Wed Aug 17 20:14:49 2011 jamie, kiwamu, suresh, jenne, keiko, anamariaUpdateGeneralin-vacuum work status, prep for pump

This afternoon's work:

  • OSEMs were adjusted on all suspended optics.
  • X and Y-arms were aligned to green.
  • Once that was done, the input pointing was adjusted with the PZTs to get the IR beam centered on ITMY and ETMY.
  • Once the input pointing was ok we extracted IPANG and IPPOS.
  • BS, ITMY, PRM, SRM optical levers (oplevs) were extracted.
  • Prepared rails and stops for TMs for morning drag wiping.

TODO before drag wiping:

  • Check full IFO alignment.
  • Readjust OSEMs if needed.
  • Extract ITMX oplev.
  5259   Thu Aug 18 00:53:48 2011 jamie, kiwamu, suresh, jenneUpdateGeneralPUMP PLAN ABORTED; need to work more on IFO alignment

We have decided that the IFO alignment is bad enough that we're not ready to pump down.  PUMP ABORTED.

The IFO alignment is somewhat OK, in that the green and IR beams are flashing in the arms, and the return beams are overlapping at the BS.  However the beams appear to be not centered on any of the optics at the moment.  They are all displaced in yaw by ~0.5 to 1 cm or so in various directions.

From this we have decided that we need to step back and reattack the IFO alignment from square one.  Here is our current suggested procedure:

  1. check ETM positions relative to what we think they should be on the drawings.  This is to verify that the ETMs were not placed in the wrong places laterally.
  2. translate Y green axis north, centering green on ETMY and ITMY (by looking at cards).  North is the opposite direction from how the beams are displaced from the TM centers.
  3. steer input pointing to overlap IR on green beam at BS, ITMY, and ETMY.  IR should visibly overlap green at both BS and ITMY, and we should be able to see IR on target in front of ETMY with ETMY face camera, and in ETMY trans camera.
  4. center IR on ETMX by steering BS with DC bias.
  5. align Y arm cavity for green resonance by adjusting ITMY.
  6. adjust ITMX to achieve michelson fringes at AS
  7. adjust PRM lateral translation to center beam on PRM, if needed
  8. adjust SRM lateral translation to center beam on SRM, if needed
  9. align PRC to see fringes
  10. align SRC to see fringes
  11. extract AS (no clipping)

 Once this is done, we will need to check the following:

  • IPANG/IPPOS extraction
  • pick-off extraction
  • OPLEVs
  • OSEMs
  • green periscopes and green beam extraction at PSL

We've decided to stop for the night, get a good nights rest, and attack all of this tomorrow morning.

Beam_spot_shifts.png

  5267   Fri Aug 19 01:46:06 2011 SureshUpdateGeneralIFO alignment

[Keiko, Jamie, Kiwamu, Anamaria,

We followed the procedure that we laid-out in our elog of yesterday.  We completed the first six steps and we now have the y-arm well aligned to the green beam which passes through the center of of both ETMY and ITMY. 

The IR beam was steered with the PZTs to coincide with the green beam.  The BS was adjusted to see IR beam scatter on a target placed near the center of the ETMX.  And then the AS IR beam was steered to the AS camera by adjusting several components along OM path ( we touched OM1, OM2, OM3, OM4, OM5, OMPO and OM6).  We then looked for IR fringes in the AS port from the Y-arm. But no luck there.  We need to realign the IR beam into the Y-arm cavity axis using the pzts.

We aligned ITMX and PRM to get power recycled Michelson fringes at the AS.

 

  5270   Fri Aug 19 15:31:53 2011 steveUpdateGeneralpower interruption rescheduled to 10-1-2011

                UTILITY & SERVICE INTERRUPTION

**PLEASE POST**

 

Building:               Central Engineering Services (C.E.S.)

          LIGO Gravitational Physics building adjacent to C.E.S. 40M- Lab

          Safety Storage adjacent to CES

          Steele House 

          Keck Lab

 

Date:                   Saturday, October 1, 2011

Time:                   8:00 a.m. To 9:00 a.m.            

Interruption:   Electricity

Contact:                Mike Anchondo ext. 4999  Tom Brennan 4984

*This interruption is required for maintenance of high voltage switchgear in Campus Sub Station.

(If there is a problem with this Interruption, please notify

 the Service Center X-4717 or the above Contact as soon as possible.

 If no response is received we will proceed with the interruption.)

         

                                Reza Ohadi,

                                Director, Campus Operations & Maintenance

  5275   Mon Aug 22 02:43:28 2011 kiwamuUpdateGeneralin-vac work

[Kieko / Kiwamu]

 The AS beam shows a little bit of clipping and right now this is the only concern for the alignment of the interferometer.

Other than that everything is okay including :

  + Beam centering on all the suspended optics

  + Arm cavity alignments. The Fabry-Perot fringes from both arm cavity were found on the AS camera.

  + POX/POY/POP, they are still successfully coming out from the chambers

  + IPPOS/ANG

  + Alignment of the green beams and the associated optics. Both green are reaching to the PSL table

 

We need to check/fix the AS beam clipping and once it's done we will readjust the OSEM mid range and the oplevs.

Then it will be ready for the drag/wipe and door closing.

  5277   Mon Aug 22 16:28:44 2011 SureshUpdateGeneralSRM tower shifted on Friday

[Kiwamu, Suresh]   This is a belated elog entry from last Friday night + Saturday morning! 

    We shifted the SRM tower across the beam and away from the door by about 5mm. 

                  After the input beam from MC was aligned to the Y-arm,  Kiwamu noticed that the AS beam was being clipped and that the correction had to start from SRM onwards as the beam had become offcentered on the SRM.  So we shifted the SRM tower by about 5mm away from the door and transverse to the beam and rotated it by a few degrees to center the OSEM offsets.  After this we aligned all optics along the AS beam path to extract the AS beam from the chamber.    We then aligned each optic in the vertex so that their beams overlap at the AS port with the reflection from ITMY.  Then we aligned BS to center the beam on ETMX and then looked for flashes from the Y arm.

      At this point Kiwamu checked and found that the input beam from the MC had shifted.  It was landing on the ITMY about 5mm below the center.  And (inexplicably) it was still centered on ETMY!  The Y- green which traced the cavity axis (since this was still flashing) was not coincident with the IR beam.   So all the work we did in aligning the AS beam and the vertex optics work was lost..... and had to be done again.

 

 

  5279   Mon Aug 22 21:32:10 2011 kiwamuUpdateGeneralmore in-vac work : AS clipping fixed and OSEM/oplev adjustment

[Keiko / Jenne / Jamie / Kiwamu]

 We did the following things today :

  + fixed the AS clipping issue

  + realigned all the oplevs

  + checked and adjusted the all OSEM DC values, including PRM, SRM, BS, ITMs, ETMs, MC1 and MC3

 

Since we touched the OSEMs the alignment has changed somewhat.

Right now Jenne, Suresh and I are working on the "confirmation alignment".

Once we find the alignment is still good (steerable by the PZTs and the DC coil bias), tomorrow we will do the drag&wipe and door closing.

Quote from #5275

We need to check/fix the AS beam clipping and once it's done we will readjust the OSEM mid range and the oplevs.

 

  5280   Tue Aug 23 00:55:13 2011 JenneUpdateGeneralIFO ready for doors

[Kiwamu, Jenne]

After the IFO was aligned in air one final time, we tapped on a few OSEMs until we were happy with all of the centering of all of the optics' OSEMs.  All are within 0.05 of their halfway values, with the exception of one each on MC1 and MC3, one of which is within 0.06, and the other 0.08.  Because of the realignment pain of dealing with MC OSEMs, we elected to leave these alone.  Also, since we obviously didn't open the MC2 tank, we don't know how they are, although the numbers look reasonable. 

Also, we took photos (to be posted on Picasa in a day or two) of all the main IFO magnet-in-OSEM centering, as best we could.  SRM, BS, PRM all caused trouble, due to their tight optical layouts.  We got what we could.  Various people have been looking at these for the past 2 weeks, and I think they're all fine, even if we didn't get stellar photos.

We are now prepared for pumping.  For real this time.

 

  5284   Tue Aug 23 06:49:24 2011 AnamariaUpdateGeneralmore in-vac work : AS clipping fixed and OSEM/oplev adjustment

Where was the AS clipping?! Ah, the suspense...

Quote:

  + fixed the AS clipping issue

 

Quote from #5275

We need to check/fix the AS beam clipping and once it's done we will readjust the OSEM mid range and the oplevs.

 

 

 

  5285   Tue Aug 23 09:40:37 2011 kiwamuUpdateGeneralRe: AS clipping fixed

Indeed it was suspenseful.

We tried finding where the clipping happened, but we couldn't find any obvious clippings.

So we checked centering of the beams on all the optics associated with the AS path, starting from BS, SR3,... to the AS optical bench.

And during the work some of them were recentered.

At the end we found no clipping. To make sure we tested the available range (no clipping range) by exciting the angular motion of BS with AWG (f ~ 1Hz, a ~ 1000).

The beam looked successfully coming out at the most of the angular oscillation point.

Quote from #5284

Where was the AS clipping?! Ah, the suspense...

 

  5307   Fri Aug 26 10:45:17 2011 steveUpdateGeneral crane at ETMY is fixed

Quote:

The horizontal trolley drive stopped working  at the east end this morning. It is working intermittently. In the worst case we can take the door off with the manual -Genie- lift.

I'm working with Konecrane to solve  the wormgear drive problem.

 New gear box installed and tested by Fred KoneCranes.

Attachment 1: P1080190.JPG
P1080190.JPG
Attachment 2: P1080188.JPG
P1080188.JPG
  5309   Fri Aug 26 15:59:33 2011 steveUpdateGeneraldrawer cabinet moves in

manufacturer info

Jamie and Shuresh moved in Jenne's 11 drawers cabinet and relocated old note book boxes on the inside of the vac tube.

Attachment 1: P1080192.JPG
P1080192.JPG
  5310   Fri Aug 26 16:17:31 2011 JenneUpdateGeneraldrawer cabinet moves in

Quote:

Jamie and Shuresh moved in Jenne's 11 drawers cabinet and relocated old note book boxes on the inside of the vac tube.

 Barring other chores for next Wednesday, we're going to spend Wednesday afternoon populating the new cabinet with all of the optics hardware: posts, forks, dogs, everything!  It's going to be so organized and awesome!!

  5330   Wed Aug 31 16:18:25 2011 JenneUpdateGeneralNew Optics Drawers

[Kiwamu, Manuel, Jenne]

The new optics storage drawers have been populated with optics.  Each drawer is labelled.  Harsh punishments will be inflicted on anyone found disobeying the new scheme.

  5343   Tue Sep 6 11:27:19 2011 JenneUpdateGeneralAfternoon pre-pumpdown todo list

These are the things that I can think of that we need to do before we can close up:

* Take a close look at both ITMs' OSEMs, and ensure that the magnets aren't too close to either plate in the OSEMs.  Both have had funny business over the past week.

* Do a free swinging test on both ITMs.  (ITMY may not need it, if we haven't touched it since the last free swinging test, but it can't hurt to take the data)

* Confirm that POX is exiting the chamber.

* Is there anything else???

 

Our goal is to finish this work by tonight, so that we can close doors and start pumping tomorrow.

ELOG V3.1.3-